DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
1 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
APPROVED BY
SIGN / DATE
DESIGNATION
____________________
____________________
____________________
____________________
QA MANAGER
ORIGINATED BY / DATE
____________________
QA/ VALIDATION SPECIALIST (SEERPHARMA)
REVIEWED BY / DATE
____________________
JOHN ROGERS
SENIOR CONSULTANT (SEERPHARMA)
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
2 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
TABLE OF CONTENTS
1
PURPOSE ....................................................................................................................................3
2
SCOPE .........................................................................................................................................3
3
ASSOCIATED DOCUMENT & REFERENCE .........................................................................3
4
DEFINITION ...............................................................................................................................3
5
RESPONSIBILITY .....................................................................................................................4
6
EQUIPMENT/ PROCESS OVERVIEW ....................................................................................4
7
OVERVIEW OF IQ TESTS AND CHECKS .............................................................................6
8
PREREQUISITE .........................................................................................................................7
9
TESTS AND CHECKS ...............................................................................................................8
10
TEST INSTRUMENTATION...............................................................................................31
11
IDENTIFICATION OF PERSONNEL .................................................................................32
12
LIST OF DEVIATIONS........................................................................................................33
13
REVIEW AND APPROVALS ..............................................................................................34
14
LIST OF APPENDICES........................................................................................................35
15
PROTOCOL REVISION HISTORY ....................................................................................36
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
3 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
1
PURPOSE
This purpose of this Installation Qualification (IQ) is:




Establish that the critical components of Tuttnauer Sterilizer are installed correctly and
in accordance with design documentation requirements.
Ensure that all supporting documentations are in place and of suitable quality.
Verify that all critical instrumentations are calibrated.
Ensure that Autoclave 69180\1HSD-EP has been entered onto the Xxxx Planned
Preventative Maintenance database.
This Installation Qualification (IQP/QA/027/2005, revision 01) document supersedes the
original IQ document (Installation Qualification Protocol for Tuttnauer Sterilizer 69180\HSDEP, revision 00).
This updated protocol has been developed by Xxxx International to ensure that all the latest
regulatory requirements are met. The original IQ documentation (Protocol and Report) is still
valid and all information contained within auditable. This updated Protocol includes IQ
function testing that was omitted from the original documentation, the extra tests include:
Computer Validation, Instrument Tagging and Drawing Verification and Control.
2
SCOPE
The scope of this IQ is applicable to Tuttnauer Sterilizer, 69180\1HSD-EP located on level 4 of
Xxxx International, Singapore. Associated utilities are considered from the point of connection
with the equipment. It also includes all ancillary equipments/ components.
3
4
ASSOCIATED DOCUMENT & REFERENCE
 Validation Master Plan, VMP/GEN/001-2004, Rev. No. 02.
 Process Validation for Installation Qualification, SG-PRO-004, Rev. No. 01.
 Installation Qualification Protocol for Tuttnauer Sterilizer 69180\HSD-EP, revision 00.
DEFINITION
The following table contains the explanations of the abbreviations and terminology used within
this IQ.
Abbreviation
IQ
SOP
LCD
Term
Installation Qualification
Standard Operation Procedure
Liquid Crystal Display
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
4 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
5
RESPONSIBILITY
The following table defines the responsibilities of personnel involved in compilation, execution
and approval of this IQ.
Personnel
Responsibilities

Preparation of protocol against design documentation.

Submittal of completed protocol for approval prior to execution.

Execution of tests detailed in the approved protocol with Good
Documentation Practices.

Compilation of test results or raw data.

Initiate and submit Deviation report for review.

Preparation of executed Protocol Summary Report.
Reviewer

Review completed protocol to ensure correctness in execution and
completeness in compilation.
Approver

Ensure that protocol has met its defined purpose.

Approval of protocol that met all technical, compliance and quality
requirements.
Author
Executor
6
EQUIPMENT/ PROCESS OVERVIEW
The Tuttnauer, Model 69180\1HSD-EP is a Pre and Post Vacuum autoclave with 1015 Litres
of chamber volume. It is constructed from stainless steel using a double wall type structure.
Saturated steam is used as the sterilizing media; it is supplied by an external steam generator.
The autoclave has a max operating temperature of 131oC and maximum pressure of 2.3 bars.
It is equipped with one horizontal sliding door operated by a hydraulic system. The autoclave
chamber is sealed by a heat resistance silicone rubber gasket mounted on the door frame
groove. Sealing occurs when the gasket is pushed against the door by steam pressure and
thrust into the groove behind it.
The control of the autoclave is based on a microprocessor system, this controls and monitors
all of the physical parameters of the process and performs the operation sequence of the
machine according to the selected predefined programs. These programs are accessible via an
Operator Interface Terminal (OIT); it also allows users to “START” and “STOP” the autoclave
cycles. The OIT displays alarms and informative messages on its 4 line LCD screen. After
every executed cycle a “once-through” print out is generated via the attached printer.
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
5 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
Cycle control, monitoring and alarm functionality are assessed from the following instruments:






Condensate Temperature sensor
Drain Temperature sensor
Chamber Temperature sensors
Chamber Pressure Transducer
Jacket Pressure Transducer
Gasket Pressure Transducer
Three pressure gauges are installed on the autoclave, and are used for visual monitoring of the
Chamber, Jacket and Gasket pressure.
Typical sequences of the 12 pre programmer cycles are:
Air Removal
Steam
Injection
Steam
Holding
Steam
Removal
Pressure
Equilibration
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
6 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
7
OVERVIEW OF IQ TESTS AND CHECKS
The following table defines all the testing within this IQ with its objective.
Section
Reference
Test Description
Objective
9.1
Documentation Verification
To verify that all relevant lifecycle
documentations of the equipment are available
and current.
9.2
Drawings Verification
To verify that all relevant drawings of the
equipment are available and current.
9.3
Utilities Requirements Verification
To verify that all supporting utilities of the
equipment are available as specified.
9.4
Equipment and Critical
Components Verification
To verify that installed equipment and its critical
components are as specified.
9.5
Control System Verification
To verify that installed control system of the
equipment is as specified.
9.6
Product Contact Parts Material
Review
To verify that product contact parts materials are
acceptable for use.
9.7
Instrumentation Identification And
Calibration Verification
To verify that critical instrumentation of the
equipment are as specified. Ensure that all
calibrations for the instrumentations are current.
9.8
Control System Configuration And
Set point Verification
To verify that control system configurations and
system setpoints are as specified.
9.9
Safety Features Verification
To verify that system safety features are
available as specified
9.10
Planned Preventative Maintenance
System
To verify that the autoclave has been entered
onto the Planned Preventative Maintenance
System
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
7 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
8
PREREQUISITE
The following pre-requisite shall be completed prior to the protocol execution commencement.
Item
Test Description
Acceptable
(Yes/ No)
1
Site Acceptance Test/ Vendor Commissioning
Documentation
2
Completed Equipment Categorization Form
3
Completed Checklist for Installation Qualification
Requirements
Verified by
(Initial & Date)
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
8 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9
TESTS AND CHECKS
9.1
9.1.1
DOCUMENTATION VERIFICATION
OBJECTIVE
To verify that all relevant lifecycle documentations of the equipment are available and
current.
9.1.2
METHOD



9.1.3
Record the Title, Revision No. and Location in the following results table as defined.
Ensure that all listed documentations are current.
Record and list all deviation/non-conformances observed in Section 12.
TEST TOOL
Not applicable.
9.1.4
ACCEPTANCE CRITERIA
All relevant lifecycle documentations of the equipment are available and current.
9.1.5
RESULTS
Document
Type
Purchase Order
Document Reference
(Title, Doc. No., Revision No.)
Location
Acceptable
(Yes/ No)
Verified by
(Initial & Date)
Title:
Doc. No.:
Rev. No.:
User
Requirement
Specification
Title:
Design
Specification
Title:
Doc. No.:
Rev. No.:
Doc. No.:
Rev. No.:
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
9 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.1.5
RESULTS (CONTINUED)
Document
Type
Document Reference
(Title, Doc. No., Revision No.)
Operating and
Maintenance
Manual
Title:
Control System
Operating
Manual
Title:
Thermal Printer
Operating
Manual
Title:
Welding
Documentation
Title:
Location
Acceptable
(Yes/ No)
Verified by
(Initial & Date)
Doc. No.:
Rev. No.:
Doc. No.:
Rev. No.:
Doc. No.:
Rev. No.:
Doc. No.:
Rev. No.:
Vacuum
Testing
Documentation
Title:
Spare Parts List
Title:
Doc. No.:
Rev. No.:
Doc. No.:
Rev. No.:
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
10 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.1.5
RESULTS (CONTINUED)
Document
Type
Document Reference
(Title, Doc. No., Revision No.)
Vendor
Installation
Qualification
Protocol
Title:
Vendor
Operational
Performance
Qualification
Protocol
Title:
SOP Operation
Title:
Location
Acceptable
(Yes/ No)
Verified by
(Initial & Date)
Doc. No.:
Rev. No.:
Doc. No.:
Rev. No.:
Doc. No.:
Rev. No.:
SOP Maintenance
Title:
Doc. No.:
Rev. No.:
SOP Calibration
Title:
Doc. No.:
Rev. No.:
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
11 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.2
DRAWINGS VERFICATION
9.2.1
OBJECTIVE
To verify that all relevant drawings of the equipment are available and current.
9.2.2
METHOD





9.2.3
Record the Drawing No., Revision No. and Location in the following results table as
defined.
Ensure that drawings are accurate against installation by marking up with YELLOW
highlighter if content is correct and RED highlighter if content is incorrect.
Correct inaccurate drawings with RED pen with appropriate annotation.
Append all markup drawings and record in Section 13.
Record and list all deviation/ non-conformance observed in Section 12.
TEST TOOL
Not applicable.
9.2.4
ACCEPTANCE CRITERIA
All relevant drawings of the equipment are available and current.
9.2.5
RESULTS
Drawing Title
Drawing Reference
(Dwg. No., Rev No.)
General Arrangement
Dwg. No:
Rev. No.:
Piping Diagram
Dwg. No:
Rev. No.:
Electric Control
Diagram
Dwg. No:
Rev. No.:
Location
Acceptable
(Yes/ No)
Verified by
(Initial & Date)
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
12 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.2.5
RESULTS (CONTINUED)
Drawing Title
Drawing Reference
(Dwg. No., Rev No.)
Location
Acceptable
(Yes/ No)
Verified by
(Initial & Date)
Dwg. No:
Rev. No.:
Dwg. No:
Rev. No.:
Dwg. No:
Rev. No.:
Dwg. No:
Rev. No.:
Dwg. No:
Rev. No.:
Dwg. No:
Rev. No.:
Dwg. No:
Rev. No.:
Dwg. No:
Rev. No.:
Dwg. No:
Rev. No.:
Dwg. No:
Rev. No.:
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
13 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.3
9.3.1
UTILITY REQUIREMENTS VERIFICATION
OBJECTIVE
To verify that all supporting utilities of the equipment are available as specified.
9.3.2
METHOD


9.3.3
Verify supporting utilities parameters using appropriate test tool. Record details of
verification in the following results table.
Record and list all deviation/ non-conformance observed in Section 12.
TEST TOOL
Calibrated Multimeter.
9.3.4
ACCEPTANCE CRITERIA
All supporting utilities are available as specified.
9.3.5
RESULTS
Utilities
As Specified
As Found
Electricity
Voltage: 380V
Phase: 3 Phase
Frequency: 50 Hz
Voltage:
Phase:
Frequency:
Steam
Maximum Supply: 540kpa
Maximum Supply:
Compressed Pressure: 6-8 Bar
Air
Flow rate: 100 L/m
Cold Water
Pressure: 4-6 Bar
Flow rate
Peak: 28 L/m
Average: 1020 L/Hr
Temperature: 15 oC
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
Pressure:
Flow rate:
Pressure:
Flow rate
Peak:
Average:
Temperature:
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
14 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.4
9.4.1
EQUIPMENT AND CRITICAL COMPONENTS VERFICATION
OBJECTIVE
To verify that installed equipment and its critical components are as specified.
9.4.2
METHOD


9.4.3
Visually verify and record the details of the installed equipment against the specification
detailed in the results table.
Record and list all deviation/ non-conformance observed in Section 12.
TEST TOOL
Not applicable.
9.4.4
ACCEPTANCE CRITERIA
Installed equipment and its critical components are as specified.
9.4.5
RESULTS
A. AUTOCLAVE
Item
As Specified
Manufacturer
Tuttnauer
Model No.
69180 EP-1H
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
Equipment ID
Serial No.
Sterilizing
Agent/Media
Chamber Volume
Saturated Steam
Location
Level 4
1015 L
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
15 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.4.5
RESULTS (CONTINUED)
B. ELECTRIC STEAM GENERATOR
Item
As Specified
Manufacturer
Tuttnauer
Model No.
350GEN
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
Equipment ID
Serial No.
Power
60kW
C. VACUUM PUMP
Item
As Specified
Manufacturer
Speck
Model No.
V95
Equipment ID
Serial No.
Motor Power
Max Suction
Capacity
2.2kW/ 3 HP
70 m3/h
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
16 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.4.5
RESULTS (CONTINUED)
D. WATER PUMP
Item
As Specified
Manufacturer
Speck
Model No.
CY 4081
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
Equipment ID
Serial No.
Motor Power
2.2kW/ 3 HP
E. HEAT EXCHANGER
Item
As Specified
Manufacturer
Heatran
Model No.
31 ORZ2
Equipment ID
Serial No.
Type
Plate Type
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
17 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.4.5
RESULTS (CONTINUED)
F. AIR FILTER
Item
Manufacturer
As Specified
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
Domnick Hunter
Equipment ID
Filter Model No.
Lot No.
Filter Type
Filter Rating
Housing Model
No.
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
18 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.5
9.5.1
CONTROL SYSTEM VERIFICATION
OBJECTIVE
To verify that installed control system of the equipment is as specified.
9.5.2
METHOD


9.5.3
Visually verify and record the details of the control system against the specification
detailed in the results table.
Record and list all deviation/ non-conformance observed in Section 12.
TEST TOOL
Not applicable.
9.5.4
ACCEPTANCE CRITERIA
Installed control system of the equipment is as specified.
9.5.5
RESULTS
A. CONTROL PANEL
Item
As Specified
Manufacturer
CAT Technology
Model No
CAT 2007
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
Firmware Version
Serial No.
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
19 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.5.5
RESULTS (CONTINUED)
B. CONTROL BOARD
Item
As Specified
Manufacturer
CAT Technology
Model No.
CAT 2004
Version No.
No of PT100
Inputs
No of 4-20mA
Input
No of 4-20mA
Output
No of Digital
Inputs
No of Digital
Outputs
No of Serial Ports
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
6
7
2
16
24
4
C. PRINTER
Item
As Specified
Manufacturer
Seiko Instruments
Model No.
DPU-414
Serial No.
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
20 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.5.5
RESULTS (CONTINUED)
D. SOFTWARE/ APPLICATION
Item
Manufacturer
As Specified
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
CAT Technology
Programmer
File Name
Version
Backup Copy
Available
Backup File Name
Backup File
Version
Backup File
Storage Location
Yes
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
21 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.6
9.6.1
PRODUCT CONTACT PARTS MATERIALS REVIEW
OBJECTIVE
To verify that product contact parts materials are acceptable for use.
9.6.2
METHOD



9.6.3
Use one or both of the following methods:
a. Visual verification.
b. Material Certificate/ Declaration review.
Record details of observation in the following results table.
Record and list all deviation/ non-conformance observed in Section 12.
TEST TOOL
Not applicable.
9.6.4
ACCEPTANCE CRITERIA
All product contact parts materials are acceptable for use.
9.6.5
RESULTS
Item
Internal of
Chamber
Door
As Specified
Verification
Method
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
316 SS
304 SS
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
22 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.7
9.7.1
INSTRUMENTATION IDENTIFICATION AND CALIBRATION VERIFICATION
OBJECTIVE
To verify that critical instrumentation of the equipment are as specified. Ensure that all
calibrations for the instrumentations are current.
9.7.2
METHOD



9.7.3
Verify and record details of installed instruments in the following results table.
Verify that calibrations of instruments are valid. Ensure that calibration certificates are
available. Record location of where certificates are kept.
Record and list all deviation/ non-conformance observed in Section 12.
TEST TOOL
Not applicable.
9.7.4
ACCEPTANCE CRITERIA
All critical instrumentations are as specified and its calibrations are valid. Records of
calibration are correct and available.
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
23 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.7.5
RESULTS
DRAIN TEMPERATURE SENSOR
Item
As Specified
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
Tag Number
Manufacturer
MMB
Model No.
Serial No.
Type
Calibration Due
Date
Calibration Cert
No.
Calibration Cert.
Location
PT100
MEDIA TEMPERATURE SENSOR
Item
As Specified
Tag Number
Manufacturer
MMB
Model No.
Serial No.
Type
Calibration Due
Date
Calibration Cert
No.
Calibration Cert.
Location
PT100
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
24 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.7.5
RESULTS (CONTINUED)
LOAD TEMPERATURE SENSOR
Item
As Specified
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
Tag Number
Manufacturer
MMB
Model No.
Serial No.
Type
Calibration Due
Date
Calibration Cert
No.
Calibration Cert.
Location
PT100
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
25 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.7.5
RESULTS (CONTINUED)
CHAMBER PRESSURE TRANSDUCER
Item
As Specified
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
Tag Number
Manufacturer
Tecsis
Model No.
3296
Serial No.
Range
Calibration Due
Date
Calibration Cert
No.
Calibration Cert.
Location
0 – 10 bar
JACKET PRESSURE TRANSDUCER
Item
As Specified
Tag Number
Manufacturer
Tecsis
Model No.
3296
Serial No.
Tag No:
Calibration Due
Date
Calibration Cert
No.
Calibration Cert.
Location
0 – 10 bar
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
26 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.7.5
RESULTS (CONTINUED)
GASKET PRESSURE TRANSDUCER
Item
As Specified
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
Tag Number
Manufacturer
Tecsis
Model No.
3296
Serial No.
Range
Calibration Due
Date
Calibration Cert
No.
Calibration Cert.
Location
0 – 10 bar
CHAMBER PRESSURE GAUGE
Item
As Specified
Tag Number
Manufacturer
Wika
Serial No.
Range
Calibration Due
Date
Calibration Cert.
No.
Calibration Cert.
Location
0 – 500 kPa
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
27 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.7.5
RESULTS (CONTINUED)
JACKET PRESSURE GAUGE
Item
As Specified
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
Tag Number
Manufacturer
Wika
Serial No.
Range
Calibration Due
Date
Calibration Cert.
No.
Calibration Cert.
Location
0 – 500 kPa
GASKET PRESSURE GAUGE
Item
As Specified
Tag Number
Manufacturer
Wika
Serial No.
Range
Calibration Due
Date
Calibration Cert.
No.
Calibration Cert.
Location
0 – 500 kPa
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
28 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.8
9.8.1
CONTROL SYSTEM CONFIGURATION AND SETPOINT VERIFICATION
OBJECTIVE
To verify that control system configurations and system setpoints are as specified.
9.8.2
METHOD



9.8.3
Verify and record control system configuration and system setpoints in the following
results table. Ensure that recorded details are as specified.
Append system configuration and/or system setpoints printout if available.
Record and list all deviation/ non-conformance observed in Section 12.
TEST TOOL
Not applicable.
9.8.4
ACCEPTANCE CRITERIA
Control system configuration and system setpoints are as specified.
9.8.5
RESULTS
Item
As Specified
Chamber & Jacket
Safety Valve Set
point
40 psi
Gasket Safety
Valve Set point
40 psi
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
Pressure Switch
set point
Thermal Printer
SW1 Dip Switch
Thermal Printer
SW2 Dip Switch
Thermal Printer
SW3 Dip Switch
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
29 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.9
9.9.1
SAFETY FEATURES VERIFICATION
OBJECTIVE
To verify that system safety features are available as specified.
9.9.2
METHOD


9.9.3
Verify that listed safety features listed are available.
Record and list all deviation/ non-conformance observed in Section 12.
TEST TOOL
Not applicable.
9.9.4
ACCEPTANCE CRITERIA
System safety features are available as specified.
9.9.5
RESULTS
Item
Emergency Stop
(In Front)
As Specified
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
Available
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
30 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
9.10 PLANNED PREVENTATIVE MAINTENANCE VERIFICATION
9.10.1 OBJECTIVE
To verify that the autoclave has been entered onto the planned preventative maintenance
system.
9.10.2 METHOD


Verify that autoclave Tuttnauer Sterilizer 69180\1HSD-EP has been entered onto the
planned preventative maintenance system.
Record and list all deviation/ non-conformance observed in Section 12.
9.10.3 TEST TOOL
Not applicable.
9.10.4 ACCEPTANCE CRITERIA
The autoclave has been entered onto the planned preventative maintenance system.
9.10.5 RESULTS
Item
Tuttnauer
Sterilizer
69180\1HSD-EP
As Specified
As Found
Acceptable
(Yes/ No)
Verified By
(Initial &
Date)
Autoclave Tuttnauer Sterilizer
69180\1HSD-EP has been
entered onto the planned
preventative maintenance system
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
31 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
10 TEST INSTRUMENTATION
All test instrumentation used in this protocol shall be recorded in the table below. Calibration
of Test Instrumentation must be valid and copies of calibration certification shall be appended
to this document.
Instrument Type
Manufacturer
Serial Number
Calibration
Cert. No.
Cal Due
Date
Initial & Date
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
32 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
11 IDENTIFICATION OF PERSONNEL
The following is a record of each individual who is involved in the execution of this protocol.
Personnel profiles for all non-Xxxx personnel must be attached in this IQ protocol.
Company/
Title
Name
Department
Signatures
Full
Date
Initials
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
33 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
12 LIST OF DEVIATIONS
The following table details of all deviation/ non-conformance identified during the
performance of the IQ. All deviations/ non-conformances identified require the initiation of a
Deviation Report Form.
Deviation
Report No.
Ref.
Protocol
Section
Deviation Details
Corrective
Action
Implemented
By:
Date
Resolved
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
34 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
13 REVIEW AND APPROVALS
13.1 Approval To Proceed To OQ
The tests and checks specified in this protocol have been carried out in a satisfactory manner
and the results documented correctly. All remaining Deviation Reports listed in Section 12
have been considered and will not affect the correct operation of the equipment. The Tuttnauer
Sterilizer is ready for Operational Qualification.
Note: If there are no outstanding Deviation Reports proceed directly to Final Approval.
APPROVAL TO PROCEED TO OQ
Department
Title
Name
Signature
Date
Technical Services
Engineering &
Maintenance
Production
Quality Assurance
13.2 Final Approval
The tests and checks specified in this protocol have been carried out in a satisfactory manner
and the results documented correctly. All Deviation Reports listed in Section 12 have been
resolved. The Tuttnauer Sterilizer Installation Qualification is complete.
FINAL APPROVAL
Department
Title
Name
Signature
Date
Technical Services
Engineering &
Maintenance
Production
Quality Assurance
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
35 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
14 LIST OF APPENDICES
The table below documents and numbers each Appendix of IQ. If extra Appendices are
required during execution, they may be entered by hand.
Appendix
Number
Appendix Title
Protocol Test
Section(s)
Comments:
Reviewed By:
Date:
FOR USE BY AFFILIATES OF XXXX PTE LTD
THIS DOCUMENT CONTAINS PROPRIETARY INFORMATION –
IT MUST NOT BE REPRODUCED OR DISCLOSED TO OTHERS WITHOUT PRIOR WRITTEN APPROVAL
No of
Pages
DOCUMENT NUMBER
IQP/QA/027/2005
TITLE
REVISION
01
PAGE
36 of 36
INSTALLATION QUALIFICATION FOR
TUTTNAUER STERILIZER 69180\1HSD-EP
15 PROTOCOL REVISION HISTORY
Rev. No
Date
Issued By
Remarks
FOR USE BY AFFILIATES OF XXXX PTE LTD