SERVICE MANUAL DENTA-CART 404 SERVICE MANUAL DENTA-CART 404 SERVICE MANUAL DENTA-CART 404 Table of contents Entry Test......................................................................................................................................... 3 1.1 Cleaning ................................................................................................................................... 3 1.2 Case ......................................................................................................................................... 3 1.3 Power supply cable.................................................................................................................. 3 1.4 Water- and air system ............................................................................................................. 3 1.5 Handpieces .............................................................................................................................. 3 Maintenance.................................................................................................................................... 4 2.1 Annual Maintenance ............................................................................................................... 4 2.2 Three-year maintenance ......................................................................................................... 6 Troubleshooting and repair ............................................................................................................. 9 3.1 No function of the unit ............................................................................................................ 9 3.1.1 Power Supply ................................................................................................................... 9 3.1.2 Fuses ................................................................................................................................ 9 3.1.3 Converter ......................................................................................................................... 9 3.2 3.2.1 Scaler ............................................................................................................................... 9 3.2.2 Electric Motor ................................................................................................................ 10 3.2.3 Turbine / Air-driven handpiece ..................................................................................... 11 3.2.4 Three-way syringe ......................................................................................................... 12 3.3 Water system ........................................................................................................................ 12 3.3.1 Leakage of the water system......................................................................................... 12 3.3.2 Spray water-bottle overflows ........................................................................................ 12 3.4 1 Handpieces .............................................................................................................................. 9 Air system .............................................................................................................................. 12 Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 Final Test........................................................................................................................................ 13 4.1 Case ....................................................................................................................................... 13 4.2 Power supply cable................................................................................................................ 13 4.3 Water- and air system ........................................................................................................... 13 4.4 Handpieces ............................................................................................................................ 13 Parts list ......................................................................................................................................... 14 5.1 DENTA-CART 404 – PICTURE A .............................................................................................. 14 5.2 Surface Plate DENTA-CART 404 – PICTURE B ........................................................................ 15 5.3 Lower Plate 1 DENTA-CART 404 – PICTURE C ....................................................................... 16 5.4 Lower Plate 2 DENTA-CART 404 – PICTURE D ....................................................................... 17 5.5 Accessories – PICTURE E ........................................................................................................ 18 Service ........................................................................................................................................... 19 2 Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 Entry Test In order to record the condition, the following work should be carried out before each maintenance and repair: 1.1 Cleaning A basic cleaning of the unit must be carried out before starting the work. 1.2 Case Visual check for intactness. Defective parts must be replaced. 1.3 Power supply cable Visual check for intactness. A defective power supply cable D-244063 must be replaced. 1.4 Water- and air system Inspection of the water system: Visual inspection of the entire water system for leaks. Search and repair of leakage according to chapter 3.3. Inspection of the air system: Visual inspection of the entire air system for leaks. Search and repair of leakage according to chapter 3.4. 1.5 Handpieces 3 Visual inspection of the integrity of all handpieces and their hoses. Checking the function of the electric motor. If the electric motor does not work, troubleshooting and repair must be carried out in accordance to chapter 3.2.2. Check for heat development on the motor. If the motor becomes hot during operation, it must be replaced. Checking the function of the scaler. If the scaler does not work, troubleshooting and repair must be carried out in accordance to chapter 3.2.1. Checking the wear of the used scaler tips using the "EMS Tip Card". If the wear is under minimum, the tips must be replaced. Checking the function of the air driven handpieces. If the turbine does not work, troubleshooting and repair must be carried out in accordance to chapter 3.2.3. Check the function of the three-way syringe. If the three-way syringe does not work, troubleshooting and repair must be carried out in accordance to chapter 3.2.4. Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 Maintenance 2.1 Annual Maintenance All handpieces and maintenance documents have to be with the unit for maintenance. In order to guarantee the correct functioning of the device, the following items must be replaced at each annual maintenance. PIECE 1x ART. NR. D-24D207 MANDATORY PARTS Water Valve (2-way) 2x D-24D205 Check Valve 1x D-24F005 Hose 3x2 yellow 3 meters 10X D-24K003 Y-Connector 1/16“ 2x D-24F073 Hose complete for Spray water tank 2x D-24304 Gasket ring 19x3 4 PICTURE Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 1x D-244B001 Return air filter In order to guarantee the correct functioning of the unit, the following items must be replaced depending on wear and tear. PIECE 1x ART. NR. D-24H001 ADDITIONAL PARTS Syringe tip 1x D-244500 Membrane for control block, Quad Block 5 PICTURE The recommended reference time for an annual maintenance is 3h. The maintenance document must be completed in full after each service After all parts to be exchanged have been replaced and possible faults have been repaired (see chapter 3), continue with final test chapter 4. Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 2.2 Three-year maintenance All handpieces and maintenance documents have to be with the unit for maintenance. In order to guarantee the correct functioning of the device, the following items must be replaced at each three-yearly maintenance. PIECE 1x ART. NR. D-244500 MANDATORY PARTS Membrane for control block, Quad Block 1x D-24D207 Water Valve (2-way) 1x D-24D205 Check Valve 2x D-24F073 Hose complete for Spray water tank 1x D-24A009 Electric switch NO 1x D-24D206 Toggle-valve (3-Way) 6 PICTURE Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 1x D-24D211 Routing valve 1x D-24D208 Toggle Routing Selector 1x D-24D202 Shuttle valve 1x D-24F005 Hose 3x2 yellow 3 meters 1X D-24F006 Hose 3x2 transparent 3 meters 10X D-24K003 Y-Connector 1/16“ 1x D-244B001 Return air filter 7 Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 4x D-24D209 Automatic Holder Valve NC 2x D-24304 Gasket ring 19x3 1x D-24H001 Syringe tip 8 The recommended reference time for a three-year maintenance is 5h. The maintenance document must be completed in full after each service. After all parts to be exchanged have been replaced and possible faults have been repaired (see chapter 3), continue with final test chapter 4. Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 Troubleshooting and repair Before starting any repair, make sure that the unit is Set-up correct and as specified (three-way syringe, foot pedal, cables, hoses, bottles). 3.1 No function of the unit 3.1.1 Power Supply Check the power supply of the unit. Check the building plug socket, cable strips and extension cables for proper function. Check the power supply cable D-244063 for damage and plug it correctly into the device. If damaged, the power supply cable must be replaced. Switch on the unit at the main switch above the power supply cable. Switch on the pneumatic main switch D-24D206 in the front left underneath the unit. 3.1.2 Fuses Danger! Before checking the fuses, the unit must be disconnected from the power supply. Check the two fuses 6.3A D-244064 and replace in case of defect. 3.1.3 Converter Check the converter. Measuring at the upper two pins (black-white) of the converter: 22VDC, ±2VDC. Measuring at the lower two pins (black-white) of the converter: 32VDC, ±2VDC. If these values are not reached, the converter must be replaced. 3.2 Handpieces 3.2.1 Scaler 3.2.1.1 9 Scaler does not work Checking the wear of the used scaler tips using the "EMS Tip Card". If the wear is less than minimum, the tips must be replaced. Check the scaler tips for correct mounting. Fasten the tips with the EMS CombiTorque wrench to the maximum torque. Check that the pressure gauge on the back of the unit indicates a minimum pressure of 3.5 bar when using the handpiece. Otherwise, fully open the adjustment screw on the control block on the back of the unit. Check function with another handpiece. If the new handpiece works, replace old handpiece D-23D004. Connect another cable to the unit and check for function. If the scaler works with the new cable, replace cable D-23D008. Check the function of the electro-pneumatic valve D-24A009. If the valve does not switch when the scaler is started, it must be replaced. Check of the converter D-244062. Measuring at the lower two pins (black-white) of the converter: 32VDC, ±2VDC. If these values are not reached, the converter must be replaced. Replace the electric scaler board D-23D005. Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 3.2.1.2 3.2.1.3 3.2.1.4 Spray does not work Check the “Spray ON/OFF” switch on the foot pedal to be switched to “ON”. Check that the water control valve D-24D204 is set to maximum. Check the three-way syringe for outcoming water. If no water is coming out, proceed as described in chapter 3.2.4. If water is coming out, proceed according to the next test point. Check the quick couplings of the water hoses in the fresh water bottles for correct installation. Check the filters in the fresh water bottles for contamination. If the filters are dirty, replace the hoses D-24F073. Replace water-valve D-24D207. Replace check- valves D-24D205. Check the “Spray ON/OFF” switch on the foot pedal. If this no longer switches, it must be replaced. Spray does not stop Replace water-valve D-24D207. Regulation of scaler not possible Checking the wear of the used scaler tips using the "EMS Tip Card". If the wear is less than minimum, the tips must be replaced. Check the scaler tips for correct mounting. Fasten the tips with the EMS CombiTorque wrench to the maximum torque. Make sure the handpiece is properly connected to the cable. Make sure the cable is properly connected to the unit. Check function with another handpiece. If the new handpiece works, replace the old handpiece D-23D004. Connect another cable to the unit and check for function. If the scaler works with the new cable, replace cable D-23D008. Turn the potentiometer fully open and check the function. The vibration intensity must change. Replace potentiometer D-258300 if necessary. Replace the electric scaler board D-23D005. 3.2.2 Electric Motor 3.2.2.1 10 Electric motor does not work Check that the motor speed potentiometer is set to maximum. Make sure the handpiece is properly connected to the cable. Make sure the cable is properly connected to the unit. Check function without, or with another contra-angle handpiece. If the motor works, replace the contra-angle handpiece. Check that the pressure gauge on the back of the unit indicates a minimum pressure of 3.5 bar when using the handpiece. Otherwise, fully open the adjustment screw on the control block on the back of the unit. Check function with another handpiece. If the new handpiece works, replace the old handpiece D-231301. Connect another cable to the unit and check for function. If the motor works with the new cable, replace cable D-258201. Check of the converter D-244062. Measuring at the lower two pins (black-white) of the converter: 32VDC, ±2VDC. If these values are not reached, the converter must be replaced. Replace the electric board D-231303. Set the microswitches on the new board in the same positions as they were on the old board. Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 3.2.2.2 3.2.2.3 3.2.2.4 3.2.2.5 3.2.2.6 Spray does not work Check the “Spray ON/OFF” switch on the foot pedal to be switched to “ON”. Check that the water and the air control valves D-24D204 are set to maximum. Check the three-way syringe for outcoming water. If no water is coming out, proceed as described in chapter 3.2.4. If water is coming out, proceed according to the next test point. Check the quick couplings of the water hoses in the fresh water bottles for correct installation. Check the filters in the fresh water bottles for contamination. If the filters are dirty, replace the hoses D-24F073. Replace water-valve D-24D207. Replace check-valve D-24D205. Check the “Spray ON/OFF” switch on the foot pedal. If this no longer switches, it must be replaced. Spray does not stop Replace water-valve D-24D207. Motor switches on automatically when the scaler is used Membrane in the control block is leaking and must be replaced. Replace the membrane D-244501. Motor speed variations Check function without, or with another contra-angle handpiece. If the motor works correct, replace the contra-angle handpiece. If the speed of the electric motor is turned down at the speed potentiometer, but the motor increases the speed again automatically after some time, the control board D-231303 must be replaced. Display If a fault message appears on the display, the Bien-Air operating instructions must be consulted and followed up with the relevant procedure. 3.2.3 Turbine / Air-driven handpiece 3.2.3.1 3.2.3.2 11 Turbine does not work Make sure the handpiece is properly connected to the cable. Make sure the cable is properly connected to the unit. Check that the pressure gauge on the back of the unit indicates a minimum pressure of 3.5 bar when using the handpiece. Otherwise, fully open the adjustment screw on the control block on the back of the unit. Check the return air filter D-244B01 for free flow and replace if necessary. Check function with another handpiece. If the new handpiece works, replace the old handpiece. Connect another cable to the unit and check for function. If the turbine works with the new cable, replace cable D-24F001. Light does not work Check, if the lamp inside the coupling is intact. If necessary, replace the lamp. Attention: Installation possible in two directions, but function only on one direction. If the new lamp does not work, try the other direction. Connect another cable to the unit and check for function. If the light works with the new cable, replace cable D-24F001. Replace light module D-258100. Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 3.2.4 Three-way syringe 3.2.4.1 Three-way syringe does not work Check the quick couplings of the water hoses in the fresh water bottles for correct installation. Check the filters in the fresh water bottles for contamination. If the filters are dirty, replace the hoses D-24F073. Check the syringe tip D-24H001 for free flow and replace if necessary. Replace check-valves D-24D205. Check function of three-way syringe D-24H002 and replace if necessary. 3.3 Water system 3.3.1 Leakage of the water system Let the compressor run until it has reached the maximum pressure and switches off. Put each handpiece into operation one after the other using the water spray. While doing so, check all water-carrying parts of the handpiece for leaks. Replace leaking parts. 3.3.2 Spray water-bottle overflows Replace check-valves D-24D205. 3.4 Air system 12 Check the entire air system (pressure regulator, hoses, valves, foot pedal, foot pedal connections, three-way syringe connections, buttons on three-way syringe) for leaks with leak detection spray and replace the leaking parts. Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 Final Test In order to record the condition, the following work should be carried out after each maintenance and repair: 4.1 Case Visual check for intactness. Defective parts must be replaced. 4.2 Power supply cable Visual check for intactness. A defective power supply cable D-244063 must be replaced. 4.3 Water- and air system Inspection of the water system: Visual inspection of the entire water system for leaks. Search and repair of leakage according to chapter 3.3. Inspection of the air system: Visual inspection of the entire air system for leaks. Search and repair of leakage according to chapter 3.4. 4.4 Handpieces 13 Visual inspection of the integrity of all handpieces and their hoses. Checking the function of the electric motor. If the electric motor does not work, troubleshooting and repair must be carried out in accordance to chapter 3.2.2. Check for heat development on the motor. If the motor becomes hot during operation, it must be replaced. Checking the function of the scaler. If the scaler does not work, troubleshooting and repair must be carried out in accordance to chapter 3.2.1. Checking the wear of the used scaler tips using the "EMS Tip Card". If the wear is under minimum, the tips must be replaced. Checking the function of the air driven handpieces. If the turbine does not work, troubleshooting and repair must be carried out in accordance to chapter 3.2.3. Check the function of the three-way syringe. If the three-way syringe does not work, troubleshooting and repair must be carried out in accordance to chapter 3.2.4. Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 Parts list 5.1 DENTA-CART 404 – PICTURE A 14 Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 5.2 Surface Plate DENTA-CART 404 – PICTURE B 15 Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 5.3 Lower Plate 1 DENTA-CART 404 – PICTURE C 16 Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 5.4 Lower Plate 2 DENTA-CART 404 – PICTURE D 17 Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 18 D-23B003 D-23B007 EMS D-23B005 SATELEC D-23B006 SATELEC EMS 5.5 Accessories – PICTURE E Rev.1 – 04/2019 SERVICE MANUAL DENTA-CART 404 Service If problems occur with this product or if a fault occurs, contact your service point or the manufacturer at following contact address: Manufacturer service point: BPR Swiss GmbH Weststrasse 16 CH-3672 Oberdiessbach 0041 31 506 06 06 www.bpr-swiss.com info@bpr-swiss.com © BPR SWISS GMBH 2019 19 Rev.1 – 04/2019