Journal for the Measurement of Physical Behaviour, 2018, 1, 165-173 https://doi.org/10.1123/jmpb.2018-0022 © 2018 Human Kinetics, Inc. ORIGINAL RESEARCH The Accuracy of the Placement of Wearable Monitors to Classify Sedentary and Stationary Time Under Free-Living Conditions Alberto Flórez-Pregonero Pontificia Universidad Javeriana Matthew S. Buman and Barbara E. Ainsworth Arizona State University Background: Published accelerometer cut-points have limited accuracy in measuring sedentary (SED) and stationary time (STA) despite hip or wrist placement. Few studies have evaluated established cut-points to measure SED and STA in free-living settings. Methods: This study evaluated published uniaxial and triaxial cut-points of accelerometers and identified optimal cutpoints to measure SED and STA. Twenty participants, ages 18–65, wore three ActiGraph GT3X+ (one on each wrist and the waist) and two GENEActiv accelerometers (one on each wrist) for one weekday and one weekend day during simultaneous direct observation of movement. ActiGraph uniaxial cut-points (50, 100, 150, and 500 counts per minute [cpm]) and GENEActiv vector magnitude cut-points (VMCP; 217 and 386 cpm) were compared against the criterion measure of direct observation. As compared to the criterion, accuracy was determined with mean percent error, Bland-Altman plots, kappa coefficient, sensitivity, and specificity. Receiver operating characteristic curves identified cut-points with greatest discrimination to detect SED and STA. Results: For the GENEActiv, the 217 VMCP was most accurate for measuring SED and STA regardless of which arm wore the monitor. The ActiGraph was most accurate worn on the right hip using 100 and 150 uniaxial cpm to measure STA and 50 cpm to measure SED. Optimal ActiGraph VMCP cut-points to classify SED and STA were ActiGraph 2,000 cpm (left-wrist) and 63 cpm (right hip), respectively. Conclusion: Accuracy of ActiGraph uniaxial cut-points and GENEActiv VMCP is limited in assessing SED in free-living settings. Newer cut-points may increase the accuracy of measuring SED and STA from monitors in free-living settings. Keywords: accelerometers, activity classification, computational methods, sedentary behavior, validation Sedentary behaviors are a risk factor for several chronic diseases and mortality (Biswas et al., 2015; Dunstan, Howard, Healy, & Owen, 2012). Interrupting periods of sedentary behaviors with light-intensity physical activity may be beneficial for health (Chastin, Egerton, Leask, & Stamatakis, 2015). Furthermore, increasing standing time might attenuate the health risks of prolonged sitting (Katzmarzyk, 2014; van der Ploeg et al., 2014). Sedentary behaviors are defined as any waking behavior characterized by an energy expenditure ≤1.5 METs while in a sitting, reclining, or lying posture (e.g., watching TV); alternatively, stationary behaviors are any waking behavior done while lying, reclining, sitting, or standing, with no ambulation, irrespective of energy expenditure (e.g., standing) (Tremblay et al., 2017). Accelerometer-based wearable monitors are used in sedentary behavior and public health research as they can be used to measure total volume and breaks in sedentary time (Atkin et al., 2012). Among these monitors, the activPAL has shown high accuracy to measure sedentary time (Kozey-Keadle, Libertine, Lyden, Staudenmayer, & Freedson, 2011; Sellers, Dall, Grant, & Stansfield, 2016). This type of monitor attaches to the thigh using adhesive tape which may increase participant burden when Flórez-Pregonero is with the Departamento de Formación, Pontificia Universidad Javeriana, Bogota, Colombia. Buman and Ainsworth are with the College of Health Solutions, Arizona State University, Phoenix, AZ. Ainsworth (Barbara.ainsworth@ asu.edu) is corresponding author. compared to wrist or hip mounted monitors (Carson, Saunders, & Tremblay, 2016). Having to attach the activPAL to the thigh may limit its use for some individuals. Compliance in wearing monitors is improved by wearing a monitor on the wrist as opposed to other parts of the body (Freedson, & John, 2013). The ActiGraph and the GENEActiv are monitors that can be worn on the wrist. The wrist-worn ActiGraph (Koster et al., 2016) and GENEActiv (Pavey, Gomersall, Clark, & Brown, 2016) have been evaluated to measure sedentary time but not stationary time. Thus, it is important to confirm the accuracy of such monitors to measure sedentary and stationary behaviors. The most common method used to process accelerometerbased wearable monitors is to translate accelerometer output into measures of metabolic equivalents (METs) expenditure that reflect thresholds (cut-points) for specified levels of activity (Watson, Carlson, Carroll, & Fulton, 2014). Cut-points are numerical values for the acceleration of movement intensity (activity counts) that reflect thresholds in the energy cost of movement; higher numerical cut-points reflect higher energy costs and vice versa. Most cutpoints are derived from prediction equations in which accelerometer counts are regressed against energy expenditure values in kilocalories or in oxygen uptake values (Crouter, Clowers, & Bassett, 2006); other cut-points are proposed based upon researchers’ experiences (e.g., 100 or 150 counts per minute). Cut-points for sedentary behaviors use activity counts that reflect activities with an energy expenditure ≤1.5 METs and are commonly summarized using 60-second data-collection epochs. To date, several 165 Downloaded by on 02/04/19 166 Flórez-Pregonero, Buman, and Ainsworth uniaxial cut-points for classifying sedentary time with a hipmounted ActiGraph have been proposed including 50 (Crouter et al., 2006), 100 (Matthews et al., 2008), and 150 counts per minute (cpm) (Kozey-Keadle et al., 2011). Triaxial cut-points (i.e., vector magnitude cut-points [VMCP]) that classify sedentary time from a GENEActiv have been proposed including 217 and 386 cpm for left-wrist and right-wrist respectively (Esliger et al., 2011). Despite their extensive use, cut-points reflecting sedentary time have limitations as few cut-points have been tested in freeliving conditions (Healy et al., 2011; Judice, Santos, Hamilton, Sardinha, & Silva, 2015; Pavey et al., 2016). Cut-points-derived estimates of sedentary time may reflect little-to-no movement and/ or movements greater than 1.5 METs. Light intensity physical activities such as dusting or washing dishes with little movement may be classified as sedentary time (Kozey, Lyden, Howe, Staudenmayer, & Freedson, 2008). Thus, it is likely that existing uniaxial sedentary cut-points reflect stationary behaviors rather than exclusively sedentary behaviors. Accordingly, the cut-points can result in inaccurate estimates of sedentary time (Crouter et al., 2006; Lyden, 2012). Likewise, as monitor cut-points are specific to where they are worn, cut-points should be considered locationspecific; it is unlikely that a cut-point validated for the hip will lead to accurate classifications at the wrist. Research suggests that VMCP may be an accurate method to estimate time spent in sedentary behaviors in older adults in freeliving environments (Aguilar-Farias, Brown, & Peeters, 2014). However, researchers have not yet developed VMCP to assess sedentary and/or stationary behaviors with the ActiGraph in healthy adults. While, sedentary VMCP have been developed under laboratory settings for the GENEActiv, the cut-points have not been extensively tested in free-living settings (Esliger et al., 2011). The distinction of sedentary and stationary behaviors may have important implications in the relationships with health outcomes. As one’s cardiometabolic health may be enhanced differentially by interrupting prolonged sitting time with frequent brief bouts of light-intensity activity and/or standing, valid measurement of sedentary and stationary behaviors is important. The aim of this study is twofold: (1) to test the accuracy of published uniaxial and VMCP to classify sedentary and stationary time in free-living conditions in different body locations (left/right wrist and right hip) as compared to direct observation, and (2) to develop optimal VMCP to classify sedentary and stationary time based upon data collected under free-living conditions from the ActiGraph and GENEActiv monitors. Methods Participants A convenience sample of 20 adults enrolled in the study. Eligible participants were (a) 18–65 years of age; (b) normal-to-overweight body mass index (BMI) of 18.5 to 29.9 kg/m2; and (c) without disease or disability that could inhibit daily physical activity, assessed by completing the Physical Activity Readiness Questionnaire (PAR-Q; Shephard, 1988). Participants were recruited through e-mail announcements and fliers placed on the university campus. All enrolled participants provided informed consent prior to participation and the University Office of Research Integrity and Assurance approved the study protocol. One author (AFP) visited each participant three times. The first visit included a brief interviewer-administered questionnaire to collect demographics, an agreement on the times and places for visits two and three, and an explanation of the use of wearable monitors. Visits 2 and 3 consisted of data collection of free-living movement. Wearable Monitors The ActiGraph GT3X+ (ActiGraph LLC, Pensacola, FL, USA) and the GENEActiv (Activinsights, Kimbolton, Cambridgeshire, United Kingdom) wearable monitors were used in this study. The ActiGraph is a triaxial wearable monitor capable of recording accelerations in three axes (vertical, anteroposterior, and mediolateral). The ActiGraph GT3X+ measures accelerations ranging from 30 Hz up to 100 Hz in response to a magnitude range of ±6 g. The GENEActiv is a triaxial wearable monitor capable of recording accelerations in three axes; it measures accelerations ranging from 10 Hz up to 100 Hz in response to a magnitude range of ±8 g. Before field data collection, the investigators tested the monitors for inter-monitor reliability by placing them in the same body location (left arm) during five different activities performed in a 10-minute period. Intraclass correlation coefficients (ICC) were calculated to assess inter-monitor reliability in the vertical axis using 60-seconds data-collection epochs (ICC = 0.95 for ActiGraph, and 0.96 for GENEActiv wearable monitors). Before visits 2 and 3, study staff initialized the monitors at 100 Hz and downloaded the data to. csv files in 1-, 15-, and 60second epochs using ActiLife® software 6.11.5 and GENEActiv software 2.9 for ActiGraph and GENEActiv, respectively. During visits 2 and 3, participants wore the monitors simultaneously for 6 hours and were observed by trained staff members. One staff member placed the monitors on participants. Participants wore one ActiGraph GT3X+ on each wrist in the most proximal position using the manufacturer’s adjustable wristband; the monitors were oriented so the participant could read the brand logo. Participants wore another ActiGraph on the hip over the right anterior superior iliac spine mounted with an elastic belt and with the USB cap oriented towards the participant’s head. Participants wore a GENEActiv on each wrist in the most distal position and next to the ActiGraph with an orientation so the participants could read the serial number. Criterion Measure: Direct Observation The investigators used direct observation with focal sampling and duration coding to collect criterion data in real-time occurrence in free-living conditions. They coded activities into six mutually exclusive categories and/or three non-activities as described below. Three non-activity categories included times when the participants’ needed privacy, when activities were unobserved, and when there was an error in recording the data. • Private: Time when participants required private time (e.g., restrooms use). Direct observations resumed when the participant finished the private activity. • Unobserved: Time when participants were available for observations but out of sight of the observers (e.g., turning corners). • Error: Time when observers made an error in coding or were unable to determine an accurate code for an activity performed. • Walking: Walking for all locomotion purposes, walking in flat or inclined surfaces, and walking up or down stairs. Incidental or incomplete steps that did not result in moving JMPB Vol. 1, No. 4, 2018 Downloaded by on 02/04/19 Wearable Monitors Classification from one place to another were not included in this category (e.g., weight shifting). • Running: Continuous and short bouts of running or jogging (e.g., jogging for exercising or short runs to cross the street). • Sports and conditioning exercise: Playing sports or performing continuous or intermittent conditioning exercises. Exercises that differed from running or jogging were in this category (e.g., weight lifting, yoga, Pilates, or gym classes). • Household chores: Housekeeping activities such as dish washing, gardening, vacuuming, and doing the laundry. • Standing: Standing with or without upper body movements while bearing the body weight in one or both lower limbs. Incidental or incomplete steps that did not result in moving from one place to another were included in this category. • Sitting and lying down: Includes various body positions in which the participant’s buttocks, thighs, or back supported the body weight. This category included sitting in a chair or laboratory stool, reclining, and lying down in a supine and/ or prone position. Simultaneously, two investigators directly observed a participant’s movement in their free-living environment for six hours per day during two separate days; on a weekday (visit 2) and a weekend day (visit 3). Each observer independently recorded activities in an iPad tablet. Every time a participant changed the activity, observers made an annotation reflecting the new activity. Annotations were made with a commercially available app, Timestamped Field Notes app (TFNA) (Neukadye, 2017). TFNA allows configuration of colored buttons for predefined observation categories. TFNA stores annotations in an offline database that later can be downloaded as a. txt file containing the timestamp and the observations made at each time point. Tablets were time synchronized with the same computer that initialized and downloaded the monitor data. Investigators downloaded and exported data from the tablet into a text file (.txt) at the end of each direct observation period. The criterion variable for sedentary behaviors was a dichotomous variable calculated by including sitting and lying down for those observations in which both researchers had 100% agreement. A minute was considered sedentary when most of its seconds were sedentary (i.e., between 31–60 seconds per minute). The criterion variable for stationary behaviors was calculated with the same criteria except that standing activities were considered as stationary in addition to sitting and lying down postures. To address the second aim of this study, two dichotomous 15-second criterion variables (sedentary and stationary) and two 1-second criterion variables (sedentary and stationary) were computed. The 15-second criterion variable for sedentary behaviors was a dichotomous variable calculated by including sitting and lying down for those observations in which both researchers had 100% agreement. A 15-second epoch was considered sedentary when most of its seconds were sedentary (i.e., between 8–15 seconds per minute). The 15-second criterion variable for stationary behaviors was calculated with the same criteria noted above except that standing activities were considered as stationary in addition to sitting and lying down postures. The 1-second criterion variable for sedentary behaviors was a dichotomous variable including the seconds of sitting and lying down. The 1-second criterion variable for stationary activities was calculated with the same criteria as sedentary except that standing activities were considered as stationary in addition to sitting and lying down postures. 167 Observer Training Two observers completed 24 hours of one-to-one supervised training consisting of: • Two hours to become familiarized with the study protocols including thorough explanations of activity categories and use of tablets to record data. • Two hours of training in direct observation techniques designed not to disrupt, disturb, or modify the participant’s natural behavior to every extent possible. This included how the observer placed him- or herself in small spaces to observe movement but not disturb the participant. • Ten hours of direct observation practice using the tablets to record observations while watching a set of training videos made for this study with people doing various activities at work and in home environments. • Ten hours of direct observation practice using the tablets in real-time occurrence with people doing their own activities at work and home environments. After the training, researchers completed a testing session of direct observation of eight video clips with different activity scenarios. The total duration of the video clips was 40 minutes and referred to as the testing set. Two investigators (MSB and BEA) coded the videos independently; they discussed and modified the codes assigned as needed to assure full agreement. To be able to collect field data, the two observers were required to have high agreement with the testing set determined as an ICC > 0.80. Cut-points To achieve the first aim of the study, ActiGraph uniaxial cut-points of 50 (Crouter et al., 2006), 100 (Matthews et al., 2008), 150 (Kozey-Keadle et al., 2011), and 500 cpm (Silva, Aires, Santos, Vale, Welk, & Mota, 2011) were tested for the accuracy of SED and STA time at each wear location. For the GENEActiv, VMCP of 217 and 386 cpm were tested for accuracy of SED and STA time at each wear location (Esliger et al., 2011). To achieve the second study’s aim, analyses used receiver operating characteristic (ROC) curves to estimate the cut-points for the greatest accuracy of SED and STA time for the ActiGraph using 1-, 15-, and 60-second epochs and for the GENEActiv using 1- and 15-second epochs at each wear location. Statistical Analysis Descriptive statistics characterized the sample by sex, age, and BMI. The ICC identified agreement between the observers’ field observations. Several computations identified the accuracy (first aim) of the tested cut-points used to classify an activity as sedentary for each one of the aforementioned wearable monitors and locations. Percent error (PE) assessed the proportion of error for each of the tested cut-points relative to the sedentary criterion. PE was calculated as PE = [(Monitor Total Sedentary Minutes – Criterion Sedentary Minutes) / Criterion Sedentary Minutes] × 100. A positive PE indicated an overestimation of sedentary time whereas a negative PE indicated an underestimation of sedentary time. Kappa was used to observe agreement between each cut-point and the sedentary criterion value for classifying activities as sedentary while taking into account the agreement occurring by chance (Cohen, 1960). Landis and Koch published categories to interpret the kappa values as follows: 0–0.2 = slight agreement, JMPB Vol. 1, No. 4, 2018 Downloaded by on 02/04/19 168 Flórez-Pregonero, Buman, and Ainsworth 0.2–0.4 = fair agreement, 0.4–0.6 = moderate agreement, 0.6–0.8 = substantial agreement, and 0.8–1.0 = almost perfect agreement (Landis & Koch, 1977). Kappa was calculated on a minute-byminute basis. Sensitivity and specificity measure the accuracy of the tested cut-points to classify an activity as sedentary or non-sedentary. Sensitivity measured the accuracy of a cut-point to classify an observed sedentary activity as sedentary (true positives proportion). Sensitivity was calculated using the formula Sensitivity = True positives / (True positives + False negatives). A sensitivity value close to 1.0 shows that a cut-point correctly classifies a high proportion of observed sedentary activities as sedentary; a sensitivity value close to 0.0 indicates that the cut-point fails to classify observed sedentary activities as sedentary. Specificity measures the ability of a cut-point to classify observed non-sedentary activities as non-sedentary (true negatives proportion). Specificity was calculated using the formula Specificity = True negatives/(False positives + True negatives). A specificity value close to 1.0 shows that the cut-point correctly classify a high proportion of the observed non-sedentary activities as non-sedentary. A specificity value close to 0.0 indicates that the cut-point fails to classify observed non-sedentary activities as non-sedentary. The investigators also used PE, kappa, sensitivity, and specificity calculate the accuracy (first aim) of the tested cut-points to classify activities as stationary for each wearable monitor worn at different locations on the body as compared with the stationary behavior criterion. To develop VMCP (second aim), the investigators randomly divided the observations into training (50%) and testing (50%) datasets. ROC curves were created using the training sedentary and stationary criterion dataset. To determine the cut-points, the investigators used the minimum distance method which identifies the closest value to the optimal point at the upper-left corner of the ROC plot where Sensitivity = 1 and 1 − Specificity = 0. The area under the ROC curve (AUC), defined as an index of the accuracy of the ROC curve was calculated for each of the estimated cut-points (Hajian-Tilaki, 2013). An AUC of 1.0 indicates an estimated cut-point is perfect in the classification of activities. An AUC of 0.5 indicates an estimated cut-point is no better than chance in the classification activities. An AUC of 0.0 indicates an estimated cutpoint incorrectly classifies all activities. To test the accuracy of the estimated cut-points further, the investigators computed PE, simple kappa coefficient, sensitivity, and specificity in the testing dataset to compare the classification of sedentary and stationary behaviors obtained from the estimated cut-points with direct observation. ROC curve analyses were conducted using ROCPLOT macro for SAS (2005). The investigators used SAS version 9.4 to analyze the data. Results All 20 participants completed the study. Fifty percent of the participants were female. The mean age and BMI were 30.25 (±6.43) years and 22.7 (±3.1) kg/m2, respectively. All participants enrolled in the study were right-handed. Due to an error in the GENEActiv worn on the right wrist, 5.99 hours of movement for one participant was missing for this device and wrist placement. The observers directly recorded 241.32 hours of free-living data. The average length of free-living observation sessions was 5.97 ± 0.26 hours. Table 1 displays the distribution of direct observation classification categories and movement contexts. There was a substantial agreement between researchers’ direct observation codes (ICC = 0.76, 95% CI = 0.75–0.77). Cut-points Accuracy To assess the first study aim, investigators tested the accuracy of published uniaxial ActiGraph and vector magnitude GENEActiv cut-points to classify SED and STA as compared to direct observation of time spent in sedentary (sitting and lying down) and stationary (sitting, lying, and standing) behaviors in different body locations. Tables 2 and 3 show percent error, kappa coefficient, Table 1 Minutes (Standard Deviation [SD]) and Percent of the Observation Period Stratified by Activity Categories, Context, and Days of the Week Total observation time (minutes) Average observation time per participant (minutes) Activity categories Sitting/lying down Standing Other non-sedentary Unobserved Private time Context Sports/conditioning Household Transportation Occupation Leisure Non-agreement Weekdays Weekends Combined 7,180 359 7,314 365.7 14,494 362.35 Minutes (SD) 187.4 (102.1) 64.6 (68.4) 28.9 (27.5) 5.8 (6.8) 4.9 (6.0) Minutes (SD) 15.9 (38.4) 2.3 (8.8) 16.4 (27.1) 241.9 (115.1) 28.6 (42.5) Minutes (SD) 67.4 (68.7) % (SD) 52.3 (28.8) 18.1 (19.8) 8.0 (7.5) 1.6 (1.8) 1.3 (1.6) % (SD) 4.4 (10.5) 0.66 (2.5) 4.7 (7.6) 67.4 (31.9) 8.0 (12.0) % (SD) 18.5 (19.0) Minutes (SD) 140.5 (69.1) 79.0 (51.2) 23.8 (22.2) 3.6 (7.5) 8.0 (13.9) Minutes (SD) 19.8 (44.4) 2.9 (13.2) 21.2 (25.8) 88 (140.9) 148.8 (120.2) Minutes (SD) 110.9 (70.5) JMPB Vol. 1, No. 4, 2018 % (SD) 38.5 (19.0) 21.6 (13.7) 6.5 (6.0) 1.0 (2.0) 2.2 (3.8) % (SD) 5.2 (11.4) 0.8 (3.6) 5.8 (7.0) 23.9 (38.4) 41.1 (33.6) % (SD) 30.3 (19.4) Minutes (SD) 163.9 (89.3) 71.7 (60.1) 26.4 (24.8) 4.7 (7.2) 6.4 (10.7) Minutes (SD) 17.9 (41.0) 2.6 (11.1) 18.8 (26.1) 164.9 (148.9) 88.7 (107.9) Minutes (SD) 89.1 (72.2) % (SD) 45.4 (25.1) 19.8 (16.9) 7.26 (6.7) 1.3 (1.9) 1.8 (2.9) % (SD) 4.7 (10.9) 0.8 (3.1) 5.2 (7.3) 45.7 (41.3) 24.7 (30.0) % (SD) 24.5 (19.8) Wearable Monitors Classification 169 Table 2 Percent Error, Simple Kappa, Sensitivity, and Specificity for Published Sedentary Cut-Points as Compared to the Sedentary Criterion Monitor Location ActiGraph ActiGraph ActiGraph ActiGraph GENEActiv GENEActiv ActiGraph ActiGraph ActiGraph ActiGraph GENEActiv GENEActiv ActiGraph ActiGraph ActiGraph ActiGraph Left-wrist Left-wrist Left-wrist Left-wrist Left-wrist Left-wrist Right-wrist Right-wrist Right-wrist Right-wrist Right-wrista Right-wrista Right-hip Right hip Right hip Right hip Tested Cut-point Number of Axis Percent Error Kappa (95% CI) 50 100 150 500 217 386 50 100 150 500 217 386 50 100 150 500 1 1 1 1 3 3 1 1 1 1 3 3 1 1 1 1 −73.04 −66.66 −61.16 −22 −0.66 39.16 −72.05 −65.01 −58.96 −25.85 −15.73 28.38 18.37 35.53 45.54 72.71 0.06 (0.05 to 0.07) 0.08 (0.07 to 0.10) 0.10 (0.09 to 0.12) 0.24 (0.23 to 0.26) −0.29 (−0.31 to −0.28) −0.36 (−0.37 to −0.34) 0.08 (0.06 to 0.09) 0.10 (0.09 to 0.11) 0.13 (0.11 to 0.14) 0.26 (0.24 to 0.27) −0.26 (−0.28 to −0.25) −0.36 (−0.38 to −0.35) 0.27 (0.26 to 0.29) 0.29 (0.28 to 0.31) 0.30 (0.28 to 0.31) 0.25 (0.24 to 0.26) Sensitivity (95% CI) 0.15 0.19 0.23 0.48 0.61 0.82 0.17 0.21 0.25 0.47 0.53 0.78 0.69 0.78 0.83 0.93 (0.15 (0.18 (0.22 (0.47 (0.60 (0.81 (0.16 (0.20 (0.24 (0.46 (0.52 (0.77 (0.68 (0.77 (0.82 (0.92 to to to to to to to to to to to to to to to to 0.16) 0.20) 0.24) 0.50) 0.62) 0.83) 0.17) 0.22) 0.26) 0.49) 0.54) 0.79) 0.70) 0.79) 0.84) 0.94) Specificity (95% CI) 0.90 0.88 0.87 0.76 0.68 0.53 0.91 0.88 0.87 0.78 0.74 0.58 0.59 0.52 0.48 0.34 (0.90 (0.88 (0.86 (0.75 (0.67 (0.52 (0.90 (0.88 (0.86 (0.77 (0.73 (0.57 (0.58 (0.51 (0.47 (0.33 to to to to to to to to to to to to to to to to 0.91) 0.89) 0.88) 0.76) 0.69) 0.54) 0.91) 0.89) 0.88) 0.79) 0.75) 0.59) 0.60) 0.53) 0.49) 0.35) a Downloaded by on 02/04/19 Due to device malfunctioning there was 5.99 missing hours on this device; accordingly analyses include only 235.33 hours. Table 3 Percent Error, Simple Kappa, Sensitivity, and Specificity for Published Sedentary Cut-Points as Compared to the Stationary Criterion Kappa Monitor Location ActiGraph ActiGraph ActiGraph ActiGraph GENEActiv GENEActiv ActiGraph ActiGraph ActiGraph ActiGraph GENEActiv GENEActiv ActiGraph ActiGraph ActiGraph ActiGraph Left-wrist Left-wrist Left-wrist Left-wrist Left-wrist Left-wrist Right-wrist Right-wrist Right-wrist Right-wrist Right-wrista Right-wrista Right-hip Right-hip Right-hip Right-hip Tested Cut-point Number of Axis Percent Error (95% CI) 50 100 150 500 217 386 50 100 150 500 217 386 50 100 150 500 1 1 1 1 3 3 1 1 1 1 3 3 1 1 1 1 −81.4 −77 −73.15 −46.2 −31.35 −41.8 −80.74 −75.9 −71.61 −48.84 −3.96 −11.33 −18.26 −6.38 0.44 19.25 0.02 (0.01 to 0.03) 0.03 (0.02 to 0.04) 0.04 (0.03 to 0.05) 0.10 (0.09 to 0.12) −0.14 (−0.15 to −0.12) −0.11 (−0.12 to −0.09) 0.01 (0.00 to 0.01) 0.02 (0.01 to 0.03) 0.03 (0.02 to 0.04) 0.09 (0.08 to 0.10) −0.18 (−0.19 to −0.16) −0.16 (−0.17 to −0.14) 0.20 (0.18 to 0.21) 0.25 (0.23 to 0.26) 0.28 (0.26 to 0.29) 0.31 (0.30 to 0.33) Sensitivity Specificity (95% CI) 0.13 0.16 0.19 0.40 0.50 0.42 0.13 0.17 0.20 0.37 0.70 0.65 0.61 0.70 0.75 0.88 (0.12 (0.16 (0.19 (0.39 (0.49 (0.41 (0.12 (0.16 (0.19 (0.36 (0.70 (0.64 (0.60 (0.69 (0.75 (0.87 to to to to to to to to to to to to to to to to 0.14) 0.17) 0.20) 0.41) 0.51) 0.43) 0.14) 0.17) 0.21) 0.38) 0.71) 0.65) 0.62) 0.71) 0.76) 0.89) (95% CI) 0.89 0.87 0.86 0.73 0.65 0.69 0.88 0.86 0.84 0.73 0.51 0.54 0.60 0.55 0.52 0.40 (0.88 (0.87 (0.85 (0.72 (0.63 (0.68 (0.87 (0.85 (0.83 (0.72 (0.50 (0.52 (0.59 (0.54 (0.51 (0.39 to to to to to to to to to to to to to to to to 0.90) 0.88) 0.87) 0.74) 0.66) 0.70) 0.89) 0.87) 0.85) 0.75) 0.52) 0.55) 0.62) 0.57) 0.54) 0.42) a Due to device malfunctioning there was 5.99 missing hours on this device; accordingly analyses include only 235.33 hours. sensitivity, and specificity for the tested cut-points for the sedentary and stationary criteria, respectively. When compared to the sedentary criterion (sitting and lying down), none of the cut-points tested had outstanding accuracy (PE ranging from −73% to 73%; kappa <0.30; sensitivity <0.53; and specificity <0.91). Overall, the ActiGraph hip cut-points showed better accuracy than the wrist cut-points. The left-wrist cut-points tended to be less accurate than the right-wrist and right-hip cutpoints. Further, the left-wrist cut-points tended to have high negative percent error (except for the GENEActiv 217 and GENEActiv 386), slight agreement (except for ActiGraph 500), low-to-moderate sensitivity (except for GENEActiv 217 and GENEActiv 386), and high specificity (except for GENEActiv 217 and GENEActiv 386). With some exceptions, the right-wrist cut-points tended to have JMPB Vol. 1, No. 4, 2018 170 Flórez-Pregonero, Buman, and Ainsworth high negative percent error in excess of −25%, slight agreement with kappa’s <0.17 (except for ActiGraph 500), low-to-moderate sensitivity <0.53 (except for GENEActiv 386), and high specificity >0.74 (except for GENEActiv 386). The right-hip cut-points tended to have moderate positive percent error (except for ActiGraph 500), fair agreement, high sensitivity, and moderate specificity. When stationary behaviors (standing, sitting and lying down) were included in the criterion variable, the ActiGraph hip cutpoints were more accurate than the wrist cut-points. When compared to the wrist cut-points (PE −81% to −46%; kappa 0.01 to 0.10; sensitivity 0.13 to 0.40; and specificity 0.73 to 0.89), the ActiGraph hip cut-points had a lower percent error and higher values for kappa, sensitivity, and specificity (PE −18% to 19%; kappa 0.20 to 0.31; sensitivity 0.61 to 0.88; and specificity 0.40 to 0.60). The right-wrist GENEActiv cut-points (PE −11% to −4%; kappa −0.18 to −0.16; sensitivity 0.65 to 0.70; and specificity 0.51 to 0.54) were more accurate that the left-wrist cut-points. Among the cut-points tested, the ActiGraph 150 cpm right hip was the most accurate stationary uniaxial cut-point (PE 0.4%; kappa 0.28; sensitivity 0.75; and specificity 0.52). Regardless the location (left or right) or the criterion (sedentary, stationary) all of the kappa values for GENEactiv were negative., This is less agreement than would have been expected by chance. Downloaded by on 02/04/19 Developing Vector Magnitude Cut-points To address the second aim, the investigators estimated VMCP to classify SED based on the sedentary criterion (sitting + lying down) and STA based on the stationary criterion (standing + sitting + lying down). Estimated cut-points included a 60-seconds epoch for the ActiGraph worn on the left-wrist, right-wrist, and right hip. The investigators also estimated a 15-second and a 1-second epoch for the ActiGraph worn on the left-wrist, right-wrist, and right hip and for the GENEActiv worn on the left- and right-wrist. Tables 4 and 5 show values for AUC, PE, kappa, sensitivity, and specificity for the estimated SED and STA cut-points, respectively. The Online Supplementary Material presents all ROC graphs. For the vector magnitude cut-point (VMCP) estimated from the sedentary criterion (sitting + lying down), overall accuracy metrics were better for the 60-seconds epoch (AUC 0.702 to 0.729; PE 13% to 20%; kappa 0.33 to 0.37; sensitivity 0.69 to 0.75; and specificity 0.62 to 0.64) as compared to the 1-second and 15-seconds epochs (AUC 0.646 to 0.699; PE 3% to 61%; kappa −0.30 to 0.31; sensitivity 0.69 to 0.75; and specificity 0.40 to 0.63). Overall accuracy metrics also were better for wrist- worn monitor cut-points (AUC 0.647 to 0.723; PE 3% to 19%; kappa −0.30 to 0.35; sensitivity 0.62 to 0.71; and specificity 0.58 to 0.66) as compared to the ActiGraph hip worn monitor. The overall SED accuracy metrics were higher for all of the estimated VMCP (PE 3% to 61%; kappa −0.30 to 0.37; sensitivity 0.62 to 0.88; and specificity 0.40 to 0.66) as compared to the published uniaxial cut-points (PE −73% to 73%; kappa −0.36 to 0.30; sensitivity 0.15 to 0.93; and specificity 0.34 to 0.91). Among the estimated VMCP, the ActiGraph 2000 cpm worn on the left wrist was the most accurate SED cut-point. The Supplementary Material [available online] presents all of the AUC plots. The VMPC estimated from the stationary criterion (standing + sitting + lying down) accuracy metrics were similar across the different time epoch lengths ranging from 1 minute to 1 second. The overall accuracy metrics were best for the estimated VMCP (PE −20% to 12%; kappa −0.16 to 0.25; sensitivity 0.59 to 0.81; and specificity 0.42 to 0.56) as compared to the uniaxial cut-points (PE −81% to 19%; kappa −0.18 to 0.31; sensitivity 0.13 to 0.88; and specificity 0.40 to 0.89). Among the estimated VMCP, the ActiGraph 63 cpm worn on the right hip was the most the accurate STA cut-point (AUC = 0.638). The Online Supplementary Material presents all of the AUC plots. Discussion This study had two aims: (1) to test the accuracy of published cutpoints to classify sedentary and stationary time in free-living conditions and (2) to develop VMCP to classify sedentary and stationary time. Results showed four observations. First, there was Table 4 Percent Error, Kappa, Sensitivity and Specificity for Estimated Vector Magnitude Cut-Points Sedentary Criterion Kappa Monitor Location ActiGraph ActiGraph ActiGraph GENEActiv GENEActiv ActiGraph ActiGraph ActiGraph GENEActiv GENEActiv ActiGraph ActiGraph ActiGraph Left-wrist Left-wrist Left-wrist Left-wrist Left-wrist Right-wrist Right-wrist Right-wrist Right-wrista Right-wrista Right-hip Right-hip Right-hip VMCP Epoch Length (sec) 2000 455 5 65 2 2358 495 8 61 3 249 15 0 60 15 1 15 1 60 15 1 15 1 60 15 1 AUC Percent Error (95% CI) 0.702 0.672 0.647 0.685 0.664 0.723 0.689 0.666 0.686 0.661 0.729 0.699 0.646 12.98 16.17 16.39 19.25 14.41 13.86 11.88 11.77 3.08 17.6 19.8 16.72 61.05 0.33 (0.30 to 0.35) −0.27 (−0.28 to −0.26) 0.26 (0.26 to 0.26) −0.29 (−0.30 to −0.28) −0.25 (−0.25 to −0.25) 0.35 (0.33 to 0.37) −0.30 (−0.31 to −0.29) 0.29 (0.28 to 0.29) −0.28 (−0.29 to −0.27) −0.25 (−0.26 to −0.25) 0.37 (0.35 to 0.39) 0.31 (0.30 to 0.32) 0.27 (0.26 to 0.27) Sensitivity Specificity (95% CI) 0.69 0.67 0.67 0.70 0.65 0.71 0.67 0.66 0.62 0.67 0.75 0.70 0.88 (0.68 (0.67 (0.67 (0.69 (0.65 (0.70 (0.66 (0.66 (0.61 (0.67 (0.74 (0.69 (0.88 to to to to to to to to to to to to to 0.71) 0.68) 0.67) 0.71) 0.66) 0.73) 0.68) 0.67) 0.63) 0.67) 0.77) 0.71) 0.88) (95% CI) 0.63 0.60 0.59 0.59 0.60 0.64 0.63 0.63 0.66 0.58 0.62 0.62 0.40 (0.62 (0.59 (0.59 (0.59 (0.59 (0.63 (0.62 (0.62 (0.66 (0.58 (0.61 (0.61 (0.40 to to to to to to to to to to to to to 0.65) 0.61) 0.60) 0.60) 0.60) 0.66) 0.64) 0.63) 0.67) 0.59) 0.64) 0.62) 0.40) a Due to device malfunctioning there was 5.99 missing hours on this device; accordingly analyses include only 235.33 hours. VMCP = Vector Magnitude Cut-point; AUC = Area Under the Curve. JMPB Vol. 1, No. 4, 2018 Wearable Monitors Classification 171 Table 5 Percent Error, Kappa, Sensitivity and Specificity for Estimated Vector Magnitude Cut-Points Stationary Criterion Kappa Monitor Location ActiGraph ActiGraph ActiGraph GENEActiv GENEActiv ActiGraph ActiGraph ActiGraph GENEActiv GENEActiv ActiGraph ActiGraph ActiGraph Left-wrist Left-wrist Left-wrist Left-wrist Left-wrist Right-wrist Right-wrist Right-wrist Right-wrista Right-wrista Right-hip Right-hip Right-hip VMCP Epoch Length (sec) 2365 523 6 77 3 2411 630 18 91 4 423 63 0 60 15 1 15 1 60 15 1 15 1 60 15 1 AUC Percent Error (95% CI) 0.611 0.603 0.600 0.620 0.613 0.601 0.598 0.599 0.602 0.597 0.645 0.638 0.626 −13.31 −14.85 −18.37 −10.34 −11 −20.13 −15.51 −16.39 −10.89 −11.44 −5.5 −2.2 11.77 0.19 (0.17 to 0.21) −0.15 (−0.16 to −0.14) 0.17 (0.17 to 0.18) −0.16 (−0.17 to −0.15) −0.15 (−0.15 to −0.15) 0.17 (0.15 to 0.19) −0.15 (−0.16 to −0.14) 0.17 (0.17 to 0.17) −0.15 (−0.16 to −0.14) −0.14 (−0.14 to −0.14) 0.25 (0.22 to 0.27) 0.22 (0.21 to 0.23) 0.24 (0.24 to 0.25) Sensitivity Specificity (95% CI) 0.64 0.62 0.60 0.65 0.64 0.59 0.61 0.61 0.64 0.63 0.71 0.72 0.81 (0.63 (0.61 (0.59 (0.64 (0.64 (0.57 (0.60 (0.61 (0.64 (0.63 (0.70 (0.71 (0.81 to to to to to to to to to to to to to 0.65) 0.62) 0.60) 0.66) 0.64) 0.60) 0.62) 0.61) 0.65) 0.63) 0.72) 0.72) 0.81) (95% CI) 0.56 0.56 0.59 0.54 0.53 0.59 0.56 0.58 0.53 0.53 0.54 0.51 0.42 (0.54 (0.55 (0.59 (0.53 (0.53 (0.58 (0.55 (0.57 (0.53 (0.52 (0.52 (0.50 (0.42 to to to to to to to to to to to to to 0.58) 0.57) 0.59) 0.55) 0.54) 0.61) 0.57) 0.58) 0.54) 0.53) 0.56) 0.52) 0.43) Downloaded by on 02/04/19 a Due to device malfunctioning there was 5.99 missing hours on this device; accordingly analyses include only 235.33 hours. AUC = Area Under the Curve; VMCP = Vector Magnitude Cut-point. an overall lack of accuracy for the published sedentary uniaxial cutpoints regardless of the accelerometer location and cut-points used. Second, the ActiGraph 150 cpm worn on the right hip demonstrated moderate accuracy to differentiate STA (PE 0.44%; kappa 0.28; sensitivity 0.75; and specificity 0.52) but not SED (PE 46%; kappa 0.30; sensitivity 0.83; and specificity 0.48). Third, established ActiGraph uniaxial cut-points used to measure SED were more accurate when the monitor was worn on the right hip had than when the monitor worn on the left- and right wrist. Fourth, the estimated VMCP had the highest accuracy in measuring SED and STA regardless of the monitor location and cut-point used. Overall, accuracy of established uniaxial cut-points to classify SED and STA was low regardless of the location worn and the cutpoint used. The low accuracy supports the ascertainment that there is not an ideal uniaxial cut-point to measure SED. Interestingly; uniaxial cut-points for the right wrist were more accurate than the left wrist. This might be an effect of handedness; however, we could not test this hypothesis, as all of the participants were righthanded. Future studies that test handedness on the accuracy of a wrist mounted wearable monitors should recruit participants with right- and left-handed dominance. The ActiGraph 100 cpm and ActiGraph 150 cpm uniaxial cutpoints for the monitors worn on the right hip were most accurate to assess STA (standing, sitting, and lying down) but not SED (sitting and lying down). As wearable monitors measure body movements using changes in acceleration that are used to estimate the intensity of physical activities over time (Chen, & Bassett, 2015), these findings are not surprising and suggest caution interpreting wearable monitors-derived measures of SED and its associations to health-related outcomes. In 2011, Kozey-Keadle et al. (2011) reported that the ActiGraph 100 cpm and ActiGraph 150 cpm cut-points for monitors worn on the right hip had a similar error magnitude in measuring SED. We observed the same cut-points to be best for measuring STA, but not SED. It is possible that the metrics used and methodological differences between KozeyKeadle et al and the current study explain some differences in the results. For example, Kozey-Keadle et al used the low-frequency extension for the ActiGraph; we did not apply additional filters to the wearable monitors’ signal. It should be noted that the use of the low-frequency extension in ActiGraph monitors may overestimate the number of steps taken throughout the day (Feito, Garner, & Bassett, 2015), which may introduce error in the measurement of sedentary and stationary behaviors as some incidental movements without ambulation may be misclassified as light intensity physical activity when they are stationary behaviors. Kozey-Keadle et al derived their criterion measure from observations of a single researcher whereas the in the current study, two observers coded the free-living data. Regardless of the differences in the methods and observations of Kozey-Keadle et al.’s and the current study, the results contribute to the ongoing debate about what is most accurate uniaxial cut-point to use in classifying SED from the ActiGraph and whether the cut-points approach is more reflective of stationary types of behaviors versus sedentary behaviors. To increase consistency of results, we suggest future studies compare the accuracy of existing cut-points to measure SED and STA under free-living conditions. All the established uniaxial cut-points for the ActiGraph placed on the hip showed better accuracy for SED than those for wrist locations regardless the cut-point used. This finding may be related the participants’ arms movements during sedentary activities (e.g., typing) that increased false negative classification of SED. Investigators developed the ActiGraph uniaxial cut-points with the monitor worn on the waist (Matthews et al., 2008); hence, it is likely the cut-points are specific to hip displacement as opposed to arm movement. In support of this latter explanation, the wristmounted GENEActiv VMPC cut-points developed for the wristworn monitor had a lower PE than the wrist-mounted ActiGraph uniaxial cut-points developed for a waist-worn monitor (Esliger et al., 2011). The poor accuracy of the wrist-mounted ActiGraph wearable monitors to measure SED is an issue that may have impact in physical activity and sedentary behaviors surveillance and requires further research. For example, the U.S. NHANES is using data from wrist-mounted ActiGraph wearable monitors to estimate sedentary time at the population level (Fulton et al., 2016). JMPB Vol. 1, No. 4, 2018 Downloaded by on 02/04/19 172 Flórez-Pregonero, Buman, and Ainsworth As compared to uniaxial cut-points, the ActiGraph estimated VMCP improved the accuracy of measuring SED and STA considerably by reducing the overall PE and increasing kappa, sensitivity, and specificity values. Among the estimated VMCP, the ActiGraph 2000 cpm left-wrist was most accurate in measuring SED. On the other hand, the ActiGraph 63 cpm worn on the right hip was most accurate in measuring STA. Having cut-points that accurately differentiate standing, sitting and lying down from other types of activities may be of interest for some researchers depending on the goal of their research. Despite the limited accuracy of using cutpoints to assess SED, the cut-point approach remains the method of choice for many researchers and practitioners due to its simplicity and relatively low cost. Moreover, until machine learning-based approaches that require computational skills and resources for the researchers than current methods, are easily accessible to researchers and practitioners to score wearable monitors data; investigators should use the most accurate cut-points available. When selecting a monitor to assess sedentary or stationary time, researchers should consider several aspects to compute a summary score. The cut-points approach is the most common method used to assess sedentary behaviors and is easy to use and understand. However, its use encompasses some limitations that need to be addressed. The most common cut-point, 100 cpm, was not empirically derived (Atkin et al., 2012); however, there are several cut-points to measure sedentary behaviors with 150 cpm considered to be the most accurate (Kozey-Keadle et al., 2011). The 100 cpm cut-point was proposed for use with data from accelerometers worn at the right hip, however, it is being used in large scale studies while worn on the wrist. This approach, although proposed as a way to increase monitors wear compliance, may be inaccurate to detect sedentary behaviors as one can mimic movement with their arms during sedentary behaviors (e.g., seated while typing) (Freedson, & John, 2013). Accordingly, the most accurate, location-specific cut-points should be used when measuring sedentary behaviors. Monitors such as the activPAL worn on the thigh have shown high accuracy to measure sedentary and stationary time. Accordingly, the activPAL is the optimal device if the primary goal of a research study is to estimate sedentary and/or stationary time. However, if the location of the activPAL as attached to the leg using adhesive tape increases participant burden, its use may be limited. Therefore, researchers should consider other monitors or approaches to measure sedentary or stationary time (Calabró, Lee, Saint-Maurice, Yoo, & Welk, 2014). Machine learning techniques offer promising approaches to estimating sedentary or stationary behaviors with increased precision (Dutta, Ma, Buman, & Bliss, 2016; Ellis, Kerr, Suneeta, Staudenmayer, & Lanckriet, 2016; Lyden, Kozey-Keadle, Staudenmayer, & Freedson 2014; Staudenmayer, He, Hickey, Sasaki, & Freedson, 2015; Zhang, Rowlands, Murray, & Hurst, 2012). Algorithms have been developed and validated; however, such methods are complex and not practical for most researchers and practitioners to measure sedentary behaviors at this time. A strength of the current study is the observers’ intensive training that resulted in a substantial agreement between their field observations. This agreement yielded a valid criterion with high agreement between observers. In addition, the participants’ movements were observed in free-living settings for two days (weekday and weekend day) allowing observers to capture a broad range of observations in different contexts. An important limitation of the current study is was the lack of a criterion measure for energy expenditure, such as indirect calorimetry to classify SED as ≤1.5 METs that could have led to erroneous activity classifications. In addition, when observers agreed on an observation coding it is not possible to know whether each observer coded incorrectly, which could be a source of error in the criterion. The study sample included healthy right-handed young adults; this limits generalization of the results to other populations (e.g., left-handed and/or older adults). Last, we lost data due to a monitor error resulting in a GENEActiv right-wrist recording only 235.33 hours of movement from all study participants as compared to the other monitors that recorded 241.32 hours. To assure good external validity, additional studies should evaluate the cut-points from this study in different population groups. Conclusion This study showed that the ActiGraph uniaxial cut-points of 50, 100, 150, and 500 cpm and the GENEActiv VMCP of 217 and 386 cpm had limited accuracy to assess SED in free-living settings (PE ranging from −73% to 73%; kappa 0.01 to 0.10; sensitivity 0.13 to 0.40; and specificity 0.73 to 0.89). The ActiGraph 100 cpm and the ActiGraph 150 cpm uniaxial cut-points worn on the right hip demonstrated the best accuracy in differentiating STA (standing, sitting, and lying down; PE 0.4%; kappa 0.28; sensitivity 0.75; and specificity 0.52) but not SED (sitting and lying down). The estimated 60-seconds epoch ActiGraph 2000 cpm VMCP with the monitor worn on the left wrist and the 15-seconds epoch ActiGraph 63 cpm VMCP with the monitor worn on the right hip were most accurate in classifying SED and STA in free-living settings, respectively. References Aguilar-Farias, N., Brown, W.J., & Peeters, G.M. (2014). ActiGraph GT3X+ cut-points for identifying sedentary behaviour in older adults in free-living environments. Journal of Science and Medicine in Sport, 17(3), 293–299. PubMed ID: 23932934 doi:10.1016/j.jsams. 2013.07.002 Atkin, A.J., Gorely, T., Clemes, S.A., Yates, T., Edwardson, C., Brage, S., . . . Biddle, S.J. (2012). Methods of measurement in epidemiology: Sedentary behaviour. International Journal of Epidemiology, 41(5), 1460–1471. doi:10.1093/ije/dys118 Biswas, A., Oh, P.I., Faulkner, G.E., Bajaj, R.R., Silver, M.A., Mitchell, M.S., & Alter, D.A. (2015). Sedentary time and its association with risk for disease incidence, mortality, and hospitalization in adults: A systematic review and meta-analysis. Annals of Internal Medicine, 162(2), 123–132. PubMed ID: 25599350 doi:10.7326/M14-1651 Calabró, M., Lee, J.M., Saint-Maurice, P.F., Yoo, H., & Welk, G.J. (2014). Validity of physical activity monitors for assessing lower intensity activity in adults. International Journal of Behavioral Nutrition and Physical Activity, 11(119), 1–9. doi:10.1186/s12966-014-0119-7 Carson, V., Saunders, T., & Tremblay, M. (2016). Can epidemiologist learn more from sedentary behaviour than from the measurement of physical activity? In C. Tudor-Locke & R.J. Shephard (Ed.), The objective monitoring of physical activity: Contributions of accelerometry to epidemiology, exercise science and rehabilitation (Vol. 1, pp. XII, 383). Switzerland: Springer International Publishing. Chastin, S., Egerton, T., Leask, C., & Stamatakis, E. (2015). Meta-analysis of the relationship between breaks in sedentary behavior and cardiometabolic health. Obesity, 23(9), 1800–1810. PubMed ID: 26308477 doi:10.1002/oby.21180 Chen, K., & Bassett, D.R., Jr. (2015). The technology of accelerometrybased activity monitors: Current and future. Medicine & Science in Sports & Exercise, 37(Suppl. 11), S490–S500. JMPB Vol. 1, No. 4, 2018 Downloaded by on 02/04/19 Wearable Monitors Classification Cohen, J. (1960). A coefficient of agreement for nominal scales. Educatonal and Psychological Measurement, 20(1), 37–46. doi:10.1177/ 001316446002000104 Crouter, S., Clowers, K.G., & Bassett, D.R., Jr. (2006). A novel method for using accelerometer data to predict energy expenditure. Journal of Applied Physiology, 100(4), 1324–1331. PubMed ID: 16322367 doi:10.1152/japplphysiol.00818.2005 Dunstan, D., Howard, B., Healy, G.N., & Owen, N. (2012). Too much sitting–a health hazard. Diabetes Research and Clinical Practice, 97(3), 368–376. PubMed ID: 22682948 doi:10.1016/j.diabres.2012.05.020 Dutta, A., Ma, O., Buman, M.P., & Bliss, D.W. (2016). Learning approach for classification of GENEActiv accelerometer data for unique activity identification. Paper presented at theIEEE 13th International Conference on Wearable and Implantable Body Sensor Networks (BSN). San Francisco, CA, 359–364. doi:10.1109/BSN. 2016.7516288 Ellis, K., Kerr, J., Suneeta, G., Staudenmayer, J., & Lanckriet, G. (2016). Hip and wrist accelerometer algorithms for free-living behavior classification. Medicine & Science in Sports & Exercise, 48(5), 933–940. PubMed ID: 26673126 doi:10.1249/MSS.0000000000000840 Esliger, D., Rowlands, A.V., Hurst, T.L., Catt, M., Murray, P., & Eston, R.G. (2011). Validation of the GENEA accelerometer. Medicine & Science in Sports & Exercise, 43(6), 1085–1093. PubMed ID: 21088628 doi:10.1249/MSS.0b013e31820513be Feito, Y., Garner, H.R., & Bassett, D.R. (2015). Evaluation of ActiGraph’s low-frequency filter in laboratory and free-living environments. Medicine & Science in Sports & Exercise, 47(1), 211–217. PubMed ID: 24870583 doi:10.1249/MSS.0000000000000395 Freedson, P., & John, D. (2013). Comment on “estimating activity and sedentary behavior from an accelerometer on the hip and wrist”. Medicine & Science in Sports & Exercise, 45(5), 962–963. PubMed ID: 23594509 doi:10.1249/MSS.0b013e31827f024d Fulton, J., Carlson, S.A., Ainsworth, B.E., Berrigan, D., Carlson, C., Dorn, J.M., . . . Wendel, A. (2016). Strategic priorities for physical activity surveillance in the United States. Medicine & Science in Sports & Exercise, 48(10), 2057–2069. PubMed ID: 27187094 doi:10.1249/ MSS.0000000000000989 Hajian-Tilaki, K. (2013). Receiver Operating Characteristic (ROC) cureve analysis for medical diagnostic test evaluation. Caspian Journal of Internal Medicine, 4(2), 627–635. PubMed ID: 24009950 Healy, G.N., Clark, B.K., Winkler, E.A., Gardiner, P.A., Brown, W.J., & Matthews, C.E. (2011). Measurement of adults’ sedentary time in population-based studies. American Journal of Preventive Medicine, 41(2), 216–227. PubMed ID: 21767730 doi:10.1016/j.amepre.2011. 05.005 Judice, P., Santos, D.A., Hamilton, M.T., Sardinha, L.B., & Silva, A.M. (2015). Validity of GT3X and Actiheart to estimate sedentary time and breaks using ActivPAL as the reference in free-living conditions. Gait and Posture, 41(4), 917–922. PubMed ID: 25852024 doi:10. 1016/j.gaitpost.2015.03.326 Katzmarzyk, P. (2014). Standing and mortality in a prospective cohort of Canadian adults. Medicine & Science in Sports & Exercise, 46(5), 940– 946. PubMed ID: 24152707 doi:10.1249/MSS.0000000000000198 Koster, A., Shiroma, E.J., Caserotti, P., Metthews, C.E., Chen, K.Y., & Glynn, N.W. (2016). Comparison of Sedentary Estimates between activPAL and Hip- and Wrist-Worn ActiGraph. Medicine & Science in Sports & Exercise, 48(8), 1514–1522. PubMed ID: 27031744 doi:10.1249/MSS.0000000000000924 Kozey, S., Lyden, K., Howe, C.A., Staudenmayer, J.W., & Freedson, P.S. (2008). Accelerometer output and MET values of common physical activities. Medicine & Science in Sports & Exercise, 42(9), 1776– 1784. doi:10.1249/MSS.0b013e3181d479f2 173 Kozey-Keadle, S., Libertine, A., Lyden, K., Staudenmayer, J., & Freedson, P.S. (2011). Validation of wearable monitors for assessing sedentary behavior. Medicine & Science in Sports & Exercise, 43(8), 1561– 1567. PubMed ID: 21233777 doi:10.1249/MSS.0b013e31820ce174 Landis, J., & Koch, GG. (1977). The measurement of observer agreement for categorical data. Biometrics, 33(1), 159–174. PubMed ID: 843571 doi:10.2307/2529310 Lyden, K., Kozey-Keadle, S., Staudenmayer, J., & Freedson, P.S. (2014). A method to estimate free-living active and sedentary behavior from an accelerometer. Medicine & Science in Sports & Exercise, 46(2), 386–397. doi:10.1249/MSS.0b013e3182a42a2d Lyden, K., Kozey-Keadle, S.L., Staudenmayer, J.W., & Freedson, P.S. (2012). Validity of two wearable monitors to estimate breaks from sedentary time. Medicine & Science in Sports & Exercise, 44(11), 2243–2252. Matthews, C., Chen, K.Y., Freedson, P.S., Buchowski, M.S., Beech, B.M., Pate, R.R., & Troiano, R.P. (2008). Amount of time spent in sedentary behaviors in the United States, 2003–2004. American Journal of Epidemiology, 167(7), 875–881. PubMed ID: 18303006 doi:10.1093/ aje/kwm390 Neukadye. (2017). Timestamped Field Notes. Retrieved from https:// www.neukadye.com/mobile-applications/timestamped-field-notes/ Pavey, T., Gomersall, S.R., Clark, B.K., & Brown, W.J. (2016). The validity of the GENEActiv wrist-worn accelerometer for measuring adult sedentary time in free living. Journal of Science and Medicine in Sport, 19(5), 395–399. PubMed ID: 25956687 doi:10.1016/j.jsams. 2015.04.007 SAS. (2005). Plot ROC curve with cutpoint labeling and optimal cutpoint analysis. Retrieved from http://support.sas.com/kb/25/018.html Sellers, C., Dall, P., Grant, M., & Stansfield, B. (2016). Agreement of the activPAL3 and activPAL for characterising posture and stepping in adults and children. Gait and Posture, 48, 209–214. PubMed ID: 27318305 doi:10.1016/j.gaitpost.2016.05.012 Shephard, R. (1988). PAR-Q Canadian home fitness test and exercise alternatives. Sports Medicine, 5(3), 185–195. PubMed ID: 3368685 doi:10.2165/00007256-198805030-00005 Silva, P., Aires, L., Santos, R.M., Vale, S., Welk, G., & Mota, J. (2011). Lifespan snapshot of physical activity assessed by accelerometry in Porto. Journal of Physical Activity and Health, 8(3), 352–360. PubMed ID: 21487134 doi:10.1123/jpah.8.3.352 Staudenmayer, J, He, S., Hickey, A., Sasaki, J., & Freedson, P. (2015). Methods to estimate aspects of physical activity and sedentary behavior from high-frequency wrist accelerometer measurements. Journal of Applied Physiology, 119(4), 396–403. PubMed ID: 26112238 doi:10.1152/japplphysiol.00026.2015 Tremblay, M., Aubert, S., Barnes, J.D., Saunders, J., Carson, V., LatimerCheung, A.E., . . . Chinapaw, M.J.M. (2017). Sedentary Behavior Research Network (SBRN) - Terminology Consensus Project process and outcome. International Journal of Behavioral Nutrition and Physical Activity, 17(75), 1–17. doi:10.1186/s12966-017-0525-8 van der Ploeg, H., Chey, T., Ding, D., Chau, J.Y., Stamatakis, E., & Bauman, A.E. (2014). Standing time and all-cause mortality in a large cohort of Australian adults. Preventive Medicine, 69, 187–191. PubMed ID: 25456805 doi:10.1016/j.ypmed.2014.10.004 Watson, K., Carlson, S.A., Carroll, D.D., & Fulton, J.E. (2014). Comparison of accelerometer cut points to estimate physical activity in US adults. Journal of Sports Science, 32(7), 660–669. doi:10.1080/ 02640414.2013.847278 Zhang, S., Rowlands, A.V., Murray, P., & Hurst, T.L. (2012). Physical Activity Classification using the GENEA wrist-worn accelerometer. Medicine & Science in Sports & Exercise, 44(4), 742–748. doi:10. 1249/MSS.0b013e31823bf95c JMPB Vol. 1, No. 4, 2018