Phadia 250 EliA anti-TPO Test-Specific Reagents EliA anti-TPO Well (Art. No. 14-5641-01) anti-TPO Well; short name: tpo anti-TPO FLUOROENZYMEIMMUNOASSAY FOR THYROID PEROXIDASE ANTIBODIES FOR IN VITRO DIAGNOSTIC USE Human serum in PBS containing BSA, detergent and sodium azide (0.095 %); symbol: pos EliA uses a modular reagent system. All information needed to understand the use of the EliA tests can be found in the analyte specific DfU and the corresponding EliA Control DfU. INTENDED USE EliA anti-TPO is intended for the in vitro quantitative measurement of IgG antibodies directed to thyroid peroxidase (TPO) in human serum and plasma (Li-heparin, EDTA) as an aid in the clinical diagnosis of autoimmune thyroiditis and Graves' disease in conjunction with other laboratory and clinical findings. EliA anti-TPO uses the EliA IgG method on the instrument Phadia 250. ready for use; store dry at 2-8 °C until expiration date Control containing IgG antibodies to TPO and TG 6 single-use vials (0.3 ml each); sufficient for 2 determinations per vial Ready for use; store at 2-8 °C until expiration date EliA Thyroid Positive Control 250 is prepared from selected pooled human sera. EliA IgG/IgM/IgA Negative Control 250 (Art. No 83-1037-01) Human serum in PBS containing BSA, detergent and sodium azide (0.095 %); symbol: neg SUMMARY AND EXPLANATION OF THE TEST Thyroid autoimmunity comprises a number of distinct but pathogenically related autoimmune disorders of the thyroid gland, such as Graves’ disease or Hashimoto’s thyroiditis. Among other things, these diseases are characterized by the presence of autoantibodies that are directed against thyroid antigens, such as TSH receptor, thyroglobulin (TG) or thyroid peroxidase (TPO, formerly known as the thyroid microsomal antigen). Thyroid peroxidase (TPO) is an integral membrane hemoprotein (100 kDa) which catalyzes the iodination of thyroglobulin (TG), as well as the coupling of two di-iodotyrosine residues in the TG molecule to form thyroxine.1 TPO antibodies are present in the serum of most patients with Hashimoto's thyroiditis, in more than 70% of those with Graves' disease, and, to a variable degree, in patients with non-thyroid autoimmune diseases and some normal subjects.2 The titers do not correlate with thyroid functional status. Individual patients with either disorder may have decreases in TPO antibody titers during therapy, but others have no change.3 On the other hand, there is a good correlation between the degree of lymphocytic infiltration of the thyroid gland and the presence of TPO antibodies.2 For the development of an immunoassay it is necessary to obtain reproducible qualities and amounts of a highly purified antigen. The TPO in the EliA anti-TPO assay is synthesized in the eukaryotic Baculovirus/Sf9-cell system to achieve maximum antigen purity and quality.4 Multiparameter control containing normal sera from healthy donors 6 single-use vials (0.3 ml each); sufficient for 2 determinations per vial ready for use; store at 2-8 °C until expiration date EliA IgG/IgM/IgA Negative Control 250 is prepared from selected pooled human sera. EliA Method-Specific Reagents (Phadia 250) EliA Sample Diluent (Art. No 83-1023-01) Sample Diluent (yellow colored); PBS containing BSA, detergent and sodium azide (0.095 %) 6 bottles (48 ml each); sufficient for ≥ 6 x 180 dilutions ready for use; store at 2-8 °C until expiration date EliA IgG Conjugate 50 (Art. No 83-1017-01) IgG Conjugate (blue colored); ß-Galactosidase anti-IgG (mouse monoclonal antibodies) in PBS containing BSA and sodium azide (0.06 %); symbol: El-G PRINCIPLES OF THE PROCEDURE 6 wedge shaped bottles (5 ml each); sufficient for 6 x 50 determinations ready for use; store at 2-8 °C until expiration date DO NOT FREEZE DO NOT REUSE EliA IgG Conjugate 200 (Art. No 83-1018-01) The EliA anti-TPO Wells are coated with human recombinant TPO antigen. If present in the patient's specimen, antibodies to TPO bind to their specific antigen. After washing away non-bound antibodies, enzyme-labeled antibodies against human IgG antibodies (EliA IgG Conjugate) are added to form an antibody-conjugate complex. After incubation, non-bound conjugate is washed away and the bound complex is incubated with a Development Solution. After stopping the reaction, the fluorescence in the reaction mixture is measured. The higher the response value, the higher the amount of antibody bound and detected in the sample tested. To evaluate test results, the response for patient samples is compared directly to the response for calibrators. IgG Conjugate (blue colored); ß-Galactosidase anti-IgG (mouse monoclonal antibodies) in PBS containing BSA and sodium azide (0.06 %); symbol: El-G 6 wedge shaped bottles (19 ml each); sufficient for 6 x 200 determinations ready for use; store at 2-8 °C until expiration date DO NOT FREEZE DO NOT REUSE EliA IgG Calibrator Strips (Art. No 83-1015-01) human IgG (0, 4, 10, 20, 100, 600 μg/l); in PBS containing BSA, detergent and sodium azide (0.095 %) REAGENTS / MATERIAL EliA reagents are available as modular packages, each purchased separately. All packages except for the EliA Thyroid Positive Control 250 and the EliA IgG/IgM/IgA Negative Control 250 are required to carry out an EliA anti-TPO test. The EliA anti-TPO Wells are packed in carriers which are stored in sealed aluminium foil bags containing a desiccant. Issued March 2016 4 carriers (16 wells each); sufficient for 64 determinations EliA Thyroid Positive Control 250 (Art. No 83-1113-01) DIRECTIONS FOR USE CONTENTS 250-5641- 020 / US coated with human recombinant thyroid peroxidase antigen 5 strips 6 single-use vials per strip (0.3 ml each); sufficient for one calibration curve (double determination) ready for use; store at 2-8 °C until expiration date Manufactured from human sera. 1/5 Published 2016-04-15 anti-TPO WARNING! Reagents contain sodium azide (NaN3) as a preservative. NaN3 may be toxic if ingested or absorbed by skin or eyes. NaN3 may react with lead and copper plumbing to form highly explosive metal azides. On disposal, flush with a large volume of water to prevent azide build-up. Please refer to decontamination procedures as outlined by CDC or other local and national guidelines. Waste Bottle and ImmunoCAP/EliA Well Waste Container may be contaminated by potentially infectious material. Use appropriate safety measures and wear gloves. EliA IgG Curve Control Strips (Art. No 83-1016-01) human IgG (20 μg/l); in PBS containing BSA, detergent and sodium azide (0.095 %) symbol: CC-1 5 strips Each strip contains 6 x 0.3 ml CC-1 (double determination) ready for use; store at 2-8 °C until expiration date Manufactured from human sera. EliA IgG Calibrator Well (Art. No 14-5509-01) IgG Calibrator Well coated with mouse monoclonal antibodies; short name: Gcal 4 carriers (12 wells each); sufficient for 48 determinations ready for use; store dry at 2-8 °C until expiration date Indication of Instability Phadia 250 Instrument Software has built-in acceptance limits for the calibration curve and the curve control. EliA Wells are moisture sensitive. An activity loss that might occur due to inappropriate handling can be detected using the appropriate EliA Control. For more information see Phadia 250 User's Guide/Reference Manual. Phadia 250 General Reagents Development Solution (Art. No. 10-9440-01) Development Solution 0.01 % 4-Methylumbelliferyl-β-D-galactoside, <0.0010% preservative* 6 bottles (17 ml each); sufficient for 6 x >170 determinations ready for use; store at 2-8 °C until expiration date DO NOT FREEZE INSTRUMENT The Phadia 250 Instrument processes all steps of the test. For further information regarding test set-up, instrumentation and software etc. see Phadia 250 User's Guide/ Reference Manual. Development Solution (Art. No. 10-9441-01) Development Solution 0.01 % 4-Methylumbelliferyl-β-D-galactoside, <0.0010% preservative* 6 bottles (11 ml each); sufficient for 6 x >110 determinations ready for use; store at 2-8 °C until expiration date DO NOT FREEZE SPECIMEN COLLECTION, HANDLING AND PREPARATION The procedure can be performed with serum or plasma specimens. Lipemic, hemolyzed or microbially contaminated samples may give poor results and should not be used. CLSI-Document H18-A4 recommends the following storage conditions for samples: • Separated serum/plasma should remain at room temperature for no longer than eight hours. • If assays will not be completed within eight hours, serum/plasma should be refrigerated (2 to 8°C). • If assays are not completed within 48 hours, or the separated serum/plasma will be stored beyond 48 hours, serum/plasma should be frozen at or below -20°C. Avoid repeated freezing and thawing. Note: It is the responsibility of the individual laboratory to use all available references and/or its own studies to determine specific stability criteria for its laboratory. In general, laboratories should perform validation studies before implementing a change in specimen acceptance criteria. Stop Solution (Art. No. 10-9442-01) Stop Solution 4 % Sodium Carbonate 6 bottles (119 ml each); sufficient for 6 x >560 determinations ready for use; store at 2-32 °C until expiration date Stop Solution (Art. No. 10-9479-01) Stop Solution 4 % Sodium Carbonate 6 bottles (65 ml each); sufficient for 6 x >292 determinations ready for use; store at 2-32 °C until expiration date Dilution Plates (Art. No. 12-3907-08) MicroWell ™ plates with 96 wells, 0.5 ml each; Polypropylene 100 plates per package; sufficient for 100 x 96 samples ready for use DO NOT REUSE * Preservative: mixture of 5-chloro-2-methyl-2H-isothiazol-3-one [EC no. 247-500-7] and 2-methyl-2H-isothiazol-3-one [EC no. 220-239-6] (3:1). Washing Solution (Art. No. 10-9422-01/10-9202-01) For information see separate Washing Solution package insert. The expiration date for each of the complete packages is stated on the outer label. However, each component is stable until the date stated on the respective vial label. Sample Dilution Samples must be diluted with EliA Sample Diluent. A 1:200 dilution of the samples is required for the EliA anti-TPO test. Samples can be diluted manually, but instrument dilution is recommended and is a default setting in Phadia 250 Software. Material required but not provided by Phadia AB: • Measuring cylinder 1000 ml • Clinical laboratory reagent water (CLRW, may be used in place of Type I and Type II water) as defined by CLSI Standards Document, C3-A4: Preparation and Testing of reagent Water in the Clinical Laboratory; Approved Guideline-Fourth Edition PROCEDURE Handling of EliA anti-TPO Well In the Phadia 250 storage chamber, carriers are stable for up to 28 days. If you are not expecting to use them up within this time, the carriers should be loaded via the Phadia 250 Loading Tray and, for stability reasons, must be put back into the desiccant-containing foil bag directly after the run. Because it is important to store the wells in dry conditions at 2-8°C, the bag must be properly resealed. If stored under these conditions, the shelf-life from the date of first opening is 9 months, if not limited by the expiry date stated on the carrier and foil bag. WARNINGS AND PRECAUTIONS • For in vitro diagnostic use. • Do not use reagents beyond their expiration dates. • We do not recommend to pool reagents. • Some of the reagents are manufactured from human blood components. The source materials have been tested by FDA approved immunoassay for hepatitis B surface antigen, for antibodies to HIV1, HIV2 and hepatitis C virus and found negative. Nevertheless, all recommended precautions for the handling of blood derivatives should be observed. Please refer to Human Health Service (HHS) Publication No. (CDC) 93-8395 or local and national guidelines on laboratory safety procedures. • Caution: Federal Law restricts this device to sale by or on the order of a licensed practitioner. 250-5641- 020 / US Issued March 2016 Lot specific barcode Use the built-in barcode reader to enter the lot specific information of EliA anti-TPO Well, EliA IgG Calibrator Well and EliA IgG Conjugate. In case of manual handling make sure to enter the characters below the barcode. 2/5 Published 2016-04-15 anti-TPO On-board stability of reagents CALIBRATION AND REFERENCE MATERIAL • EliA Wells EliA Well carriers can be stored on-board for 28 days at 2-8°C or 24 hours at room temperature. The calibration curve is obtained with EliA IgG Calibrators which are run in duplicate. The curve is stored and subsequent tests are evaluated against the stored curve using only the EliA IgG Curve Control (run in duplicate). The IgG Calibrators are traceable via an unbroken chain of calibrations to the International Reference Preparation (IRP) 67/86 of Human Serum Immunoglobulins A, G and M from World Health Organization (WHO). • EliA Calibrator Strips, EliA Curve Control Strips Can be stored on-board for 28 days. • EliA Sample Diluent Can be stored on-board for 7 days at room temperature. Re-cap bottles every night. A new calibration curve must be run when: • the last calibration was made more than one month ago or • a new lot of EliA IgG Conjugate is introduced or • when the EliA IgG Curve Control is outside the specified limits (defined in Phadia 250 Instrument Software). • EliA Conjugate Single use reagent, open vials must not be stored. • Development Solution Can be stored on-board for a total of 40h at room temperature. Can be used 5 times during shelf life and be stored at room temperature for 8 hours on each occasion. Recap bottles every night. During weekends or longer interval between instrument usage it is recommended to store bottles at 2-8°C. • Stop Solution Can be stored on-board for 7 days at room temperature. Re-cap bottles every night. EliA anti-TPO is calibrated against the NIBSC research standard for anti-thyroid microsome serum code 66/387. Results are given in International Units (IU/ml). QUALITY CONTROL Control Specimens Good laboratory practice requires that quality control specimens should be included in every run. Any material used should be assayed repeatedly to establish mean values and acceptance ranges. EliA Controls are available for the quality control of the measurements. • Washing Solution Prepared solution can be stored on-board for 7 days at room temperature. Discard every seventh day and perform weekly maintenance according to instrument user manual. CALCULATION AND INTERPRETATION OF RESULTS Volumes per determination Reagent volumes per determination Calibrator 90 μl EliA IgG Conjugate 90 μl Development Solution 90 μl Stop Solution 200 μl Presentation of Results Phadia 250 measures specific IgG concentrations in μg/l. By using a conversion factor given by the lot-specific code of the EliA anti-TPO Well, the results are automatically converted to IU/ml. Interpretation of Test Results The ranges (negative, equivocal, positive) recommended for the evaluation of the results are given in the table below. Sample volumes per determination Manual dilution: 90 μl of diluted sample Instrument dilution (1:200): 10 μl of non diluted sample For tube-specific dead volumes see Phadia 250 User's Guide/Reference Manual. negative equivocal positive IU/ml < 25 25 – 35 > 35 LIMITATIONS * The residual volume depends on the number of samples and dilution method used. A definitive clinical diagnosis should not be based on the results of a single diagnostic method, but should only be made by the physician after all clinical and laboratory findings have been evaluated. Procedural comments • From one sample diluted by the instrument (1:200), up to 11 determinations can be made. • When using software default, samples are run in single determination. • Washing Solution must be at room temperature when used. • The first result is available after approx. 2 hours and further results at one minute intervals afterwards. Up to 5 x 10 samples can be loaded continuously and are processed by random access. • Incubations are automatically performed at 37 °C (98.6 °F). Issued March 2016 Unit EliA anti-TPO Good laboratory practice requires that each laboratory establishes its own range of expected values. In case of equivocal results, we recommend to retest the patient after 8 –12 weeks. Reagent volumes per 200 determinations Washing Solution 5-7 l* Rinse Solution 5-6 l* 250-5641- 020 / US Test EXPECTED VALUES Antibody prevalence in autoimmune patients varies widely depending on disease area. A significant percentage (>10%) of healthy subjects was reported to have anti-thyroid peroxidase antibodies.5 Expected values may vary depending on the population tested. 3/5 Published 2016-04-15 anti-TPO Results Obtained for Healthy Subjects The frequency distribution for TPO antibodies was investigated in a group of apparently healthy subjects equally distributed by age and gender, using sera from Caucasian, African American, Hispanic and Asian population obtained from a blood bank. The results are given in the table below. Test EliA anti-TPO Unit IU/ml No. of samples Median value 90th percentile 95th percentile 604 4.4 14.2 59.9 METHOD COMPARISON For analysis of agreement with the predicate device 498 sera were measured. n=498 297 167 331 498 Sample Unit Mean value Diagnostic groups Graves` Disease (= Thyroid-AID) Autoimmune thyroiditis (= Thyroid-AID) Non-AI Thyroid Disease Thyroid cancer Multi-nodular goiter Connective Tissue Disease Crohn`s disease Ulcerative colitis Celiac disease Primary Biliary Cirrhosis HIV infection HCV infection HBV infection Other infection Cancer6,7 Rheumatoid arthritis Hypergammaglobulinemia Systemic lupus erythematosus Sjögren`s syndrome Type I diabetes mellitus8 Type II diabetes mellitus Pregnant women (all trimesters)9 Pre-eclampsia10 Miscarriage10 Myasthenia gravis Pernicious anemia 1 IU/ml 2 Coefficients of variation (%) Intra-Run Inter-Run 15.7 5.7 2.3 IU/ml 27.4 4.6 1.5 3 IU/ml 39.4 3.8 0.5 4 IU/ml 66.7 3.9 1.0 5 IU/ml 218.0 4.6 1.3 6 IU/ml 480.1 4.7 1.0 7 IU/ml 814.9 3.9 0.4 8 IU/ml 1212.6 4.0 0.0 Issued March 2016 100% x 167/167 Agreement [%] 95% CI 100.0 97.8 – 100.0 100% x 297/331 89.7 85.9 – 92.8 100% x (167+297)/498 93.2 90.6 – 95.2 CLINICAL PERFORMANCE Precision To determine the precision of the assay, the variability was assessed in a study with 21 runs by examining the samples in 252 replicates on 3 instruments over 7 days with a calibration curve included into each run. 250-5641- 020 / US 201 297 Total Agreement Interferences Undiluted samples were analyzed for interference with EliA anti-TPO. No interference could be observed for the following substances at the levels indicated: Bilirubin C/F (201/192 mg/dl), Lipemic factor (ClinOleic) (1%), Hemoglobin (4960 mg/dl), Rheumatoid factor (RF) (500 IU/ml), Thyroxine (2.5 ng/dl) and Iodide (3 mM). EliA anti-TPO 34 0 Negative Percent Agreement Specificity The EliA anti-TPO test permits the determination of IgG antibodies directed against the TPO antigen as described in section “Reagents”. Test 167 EliA anti-TPO Negative (≤ 35 IU/ml) Positive Percent Agreement Measuring Range The technical measuring range (Limit of Detection, upper limit of the calibration curve) for EliA anti-TPO is from 4 to 1542 IU/ml. Results above the upper limit are generally reported as “above”. No hook effects could be observed for concentrations up to 20725 IU/ml. Linearity was shown for samples in the range from 4 – 1542 IU/ml (linear range). Please note that due to differing binding characteristics of the antibodies in patient samples, not all sera can be diluted linearly within the technical measuring range. Total EliA anti-TPO Positive (> 35 IU/ml) Total PERFORMANCE CHARACTERISTICS Predicate Device VarelisA TPO Positive Negative (> 100 IU/ml) (≤ 100 IU/ml) 4/5 Published 2016-04-15 n 100 115 48 21 4 15 20 20 15 10 20 20 20 20 56 12 15 24 14 15 15 90 5 9 7 8 No (%) Positive EliA anti-TG 81 (81%) 96 (83.5%) 5 (10.4%) 2 (9.5%) 0 (0%) 1 (6.7%) 0 (0%) 0 (0%) 2 (13.3%) 0 (0%) 1 (5%) 1 (5%) 0 (0%) 0 (0%) 11 (19.6%) 1 (8.3%) 0 (0%) 1 (4.2%) 1 (7.1%) 2 (13.3%) 2 (13.3%) 10 (11.1%) 4 (80%) 2 (22.2%) 1 (14.3%) 1 (12.5%) No (%) Positive VarelisA TG 68 (68%) 88 (76.5%) 5 (10.4%) 1 (4.8%) 0 (0%) 1 (6.7%) 0 (0%) 0 (0%) 2 (13.3%) 0 (0%) 0 (0%) 1 (5%) 0 (0%) 0 (0%) 5 (8.9%) 1 (8.3%) 0 (0%) 0 (0%) 1 (7.1%) 2 (13.3%) 2 (13.3%) 10 (11.1%) 3 (60%) 1 (11.1%) 0 (0%) 0 (0%) anti-TPO WARRANTY The performance data presented here was obtained using the procedure indicated. Any change or modification in the procedure not recommended by Phadia AB may affect the results, in which event Phadia AB disclaims all warranties expressed, implied or statutory, including the implied warranty of merchantability and fitness for use. Phadia AB and its authorized distributors, in such event, shall not be liable for damages, indirect or consequential. REFERENCES 1. Trotter WR, Belyavin G, Wadhams A. Precipitating and complement-fixing antibodies in Hashimoto's disease. Proc R Soc Med 1957; 50:961 2. Portmann L, Hamada N, Heinrich G, de Groot LJ. Anti-thyroid peroxidase antibody in patients with autoimmune thyroid disease: Possible identity with antimicrosomal antibody. J Clin Endocrinol Metab 1985; 61:1001-1003 3. Mariotti S, Caturegli P, Piccolo P, et al. Antithyroid Perixodase autoantibodies in thyroid disease. J Clin Endocrinol Metab 1990; 71:661-9 4. Berthold H, Steffens U, Northemann W. Human Thyroid Peroxidase: Autoantibody Recognition Depends on the Natural Conformation. J Clin Lab Anal 1993; 7:401-4 5. Hollowell JG, Staehling NW, Flanders WD, et al. Serum TSH, T4, and Thyroid Antibodies in the United States Population (1988 to 1994): National Health and Nutrition Examination Survey (NHANES III) J Clin Endocrinol Metabol 2002; 87: 489 6. Turken O, Narin Y, Demirbas S, et al. Breast cancer in association with thyroid disorders. Breast Cancer Research Vol 5 2003; No 5 7. Jiskra J, Límanová Z, Barkmanová J, et al. Autoimmune Thyroid Diseases in Women with Breast Cancer and Colorectal Cancer. Physiol R 2004; 53:693-702 8. Grzelka A, Araszkiewicz A, Uruska A, et al. Prevalence of Anti-Thyroid Peroxidase in Adults with Type 1 Diabetes Participating in PoznaĆ Prospective Study. Adv Clin Med 2015; 24(1):79-84 9. Stagnaro-Green A, Abalovich M, Alexander E, et al. Guidelines of the American Thyroid Association for the Diagnosis and Management of Thyroid Disease During Pregnancy and Postpartum. Thyroid 2011; 21(10): 1081-1125 10. Karakosta P, Alegakis D, Georgiou V, et al. Thyroid Dysfunction and Autoantibodies in Early Pregnancy Are Associated with Increased Risk of Gestational Diabetes and Adverse Birth Outcomes. J Clin Endocrinol Metab 2012; 97(12):4464–4472 ßLOTß Ä ' ( ßIVDß Batch code À Contains x determinations Biological Risk § Read Directions for Use Store at 2-8°C/35-46°F & Manufactured by Expiration date " Do not reuse in a second run For in vitro diagnostic use & Phadia AB ) Rapsgatan 7P P.O. Box 6460 751 37 Uppsala Sweden Tel: +46-18-16 50 60 Fax: +46-18-14 03 58 Autoimmunity@phadia.com www.phadia.com © Phadia GmbH, Freiburg, Germany 250-5641- 020 / US Issued March 2016 5/5 Published 2016-04-15 anti-TPO