Medical Gases Systems Presented By: Mohamed Salah THE FIFTH LECTURE. 2005628 03/05/12 Page 1 of 9 Issue 3 mAIR Medical Air, cAIR Combined Air & sAIR Surgical Air Systems – EN ISO 7396-1/HTM 02-01 and HTM2022 EurPh 400V 50Hz, 4 Bar, 7 Bar & 11 Bar Outlet mAIR MEDICAL AIR SYSTEMS In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. SPECIFICATION Air Plant System Compressors Surgical air systems shall have a ‘duplexed’ dryer system and a simplex compressor. Additional compressors shall be available to fully ‘duplex’ the surgical air system, such that any single functional component failure will not affect the integrity of the air supply. Variable Speed Drive - Optional The Medical Air system shall conform to EN ISO 7396-1 and NHS Health Technical Memorandum HTM02-01. Medical quality air to the European Pharmacopoeia monograph shall be delivered at pressures of 400 kPa (4 bar), 700kPa (7 bar) or 1100 kPa (11 bar) gauge for supply of the hospital medical or surgical air systems. The entire system shall be ‘duplexed’ such that any single functional component failure will not affect the integrity of the medical compressed air supply. Sources Of Supply - HTM02-01/EN ISO 7396-1 Triplex or quadruplex compressor configurations will produce the primary supply with two compressors in standby. Each compressor will be capable of supplying the specified volumetric flow for duplex and triplex plant, and half flow for quadruplex. For duplex plant the secondary supply shall be from an automatic manifold capable of supplying at average demand for 4 hours. For triplex and quadruplex plant the standby compressors will form the secondary supply. Sources Of Supply - HTM2022 EurPh Duplex, triplex or quadruplex compressor configurations will produce the primary supply with one compressor on standby. A secondary supply shall be from an automatic manifold capable of supplying at average demand for 4 hours. Control System The central control system shall provide an intelligent human machine interface incorporating on board flash memory and real-time clock for recording operational parameters in the in built event log. The central control system shall operate at low voltage and include BMS connection for plant fault, plant emergency, reserve fault and pressure fault. Visualisation of plant inputs, outputs and status through a web browser, using a simple Ethernet connection shall be available. The central control unit shall incorporate a user friendly 5.7” high-definition colour display with clear pictograms and LED indicators, providing easy access to system operational information. A mechanical back-up facility shall ensure continued operation in the event of a control system malfunction. The control system shall normally employ automatic rotation of the lead compressor to maximise life and ensure even wear. Optional Control Equipment An advanced monitoring system shall be available to give immediate access to valuable information such as system status, trends, historical data and system performance. Data collected from all pumps shall be made available in real-time visualisation pages and shall be accessed through the hospital’s LAN, such that total data security is assured. The Airconnect™ monitoring system shall also include :- Compressors shall be Atlas Copco GA-MED oil injected rotary screw compressors suitable for both continuous and frequent start/stop operation at a nominal outlet pressure of 750 kPa (7.5 bar), 950 kPa (9.5 bar) or 13000 kPa (13 bar) gauge. Compressors shall be supplied with a block and fin style after cooler with a dedicated quiet running fan to maximise cooling and efficiency. A multistage oil separator capable of achieving 2ppm oil carry over shall be fitted to minimise contamination and maintenance. EFF1 (CEMEP) rated TEFC, IP55 class F electric motors shall be used and incorporate maintenance-free greased for life bearings. Motors with lower efficiency ratings are not acceptable. Variable Speed Drive (VSD) shall be available on compressor(s). By including an AC-DC converter, along with associated control hardware and software it will enable the compressor to match it’s running speed on a 1:1 ratio with the flow demand required by hospital. By using such technology energy of savings of up to 35% shall be achievable as well as prolonged machine life. Start currents will be reduced and the motor will run from 0-100% speed increasing or decreasing it’s speed as necessary to continuously match the required demand. The compressor shall operate from 400-1300kPa (4-13 bar) gauge. ‘Full VSD’ air plant shall incorporate VSD controllers on all compressors, cycling the lead compressor to ensure even wear as per HTM02-01 requirements. ‘Mixed VSD’ air plant shall incorporate a VSD controller on the lead compressor and the remaining compressors will be fixed speed. Dryer/Filter/Regulator System The duplexed filter and dryer module shall incorporate high efficiency water separators, oil filters, heatless regenerative desiccant dryer, dust/activated carbon filters, hopcolite filters and bacterial filters with autoclavable element. Electrical contacts shall be installed on the filters to provide warning alarms on the dryer controller in the event of high pressure drop (ie blockage) and shall also include connections for BMS. Contaminants in the delivered air downstream of the bacterial filters shall be maintained at levels below those shown in the following table: Contaminant Threshold H2O 67 ppm v/v Dry particulates Free from visible particulates in a 75 litre sample Oil (droplet or mist) 0.1 mg/m³ CO 5 ppm v/v CO2 500 ppm v/v SO2 1 ppm v/v NO 2 ppm v/v NO2 2 ppm v/v • Logging and trending for an accurate performance status of the system. • Desktop event notification to avoid constant status checking. • E-mail and SMS event notification for additional convenience. BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982 Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450 In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. Dryer Purge Control - Optional The dryer control system shall incorporate a Purge Saver Energy Management system that freezes the regeneration of the desiccant once adequate dew point is reached in the inactive tower. Only when the dewpoint level in the active tower deteriorates to an unacceptable level will the intelligent controller switch towers. This shall be achieved by including an additional dew point sensor and associated software in the dryer controller to effectively manage the system as well as providing on screen measurements of purge savings. Dew Point Monitoring The dryer shall incorporate a ceramic dew point hygrometer with an accuracy of ±10C in the range -20 to -800C atmospheric dew point and 4-20mA analogue output. Aluminium oxide or palladium wire sensors are not acceptable. An alarm condition shall trigger on the dryer control panel if the dew point exceeds a -460C atmospheric set point. The plant control unit shall incorporate a multifunctional LCD displaying, amongst other things, the dew point of the delivered air to enable monitoring of the air quality by the hospitals estates department. Volt free contacts shall be included to enable the dew point alarm signal to be connected to a central medical gas alarm system and/or building management system (BMS). To enable periodic calibration of the dew point sensor element, the hygrometer shall be remotely connected downstream of the dryer via a micro-bore tube. It is not acceptable to install the sensor directly into the medical air supply pipeline. 2005628 03/05/12 Page 2 of 9 Issue 3 Oil-free Compressors - Optional Medical air plant, surgical air plant 7 bar and combined medical and surgical air plant 7 bar can be supplied with oil-free compressors. Incorporating either Atlas Copco ZT Claw compressors or SF Scroll compressors - all available upon special request. CE Marking The standard range of BeaconMedæs Medical Air plant systems are ‘CE’ marked under the Medical Devices Directive 93/42/EEC with approval from notified body no. 0088 (Lloyd’s Register Quality Assurance). Under this directive, the specified products are classified as Class IIa Medical Devices. Typical Layout Receivers Compressors Receiver Assembly Air receivers shall comply with DIRECTIVE 97/23/EC, supplied with relevant test certificates. Each air receiver shall be hot dip galvanised inside and out and fitted with a zero loss electronic drain valve. Float type drain valves are not acceptable. The receiver assembly shall be fitted with a pressure safety valve capable of passing the maximum flow output of the compressor at 10% receiver overpressure. The receiver shall be further protected by a safety pressure relief valve and include a pressure gauge. Optional Items There shall be the followings options available for enhanced operation of the air plant system:• Phase sequence relays that prevent unintentional reverse operation of the compressors • OCS electronic water/oil drains for the air plant system • EWD zero loss electronic water drains for the dMED dryer including secure mounting to the dryer base Dryer Note:Inter connecting pipework (blue illustration) between components to be made on site and provided by the installer. Controller CAN cables are provided as a 10m assembly with each compressor which can be shortened on site if required. Combined Air Plant Sizing Guide In HTM02-01, the relative size of receiver capacity and compressor capacity on surgical air or combined medical/surgical air systems changes according to the design flow rate. In order to correctly calculate the receiver capacity and compressor capacity, both the medical and surgical design flow-rates (DF’s) are required. It should be noted that for all combined air systems, an additional duplex regulating station (ordered separately) is needed to supply the medical air pipeline. • VSD compressors and optional harmonics filter for installation close to highly sensitive equipment • Synthetic oil for increased compressor life • QDT saturation indicators to give clear visual indication of oil carry over to the activated carbon tower Surgical Air Compressors Design Flow (l/min) Value ‘A’ FAD (l) <500 0.33 x DF 500-3500 0.66 x DF >3500 0.5 x DF Table 1: Surgical Air Plant Flow Rate Multiplier Value ‘A’. • Tropical thermostatic sensors for countries with high humidity • Heavy duty inlet filters for compressors installed in areas of highly concentrated dust levels • Painted hot dipped galvanised vessels • CO and CO2 monitors including full integration into the ES-Med central controller giving alarm warnings when unacceptable CO and CO2 levels are present Surgical Air Receivers Design Flow (l/min) Value ‘B’ Receiver water capacity (l) <500 1 x 200% x DF 500-2000 2 x 66.6% x DF 2001-3500 2 x 50% x DF >3500 3 x 33.3% x DF Table 2: Surgical Air Receiver Multiplier Value ‘B’. BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982 Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450 In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. 2005628 03/05/12 Page 3 of 9 Issue3 Example 1 - Small Day Treatment Centre (Upgrade) Example 2 - Large District Hospital Flow Rate and Dryer Sizing Medical Air DF = 555 l/min (FAD) (4 Bar) Surgical Air DF = 1138 l/min (FAD) (7 Bar) Combined/total DF = 1693 l/min (FAD) (11 Bar high pressure system) Flow Rate and Dryer Sizing Medical Air DF = 4920 l/min (FAD) (4 Bar) Surgical Air DF =2888 l/min (FAD) (11 Bar) Combined/total DF = 7808 l/min (FAD) (11 Bar high pressure system) A dryer greater than 1693 l/min outlet flow should be selected (outlet flow is the inlet flow minus purge losses) = dMED25 inlet flow 2025 l/min, outlet flow 1755 l/min A dryer greater than 7808 l/min should be selected (outlet flow is the inlet flow minus purge losses) = dMED145 inlet flow 11745 l/min, outlet flow 10092 l/min Flow Rate and Compressor Sizing From Table 1 surgical air DF is between 500-3500 l/min, so the multiplying factor ‘A’ = 0.66 Flow Rate and Compressor Sizing From Table 1 surgical air DF is between 500-3500 l/min, so the multiplying factor ‘A’ = 0.66 Compressor flow rate = Med. DF + (Surg. DF x A) = 555 + (1138 x 0.66) = 555 + 751 = 1306 l/min Plant flow rate We also need to add the purge losses to the compressor output. For additional purge consumption use:- We also need to add the purge losses to the compressor output. For additional purge consumption use:- dMED inlet - dMED outlet = purge losses lpm = 2025 - 1755 = 270 l/min dMED inlet - dMED outlet = purge losses lpm = 11745 - 10092 = 1653 l/min Compressors should be selected with a flow rate greater than 1306 l/min + 270 l/min = 1576 l/min Compressors should be selected with a flow rate greater than 6827 l/min + 1653 l/min = 8480 l/min Receiver Sizing From Table 2 surgical air DF is between 500-2000 l/min, so the multiplying factor ‘B’ = 2 x 2/3 Receiver Sizing From Table 2 surgical air DF is between 2001-3500 l/min, so the multiplying factor ‘B’ = 2 x 1/2 Capacity Capacity = (Med. DF x 0.5) + (Surg. DF x B) = (555 x 0.5) + (1138 x 2 x 2/3) = 278 + 1518 = 1796 litres = Med. DF + (Surg. DF x Value ‘A’) = 4920 + (2888 x 0.66) = 4920 + 1907 = 6827 l/min = (Med. DF x 0.5) + (Surg. DF x B) = (4920 x 0.5) + (2888 x 2 x 1/2) = 2460 + 2888 = 5348 litres A combination of receivers with a minimum number of 2 should be selected A combination of receivers with a minimum number of 2 should be selected Selected receiver capacity = 2000 litres (2 x 1000 litre) Selected receiver capacity = 6000 litres (3 x 2000 litre) Plant System Selection Selected plant part number = cAIR-1755-TGA11 Plant System Selection Selected plant part number = n/a If no standard model is available for selection from the standard range a bespoke configuration of dryer, compressors and receivers are available and can be quoted by our sales and sales support teams. If no standard model is available for selection from the standard range a bespoke configuration of dryer, compressors and receivers are available and can be quoted by our sales and sales support teams. Receiver Selection Table * Receiver Capacity (litres) 250 500 1000 1500 2000 Maximum working pressure (bar) 16 16 16 16 16 Individual Receiver Dimensions (diameter, height, mm) 500/1950 600/2350 800/2550 1000/2525 1150/2605 Receiver Weight (kg) 80 160 304 445 557 Receiver pipe size (mm) 22 22 28 28 35 Receiver Part Number Accessory Kit Part Number * 8101 0211 97 Specific to plant 8101 0212 62 Specific to plant 8101 0213 12 Specific to plant 8102 0213 38 Specific to plant 8101 0213 53 Specific to plant Drawing Number tbc tbc tbc tbc tbc Accessory kit for medical air receiver complete with plant data plate, test certificate, pressure safety valve, zero-loss electronic drain valve (with isolation and bypass valve), pressure gauge (with isolation valve), safety pressure relief valve and inlet and copper outlet union connection pipes (each with isolation valve). BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982 Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450 In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. 2005628 03/05/12 Page 4 of 9 Issue 3 Compressor Selection Table - Fixed Speed Model Name GA-MED 5 GA-MED 7 GA-MED 11 GA-MED 15 GA-MED 18 GA-MED 22 Output flow (litres/minute) * 7.5 bar variant 900 1308 1842 2580 3150 3612 Output flow (litres/minute) * 10 bar variant 702 1032 1560 2178 2610 3102 Output flow (litres/minute) * 13 bar variant 504 852 1320 1806 2232 2700 Footprint L x W x H (mm) 1140 x 700 x 1240 1140 x 700 x 1240 1140 x 700 x 1240 1285 x 680 x 922 1285 x 680 x 922 1285 x 680 x 922 Compressor weight (kg) 270 284 310 375 395 410 Service connection (mm) 22 22 22 22 22 22 Noise level/pump (dB[A]) 60 61 62 72 73 74 Maximum ambient temperature (0C) 46 46 46 46 46 46 Supply voltage (v) 400 400 400 400 400 400 Supply frequency (Hz) 50 50 50 50 50 50 Nominal motor rating (kW) 5 7 11 15 18 22 Full load current per compressor (A) 17 22 32 36 43 54 Starting current (A) 98.6 131 188 252 302 383 Cooling air flow per compressor (m3/s) 0.8 0.8 1 1.1 1.15 1.2 Part Number - 7.5 bar 8153 0344 86 8153 0344 94 8153 0345 02 8153 0350 04 8153 0350 38 8153 0350 61 Part Number - 10 bar 8152 0341 89 8153 0342 05 8153 0342 21 8153 0350 12 8153 0350 46 8153 0350 79 Part Number - 13 bar 8153 0341 97 8153 0342 13 8153 0342 39 8153 0350 20 8153 0350 53 8153 0350 87 Drawing Number 9828 4969 26 9828 4969 26 9828 4969 26 9820 6060 17 9820 6060 17 9820 6060 17 * Output flow stated at reference conditions Compressor Selection Table - Variable Speed Drive * Model Name GA-MED VSD 5 GA-MED VSD 7 GA-MED VSD 11 GA-MED VSD 15 Output flow (litres/minute) 7.5 bar , 10 bar and 13 bar variants * 900/ 792 / 600 1218 / 1008 / 828 1842/ 1446 / 1242 2226 / 1854 / 1488 Footprint L x W x H (mm) 1345 x 700 x 1240 1345 x 700 x 1240 1345 x 700 x 1240 1345 x 700 x 1240 Compressor weight (kg) 290 295 308 315 Service connection (mm) 22 22 22 22 Noise level/pump (dB[A]) 62 64 66 69 Maximum ambient temperature (0C) 46 46 46 46 Supply voltage (v) 400 400 400 400 Supply frequency (Hz) 50 50 50 50 Nominal motor rating (kW) 5 7 11 15 Full load current per compressor (A) 20.9 24.3 36.3 45.3 Cooling air flow per compressor (m3/s) 0.8 0.8 1 1 Part Number 8153 0343 61 8153 0343 79 8153 0343 87 8153 0343 95 Drawing Number 9828 4969 27 9828 4969 27 9828 4969 27 9828 4969 27 Variable speed drive compressor operate from 4-13 bar at 0-100% speed - start currents nominal for VSD BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982 Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450 In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. 2005628 03/05/12 Page 5 of 9 Issue 3 Dryer Selection Table dMED145 Inlet flow at 7.5 bar (litres/minute) 420 780 1500 2100 3000 4200 4800 6000 8700 Output flow (litres/minute) at 4 bar line pressure * 340 632 1230 1722 2430 3444 3936 4860 7047 Inlet flow at 10 bar (litres/minute) 504 936 1800 2520 3600 5040 5760 7200 10440 Output flow (litres/minute) at 7 bar line pressure * 424 788 1530 2142 3030 4284 4896 6060 8787 Inlet flow at 13 bar (litres/minute) 567 1053 2025 2835 4050 5670 6480 8100 11745 Output flow (litres/minute) at 10 bar line pressure * 487 905 1755 2457 3480 4916 5616 6960 10092 Footprint L x W x H (mm) 1920 x 600 x 1590 1920 x 600 x 1590 1920 x 600 x 1590 1920 x 600 x 1590 1920 x 600 x 1870 2160 x 1250 x 1850 2160 x 1250 x 1850 2160 x 1250 x 1850 2160 x 1250 x 1870 Dryer weight (kg) 255 275 325 345 415 615 700 705 895 Inlet and outlet connections (mm) 15 15 15 28 28 28 28 28 28 Supply voltage (v) 230 230 230 230 230 230 230 230 230 Supply frequency (Hz) 50 50 50 50 50 50 50 50 50 Central control supply - single phase (mm2/Amps) 1.5 (2) 1.5 (2) 1.5 (2) 1.5 (2) 1.5 (2) 1.5 (2) 1.5 (2) 1.5 (2) 1.5 (2) Part Number - dryer at 7 bar outlet + QDT hopcolite filter for EurPh 8102 1433 54 0000 0203 58 8102 1433 62 0000 0203 58 8102 1433 70 0000 0203 71 8102 1433 88 0000 0203 71 8102 1433 96 0000 0203 72 8102 1434 04 0000 0203 73 8102 1434 12 0000 0203 73 8102 1434 20 0000 0203 74 8102 1434 38 0000 0203 75 Part Number - dryer at 10 bar outlet + QDT hopcolite filter for EurPh 8102 1434 46 0000 0203 58 8102 1434 53 0000 0203 58 8102 1434 61 0000 0203 71 8102 1434 79 0000 0203 71 8102 1434 87 0000 0203 72 8102 1434 95 0000 0203 73 8102 1435 03 0000 0203 73 8102 1435 11 0000 0203 74 8102 1435 29 0000 0203 75 Drawing Number 9827 7000 08 9827 7000 09 9827 7000 10 9827 7000 11 9827 7000 12 9827 7000 13 9827 7000 14 9827 7000 15 9827 7000 16 Part Number - dryer at 4 bar outlet + QDT hopcolite filter for EurPh 8102 1432 55 0000 0203 75 dMED100 8102 1432 48 0000 0203 74 dMED80 8102 1432 30 0000 0203 73 dMED70 8102 1432 22 0000 0203 73 dMED50 8102 1432 14 0000 0203 72 dMED35 8102 1432 06 0000 0203 71 dMED25 8102 1431 98 0000 0203 71 dMED13 8102 1431 80 0000 0203 58 dMED7 8102 1431 72 0000 0203 58 Model Name * Output flow stated includes calculated purge lost during the regenration process of between 15-19% depending on model and inlet pressure. Notes on plant • • • • • Design flow in terms of free air delivered after losses at working pressure with the reserve compressor(s) on standby. Tolerance ±5%. Component dimensions supplied do not include maintenance access space, and are provided to allow customer to arrange plant components within plant room. Complete installation drawings are available on request. Quote the drawing number required. Duplex systems must be installed with a manifold as the third source of supply for HTM02-01 complaince. Mean sound level in accordance with ISO 2151. Electrical details are provided for guidance only. Site conditions may impose a larger cable size. For exact cable sizing, and fuse / MCB ratings, consult a qualified electrical engineer. BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982 Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450 2005628 In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. 03/05/12 Page 6 of 9 Issue 3 Standard Models mAIR-1465-TGF4 mAIR-1720-TGF4 mAIR-2000-TGF4 mAIR-2410-TGF4 mAIR-2855-TGF4 mAIR-3445-QGF4 mAIR-3935-QGF4 mAIR-4860-QGF4 mAIR-5570-QGF4 mAIR-6000-PGF4 mAIR-7045-PGF4 1230 1465 1720 2000 *2430 2855 3445 3935 4860 5570 6000 7045 Number of compressors 3 3 3 3 3 3 3 3 3 3 4 4 4 4 5 5 Type of compressor GAMED 5 GAMED 5 GAMED 5 GAMED 7 GAMED 11 GAMED 11 GAMED 15 GAMED 15 GAMED 18 GAMED 22 GAMED 15 GAMED 15 GAMED 18 GAMED 22 GAMED 15 GAMED 18 Type of dryer dMED 7 dMED 13 dMED 13 dMED 25 dMED 25 dMED 35 dMED 35 dMED 50 dMED 50 dMED 70 dMED 70 dMED 80 dMED 100 dMED 145 dMED 145 dMED 145 Number of receivers 1 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 Receiver volume (l) (each) 250 250 250 250 500 500 500 500 1000 1000 1000 1000 1500 1500 1500 2000 8153 0353 01 9827 8413 00 8153 0353 19 9827 8413 00 8153 0353 27 9827 8414 00 8153 0353 35 9827 8414 00 8153 0353 43 9827 8415 00 8153 0353 84 9827 8421 00 8153 0353 92 9827 8421 00 8153 0354 00 9827 8421 00 8153 0354 18 9827 8421 00 8153 0354 26 9827 8422 00 8153 0354 59 9827 8424 00 mAIR-1230-TGF4 1000 8153 0354 42 9827 8422 00 mAIR-1000-TGF4 630 8153 0354 34 9827 8422 00 mAIR-630-TGF4 500 8153 0353 76 9827 8420 00 mAIR-500-TGF4 340 8153 0353 68 9827 8420 00 mAIR-340-TGF4 Design flow (litres/minute) 8153 0353 50 9827 8415 00 Model Number Part Number Drawing Number HTM02-01 Medical Air 4 bar cAIR-1000-TGF7 cAIR-1530-TGF7 cAIR-2140-TGF7 cAIR-2450-TGF7 cAIR-3030-QGF7 cAIR-4285-QGF7 cAIR-4895-QGF7 cAIR-6000-QGF7 cAIR-8000-PGF7 cAIR-8790-PGF7 790 998 1530 2140 2450 3030 4285 4895 6000 8000 8790 524 664 1269 1776 2035 2515 3556 4064 4980 6000 6593 985 1248 1402 1962 2245 2777 3213 3672 4500 6000 6593 3 3 3 3 3 3 4 4 4 4 5 5 Type of compressor GAMED 5 GAMED 5 GAMED 7 GAMED 11 GAMED 15 GAMED 18 GAMED 11 GAMED 15 GAMED 18 GAMED 22 GAMED 18 GAMED 18 Type of dryer dMED 7 dMED 13 dMED 25 dMED 25 dMED 35 dMED 50 dMED 50 dMED 70 dMED 80 dMED 100 dMED 145 dMED 145 1 2 1 2 2 2 2 2 2 3 3 4 1000 500 1500 1000 1000 1500 1500 2000 2000 1500 2000 2000 8153 0355 33 9827 8424 00 8153 0355 41 9827 8424 00 8153 0355 58 9827 8423 00 8153 0355 66 9827 8425 00 8153 0355 74 9827 8426 00 Part Number Drawing Number Number of receivers Receiver volume (l) (each) 8153 0355 25 9827 8419 00 Number of compressors 8153 0355 17 9827 8422 00 530 8153 0355 09 9827 8421 00 282 Required receiver volume * 8153 0354 91 9827 8417 00 Required compressor output * 8153 0354 83 9827 8418 00 420 8153 0354 75 9827 8415 00 Design flow (litres/minute) 8153 0354 67 9827 8416 00 Model Number cAIR-420-TGF7 cAIR-790-TGF7 HTM02-01 Combined Medical & Surgical Air 7 bar * Flow output based on a 50/50 split of medical and surgical air design flow (D.F) For examples on how to calculate the required design flow, compressor flow and receiver volume refer to earlier page, in line with HTM02-01 requirements and tables. BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982 Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450 2005628 In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. 03/05/12 Page 7 of 9 Issue 3 * cAIR-1260-TGF11 cAIR-1755-TGF11 cAIR-2180-TGF11 cAIR-2460-TGF11 cAIR-3480-QGF11 cAIR-4470-QGF11 cAIR-4915-QGF11 cAIR-5460-QGF11 cAIR-6000-PGF11 cAIR-6955-PGF11 cAIR-8000-PGF11 cAIR-8760-QGF11 905 1260 1755 2180 2460 3480 4470 4915 5460 6000 6955 8000 8760 Required compressor output * 324 532 601 1046 1457 1809 2039 2888 3710 4079 4532 4980 5773 6000 6570 Required receiver volume * 609 1000 1130 1155 1609 1998 2252 3190 3353 3686 4095 4500 5217 6000 6570 Number of compressors 3 3 3 3 3 3 3 4 4 4 4 5 5 5 4 Type of compressor GAMED 5 GAMED 7 GAMED 7 GAMED 11 GAMED 15 GAMED 18 GAMED 22 GAMED 15 GAMED 18 GAMED 22 GAMED 22 GAMED 18 GAMED 22 GAMED 22 GAMED 30 Type of dryer dMED 7 dMED 13 dMED 13 dMED 25 dMED 25 dMED 35 dMED 35 dMED 50 dMED 70 dMED 70 dMED 80 dMED 100 dMED 100 dMED 145 dMED 145 Number of receivers 1 1 2 2 2 2 2 2 2 2 3 3 3 3 4 Receiver volume (l) (each) 1000 1000 1000 1000 1000 1000 1500 2000 2000 2000 1500 1500 2000 2000 2000 8153 0355 82 9827 8416 00 8153 0355 90 9827 8416 00 8153 0356 16 9827 8417 00 8153 0356 24 9827 8421 00 8153 0356 32 9827 8421 00 8153 0356 57 9827 8424 00 8153 0356 65 9827 8424 00 8153 0356 73 9827 8424 00 8153 0356 81 9827 8423 00 8153 0356 99 9827 8423 00 8153 0357 07 9827 8425 00 n/a special cAIR-905-TGF11 800 8153 0357 15 9827 8425 00 cAIR-800-TGF11 485 8153 0356 40 9827 8422 00 cAIR-485-TGF11 Design flow (litres/minute) 8153 0356 08 9827 8417 00 Model Number Part Number Drawing Number HTM02-01 Combined Medical & Surgical Air 11 bar Flow output based on a 50/50 split of medical and surgical air design flow (D.F). For examples on how to calculate the required design flow, compressor flow and receiver volume refer to earlier page, in line with HTM02-01 requirements and tables. sAIR-485-SGF11 sAIR-750-SGF11 sAIR-905-SGF11 sAIR-1445-SGF11 sAIR-1755-SGF11 sAIR-2460-SGF11 sAIR-3225-SGF11 sAIR-3890-SGF11 sAIR-4915-SGF11 sAIR-5615-SGF11 HTM02-01 Surgical Air 11 bar Design flow (litres/minute) 485 750 905 1445 1755 2455 3225 3890 4915 5615 Required compressor output * 161 495 597 954 1158 1642 2129 1945 2458 2808 Required receiver volume * 974 1000 1207 1927 2340 2460 3225 3890 4915 56105 Model Number 1 Type of compressor GA-MED 30 Type of dryer dMED 7 dMED 13 dMED 13 dMED 25 dMED 25 dMED 35 dMED 50 dMED 70 dMED 70 dMED 80 1 2 2 2 2 2 2 3 3 3 1000 500 1000 1000 1500 1500 2000 1500 2000 2000 Part Number Drawing Number Number of receivers Receiver volume (l) (each) n/a special 1 GA-MED 30 n/a special 1 GA-MED 22 8153 0357 98 9827 8423 00 1 GA-MED 22 8153 0357 80 9827 8424 00 1 GA-MED 18 8153 0357 72 9827 8422 00 1 GA-MED 15 8153 0357 64 9827 8422 00 1 GA-MED 11 8153 0357 56 9827 8417 00 1 GA-MED 7 8153 0357 49 9827 8417 00 1 GA-MED 7 8153 0357 31 9827 8415 00 1 GA-MED 5 8153 0357 23 9827 8416 00 Number of compressors * For examples on how to calculate the required design flow, compressor flow and receiver volume refer to earlier page, in line with HTM02-01 requirements and tables. BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982 Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450 In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. 2005628 03/05/12 Page 8 of 9 Issue 3 HTM02-01 Combined Medical & Surgical Air 11 bar - Varible Speed Drive Available upon request Model Number mAIR-340-DGF4 mAIR-500-DGF4 mAIR-630-DGF4 mAIR-1000-DGF4 mAIR-1230-DGF4 mAIR-1465-DGF4 mAIR-1720-DGF4 mAIR-2000-DGF4 mAIR-2430-DGF4 mAIR-2855-DGF4 mAIR-3445-TGF4 mAIR-3935-TGF4 mAIR-4860-TGF4 mAIR-5570-TGF4 mAIR-6000-QGF4 mAIR-7045-QGF4 Design flow (litres/minute) 340 500 630 1000 1230 1465 1720 2000 2430 2855 3445 3935 4860 5570 6000 7045 Number of compressors 2 2 2 2 2 2 2 2 2 2 3 3 3 3 4 4 Type of compressor GAMED 5 GAMED 5 GAMED 5 GAMED 7 GAMED 11 GAMED 11 GAMED 15 GAMED 15 GAMED 18 GAMED 22 GAMED 15 GAMED 15 GAMED 18 GAMED 22 GAMED 15 GAMED 18 Type of dryer dMED 7 dMED 13 dMED 13 dMED 25 dMED 25 dMED 35 dMED 35 dMED 50 dMED 50 dMED 70 dMED 70 dMED 80 dMED 100 dMED 145 dMED 145 dMED 145 Number of receivers 1 1 1 1 1 1 1 1 1 1 1 1 2 2 2 2 Receiver volume (l) (each) 250 250 500 500 1000 1000 1000 1000 1500 1500 2000 2000 1500 1500 1500 2000 Part Number Drawing Number 8153 0358 06 9827 8427 00 8153 0358 14 9827 8427 00 8153 0358 22 9827 8428 00 8153 0358 30 9827 8428 00 8153 0358 48 9827 8429 00 8153 0358 55 9827 8429 00 8153 0358 63 9827 8430 00 8153 0358 71 9827 8430 00 8153 0358 89 9827 8431 00 8153 0358 97 9827 8431 00 8153 0359 05 9827 8434 00 8153 0359 13 9827 8434 00 8153 0359 21 9827 8432 00 8153 0359 39 9827 8432 00 8153 0359 47 9827 8432 00 8153 0359 54 9827 8435 00 HTM2022 EurPh Medical Air 4 bar cAIR-425-DGF7 cAIR-555-DGF7 cAIR-785-DGF7 cAIR-1000-DGF7 cAIR-1290-DGF7 cAIR-1530-DGF7 cAIR-1800-DGF7 cAIR-2140-DGF7 cAIR-2535-DGF7 cAIR-3000-TGF7 cAIR-3600-TGF7 cAIR-4285-TGF7 cAIR-4895-TGF7 cAIR-6000-QGF7 cAIR-7655-QGF7 cAIR-8790-PGF7 Design flow (litres/minute) 425 555 785 1000 1290 1530 1800 2140 2535 3000 3600 4285 4895 6000 7655 8790 Number of compressors 2 2 2 2 2 2 2 2 2 3 3 3 3 4 4 5 Type of compressor GAMED 5 GAMED 5 GAMED 7 GAMED 11 GAMED 11 GAMED 15 GAMED 15 GAMED 18 GAMED 22 GAMED 15 GAMED 15 GAMED 18 GAMED 22 GAMED 18 GAMED 22 GAMED 18 Type of dryer dMED 7 dMED 13 dMED 13 dMED 25 dMED 25 dMED 25 dMED 35 dMED 35 dMED 50 dMED 50 dMED 70 dMED 70 dMED 80 dMED 100 dMED 145 dMED 145 Number of receivers 1 1 1 1 1 1 1 1 1 1 1 2 2 2 2 3 Receiver volume (l) (each) 250 500 500 500 1000 1000 1000 1500 1500 1500 2000 1500 1500 1500 2000 1500 8153 0359 70 9827 8428 00 8153 0359 88 9827 8428 00 8153 0359 96 9827 8428 00 8153 0360 02 9827 8429 00 8153 0360 10 9827 8430 00 8153 0360 28 9827 8430 00 8153 0360 36 9827 8431 00 8153 0360 44 9827 8431 00 8153 0360 51 9827 8431 00 8153 0360 69 9827 8434 00 8153 0360 77 9827 8432 00 8153 0360 85 9827 8432 00 8153 0360 93 9827 8432 00 8153 0361 01 9827 8435 00 8153 0361 19 9827 8433 00 Part Number Drawing Number Model Number 8153 0359 62 9827 8427 00 HTM2022 EurPh Combined Medical & Surgical Air 7 bar BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982 Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450 In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. 2005628 03/05/12 Page 9 of 9 Issue 3 • • • • • Model Number cAIR-425-DGF11 cAIR-500-DGF11 cAIR-705-DGF11 cAIR-905-DGF11 cAIR-1000-DGF11 cAIR-1535-DGF11 cAIR-1755-DGF11 cAIR-2325-DGF11 cAIR-3000-TGF11 cAIR-3710-TGF11 cAIR-4645-TGF11 cAIR-4915-QGF11 cAIR-5615-QGF11 cAIR-6960-QGF11 cAIR-7270-PGF11 cAIR-8790-PGF11 Design flow (litres/minute) 425 500 705 905 1000 1535 1755 2325 3000 3710 4645 4915 5615 6960 7270 8790 Number of compressors 2 2 2 2 2 2 2 2 3 3 3 4 4 4 5 5 Type of compressor GAMED 5 GAMED 7 GAMED 7 GAMED 11 GAMED 11 GAMED 15 GAMED 18 GAMED 22 GAMED 15 GAMED 18 GAMED 22 GAMED 18 GAMED 18 GAMED 22 GAMED 18 GAMED 22 Type of dryer dMED 7 dMED 13 dMED 13 dMED 13 dMED 25 dMED 25 dMED 25 dMED 35 dMED 50 dMED 70 dMED 70 dMED 70 dMED 80 dMED 100 dMED 145 dMED 145 Number of receivers 1 1 1 1 1 1 1 1 1 1 2 2 2 2 2 3 Receiver volume (l) (each) 250 250 500 500 500 1000 1000 1500 1500 2000 1500 1500 1500 2000 2000 2000 Part Number Drawing Number 8153 0361 27 9827 8427 00 8153 0361 35 9827 8427 00 8153 0361 43 9827 8428 00 8153 0361 50 9827 8428 00 8153 0361 68 9827 8428 00 8153 0361 76 9827 8430 00 8153 0361 84 9827 8430 00 8153 0361 92 9827 8431 00 8153 0362 00 9827 8431 00 8153 0362 18 9827 8434 00 8153 0362 26 9827 8432 00 8153 0362 34 9827 8432 00 8153 0362 42 9827 8432 00 8153 0362 59 9827 8435 00 8153 0362 67 9827 8435 00 8153 0362 75 9827 8436 00 HTM2022 Combined Medical & Surgical Air 11 bar Design flow in terms of free air delivered after losses at working pressure with the reserve compressor(s) on standby. Tolerance ±5%. Component dimensions supplied do not include maintenance access space, and are provided to allow customer to arrange plant components within plant room. Complete installation drawings are available on request. Quote the drawing number required. Duplex systems must be installed with a manifold as the third source of supply for HTM02-01 complaince. Mean sound level in accordance with ISO 2151. Electrical details are provided for guidance only. Site conditions may impose a larger cable size. For exact cable sizing, and fuse / MCB ratings, consult a qualified electrical engineer. BeaconMedæs Telford Crescent, Staveley, Derbyshire S43 3PF, England Tel: +44 (0) 1246 474 242 Fax: +44 (0) 1246 472 982 www.beaconmedaes.com E-mail: gbn.info@beaconmedaes.com 0088 2005161 08/11/11 Page 1 of 4 Issue 6 mVAC Medical Vacuum System 400v 50Hz – EN ISO 7396-1 & HTM 02-01 mVAC MEDICAL VACUUM SYSTEM In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. SPECIFICATION mVAC Medical Vacuum The mVAC Medical Vacuum System shall conform to EN ISO 7396-1 and NHS Health Technical Memorandum No. 02-01 (HTM 02-01). The Medical Vacuum System shall ensure the minimum pipeline vacuum level of 450mmHg is maintained at the plant service connection point at the rated volumetric ‘free air’ flow rate with two pumps in standby. The bacteria filtration system shall be ‘duplexed’ such that each filter can be isolated for replacement of the filter cartridge. Vacuum Pumps Vacuum pumps shall be air-cooled, oil lubricated rotary vane type suitable for both continuous and frequent start/stop operation at nominal inlet vacuum levels of between 578mmHg and 728mmHg. Composite carbon fibre rotor blades shall be fitted to minimise the cost of maintenance. Rotors shall be driven by directly coupled TEFV, IE2 efficiency electric motors. Pump inlets shall include a wire mesh filter and integral non-return valve to prevent oil suck back and pressure increases in the vacuum system. Each vacuum pump shall have an integral separator filter to ensure a virtually oil-free exhaust. Each pump shall be fitted with anti-vibration pads between the pump foot and mounting frame. Bacteria Filters The duplex bacteria filter system shall incorporate high efficiency filter elements. A differential vacuum indicator shall be installed across the filter to indicate blockage. Additional pressure sensors shall be installed at the inlet and outlet of the filter to measure the pressure drop across the filters. Each filter shall be designed and sized to carry the full plant design flow capacity with a pressure drop not exceeding 33mbar (25mmHg). Bacteria Filter elements shall have penetration levels not exceeding 0.005% when tested by the sodium flame method in accordance with BS 3928:1969 and utilising particles in the 0.02 to 2 micron size range. Drain flasks shall be connected to each filter. Drain flasks shall be manufactured from transparent Pyrex® with a polymer coating on the inner and outer surfaces in order to maintain a seal in the event of inadvertent breakage of the Pyrex® flask. All drain flasks shall be suitable for sterilisation and be connected via a manual isolating valve. Control System The central control system shall provide an intelligent human machine interface incorporating on board flash memory and real-time clock for recording operational parameters in the in built event log. The central control system shall operate at low voltage and include BMS connection for commont fault. Visualisation of plant inputs, outputs and status through a web browser, using a simple Ethernet connection shall be available. The central control unit shall incorporate a user friendly 5.7” high-definition colour display with clear pictograms and LED indicators, providing easy access to system operational information. Optional Control Equipment An advanced monitoring system shall be available to give immediate access to valuable information such as system status, trends, historical data and system performance. Data collected from all pumps shall be made available in real-time visualisation pages and shall be accessed through the hospital’s LAN, such that total data security is assured. The Airconnect™ monitoring system shall also include :• Logging and trending for an accurate performance status of your system. • Desktop event notification to avoid constant status checking. • E-mail and SMS event notification for additional convenience. Optional Items There shall be the followings options available for enhanced operation of the vacuum plant system:• Synthetic oil for increased pump life. • Painted hot dipped galvanised vessels. • Oil level switch and integrated ES-VAC alarm for notification of low oil level in the pumps. Vacuum Receiver(s) Vacuum receiver(s) shall be supplied with relevant test certificates and have a total volume of at least 100% of the plant output in 1 minute in terms of free air aspired at normal working pressure. Each vacuum receiver shall be hot dip galvanised inside and out. Pyrex® is a registered trademark of Corning Glass. CE Marking The standard range of BeaconMedæs Medical Vacuum plant systems are ‘CE’ marked under the Medical Devices Directive 93/42/EEC with approval from notified body no. 0088 (Lloyd’s Register Quality Assurance). Under this directive, the specified products are classified as Class IIa Medical Devices. Cascading of vacuum pumps shall be achieved by measuring the vacuum level at the plant inlet with a pressure transducer. A mechanical back-up facility shall ensure continued operation in the event of a control system malfunction. The control system shall normally employ automatic rotation of the lead pump to maximise pump life and ensure even wear. BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982 Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450 2005161 In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. 08/11/11 Page 2 of 4 Issue 6 Typical Layout Model L (ength) W (idth) H (eight) LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø mVAC250 T 2300 980 1650 3600 1980 500 800 54 54 n/a mVAC330 T 2300 980 1650 3600 1980 500 800 54 54 n/a mVAC500 T 2400 980 1650 3700 1980 500 800 54 54 n/a Model L (ength) W (idth) H (eight) ØD LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø mVAC660 Q 1910 980 2100 660 3210 3740 500 800 54 54 54 outl. unit - 54 inl. vessel mVAC1000 Q 1910 980 2100 660 3210 3740 500 800 54 54 54 outl. unit - 54 inl. vessel mVAC2560 Q 2600 1200 2400 1060 3900 4360 500 800 76 54 76 outl.unit - 54 inl. vessel mVAC3300 Q 2600 1200 2500 1230 3900 4530 500 800 76 54 76 outl.unit - 76 inl. vessel mVAC3840 P 4100 1250 2500 1230 5990 4580 500 800 76 54 76 outl.unit - 76 inl. vessel BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982 Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450 In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. Model L (ength) W (idth) H (eight) ØD LMA WMA F G mVAC1280 T 2200 1100 mVAC4950 P 4100 1250 2100 660 4280 3860 500 2500 1230 5990 4580 500 mVAC6000 H 4100 1250 2500 1230 7220 4580 500 Model mVAC6600 H L (ength) W (idth) H (eight) ØD LMA WMA 4100 1250 2500 1230 7220 4580 2005161 08/11/11 Page 3 of 4 Issue 6 I (nlet) Ø E (xhaust) Ø V (essel con.) Ø 800 76 54 54 outl. unit - 54 inl. Vessel 800 108 54 108 outl. unit - 76 inl. Vessel 800 108 54 108 outl.unit - 76 inl. Vessel F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø 500 800 108 54 108 outl.unit - 76 inl. Vessel Note! mVAC-6600-H includes 4 x vessels Drawing shows other models not used in this range and are for illustration purposes only. BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982 Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450 • • • • • mVAC-250-T mVAC-330-TH mVAC-330-T mVAC-500-TH mVAC-500-T mVAC-660-Q mVAC-1000-Q mVAC-1280-T mVAC-2560-Q mVAC-3300-Q mVAC-3840-P mVAC-4950-P mVAC-6000-H mVAC-6600-H 08/11/11 Page 4 of 4 Issue 6 Model Number Plant Output (litres/minute) 250 250 330 330 500 500 660 1000 1280 2560 3300 3840 4950 6000 6600 System Flow (m3/h - l/m) 40 665 40 665 52 865 52 865 79 1315 79 1315 105 1750 159 2650 203 3380 406 6765 524 8730 609 10150 786 13100 952 15865 1047 17450 Number of pumps 3 3 3 3 3 3 4 4 3 4 4 5 5 6 6 Number of receivers 1(H) 1 1(H) 1 1(H) 1 2 2 3 2 2 2 3 3 4 Total receiver volume (litres) 500 500 500 500 500 500 1000 1000 1500 3000 4000 4000 6000 6000 8000 Receiver connection(s) (mm) n/a 54 n/a 54 n/a 54 54 54 54 54 76 76 76 76 76 Inlet/service connection (mm) 54 54 54 54 54 54 54 54 54 76 76 76 108 108 108 Exhaust connection (mm, per pump) 54 54 54 54 54 54 54 54 54 54 54 54 54 54 54 Maximum exhaust back pressure (mbar) 60 60 60 60 60 60 60 60 60 60 60 60 60 60 60 Noise level/pump (dB[A]) 67 67 69 69 70 70 69 70 72 72 75 72 75 75 75 3820 7.5 7.5 Motor cable size (mm2/Amps) 1.5 (13) 1.5 (13) 1.5 (13) 1.5 (13) 1.5 (13) 1.5 (13) 1.5 (13) 1.5 (13) 1.5 (13) 1.5 (13) 2.5 (17) 1.5 (13) 2.5 (17) 2.5 (17) 2.5 (17) Motor rated supply per pump (A) 10 10 10 10 10 10 10 10 20 20 20 20 20 20 20 FLC per pump (A) 2.8 2.8 3.6 3.6 5.1 5.1 3.6 5.1 12 12 14.9 12 14.9 14.9 14.9 Starting current (A) 14 14 18 18 26 26 18 26 60 60 75 60 75 75 75 Central control supply - single phase (mm2/Amps) 1.5 (5) 1.5 (5) 1.5 (5) 1.5 (5) 1.5 (5) 1.5 (5) 1.5 (5) 1.5 (5) 1.5 (5) 1.5 (5) 1.5 (5) 1.5 (5) 1.5 (5) 1.5 (5) 1.5 (5) Maximum Inlet Temperature (0C) 40 40 40 40 40 40 40 40 40 40 40 40 40 40 40 Cooling air flow per pump (m3/s) 0.1 0.1 0.1 0.1 0.2 0.2 0.1 0.2 0.3 0.3 0.4 0.3 0.4 0.4 0.4 9820 6370 00 8152 1306 22 3390 7.5 9820 6369 00 8152 1306 20 3190 5.5 9820 6369 00 8152 1306 18 2660 7.5 9820 6368 00 8152 1306 16 2335 5.5 9820 6368 00 8152 1306 14 2215 5.5 9820 6368 00 8152 1306 12 1475 2.2 9820 6369 00 8152 1306 10 1040 1.5 9820 6368 00 8152 1306 08 960 2.2 9820 6368 00 8152 1306 06 - 2.2 8152 1306 50 750 1.5 9820 6365 00 8152 1306 04 - 1.5 81521 306 49 690 1.1 9820 6365 00 81521 306 02 - 1.1 8152 1306 48 650 9820 6365 00 8152 1306 00 Weight (kg) Motor rating (kW) Installation Proposal Part Number • 2005161 mVAC-250-TH In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. Plant Output in terms of free air aspired at a vacuum of 450 mmHg at the inlet connection with two pumps on standby and with a tolerance of ±10%. System Flow at atmospheric pressure at the inlet connection with two pump on standby and with a tolerance of ±10%. Plant dimensions include the required space around the plant for maintenance access. Mean sound level measured at a distance of 1m as measured to ISO 2151 / DIN 45635. Electrical details are provided for guidance only and are referenced at 400C ambient temperature. Site conditions may impose a larger cable size. For exact cable sizing, and fuse / MCB ratings, consult a qualified electrical engineer. Plant weight includes packaging for shipping purposes. BeaconMedæs Telford Crescent, Staveley, Derbyshire S43 3PF, England Tel: +44 (0) 1246 474 242 Fax: +44 (0) 1246 472 982 www.beaconmedaes.com E-mail: gbn.info@beaconmedaes.com 2005208 16/03/12 Page 1 of 3 Issue 7 Anaesthetic Gas Scavenging System HTM02-01/HTM2022/ISO7396-2 SPECIFICATION Anaesthetic Gas Scavenging System The Anaesthetic Gas Scavenging (AGS) System shall comply with HTM 02-01, HTM2022 and either EN ISO 7396-2 or BS 6834. The AGS system shall be a dedicated, specifically designed active extraction and disposal system for waste anaesthetic gas. It shall provide a maximum flow rate of 80 l/min (EN ISO 7396-2) or 130 l/min (BS 6834) with a 1 kPa resistance to flow, and a minimum of 50 l/min (EN ISO 7396-2) or 80 l/min (BS 6834) with a 2 kPa (EN ISO 7396-2) or 4 kPa (BS 6834) resistance to flow at each terminal unit, irrespective of the number of terminal units in use. The AGS system shall use dedicated radial blowers in a simplex or duplex configuration. The AGS pump assemblies shall be skid mounted and included on the skid shall be the simplex or duplex pump(s), motor control unit(s) with starter/isolator, moisture drain flask and flexible connector(s) to connect the plant to the pipeline. Each pump shall include an electric motor and directly coupled impeller assembly. Impeller bearings in the pump(s) shall not require lubrication. The pump(s) shall be air cooled and rated for continuous operation. Vacuum/Flow Regulating Valve A vacuum/flow regulating valve shall be provided and positioned at the pump, comprised of a spring-loaded plate valve and inlet silencer. The valve should be changeable with the pipeline inlet in order to provide flexibility on site. The plate shall control air ingress into the pipeline system, thereby controlling the vacuum level within. An optional air inlet filter shall be available should the air quality be poor/dusty offering further protection against dirt ingress into the pump. Additional in line vacuum/flow regulating valves may be installed if required and shall be determined by the pipeline designer. The vacuum/flow regulating valve shall ensure a maximum vacuum of 200mb below atmospheric pressure is not exceeded and shall be factory preset at 150mb. Control System Each motor control panel shall incorporate an emergency panel isolation switch facility, which controls all electrical power to the exhauster unit, remote start switch panels and system indication lights. All control and status indication circuitry shall be limited to 24V a.c. A green ‘POWER ON’ indicator shall be fitted to the starter/ isolator panel, and shall illuminate whenever power is available to the 24V control and indication circuit. A ‘HAND/OFF/AUTO’ switch shall be provided to control operation of the pump, running the pump continuously when selected to ‘HAND’. When selected to ‘AUTO’, control of the pump shall be passed to the remote start switch panels. Operation of any of the remote start switches shall activate the pump. The pump shall continue to run until all remote switches are selected ‘OFF’. ANAESTHETIC GAS SCAVENGING In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. Simplex starter/isolator panels c/w alarm pressure switch and duplex units incorporate line pressure switch. This line pressure switch monitors vacuum levels and provides an additional control of the remote start switch and starter/isolator panel green ‘RUNNING’ indicators. The pressure switch shall also include a digital display providing an accurate readout of the vacuum level in the pipeline in order to assist with installation/commissioning and annual re-commissioning. Simplex installations shall use remote start switches that include a red ‘PLANT EMERGENCY’ indicator. This indicator shall illuminate on all remote start switch panels if the vacuum level falls below the pressure switch set point level when the pump has been called, or if the overload trips. The on/off rocker switch shall include a green illuminated surround to indicate ‘mains on’. Duplex installations shall use remote start switches that include an amber ‘PLANT FAULT’ indicator. This shall illuminate, if either pump is set to ‘HAND’, or if one of the overloads trip. A red ‘PLANT EMERGENCY’ indicator shall also be provided and shall illuminate on all remote start switch panels if the vacuum level falls below the pressure switch set point level when the pump has been called. The on/off rocker switch shall include a green illuminated surround to indicate ‘mains on’. Where a duplex system is installed each pump shall be controlled by a separate motor control panel to enable servicing of either pump or control gear whilst maintaining system operation. Volt free relay kits for replicating alarm conditions to BMS shall be available as an optional extra. To be either installed either at factory or as a retro-fit kit for onsite installation. Terminal Units Terminal unit shall be provided with an adjustable orifice to allow balancing of the terminal unit flows during commissioning. Venturi style terminal units are not acceptable. Terminal units shall not be connected to the medical vacuum system. CE Marking The standard range of BeaconMedæs Anaesthetic Gas Scavenging System are ‘CE’ marked under the Medical Devices Directive 93/42/EEC with approval from notified body no. 0088 (Lloyd’s Register Quality Assurance). Under this directive, the specified products are classified as Class IIb Medical Devices. The starter/isolator panel shall incorporate a thermal protection overload device. The thermal protection overload device shall also monitor the electrical power supply and phase input. In the event of a fault, the overload device shall break the circuit to the pump, preventing operation until the system is manually re-set. Operation of the overload device shall also break the circuit to the remote start switch panels, extinguishing the green running indicator. BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982 Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450 2005208 In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. 16/03/12 Page 2 of 3 Issue 7 FLC/pump (A) Motor start current (A) Min. cable size (mm2) Motor rated fused supply/pump (A) Length (mm) Width (mm) Height (mm) Weight (kg) Motor plate (kW) Noise level per pump @ 1m dB(A) Connection (mm) 2.5 16 900 525 1160 96 0.75 60 54 400 3 3.3 23 2.5 16 900 525 1160 121 1.5 64 54 AGS2860-D/3 400 3 7.2 48 4 16 900 525 1160 139 3 69 54 4153 1250 00 AGS520-S/1 230 1 5 20 2.5 16 525 525 1160 56 0.75 60 54 4153 1251 00 AGS520-S/3 400 3 1.8 9 2.5 16 525 525 1160 56 0.75 60 54 4153 1252 00 AGS1560-S/3 400 3 3.3 23 2.5 16 525 525 1160 68 1.5 64 54 4153 1253 00 AGS2860-S/3 400 3 7.2 48 2.5 16 525 525 1160 77 3 69 54 4153 1261 00 AGS650-D/3 380 3 1.8 10 2.5 16 900 525 1160 96 0.9 64 54 4153 1262 00 AGS2210-D/3 380 3 3.3 23 2.5 16 900 525 1160 121 1.8 69 54 4153 1263 00 AGS3770-D/3 380 3 7.2 48 4 16 900 525 1160 139 3.6 70 54 4153 1257 00 AGS650-S/1 220 1 5.5 18 2.5 16 525 525 1160 56 0.9 64 54 4153 1258 00 AGS650-S/3 380 3 1.8 10 2.5 16 525 525 1160 56 0.9 64 54 4153 1259 00 AGS2210-S/3 380 3 3.3 23 2.5 16 525 525 1160 68 1.8 69 54 4153 1260 00 AGS3770-S/3 380 3 7.2 48 2.5 16 525 525 1160 77 3.6 70 54 50 Hz 60 Hz Duplex Simplex 4153 1256 00 Duplex AGS1560-D/3 Simplex 4153 1255 00 Supply frequency (Hz) 9 AGS520-D/3 Voltage (V) 1.8 4153 1254 00 Format 3 Model 400 Part number Phase ~ AGS Pump Assembly Details Notes… 1. Noise levels are averages, measured in accordance with ISO 517:1996 Free Air Aspiration and ISO 2151 Mean Sound Level 2. Service connections are copper to BS EN 13348 and relate to the external pipe diameter BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982 Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450 2005208 In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. 16/03/12 Page 3 of 3 Issue 7 Typical AGS Pump Assembly Layouts Duplex Pump Assembly 2 10 7 1 1 3 1160 7 8 Ø54 6 Ø54 10 ALT. 2 INLET 5 ALT. 1 INLET 8 9 9 40 445 525 15 870 900 2 CABLE ENTRANCE MOUNTING BY 4x M8 BOLT Simplex Pump Assembly Ø54 2 10 7 3 10 ALT. 1 INLET ALT. 2 INLET 9 1 1160 7 9 8 2 CABLE ENTRANCE 6 MOUNTING BY 4x M8 BOLT 5 8 40 445 525 15 REF. 1 2 3 4 5 6 7 8 9 10 495 525 DESCRIPTION FLEXIBLE HOSE FLOW/REGULATING VALVE PRESSURE SWITCH NON-RETURN VALVE DRAIN FLASK ISOLATION VALVE MOTOR CONTROL PANELS PUMPS INLET CONNECTION EXHAUST CONNECTIONS Note: Items 2 & 9, Flow regulating valve and inlet connection, can be re positioned on site if required BeaconMedæs Part of the Atlas Copco Group Telford Crescent, Staveley, Derbyshire, S43 3PF, England Tel: +44 (0) 1246 474 242 www.beaconmedaes.com 0088