BLOOD COMPONENTS Noninvasive pH monitoring of platelet concentrates: a large field test Effimia Gkoumassi, Christa Klein-Bosgoed, Margriet J. Dijkstra-Tiekstra, Dirk de Korte, and Janny de Wildt-Eggen BACKGROUND: Developing new quality control methods for platelet concentrates (PCs) can contribute to increasing transfusion safety and efficiency. The aim of this study was to investigate in a large field test the quality of expired PCs and whether 100% noninvasive pH monitoring can be used to predict PC quality. STUDY DESIGN AND METHODS: The pH of 13,693 PCs produced for transfusion was monitored daily using Blood Storage, Inc.’s pH sterile, automated fluoroscopic evaluation technology. Upon indication of compromised quality or expiration, PCs were returned and in vitro tests were performed. RESULTS: A total of 998 PCs were returned, of which 962 outdated, 26 had a positive BacT/ALERT reaction, seven had aggregates, one was without swirl, one had low pH, and one had high pH. BacT/ALERT was faster in identifying bacterial contamination than pH measurements. The pH at the end of the storage period was significantly lower than at the beginning. In vitro tests indicated that while the PC quality was acceptable upon expiration, it rapidly declined after expiration. CONCLUSION: In this setting where the vast majority of PCs were of good quality and within acceptable pH limits, daily, noninvasive routine pH measurement has limited added value in identifying quality-compromised PCs. P latelets (PLTs) are sensitive to pH variations and the ideal storage pH37°C is between 6.8 and 7.4, although acceptable storage limits are set as pH37°C > 6.3 (pH22°C 6.4).1,2 A pH below 6.4 is associated with a lower PLT viability.3 Rapid lactate accumulation can occur in a PLT concentrate (PC) with high PLT content and consequently pH can quickly decrease.4 A rapid decrease in pH can also be the consequence of bacterial contamination with acid-producing bacteria.5 Bacterial contamination of blood products can cause bacterial-induced transfusion-related sepsis, which in turn increases patient transfusion-related morbidity and mortality in patients. Despite recently introduced safety measures, bacterial contamination still represents one of the most frequent infectious risks of transfusion.6-8 Storage lesion of PLTs can also be induced through poor storage conditions, such as gas-impermeable storage bags, high or low temperatures, too rigorous shaking, or absence of shaking for more than 24 hours.3,9-13 Consequently, monitoring the pH and assuring that it is within acceptable limits could be of added value to blood banking. Most methods of pH measurement (blood gas analyzers [BGA], handheld pH meters, and colorimetric dipsticks) are invasive, expensive, laborious, and inaccurate (as in the case of dipsticks) or require a large sample volume, ABBREVIATIONS: BCSI = Blood Storage, Inc.; BGA(s) = blood gas analyzer(s); PC(s) = platelet concentrate(s); SAFE = sterile, automated fluoroscopic evaluation. From the Department Transfusion Monitoring, Sanquin Research, Groningen; and the Department Product and Process Development, Sanquin Blood Supply, Amsterdam, the Netherlands. Address reprint requests to: Effimia Gkoumassi, Department Transfusion Monitoring, Sanquin Research, PO Box 1191, 9701 BD Groningen, the Netherlands; e-mail: E.Gkoumassi@gmail.com. Received for publication July 25, 2012; revision received November 29, 2012, and accepted November 30, 2012. doi: 10.1111/trf.12099 TRANSFUSION 2013;53:2287-2292. Volume 53, October 2013 TRANSFUSION 2287 GKOUMASSI ET AL. resulting in product wastage. A new system, Blood Storage, Inc. (BCSI)’s pH sterile, automated fluoroscopic evaluation (SAFE) reader (also known as the pH 1000; BCSI, Seattle, WA), offers the ability to monitor pH in a noninvasive manner, which could be used on 100% of PCs to identify those of compromised quality.5,7,14,15 Over a period of 15 months (June 2009 to September 2010), a large field test was performed with PCs stored in a bag containing an integrated pH sensor, allowing noninvasive pH monitoring through the use of the BCSI pH SAFE reader until just before transfusion. Furthermore, the quality of PCs after expiration was investigated as well as whether pH could be used to identify PCs unsuitable for transfusion. MATERIALS AND METHODS PC preparation Whole blood–derived, multiple-donor PCs were made for clinical use, complying with European guidelines, using validated production protocols, as previously described.2,16 In short, PCs were made by pooling the buffy coats of five volunteer donors together with 1 plasma unit from one of those donors. Subsequently this pool was gently centrifuged (900rcf; ACE 1.27 ¥ 109, Break 3) in a centrifuge (Sorvall 12BP, Thermo Fisher Scientific, Inc., Waltham, MA) and the resulting PLT-rich plasma was slowly extracted with an automated separator (Compomat G4, Fresenius Kabi, Emmer-Compascuum, the Netherlands) into a polyvinylchloride-citrate storage bag with an integrated BCSI pH probe. The PC was stored in a shaking incubator (one cycle per second) at 22 ⫾ 2°C until transfusion (maximally 7 days after production) or in vitro testing (in case of expiration or recall; maximally 11 days after production). There was no quarantine time for the PCs before release. Quality control A bacterial detection system (BacT/ALERT 3D system, bioMérieux, Marcy l’Etoile, France) was used for bacterial contamination screening of all PCs, with incubation of 7.5 mL of concentrate in an aerobic and 7.5 mL in an anaerobic bottle for 7 days at 35°C. PLT concentration was measured using a hematology analyzer (Sysmex XT1800i, Toa, Japan) and used to calculate the amount of PLTs per concentrate in combination with the volume. Only PCs with more than 250 ¥ 109 PLTs, the presence of swirling effect, and negative-to-date for bacterial contamination were released for transfusion. From Day 2 onward and until transfusion, the pH of the PLTs was measured daily (if possible) using the BCSI pH SAFE reader. The measurements took place at the blood bank, three blood bank distribution centers, and four hospitals (Isala Klinieken, Zwolle; Medisch Centrum 2288 TRANSFUSION Volume 53, October 2013 Leeuwarden; Medisch Centrum Twente, Enschede; Universitair Medisch Centrum, Groningen, the Netherlands). PCs, either in the blood bank or in the hospital inventory, were sent to the research department if they exhibited absence of swirl, presence of aggregates, pH outside the acceptable range, or upon their expiration (7 days after blood collection). During the study, the acceptable pH range was pH37°C 6.3 to 7.3. Successfully recalled PCs with a positive BacT/ALERT were also sent to the research department for in vitro testing. In vitro testing All assays were performed with validated standard methods available in our laboratory. Of all PCs returned, the pH37°C was measured daily, on the day of return and on Day 9, 10, or 11, using the BCSI pH SAFE reader, as previously described.5,14,15 In short, the BCSI pH SAFE method uses a pH-sensitive fluorescent membrane fixed to a clear window inside a small sensor tube which is welded in the rim of the PC container. Fluorescence of the sensor membrane is measured through the window using a fiber optic probe and the BCSI pH SAFE reader calculates pH based on the ratio of yellow and red fluorescence measured. Returned PCs were sampled using a syringe and luerlock adapter and tested on a BGA (RapidLab, Siemens Medical Solutions Diagnostics BV, Breda, the Netherlands) for pH37°C, pO2, pCO2, glucose, and lactate concentration on Day 9, 10, or 11. The returned PCs were also tested for annexin V binding and CD62P surface expression using flow cytometry (FACSCalibur, Becton Dickinson, San Jose, CA), as previously described.17 In case of a successful recall due to an initial positive BacT/ALERT reaction, PCs were first aseptically sampled and then cultured for a second time in BacT/ALERT, before proceeding to in vitro testing. Only the PCs where the second sample gave a positive BacT/ALERT reaction were considered to be confirmed positively contaminated. Statistical analysis Results are presented as mean ⫾ standard deviation (SD). Significance was determined using unpaired t test, and analysis of variance (with Tukey posttest) with a p value of less than 0.05 used to indicate significance. All statistical tests were performed using commercially available software (Microsoft Excel, Microsoft Corp., Redmond, WA; Instat, GraphPad Software, La Jolla, CA). RESULTS Produced, returned, and recalled PCs In total, 13,693 PCs were produced during the period of the study. The distribution of PCs per study hospital was NONINVASIVE pH MONITORING OF PCs as follows: 20% to Isala Klinieken, Zwolle; 7% to Medisch Centrum, Leeuwarden; 30% to Medisch Centrum Twente, Enschede; and 43% to Universitair Medisch Centrum, Groningen. On the production day (Day 1) PLT count (341 ¥ 109 ⫾ 53 ¥ 109) and volume (346 ⫾ 17 mL) were in compliance with national guidelines. During the storage and distribution of the PCs (Day 2 to 7), there were 37,307 pH readings collected with the pH SAFE readers. Of all PCs, 998 were returned: 962 due to outdating, 26 recalled successfully due to an initial positive BacT/ ALERT reaction, seven due to presence of aggregates (pH was >7.0 and swirl normal in all seven cases), one due to absence of swirl, one due to low pH (<6.2), and one due to high pH (pH 7.42). No adverse transfusion reactions to PC transfusions or cases of transfusion-related sepsis were reported over the course of the study. Low and high pH37°C measurements Low pH (pH37°C < 6.3) was measured on the BCSI pH SAFE reader in only 10 (0.073%) of all PCs before expiration (Days 2-7) and four of these were also confirmed with positive bacterial contamination (0.029%). One of the PCs (0.007%) that developed a low pH37°C (pH37°C 6.58 on Day 3, return day; pH37°C < 6.3 on Day 6) contained a high amount of PLTs (629 ¥ 109 per bag) and was returned due to absence of swirl. One PC with a characteristic deviating brown color was returned due to low pH (on Day 7; BCSI pH SAFE value < 6.3). The following day the pH was determined on the BGA (pH 7.133), indicating that it still fell within the acceptable range, although not according to the BCSI pH SAFE system (pH < 6.3). No positive culture reaction was detected, swirl was normal, and all in vitro tests (performed in this case on Day 8) gave normal results. Consequently, it was speculated that the brown color interfered with the BCSI pH SAFE measurement. For two PCs (0.015%) low pH37°C measurements could be explained by handling error of the operator or mechanical fault of the apparatus, based on the fact that a repeated pH measurement gave normal results. Two cases were unexplainable. PLT content was not deviating and since the PCs were transfused no further in vitro tests were performed. No adverse transfusion reactions were reported. All five PCs with confirmed low pH (four with confirmed bacterial contaminations and one with high PLT content) were of extremely compromised quality and unsuitable for transfusion. Over the whole study, one PC was returned as a result of a high pH. On Day 4, BCSI pH SAFE indicated a pH value of 7.42, which was above the desirable upper limit. The swirl and in vitro tests (performed on Day 9) were normal and no interfering variables could be identified for this PC. after an initial increase (on Day 3 pH was 7.31 ⫾ 0.09). However, the decrease was only minimal from Day 4 to Day 7 (Fig. 1A). The main decrease was found after expiration, between Day 7 and Day 10 (Fig. 1A) but pH remained greater than 6.8 in most cases. On Day 11 the pH was significantly higher than on Day 10, probably because not all PCs were measured on all days. pH37°C BCSI pH SAFE versus pH37°C BGA Upon returning to the laboratory, the in vitro tests were performed and the pH37°C was measured on the same day using both the BCSI pH SAFE reader and the BGA. The pH measured with the BCSI pH SAFE reader (7.09 ⫾ 0.009) was significantly higher than the pH measured with the BGA (6.99 ⫾ 0.006; p < 0.05), implying that the BCSI pH SAFE readers used in this test slightly overestimated the pH. In vitro tests of expired PCs In vitro tests of expired PCs were performed in most cases on Day 9, 10, or 11. On Day 9 (n = 540) the mean CD62P expression and annexin V binding were significantly lower than on Day 10 (n = 79) and Day 11 (n = 250), but no significant difference was found between Day 10 and Day 11 (Fig. 2A). CD62P expression was greater than 50% for 15 (2.8% of PCs), 18 (23.4% of PCs), and 23 (9.7% of PCs) PCs on Days 9, 10, and 11, respectively. Annexin V binding was greater than 40% for 44 (8.2% of PCs), 24 (31.2% of PCs), and 69 (29.2% of PCs) PCs on Days 9, 10, and 11, respectively. In accordance with reduced but still active metabolism, pO2 was higher on Day 11 than on Day 9, pCO2 was significantly lower on Day 11 than on Days 9 and 10. Glucose was significantly higher and lactate concentration was significantly lower on Day 9 than on Days 10 and 11 (Fig. 2B). Lactate was greater than 20 mmol/L in 23% of the PCs on Day 9, 46% of the PCs on Day 10, and 54% of the PCs on Day 11. Glucose remained present (>5 mmol/L) during storage for all tested PCs. On Day 9, on average, the PCs exhibited swirl and acceptable pH; however, CD62P expression, annexin V binding, and lactate concentration were high. In total, at least 125 PCs were of compromised quality on Day 9 (based on in vitro tests). If Day 7 pH was less than 7.1, CD62P expression, annexin V binding, and lactate concentration on Day 9 were higher than for units with pH values of greater than 7.1 (n = 12). However, due to the low sample size, it cannot be claimed that pH on Day 7 is predictive of the PC quality on Day 9. pH progress during storage BacT/ALERT and pH In accordance with previous studies, pH decreased during storage from pH 7.27 ⫾ 0.08 on Day 2 to pH 7.08 ⫾ 0.21 Of all PCs, 48 (0.351%) gave an initial positive culture reaction of which 26 were successfully recalled. Of these 26 Volume 53, October 2013 TRANSFUSION 2289 GKOUMASSI ET AL. A 7.4 7.2 pH37°C 7.0 6.8 6.6 6.4 6.2 2 9993 3 8964 4 6871 5 5435 6 4060 7 1984 8 795 9 598 10 96 11 567 Day Units tested per day (n) B 7.4 pH37°C 7.2 * * * ** 7.0 * 6.8 6.6 6.4 6.2 2 3 4 5 6 7 8 9 Day Fig. 1. (A) The pH37°C during storage in “normal” PCs (mean ⫾ SD). Not all PCs were measured all days. X-axis: top row, PC age in days; bottom row, number per day. (B) Mean pH decrease during storage in normal PCs (䉱; number as in A), PCs that gave a positive BacT/ALERT reaction (•; n = 48) and PCs that were confirmed positive (䊏; n = 9).*p < 0.05 normal PCs versus BacT/ALERT positive PCs. PCs, nine (0.066%) were confirmed positive by a second culture, with all confirmed positive cases originating from skin flora microorganisms. The pH remained within the acceptable range on the day of the first positive culture reaction; however, in vitro tests indicated deviating values (data not shown). Three PCs gave a positive reaction within 24 hours of inoculation (Day 2) and one of these also exhibited pH37°C of less than 6.3 (Staphylococcus aureus). In total, the pH of four bacterially infected PCs (Bacillus sp., S. aureus, S. aureus, and hemolytic Streptococcus) dropped to less than 6.3 on Days 7, 7, 2, and 4, respectively (not shown). BacT/ALERT was approximately 3 days faster in identifying suspected contamination than pH measurements, although it must be kept in mind that BacT/ALERT is cul2290 TRANSFUSION Volume 53, October 2013 turing a sample at elevated temperatures whereas pH is measured in the real product to be transfused. It was observed that pH decrease in PCs with a positive BacT/ ALERT reaction, whether confirmed positive or not, was significantly more rapid than in the normal PCs (Fig. 1B). The pH of confirmed positive bacterially contaminated PCs appears to decline drastically but due to the low frequency of confirmed contaminations, the difference is not significant (Fig. 1B). DISCUSSION During this large field study for monitoring pH of PCs the great majority of all PCs were of high quality, meeting European standards. Noninvasive pH measurement could help with monitoring the quality of PCs containing high initial amounts of PLTs, taking into account that a high NONINVASIVE pH MONITORING OF PCs A new software upgrade available for the BCSI pH SAFE reader, developed 50 after the end of this study (and partly as a * result of this study), makes user calibra* 40 tion to a reference (e.g., BGA) possible and could potentially increase accuracy. 30 It has been suggested recently 20 that present PC quality control (QC) methods, including bacterial contami10 nation screening are insufficient, often resulting in product wastage and conse0 quently new techniques should be D9 D10 D11 D9 D10 D11 employed to further reduce bacterial B risk in PCs.20 Bacterial screening systems cannot 25 * be substituted with pH measurements, however, and while on the one hand it is 20 * of interest that pH trending of potentially contaminated PCs was found to be † 15 distinctly different than normal PCs, on the other hand, very low concentrations 10 of microorganisms in PCs could poten‡ † tially grow in BacT/ALERT (at 35°C) and 5 result in an initial positive reaction but fail to survive in the PC which is stored at a lower temperature. 0 D9 D10 D11 D9 D10 D11 D9 D10 D11 D9 D10 D11 The BCSI pH SAFE system could be deemed suitable for use until just Fig. 2. In vitro tests of expired PCs (mean ⫾ SD). (A) Percentage CD62P membrane before transfusion by hospital staff, but expression (䉬) and annexin V (䊏) binding. (B) pO2 in kPa (䉬), pCO2 in kPa (䊐), not as a blood bank monitoring instruglucose in mmol/L (䉱), and lactate content in mmol/L (䊊) during Days 9, 10, and 11. ment. However, it must be kept in mind *p < 0.05 Day 9 versus Days 10 and 11. †p < 0.05 Day 9 versus Day 11. ‡p < 0.05 Day that the overall process of whole blood 10 versus Day 9 and Day 11. Not all PCs were measured all days. collection and processing of PCs was very standardized in this study, leading to restricted PLT concentration can lead to quicker depletion of variation in the volume and PLT content of the PCs. This glucose and a rapid increase of lactate, resulting in lower was reflected in the results of the study, with only a pH and reduced PLT viability. The amount of PCs prominimal amount of aberrant pH values in the PCs during duced with high PLT concentrations was limited in the storage for up to 7 days. The use of the pH SAFE techcurrent study, but earlier studies have shown this correlanology confirmed that the routine production and distrition.4 However, rapid deterioration of PCs containing high bution setting of this study is well controlled and capable amounts of PLTs can also be prevented by releasing (and of producing PCs of high quality, with very few PCs with transfusing) such PCs more quickly, but that requires low pH values during their 7-day storage. Under these 100% counting of PCs. conditions, daily pH monitoring does not have added All in vitro variables measured on Days 9, 10, and 11 value to currently implemented QC tools (such as swirl were in line with the reducing quality of the PCs and comand BacT/ALERT) in identifying PCs unsuitable for transparable with what is previously reported for PLTs of this fusion. Under less standardized conditions, daily pH age.17 monitoring might identify more deviating units and be of The BCSI pH SAFE system was user-friendly and did additional value. not result in product wastage when compared to BGAs. However, pH measurements with BGAs were more accurate and were comparable to previous studies. ACKNOWLEDGMENTS An AABB Standard characterizes pH, glucose concentration measurement, and microscopy as less sensitive The authors thank the employees of the transfusion monitoring, methods for detecting bacterial contamination of PCs and blood processing and distribution, and client service departconsequently discourages their use, which is confirmed by ments of Sanquin Blood Supply in Groningen for their collaboraour results.18,19 tion and E.J. Klip and S. Geelhood from BCSI for their advice. pO2 (kPa), pCO2 (kPa), glucose (mmol/L) and lactate (mmol/L) % Annexin V binding % CD62P expression A Volume 53, October 2013 TRANSFUSION 2291 GKOUMASSI ET AL. CONFLICT OF INTEREST At the time of this work DdK was a member of the scientific advisory committee of Blood Cell Storage, Inc., Seattle, WA (BCSI). [Correction added after online publication 30-Jan-2013. The company name has updated.] This committee is presently inactive. The remaining authors declare that they have no conflicts of interest relevant to the manuscript submitted to TRANSFUSION. 9. Murphy S, Gardner FH. Effect of storage temperature on maintenance of platelet viability—deleterious effect of refrigerated storage. N Engl J Med 1969;280:1094-8. 10. Murphy S, Sayar SN, Gardner FH. Storage of platelet concentrates at 22 degrees C. Blood 1970;35:549-57. 11. Slichter SJ, Harker LA. Preparation and storage of platelet concentrates. II. Storage variables influencing platelet viability and function. Br J Haematol 1976;34:403-19. 12. Tynngard N. Preparation, storage and quality control of REFERENCES 1. Dekkers DW, De Cuyper I, van der Meer PF, Verhoeven AJ, de Korte D. Influence of pH on stored human platelets. Transfusion 2007;47:1889-95. 2. Council of Europe. Guide to the preparation, use and platelet concentrates. Transfus Apher Sci 2009;41: 97-104. 13. van der Meer PF, de Korte D. Platelet preservation: agitation and containers. Transfus Apher Sci 2011;44:297304. 14. Krause M, Doescher A, Zimmermann B, Muller TH. Nonin- quality assurance of blood components. Reccomendation vasive pH measurement to monitor changes during subop- R (95) 15. 16th ed. Strasbourg: Council of Europe Publishing; 2010. timal storage of platelet concentrates. Transfusion 2010;50: 2185-92. 3. Murphy S, Gardner FH. Platelet storage at 22 degrees C: 15. Reed MW, Geelhood S, Barker LM, Pfalzgraf R, Vlaar R, role of gas transport across plastic containers in maintenance of viability. Blood 1975;46:209-18. 4. Dijkstra-Tiekstra MJ, Kuipers W, Setroikromo AC, de Wildt- Gouwerok E, De Cuyper I, Harris P, Verhoeven AJ, de Korte D. Noninvasive measurement of pH in platelet concentrates with a fiber optic fluorescence detector. Transfusion Eggen J. Platelet capacity of various platelet pooling systems for buffy coat-derived platelet concentrates. Transfusion 2008;48:2114-21. 2009;49:1233-41. 16. Dijkstra-Tiekstra MJ, Kuipers W, Setroikromo AC, de WildtEggen J. Overnight or fresh buffy coat-derived platelet con- 5. Barker LM, Nanassy OZ, Reed MW, Geelhood SJ, Pfalzgraf RD, Cangelosi GA, de Korte D. Multiple pH measurement during storage may detect bacterially contaminated platelet concentrates. Transfusion 2010;50:2731-7. 6. de Korte D, Curvers J, de Kort WL, Hoekstra T, van der Poel CL, Beckers EA, Marcelis JH. Effects of skin disinfection method, deviation bag, and bacterial screening on clinical safety of platelet transfusions in the Netherlands. Transfusion 2006;46:476-85. 7. de Korte D. 10 years experience with bacterial screening of platelet concentrates in the Netherlands. Transfus Med Hemother 2011;38:251-4. 8. Hillyer CD, Josephson CD, Blajchman MA, Vostal JG, Epstein JS, Goodman JL. Bacterial contamination of blood components: risks, strategies, and regulation: joint ASH and AABB educational session in transfusion medicine. Hematology Am Soc Hematol Educ Program 2003;575-89. 2292 TRANSFUSION Volume 53, October 2013 centrates prepared with various platelet pooling systems. Transfusion 2008;48:723-30. 17. Dijkstra-Tiekstra MJ, Pietersz RN, Huijgens PC. Correlation between the extent of platelet activation in platelet concentrates and in vitro and in vivo parameters. Vox Sang 2004;87:257-63. 18. American Association of Blood Banks. Association Bulletin #10-05—suggested options for transfusion services and blood collectors to facilitate implementation of BB/TS interim standard 5.1.5.1.1. Bethesda (MD): American Association of Blood Banks; 2010. 19. American Association of Blood Banks. Association Bulletin #10-02—interim standard 5.1.5.1.1. Bethesda (MD): American Association of Blood Banks; 2010. 20. Tomasulo P, Su L. Is it time for new initiatives in the blood center and/or the hospital to reduce bacterial risk of platelets? Transfusion 2011;51:2527-33.