VATECH Implagraphy with Cephalo Assembly Guide 1. Make the system in upright position. * Be careful for Sensor part! Fragile! Do not push! 2. Open rotating unit upper case and release 1bolt!!! 3. Remove the carrying bar and fix the the hole with two bolts 4. Connect Cephalo to Column Column Connect Cephalo 4bolts Pull side cover by hand forcibly and put cables inside Connect Cephalo two cables Please make sure if all cables are connected ** We tried to keep the cable as they are without disconnecting 5. Attach the back cover to column and connect UP/DOWN Switch. 6. Cable connection CT RECON USB Lan Cable Grabber Card Cable Communication Cable(Serial Cable) Router : PC, Pano, Cep * PLEASE USE THIS PC for DENTAL XRAY PURPOSE ONLY!!! DO NOT INSTALL OTHER SOFTWARE OR DAMAGE THE SOFTWARE INSIDE. This is BACKUP SOFTWARE USB and CT RECON KEY. 7. In carton box, we put Manual, Bites, PC & Monitor, cables, and etc *** D drive in PC has all backup files. Also, we put BACKUP USB. Please keep them well just in case. Please do not press this EMERGENCY BUTTON. Up/Down switch will not work. PANO&Cep SENSOR and PANO&Cep SHOT PREPARATION. ** Please always connect Panorama sensor and when you take CT shot, please disconnect Panoarama sensor and hang it to the holder. ** Sensor Lamp check before taking Panorama shot POWER STATUS 1 STATUS 2 => After this lamp is blinking you can take Panorama shot. Please do the same to Cephalo Sensor change and shot preparation as Panorama shot * ALWAYS TURN ON PC FIRST AND THEN PAX I 3D MAINBODY WHEN YOU TURN OFF, PAX I 3D MAINBODYD FIRST AND THEN PC Warning!! ** Change the power cable connectors to your country's standards so that they can be grounded. *** After you take 3~4 consecutive Xray shot, dental Xray should be cooling off for 10~20min according to room temperature. Or it will automatically cool off by itself. Thank you!