CASS TEST CHAMBER as per following specifications. 1. Performance : To carry out CASS Test as per IS 5528: 1985 2. Work Space : 3. Temperature range : +50C ±1C (without refrigeration system) Perfect Operation of the chamber is only guaranteed at an ambient temperature of +18°C to +28°C. The performance and accuracies are in steady state condition without samples/specimens at an ambient temperature of +23°C(±2°C). 4. Humidity : Saturated Condition. 5. Fog collection : Between 0.5 to 2 ml per hour 6. Construction : The chamber will be fabricated out of anticorrosive material and the interior is of 6mm thick FRP sheet. Exterior is of 6 mm thick PVC sheets all over. Top opening lid will be provided of acrylic/polycarbonate sheet of suitable thickness. Top lid will be tapered at 110° to avoid falling of condensation on the item under test. Inspection lamp will be provided located on the top lid. Water seal will be provided for the chamber door to rest to avoid leakage of salt fumes, if any. Control panel will be provided which will house the ON/OFF switches, indicating lamps, temperature indicator controller, fuses, contactors, MCBs, etc. with wiring as per IS code. The wiring will be channel type wiring using suitable current rated copper wires. The chamber will be floor mounting type with castors. Indirect heating system with wall heaters embedded in walls will be provided with thermostat to achieve the temperature in the work space. The heaters will heat the wall and then the wall will heat the work space. 7. Preheater : Preheater will be provided to preheat the air before it is admitted to the air saturator tower with thermostat and is housed on the right hand side below the control panel. 8. Humidity system/Air saturator tower: Air saturator tower will be provided to saturate the air. The air saturator tower will be provided with air inlet and outlet connections, water connection with automatic refilling system and air bubbling arrangement with filter. The air will be bubbled through the air saturator tower before it is admitted to the nozzle to ensure full saturation of the air being supplied to the spray nozzle. Air saturator tower will be provided with thermostat to control water temperature. 9. Spray nozzle : Spray nozzle, made out of glass will be provided to give uniform fog collection rate. A) 1500mm L x 600mm W x 500mm D (450 ltrs) B) 900mm L x 750mm W x 600mm D ( 400 ltrs) Manufacturers of Entire range of Environmental Test Chambers as per national & international standards. 10. Air Filter Regulator : To control the air pressure and ensure clean and oil free air to the spray nozzle, an air pressure filter regulator with micro filter and activated carbon filter will be provided. Compressed air will be provided by you at inlet of Air Filter Regulator. 11. a) Instrumentation: Dual loop Self-tune microprocessor based PID temperature humidity controller will be provided with PT 100 as temperature sensor for indication and control of temperature & humidity with direct RH display. The controller will be L7T make. 4.3” Touchscreen will be provided as interface to the controller. The Tocuhscreen will be mounted in front next to the chamber so all the data can be seen standing near the door. b) Solid state relay for control of heaters. Calibration certificate with traceability certificate will be provided. 12. Control Panel & Wiring : Control panel will be provided which will house the temperature indicator controller, Mains On/Off switches, fuses, contactors, indicating lamps, etc. All the wiring will be as per IS code. Channel type wiring will be done with suitable current rated copper wires with marking ferrules and crimped dowel terminals. 13. Utilities : Salt Solution tank of 50 liters and water storage tank of 50 liters will be provided. The corrosion test chamber will be provided with PVC rod for hanging the components as well as PVC racks with 15 deg. and 30 deg. angles for mounting the panel for testing. Beaker and measuring cylinder will be provided for measuring of fog collection. Inspection lamp located on the top lid frame. Electronic timer to ON/OFF Spray after elapsed time. Safety protection : Back up fuses protection for mains and individual circuit. Over temperature safety cut off thermostat with audio visual alarm. Water level controller with automatic refilling arrangement for air saturator tower. 14. 15. Power supply : 415 Volts, ±5%, 3 phase neutral, 50 c/s, AC supply. 16. Operational Instruction Manual : Three copies of operational instruction manual will be provided along with the chamber. Manufacturers of Entire range of Environmental Test Chambers as per national & international standards.