Uploaded by zury1823

MONITOR DE SIGNOS VITALES VITAL GUARD 405 C

advertisement
Vital-Guard 450C
Patient Monitor
®
SpO2
2684-01-20
with Masimo SET
Operation and Maintenance Manual
User Responsibility
This product will perform in conformity with the description thereof contained in this Operation and Maintenance
Manual and accompanying labels and/or inserts, when assembled, operated, maintained and repaired in accordance
with the instructions provided. This product must be checked periodically. A defective Product should not be used.
Parts that are broken, missing, plainly worn, distorted or contaminated should be replaced immediately. Should such
repair or replacement become necessary, IVY Biomedical Systems, Inc. recommends that a telephone call or written
request for service advice be made to IVY Biomedical Systems, Inc. Service Department. This product or any of its
parts should not be repaired other than in accordance with instructions provided by IVY Biomedical Systems, Inc.
trained personnel. The product must not be altered without the prior written approval of IVY Biomedical Systems,
Inc. Quality Assurance Department. The user of this Product shall have the sole responsibility for any malfunction,
which results from improper use, faulty maintenance, improper repair, damage or alteration by anyone other than
IVY Biomedical Systems, Inc.
CAUTION: US Federal law restricts this device to sale by or on the order of a licensed medical practitioner.
Ivy Biomedical Systems, Inc. has declared that this product conforms with the Eurpean Council Directive 93/42/EEC
Medical Device Directive when its used in accordance with the instructions provided in the Operation and
Maintenance Manual.
Ivy Biomedical Systems, Inc.
11 Business Park Drive
Branford, Connecticut 06405
(203) 481-4183  (800) 247-4614  FAX (203) 481-8734
www.ivybiomedical.com e-mail:ivybio@ivybiomedical.com
The following are trademarks of Masimo Corporation: Masimo®, Masimo SET® and LNOP®.
Covered by one or more of the following U.S. Patents: 5,482,036; 5,490,505; 5,632,272; 5,685,299; 5,758,644; 5,769,785; 6,002,952;
6,036,642; 6,067,462; 6,206,830; 6,157,850 and international equivalents. U.S.A. and international patents pending.
Possession or purchase of this device does not convey any express or implied license to use the device with replacement parts which would,
alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.
S450C-4May2010
2684-01-20 Rev.03
Declaration of Conformity
Manufacturer:
Ivy Biomedical Systems, Inc.
11 Business Park Drive
Branford, CT 06405
Authorized Representative:
Cavendish Scott Ltd.
Starlings Bridge, Nightingale Road
Hitchin, Herts, SG5 1FW, England
Type of Equipment:
Physiological Monitors
Models:
450C
We, Ivy Biomedical Systems, Inc., hereby declare that the devices mentioned above
comply with the Swedish National Board of Health and Welfare Regulation and
guidelines on medical devices LVFS 2003:11 (M) 28 October 1994 – transposing
European Medical Devices Directive 93/42/EEC.
Date of Validity:
March 30, 2010
Classification:
IIb
Conformity Assessment
Procedure:
Notified Body:
10
Annex II
Intertek SEMKO AB
Name of Authorized Signatory:
Position held in Company:
Signature
According to rule No.
Notified Body No.
Dick Listro
Director of Regulatory
0413
This page is intentionally left blank.
TABLE OF CONTENTS
TABLE OF CONTENTS
WARRANTY ............................................................................................................................................................... v
INTRODUCTION ....................................................................................................................................................... 1
SAFETY ....................................................................................................................................................................... 2
Electrical ............................................................................................................................................................. 2
Explosion ............................................................................................................................................................. 2
Patient Connections ........................................................................................................................................... 3
MRI ..................................................................................................................................................................... 3
Pacemakers ......................................................................................................................................................... 3
Electrosurgery .................................................................................................................................................... 3
Defibrillation Protection .................................................................................................................................... 3
EMC .................................................................................................................................................................... 3
Electromagnetic Compatibility IEC 60601-1-2:2001 ..................................................................................... 4
Description of Warning Labels ......................................................................................................................... 8
For monitors used in Australia ......................................................................................................................... 8
MONITOR DESCRIPTION ....................................................................................................................................... 9
Classification....................................................................................................................................................... 9
Specifications .................................................................................................................................................... 10
Controls and Indicators ................................................................................................................................... 16
Display ............................................................................................................................................................... 20
Alarms ............................................................................................................................................................... 21
Priority Alarms Feature (If Present) .............................................................................................................. 24
Rear Panel......................................................................................................................................................... 24
Menu Structure ................................................................................................................................................ 25
The PATIENT SETUP Key......................................................................................................................... 26
The SYSTEM SETUP Key .......................................................................................................................... 27
MONITOR SETUP ................................................................................................................................................... 28
Adult/Neonate Modes ...................................................................................................................................... 28
Heart Rate Source ............................................................................................................................................ 29
Trace Speed ...................................................................................................................................................... 29
Set Time and Date ............................................................................................................................................ 29
Display Setup .................................................................................................................................................... 30
Data and Waveforms Color Setup .................................................................................................................. 30
Audio Setup ...................................................................................................................................................... 31
Display Limits ................................................................................................................................................... 31
Trend Displays.................................................................................................................................................. 31
Default Settings ................................................................................................................................................ 33
BATTERY OPERATION ......................................................................................................................................... 34
Internal Battery ................................................................................................................................................ 34
Status of the Battery ........................................................................................................................................ 34
Removing the Battery ...................................................................................................................................... 34
Vital-Guard 450C Operation and Maintenance Manual
i
TABLE OF CONTENTS
RECORDER OPERATION ..................................................................................................................................... 35
Changing Paper ................................................................................................................................................ 35
Recorder Menus ............................................................................................................................................... 36
Recorder Modes ............................................................................................................................................... 36
Recorder Messages ........................................................................................................................................... 39
Example Printouts ............................................................................................................................................ 39
ECG MONITORING ................................................................................................................................................ 42
Safety Considerations ...................................................................................................................................... 42
Patient Connections ......................................................................................................................................... 43
ECG Waveform Amplitude ............................................................................................................................. 44
Heart Rate (HR) Alarm Limits ....................................................................................................................... 44
Lead Selection ................................................................................................................................................... 44
ECG Filter (Adult Mode Only) ....................................................................................................................... 44
ECG On/Off ...................................................................................................................................................... 45
Pacemakers ....................................................................................................................................................... 45
RESPIRATION MONITORING ............................................................................................................................. 46
Respiration Size ................................................................................................................................................ 46
Apnea Delay ...................................................................................................................................................... 47
Coincidence On/Off.......................................................................................................................................... 47
Respiration Rate (RR) Alarm Limits.............................................................................................................. 47
Respiration On/Off .......................................................................................................................................... 47
END TIDAL CO2 MONITORING........................................................................................................................... 48
General Description ......................................................................................................................................... 48
Preparation for Monitoring ............................................................................................................................ 53
Response Setting ............................................................................................................................................... 55
Compensation Settings ..................................................................................................................................... 56
Barometric Pressure ...................................................................................................................................... 56
N2O Compensation ........................................................................................................................................ 56
O2 Compensation ........................................................................................................................................... 56
Trace Scale Setting ........................................................................................................................................... 57
CO2 Alarm Limits ............................................................................................................................................ 57
Respiration Rate Alarm Limits ....................................................................................................................... 57
Respiration On/Off .......................................................................................................................................... 57
Gas Calibration ................................................................................................................................................ 58
INVASIVE PRESSURE MONITORING ................................................................................................................ 59
Pressure Transducer ........................................................................................................................................ 59
Connecting the Transducer ............................................................................................................................. 59
Zeroing the Transducer ................................................................................................................................... 60
Blood Pressure Scale ........................................................................................................................................ 60
Pressure Alarm Type ....................................................................................................................................... 61
Pressure Alarm Limits ..................................................................................................................................... 61
Expanded Pressure Scale ................................................................................................................................. 61
Alternate Pressure Site Labeling and Automatic Scale/Limit Adjustment Feature (If Present) .............. 62
MASIMO SET® PULSE OXIMETRY MONITORING ........................................................................................ 63
Overview ........................................................................................................................................................... 63
Principles of Operation .................................................................................................................................... 63
Pulse Oximetry Sensors ................................................................................................................................... 64
Monitoring Procedure ..................................................................................................................................... 64
SpO2 Alarm Limits........................................................................................................................................... 66
ii
Vital-Guard 450C Operation and Maintenance Manual
TABLE OF CONTENTS
Response Mode Settings .................................................................................................................................. 66
Default Values .................................................................................................................................................. 66
Pulse Oximetry with ECG Off ........................................................................................................................ 67
Pulse Oximetry Considerations ....................................................................................................................... 67
NIBP MONITORING ............................................................................................................................................... 68
General Description ......................................................................................................................................... 68
NIBP Monitoring Considerations ................................................................................................................... 68
Cuff Placement ................................................................................................................................................. 69
Monitoring Procedure ..................................................................................................................................... 69
NIBP Graph ...................................................................................................................................................... 70
NIBP Messages ................................................................................................................................................. 71
Tabular Scrolling ............................................................................................................................................. 71
Clear Tabular Data .......................................................................................................................................... 72
NIBP Mode Select ............................................................................................................................................ 72
NIBP Alarm Settings........................................................................................................................................ 72
NIBP Calibration ............................................................................................................................................. 72
TEMPERATURE MONITORING .......................................................................................................................... 73
Temperature Alarms Description ................................................................................................................... 73
ALARM MESSAGES ............................................................................................................................................... 74
MONITOR TESTING............................................................................................................................................... 75
MAINTENANCE AND CLEANING ....................................................................................................................... 76
The Monitor ...................................................................................................................................................... 76
Patient Cables ................................................................................................................................................... 76
Reusable Pressure Transducer........................................................................................................................ 76
Cleaning and reuse of SpO2 sensors ............................................................................................................... 77
Reusable Temperature Probes ........................................................................................................................ 77
ETCO2 Sensor and Airway Adapter ............................................................................................................... 77
Preventive Maintenance .................................................................................................................................. 77
ACCESSORIES ......................................................................................................................................................... 79
ECG/Respiration .............................................................................................................................................. 79
EtCO2................................................................................................................................................................ 79
Invasive Pressure.............................................................................................................................................. 79
Non-Invasive Pressure (NIBP) ........................................................................................................................ 80
Temperature ..................................................................................................................................................... 80
Recorder............................................................................................................................................................ 80
SpO2 Reusable sensors ...................................................................................................................................... 81
SpO2 Disposable sensors ................................................................................................................................... 81
SpO2 Patient cables ........................................................................................................................................... 81
SpO2 Starter kits ............................................................................................................................................... 81
Disposal ............................................................................................................................................................. 81
REAR PANEL OUTPUTS ........................................................................................................................................ 82
Analog Output Connector ............................................................................................................................... 82
COM1 & COM2 Connectors .......................................................................................................................... 83
RJ45 Connector (option in place of COM2) .................................................................................................. 83
Configuration Menu ........................................................................................................................................ 84
Vital-Guard 450C Operation and Maintenance Manual
iii
TABLE OF CONTENTS
TESTING AND TROUBLESHOOTING ................................................................................................................ 85
General .............................................................................................................................................................. 85
ECG Test ........................................................................................................................................................... 85
Pressure Test .................................................................................................................................................... 86
Respiration Test ............................................................................................................................................... 87
Temperature Test ............................................................................................................................................. 87
Pulse Oximeter Test ......................................................................................................................................... 87
EtCO2 Test ....................................................................................................................................................... 88
NIBP Test .......................................................................................................................................................... 88
Recorder Test ................................................................................................................................................... 88
Battery Test ...................................................................................................................................................... 89
Isolation and Leakage Tests ............................................................................................................................ 89
System Error Messages.................................................................................................................................... 90
Simplified Block Diagram................................................................................................................................ 91
REPLACEMENT PARTS ........................................................................................................................................ 92
Printed Circuit Boards .................................................................................................................................... 92
Modules and Assemblies .................................................................................................................................. 92
Fuses .................................................................................................................................................................. 92
Cables and Harnesses ...................................................................................................................................... 93
Labels ................................................................................................................................................................ 93
Software ............................................................................................................................................................ 93
DISASSEMBLY AND ASSEMBLY ....................................................................................................................... 94
Covers............................................................................................................................................................... 94
Printed Circuit Boards ................................................................................................................................... 95
450C Mother (5143) Board ........................................................................................................................... 95
EtCO2 Interface (5108) Board ...................................................................................................................... 95
EtCO2 Module .............................................................................................................................................. 95
Masimo SpO2 Module .................................................................................................................................. 95
Isolated Input (5144) Board .......................................................................................................................... 96
Front Panel ....................................................................................................................................................... 96
EOS Power Supply ........................................................................................................................................... 97
Recorder Assembly .......................................................................................................................................... 97
NIBP Module ................................................................................................................................................... 97
Battery Removal ............................................................................................................................................... 97
Display Assembly.............................................................................................................................................. 98
Speaker.............................................................................................................................................................. 98
BOARD LAYOUT DIAGRAMS ............................................................................................................................... A
SCHEMATIC DIAGRAMS....................................................................................................................................... B
iv
Vital-Guard 450C Operation and Maintenance Manual
WARRANTY
WARRANTY
All products manufactured by Ivy Biomedical Systems, Inc. are warranted to be free from defects in material and
workmanship and to operate within published specifications, under normal use, for a period of one year from date of
original shipment.
All accessories supplied by Ivy Biomedical Systems, Inc. are warranted to be free from defects in material and
workmanship and to operate within published specifications, under normal use, for a period of 90 days from date of
original shipment.
If an examination by Ivy Biomedical Systems, Inc. discloses such products or component parts to have been
defective, then our obligation is limited to repair or replacement (at our option). Fuses and batteries are not covered
under this warranty.
If products need to be returned to the manufacturer for repair or examination contact customer service personnel at
IVY Biomedical Systems, Inc. to obtain the Return Authorization Number (RMA) and proper packing instructions.
Vital-Guard 450C Operation and Maintenance Manual
v
WARRANTY
This page is intentionally left blank.
vi
Vital-Guard 450C Operation and Maintenance Manual
INTRODUCTION
INTRODUCTION
This manual is to provide information on the correct use of the Vital-Guard 450C patient monitor. It is up to the
user to ensure that any applicable regulations regarding the installation and operation of the monitor are observed.
The Information regarding circuit diagrams, board layouts and replacement parts is contained in the Operation and
Maintenance Manual. For instructions on how to obtain the Operation Manual or the Operation and Maintenance
Manual please contact customer service personnel at IVY Biomedical Systems, Inc.
The model 450C is to be operated by qualified personnel only.
Using This Manual
We recommend that you read this entire manual before operating the equipment. This manual is written to include
all parameters for a fully configured monitor. If the configuration of your monitor does not include all parameters,
menu selections and display data for those parameters will not appear on your monitor.
Use the Monitor Description section for general descriptions of controls and displays. For details on monitoring
each parameter, refer to the section of the manual dealing with the appropriate parameter.
NOTE: If the monitor beeps when you press a key, it indicates an invalid selection for the current
mode or function.
Boldface type is used in text to refer to the labeling on user controls. Special brackets [
selections used with the programmable keys.
] surround menu
Manufacturer’s Responsibility
The manufacturer of this equipment is responsible for the effects on safety, reliability, and performance of the
equipment only if:
•
Assembly operations, extensions, re-adjustments, or repairs are carried out by persons authorized by the
manufacturer
•
The electrical installation complies with all applicable regulations
•
The equipment is used in accordance with the instructions in this manual
Incorrect operation or failure of the user to maintain the monitor in accordance with proper maintenance procedures
relieves the manufacturer or his agent from all responsibility for consequent non-compliance, damage, or injury.
For technical and service information, please refer to the Vital-Guard 450C Operation and Maintenance Manual
or contact:
Ivy Biomedical Systems, Inc.
11 Business Park Drive
Branford, Connecticut 06405
(203) 481-4183 or (800) 247-4614
e-mail: techline@ivybiomedical.com
This manual explains how to set up and use the Model 450C. Important safety information relating to general use of
the pulse oximeter appears in this manual. Other important safety information is located throughout the manual
where appropriate. READ THE ENTIRE SAFETY INFORMATION SECTION BEFORE YOU OPERATE THE
MONITOR.
Vital-Guard 450C Operation and Maintenance Manual
1
SAFETY
SAFETY
Electrical
This product is intended to be operated from a mains power source of nominally 100 to 230 V ~, 47-63 Hz, or
from its internal battery.
WARNING: To prevent electrical hazards to all personnel, this monitor must be properly grounded. Connect the
monitor only to a three-wire, grounded, hospital grade receptacle. The three-conductor plug must be inserted into a
properly wired three-wire receptacle; if a three-wire receptacle is not available, a qualified electrician must install
one in accordance with the governing electric code.
WARNING: Do not under any circumstances remove grounding conductor from the power plug.
WARNING: The power cable supplied with this equipment provides for this protection. Do not attempt to defeat
this protection by modifying the cable or by using ungrounded adapters or extension cables. The power cord and
plug must be intact and undamaged. To disconnect the equipment from the mains power unplug the power cord.
WARNING: Do not connect to an electrical outlet controlled by a wall switch or dimmer.
WARNING: If there any doubt about the integrity of the protective earth conductor arrangement, operate the
monitor on internal battery power until the AC power source protective conductor is fully functional.
WARNING: Do not place the monitor in any position that may cause it to fall on the patient. Do not lift the monitor
by the power supply cord or patient cable, use only the handle on the monitor.
WARNING: Electric shock hazard! Do not remove covers or panels. Refer service to qualified service
personnel.
WARNING: To avoid electrical shock, disconnect the monitor from its power source before changing fuses.
Replace fuses only with same type and rating (1.6ASB, Metric 5x20mm, 250V).
WARNING: Do not clean monitor while it is on and/or plugged into a power source.
WARNING: If unit is accidentally wet, discontinue use until dry and then test unit for proper operation before reuse
on patient.
WARNING: This unit uses a common isolation path for the ECG leads. Do not connect any non-isolated
accessories to the ECG input when connected to a patient, as this may compromise the safety of the unit. When
attached to other devices, insure that the total chassis leakage currents of all units do not exceed 300 μA.
Explosion
DANGER: Explosion hazard! Do not use this equipment in the presence of flammable anesthetics or other
flammable substance in combination with air, oxygen-enriched environment or nitrous oxide.
2
Vital-Guard 450C Operation and Maintenance Manual
SAFETY
Patient Connections
Patient connections are electrically isolated. For all connections use isolated probes. Don’t let patient connections
contact other conductive parts, including earth. See instructions for patient connections in this manual.
Carefully route patient cables to reduce the possibility of patient entanglement or strangulation.
Leakage current is limited internally by this monitor to less than 10 μA. However, always consider cumulative
leakage current that can be caused by other equipment used on the patient at the same time as this monitor.
To ensure that the leakage current protection remains within the specifications, use only the patient cables specified
in this manual. This monitor is supplied with protected lead wires. Do not use cables and leads with unprotected
lead wires having exposed conductors at the cable end. Unprotected lead wires and cables may pose an
unreasonable risk of adverse health consequences or death.
If an alarm condition occurs while the alarms are set to off, the only alarm indication will be visual displays and
symbols related to the alarm condition.
MRI
The model 450C should not be used within the magnetic field during Magnetic Resonance Imaging.
Pacemakers
Rate meters might continue to count the pacemaker rate during occurrences of cardiac arrest or some arrhythmias.
Do not rely on rate meter alarms. Keep pacemaker patients under close surveillance.
Electrosurgery
To avoid the potential of electrosurgery burns at monitoring sites, ensure proper connection of the electrosurgery
return circuit as described by manufacturer’s instructions. If improperly connected, some electrosurgery units might
allow energy to return through the ECG electrodes, SpO2 sensors, Pressure Transducers, EtCO2 sensors, and
Temperature probes. To further reduce the potential of burns, locate the SpO2 sensors, Pressure Transducers, EtCO2
sensors, and Temperature probes as described in the manufacturer’s instructions.
Defibrillation Protection
This equipment is protected against 360 J discharge and electrosurgery potentials. The monitor is internally
protected to limit current through the electrodes to prevent injury to the patient and damage to the equipment as long
as the defibrillator is used in conformance with the manufacturer’s instructions.
EMC
This equipment has been certified to be protected to emissions and immunity according to IEC-60601-1-2.
Vital-Guard 450C Operation and Maintenance Manual
3
SAFETY
Electromagnetic Compatibility IEC 60601-1-2:2001
CAUTION: Medical Equipment needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided in the Operation Manual.
CAUTION: Portable and mobile RF communications equipment can affect medical electrical equipment.
WARNING: The model 450C should not be used adjacent to or stacked with other equipment and that if adjacent or
stacked use is necessary, the Model 450C should be observed to verify normal operation in the configuration in
which it will used.
Accessories
WARNING: The use of accessories other than those specified below may result in increased emissions or decreased
immunity of the equipment.
Part Number
Description
590170 Three lead patient cable
590162 Set of three lead wires
590167 Five lead patient cable
590168 Set of five lead wires
590198 Set of three lead wires for prewired electrodes
590317 Low noise three lead patient cable
590318 Set of three radiotranslucent lead wires
590323 Low noise three lead patient cable with 1kohm resistors.
The minimum amplitude or value patient physiological signal is 0.5 mV (AAMI EC-13 3.2.6.1).
WARNING: The use of the Model 450C below the following amplitude values may cause inaccurate results:
4
Vital-Guard 450C Operation and Maintenance Manual
SAFETY
Guidance and manufacturer’s declaration – Electromagnetic emissions
The Model 450C is intended for use in the electromagnetic environment specified below. The
customer or the user of the Model 450C should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
Group 1
The Model 450C uses RF energy only for its internal
CISPR 11
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
Class A
The Model 450C is suitable for use in all
CISPR 11
establishments other than domestic and those
directly connected to the public low-voltage power
Harmonic emissions
Class A
supply network that supplies buildings used for
IEC 61000-3-2
domestic purposes.
Voltage fluctuations/
Complies
flicker emissions
IEC 61000-3-3
Vital-Guard 450C Operation and Maintenance Manual
5
SAFETY
Guidance and manufacturer’s declaration – Electromagnetic immunity
The Model 450C is intended for use in the electromagnetic environment specified below. The
customer or the user of the Model 450C should assure that it is used in such an environment.
Immunity test
IEC 60601 test
Compliance level Electromagnetic environment –
level
guidance
Electrostatic
±6 kV contact
±6 kV contact
Floors should be wood, concrete,
discharge (ESD)
or ceramic tile. If floors are
IEC 61000-4-2
±8kV air
±8kV air
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
±2 kV for power
±2 kV for power
Mains power quality should be
Transient/burst
supply lines
supply lines
that of a typical commercial or
IEC 61000-4-4
hospital environment.
±1 kV for
±1 kV for
input/output lines
input/output lines
Surge
±1 kV differential
±250V
Remove equipment that can cause
IEC 61000-4-5
mode
differential mode
electrical disturbance from the
vicinity of the input/output
±2 kV common
±2 kV common
cables.
mode
mode
Voltage dips, short
<5 % UT
<5 % UT
Mains power quality should be
interruptions, and
(>95 % dip in UT)
(>95 % dip in UT) that of a typical commercial or
voltage variations
for 0.5 cycle
for 0.5 cycle
hospital environment. If the user
on power supply
of the Model 450C requires
input lines
40 % UT
40 % UT
continued operation during power
IEC61000-4-11
(60 % dip in UT) for (60 % dip in UT)
mains interruptions, it is
5 cycles
for 5 cycles
recommended that the Model
450C be powered from an
70 % UT
70 % UT
uninterruptible power supply.
(30 % dip in UT) for (30 % dip in UT)
25 cycles
for 25 cycles
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
6
<5 % UT
(>95 % dip in UT)
for 5 sec cycle
3 A/m
<5 % UT
(>95 % dip in UT)
for 5 sec cycle
Not applicable
Not applicable
Vital-Guard 450C Operation and Maintenance Manual
SAFETY
Guidance and manufacturer’s declaration – Electromagnetic immunity
The Model 450C is intended for use in the electromagnetic environment specified below. The
customer or the user of the Model 450C should assure that it is used in such an environment.
Immunity test
IEC 60601 test
Complianc Electromagnetic environment – guidance
level
e level
Portable and mobile RF communications
equipment should be used no closer to any
part of the Model 450C, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
d = 1.2
p
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d = 1.2
p
80 MHz to 800 MHz
d = 2.3
p
800 MHz to 2.5 GHz
Where p is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey a, should be less than the
compliance level in each frequency range b
Interference may occur in the vicinity of
the equipment marked with the following
symbol:
NOTE 1 – At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 – These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radios, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, and electromagnetic site survey should be
considered. If the measured field strength in the location in which the Model 450C is used exceeds the applicable RF
compliance level above, the Model 450C should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the Model 450C.
b
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
Vital-Guard 450C Operation and Maintenance Manual
7
SAFETY
Description of Warning Labels
Attention, consult ACCOMPANYING DOCUMENTS before attempting to change
power supply selection or carry out interconnections. Equipment connected should
comply with IEC-60601-1 or IEC-950 with configuration to IEC-60601-1-1.
 ♥ 
Type CF applied part, Defibrillator proof.
Equipotential earth connector adjacent to this symbol.
Fuse type/rating.
Output signal.
ON
Input signal.
_______
Stand By (STBY)
♥
Type CF applied part.
∼
Alternate Current (AC)
WEEE Compliance
For monitors used in Australia
For continued Defibrillation protection and proper Respiration monitoring use only patient cables with 1kΩ series
resistors. It is important to use only patient cables with 1kΩ series resistors because the respiration circuit is tuned to
operate with series resistors.
For 3-lead configuration monitors, use 3-lead patient cable with 1kΩ series resistors. Ivy reorder no. 590197.
For 5-lead configuration monitors, use 5-lead patient cable with 1kΩ series resistors. Ivy reorder no. 590207.
8
Vital-Guard 450C Operation and Maintenance Manual
MONITOR DESCRIPTION
MONITOR DESCRIPTION
The Model 450C Patient Monitor is a color four-trace adult, pediatric, or neonatal monitor for:
ECG
Heart Rate
Respiration - Impedance
EtCO2
Two Invasive Pressures
Two Temperatures
Non-Invasive Blood Pressure (NIBP)
Pulse Oximetry
The Model 450C is available in several configurations, therefore, not all parameters are included in all monitors.
The use of the Model 450C is restricted to one patient at a time. The monitor is intended for use in emergency room,
recovery room, intensive care, and surgical applications and transport with 2.5 hours battery operation.
The Model 450C is suitable for use in presence of electrosurgery.
The Model 450C is not intended for use with any other physiological monitoring unit.
An optional integral recorder is available. Recorder settings are made through the monitor menus. Battery operation
consists of two rechargeable lead acid cells.
The display has four traces available and alphanumeric displays for data, alarms, and user information, such as
menus. The monitor automatically senses which devices are connected and acquires and displays data for the
parameters being used. Data for parameters not being used is not displayed. Areas for display of those parameters
are left blank.
Classification (in accordance with IEC-60601-1)
Protection against electric shock:
Class 1 and internally power equipment
Degree of protection against electric shock:
Type CF applied part. Defibrillator proof:
ECG
Type CF applied part: SpO2, IBP, Temperature,
NIBP, & EtCO2.
Degree of protection against harmful ingress of water:
Ordinary equipment IPX0 per IEC-60529
Methods of Maintenance and Cleaning:
See page 72
Degree of safety of application in the presence of a
flammable anesthetic mixture with air or with oxygen
or nitrous oxide:
Equipment not suitable for use in the presence of a
flammable anesthetic mixture
Mode of operation:
Continuous
Vital-Guard 450C Operation and Maintenance Manual
9
MONITOR DESCRIPTION
Specifications
ECG
Lead Selection:
Standard Three-Lead:
Five-Lead Option:
ECG Amplitud range:
1 to 100 mm/mv
Patient Cable:
6-Pin AAMI Standard
Isolation:
Isolated from ground related circuits by ≥4 kV rms, 5.5 kV peak
CMRR:
≥90 dB with patient cable and 51 kΩ/47 nF imbalance
Input Impedance:
≥20 MΩ at 10 Hz with patient cable
Frequency Response
Display, at 50 mm/sec:
LI, LII, LIII,
LI, LII, LIII, aVR, aVL, aVF, V
Filtered:
Diagnostic:
0.5 to 35 Hz
0.05 to 100 Hz
Frequency Response
X1000 Output:
Filtered:
Diagnostic:
0.5 to 35 Hz
0.05 to 100 Hz
Input Bias Current:
Any lead configuration <200 nA dc maximum
Electrode Offset Potential:
± 0.5 V
Noise:
<20 µV peak-to-peak, referred to the input with all leads
connected through 51 kΩ/47 nF to ground
Defibrillator Protection:
Protected against 360 J discharge.
Leakage Current:
<10 µA at 120 V ~, 60 Hz
Electrosurgical Interference
Protection:
Standard
Lead Fault Detection:
Lead fault detection in any lead configuration with any of the leads open
NOTE: The Phase Delay from patient electrode to rear panel outputs is < 3ms
Cardiotach
10
Range:
15 to 300 bpm
Accuracy:
±1%
Resolution:
1 bpm
Sensitivity:
Adult:
Neonate:
Pacer Rejection Width:
0.1 to 2 ms at ±2 to ±700 mV
Tall T Wave Rejection:
Rejects T waves ≤R wave
300 µV peak
100 µV peak
Vital-Guard 450C Operation and Maintenance Manual
MONITOR DESCRIPTION
Temperature (Two Channels)
Thermistor:
YSI Series 400 or equivalent
Range:
19°C to 45°C
Accuracy:
±0.1°C from 25°C to 40°C
Resolution:
0.1°C
Isolation:
Isolated from ground related circuits by ≥4 kV rms, 5.5 kV peak
Leakage:
<10 µA at 120 V ~, 60 Hz
Respiration
Excitation Current:
<400 µA rms
Excitation Frequency:
57 kHz
Maximum Electrode
Impedance:
4 kΩ
Size Adjustment:
Manual
Respirotach Range:
Neonate:
Adult:
4 to 150 br/min
4 to 60 br/min
Sensitivity Range:
Neonate:
Adult:
0.1 to 10 Ω, 4 to 150 br/min
0.3 to 10 Ω, 4 to 60 br/min
Pace Pulse
Rejection Width:
0.1 to 2 ms at ±2 to ±700 mV
End Tidal CO2
Measurement Range:
0.0 to 100.0 Torr
Reporting Range:
–5.0 to 120.0 Torr
Resolution:
0.1 Torr (0.01%)
Bias:
0.0 to 40.0 Torr:
40.1 to 100.0 Torr:
±2.0 Torr Abs
±5.0% Relative
Precision:
0.0 to 40.0 Torr
40.1 to 100 Torr
<0.5 Torr (RMS)
<1.0 Torr (RMS)
CO2 Compensation:
Local Barometric Pressure: 500 to 850 Torr (in whole units)
0 to 100% (in whole units)
N2O
0 to 100% (in whole units)
O2
Vital-Guard 450C Operation and Maintenance Manual
11
MONITOR DESCRIPTION
MASIMO SET® Pulse Oximetry
Range:
Saturation (%SpO2)
Pulse Rate (bpm)
1% - 100%
25 - 240
Accuracy:
Saturation (%SpO2) – During no motion conditions
Adults
70% - 100% ± 2 digits
0% - 69% unspecified
Neonates
70% - 100% ± 3 digits
Saturation (%SpO2) – During motion conditions
Adults
70% - 100% ± 3 digits
0% - 69% unspecified
Pulse (bpm) – During no motion conditions
25 to 240 ± 3 digits
Pulse (bpm) – During motion conditions
25 to 240 ± 5 digits
Resolution:
Saturation (%SpO2)
Pulse Rate (bpm)
1%
1
Non-Invasive Blood Pressure
Measurement Technique:
Oscillometric method determines mean arterial
pressure (MAP), systolic, and diastolic pressure.
Resolution:
1 mmHg
Measuring Range:
Systolic
Mean
Diastolic
Pulse Rate
12
Adult
(mmHg)
60–250
45–235
40–220
30 –180
Neonatal
(mmHg)
40–130
35–105
20– 90
30 –240
Accuracy:
Over the blood pressure measurement range stated above, for systolic and
diastolic pressures treated separately, the mean difference of the paired
measurement of the test system and the comparison system shall be
±5mmHg or less, with a standard deviation of 8mmHg or less.
Inflation speed:
At sea level <10 seconds.
Maximum leak rate:
5 mmHg/3 min.
Over pressure protector:
(Maximum cuff pressure)
Adult/Child
Neonate
300mmHg
150mmHg
Vital-Guard 450C Operation and Maintenance Manual
MONITOR DESCRIPTION
Modes:
Auto: Measurements are automatically repeated at timed
programmed intervals of 1, 2, 3, 4, 5, 10, 15, 20, 30, or 60 minutes
Stat: A series of consecutive measurements; the series lasts for 5
minutes.
Manual: A single reading each time START is pressed.
Alarms:
High and low pressure alarms derived from the NIBP systolic,
diastolic, or mean
Invasive Pressure (Two Channels)
Frequency Response:
dc to 12 Hz
Input Impedance:
500 kΩ
Transducer Sensitivity:
50 µV/V/cm Hg
Excitation Voltage:
+5 V dc
Zero:
Auto pushbutton, ±150 mmHg
Drift:
0.1 mmHg/oC
Accuracy:
±2% of reading or ±1 mmHg, whichever is greater
Isolation:
Isolated from ground related circuits by ≥4 kV rms, 5.5 kV peak
Test Waveform:
150/50 mmHg
NOTE: Refer to the manufacturer’s documentation for technical specifications on any particular pressure transducer.
Alarms
Heart Rate:
High and Low: 15 to 300 bpm in 5 bpm increments
SpO2:
High: 70 to 99% in 1% increments
Low: 50 to 95% in 1% increments
Asystole:
R to R interval >6 seconds
No Pulse:
When triggering from pleth or invasive pressure, interval >6 seconds
Lead Off:
Detached lead or offset potential ≥0.5 V
Check Leads:
Electrode impedance is >4 kΩ
Coincidence:
When QRS and respiration triggers are synchronized for eight
consecutive events
Apnea:
Absence of detectable respiration activity for 5, 10, 15,, or 20 seconds
Vital-Guard 450C Operation and Maintenance Manual
13
MONITOR DESCRIPTION
Alarms
Resp Rate:
High and Low: 5 to 150 br/min in 5 br/min increments
Invasive Pressure:
High and low; systolic, diastolic, or mean:
-40 to 300 mmHg in 2 mmHg increments
NIBP:
High and low; systolic, diastolic, or mean:
10 to 300 mmHg in 2 mmHg increments
For additional NIBP alarm conditions, see the NIBP Monitoring section of this manual.
Trends
NIBP:
Last 200 readings, tabular
Cardiorespirogram:
Up to 4 minutes beat-to-beat heart rate and respiration
Trend Waveforms:
Up to 24 hours of heart rate, respiration rate, SpO2, and invasive
pressure
Review Trend:
6 hour view: Stored the last 24 hours of heart rate, respiration rate, SpO2
and IBP. Six hours of data are displayed per window.
(1 pixel = 1 minute).
30 minute view: Stored the last 24 hours of heart rate, respiration rate,
SpO2 and IBP. Thirty minutes of data are displayed per window.
(1 pixel = 5 seconds).
Real Time Clock
Resolution:
1 minute
Display:
24 hours
Power Requirement:
The real time clock keeps time if the monitor has power or not.
The clock is powered by a dedicated battery whose life is minimum
4 years at a temperature of 25oC
Test Mode
ECG:
1 mV/100 ms @ 70 bpm
Respiration:
70 br/min, coincidence check
Invasive Pressures:
150/50 mmHg
Temperatures:
40oC
Type:
Active Matrix TFT Color LCD
Trace:
Four Traces, moving bar
Screen size (viewing area):
Height:
Width:
Display
14
158.4 mm
211.2 mm
Vital-Guard 450C Operation and Maintenance Manual
MONITOR DESCRIPTION
Display
Sweep Speed:
6.25, 12.5, 25, 50 mm/s
Character Size:
Large:
Small:
Recorder
Writing Method:
Number of Traces:
Modes:
Paper Speeds:
Trends:
Resolution:
Frequency Response:
Data Rate:
13x8 mm
4x3 mm
Direct Thermal
2
Direct - Manual Recording
Timed - Print button initiates a 30 second recording
Delay - Recorder is activated by an alarm
Auto - Parameters to be printed are selected based on the alarm
CRG - Prints Cardiorespirogram
Remote - Allows activation of a remote recorder
1 mm/s in CRG mode, 6.25, 12.5, 25 and 50 mm/s
Printed as selected for the display
NIBP tabular - prints up to 200 measurements as shown on the display
Vertical - 200 dots/in.
Horizontal - 600 dots/in. at ≤25 mm/s
400 dots/in. at >25 mm/s
>100 Hz at 50 mm/s
400 samples/s/trace
Mechanical
Size:
Height:
Width:
Depth:
Weight:
Basic unit: 12.8 lbs. (5.8 kgs.)
Environmental
Operating Temperature Range:
11.0 in. (28.0 cm)
11.1 in. (28.2 cm)
7.5 in. (19.1 cm)
15oC to 35oC
Storage Temperature Range:
-5oC to 55oC
Relative Humidity:
0-90% non-condensing
Power Requirements
Voltage Input:
100 - 230 V ~; 47 - 63 Hz, +10% –15%
Fuses Type and Rating:
1.6ASB, Metric 5x20mm, 250V
Maximum ac Power
Consumption:
145 VA
Battery:
Run Time:
2.5 hours – (NIBP measurement every 15 minutes and
30-second recorder print-out every 30 minutes)
Charge Time: <12 hours
Type:
Replaceable sealed gel cell, DOT and IATA
approved
Vital-Guard 450C Operation and Maintenance Manual
15
MONITOR DESCRIPTION
Controls and Indicators
Setup Keys
PATIENT
SETUP
SYSTEM
SETUP
ALARM
SETUP
EXIT
Causes menu to be displayed for settings involving the parameters.
Displays menus for general settings, such as display and recorder setup, adult/neonatal mode,
and audio and clock settings.
To set limits for all parameters
To leave a menu and return to the normal display.
See the Menu Structure section of this manual for a description of the menus.
PROGRAMMABLE
KEYS
COLOR
DISPLAY
SETUP
KEYS
BASIC KEYS
16
Vital-Guard 450C Operation and Maintenance Manual
MONITOR DESCRIPTION
Basic Keys
Generates a 1 mV pulse at 70 bpm, respiration signal at 70 br/min, (coincidence check),
invasive blood pressures at 150/50, and temperatures at 40°C displayed on the screen and sent
to the rear panel. This key is used to verify that the monitor is functioning correctly. In TEST,
heart rate is always computed from the generated ECG.
Turns the alarm audio on and off. When off, the ALARMS OFF tag is displayed on the
screen.
Disables the audible and visual alarms for a two-minute period to allow the operator to
perform procedures that would otherwise set off the alarms. This avoids the problem of
turning off the alarms and forgetting to turn them back on. A message on the screen indicates
when this function is active. Press the ALARM PAUSE key again to return the alarms to
normal before the two minutes have expired.
Respiration Size
RESP SIZE
The up and down arrows control the relative size of the respiration signal. Size ranges are from
0 to 100%. This controls both the display size and the trigger sensitivity.
NOTE: If the respiration size is set for a very sensitive range, cardiac artifact can generate false triggers, causing a
coincidence alarm.
On
When the monitor is either plugged into an AC power source or operating on batteries,
the switch toggles from Stand By (off position) to ON providing power to the monitor’s
electronic circuits. The monitor must be connected to an AC source to be able to charge
the internal batteries.
NOTE: To disconnect the monitor from the mains power unplug the power cord.
Stand By
NIBP Control Keys
START
STOP
CYCLE
Starts a single non-invasive blood pressure measurement in cycle off (manual mode) or
repetitive measurements in cycle mode or stat mode. Measurements in any mode can be
canceled by pressing STOP.
Stops any non-invasive blood pressure measurements in cycle off (manual mode), automatic
mode, or stat mode and cancels automatic or stat mode.
Selects the NIBP mode of operation. Selections are: automatic cycle mode, stat mode or
cycle off (manual mode). In the automatic cycle mode inflation intervals of 1, 2, 3, 4, 5,
10, 15, 20, 30 or 60 minutes can be selected.
Vital-Guard 450C Operation and Maintenance Manual
17
MONITOR DESCRIPTION
Programmable keys
Each of the keys to the right of the display has a normal function, printed on the key, and several menu functions
displayed on the screen in the setup mode. The normal functions are described below. Menu functions are described
in the menu structure section of this manual.
DISPLAY
LIMITS
Displays the current alarm limit settings. Press DISPLAY LIMITS again to return to normal
display.
FREEZE
Stops the movement of the all waveforms on the display for closer evaluation. Press FREEZE
again to resume movement.
ALARM
RESET
Resets the audible and visual indicators for an alarm that has been activated.
PRINT
TREND
18
Starts a printout at the optional recorder. The recorder normally plots waveforms. To define
the waveform printout, use the menus accessed by pressing the SYSTEM SETUP key.
Displays Trend data. To define the trend display, see the trend display setup instructions.
Vital-Guard 450C Operation and Maintenance Manual
MONITOR DESCRIPTION
Side Panel Connector
SpO2
ECG/RESP
PRESS 1
PRESS 2
TEMP 1
TEMP 2
NIBP
ETCO2
Vital-Guard 450C Operation and Maintenance Manual
19
MONITOR DESCRIPTION
Display
Waveform Display
When the monitor is configured for all parameters, any of the following waveforms can be displayed:
ECG
Respiration
Two Invasive Pressures (P1, P2)
Pleth (Pulse Oximetry)
EtCO2
Waveforms are updated with a moving bar at a rate of 6.25, 12.5, 25, or 50 mm/s, selected by the user. Waveform
amplitudes are adjusted using the ECG Size selection in the menus and the RESP SIZE keys on the front panel.
Invasive blood pressure scales are changed through the P1/P2 menu. A typical display is shown below.
The display contains five zones, selectable through the menus.
ZONE 1: Can only display the ECG waveform or no waveform (OFF)
ZONE 2: Selectable to display waveforms of RESP, EtCO2, SpO2, P1, P2, or OFF
ZONE 3: Selectable to display waveforms of RESP, EtCO2, SpO2, P1, P2, or OFF
ZONE 4: Selectable to display waveforms of RESP, EtCO2, SpO2, NIBP, P1, P2, Zones 3 & 4 P1
EXPAND, Zones 3 & 4 P2 EXPAND, Zones 3 & 4 P1/P2 EXPAND, or OFF
ZONE 5 (numeric only): Displays temperature readings with an area reserved to selectively
display RESP, EtCO2, SpO2, or OFF
If invasive pressure is being monitored, Zones 3 and 4 combined can display one or both pressures on a single
expanded scale that is twice the height of the normal pressure scale.
ZONE 1
ZONE 2
ZONE 3
PR=70
ZONE 4
TIME OF DAY
ZONE 5
20
BATTERY
INDICATOR
Vital-Guard 450C Operation and Maintenance Manual
MONITOR DESCRIPTION
Temperature data displayed in Zone 5
P1 displayed in Zone 3 and P2 displayed in Zone 4
PR=70
PR=70
Expanded P1 displayed in Zones 3 and 4
With NIBP Graph in Zone 4.
PR=70
With PATIENT SETUP selections displayed
PR=70
With SYSTEM SETUP selections displayed
PR=70
NIBP Graph
To display the NIBP graph, press the PATIENT SETUP key, select [NIBP], and select [DATA FORMAT] until it is
set at GRAPH.
The NIBP graph is shown in Zone 4 in one of the above illustrations. This oscillation profile assists the clinician in
determining the validity of the NIBP measurement. An erratic or flat profile suggests that excessive motion or other
artifact may have affected the measurement. A profile shaped like a bell curve represents a measurement not
affected by motion or other artifact.
Vital-Guard 450C Operation and Maintenance Manual
21
MONITOR DESCRIPTION
Alphanumeric Data
The following data appears on the display when the appropriate parameters are in use. Whenever possible, numeric
data is displayed near related waveforms. When you select a display zone for a waveform, numeric data related to
that waveform is also displayed in that zone. To display data without displaying the waveform, use Zone 5.
Time of Day- At the lower right-hand corner of the display.
Heart Rate- Upper left corner of the display with a symbol indicating the source of the heart rate.
Respiration rate and flashing respiration indicator- To the left of the waveform. if the waveform is displayed.
Otherwise, at the bottom of display in Zone 5. Respiration rate can be computed from either impedance or EtCO2.
End Tidal CO2 (EtCO2 ) - To the left of the waveform if the waveform is displayed. Otherwise, at the bottom of
display in Zone 5.
Oxygen Saturation (SpO2) and pulse oximeter bar graph- To the left of the waveform if the waveform is displayed.
Otherwise at the bottom of the display in Zone 5.
Pressure Readings-one or two sets of readings- To the left of the waveform in the form:
syst/dias
(mean)
Current NIBP Reading- At the lower left in the form:
syst/dias
(mean)
Temperature- Three readings at the bottom of the display: Temp 1, Temp 2 and the difference between the two
temperatures (∆ Temp).
The data display is blank for any parameter not in use.
Alarm Limits Display
A numeric display of current alarm limits settings. The alarm limits appear just below the numeric value of the
corresponding parameter. This display appears when the DISPLAY LIMITS key is pressed.
Alarm Displays
When an alarm setting has been exceeded, a flashing alarm tag appears on the screen, indicating which alarm has
been triggered. Each alarm tag appears in the same location where the limits for that alarm are displayed.
Trend Displays
NIBP Trend: A tabular display of the readings at the time of each NIBP measurement. The latest five readings are
displayed, and the display can be scrolled to view previous readings.
CRG: Graphic Display of up to four minutes of beat-to-beat heart rate, SpO2 and respiration effort. Whenever EtCO2
is being monitored, the EtCO2 waveform is used for the respiration portion of the CRG.
Trend: Graphic Display of up to 24 hours of averaged heart rate, SpO2, respiration rate, and pressure readings.
Review Trend: Graphic and Numeric display of the most recent 24 hours of heart rate, SpO2, respiration rate, and
pressure readings at two different views: 6 Hour View displays 6 hours of data per window; and 30 Minute View
displays 30 minutes of data per window. Data can be moved under the cursor to obtain parameters readings at
specific point in time. Numeric value of heart rate, SpO2, respiration rate, and invasive pressures is displayed over
the cursor.
22
Vital-Guard 450C Operation and Maintenance Manual
MONITOR DESCRIPTION
Alarms
The following is a list of alarms and ranges for which they can be set. When the alarm triggers, the alarm indicator
flashes on the display and the audio alarm tone sounds. Pressing ALARM RESET turns off any alarms that have
triggered. If the measurement returns to within the limits, the alarm turns off.
The ALARM PAUSE key disables the alarms for two minutes or until ALARM PAUSE is pressed again.
High heart rate: 15 to 300 bpm in 5 bpm increments
Low Heart Rate: 15 to 300 bpm in 5 bpm increments
High Resp Rate: 5 to 150 br/min in 5 br/min increments
Low Resp Rate: 5 to 150 br/min in 5 br/min increments
Temp High: 35ºC to 45ºC in 0.1ºC increments
Temp Low: 35ºC to 45ºC in 0.1ºC increments
NOTE: For additional Temp alarm conditions, see the Temperature Monitoring section of this manual.
EtCO2 High: 10 to 100 Torr in 2 Torr increments
EtCO2 Low: 10 to 100 Torr in 2 Torr increments
NIBP High: Selectable systolic, diastolic, or mean 10 to 300 mmHg in 2 mmHg increments
NIBP Low: Selectable systolic, diastolic, or mean 10 to 300 mmHg in 2 mmHg increments
NOTE: For additional NIBP alarm conditions, see the NIBP Monitoring section of this manual.
IBP High: Selectable systolic, diastolic, or mean; -40 to 300 mmHg in 2 mmHg increments
IBP Low: Selectable systolic, diastolic, or mean; -40 to 300 mmHg in 2 mmHg increments
NOTE: Invasive and non-invasive pressures can be set to alarm on systolic, diastolic, or mean.
High SpO2: 50 to 100%in 1% increments
Low SpO2: 50 to 100% in 1% increments
Asystole: R-to-R interval >6 seconds
No Pulse: When triggering from pleth or invasive pressure, the interval between heartbeats exceeds six
seconds
Lead Off: Detached lead or offset potential ≥0.5 V. The ALARM RESET key does not reset this alarm
Check Leads: Electrode impedance is >4 kΩ
Coincidence: When QRS and respiration triggers are synchronized for eight consecutive events
Apnea: Absence of detectable respiration activity for 5, 10, 15, or 20 seconds, selectable.
Vital-Guard 450C Operation and Maintenance Manual
23
MONITOR DESCRIPTION
Priority Alarms Feature (If present)
Note: Priority alarms may be enabled via the CONFIGURATION MENU (see operation and maintenance
manual).This section applies only if priority alarms are PRESENT (default is ABSENT)
There are two types of Priority alarms schemes. Scheme 1 prioritized the alarms into 3 groups while Scheme 2
prioritized the alarms into 4 groups. The table below describes the color and audio associated with each alarm and
scheme.
Priority
High
Medium
SpO2 High
Low
Color
Red
Yellow
N/A
White
SCHEME 1
Audio Type
Rapid Beeps
Slower Beeps
N/A
Longest Beeps
SCHEME 2
Color
Audio Type
Red
Siren
Yellow
Warble
Light Blue
Alternate tones
White
Tone
The alarms are grouped as follows:
High Priority:
Asystole / No Pulse
Apnea / No Resp
Heart Rate (high/low)
Medium Priority:
Respiration rate (high/low)
SpO2% (high/low). If Scheme 2 is selected this alarm becomes a separate priority.
NIBP pressures (high/low)
Invasive pressures (high/low)
EtCO2 (high/low)
InCO2 (high)
Low Priority:
Temperature (high/low)
Miscellaneous alerts (i.e.: Coincidence)
All non-physiological alerts (i.e.: Lead Off, Check Cuff, etc)
Note: The Priority Alarm feature incorporated into the Vital-Guard 450C is intended to provide the user
with a graded visual and audible indication of both the Patient’s and the Monitor’s status. It is the
responsibility of the users to ensure that they use good clinical judgement to determine how they should
respond to any alarm.
Rear Panel
The following are located on the rear panel.
ANALOG OUTPUT: An analog output that can be used for connecting to an external device. Signals on this
connector include ECG, respiration, invasive pressures, pleth, heart rate, SpO2, alarm out, remote recorder, and
optional nurse call.
Do not attempt to connect a cable to this connector without contacting your Biomedical Engineering Department or
the Ivy Service Department. This is to insure the connection complies with leakage-current requirements of one of
the following applicable standards: Underwriters Laboratories UL 2601-1, Canadian Standards Association CSA
601.1 No. M90 or Internal Electrotechnical Commission IEC 60601-1. The maximum non-destructive voltage that
may be applied to this connector is 5 V dc.
ECG X1000: A ¼-inch phone socket that can be used for connecting to an external device using the ECG waveform
and LEAD OFF indicator.
24
Vital-Guard 450C Operation and Maintenance Manual
MONITOR DESCRIPTION
COM 1: A digital interface for network communication.
COM 2 / RJ45: A digital interface for network communication.
TELEMETRY ANTENNA: A BNC connector for the Telemetry System antenna.
CAUTION: When the Telemetry option is selected and installed in the Model 450C due to the technical
limitations inherent with all wireless communications, IVY Biomedical Systems, Inc. cannot guaranty that all
signals transmitted via the antenna will be received by the appropriate receiver or receiving device.
PEQ GROUND: Potential Equalization- Use this ground connection to ensure that no potential differences can
develop between this equipment and the other electrical equipment.
TELEMETRY
ANTENNA
COM 1
COM 2 / RJ45
ANALOG OUTPUT
PEQ GROUND
ECG X1000 OUTPUT
MAINS POWER INPUT
FUSES BEHIND THE PANEL
Note: Parts and accessories used must meet the requirements of the applicable IEC 60601 series safety standards and
essential performance standards, and/or the system configuration must meet the requirements of the IEC 60601-1-1
medical electrical systems standard.
ACCESSORIES (EQUIPMENT) – The use of ACCESSORY equipment not complying with the equivalent safety
requirements of this equipment may lead to a reduced level of safety of the resulting system. Consideration relating
to the choice shall include:
• use of the accessory in the PATIENT VICINITY; and
• evidence that the safety certification of the ACCESSORY has been performed in accordance to the appropriate
IEC 60601-1 and/or IEC 60601-1-1 harmonized national standard.
Menu Structure
Settings for all options for the Model 450C are made through a series of menus that appear at the right side of the
display. You access the menus by pressing the PATIENT SETUP or SYSTEM SETUP key. The PATIENT
SETUP key is for settings involving the parameters, while the SYSTEM SETUP key is for general monitor settings.
When menus are displayed, if no key is pressed for a period of 30 seconds, the monitor returns to normal display.
NOTE: If the monitor beeps when you press a key, it indicates an invalid selection for the current mode or function.
Vital-Guard 450C Operation and Maintenance Manual
25
MONITOR DESCRIPTION
The PATIENT SETUP Key
Press PATIENT SETUP to display the main menu, which allows you to select ECG, RESP, SpO2, P1/P2, or NIBP
menus. See the menu structure diagram below.
When the main menu is displayed, press EXIT to return to normal display.
PATIENT
SETUP
ECG/HR
RESP/CO2
SpO2
NIBP
INVASIVE BP
SpO2 ALARM
ON
RESP
IMPEDANCE
RESP
CO2
AVERAGING
8 SEC
COINCIDENCE
ON
SENSITIVITY
NORMAL
APNEA DELAY
20 sec
ECG
ON
ECG
OFF
RESPONSE
NORMAL
LEAD
I
HR SOURCE
OXIMETRY
ZERO
SENSOR
FILTER
ON
RESP
OFF

SPAN
CHECK
ECG SIZE
10 m m/mv

SETTINGS
PARAMETER
P1
ZERO
TRANSDUCER
ALARM TYPE
MEAN
TRACE SCALE
0 TO 240
ALARM TYPE
MEAN
ALARM TYPE
MEAN
ALARM TYPE
MEAN
ALARM TYPE
MEAN
NEXT
SETTING





SCROLL
SCROLL
INITIAL CUFF
180
INITIAL CUFF
180
BAR PRESS
760
DATA FORMAT
GRAPH
DATA FORMAT
( NONE )
DEFAULT
760 MMHG


DATA FORMAT
NIBP TABLE
DATA FORMAT
IBP / TEMP



TRACE SCALE
0 TO 50
SITE LABEL
P1
PATIENTSETUP KEY SELECTION
SpO2
ECG/HR
SpO2 ALARM; ON, OFF
AVERAGING:6, 8, 10, 12,14, 16 SEC
SENSITIVITY: NORMAL, HIGH
ECG: ON, OFF
LEAD: I, II, III, aVR, aVL, aVF, V
FILTER: ON, OFF
ECG SIZE: 1 TO 100 m m/mV
NIBP
ALARM TYPE: MEAN, SYS, DIAS, OFF
DATA FORMAT: NIBP TABLE, IBP/TEMP, GRAPH, NONE
INITIAL CUFF: 50-250 mm Hg (ADULT)
50-150 mm Hg (NEONATE)
RESP/CO2
RESP: IMPEDANCE, CO2, OFF
COINCIDENCE: ON, OFF
APNEA DELAY: OFF, 5, 10, 15, 20 s
NO RESP DELAY: OFF, 5, 10, 15, 20 s.
SETTINGS
NEXTSETTING: BAR PRESS, N20, 02,
NORMAL RESPONSE
TRACE SCALE:0 TO 50, 0 TO 100 mmHg
26
INVASIVE BP 1&2
PARAMETER: P1, P2
ALARM TYPE: MEAN, SYS, DIAS, OFF
TRACE SCALE: -30 TO 30, 0 TO 60, 0 TO 120,
0 TO 240, 0 TO 300 mmHg
SITE LABEL: P1, ART1, CVP1, PA1, ICP1, UAC1, UVC1
P2, ART2, CVP2, PA2, ICP2, UAC2 & UVC2
Vital-Guard 450C Operation and Maintenance Manual
MONITOR DESCRIPTION
The SYSTEM SETUP Key
Press SYSTEM SETUP to display the mode menu for setting audio volume, trace speed, display and recorder
settings.
When the mode menu is displayed, press EXIT to return to normal display.
SYSTEM
SETUP
ALARM
SETUP
DISPLAY
SETUP
TREND
SETUP
RECORDER
SETUP
SET
VOL/CLOCK
MODE
ADULT
PARAMETER
HR

HIGH
140

MODE
DIRECT
TREND MODE
1 HOUR
CONFIRM
CLEAR
TRACE SPEED
25 MM/SEC
CHANNEL 1
ECG
CLEAR
TRENDS
CANCEL
ZONE 2
RESP/CO2

LOW
40

REVIEW
TREND
CO2 TREND
ETCO2

SELECT
MONTH
HR VOL
40%


< ----
< ----
---->
---->

VIEW IN
(30 MIN)
VIEW OUT
(6 HOUR)

ENTER
PRINT
PAGE
PRINT
PAGE
SET
CLOCK
CANCEL
2

ALARM VOL
40%
SYSTEM SETUP KEY SELECTION
DISPLAY SETUP
TRACE SPEED: 6.25, 12.5, 25, 50 mm/s
ZONE 2: OFF, RESP/CO2, SpO2, P1, P2
ZONE 3: OFF, RESP/CO2, SpO2, P1, P2
ZONE 4: OFF, RESP/CO2, SpO2, NIBP, P1, P2
EXPANDED 3&4: P1, P2, P1/P2
ZONE 5: OFF, RESP, SpO2, NIBP
TREND SETUP
RECORDER SETUP
MODE: DIRECT,TIMED,DELAY, AUTO, CRG, REMOTE
CHANNEL 1: ECG, RESP/CO2, PLETH, P1, P2
CHANNEL 1: ECG, RESP/CO2, PLETH, P1, P2, OFF
SET VOL/CLOCK
HR VOL: OFF, 20, 40, 60, 80, 100%
ALARM VOL: 20, 40, 60, 80, 100%
SETCLOCK
TREND MODE: CRG, 1, 8, 12, 24 hrs.
CO2 TREND: ETCO2, RR
SELECT: MONTH, DAY, YEAR, HOUR, MINUTE
ALARM SETUP KEY SELECTIONS
PARAMETER: HR, RR, SpO2, NIBP, P1, P2, ETCO2,
INCO2, T1 & T2 (if active)
Vital-Guard 450C Operation and Maintenance Manual
27
MONITOR SETUP
MONITOR SETUP
To setup the instrument for operation:
1.
Plug the ac line cord into a power source providing the proper voltage.
2.
Press the ON/STANDBY key at the bottom of the front panel to turn the unit on.
3.
Connect the patient cable to the ECG/RESP connector on the right side panel.
4.
If end tidal CO2 is to be monitored, connect the Sensor cable to the ETCO2connector on the right side panel.
5.
If pulse oximetry is to be monitored, connect the sensor and the pulse oximetry cable to the SpO2 connector
on the right side panel.
6.
If NIBP is to be monitored, connect the NIBP hose to the NIBP connector on the right side panel.
7.
If invasive pressure is to be monitored, connect the pressure transducer cable to the PRESS 1 or PRESS 2
right side panel connector.
8.
If temperature is to be monitored, connect the temperature sensor cable to the TEMP 1 or TEMP 2 right
side panel connector.
Adult/Neonate Modes
The Model 450C has mode settings for both adult/pediatric and neonatal monitoring. In general, the neonate mode is
used for patients less than one year old. All other patients should be monitored using the adult mode.
When monitoring NIBP, set the monitor to neonate mode whenever a neonatal cuff is being used.
Select Adult or Neonate monitoring mode as follows:
1.
Press the SYSTEM SETUP key to display the mode menu.
2.
Use the [MODE] menu selection to select either ADULT or NEO. The current selection is displayed on the
left side of the screen near the bottom.
3.
Press EXIT to return to the normal display.
28
Vital-Guard 450C Operation and Maintenance Manual
MONITOR SETUP
Heart Rate Source
Normally, heart rate is computed using the ECG waveform. If ECG is set to OFF, heart rate can be computed using
the oximetry, P1, or P2 waveforms.
1.
Press the PATIENT SETUP key.
2.
Select [ECG/HR] and select [ECG], setting it to OFF. The [HR SOURCE] selection appears.
3.
Use the [HR SOURCE] selection to select OXIMETRY, P1, P2, or NONE.
Trace Speed
1.
Press the SYSTEM SETUP key to display the mode menu. Then select [DISPLAY SETUP].
2.
Use the [TRACE SPEED] selection to select the trace speed. Selections are 6.25, 12.5, 25, and 50 mm/s.
3.
Press EXIT to return to the normal display.
Set Time and Date
1.
Press the SYSTEM SETUP key to display the mode menu. Then select [SET VOL/CLOCK] and select
[SET CLOCK] from the next menu.
2.
The first setting is for MONTH. Use the
3.
Press [SELECT] to move to the DAY setting. Use the
setting.
4.
Press [SELECT] to move to the YEAR setting. Use the
setting.
and
keys to increase or decrease the year
5.
Press [SELECT] to move to the HOUR setting. Use the
setting.
and
keys to increase or decrease the hour
6.
Press [SELECT] to move to the MINUTE setting. Use the
minute setting.
7.
When all date and clock settings are correct, select [ENTER] to enter the settings into the monitor’s
memory.
and
keys to increase or decrease the month setting.
Vital-Guard 450C Operation and Maintenance Manual
and
keys to increase or decrease the day
and
keys to increase or decrease the
29
MONITOR SETUP
Display Setup
1.
Press the SYSTEM SETUP key to display the mode menu.
2.
Select [DISPLAY SETUP] to display the Zone Select menu.
3.
Use the key next to each zone to select the display for the corresponding zone. Zone 5 is numeric only.
Selections are as follows.
Zone 1
Zone 2
Zone 3
Zone 4
Zone 5
ECG
Off
Resp/CO2
SpO2 (Pleth)
P1
P2
Off
Resp/CO2
SpO2 (Pleth)
P1
P2
Off
Resp/CO2
SpO2
NIBP
P1
P2
P1 Expanded
P2 Expanded
P1/P2 Expanded
Off
Resp/CO2
SpO2
NIBP
Each zone can also be set to off, except Zone 1, which is always ECG.
Temperatures are displayed in a portion of Zone 5.
NOTE: The above selections are available when the indicated parameters are included in the monitor’s
configuration and are being used.
4.
Press EXIT to return to the normal display.
5.
To display expanded invasive pressure waveforms, use the Zone 4 setting.
Data and Waveforms Color Setup
1.
Turn the monitor OFF.
2.
Simultaneously press the SYSTEM SETUP key and the ON/STANDBY button.
3.
Use the
4.
Press [CHANGE COLOR] to change the color.
5.
Press [DEFAULTS] to reset the colors.
6.
Press ON/STANDBY button twice to return to normal display.
30
and
keys to change the parameter.
Vital-Guard 450C Operation and Maintenance Manual
MONITOR SETUP
Audio Setup
1.
Press the SYSTEM SETUP key to display the mode menu.
2.
Select [SET VOL/CLOCK] to display the menu for setting HR and alarm volume.
3.
Adjust the HR volume using [HR
4.
Adjust the alarm volume using [ALARM
5.
Press EXIT to return to the normal display.
and
] keys.
and
] keys.
Display Limits
To avoid clutter on the display, limits for parameter alarms are not displayed unless the Limits Display selection is
set to ON.
To enable or disable Limits Display, press the DISPLAY LIMITS key during normal display.
Trend Displays
The 450C monitor has a trend display showing heart rate, SpO2, P1, P2, and respiration rate. Time selections for this
trend display are 1, 8, 12, and 24 hours. Another trend display is cardiorespirogram (CRG). The CRG displays the
last four minutes of beat-to-beat heart rate, a compressed respiration signal, and SpO2 trend, if the SpO2 waveform is
being displayed in the normal display.
Trend Setup
1.
Press the SYSTEM SETUP key. The mode menu appears.
2.
Select [TREND SETUP] to display the trend selection menu.
3.
Use the [TREND MODE] selection to select the trend mode: 1, 8, 12, or 24 hours or CRG.
4.
Press EXIT to return to the normal display.
Trend Display
1.
To display the trend selected in the trend setup, press the TREND key.
The trends appear on the display for the parameters selected in display setup.
2.
Press TREND again to return to the normal display.
NOTE: The monitor continuously accumulates trend data for all parameters in use. Any trend can be displayed by
changing the setup in the menus.
Vital-Guard 450C Operation and Maintenance Manual
31
MONITOR SETUP
CRG Display
1 Hour Trend Display
Review Trend Displays
Review trend collects the most recent 24 hours of heart rate, SpO2, invasive pressures, and respiration rate. The
monitor exhibits them at two different trend views:
1.
The 6 Hour view trend displays 6 hours of data per screen (1 dot = 1 minute).
2.
The 30 Minute view shows 30 minutes of data per screen (1 dot = 5 seconds).
In either case, a cursor indicates the numeric value of the parameters. The data can be moved under the cursor to
review the accumulated information at any specific time within the last 24 hours.
NOTE: The 450C monitor is constantly gathering data of heart rate, SpO2, invasive pressures, and respiration rate to
be used in the Review Trend. However, if the monitor is turned off the data will be erased.
To access the Expanded Trend mode:
1.
Press the SYSTEM SETUP key. The setup menu appears.
2.
Select [TREND SETUP] to display the trend selection menu.
3.
Select [REVIEW TREND] to enter the expanded trend display.
4.
Press [←] or [→] to move the cursor a long the trend.
5.
Select [VIEW IN/OUT] to toggle between 6 hour view and 30 minute view displays.
6.
Press EXIT to return to the normal display.
NOTE: The 450C monitor should be ON for at least one minute before Review Trend can be accessed.
32
Vital-Guard 450C Operation and Maintenance Manual
MONITOR SETUP
6 Hour View
30 Minute View
Clear Trend Data
Use the following procedure to delete all trend data in memory (including NIBP tabular trend data) without turning
power off.
1.
Press the SYSTEM SETUP key. The mode menu appears.
2.
Select [TREND SETUP] to display the trend selection menu.
3.
Select [CLEAR TRENDS]. The selections [CONFIRM] and [CANCEL] appear.
4.
Select [CANCEL] to abort the clear trends procedure. To clear the trend data, select [CONFIRM].
Default Settings
User settings are stored in non-volatile memory. The monitor powers up with the same settings that were in effect
when power was last turned off.
To reset the monitor to the default settings press TEST key and ALARM RESET key simultaneously while
applying power to the monitor by pushing the ON/STAND BY key.
Default Settings
System Setup
Trace Speed: 25 mm/sec
Displayed parameters:
Zone 1: ECG
Zone 2: RESP
Zone 3: SpO2
Zone 4: NIBP
Zone 5: Off
Trend Mode: One hour
Recorder: direct
HR Volume: 40%
Alarm Volume: 100%
Monitoring Mode: Adult
ECG
Lead: I
Filter: ON
ECG size 10mm/mv
HR limits (140-40)
RESPIRATION
Resp: Impedance
Apnea delay: 20 sec
Coincidence: ON
RR limits (60-10)
SpO2
SpO2 Alarm: ON
Averaging: 8 sec
Sensitivity: Normal
Alarm limits (100-90)
NIBP
Cycle: OFF
Data Format: NIBP table
Alarm Type: Mean
Alarm limits (160-80)
TEMP (Both channels)
Alarm limits: Off
Vital-Guard 450C Operation and Maintenance Manual
IBP 1
Alarm Type: Mean
Alarm limits (120-60)
Trace Scale: 0 to 240 mmHg
Site Label: P1
IBP 2
Alarm Type: Mean
Alarm limits (120-60)
Trace Scale: 0 to 240 mmHg
Site Label: P2
EtCO2
No Resp Delay: 20 sec
Bar. Press: 760mmHg
Trace Scale: 0 to 50
Alarm limits (50-300)
33
BATTERY OPERATION
BATTERY OPERATION
Internal Battery
The model 450C is equipped with two internal batteries. This allows the monitor to operate for 2½ hours under its
own power.
In the event of a mains power failure, the model 450C will automatically switch to battery operation.
A BAT indicator in the lower right corner of the display graphically displays the approximate remaining charge of
the battery: full, 75 %, 50 %, 25 %.
When the remaining battery charge is less than 20 minutes, the BAT indicator flashes the word LOW and an audio
tone sounds every 15 seconds.
The battery is continuously charged whenever the ac power is connected, even when the monitor is turned off.
Charging time when the battery is fully discharged is 12 hours.
If monitor is not in regular service, i.e. weekly use, to ensure maximum battery life, it is recommended that, at least
once a month, the unit be run on battery until it turns itself off and then recharged.
Remove the batteries if the monitor will not be in regular use for more than six months.
Status of the Battery
The model 450 has three LED’s (label as Charged, Charging and Battery Fault) to indicate the status of the battery.
They are located in the bottom left corner of the faceplate. The following table shows the status of the battery
depending on the action (ON, OFF or FLASHING) of the LED’s.
Status
of the Battery
Charged
LED
Charging
LED
Battery Fault
LED
Battery Absent
Pre-charge qualification
Fast charging
Maintenance charging
*Charge Pending (Temperature out of range)
**Fault
OFF
OFF
OFF
ON
X
X
OFF
FLASHING
ON
OFF
X
X
ON
OFF
OFF
OFF
FLASHING
ON
*If the temperature of the battery exceeds 116 oF (47 oC) the battery will not charge and will remain in a Charge Pending status.
**If Fault condition occurs replace the batteries immediately.
Removing the Battery
Refer to Disassembly and Assembly section of the Operation and Maintenance Manual to find out how to remove the
batteries.
WARNING: To avoid risk of fire, replace only with the same type battery (Ivy Part No. 110007).
34
Vital-Guard 450C Operation and Maintenance Manual
RECORDER OPERATION
RECORDER OPERATION
Changing Paper
Replace the roll of thermal paper as follows.
1.
Press the paper eject button to open the door at the front of the recorder.
Press Here
If the door does not open completely, pull it toward you until it is completely open.
2.
Reach in and remove the spent paper core by pulling it gently toward you.
3.
Place a new paper roll between the two round tabs of the paper holder.
4.
Pull some paper from the roll. Make sure the sensitive (shiny) side of the paper faces the print head. The
shiny side of the paper normally faces inside the roll.
5.
Align the paper with the pinch roller on the door.
Print
Head
Pinch
Roller
Vital-Guard 450C Operation and Maintenance Manual
35
RECORDER OPERATION
6.
Hold the paper against the pinch roller and close the door.
Recorder Menus
The recorder user interface is through the menu structure of the 450C.
1. Press the SYSTEM SETUP key to display the system menu.
2. Select [RECORDER SETUP] to display the recorder menu.
NOTE: At the start of each printout, the printer produces a header consisting of the current readings for all
parameters in use at the time of the printout, whether the parameter is displayed or not.
Recorder Modes
From the recorder menu, use (MODE) to select the printing mode to be used. Selections are DIRECT, TIMED,
DELAY, AUTO, CRG, and REMOTE.
The print mode is indicated in the lower right corner of the display.
Direct
To print in direct, press the PRINT key. Press PRINT again to stop printing.
The plot is preceded by a header which contains all parameter readings and the time/date.
The speed of the plot and vertical resolution are the same as the display. The plot is labeled with the speed of the plot
in mm/s, the recorder mode, and the parameters.
The following setup allows you to select which parameter is plotted in which channel of the recorder. If only one
waveform is plotted or if the same waveform is selected for both channels, the recorder is scaled to use the entire 40
mm. If two different waveforms are selected, the recorder splits the plot into two 20 mm channels.
36
Vital-Guard 450C Operation and Maintenance Manual
RECORDER OPERATION
Use the following procedure to select the waveforms for each channel.
1.
From the recorder menu, press the [CHANNEL 1] menu key to select the waveform to be printed on
Channel 1.
2.
Press the [CHANNEL 2] menu key to select the waveform to be printed on Channel 2.
Selections for each channel are ECG, Respiration, ETCO2, SpO2 (Pleth), P1, and P2. Channel 2 can also be
set to OFF. Whenever ETCO2 is being monitored, the ETCO2 waveform is used for respiration.
Timed
TIMED mode starts by pressing PRINT and prints for 30 seconds.
The speed of the plot and vertical resolution are the same as the display. The plot is labeled with the speed of the plot
in mm/s, the recorder mode, and the parameters.
Delay
Delay mode plots the 15 seconds before and 15 seconds after the occurrence of an alarm condition. The parameters
printed are those selected as described in the Direct Mode section.
The speed of the plot and vertical resolution are the same as the display. The plot is labeled with the speed of the plot
in mm/s, the recorder mode, and the parameters.
With no alarm, pressing PRINT starts a printout that includes the 15 seconds before and the 15 seconds after
pressing the key. An asterisk marks the time when the key was pressed or the alarm occurred.
The plot is preceded by a header which contains all parameter readings and the time/date.
Whenever delay mode is enabled, REC DELAY is displayed in the lower right corner of the display.
Auto
Parameters to be printed are selected based on the alarm. The printout is the same as in delay mode except that when
an alarm triggers the recorder, the parameters are selected automatically. If the recording is started by pressing the
PRINT key, the selected parameters are printed.
The following are the parameters printed for each type of alarm.
Type of Alarm
ASYSTOLE
HR
NO PULSE
APNEA
COINCIDENCE
RR
SpO2
NIBP
IBP1
IBP2
Traces Printed
Single-trace ECG
Single-trace ECG, SpO2 (Pleth), P1, or P2 (based on HR source)
Single-trace SpO2 (Pleth), P1, or P2 (based on HR source)
ECG, Respiration
ECG, Respiration
ECG, Respiration
ECG, SpO2 (Pleth)
Single-trace ECG
ECG, P1
ECG, P2
Vital-Guard 450C Operation and Maintenance Manual
37
RECORDER OPERATION
CRG
The recorder plots the CRG as two trend waveforms. The heart rate trend occupies the top half, and the compressed
respiration occupies the lower half. If CRG is selected and respiration is not being monitored, HR only is printed.
The trend is plotted at 1 mm/s. The plot is preceded by a header containing all the parameter readings and time/date.
From the recorder menu, use [MODE] to select the CRG mode. While the recorder is in CRG mode, REC
CRG is displayed in the lower right portion of the display.
When menus are not displayed, press PRINT. The recorder then prints the real time CRG at 1 mm/s until you press
PRINT again.
Remote
REMOTE mode will activate the recorder at a central station.
Print Trend
To print trends, select [PRINT TREND] from the [RECORDER SETUP] menu.
The recorder prints the trends as they appear on the display. The trend is scaled to appear as it does on the display.
The strip starts with a header, which contains the time and date and prints current readings for all measured
parameters. Each trend is then plotted sequentially. The parameter, vertical scaling, and total time of the trend is
printed with each trend. The vertical axis is printed at the start of the plot, and the horizontal time divisions are
indicated. The resolution of the trend printout is:
1 hour
5 min/cm
8 hours
30 min/cm
12 hours
45 min/cm
24 hours
90 min/cm
The contents of the trend buffers at the settings currently selected in the trend menu are printed. The system prints
the selected trends sequentially. Once the contents of the print trend buffer have been printed, the recorder exits this
mode.
Print Review Trend
To print Review trends:
1.
Press the SYSTEM SETUP key and then select [TREND SETUP] from the setup menu.
2.
Select [REVIEW TREND] to enter the Review trend mode.
3.
Press [PRINT PAGE], the recorder will start printing.
The resolution of the Review trend printout is:
6 Hour view View (6 hours):
30 Minute view (30 minutes):
0.5 hr / cm
2.5 min / cm
The recorder prints the trends as they appear on the display. The trend is scaled to appear as it does on the display.
The strip starts with a header, which contains the time and date and prints current readings for all measured
parameters. Each trend is then plotted sequentially. The parameter, vertical scaling, and total time of the trend is
printed with each trend. The vertical axis is printed at the start of the plot, and the horizontal time divisions are
indicated.
38
Vital-Guard 450C Operation and Maintenance Manual
RECORDER OPERATION
Print NIBP Tab
From [RECORDER SETUP] select [PRINT NIBP TAB] to print the entire contents of the tab buffer in columns as
shown on the display.
Recorder Messages
The following are messages that appear on the display to indicate the status of the recorder.
REC DIRECT: The recorder is ready to print in direct mode.
REC DELAY: The recorder is set for delay mode.
REC TIMED: The recorder is set for timed mode.
REC REMOTE: The recorder is set for remote mode.
DOOR OPEN: The recorder indicates that the door is open.
PAPER OUT: The recorder is out of paper.
REC ERROR: The monitor has detected a recorder failure.
REC INOP: Low battery voltage. Insufficient battery charge to operate the recorder.
REC HR: CRG is selected when respiration is not being monitored. HR only is printed.
Example Printouts
5/11/99
Direct Mode: ECG on Channel 1 and Respiration on Channel 2
5/11/99
Direct Mode: ECG on Channel 1 and SpO2 (Pleth) on Channel 2
Vital-Guard 450C Operation and Maintenance Manual
39
RECORDER OPERATION
5/11/99
Direct Mode: ECG on Channel 1 and P1 on Channel 2
5/11/99
Direct Mode: ECG on Channel 1 and P2 on Channel 2
5/11/99
Direct Mode: ECG on Channel 1 and Channel 2 off
40
Vital-Guard 450C Operation and Maintenance Manual
RECORDER OPERATION
5/11/99
Delayed Mode: ECG on Channel 1 and Channel 2 off
5/11/99
CRG Mode
NIBP Table
Vital-Guard 450C Operation and Maintenance Manual
IBP/TEMP Table
41
ECG MONITORING
ECG MONITORING
When ECG monitoring, the display shows the ECG waveform, the heart rate, heart rate limits (if DISPLAY
LIMITS is selected), lead selection, adult/neonate mode, and alarm indications. Also, a heart symbol, pulse, or BP
symbol flashes each time a heartbeat is detected, depending on whether the rate is being triggered by the ECG,
pleth, or blood pressure waveform.
Safety Considerations
Disposable products are intended for single-use only. Do not attempt to re-use these products.
. For ECG connections use insulated probes. Don’t
ECG Patient connections are electrically isolated Type CF
let patient connections contact other conductive parts, including earth. See instructions for patient connections in
this manual.
This monitor is supplied with protected lead wires. Do not use cables and leads with unprotected lead wires having
exposed conductors at the cable end. Unprotected lead wires and cables may pose an unreasonable risk of adverse
health consequences or death.
Leakage current is limited internally by this monitor to less than 10 μA. However, always consider cumulative
leakage current that can be caused by other equipment used on the patient at the same time as this monitor.
To avoid the potential of electrosurgery burns at ECG monitoring sites, ensure proper connection of the
electrosurgery return circuit as described by manufacturer’s instructions. If improperly connected, some
electrosurgery units might allow energy to return through the electrodes.
Typical Electrode Placement
during Surgical Procedure.
Typical Electrode Placement.
RA
LA
V
RL
RA
LA
RL
LL
LL
Rate meters might continue to count the pacemaker rate during occurrences of cardiac arrest or some arrhythmias.
Do not rely on rate meter alarms. Keep pacemaker patients under close surveillance.
For monitors used in Australia. For continued defibrillation protection and proper respiration monitoring use only
patient cables with 1kΩ series resistors. Respiration circuit is tuned to operate with series resistors.
42
Vital-Guard 450C Operation and Maintenance Manual
ECG MONITORING
Patient Connections
To ensure compliance with safety and performance specifications, always use the patient cables and leads supplied
by Ivy. Other cables might not produce reliable results.
NOTE: We recommend that you always use five-lead cables with monitors having the five-lead option. The use of
many three-lead cables can cause poor quality signals and can inhibit the LEAD OFF alarm. When monitoring three
leads, use the five-lead cable and remove the two unnecessary patient leads.
Use only high quality silver/silver-chloride electrodes or equivalent.
Use the following procedure for ECG monitoring:
1.
Prepare each electrode site and apply the electrodes.
2.
Connect the patient cable to the monitor’s right side panel ECG input.
3.
Attach the leads to the electrodes.
Model 450C monitors with five-lead option can accommodate both three- and five-lead configurations.
4.
Use the procedures described in the following sections for alarm limit settings, lead selection, amplitude
adjustment, and enabling or disabling the filter. See the menu diagram below.
NOTE: For monitors designed to use both three- and five-lead cables, the LEAD OFF alarm tests only those leads
that are relevant to the selected lead setting.
PATIENT
SETUP
ECG/HR
RESP/CO2
SpO2
NIBP
INVASIVE BP
ECG
ON
ECG
OFF
LEAD
I
HR SOURCE
OXIMETRY
FILTER
ON

ECG SIZE
10 m m/mv

Vital-Guard 450C Operation and Maintenance Manual
43
ECG MONITORING
ECG Waveform Amplitude
Use the following procedure to adjust the amplitude (size) of the displayed ECG waveform.
1.
Press the PATIENT SETUP key to display the main menu. Select [ECG/HR] to display the ECG menu.
2.
Use the [ECG SIZE
3.
Press EXIT to return to the normal display.
and
] selections to adjust the ECG size.
HR Alarm Limits
1.
Press the ALARM SETUP key to display the alarm limits menu.
2.
If necessary select [PARAMETER] to display HR limits.
3.
Use the [HIGH
and
] selections to set the high heart rate limit.
4.
Use the [LOW
and
] selections to set the low heart rate limit.
Each time you press a key, the corresponding limit changes by 5 bpm. The current settings are shown on the
display when the front panel DISPLAY LIMITS selection is on.
5.
Press EXIT to return to the normal display.
Lead Selection
1.
Press the PATIENT SETUP key to display the main menu. Select [ECG/HR] to display the ECG menu.
2.
Use the [LEAD] selection to set the lead selection.
Available lead selections for five-lead ECG are I, II, III, aVR, aVL, aVF, and V. Selections for three-lead
ECG are I, II, and III. If you select any other setting with only three leads connected, the monitor activates
the LEAD OFF alarm.
3.
Press EXIT to return to the normal display.
ECG Filter (Adult Mode Only)
1.
Press the PATIENT SETUP key to display the main menu. Select [ECG/HR] to display the ECG menu.
2.
Use the [FILTER] selection to set the filter to on or off. The filter on indication is FILT; the filter off
indication is DIAG (diagnostic). The filter sets the frequency response of the displayed waveform as
follows:
Filtered:
Unfiltered (diagnostic):
0.5 to 35 Hz
0.05 to 100 Hz
In neonate mode, the filter is always on.
3.
44
Press EXIT to return to the normal display.
Vital-Guard 450C Operation and Maintenance Manual
ECG MONITORING
ECG On/Off
The ECG can be turned off to monitor other parameters and not generate a LEAD OFF alarm. LEAD OFF is still
functional for respiration.
1.
Press the PATIENT SETUP key to display the main menu. Select [ECG/HR] to set the ECG to ON or
OFF.
2.
Press EXIT to return to the normal display.
Pacemaker
When a pacemaker has been detected, a P will start flashing in the heart symbol.
Rate meters might continue to count the pacemaker rate during occurrences of cardiac arrest or some arrhythmias.
Do not rely on rate meter alarms. Keep pacemaker patients under close surveillance.
Vital-Guard 450C Operation and Maintenance Manual
45
RESPIRATION MONITORING
RESPIRATION MONITORING
The monitor obtains the respiration signal through the same electrodes used for ECG. The lead selection does not
affect the respiration signal.
WARNING: This monitor is supplied with protected lead wires. Do not use cables and leads with unprotected lead
wires having exposed conductors at the cable end. Unprotected lead wires and cables may pose an unreasonable risk
of adverse health consequences or death.
CAUTION: Do not use patient cables with internal resistors for defibrillation protection when monitoring
respiration. Use of such cables might cause erratic respiration readings.
For proper respiration monitoring, cable capacitance between the RA and LA leads must be 400 ±100 pF.
The following diagram shows the respiration menus.
PATIENT
SETUP
ECG/HR
RESP/CO2
SpO2
NIBP
INVASIVE BP
RESP
IMPEDANCE
RESP
CO2
COINCIDENCE
ON
RESPONSE
NORMAL
APNEA DELAY
20 sec
ZERO
SENSOR
RESP
OFF
SPAN
CHECK
SETTINGS
Respiration Size
The front panel up and down arrows control the relative size of the respiration signal.
Adjust the size so that it is not so large that the peaks of the waveform are clipped on the display but large enough so
that the signal triggers the rate counter.
NOTE: The respiration detector requires the proper size setting to correctly detect respiration activity.
46
Vital-Guard 450C Operation and Maintenance Manual
RESPIRATION MONITORING
Apnea Delay
1.
Press the PATIENT SETUP key to display the main menu. Select [RESP/CO2] to display the respiration
menu.
2.
Use [APNEA DELAY] to set the delay at 5, 10, 15, 20 seconds, or off. Once this alarm is triggered, it
remains on until a respiration is detected or ALARM RESET is pressed.
3.
Press EXIT to return to the normal display.
Coincidence On/Off
In some cases, the true respiration rate and heart rate are the same. If this condition occurs for some time, the
coincidence alarm can become a nuisance. To avoid this problem, the audio and visual coincidence alarms can be
turned off.
Use the following procedure to enable or disable the coincidence alarm.
1.
Press the PATIENT SETUP key to display the main menu. Select [RESP/CO2] to display the respiration
menu.
2.
Use the [COINCIDENCE] selection to turn the coincidence alarm on or off.
3.
Press EXIT to return to the normal display.
Respiration Rate Alarm Limits
1.
Press the ALARM SETUP key to display the alarm limits menu.
2.
Use the [PARAMETER] selection to display respiration rate (RR) limit settings.
3.
Use the [HIGH
and
] selections to set the high respiration rate limit.
4.
Use the [LOW
and
] selections to set the low respiration rate limit.
Each time you press a key, the corresponding limit changes by 5 br/min. The current settings are shown on
the display when the front panel DISPLAY LIMITS selection is on.
5.
Press EXIT to return to the normal display.
Respiration On/Off
1.
Press the PATIENT SETUP key to display the main menu. Select [RESP/CO2] to display the respiration
menu.
2.
Use the [RESP] selection to turn respiration monitoring on or off.
3.
Press EXIT to return to the normal display.
Vital-Guard 450C Operation and Maintenance Manual
47
ETCO2 MONITORING
END TIDAL CO2 MONITORING
General Description
The Ivy Model 450C monitor uses NonDispersive InfraRed (NDIR) gas measurement to determine CO2 gas
concentration. The NDIR measurement method uses fixed (non-scanning) frequencies of infrared (IR) light. NDIR
measurement is based on the phenomenon that higher gas concentrations in a given volume of gas in the path of an
IR light signal create more absorption of the IR light. Therefore, higher concentrations of an IR absorbing gas create
a measurably lower transmission of the IR light.
Components
The following components make up an NDIR instrument. See the illustration below.
IR Source: A source of infrared light.
The infrared light source actually produces a wide range of light at frequencies covering the IR band and extending
in either direction into the visible and ultraviolet spectra.
Sample Cell: The gas to be measured is transported into a sample cell with windows that allow the IR light to pass
through the gas sample. The Sensor Head is mounted on the sample cell, called the Airway Adapter. The Airway
Adapter has a constant volume and path length as seen by the Sensor Head. The Sensor Head and Airway Adapter
windows are chosen for their non-IR absorbing quality.
IR Filter: An optical bandpass filter is used to select a specific band of IR light centered on an exact wavelength.
The optical IR filter’s center wavelength is chosen for the specific gas to be measured (CO2). The monitor also needs
a reference channel which requires a second IR filter.
IR Detector: A lead selenide IR detector converts the IR light transmitted through the sample cell into an output
voltage.
48
Vital-Guard 450C Operation and Maintenance Manual
ETCO2 MONITORING
Sensor Head Components
The diagram below represents the components of the Sensor Head.
Sensor Head Case – The Sensor Head Case is sealed to control internal ambient CO2 concentration and to provide
sufficient moisture resistance for cleaning.
Hybrid Processing Board – The Hybrid Processing Board includes programmed control of:
Lamp Power on/off
Window Heating and Temperature Monitoring
Detector Block Heating and Temperature Monitoring
Multiplexing of Analog Signals
Non-Volatile Memory containing all unit-specific data
Bi-directional Digital Communications with the Signal Processing Board
Vital-Guard 450C Operation and Maintenance Manual
49
ETCO2 MONITORING
Lamp and Reflector Assembly – The lamp is pulsed on and off. Its collimated light beam travels through one Sensor
Head Window, through the Airway Adapter containing the patient sample, through the second Sensor Head Window,
and through the beam splitter.
Heated Windows – The Sensor Head Windows are heated to reduce condensation. A thermistor monitors window
temperature.
Beam Splitter – The Beam Splitter directs part of the light falling on it to the CO2 IR Detector and the balance of the
light to the Reference Channel Detector.
IR Filters – An optical band pass filter is located in front of both IR Detectors. Each filter passes to its detector only
the wavelength of light that is appropriate.
IR Detectors – The IR Detectors convert the light falling on them into analog signals. The CO2 IR Detector’s output
is relative to the CO2 concentration in the Airway Adapter, while the reference channel IR Detector is insensitive to
CO2 concentration.
Detector Preamplifier Board – The outputs of the IR Detectors are amplified and filtered. These analog signals are
fed through the Hybrid Processing Board’s multiplexer to the Signal Processing Board.
Heated Detector Block – The Detector Block is heated to stabilize the IR filter temperature. A thermistor monitors
Detector Block temperature.
50
Vital-Guard 450C Operation and Maintenance Manual
ETCO2 MONITORING
To demonstrate the absorption of specific IR wavelengths by CO2 gas, the chart below shows the amount of IR
transmission for various wavelengths of IR light through CO2. The CO2 filter passes infrared light in the range
between 4.22 and 4.32 microns.
The chart below shows that as the CO2 concentration increases from 2% to 10%, infrared light transmission
decreases. IR light transmission is inversely proportional to the IR light absorption of the gas. By measuring how
much light gets through the sample gas and the CO2 optical filter, we can calculate how much CO2 is present in the
gas sample.
Vital-Guard 450C Operation and Maintenance Manual
51
ETCO2 MONITORING
Complications of NonDispersive InfraRed (NDIR) Monitoring
IR Light Amplitude Changes – Changes in the output of the IR light source directly affect the IR detector’s voltage
output for a given gas sample.
In this monitor, the IR lamp pulses on and off. The on phase is created by a constant voltage, providing a consistent
light output. A consistent IR lamp output is more important than a specific output level.
Short and long term IR lamp amplitude changes are compensated for by use of a reference channel. A reference
channel output change can be assumed to be non-gas related, and common to the CO2 channel. Software data
processing compensates for IR lamp output changes. This same mechanism compensates for any other common
mode source of signal degradation, such as light beam obscuration by patient secretions.
Sample Gas Pressure Changes – Because a change in gas pressure in the Airway Adapter reduces the number of
CO2 molecules in the IR path, without airway adapter pressure compensation, changes in gas pressure would appear
as changes in CO2 concentration.
Typically, the monitor measures CO2 only during the patient’s expired breath phase. At this time, airway adapter
pressure is very close to local barometric pressure. The sensors are factory characterized and calibrated at sea level
where local barometric pressure, and airway adapter pressure is 760 Torr. The system also has the ability to set the
local barometric pressure into the monitor. The Sensor Head stores this pressure value and uses it to correct for the
difference in airway adapter pressure between calibration time and normal use.
Temperature Changes – Changes in the temperature of the IR filters, IR detectors, or sample gas (patient breath or
calibration gas) directly affect the IR detector’s voltage output for a given gas sample. Each of these must be either
held at a constant temperature or monitored for temperature compensation.
During the expired phase, the patient’s breath is slightly lower than body temperature (about 34°C). This relatively
constant temperature relieves the monitor of the need to control sample gas temperature or correct for its changes.
Changes in sample gas flow rate can cause small sample gas temperature changes. The sensors are factory
characterized and calibrated at 4.0 liters per minute.
NOTE: When using bottled calibration gas for span calibration or troubleshooting, its temperature might not be
34°C. The difference between calibration gas temperature and 34°C can cause CO2 measurement errors.
The IR filters and detectors are mounted on the Sensor Head’s detector block. Attached to the detector block is a
thermoelectric heater strip and a thermistor. The heater temperature holds the Detector Block, IR Filters, and IR
Detectors at a constant temperature (approximately 43°C).
Multiple IR Absorbing Gases – N2O, water vapor, and anesthetic agents also absorb IR light and might be present in
some applications. Each of these gases has different IR absorption characteristics, and choosing the correct CO2 and
reference channel IR filters reduces the “cross gas interference” effect.
Non-IR Absorbing Gases Present – The sample gas might also contain O2, N2O, or N2. Although O2 and N2 don’t
absorb IR, they can create NDIR measurement problems via “collision line broadening.”
The monitor uses assumed values for O2 and N2O and uses these values to compensate for “collision line
broadening.” The default values are 21.0% O2 and 0.0% N2O. Different values can be entered and stored in the
Sensor Head.
NOTE: When using a calibration gas for span calibration or troubleshooting, O2 and N2O concentrations must also
be entered.
52
Vital-Guard 450C Operation and Maintenance Manual
ETCO2 MONITORING
Zero Baseline and Span Calibration – The voltage output of the IR Detector has no absolute meaning until it is
established what CO2 detector voltage represents 0.00% CO2 gas concentration and what other voltage represents full
scale CO2 gas concentration (10.00%).
The voltage level when the IR Lamp is off, known as the “dark level,” is the minimum voltage output of the IR
Detector, and is used as the common reference for measuring all other IR Detector output voltages. When there is
0% CO2 in the sample gas, there is minimum IR absorption, and therefore, maximum IR transmission. This
measurement is the maximum output voltage of the IR Detector, and is stored in the system at each Zero Calibration.
All CO2 measurements are between these two values. The factory characterization of each unit includes the
calculation of the system’s CO2 response curve. A field Span Calibration calculates simple multipliers that allow
correction of long term component performance changes.
Nonlinearity – As the IR Detector output voltage decreases with increasing CO2 gas concentration, some
nonlinearities of NDIR technology must be managed.
The nonlinearity of the system’s response curves was determined during its development. This nonlinearity is
described in the resident CO2 response curve, and is automatically compensated for during data processing.
Preparation for Monitoring
Choosing an Airway Adapter
Select one of the two Airway Adapters that are available. Use the Adult Airway Adapter for monitoring patients
with endotracheal tube sizes greater than 4 cc. The Adult Airway Adapter is shown below.
Use the Pediatric/Neonatal Airway Adapter for all other patients. The Pediatric/Neonatal Airway Adapter is shown
below.
Zeroing the Airway Adapter
This procedure should be done at the start of monitoring or any time the message ZERO SENSOR is displayed on
the monitor.
Vital-Guard 450C Operation and Maintenance Manual
53
ETCO2 MONITORING
1.
Insert the Airway Adapter into the Sensor as shown below.
2.
Press the PATIENT SETUP key on the monitor to display the Main Menu, then select [RESP] so that CO2
is displayed.
3.
Make sure that the Airway Adapter is exposed to room air and is not in the patient breathing circuit. Keep
the Sensor and Airway Adapter away from all sources of CO2, such as ventilator exhaust and your own
breath.
4.
Select [ZERO SENSOR].
5.
After approximately 15 seconds, the message ZERO OK appears on the display.
If an error message is displayed, either the Airway Adapter windows are occluded or the Sensor is
defective. Replace the Airway Adapter and repeat the zeroing procedure.
6.
Press the EXIT key to return to the normal display.
Checking Sensor Calibration
1.
Allow the sensor temperature to stabilize.
2.
Place the sensor on the Span Check Cell attached to the sensor cable as shown below. Make sure that the
sensor is fully seated on the cell.
3.
Press the PATIENT SETUP key on the monitor to display the Main Menu, then select [RESP] so that CO2
is displayed.
54
Vital-Guard 450C Operation and Maintenance Manual
ETCO2 MONITORING
4.
Select [CHECK SPAN]. The test takes approximately 15 seconds.
The message SPAN OK appears on the display. If SPAN FAIL appears, do not use the sensor. Have the
sensor checked and calibrated by a biomedical engineer or contact Ivy Biomedical Systems, Inc.
5.
Press the EXIT to return to the normal display.
The Patient Airway Connection
1.
Install the Sensor/Airway Adapter into the patient airway at the proximal end of the airway circuit between
the endotracheal tube and the ventilator circuit wye. The Sensor cable should face away from the patient.
2.
Connect the Sensor cable connector to the ETCO2 input on the monitor.
3.
Verify the presence of a CO2 waveform on the monitor.
Response Setting
Three response settings are available on the Model 450C monitor, slow, fast, or normal. Use the following procedure
to set the response.
1.
Press the PATIENT SETUP key on the monitor to display the Main Menu, select [RESP/CO2], then select
[RESP] so that CO2 menu is displayed.
2.
Select [SETTINGS] to display next menu, then select [NEXT SETTING] until Response is shown.
3.
Select [RESPONSE] until the desired response setting is displayed.
4.
Press the EXIT key to return to the normal display.
Vital-Guard 450C Operation and Maintenance Manual
55
ETCO2 MONITORING
Compensation Settings
As described earlier in this section, settings are available for entering barometric pressure. N2O, and O2 in case the
actual values differ from the default settings, which were entered into the Sensor at the factory. Default settings are
barometric pressure = 760 Torr, N2O = 0.0%, and O2 = 21.0%. If necessary, use the following procedures to change
these settings.
Barometric Pressure
1.
Press the PATIENT SETUP key on the monitor to display the Main Menu, select [RESP/CO2], then select
[RESP] so that CO2 menu is displayed. Then select [SETTINGS].
2.
Use the [BAR PRESS
3.
To return to the default setting (760 Torr) select [DEFAULT].
4.
Press the EXIT key to return to the normal display.
and
] selections to set the barometric pressure to the desired value.
N2O Compensation
1.
Press the PATIENT SETUP key on the monitor to display the Main Menu, select [RESP/CO2], then select
[RESP] so that CO2 menu is displayed. Then select [SETTINGS].
2.
Select [NEXT SETTING] until N2O is displayed.
3.
Use the [N2O
4.
Press the EXIT key to return to the normal display.
and
] selections to set the N2O to the desired value.
O2 Compensation
1.
Press the PATIENT SETUP key on the monitor to display the Main Menu, select [RESP/CO2], then select
[RESP] so that CO2 menu is displayed. Then select [SETTINGS].
2.
Select [NEXT SETTING] until O2 is displayed.
3.
Use the [O2
4.
Press the EXIT key to return to the normal display.
56
and
] selections to set the O2 to the desired value.
Vital-Guard 450C Operation and Maintenance Manual
ETCO2 MONITORING
Trace Scale Setting
Two trace scale settings are available for waveform display, 0 to 50 and 0 to 100. Use the following procedure to set
the trace scale.
1.
Press the PATIENT SETUP key on the monitor to display the Main Menu, select [RESP/CO2], then select
[RESP] so that CO2 menu is displayed. Then select [SETTINGS].
2.
Use the [TRACE SCALE] selection to select the desired trace scale setting.
3.
Press the EXIT key to return to the normal display.
CO2 Alarm Limits
1.
Press the ALARM SETUP key to display the alarm limits menu.
2.
If necessary select [PARAMETER] until ETCO2 is displayed.
3.
Use the [HIGH
and
] selections to set the high CO2 limit.
4.
Use the [LOW
and
] selections to set the low CO2 limit.
Each time you press a key, the corresponding limit changes by 2 Torr. The current settings are shown on the
display when the front panel DISPLAY LIMITS selection is on.
5.
Press EXIT to return to the normal display.
Respiration Rate Alarm Limits
1.
Press the ALARM SETUP key to display the alarm limits menu.
2.
Use the [PARAMETER] selection to display respiration rate (RR) limit settings.
3.
Use the [HIGH
and
] selections to set the high respiration rate limit.
4.
Use the [LOW
and
] selections to set the low respiration rate limit.
5.
Press EXIT to return to the normal display.
Respiration On/Off
1.
Press the PATIENT SETUP key to display the main menu. Select [RESP/CO2], then select [RESP] to set
respiration to OFF.
2.
Press EXIT to return to the normal display.
Vital-Guard 450C Operation and Maintenance Manual
57
ETCO2 MONITORING
Gas Calibration
The Sensor should never need to be calibrated. However, if hospital regulations require a periodic calibration, use
the following procedure.
1.
Make sure the Sensor has been operating for at least 30 minutes with no error conditions reported.
2.
Zero the Sensor on an Adult Airway Adapter. See “Zeroing the Sensor” earlier in this section.
3.
Press the PATIENT SETUP key on the monitor to display the Main Menu, select [RESP/CO2], then select
[RESP] so that CO2 menu is displayed. The select [SETTINGS].
4.
Select [TRACE SCALE] and press TEST at the same time to display a “hidden” menu.
5.
Introduce a 1.0 ±0.1 liter per minute flow of 5% ±0.1% CO2 balance Nitrogen into the Airway Adapter. The
calibration gas should be held at a constant 34°C ±3°C. The flow of gas out of the Airway Adapter must be
free and unobstructed. The reading should be 38 ±2 Torr. If 10% ±0.1% CO2 balance Nitrogen is available,
check the accuracy with this. The reading should be 76 ±3 Torr.
6.
Press the EXIT key to return to the normal display.
58
Vital-Guard 450C Operation and Maintenance Manual
INVASIVE PRESSURE MONITORING
INVASIVE PRESSURE MONITORING
Model 450C monitors configured for invasive pressure monitoring have two channels available for monitoring
pressure. Procedures and specifications are the same for both channels. These procedures refer primarily to P1,
which uses the PRESS 1 connector (the lower pressure connector). Text in parentheses ( ) refers to P2.
Pressure Transducer
To ensure conformance with all safety and performance specifications, use only the recommended pressure
transducers and cable. These are available from Ivy Biomedical Systems. Transducers are not to be immersed in
liquid unless disconnected from the monitor.
WARNING: Disposable transducers are for single use only. Do not attempt to re-use these products.
WARNING: Disposable transducers and accessories identified as STERILE shall be handled in accordance to the
manufacturer’s instructions.
Connecting the Transducer
1.
Connect the pressure cable to the PRESS 1 (PRESS 2) connector on the monitor’s side panel.
2.
Make sure that the catheter tip is at the same height as the dome of the pressure transducer during both
zeroing and monitoring.
3.
If a continuous flow catheter flush system is used, fill the transducer dome and set the transducer at the
desired level before zeroing.
4.
Always do a final zeroing of the transducer after it is connected to the fluid system, positioned, and allowed
to stabilize for a few minutes.
5.
Always follow the instructions that come with the pressure transducer to ensure proper calibration and
operation.
Vital-Guard 450C Operation and Maintenance Manual
59
INVASIVE PRESSURE MONITORING
Zeroing the Transducer
If the pressure transducer has not been zeroed, the monitor displays the message ZERO TRANSDUCER. Use the
following procedure to zero the transducer. See the menu diagram below.
PATIENT
SETUP
ECG/HR
PARAMETER
P1
RESP/CO2
ZERO
TRANSDUCER
SpO2
ALARM TYPE
MEAN
NIBP
TRACE SCALE
0 TO 240
INVASIVE BP
SITE LABEL
P1
1.
Vent the pressure transducer to air
2.
Press the PATIENT SETUP key to display the main menu. Select [INVASIVE BP] to display the pressure
select menu.
3.
Use the [PARAMETER] selection to select P1 or P2 for the transducer being zeroed.
4.
Select [ZERO TRANSDUCER] to zero the transducer.
Once the system has been correctly zeroed, the ZERO TRANSDUCER message disappears.
If the transducer has an offset greater than ±150 mmHg or a varying signal, the monitor does not zero the
transducer, and a tone sounds when you select [ZERO TRANSDUCER].
4.
Press EXIT to return to the normal display.
NOTE: If Barometric Pressure changes then the transducer must be zero again.
Blood Pressure Scale
1.
Press the PATIENT SETUP key to display the main menu. Select [INVASIVE BP] to display the pressure
select menu.
2.
Use the [PARAMETER] selection to select P1 or P2 for the pressure channel being set.
3.
Use the [TRACE SCALE] selection to select the scale. Selections are ±30, 60, 120, 240, and 300.
4.
Press EXIT to return to the normal display.
60
Vital-Guard 450C Operation and Maintenance Manual
INVASIVE PRESSURE MONITORING
Pressure Alarm Type
1.
Press the PATIENT SETUP key to display the main menu. Select [INVASIVE BP] to display the pressure
select menu.
2.
Use the [PARAMETER] selection to select P1 or P2 for the pressure channel being set.
3.
Use the [ALARM TYPE] selection to select the pressure to be used to trigger the blood pressure alarm.
Selections are systolic (SYS), diastolic (DIAS), MEAN, and OFF.
4.
Press EXIT to return to the normal display.
Pressure Alarm Limits
1.
Press the ALARM SETUP key to display the alarm limits menu.
2.
If necessary select [PARAMETER] to display P1 or P2 limits. If the alarm for P1 or P2 has been set to off,
that selection does not appear in the [PARAMETER] selections.
3.
Use the [HIGH
and
] selections to set the high blood pressure limit.
4.
Use the [LOW
and
] selections to set the low blood pressure limit.
Each time you press a key, the corresponding limit changes by 2 mmHg. The current settings are shown on
the display when the front panel DISPLAY LIMITS selection is on.
5.
Press EXIT to return to the normal display.
Expanded Pressure Scale
Use the following procedure to display one or both pressure waveforms on an expanded scale in Zones 3 and 4.
1.
Press the SYSTEM SETUP key to display the mode menu. Then select [DISPLAY SETUP].
2.
Use the [ZONE 4] selection to select expanded pressure display. Selections are Zones 3 and 4 expanded
P1, Zones 3 and 4 expanded P2, and Zones 3 and 4 expanded P1/P2.
3.
Press EXIT to return to the normal display.
Vital-Guard 450C Operation and Maintenance Manual
61
INVASIVE PRESSURE MONITORING
Alternate Pressure Site Labeling and Automatic Scale/Limit Adjustment
Feature (If present)
NOTE: This feature must be enabled by setting the Pressure Site Labels option in the Configuration Menu (see
service manual) to Present. When enabled, the user may select alternate labels by using the SITE LABELING soft
key in the Invasive Pressure Setup menu.
Labels for invasive blood pressures (P1 and P2) may be changed to one of several alternate labels at the users
discretion.
Site label selections for P1 are: P1, ART1, CVP1, PA1, ICP1, UAC1, and UVC1.
Site label selections for P2 are: P2, ART2, CVP2, PA2, ICP2, UAC2 and UVC2.
These labels will appear on both the screen, and the chart recorder printouts.
When a pressure site label is changed, the alarm limits are also automatically adjusted to reflect that change. Limits
are stored separately for each site label.
The default limits are as follows (channel 2 defaults are the same as the channel 1 defaults). Note that one set of
invasive pressure alarm limits is stored regardless of whether Adult or Neonatal mode has been selected.
Site Label
P1
ART1
CVP1
PA1
ICP1
UAC1
UVC1
SYS
High Low
150
80
150
80
-10
10
150
80
20
4
100
40
20
4
DIAS
High Low
100
50
100
50
-10
10
100
50
20
4
60
20
20
4
MEAN
High Low
120
60
120
60
-10
10
120
60
20
4
70
30
20
4
In addition, pressure scales for waveforms and trends are automatically adjusted whenever the user changes pressure
site labels.
The default scale selections are as follows (again, P2 scales are the same choices as P1 scales):
Site Label
P1
ART1
CVP1
PA1
ICP1
UAC1
UVC1
Scale
0 to 240 mmHg
0 to 240 mmHg
-30 to 30 mmHg
0 to 240 mmHg
-30 to 30 mmHg
0 to 120 mmHg
0 to 60 mmHg
PATIENT
SETUP
62
ECG/HR
PARAMETER
P1
RESP/CO2
ZERO
TRANSDUCER
SpO2
ALARM TYPE
MEAN
NIBP
TRACE SCALE
0 TO 240
INVASIVE BP
SITE LABEL
ART1
Vital-Guard 450C Operation and Maintenance Manual
MASIMO SET® PULSE OXIMETRY MONITORING
MASIMO SET® PULSE OXIMETRY MONITORING
Overview
Pulse oximetry allows you to continuously and no invasively monitor a patient’s hemoglobin oxygen saturation. The
oximetry sensor contains two light emitting diodes (LEDs) that transmit specific wavelengths of light, which are
received by a photodetector.
Oxygen saturated blood absorbs light differently as compared to unsaturated blood. Thus the amount of light
absorbed by blood can be used to calculate the ratio of oxygenated hemoglobin to total hemoglobin in arterial blood.
The monitor displays this ratio as percent SpO2. Normal values typically range from 95 to 100% at sea level.
Principles of Operation
The Model 450C Pulse Oximeter is based on three principles:
1.
Oxyhemoglobin and deoxyhemoglobin differ in their absorption of red and infrared light (spectrophotometry).
2.
The volume of arterial blood in tissue and the light absorbed by the blood changes during the pulse
(plethysmography).
3. Arterio-venous shunting is highly variable and that fluctuating absorbance by venous blood is a major
component of noise during the pulse.
The Model 450C Pulse Oximeter as well as traditional pulse oximetry determines SpO2 by passing red and infrared
light into a capillary bed and measuring changes in light absorption during the pulsatile cycle. Red and infrared lightemitting diodes (LEDs) in oximetry sensors serve as the light sources, a photodiode serves as the photodetector.
Traditional pulse oximetry assumes that all pulsations in the light absorbance signal are caused by oscillations in the
arterial blood volume. This assumes that the blood flow in the region of the sensor passes entirely through the
capillary bed rather than through any arterio-venous shunts. The traditional pulse oximeter calculates the ratio of
pulsatile absorbance (AC) to the mean absorbance (DC) at each of two wavelengths, 660 nm and 940 nm.
S(660) = AC(660)/DC(660)
S(940) = AC(940)/DC(940)
The oximeter then calculates the ratio of these two arterial pulse-added absorbance signals:
R = S(660)/S(940)
The value of R is used to find the saturation SpO2 in a look-up table built into the oximeter’s software. The values in
the look-up table are based upon human blood studies against a laboratory co-oximeter on healthy adult volunteers in
induced hypoxia studies.
Vital-Guard 450C Operation and Maintenance Manual
63
MASIMO SET® PULSE OXIMETRY MONITORING
The Model 450C Pulse Oximeter assumes that arterio-venous shunting is highly variable and that fluctuating
absorbance by venous blood is the major component of noise during the pulse. The Model 450C monitor
decomposes S(660) and S(940) into an arterial signal plus a noise component and calculates the ratio of the arterial
signals without the noise:
S(660) = S1 + N1
S(940) = S2 + N2
R = S1/S2
Again, R is the ratio of two arterial pulse-added absorbance signals and its value is used to find the saturation SpO2
in an empirically derived equation into the oximeter’s software. The values in the empirically derived equation are
based upon human blood studies against a laboratory co-oximeter on healthy adult volunteers in induced hypoxia
studies.
The above equations are combined and a noise reference (N’) is determined:
N' = S(660) - S(940) x R
If there is no noise N' =0; then S(660) = S(940) x R which is the same relationship for the traditional pulse oximeter.
The equation for the noise reference is based on the value of R, the value being seeked to determine the SpO2. The
Model 450C software sweeps through possible values of R that correspond to SpO2 values between 1% and 100%
and generates an N' value for each of these R values. The S(660) and S(940) signals are processed with each possible
N' noise reference through an adaptive correlation canceller (ACC) which yields an output power for each possible
value of R (i.e., each possible SpO2 from 1% to 100%). The result is a Discrete Saturation Transform (DSTTM) plot
of Relative output power versus possible SpO2 value as shown in the following figure, where R corresponds to SpO2
= 97%.
Relative
The red light power range is 1mw to 3 mw.
The infrared light power range is 0.7mw to 3mw.
64
Vital-Guard 450C Operation and Maintenance Manual
MASIMO SET® PULSE OXIMETRY MONITORING
The DST plot has two peaks: the peak corresponding to the higher saturation is selected as the SpO2 value. This
entire sequence is repeated once every two seconds on the most recent four seconds of raw data. The Model 450C
SpO2 therefore corresponds to a running average of arterial hemoglobin saturation that is updated every two seconds.
WARNING: The pulse oximeter should NOT be used as an apnea monitor.
WARNING: The pulse oximeter should be considered an early warning device. If a trend towards patient
deoxygenation is indicated, blood samples should be analyzed by a laboratory co-oximeter to completely
understand the patient’s condition.
Caution: Use only sensors specified by Ivy Biomedical Systems, Inc. If you use sensors other than those
manufactured by the Masimo corp., it might degrade performance and could damage the pulse oximeter.
Caution: Tissue damage can be caused by incorrect application or use of the sensor. Check the sensor site
frequently. Do not allow the sensor to remain on one site for a prolonged period of time, especially when
monitoring neonates. Refer to the sensor’s specific instructions.
Caution: Never attach a SpO2 sensor on a limb being monitored with a blood pressure cuff or a limb with
restricted blood flow.
Caution: A poorly applied sensor might give incorrect saturation values. Reapply the sensor.
Caution: Choose a site with sufficient perfusion to ensure accurate oximetry values.
Caution: Certain nail aberrations, nail polish, fungus, etc. might give inaccurate oximetry values
Caution: Do not use the Model 450C Pulse Oximeter or sensors during Magnetic Resonance Imaging (MRI).
Pulse Oximetry Sensors
To ensure conformance with all safety and performance specifications, use only the recommended pulse oximetry
sensors.
Monitoring Procedure
Use the following procedure for monitoring pulse oximetry:
1.
Choose a site that is well perfused and provides proper alignment of the LEDs and receiving photodetector
Select a site that has unrestricted blood flow
Do not restrict blood flow when securing sensor with tape
Select an appropriate sensor and apply it to the patient, following the directions for use provided with the sensor
Do not select a site near potential electrical interference (electrical cords, for example).
2.
Connect the cable to the monitor’s SpO2 input.
3.
Plug the sensor into the extension cable.
4.
Use the procedures described in the following sections for alarm limit settings, response mode settings, and
enabling or disabling the SpO2 alarm.
Vital-Guard 450C Operation and Maintenance Manual
65
MASIMO SET® PULSE OXIMETRY MONITORING
PATIENT
SETUP
ECG/HR
SpO2 ALARM
ON
RESP/CO2
AVERAGING
8 SEC
SpO2
SENSITIVITY
NORMAL
NIBP
INVASIVE BP
SpO2 Alarm Limits
1.
Press the ALARM SETUP key to display the alarm limits menu.
2.
Use the [PARAMETER] selection to display SpO2 % limit settings.
3.
Use the [HIGH
and
] selections to set the high SpO2 % limit.
4.
Use the [LOW
and
] selections to set the low SpO2 % limit.
Each time you press a key, the corresponding limit changes by 1 %. The current settings are shown on the
display when the front panel DISPLAY LIMITS selection is on.
5.
Press EXIT to return to the normal display.
Response Mode Settings
1.
Press PATIENT SETUP to display the menu. Select the [SpO2] to display the SpO2 menu..
2.
Use the [SpO2 ALARM] selection to set the Alarms on/off.
3.
Use the [AVERAGING] selection to set the response time – 6, 8, 10, 12, 14 or 16 sec.
4.
Use the [SENSITIVITY] selection to set the sensitivity level. (Options – Normal and High). The High
sensitivity level should be used when the clinician wants to have absolute low perfusion performance of the
Masimo SET® and is willing to sacrifice some sensor off detection capability.
5.
Press EXIT to return to the normal display.
Default Values
The following defaults are set:
1.
High SpO2 limit – set to 100.
2.
Low SpO2 limit – set to 90.
3.
Averaging – set to 8 sec.
4.
Sensitivity – set to Normal.
66
Vital-Guard 450C Operation and Maintenance Manual
MASIMO SET® PULSE OXIMETRY MONITORING
Pulse Oximetry with ECG Off
When you are monitoring oxygen saturation but not ECG, set the ECG/HR menu setting to OFF to inhibit unwanted
LEAD OFF alarms. With these settings, the heart rate on the display is derived from the pulse waveform and has a
flashing pulse symbol for each detected heartbeat.
Use the following procedure to set ECG to OFF.
1.
Press the PATIENT SETUP key to display the main menu. Select [ECG/HR] to display the ECG/HR menu.
2.
Use the [ECG] selection to set ECG to off. When ECG is off, heart rate can be computed from the pleth
waveform.
NOTE: To set heart rate limits when using any heart rate source, use the ALARM SETUP menus as described in the
ECG Monitoring section of this manual.
Pulse Oximetry Considerations
If the accuracy of any measurement does not seem reasonable, first check the patient’s vital signs by alternate means
and then check the Model 450C Pulse Oximeter for proper functioning.
NOTE: The waveform displayed by the SpO2 signal is not proportional to the pulse volume.
Inaccurate measurements may be caused by:
•
Incorrect sensor application or use
•
Significant levels of dysfunctional hemoglobin (e.g., carboxyhemoglobin or methemoglobin) – see note.
•
Intravascular dyes such as indocyanine green or methylene blue – see note.
•
Exposure to excessive illumination, such as surgical lamps (especially ones with a xenon light source),
bilirubin lamps, fluorescent lights, infrared heating lamps, or direct sunlight (exposure to excessive
illumination can be corrected by covering the sensor with a dark or opaque material)
•
Excessive patient movement
•
Venous pulsations
•
Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line
The Model 450C can be used during defibrillation, but the readings may be inaccurate for a short time.
Loss of pulse signal can occur in any of the following situations:
•
The sensor is too tight
•
There is excessive illumination from light sources such as a surgical lamp, a bilirubin lamp or sunlight
•
A blood pressure cuff is inflated on the same extremity as the one with a SpO2 sensor attached
•
The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia
•
There is arterial occlusion proximal to the sensor
•
The patient is in cardiac arrest or is in shock
NOTE: A pulse oximeter measures functional hemoglobin saturation. This is different to an arterial blood gas that
measures fractional hemoglobin saturation. Differences can occur when significant levels of HbCO
(carboxyhemoglobin), HbMET (methomoglobin) indocynanine green or methylene blue exist.
Vital-Guard 450C Operation and Maintenance Manual
67
NIBP MONITORING
NIBP MONITORING
General Description
The Model 450C uses the oscillometric method for measuring non-invasive blood pressure (NIBP). The monitor
automatically inflates and then slowly deflates the pressure cuff. Pressure fluctuations from the arterial pulse are
detected by the monitor and used to determine the systolic, diastolic, and mean pressure. The monitor displays
numeric values for this measurement on the left side of the monitor. No waveform is displayed.
For proper NIBP monitoring, use the accessories listed in the Accessories section of this manual.
NIBP has three modes of operation: manual, automatic, and stat. Press the CYCLE key to set the desired mode. In
manual mode, a single measurement is taken by pressing the START key. Automatic mode generates measurements
at regular intervals until the STOP key is pressed. Intervals are programmable for 1, 2, 3, 4, 5, 10, 15, 20, 30, or 60
minutes.
In STAT mode, pressing the START key initiates a series of measurements taken for a five-minute period.
A tabular display shows data as it is measured at the time of each NIBP reading. The display contains columns for
time of day; systolic, diastolic, and mean NIBP; peripheral pulse measured from the NIBP cuff; respiration rate;
SaO2; invasive blood pressures and temperature. The latest four readings are displayed, and a scrolling procedure
allows you to display previous readings.
NIBP Monitoring Considerations
•
Do not use this device on a patient who has an artificial heart or lung.
•
Measurement might be impossible in case of continuous noise or when the patient’s pulse is highly irregular.
•
When this device is used to measure the blood pressure of critically conditioned patients, the pressures should
be measured again by stethoscope or palpation if the indicated values are doubted.
•
Since the pulse rate is measured at a distal part of the body, it might be different from the rate computed from
the ECG.
•
Use of the correct size cuff and proper wrapping help ensure accurate measurements.
•
Do not restrict or compress the external tubing.
•
Compression of the cuff or hose can cause noise, which might interfere with the measurement.
•
Try to keep the patient relaxed and quiet during the measurement. Muscle tension or talking can interfere with
the measurement.
•
Check to be sure that use of the equipment does not result in prolonged impairment of the circulation of the
patient.
•
The repeated use of the Short Term Automatic Mode (STAT MODE) for more than five minutes can caused
tissue damaged.
68
Vital-Guard 450C Operation and Maintenance Manual
NIBP MONITORING
Cuff Placement
Proper cuff size and placement are essential to assure accurate blood pressure measurement.
The American Heart Association recommends that cuff sizes should be at a length-to-width ratio of about 2:1,
ensuring that if the bladder width is 40% of the arm circumference, the bladder length will encircle 80% of the arm.
Also, care should be taken to center the bladder directly over the brachial artery.
Cuff too narrow
Cuff proper width
On the left, the bladder width is small for the arm, and full cuff pressure is never applied to the artery which can
caused an erroneously high pressure result. On the right, the bladder width is adequate for the arm, and full cuff
pressure is applied to the brachial artery.
Cuffs for thighs are available for large patients or those where neither arm is available for cuff placement. Blood
pressure measured at the thigh is typically 20-30 mmHg higher than blood pressure measured at the upper arm.
Monitoring Procedure
1.
Connect the air hose fitting to the NIBP connector on the monitor’s side panel. To remove move the sleve
back and pull the air hose.
Also connect the other end of the air hose to the cuff fitting.
NIBP
connector
Air Hose Fitting
Air Hose Fitting
Cuff Fitting
Neonate
Sleve
Monitor’s Side
Vital-Guard 450C Operation and Maintenance Manual
Cuff’s Side
69
NIBP MONITORING
2.
Apply the cuff to the patient as follows.
•
Wrap the cuff snugly, with room between the cuff and arm for two fingers. If the cuff is wrapped too
loosely, it cannot be inflated properly causing possible errors in measured values.
•
Wrap the bare arm when possible. If clothing oppresses the arm, measured values might not be reliable.
The end of the cuff should fall inside the range marked on the inside of the cuff. If it doesn’t, use a different
size cuff.
Heart
Cushion
•
Keep the cuffed part of the arm at the same level as the heart. If it is above the level of the heart the blood
pressure measurement will be lowered. If it is below the heart level, the blood pressure will be raised (due
to the physical weight of the blood).
•
Do not compress the cuff or the rubber hose externally. Compression of the cuff or the rubber hose will
cause measurement error. Rest the arm on a cushion to prevent compression against the body.
•
Keep the patient as still as possible during the measurement.
WARNING: Always check the mode setting before starting an NIBP procedure. Monitoring a neonate in adult
mode can cause injury to the patient.
3.
Select the NIBP monitoring mode as described in the following procedures.
4.
Press the START key to initiate measurements.
5.
Press STOP/DEFLATE to deflate the cuff and end all NIBP procedures.
If the monitor cannot obtain a valid reading in manual mode, it displays a small number 1 near the NIBP mean value
display and repeats the measurement. The monitor might try up to four measurements, depending on the problem,
until a valid reading is obtained, displaying the number of tries after each measurement.
After several unsuccessful tries, the monitor cancels the measurement and displays the CHK CUFF alarm message.
NIBP Graph
To display the NIBP graph, press the PATIENT SETUP key, select [NIBP], and select [DATA FORMAT] until it
is set at GRAPH.
The oscillation profile assists the clinician in determining the validity of the NIBP measurement. An erratic or flat
profile suggests that excessive motion or other artifact may have affected the measurement. A profile shaped like a
bell curve represents a measurement not affected by motion or other artifact.
70
Vital-Guard 450C Operation and Maintenance Manual
NIBP MONITORING
NIBP Messages
If the monitor detects any problems with the measurement, one of the following messages is displayed. In most cases,
the monitor attempts to get a good reading several times before displaying the message. The number of attempts is
displayed near the NIBP readings.
Alarm Messages
CHECK CUFF:
Check the cuff and hose connector for leaks.
Check for patient movement and check the cuff for proper attachment.
MOTION ARTIFACT:
Check for patient movement.
LOW INFLATION:
Check for insufficient monitor pressure because of high blood pressure.
ABNORMAL PULSE:
Check for patient movement or arrhythmic patient.
TIME EXCEEDED:
Check the cause of obstructed air discharge, such as patient movement or bent or crimped
hose.
HIGH PULSE RATE:
Check the cause of obstructed air discharge, such as patient movement or bent or crimped
hose.
OVER PRESSURE:
Check for folded arm of patient or bent hose.
WEAK PULSE:
Check for improper attachment of cuff.
CHECK CUFF HOSE:
Check for improper cuff or improper attachment of cuff.
CHECK CUFF SIZE:
Check for neonatal cuff being used in ADULT mode.
Non-Alarm Messages
PRESS START:
Indicates that you must press the NIBP START key to resume measurements.
MVT (movement):
Indicates that the monitor detected some patient movement during the measurement and
the readings might be less reliable.
INOPERABLE:
The monitor has detected an NIBP failure.
Tabular Scrolling
Use the following procedure to scroll the tabular display to view previous readings.
1.
Press the PATIENT SETUP key and select [NIBP] to display the NIBP menu.
2.
Select [SCROLL
displayed.
3.
Press EXIT to return to normal display. The data selected remains on the display until the next NIBP
reading or until the scrolling procedure is repeated.
] to scroll back or [SCROLL
Vital-Guard 450C Operation and Maintenance Manual
] to scroll forward until the desired readings are
71
NIBP MONITORING
Clear Tabular Data
Use the following procedure to delete all tabular data in memory without turning power off.
1.
Press the SYSTEM SETUP key then select [TREND SETUP].
2.
Select [CLEAR TRENDS] to delete all trend data in the monitor, including tabular data.
3.
Select [CONFIRM] to proceed with clearing data or [CANCEL] to retain all data.
NIBP Mode Select
Press the CYCLE key in the front panel NIBP area repeatedly until the desired mode is displayed in the zone where
NIBP values are displayed.
Available modes are 1, 2, 3, 4, 5, 10, 15, 20, 30, or 60 minute automatic cycles, STAT, and OFF (MANUAL).
NIBP Alarm Settings
You must select one of the NIBP values (systolic, diastolic, or mean) to be used to trigger the NIBP alarm and then
set up upper and lower alarm limits.
NOTE:When the monitor alarms, there is a delay of <0.3s to the signal output connector for remote equipment.
1.
Press PATIENT SETUP then select [NIBP] to display the NIBP menu.
2.
Use [ALARM TYPE] to select the pressure to be used to trigger the NIBP alarm.
(systolic). DIAS (diastolic), MEAN and OFF.
3.
Press EXIT to return to the normal display.
4.
Press ALARM SETUP.
5.
Use the [PARAMETER] selection to select NIBP.
6.
Use the [HIGH
and
] selections to set the high NIBP limit.
7.
Use the [LOW
and
] selections to set the low NIBP limit.
Selections are SYS
Each time you press a key, the corresponding limit changes by 2 mmHg. The current settings are shown on
the display when the front panel DISPLAY LIMITS selection is on.
8.
Press EXIT to return to the normal display.
NIBP Calibration
1.
To enter the calibration mode, press TEST, ALARM RESET, and START at the same time.
2.
Momentarily block the NIBP input and then attach the cuff to a manometer or other pressure measuring
device.
3.
Pump in pressure, and compare the manometer reading to the reading on the monitor display.
4.
Press STOP/DEFLATE to exit the calibration mode.
5.
With a cuff attached to the NIBP input, enter the calibration mode as described above. One minute after the
monitor reaches a maximum pressure reading, leakage should be <5 mmHg/3 min.
72
Vital-Guard 450C Operation and Maintenance Manual
TEMPERATURE MONITORING
TEMPERATURE MONITORING
Use only YSI Series 400 temperature probes or equivalent. Probes and cables are available from Ivy Biomedical
Systems.
WARNING: Disposable probes are for single use only. Do not attempt to re-use these products.
1.
Connect the temperature cable to one of the TEMP connectors on the monitor’s side panel.
2.
Locate the probe on the patient and wait for the probe temperature to stabilize.
The temperature is displayed at the bottom of the screen. The label T1 refers to TEMP 1, the connector to the
left; the label T2 refers to TEMP 2.
When both temperature probes are connected to the monitor, the difference between the two temperatures (∆T)
is displayed in Zone 5.
Temperature Alarms Description
Alarm limits for each of the two temperature channels may be set via the Alarm Setup menu, in the same manner as
setting other limits.
Temperature alarms may be set to off, (The default is off).
Temperature alarm messages appear in the alarm overflow area (lower left corner of the screen).
Temperature alarms are never latched, even if the latched alert option is enabled for other parameters.
ALARM
SETUP
PARAMETER
T1

HIGH
39.2


LOW
36.0

Vital-Guard 450C Operation and Maintenance Manual
73
ALARM MESSAGES
ALARM MESSAGES
The following is a list of alarms and ranges for which they can be set. When the alarm triggers, the alarm indicator
flashes on the display and the audio alarm tone sounds. Pressing ALARM RESET turns off any alarms that have
triggered.
If the measurement returns to within the limits, the alarm turns off.
The ALARM PAUSE key disables the alarms for two minutes or until ALARM PAUSE is pressed again.
High heart rate: 15 to 300 bpm in 5 bpm increments
Low Heart Rate: 15 to 300 bpm in 5 bpm increments
High Resp Rate: 5 to 150 br/min in 5 br/min increments
Low Resp Rate: 5 to 150 br/min in 5 br/min increments
High ETCO2: 10 to 100 Torr in 2 Torr increments
Low ETCO2: 10 to 100 Torr in 2 Torr increments
NIBP High: Selectable systolic, diastolic, or mean 10 to 300 mmHg in 2 mmHg increments
NIBP Low: Selectable systolic, diastolic, or mean 10 to 300 mmHg in 2 mmHg increments
NOTE: For additional NIBP alarm conditions, see the NIBP Monitoring section of this manual.
IBP High: Selectable systolic, diastolic, or mean; -40 to 300 mmHg in 2 mmHg increments
IBP Low: Selectable systolic, diastolic, or mean; -40 to 300 mmHg in 2 mmHg increments
NOTE: Invasive and non-invasive pressures can be set to alarm on systolic, diastolic, or mean.
High SpO2: 50 to 100%in 1% increments
Low SpO2: 50 to 100% in 1% increments
Asystole: R-to-R interval >6 seconds
No Pulse: When triggering from pleth or invasive pressure, the interval between heartbeats exceeds six seconds
Lead Off: Detached lead or offset potential ≥0.5 V. This alarm is not reset by the ALARM RESET key
Check Leads: Electrode impedance is >4 kΩ
Coincidence: When QRS and respiration triggers are synchronized for eight consecutive events
Apnea: Absence of detectable respiration activity for 5, 10, 15, or 20 seconds, selectable
74
Vital-Guard 450C Operation and Maintenance Manual
MONITOR TESTING
MONITOR TESTING
Press and hold the TEST key to test the internal functions of the monitor. You should do this each time you begin
monitoring a patient.
The TEST function generates a 1 mV ECG pulse at 70 bpm, a respiration signal at 70 br/min and a coincidence
check, which are displayed on the display and are available at the rear panel connector. The monitor also displays
invasive pressure readings of 150/50 and temperature readings of 40°C. If these indications are not present, contact
qualified service personnel.
To test the visual and audio alarms turn on the monitor and press the ALARM PAUSE key. Check that the LEAD
OFF messages is displayed on the ECG channel and the audio alarm is on. While pressing the TEST key check for
the following sequence to happen: 1) LEAD OFF message disappear, 2) Monitor starts counting QRS, 3) HR<40
message appears on ECG channel, audio alarm is triggered, 4) RR<10 message appears on ECG channel, audio
alarm is triggered, 5) A few seconds after the HR<40 message disappeared, a COINCIDENCE message is displayed
on the Respiration channel. Stop pressing the TEST key and press ALARM OFF, all audio alarms should be silenced
and visual alarms should be displayed.
Under normal operation, no internal adjustment or recalibration is required. Safety tests and internal adjustments
should be done by qualified personnel only. Safety checks should be performed at regular intervals or in accordance
with local or governmental regulations. In the event that internal adjustment or recalibration is necessary, refer to the
Operation and Maintenance Manual for this equipment.
Note:
If no display is visible on the monitor, the monitor is inoperable. Contact qualified personal. When ECG input is
>0.5 V, a inoperable condition is indicated by flashing LEAD OFF indicator on the display.
Vital-Guard 450C Operation and Maintenance Manual
75
MAINTENANCE AND CLEANING
MAINTENANCE AND CLEANING
The Monitor
When necessary, clean the exterior surfaces of the monitor with a cloth or swab dampened with a warm water and
mild detergent solution. Do not allow liquids to enter the interior of the instrument.
CAUTION:
•
Do not autoclave, pressure sterilize, or gas sterilize the monitor.
•
Do not soak or immerse in any liquid.
•
Use cleaning solution sparingly. Excessive solution can flow into the monitor and cause damage to internal
components.
•
Do not touch, press or rub the display and covers with abrasive cleaning compounds, instruments, brushes,
rough surface materials, or bring them into contact with anything that could scratch the display or the
covers.
•
Do not use petroleum based or acetones solutions, or other harsh solvents, to clean the monitor.
Patient Cables
Do not autoclave the patient cables.
Wipe the cables using a mild detergent solution. Never submerge the cables in any liquid or allow liquids to enter
the electrical connections.
Reusable Pressure Transducer
1.
Remove the dome.
2.
Clean the transducer as recommended by manufacturer.
CAUTION: Do not immerse any part of the electrical connector of the transducer in the cleaning solution at any
time. Examine the outer sheath of the cable for perforations. If the outer covering is damaged in any way, do not
immerse the cable in the cleaning solution; this might result in moisture entering the transducer case, which is vented
through the cable.
WARNING: If liquids are permitted to enter the electrical connector, resistance between the electrical element and
the transducer case should be checked to ensure that the leakage current is below acceptable levels for safe use on
patients.
76
Vital-Guard 450C Operation and Maintenance Manual
MAINTENANCE AND CLEANING
Cleaning and Reuse of SpO2 Sensors
Reusable sensors can be cleaned as follows:
1.
Remove the sensor from the patient.
2.
Clean the sensor as recommended by the manufacturer.
NOTE: If the sensor fails to track the pulse consistently, the sensor might be incorrectly positioned. Reposition the
sensor or choose a different monitoring site.
CAUTION: Do not soak or immerse the sensor in any liquid solution. Do not sterilize any sensor by irradiation,
steam, or ethylene oxide.
Reusable Temperature Probes
Clean the probes as recommended by the manufacturer.
CAUTION: Probe plugs should not be immersed.
CAUTION: Never boil the temperature probe. Autoclaving might cause the insulation to fail, and might also cause
the probe to give inaccurate readings.
EtCO2 Sensor and Airway Adapters
The Sensor Head, Head Cables(s) and Airway Adapters should be cleaned as recommended by manufacturer.
CAUTION Do not immerse the Head or Cables in cleaning solution or any other liquids. The Sensor Head can not be
sterilized.
Preventive Maintenance
ECG & RESP
Check that:
•
Cables and Leads are clean and intact.
•
The LEAD OFF message is displayed when the patient cable is connected, but the patient leads are not
connected. Connecting the patient leads together should make the message disappear.
SpO2
Check that:
•
Cables and Leads are clean and intact.
•
The probe is functioning (place it on a finger).
Temperature
Check that:
•
Temperature probe and cable are clean and intact.
•
Both channels are working (connect a temperature probe).
Vital-Guard 450C Operation and Maintenance Manual
77
MAINTENANCE AND CLEANING
Invasive Blood Pressure
Check that:
•
Cables and pressure transducers are clean and intact.
•
The device recognizes cable connection (activated on the display).
•
The zeroing of transducers is working correctly.
Non Invasive Blood Pressure
Check that:
•
The cuff and hose are clean and intact.
•
A few seconds after turning the monitor on the message “PRESS START” appears in the NIBP zone.
EtCO2
Check that:
•
The cable and sensor are clean and intact.
•
The probe is functioning. Connect the sensor to the monitor and verify the light on the head of the sensor turns
on and the “WARMING UP” messages is displayed on the screen.
78
Vital-Guard 450C Operation and Maintenance Manual
ACCESSORIES
ACCESSORIES
ECG/Respiration
Standard Three-Lead
Three-lead patient cable
590170
Patient leads
590162
Optional Five-Lead
Five-lead patient cable
590167
Patient leads
590168
NOTE: For proper respiration monitoring, cable capacitance between the RA and LA leads must be 400 ±100 pF.
For units used in Australia
Three-lead patient cable with 1kΩ series resistors
Patient leads
590197
590162
Five-lead patient cable with 1kΩ series resistors
Patient leads
590207
590168
ETCO2
Adult Airway Adapter
590186
Pediatric/Neonatal Airway Adapter
590188
ETCO2 Sensor with Span Chech and one Adult Airway Adapter
2315-00-01
Invasive Pressure
Pressure transducer reusable
590252
Cable for reusable pressure transducer
590251
Disposable transducer domes (Box of 50)
590253
Disposable pressure transducer (Box of 10)
590016
Cable for disposable pressure transducer
590235
Vital-Guard 450C Operation and Maintenance Manual
79
ACCESSORIES
Non-Invasive Pressure (NIBP)
Disposable Cuffs
Neonatal Cuff, bladderless (3–6 cm)
590239
Neonatal Cuff, bladderless (6–11 cm)
590240
Neonatal Cuff, bladderless (8–15 cm)
590241
Child Small Cuff, full bladder (13–20 cm)
590267
Adult Small Cuff, full bladder (18–26 cm)
590268
Adult Cuff, full bladder (26–35 cm)
590269
Reusable Cuffs
Infant Cuff (8–14 cm)
590270
Child Cuff (13–20 cm)
590271
Adult Small Cuff (18–26 cm)
590272
Adult Cuff (26–35 cm)
590273
Adult Large Cuff (32–42 cm)
590274
Adult Thigh Cuff (42–50)
590275
Cuff Hoses
For use with all disposable cuff (yellow hose).
2686-00-01
For use with all reusable cuffs (black hose).
2689-00-01
Temperature
Reusable temperature probe – rectal/esophageal
590003
Reusable temperature probe – skin
590005
Reusable cable for YSI disposable probes
590000
Disposable temperature probe YSI Series 400 or equivalent
590001
Disposable skin temperature probe
590002
Recorder
Recorder Paper, box of 10 rolls
80
590035
Vital-Guard 450C Operation and Maintenance Manual
ACCESSORIES
SpO2 Reusable Sensors
Adult LNOP DC-1 Finger Sensor, Reusable
590227
SpO2 Disposable Sensors
Neonatal LNOP-Neo, Disposable Sensor, Box of 20
590223
Pediatric LNOP-Pdt, Slender Digit Disposable Sensor, Box of 20
590222
Neonatal Pre-term LNOP-NeoPt, Disposable Sensor, Box of 20
590224
Adult LNOP-Adt, Disposable Sensor, Box of 20
590221
SpO2 Patient Cables
Patient Cable, 8 ft PC08 with LNOP Sensor Connector
590220
Patient Cable 12 ft PC12 with LNOP Sensor Connector
590230
SpO2 Starter Kits
Starter Kit with 2 LNOP-Adt and 2 LNOP-Pdt sensors, and
12’ PC12 sensor cable
590225
Starter Kit with 2 LNOP-Neo and 2 LNOP-NeoPt sensors, and
8’ PC08 sensor cable
590226
Starter Kit with Adult LNOP-DC1 finger sensor, reusable and
PC08 sensor cable
590228
Disposal
Disposal of devices or consumables must be done in accordance with local, state, and federal laws and regulations.
WEEE Directive 2002/96/EC.- Do not dispose of WEEE products in general waste. At the end of life of product
contact IVY Biomedical Systems, Inc. customer service for return instructions.
Vital-Guard 450C Operation and Maintenance Manual
81
REAR PANEL OUTPUTS
REAR PANEL OUTPUTS
Analog Output Connector
Pin
Signal
Output
1
ECG X1000
1 V = 1 mV input
0.5 to 40 Hz filtered
0.05 to 100 Hz diagnostic (Adult Mode only)
2
ECG
Adult:
Neonate:
3
RESPIRATION
2.5 V peat-to-peak maximum (±1.25 V), variable with front panel control
4
IBP1
10 mV/mmHg, 0 mmHg = 0 mV ±50 mV (after Auto-Zero)
5
IBP2
10 mV/mmHg, 0 mmHg = 0 mV ±50 mV (after Auto-Zero)
6
PLETH
Waveform, 1.2 V peak-to-peak maximum. No sensor = +0.6 V
7
HEART RATE
16.6 mV/bpm, 300 bpm max, 60 bpm = 1 V, 120 bpm = 2 V, 300 bpm = 5 V
8
SpO2
50 mV/% SpO2 100% max, 100% = 5 V
9
No Connection
10
GROUND
Ground
11
EtCO2
50 mV/mmHg
12
ALARM OUT
High +5 V (>2.4 V) when an alarm is present
13
GRAPH OUT
Relay closure for remote recorder
14
GRAPH COM
15
UNIT ON
16
NURSE CALL
17
N.C. COMMON
18
LEAD OFF ALARM
19-25
No Connection
X1000 (1 V = 1 mV input)
X2000 (1 V = 0.5 mV input)
38mmHg = 1.90 ± 50 mV
Relay closure for Nurse Call option
High +5 V (>2.4 V) when a lead off alarm is present
For the mating connector, use Ivy P/N 188505, 25-pin plug for 22-26 AWG wire. Use RFI/EMI shielded cable
clamp, Ivy P/N 210032, 15-conductor shielded cable, Ivy P/N 600005, and a ferrite bead, Ivy P/N 150025, at the
monitor end of the cable.
82
Manual
Vital-Guard 450C Operation and Maintenance
REAR PANEL OUTPUTS
COM 1 & COM 2 Connectors
Pin
Signal
Output
1
No Connection
2
RECEIVE DATA
Not Used
3
TRANSMIT DATA
RS-232 ±9 V (±5 V min.)
4
Not Used
5
SIGNAL GROUND
6
Not Used
7
REQUEST TO SEND
Not Used
8
CLEAR TO SEND
Not Used
9
Not Used
Reference for COM 1 signals
For the mating connector, use Ivy P/N 179004, 9-pin receptacle for 22-26 AWG wire. Use RFI/EMI shielded cable
clamp, Ivy P/N 210030, and 5-conductor shielded cable, Ivy P/N 600002.
RJ45 Connector (option in place of COM 2)
Pin
Signal
1
TRANSMIT DATA
2
SIGNAL GROUND
3
RECEIVE DATA
4
Not Used
5
Not Used
6
SIGNAL GROUND
7
Not Used
8
Not Used
Vital-Guard 450C Operation and Maintenance Manual
RJ45
VIEWED FROM
REAR OF MONITOR
1
83
REAR PANEL OUTPUTS
Configuration Menu
To access the configuration menu hold both the TEST button and the EXIT button in while switching the power ON.
Use the arrow keys ( ⇑ ⇓ ) to scroll to the function to be changed.
Use the CHANGE OPTION key to select the desired option.
CAUTION:
Configuration is carefully preset at the factory based on which parameters and options have
been installed. Extreme care should be taken when altering any of these settings and all users
should be informed of changes
Configuration Menu
FUNCTION
ECG
3 LEAD 60Hz
OPTIONS
3 LEAD 50Hz
5 LEAD 60Hz
5 LEAD 50Hz
RESP
ABSENT
IMPEDANCE
CO2
IMP + CO2
SpO2
ABSENT
NELLCOR MP204
MASIMO
NELLCOR MP405
NIBP
ABSENT
COLIN
TEMP
ABSENT
SINGLE
DUAL
INVASIVE BP
ABSENT
SINGLE
DUAL
RECORDER
ABSENT
PRESENT
COM 1
ABSENT
HTS820TC
HTS820
COM2/ETHERNET
ABSENT
CPC REV C
ANSAR REV B
AUDIO OFF FEATURE
ABSENT
PRESENT
LATCHING ALARMS
ABSENT
PRESENT
PRIORITY ALARMS
ABSENT
SCHEME 1
INITIAL LIMITS
DEFAULTS
STORED
TEMP UNITS
DEGREES F
DEGREES C
KPA
MMHG
ABSENT
PRESENT
KPA
MMHG
PERCENT
3.0 KHZ
3.5KHZ
4.0KHZ
NONE
ONE HOUR
PERMANENT
DATE FORMAT
MM/DD/YYYY
DD/ MM/YYYY
KEYPAD TYPE
17 KEYS
20 NO LIGHT
PRESSURE UNITS
PRESSURE SITE LABELS
CO2 UNITS
ALARM TONE
TAB RETENTION
84
Manual
THERMALINK
SCHEME 2
TWO HOUR
20 + LIGHT
Vital-Guard 450C Operation and Maintenance
TESTING AND TROUBLESHOOTING
TESTING AND TROUBLESHOOTING
General
Press and hold the TEST key to test the internal functions of the monitor. You should do this each time you begin
monitoring a patient.
The TEST function generates a 1 mV ECG pulse at 70 bpm, a respiration signal at 70 br/min and a coincidence check,
which are displayed on the display and are available at the rear panel connector. The monitor also displays invasive
pressure readings of 150/50 and temperature readings of 40°C. If these indications are not present, continue with
troubleshooting procedures.
Under normal operation, no internal adjustment or recalibration is required. Safety tests and internal adjustments should
be done by qualified personnel only. Safety checks should be performed at regular intervals or in accordance with local or
governmental regulations.
ECG Test
Equipment Required
ECG Simulator (Fogg Model M310 or equivalent)
Initial Settings for the Test
HR high alarm limit
HR low alarm limit
Lead select
Filter
ECG Size
Alarms
150
50
I
ON
10 mm/mV
ON (ALARMS OFF message not displayed)
Test Procedure
1.
2.
Connect the patient cable and simulator to the monitor and set the simulator as follows:
•
Heart rate of 80 bpm
•
Amplitude of 1.0 mV
Observe the following on the monitor display:
•
ECG Lead I is displayed with an amplitude of approximately 10 mm and is noise free
Press FREEZE to check the waveform if necessary.
•
Heart rate of 80 ±1 bpm
•
The heart next to the heart rate indicator flashes with every QRS complex displayed.
Vital-Guard 450C Operation and Maintenance Manual
85
TESTING AND TROUBLESHOOTING
3.
Verify that ECG Leads I, II, III, aVR, aVL, aVF, and V display properly as shown below.
4.
Check the LEADS OFF alarm by removing one lead at a time from the simulator.
5.
Change the heart rate on the simulator to 40 bpm. Notice the following:
6.
•
The HR <50 message appears on the display.
•
Heart rate value of 40 ±1 bpm is displayed.
Press ON/OFF. Notice the following:
•
Audible alarm stops
•
Alarms Off message is displayed.
7.
Press ON/OFF. The Alarms Off message goes off.
8.
Set the simulator heart rate to 200 bpm. The HR >150 message is displayed, and an audible alarm sounds.
9.
Set the simulator to 80 bpm.
Pressure Test
Equipment Required
Transducer Simulator (Fogg Model BP48C or equivalent)
Test Procedure
1.
Press and hold the monitor's TEST button. The monitor's pressure reading should be 150/50.
2.
Connect the simulator output to the monitor's pressure input.
3.
Set the simulator output to 0 mmHg.
4.
Zero the pressure channel.
5.
Set the simulator to 100 mmHg output.
6.
Set the monitor's blood pressure range to 120.
7.
Check for the proper level on the monitor's display.
8.
Repeat for the second pressure channel.
86
Vital-Guard 450C Operation and Maintenance Manual
TESTING AND TROUBLESHOOTING
Respiration Test
Equipment Required
Respiration Simulator (Fogg Model M98 or equivalent)
Test Procedure
Connect the BLACK (LA) and RED (LL) leads to the black simulator terminal. Connect the WHITE (RA) lead to the
white simulator terminal.
1.
Set the monitor to NEONATAL mode and the front panel RESP SIZE to 50%.
2.
Set the simulator as follows:
Power:
Rate:
Variation:
Toggle Switch:
Base Impedance:
ON
50 br/min
0.5 Ω
NORM.
1 kΩ
The triggering should be consistent and the rate should be 50 ±1 br/min.
3.
Set the monitor to ADULT mode. Check for the displayed amplitude to drop approximately 50%.
The triggering should be consistent and the rate should be 50 ±1 br/min.
4.
Turn the simulator power off. Set the APNEA DELAY to 20 seconds. The respiration must not trigger. Check
for an APNEA alarm in 20 seconds.
Temperature Test
Equipment Required
Temperature Simulator (Fogg Model TP-69 or equivalent)
Test Procedure
1.
Connect the SERIES 400 TEMPERATURE OUTPUT of the simulator to the monitor's TEMP input.
2.
Set the output of the simulator to 37°C. The monitor's temperature reading should be 37°C ±0.1°C.
3.
Check the monitor's output at 20°C and 40°C.
4.
Repeat for the second temperature channel.
Pulse Oximeter Test
Equipment Required
Sensor with cable
Test Procedure
1.
Connect the patient cable to the monitor and the sensor to your finger.
2.
Check for waveform, SpO2 reading, pulse rate, bar indicator, and audio tone.
Vital-Guard 450C Operation and Maintenance Manual
87
TESTING AND TROUBLESHOOTING
ETCO2 Test
Equipment Required
EtCO2 sensor with cable.
Test Procedure
1.
Make sure the EtCO2 waveform is displayed in either zone 2, 3, or 4.
2.
Check for the NO SENSOR message to be displayed.
3.
Plug in the EtCO2 sensor and place the airway adaptor on it. The sensor should illuminate and the display should
indicate that the sensor is warm-up. Warm-up should conclude in 3-5 minutes.
4.
Through the RESP/CO2 menu, zero the sensor and check for the ZERO OK message to be displayed.
5.
Remove the airway adaptor from the sensor and make sure the OFF AIRWAY message is displayed.
6.
Place the span check cell on the sensor. Through the RESP/CO2 menu run the SPAN CHECK test, and wait for
the SPAN OK message.
7.
With the span check cell still on the sensor, check for the EtCO2 value and waveform to be within the percentage
value of the Barometric Pressure (760mmHg) printed on the span check cell.
(Span Check Value = %value * 760/100)
NIBP Test
Equipment Required
Cylinder, such as bottle or coffee can
NIBP cuff, hose, and Y-connector
Manometer or other pressure measuring device
Test Procedure
1.
To enter the calibration mode, press TEST, ALARM RESET, and START at the same time.
2.
Momentarily block the NIBP input and then attach the cuff to a manometer or other pressure measuring device.
3.
Pump in pressure, and compare the manometer reading to the reading on the monitor display.
4.
With a cuff attached to the NIBP input, enter the calibration mode as described above. One minute after the
monitor reaches a maximum pressure reading, leakage should be <5 mmHg/3 min.
Recorder Test
Use the following procedure to verify that the recorder is operating properly.
1.
Set the recorder to print ECG and respiration waveforms and set the recorder to TIMED mode.
2.
Press PRINT and then press and hold the TEST key for at least 30 seconds.
88
Vital-Guard 450C Operation and Maintenance Manual
TESTING AND TROUBLESHOOTING
The recorder prints waveforms that match the waveforms on the display. The recorder continues printing for 30
seconds and then stops.
Battery Test
1.
Disconnect the power cord from the monitor so that it is operating on battery.
2.
In approximately 30 seconds, the battery indicator should appear in the lower right corner of the display.
3.
Connect the power cord. Verify that the charging light starts flashing (pre-charged qualification) and that after a
few minutes either one of the following conditions occur: the charging light stays on (fast charging) or the
charged led goes on (maintenance charging).
4.
If the fault light goes on replace the battery immediately.
Isolation and Leakage Tests
Use a leakage tester or use the ac voltage range on a digital voltmeter to measure across a 1 kΩ resistor as shown below.
CAUTION: Be careful to connect the circuit below only between the measurement points indicated in the following
instructions. Do not connect this circuit directly across line voltage. Full line voltage would damage the measurement
circuit and could possibly damage the monitor and/or the digital voltmeter.
DMM
A.C. VOLTS
0.015µf
1kΩ
TO CIRCUIT
BEING
MEASURED
1.
Measure between a known ground and the monitor’s ground post on the rear panel. With the line cord properly
grounded, with the ground removed, and with the line cord reversed, the readings should be
<100 µA.
2.
With the line cord grounded, measure between shorted patient leads and the hot side of the ac line. The reading
should be <10 µA.
Vital-Guard 450C Operation and Maintenance Manual
89
TESTING AND TROUBLESHOOTING
System Error Messages
Numeric messages might appear in the lower right corner of the display if the processor detects a problem with the
monitor. If a message appears, contact Ivy Biomedical Systems Service Department.
When you call, have the following information available:
Model Number
Serial Number
Numeric Error Message
Procedure being performed when the message appeared
System Error Messages & Solutions
Error
90
Meaning
Solution
1
The EPROM checksum test has failed.
Replace U57 (2600-00-04) on Mother Board (5143-0001) board or replace U53 (2601-00-04) on Mother Board
(5143-00-01) board or replace Mother Board (5143-0001).
2
The static RAM test has failed.
Replace Mother Board (5143-00-01).
3
The non-volatile (clock chip) RAM test has failed.
Replace Lithium Battery (330115) or replace Mother
Board (5143-00-01).
4
An undefined interrupt has been detected.
Replace Mother Board (5143-00-01)
5
A system timing error has been detected.
Replace Mother Board (5143-00-01)
6
No DSP data has been received.
Replace Mother Board (5143-00-01).
7
A DSP communication error (framing, overrun,
etc) has been detected.
Replace Mother Board (5143-00-01)
8
Bad DSP data has been received (ie: a bad header
byte).
Replace Mother Board (5143-00-01)
9
An unknown error has been detected.
Replace Mother Board (5143-00-01)
Vital-Guard 450C Operation and Maintenance Manual
TESTING AND TROUBLESHOOTING
Simplified Block Diagram
ECG
&
RESP
MUX
A/D
DSP
Uart
ECG
Temp 1
Temp 2
ISOLATED
INPUT
BOARD
Resp
Press 1
Serial
Inputs
Press 2
Analog &
COM ports
Temp &
Pressure
Recorder
Processor
SpO2
EtCO2
NIBP
90-230VAC
47-63 Hz
SpO2
Module
EtCO2
Module
Front
Panel
SpO2
Power
Video
Generator
Display
EtCO2
Power
NIBP
Module
Mains
Filter
Power
Supply
Battery
Charger
BAT 1
BAT 2
450C Mother Board (5143-00-01)
Vital-Guard 450C Operation and Maintenance Manual
91
REPLACEMENT PARTS
REPLACEMENT PARTS
Printed Circuit Boards
Part
Ivy Part Number
450C Monitor Mother Board
Isolated Input Board (ECG/RESP, 2Temps & 2Press)
Isolated Input Board (ECG/RESP)
Isolated Input Board (ECG/RESP & 2Temps)
Isolated Input Board (ECG/RESP & 2Press)
EtCO2 Interface Board
5143-00-01
5144-00-01
5144-01-01
5144-02-01
5144-03-01
5108-00-01
Modules and Assemblies
Part
Ivy Part Number
Replacement Battery
EOS Power Supply 60W
Masimo Sp02 Module (MS-3)
Colin M2000 NIBP Module
EtCO2 Signal Processor Board
EtCO2 Power Board
LSI Telemetry Transmitter Module Kit
Ethernet Module
Color Display
Recorder
Rubber Feet
Pneumatic Hose for NIBP
Window
Speaker Assembly
Fan Assembly
Rear Enclosure (Rear bezel)
Front Bezel
Handle
110007
580025
580022
580024
580011
580012
580016
580026
570013
380004
680010
560000
2457-00-13
2616-00-10
2620-00-10
2566-00-11
2565-00-11
2570-00-11
Fuses
CAUTION: To ensure safety and reliability, always replace fuses with the type specified.
Part
Power Entry Module (2)
Mother Board – F1
Mother Board – F2, F3 & F4
92
Manual
Description
Ivy Part Number
1.6ASB, 5x20mm
5ASB, ASSY, SMT.
3/8ASB, ASSY, SMT
290021
290503
290502
Vital-Guard 450C Operation and Maintenance
REPLACEMENT PARTS
Cables and Harnesses
Part
Ivy Part Number
Masimo Internal Sp02 Cable
Andros Internal EtCO2 Cable
EtCO2 Signal Power Supply Harness Assembly
EtCO2 Signal Interface Harness Assembly
Backlight Inverter
NIBP Interface Ribbon
Recorder Interface Ribbon
COM 1 & COM 2 Interface Ribbon
Telemetry Interface Ribbon
Power Supply Input
Power Entry Module
Display Interface Ribbon
Analog Output Interface Ribbon
Horizontal Front Panel
Vertical Front Panel
Harness Interface Ribbon
2451-01-01
2317-01-01
2335-00-01
2336-00-01
2470-01-10
2609-00-10
2608-00-10
2614-00-10
2611-00-10
2612-00-01
2613-00-01
2615-00-10
2619-00-10
2605-00-10
2577-00-10
2466-00-10
Labels
Part
Ivy Part Number
Front Panel Label, Vertical
Front Panel Label, Horizontal
Ivy Model Number Label (Masimo)
Rear Panel Label
Input Connector Label (ECG/Resp, NIBP, Sp02, 2Temp, 2Press, EtCO2)
Input Connector Label (ECG/Resp, NIBP, Sp02, 2Temp, 2Press)
Input Connector Label (ECG/Resp, NIBP, Sp02 , 2Temp)
Input Connector Label (ECG/Resp, NIBP, Sp02 , 2 Press)
Input Connector Label (ECG/Resp, NIBP, Sp02)
2458-00-10
2596-00-10
2599-01-10
2598-00-10
2597-01-10
2597-03-10
2597-05-10
2597-07-10
2597-09-10
Software
Part
Mother Board – U57
Mother Board – U53
Mother Board – U26
Mother Board – U98
Description
Ivy Part Number
450C Program (1 of 2)
450C Program (2 of 2)
DSP software
450C Video
2600-00-04
2601-00-04
2636-00-04
2634-00-04
Vital-Guard 450C Operation and Maintenance Manual
93
DISASSEMBLY AND ASSEMBLY
DISASSEMBLY AND ASSEMBLY
Covers
WARNING: To prevent equipment damage and possible personal injury, always remove the power cord from the
instrument.
1.
Remove the four screws that secure the back half frame to the front half frame (front bezel).
2.
Carefully unplug the following harnesses from the 450C mother board: Power harness, ground connector,
COM1 & COM2 harness, ECGX1000, ANALOG OUTPUT harness, NIBP harness, Recorder harness,
SpO2 Connector, EtCO2 connector (if available) and Ethernet connector (if available).
The monitor should be open in two, the front half and the back half. Attached to the front half is the Mother board
(5143-00-01), the color display and both horizontal and vertical faceplates. The back half contains the batteries,
NIBP pump, Analog and digital outputs, the EOS power supply and the isolated input board (5144-00-01).
Rear Panel
Outputs
Fan
NIBP Pump
Batteries
Handle
EOS Power
Supply
Back Half
Frame
Ethernet
Module
450C Mother
Board
EtCO2 Interface
Board
Isolated
Input
Board
EtCO2
Module
SpO2 Module
Front Half
Frame
Speaker
94
Color
Display
Vital-Guard 450C Operation and Maintenance Manual
DISASSEMBLY AND ASSEMBLY
Printed Circuit Boards
450C Mother (5143) Board
1.
Open the monitor as describe in the previous section.
2.
Take the front half of the monitor and remove the three screws (located at each corner of the 5143 board)
that hold the board to the front half.
3.
Disconnect the front panel interface ribbon cable and the speaker harness.
EtCO2 Module Interface (5108) Board
The EtCO2 module interface board is mounted on the Mother board next to the SpO2 module and the EtCO2
module.
1.
Open the monitor as described above.
2.
Disconnect the EtCO2 signal interface harness and EtCO2 power supply harness from J204 and J202 on the
EtCO2 mother interface board.
3.
Remove the three nylon screws that hold the EtCO2 interface board and carefully pull it up to release it
from the Mother board.
EtCO2 Module
The EtCO2 module is mounted on the Mother board next to the SpO2 module.
1.
Open the monitor as described above.
2.
Disconnect the EtCO2 signal interface harness and EtCO2 power supply harness from J204 and J202 on the
EtCO2 module.
3.
Remove the three screws that hold the EtCO2 module to the Motherboard.
Masimo SpO2 Module
The Masimo module is attached to the 5143 board (the module is approximately at the middle of the 5143 board).
1.
Open the monitor as described above.
2.
Unplug the ribbon cable from the Masimo module.
3.
Remove the three nylon screws that secure the module.
Vital-Guard 450C Operation and Maintenance Manual
95
DISASSEMBLY AND ASSEMBLY
Isolated Input Board
NOTE: The screws securing the side panel to the monitor are covered by an adhesive label with the identification
for the connectors on the panel. To gain access to the screws, you should carefully lift the label to expose the screws.
If the label is damaged, it can be replaced by ordering Ivy P/N 2597-01-10.
1.
Remove the four screws that secure the front half (front bezel) and back half of the monitor and unplug all
the ribbon cables and harnesses from the 450C motherboard.
2.
Remove the nuts from the outside of the temperature jacks (if present) and peel the adhesive label.
3.
Unscrew the EtCO2 input connector (if present) and remove it from the monitor’s frame.
4.
Remove the four screws that secure the isolated input board to the monitor’s frame.
5.
Remove the two screws that hold the SpO2 input cable and pull it out from the monitor’s frame.
Front Panel
Use the following procedure to replace the horizontal front panel assembly.
1.
Remove the four screws that secure the front half (front bezel) and back half of the monitor and unplug all
the ribbon cables from the 450C motherboard.
2.
Take out the three screws that secure the motherboard to the front half and carefully remove the board.
3.
Carefully disconnect the 18 pin Interface ribbon cable and the speaker harness from the motherboard.
4.
Disconnect the 10-pin ribbon cable that connects the vertical and horizontal faceplates.
5.
Remove the four hex nut that hold the horizontal front panel to the front bezel (front half).
6.
Pull the horizontal front panel out.
CAUTION: Do not bend any pins when removing the front panel connector from the Mother Board. Bent pins
might require replacement of the Mother Board.
Use the following procedure to replace the vertical front panel assembly.
1.
Remove the four screws that secure the front half (front bezel) and back half of the monitor and unplug all
the ribbon cables from the 450C motherboard.
2.
Take out the three screws that secure the motherboard to the front half and carefully remove the board.
3.
Carefully disconnect the 18 pin Interface ribbon cable and the speaker harness from the motherboard.
4.
Disconnect the 10-pin ribbon cable that connects the vertical and horizontal faceplates.
5.
Remove the two hex nut that hold the vertical front panel to the front bezel (front half).
96
Vital-Guard 450C Operation and Maintenance Manual
DISASSEMBLY AND ASSEMBLY
6.
Pull the vertical front panel out.
CAUTION: Do not bend any pins when removing the front panel connector from the Mother Board. Bent pins
might require replacement of the Mother Board.
EOS Power Supply
1.
Open the monitor as described above.
2.
Remove the Isolated Input Board.
3.
Remove the four screws that secure the EOS Power Supply to the monitor’s back frame.
4.
Disconnect the connectors from the EOS Power Supply, including the ground wire connector.
Recorder Assembly
1.
Open the recorder door and remove the paper from the recorder.
2.
Loosen the two captive Philips-head screws inside the recorder housing.
3.
Pull the Recorder Assembly straight out from the monitor.
NIBP Module
The NIBP module is mounted in the back half frame of the unit under the NIBP mounting bracket.
1.
Remove the four screws that secure the front half (front bezel) and back half of the monitor and unplug all
the ribbon cables and harnesses from the 450C motherboard.
2.
Remove the recorder (if present) as described above.
3.
Remove the two hex-nut that hold the ground points over the NIBP mounting bracket and remove all
ground wires.
4.
Remove the four screws (located to the sides of the bracket) that secure the NIBP mounting bracket to the
back half frame and take the bracket out; the NIBP pump is attached to it.
5.
Once the bracket is out, remove the four screws placed at each corner of the bracket and release the NIBP
pump.
Battery Removal
The batteries are secure to the back half frame of the monitor by the Battery bracket. The Battery bracket has to be
removed to replaced the batteries.
1.
Remove the four screws that secure the front half (front bezel) and back half of the monitor and unplug all
the ribbon cables and harnesses from the 450C motherboard.
Vital-Guard 450C Operation and Maintenance Manual
97
DISASSEMBLY AND ASSEMBLY
2.
Unplug the wires at the batteries terminals.
3.
Remove the four screws that hold the battery bracket to the unit. Two screws are located on the rear panel
and the other two are in the bottom of the back half frame.
4.
Remove the battery bracket and the batteries.
Display Assembly
The Display Assembly is mounted on the front half frame under the motherboard. To remove the Display Assembly
follows the next steps:
1.
Remove the four screws that secure the front half (front bezel) and back half of the monitor and unplug all
the ribbon cables from the 450C motherboard.
2.
Remove the motherboard as described previously. Unplug the speaker harness and the front panel ribbon
cable from the motherboard.
3.
Remove the four screws that hold the display to the front half frame and lift the display up.
NOTE: When installing back the display, the screws must be tight at 60 lb/in
4.
Remove the four screws that secure the Display Power Supply to the front bezel.
NOTE: The Display Assembly and Display Power Supply must be replaced as a pair.
Speaker
1.
Open the monitor as described in the first section of this chapter.
2.
Remove the Mother Board and unplug the speaker harness.
3.
Pull the speaker out from its location.
98
Vital-Guard 450C Operation and Maintenance Manual
BOARD LAYOUT DIAGRAMS
BOARD LAYOUT DIAGRAMS
1. EtCO2 Interface Board
5108-00-02
2. 450C Monitor Mother Board (front)
5143-00-02
3. 450C Monitor Mother Board (back)
5143-00-02
4. Isolated Input Board (solder side)
5144-00-02
5. Isolated Input Board (connector side)
5144-00-02
Vital-Guard 450C Operation and Maintenance Manual
A
BOARD LAYOUT DIAGRAMS
This page is intentionally left blank.
Vital-Guard 450C Operation and Maintenance Manual
SCHEMATIC DIAGRAMS
SCHEMATIC DIAGRAMS
1. EtCO2 Module Interface Board
2. 450C Monitor Mother Board
a. 5143 Main
b. Analog
c. Bypass
d. COMM
e. Color CPU Main
f. DSP Processor
g. DSP Power Supply
h. ECG Front End
i. Ethernet Power Supply
j. Glue
k. CPU Memory
l. Power Supply
m. Resp
n. SpO2 Power Supply
o. Temp Pressure Front End
p. Video FPGA
3. Isolated input board
Vital-Guard 450C Operation and Maintenance Manual
5108-00-01
5143-00-01
(1 of 16)
(2 of 16)
(3 of 16)
(4 of 16)
(5 of 16)
(6 of 16)
(7 of 16)
(8 of 16)
(9 of 16)
(10 of 16)
(11 of 16)
(12 of 16)
(13 of 16)
(14 of 16)
(15 of 16)
(16 of 16)
5144-00-01
B
SCHEMATIC DIAGRAMS
This page is intentionally left blank.
Vital-Guard 450C Operation and Maintenance Manual
7
6
5
4
3
2
ISOLATED
RXD_B
VCC
4
D
VCC
5 RN1 6
1K
C
U5
CNY17-3
7 RN1 8
1K
ET_RES
B++
E7
2
D1
L1
25UH
C3
+ C2
100UF
25V
IGET
C4
1500UF
25V
+ .0068UF
1KV CD
2252-00-10
T1
2
.
SB5100
R6
620 OHM
2W 5%
3
4
R14
150 OHM
10W 5%
H.S.
.
1
64KHz
E5
C9
1UF MP
C8
100UF 25V
730MA
D
H.S.
Q4
1G
IRF540
D2
MBR350
+
Q1
2
2N3906
C10
.1UF MP
R2
10K
.047UF MP
8
7
6
5
REF
R_MID
GND-F
GND-S
IGET
COLLECTOR
COMP
V+
R_TOP
10000PF
NPO
R4
1
2
3
4
+ C6
22UF
TANT
3
Q2
S
R8
6.98K
3
Q
7
DIS
6
THR
R10
10K
U7
ICM7555
TR
R13
1K
C7
2
C5
1000PF
NPO
47PF NPO
R11
10K
CV
5
R9
10K
2
2N3903
B
Q3
3
U6
CNY17-3
C11
D
4
C1
2
IGET
TP2
1
IGET
5
2
B
2
2N3903
3
3
R12
100 OHM
B++
3
4
U2
BNX002-01
1
1 1
+5VIET
TP1
1
IGET
E4
4
2
4
VCC
E3
8
5
IGET
3 RN1 4
1K
9 RN1 10
1K
R
2
1K
C
+5VIET
E2
U4
CNY17-3
R3
R5
2K
E1
V+
+5VIET
TXD_B
V-
+5VIET
J204
173011
2
IGET
+5VIET
1
+5VIET
1 RN1 2
1K
5
IGET
1
2
3
1
5
U3
CNY17-3
R1
2K
4
J202-4
D
VCC
TTL RECEIVE DATA
TTL TRANSMIT DATA
TTL EXTERNAL RES
J202-3
DATE
8-28-97
11-4-97
6-25-99
8-9-02
NON-ISOLATED
2336-00-01
J202-1
J202-2
BY
GP
GP
GP
GP
1
D
1
REVISIONS
DESCRIPTION
REVISED PER ECN 691.
REVISED PER ECN 709.
REVISED PER ECN 957.
REVISED PER ECN 1201.
REV
01
02
03
04
1
8
R7
10K
E6
D_GND
D
DGND
TP3
33.2K
U1
LT1431CN8
IVY BIOMEDICAL SYSTEMS INC.
A
NOTES:
1. ALL RESISTORS ARE 1/4W, 1% U.O.S.
Title
ETCO2 MODULE INTERFACE BOARD ASSEMBLY
2. LAST USED:
Size
3. NOT USED:
B
Date:
8
7
6
5
A
11 BUSINESS PARK DRIVE
BRANFORD, CT. 06405
(203) 481-4183
4
3
Document Number
Rev
04
5108-00-01 (-03 SCHEMATIC)
Thursday, March 10, 2005
2
Sheet
1
of
1
1
REVISIONS
REV
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
GP
REVISED PER ECN 1274.
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
18
19
20
21
22
23
24
+6.5V
TP50
PRESS_TEST
+ 3
- 2
+5VI
11
SLF7032T-101MR45-2 from TDK
1.5V
R297
10K
L8
-6.5V
1
temp1_plus
+6.5V
13
V-
4
GND
5
V+
D
R299
10K
2
I
100UH
SMT
U87A
MAX327CSE
IN
1
1M
R300
1.2K .1%
50PPM
R301
2.8K .1%
50PPM
100UH
SMT
+ C181
6.8UF TANT
20V 3528
C182
.1UF CER
0805
I
+6.5V
I
I
7
AV = 100
8
9
16
L9
temp1_neg
R298
1.82K
.1% 50PPM
12
1
I
C178
.1UF CER
0805
+6.5V
C179
.1UF CER
0805
-6.5V
S
R296
3
1
3
2
Q32
MMBF4393
PRES_TEST
U86A
TL034ACD
4
+5_Vref
DATE
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
40C_sel
20C_sel
temp_cal#
40C_sel
20C_sel
temp_cal#
Vcc
Out1
3
In2
Out2
6
In3
Out3
11
In4
Out4
14
In1
2
7
10
15
Gate1
Gate2
Gate3
Gate4
4
Latch
I
R302
2 -
C180
1UF MP
2824
1.5V
TP11
TEMP 1
5
3 +
10K
6
U89
4 1 8 AD621AR
Temp Scaling
15 deg C = -3.5V
45 deg C = 4.21V
I
R304
931 OHM
.1%
50PPM
R303
9.31K
.1% 50PPM
temp1_out
25.7mV/deg C
-6.5V
R305
GND
10K
I
R306
9.31K
.1% 50PPM
U88
DS1640S
5
+
C217
1UF TANT
16V 3216
I
I
I
I
1.5V
L10
temp2_plus
+5VI
U87D
MAX327CSE
6
S
PRESS_EN
IN
8
R312
1
PRESS_EN
10K
temp2_neg
100UH
SMT
U87B
MAX327CSE
100UH
SMT
C193
.1UF CER
0805
100UH
SMT
L13
3
S
14
15
2
R318
12 +
13 -
40.2K
C190
.22UF MP
2824
6
U92
AD621AR
7
8
AV = 100
press_1
14
10
S
TP14
PRESS 2
R322
10 +
9 -
40.2K
C195
.22UF MP
2824
I
R323
22M
5%
SG2+
SG2+
C201
.1UF CER
0805
100UH
SMT
L15
100UH
SMT
In3
Out3
11
In4
Out4
14
2
7
10
15
Gate1
Gate2
Gate3
Gate4
4
1
2
2
Latch
8
C185
1UF MP
2824
1.5V
6
418
temp2_out
25.7mV/deg C
U91
AD621AR
Temp Scaling
15 deg C = -3.5V
45 deg C = 4.21V
I
R315
931 OHM
.1%
50PPM
R314
9.31K
.1% 50PPM
TP12
TEMP 2
5
2 -
-6.5V
R317
GND
I
U90
DS1640S
10K
R319
9.31K
.1% 50PPM
+
C218
1UF TANT
16V 3216
I
3
3
4
4
5K
5
5
12K
press_2
press_2
C229
.1UF CER
0805
6
U86C
TL034ACD
6
34.872K
7
7
8
8
12K
9
9
34.872K
10
10
195K
90.1875K
C230
.1UF CER
0805
R324
332K
R326
143K
I
4
1
5
C200
.1UF CER
0805
2
3
6
+6.5V
R328
499K
R329
499K
I
U86B
TL034ACD
5 +
6 -
7
R325
1K
1.5V
C198
.33UF NPO
0805
+6.5V
U93
AD621AR
+5VI
AV = 100
+6.5V
C202
.1UF CER
0805
9
7
3 +
10K
I
7
8
C199
.1UF CER
0805
Out2
I
R311
I
-6.5V
R327
22M
5%
+
SG2-
L14
C197
.1UF CER
0805
-
SG2-
I
In2
6
AV = 100
+5_Vref
+5VI
C196
.1UF CER
0805
1
I
11
D
U87C
MAX327CSE
I
3
+6.5V
I
I
I
R321
499K
Vcc
Out1
In1
5
9
R320
499K
40C_sel
20C_sel
temp_cal#
8
16
I
RN1
435023
R307
1.82K
.1% 50PPM
12
+5_Vref
press_1
U86D
TL034ACD
Av = 3.906
+6.5V
C194
.1UF CER
0805
I
I
IN
SG1+
C192
.1UF CER
0805
TP13
PRESS 1
9.775mV = 1mmHg
I
4
1
5
L12
C191
.1UF CER
0805
+
SG1+
SG1-
R316
22M
5%
-
SG1-
C189
.1UF CER
0805
-6.5V
+ C186
6.8UF TANT
20V 3528
C187
.1UF CER
0805
D
I
IN
R313
22M
5%
C188
.1UF CER
0805
R309
2.8K .1%
50PPM
L11
2
+5VI
I
R308
1.2K .1%
50PPM
1
Q33
MMBT3904LT1
C183
.1UF CER
0805
+6.5V
C184
.1UF CER
0805
16
D
R310
10K
3
7
100UH
SMT
NOTES:
+5VI
+5_Vref
+5_Vref
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
I
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
IGND
I
I
-6.5V
-6.5V
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
DWN:
MARK TUCCILLO 9/5/01
CHK:
APR:
GP 9/5/01
RAM 9/5/01
Title
Temp Pressure Front End
Size
D
Date:
Document Number
Rev
24
5143-00-01 (-03 SCHEMATIC)
Friday, May 27, 2005
Sheet
15
of
16
1
+12V
5
D41
MBRS340T3
Vin
+5VREF
Switch
1
2
Feedback
Q35
CMPT3906
3
1
2
2
U78
LM2587S-ADJ
C168
120UF TANT
20V SIZE-R
+
D42
8
11
4
Comp
3
VO
+15VI_SpO2
+ C166
6.8UF TANT
35V 7343
I
MBRS340T3
10
1
+
C169
120UF TANT
20V SIZE-R
D43
T5
2604-00-10
Gnd
2
MBRS340T3
U80
LM3411M5-5.0
VI
VO
R348
10K
R279
5
1K
C171
.1UF CER
0805
2
5
D
TP18
DGND
4
Out
Comp
IN
1
GND
2
-15VI_SpO2
C170
6.8UF TANT
+ 35V 7343
NC
3
TP46
+5VI_SpO2
I
+5VI_SpO2
+
C172
180UF TANT
16V SIZE-R
LOW ESR
TP47
IGND_SpO2
TP48
IGND_SpO2
1
U81
CNY17-3SM
4
3
U79
LM2990S-15
3
C173
.1UF CER
0805
TP45
-15VI_SpO2
1
U12-14
R347
49.9K
9
4
5
D
VI
MBRS340T3
7
2
R278
4.02K
D40
BZT52-C16
16V
C204
.22UF MP
100V 2220-N
GND
R346
680 OHM
2W SMT
+
12
C167
180UF TANT
16V SIZE-R
LOW ESR
TP44
+15VI_SpO2
U77
LM2940IMP-15
D39
GND
F2
3/8ASB
DGND
IGND_SpO2
I
D
R280
10K
2
VCC
Vcc
R281
1
VCC
2
C174
.1UF CER
0805
Iso_SaO2_Xmitt
3
5
GND
R283
499 OHM
D
VCC
Grounded
1K
U82
HCNW2601-300
6
SaO2_Xmitt
Q28
CMPT3906
3
8
7
R282
2K
I
Isolated
2
R284
10K
R285
1K
1
Q29
CMPT3906
VCC
2
3
SaO2_Rec
U83
HCNW2601-300
8
7
R286
1K
6
3
GND
5
Iso_SaO2_Rec
C175
.1UF CER
0805
R287
499 OHM
I
D
VCC
2
R288
10K
R289
1K
U84
HCNW2601-300
1
Q30
CMPT3906
VCC
2
3
SaO2_Rts
REVISIONS
8
7
6
3
GND
5
REV
R290
1K
Iso_SaO2_Rts
C176
.1UF CER
0805
R291
499 OHM
18
19
20
21
22
23
24
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
GP
REVISED PER ECN 1274.
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
DATE
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
I
D
VCC
NOTES:
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
2
R292
10K
1K
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
1
Q31
CMPT3906
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
U85
HCNW2601-300
3
R293
SaO2_Reset
2
8
7
R294
1K
6
3
5
Iso_SaO2_Reset
C177
.1UF CER
0805
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
DWN: MARK TUCCILLO 9/5/01
R295
499 OHM
CHK:
APR:
GP 9/5/01
RAM 9/5/01
Title
SpO2 Power Supply
I
D
Size
C
Date:
Document Number
Rev
24
5143-00-01 (-03 SCHEMATIC)
Friday, May 27, 2005
Sheet
14
of
16
REVISIONS
REV
18
19
20
21
22
23
24
R240
4.99K
+5VI
1
resp_high
100K
680PF NPO
0805
R251
R250
T4
2637-00-10
8
2
DS6
BA-CMS-90
R257
10K
C152
5
*
RA
0 OHM
0805
449008
.0068UF 1KV CD
C147
R255
4.99K
7
3
U70 7
OPA228U
C150
100PF NPO
0805
R260
1K
I
+5VI
*
R262
49.9K
3
Q27
MMBT3904LT1
I
.1UF CER
0805
R258
402 OHM
.1% 25PPM
.1UF CER
0805
200K
7
3.32K
inc#
7
1
3 +
2 -U74
D36
MMBD1503A
I
R263
6
R268
100K
I
R272
4.02K
.01UF MP
2824
R271
R273
2K
49.9K .5%
25PPM
I
R275
1.82K .1%
50PPM
C155
.1UF CER
0805
I
8
3 +
6
2 -U76
OPA228U
4
8
I
5 +
C154
.1UF CER
0805
-5VI
I
.1UF CER
0805
7
6 -
3 +
C160
.1UF CER
0805
1
I
-5VI
6
5
3
4
2
3
1
8
R355
100K
4
-5VI
R261
200K
U72A
LT1112S8
+5VI
10K
I
3
2
C161
C162
.01UF MP
2824
I
1
4
L
INC
U/D
W
CS
H
GND
2M
2.2UF MP
4036
R274
2 -
C163
D37
MMBD1503A
R270
I
.1UF MP
2824
+5VI
R265
200K
U75B
LT1112S8
10M
R269
121K
.5% 25PPM
C157
+5VI
C159
C158
330PF NPO
0805
3
u/d#
R378
R264
49.9K
R267
I
2
R266
49.9K .5%
25PPM
C156
.1UF CER
0805
1
49.9K .5%
25PPM
OPA228U
u/d#
offset_cs#
1
2
7
+5VI
4
R379
R354
100K
8
Vcc
-5VI
6
5
3
+5VI
10K
7
1
-5VI
L
INC
U/D
W
CS
H
GND
U71
DS1804Z-050
base_imp
4
TP10
BASE IMP
Vcc
1
2
7
gain_cs#
C153
.1UF CER
0805
4
8
PATIENT CABLES WITH 1K SERIES
RESISTORS.
U73
8 DS1804Z-100
10K
C228
.01UF CER
0805
U101
TLE2061ACD
6
2
R369
1K
u/d#
+5VI
3
+5VI
sclock0
R391
I
R256
+5VI
R259
1
I
*
SPECIAL FOR ACCOMMODATING
I
+5VI
C149
-5VI
I
10K
442049
I
+ C144
68UF TANT
16V 7343
R254
10K .1%
50PPM
680PF NPO
0805
C151
470K
R249
49.9K
249 OHM
-5VI
C146
2
6
1000PF NPO
0805
3
.0068UF 1KV CD
C148
680PF
3KV CD
I
I
2
LA
.1UF CER
4
0805
R253
49.9K
R248
+
C145
D35
MMBZ5234B
R252
1K
C143
C220
.33UF CER
NPO 0805
C139
.033UF MP
2824
R247
1M
-
-5VI
100PF NPO
0805
I
I
+5V_Ref
40.2K .5%
25PPM
TP43
ZERO_ADJ
C137
I
.1UF CER
0805
C141
6.8UF TANT
20V 3528
+
15PF NPO
0805
shield_dr
R244
2K
+ C140
6.8UF TANT
20V 3528
C142
resp_low
U69B
LT1112S8
R242
470K
C138
-5VI
I
C215
4
R246
1K
7
6 -
I
1
U69A
LT1112S8
2 -
R245
1K
TP42
RESP_OSC
.1UF CER
0805
8
3 +
5 +
470K
2
R243
TP8
RESP LOW
R241
3
C136
I
TP9
RESP HIGH
DATE
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
Try LT1881 as a replacement for the 1112
D34
MMBD4148SE
R239
100K
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
GP
REVISED PER ECN 1274.
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
R276
resp_test
U75A
LT1112S8
.1UF CER
0805
40.2K
150K
R277
15K
C165
I
D38
MMBD4148
C164
33UF TANT
10V 7343
I
I
I
Try LT1468 as a replacement for the 1028
+5V_Ref
+5V_Ref
+5VI
+5VI
-5VI
NOTES:
-5VI
TP15
IGND
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
Iso_GND
I
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
DWN:
MARK TUCCILLO 9/5/01
Title
Size
D
Date:
CHK:
APR:
GP 9/5/01
RAM 9/5/01
Resp
Document Number
Rev
24
5143-00-01 (-03 SCHEMATIC)
Friday, May 27, 2005
Sheet
13
of
16
REVISIONS
R198
Mot
MBRS360T3
1
6
5
18
19
20
21
22
23
24
MBRS340T3
+
5
C110
33UF TANT
35V SIZE-R
Vin
4
Switch
7
MBRS340T3
8
R392
150 OHM
2W 5%
Q17
IRF7205/SO
R199
20K
2
Feedback
12V_Switcher
Q18
MMBT3904LT1
Gnd
.1 OHM 1W
2512
R206
46.4K
C112
33UF TANT
35V SIZE-R
+
L5
27UH SMT
D23
MMBD4148
R204
2K
Battery_Pos
+
R203
261K
2
3
C111
.1UF CER
0805
R200
+
C115
470UF
16V ALUM
Thermistor_Neg
R389
0 OHM
R205
604K
+
C116
470UF
16V ALUM
+
C117
470UF
16V ALUM
C118
470UF
16V ALUM
R211
1K
1
Q19
MMBT3904LT1
R208
100K
1
2
Rec_GND
1
TP41
BAT_GND
JP6
Pump_GND
8
R210
.39 OHM 1W
2512
7
D44
MBRS340T3
6
TP40
BAT (-)
t
t
TP22
REC_GND
Q40
IRF7205/SO
5
2
RT1
33K
Vcc_BC
C113
6.8UF TANT
20V 3528
RT2
3.3K
2
+
Battery_Neg
2
1
VO
1
R207
249 OHM
3
VIN
C114
1UF TANT
16V 3216
+
R387
0 OHM
TP38
Vcc_BC
R209
100K
2
GND
8
Vcc_BC
3
4
Rec_GND
U62
LM2931M-5.0
DATE
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
D22
MBRS340T3
R202
1K
1
R201
1.2K
U61
LM2587S-ADJ
L4
68UH SMT
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
GP
REVISED PER ECN 1274.
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
TP39
VBATT
Vbat
Comp
3
+
4
22UH SMT
C109
180UF TANT
16V SIZE-R
LOW ESR
D21
3
D20
2
L3
REV
3.52K
1
TP49
EOS_OUT
Switcher_GND
Thermistor_Pos
TP2
+12V
+12V
U63
Vcc
R333
120K
R212
20K
D28
CMPD6263
TP51
BC_TMR Vcc_BC
ac
R334
100K
Switcher_off
R222
100K
1
2
3
4
5
6
7
8
R221
20K
R217
121K
C121
1000PF
NPO 0805
C123
.01UF CER
0805
C122
.01UF CER
0805
C124
.1UF CER
0805
R335
T3
Q4435-B
Vcc_BC
16
15
14
13
12
11
10
9
3.
R213
49.9K
R214
10K
R215
10K
D24
11
+12V_HD
TP4
-12V
MBRS340T3
D25
D26
BZT52-C16
16V
MBRS340T3
.10
2
+
D27
C119
180UF TANT
16V SIZE-R
LOW ESR
-12V
MBRS340T3
4
R216
200K
R383
20K
bq2031SN
TMTOLED2/DSEL
FLOAT
LED1/TSEL
Bat
MOD
Vcomp
Vcc
Icomp
Vss
IGsel
COM
SNS LED3/QSEL
TS
TPWM
2
Switch
R218
402 OHM
R219
402 OHM
R220
402 OHM
5
R224
1K
+
Power_On_Led
Vin
Gnd
1
2
1
U64
LM2587S-12
+
3
Bat_Maintence_LED
2K
-12V
L6
100UH
SMT
Output
Feedback
Comp
C126
120UF TANT
20V SIZE-R
U65
LM2575S-12
C125
120UF TANT
20V SIZE-R
Vin
Feedback
4
ON/OFF
5
+
C120
180UF TANT
16V SIZE-R
LOW ESR
GND
3
C127
.22UF NPO
0805
Bat_Charging_LED
Common_LED
+ 3
- 2
1
R385
49.9K
Q20
IRF7205/SO
R375
1
HD_GND
Q41
CMPT3906
3
R377
5
6
7
HD_GND
10K
100K
8
JP7
R223
3.01K
R376
U99A
LM393M
2
R336
1
49.9K
Q34
MMBT3904LT1
4
4
2
3
2
1
3
8
Battery_Fault_LED
100K
Vbuss
8
7
6
5
-12V
1
2
12V_BM
3
Q21
IRF7205/SO
4
D29
MBRS340T3
U66
LM2676S-5
D30
2
MMBD4148
R225
10K
C129
.1UF CER
0805
R332
10K
R227
3
GND ON/OFF
TP20
DGND
4
7
D31
MBRS340T3
Vcc
C131
150UF TANT
10V SIZE-C
JP8
3
8
D
Q
13
Q
12
R228
1
Q23
MMBT3904LT1
10K
2
CLK
R
11
S
2
9
10
14
6
S
VD
7
R
VS
4
3
2
1
2
3
2
3
Q
TP3
VCC
L7
33UH SMT
U67B
CD4013BM
DGND
DGND
R230
100K
R231
10K
+
C133
1UF TANT
16V 3216
10K
Q26
MMBT3904LT1
+
NOTES:
+
OP AMP
-
R337
100K
8
2
7
x1
3
-
4
C135
.1UF CER
0805
CLK
1
1
MMBD4148
R234
1
R237
316K
3
Q
Vsw
D33
1
C134
1UF TANT
16V 3216
C132
.33UF NPO
0805
R232
10K
1
Q25
CMPT3906
D
3
D32
MMBD4148
U67A
CD4013BM
4
5
6
7
8
R229
200K
Q24
IRF7205/SO
R233
100K
5
+
C128
.01UF CER
0805
6
Feedback
Cb
Q22
CMPT3906
1
10K
On_Off
C130
120UF TANT
20V SIZE-R
R226
200K
2
+
C219
6.8UF TANT
20V 3528
Input
+
U68
LTC1541CMS8
6
R338
100K
7
R235
732K
5
R236
100K
+
5
-
6
U99B
LM393M
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
+
C203
1UF TANT
16V 3216
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
DWN:
R238
100K
Title
Size
D
Date:
MARK TUCCILLO 9/5/01
CHK:
APR:
GP 9/5/01
RAM 9/5/01
Power Supply
Document Number
Rev
24
5143-00-01 (-03 SCHEMATIC)
Friday, May 27, 2005
Sheet
12
of
16
REVISIONS
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
REVISED PER ECN 1274.
GP
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
REV
18
19
20
21
22
23
24
DATE
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
VCC
VCC
GND
cs0
cs0
cs1
cs1
lbwe
lbwe
ubwe
ubwe
a1
a2
a3
a4
a5
a6
a7
a8
a9
a10
a11
a12
a13
a14
a15
a16
a17
ubre
ubre
pcs8-8
pcs8-8
nv_wr
nv_wr
d[0..15]
d[0..15]
12
11
10
9
8
7
6
5
27
26
23
25
4
28
29
3
2
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14
A15
A16
16
GND
U55
CY7C1009-20VC
U54
CY7C1009-20VC
U53
lbre
lbre
O0
O1
O2
O3
O4
O5
O6
O7
13
14
15
17
18
19
20
21
d0
d1
d2
d3
d4
d5
d6
d7
CE
OE
22
24
cs0
lbre
VPP
PGM
NC
1
31
30
VCC
a1
a2
a3
a4
a5
a6
a7
a8
a9
a10
a11
a12
a13
a14
a15
a16
a17
12
11
10
9
8
7
6
5
27
26
23
25
4
28
3
31
2
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14
A15
A16
16
GND
VCC 32
D0
D1
D2
D3
D4
D5
D6
D7
13
14
15
17
18
19
20
21
d0
d1
d2
d3
d4
d5
d6
d7
CS1
CS2
22
30
cs1
WE
OE
29
24
lbwe
lbre
NC
1
VCC
32
VCC
VCC
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
pcs8-8
lbre
nv_wr
20
22
27
26
E1*
G*
W*
E2
11
12
13
15
16
17
18
19
d8
d9
d10
d11
d12
d13
d14
d15
GND
U58
CY7C1009-20VC
U57
a1
a2
a3
a4
a5
a6
a7
a8
a9
a10
a11
a12
a13
a14
a15
a16
a17
12
11
10
9
8
7
6
5
27
26
23
25
4
28
29
3
2
16
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14
A15
A16
GND
O0
O1
O2
O3
O4
O5
O6
O7
13
14
15
17
18
19
20
21
d8
d9
d10
d11
d12
d13
d14
d15
CE
OE
22
24
cs0
ubre
VPP
PGM
NC
1
31
30
VCC
VCC
a1
a2
a3
a4
a5
a6
a7
a8
a9
a10
a11
a12
a13
a14
a15
a16
a17
32
12
11
10
9
8
7
6
5
27
26
23
25
4
28
3
31
2
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14
A15
A16
13
14
15
17
18
19
20
21
d0
d1
d2
d3
d4
d5
d6
d7
CS1
CS2
22
30
cs1
WE
OE
29
24
lbwe
lbre
NC
1
VCC
32
VCC
16
GND
U59
CY7C1009-20VC
d8
D0
D1
D2
D3
D4
D5
D6
D7
13
14
15
17
18
19
20
21
CS1
CS2
22
30
cs1
WE
OE
29
24
ubwe
ubre
NC
1
VCC
32
d9
d10
d11
d12
d13
d14
d15
a1
a2
a3
a4
a5
a6
a7
a8
a9
a10
a11
a12
a13
a14
a15
a16
a17
VCC
SOCKET
32 PIN
12
11
10
9
8
7
6
5
27
26
23
25
4
28
3
31
2
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14
A15
A16
16
GND
D0
D1
D2
D3
D4
D5
D6
D7
13
14
15
17
18
19
20
21
d8
d9
d10
d11
d12
d13
d14
d15
CS1
CS2
22
30
cs1
WE
OE
29
24
ubwe
ubre
NC
1
VCC
32
VCC
a18
a[0..20]
a18
Vcc
U60
SN74AHC1G04DCK
5
VCC
A
Y
GND
2
3
a[0..20]
16
D0
D1
D2
D3
D4
D5
D6
D7
28
DQ0
DQ1
DQ2
DQ3
DQ4
DQ5
DQ6
DQ7
14
BATTERY
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14
A15
A16
Vcc
FT
10
9
8
7
6
5
4
3
25
24
21
23
2
Vss
1
a1
a2
a3
a4
a5
a6
a7
a8
a9
a10
a11
a12
a13
12
11
10
9
8
7
6
5
27
26
23
25
4
28
3
31
2
a18
SOCKET
32 PIN
U56
M48T58SOH28
a1
a2
a3
a4
a5
a6
a7
a8
a9
a10
a11
a12
a13
a14
a15
a16
a17
4
NOTES:
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
DWN:
Size
C
Date:
CHK:
MARK TUCCILLO 9/5/01
Title
APR:
GP 9/5/01
RAM 9/5/01
CPU Memory
Document Number
Rev
24
5143-00-01 (-03 SCHEMATIC)
Friday, May 27, 2005
Sheet
11
of
16
REVISIONS
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
REVISED PER ECN 1274.
GP
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
REV
pcs2_8
pcs1_8
spare_in_1
a17
a16
pcs8_8
pcs5_8
pcs4_8
a19
ac
DATE
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
fp_1
VCC
fp_2
a18
a0
18
19
20
21
22
23
24
spare_in_3
host_a18
ac
host_a0
GND
fp_2
host_a19
fp_1
VccINT
host_a17
host_a16
~PIN001
GND
GND
pcs8_8
pcs5_8
pcs4_8
VCCIO
pcs2_8
pcs1_8
spare_in_1
11
10
9
8
7
6
5
4
3
2
1
84
83
82
81
80
79
78
77
76
75
spare_in_3
a9
spare_out_1
spare_out_0
25_MHz
r/w
siz0
3_6864MHz
kojak_light
64_KHz
a14
cs2
ac
ac
fp_0
fp_1
fp_2
fp_3
fp_0
fp_1
fp_2
fp_3
3_6864MHz
25_MHz
as
VCCIO
host_a9
spare_out_1
spare_out_0
16_MHz
r/w
GJND
siz0
3_6864MHz
kojak_light
64KHz
host_a14
cs2
VCCIO
spare_in_2
1_843MHz
host_a13
fp_0
cs3
GND
U52
2635-00-04
LC Color Glue
~PIN002
pcs0-8
GND
lbwe
lbre
clock_wr
alarm_out
spare_out_4
VCCIO
ubwe
ubre
host_d4
sao2_res
graph_out
rec_res
GND
etco2_res
leads_off
nibp_res
dac_wr
ecg_gain
74
73
72
71
70
69
68
67
66
65
64
63
62
61
60
59
58
57
56
55
54
pcs0_8
lbwe
lbre
nv_wr
alarm_out
spare_out_4
ubwe
VCC
ubre
d12
sao2_res
graph_out
rec_res
etco2_res
lead_off
nibp_res
ecg_gain
pcs0_8
pcs1_8
pcs2_8
pcs3_8
pcs5_8
pcs8_8
VCC
d11
d10
d9
d8
pcs3_8
d14
d15
d13
spare_in_0
spare_in_1
spare_in_2
spare_in_3
a10
spare_out_2
a15
cs2
cs3
cs2
cs3
spare_in_0
spare_in_1
spare_in_2
spare_in_3
fp_3
a12
a11
r/w
siz0
as
r/w
siz0
as
a20
spare_out_3
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
3_6864MHz
25_MHz
spare_in_2
1_8MHz
a13
fp_0
cs3
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
host_a20
spare_out_3
fp_3
host_a12
host_a11
VCCIO
host_a10
spare_out_2
host_a15
GND
VCCINT
host_d6
host_d7
host_d5
GND
host_d3
host_d2
host_d1
host_d0
pcs3_8
VCCIO
as
pccs0_8
pccs1_8
pccs2_8
pccs3_8
pccs5_8
pccs8_8
ecg_gain
ecg_gain
lbwe
ubwe
lbre
ubre
lbwe
ubwe
lbre
ubre
nv_wr
nv_wr
1_8MHz
1_8MHz
kojak_light
kojak_light
a[0..20]
a[0..20]
rec_res
sao2_res
nibp_res
etco2_res
rec_res
sao2_res
nibp_res
etco2_res
d[0..15]
d[0..15]
graph_out
alarm_out
lead_off
graph_out
alarm_out
lead_off
spare_out_0
spare_out_1
spare_out_2
spare_out_3
spare_out_4
VCC
VCC
spare_out_0
spare_out_1
spare_out_2
spare_out_3
spare_out_4
64_KHz
GND
NOTES:
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
64_KHz
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
DWN:
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
Title
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
Size
C
Date:
MARK TUCCILLO 9/5/01
CHK:
APR:
GP 9/5/01
RAM 9/5/01
GLUE
Document Number
Rev
24
5143-00-01 (-03 SCHEMATIC)
Friday, May 27, 2005
Sheet
10
of
16
REVISIONS
REV
Grounded
18
19
20
21
22
23
24
Isolated
F4
3/8ASB
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
REVISED PER ECN 1274.
GP
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
DATE
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
TP36
+5VI_E
+12V
C103
180UF TANT
16V SIZE-R
+
D17
BZT52-C16
16V
R188
4.02K
D18
ether_iso_+5V
MBRS340T3
5
1
D19
MBRS340T3
5
Switch
1
4
4
ether_IGND
8
I
T2
2532-00-10
Comp
Feedback
D
TP37
E_GND
+
7
Vin
DGND
C104
180UF TANT
16V SIZE-R
LOW ESR
2
Gnd
U47
3
LM2587S-ADJ
D
U48
LM3411M5-5.0
U12-14
+5VREF
R189
5
1K
C105
.1UF CER
0805
2
Q37
CMPT3906
4
Out
Comp
IN
1
GND
2
NC
3
U49
CNY17-3SM
4
R352
10K
1
3
1
C106
.1UF CER
0805
5
2
R351
49.9K
D
I
VCC
VCC
2
R190
10K
U50
HCNW2601-300
VCC
2
R191
Q15
CMPT3906
3
8
7
R192
2K
1
ether_iso_txd
1K
6
ether_txd
C107
.1UF CER
0805
3
5
GND
R193
499 OHM
D
I
VCC
2
R194
10K
1K
U51
HCNW2601-300
1
Q16
CMPT3906
VCC
3
ether_rxd
R195
2
8
7
R196
1K
6
3
GND
5
ether_iso_rxd
C108
.1UF CER
0805
R197
499 OHM
NOTES:
D
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
I
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
OPTIONAL
THIS CIRCUITRY IS FOR THE ETHERNET OPTION
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
DWN:
Title
Size
C
Date:
MARK TUCCILLO 9/5/01
CHK:
APR:
GP 9/5/01
RAM 9/5/01
Ethernet Power Supply
Document Number
Rev
24
5143-00-01 (-03 SCHEMATIC)
Friday, May 27, 2005
Sheet
9
of
16
REVISIONS
REV
+6.5V
TP6
SHIELD_DR
U36A
TL062ACD
R126
shield_dr
3
-6.5V
D11
MMBD1503A
I
U42C
LT1058SW
R140
10K .1%
50PPM
leads0
leads1
leads2
R141
10K .1%
50PPM
1
R342
22M
5%
10
D3
11
S3
R159
22M
5%
I
R160
10K .1%
50PPM
RA_ref
R163
10K .1%
50PPM
leads0
leads1
leads2
3
C101
220PF NPO
0805
13
D4
S4
V
5K MELF 1W
200K
AVL = LA - 0.5(LL + RA)
ON
AVF = LL - 0.5(LA + RA)
ON
ON
V = V - 0.33(LA + RA + LL)
INH
A
B
C
U42A
LT1058SW
418
U40
AD621AR
100K
2.2UF MP
4036
R151
178K
R150
1.62M
C95
1000PF NPO
0805
+5VI
R166
24.9K
R144
9.31K
.1% 50PPM
U41A
LT1057S8
1 +
8 -
7
-6.5V
I
C93
.1UF MP
2824
R147
R148
9.31K
.1%
50PPM
2.49K
C98
C96
.22UF MP
2824
R168
10K
7
D4
6
S4
In4
8
TP7
ECGX1000
5
ecgX1000
U41B
LT1057S8
I
I
R157
+5VI
R156
32.4K
R374
20 OHM
+6.5V
+5VI
R158
10K
5 +
2
U45A
4 LM339AM
C99
I
.01UF MP
2824
R164
21.5K
R165
11K
-6.5V
+5VI
diag_bw\
I
R176
10K
7 +
U45B
6 -
5
R178
24.9K
3
680PF NPO
0805
165K
.1% 50PPM
I
9 +
14
U45C
8 LM339AM
11 +
13
U45D
10 LM339AM
4
I
R155
1.65K
.1%
50PPM
+5VI
I
LL_Lead
R145
6
1
R177
2
10K
A Vcc
B
D15
MMBD4148
Y
GND
4
R171
32.4K
R169
1
pacer_det
10K
LM339AM
U46
SN74AHC1G08DCK
3
-5VI
I
MAX327CSE
U37D
I
-5VI
2
U44C
MAX327CSE
U44B
MAX327CSE
1
1
3
-5VI
5
6
C102
220PF NPO
0805
DS4
BA-CMS-90
ON
ON
ON
15
D2
14
S2
In2
16
10
D3
11
S3
In3
9
3
5K MELF 1W
ON
ON
ON
LL_ref
R181
22M
5%
R184
6
11
10
9
D16
MMBD1503A
2
*R183
ON
ON
V
+5VI
R180
22M
5%
*
R182
ON
AVF
U44D
MAX327CSE
R172
200K
I
R179
22M
5%
AVL
-6.5V
-6.5V
8
In4
AVR = RA - 0.5(LA + LL)
-6.5V
.1UF MP
2824
1
2
6
III = LL - LA
ON
3
X
I
4
3
R175
7
OFF
ON
U43
CD4051BCM
7
3
+6.5V
-5VI
R343
22M
5%
ON
ON
C94
1
C100
1
1
I
leads0
leads1
leads2
X0
X1
X2
X3
X4
X5
X6
X7
-5VI
2
R170
22M
5%
200K
R162
10K .1%
50PPM
D14
MMBD1503A
2
DS3
BA-CMS-90
R161
10K .1%
50PPM
13
14
15
12
1
5
2
4
R149
100K
4
16
U37C
MAX327CSE
+5VI
R167
22M
5%
OFF
ON
I
5
2 -
U39A
LT1112S8
3 +
2 -
IVCC
9
In3
ON
AVR
+6.5V
7
3 +
8
16
15
II = LL - RA
III
ECGX10
VDD
14
LEAD MIX
+6.5V
R143
100K
C92
.01UF MP
2824
U42D
LT1058SW
C97
220PF NPO
0805
I = LA - RA
OFF
7
I
RA_Lead
I
INH
A
B
C
6 -
+6.5V
-5VI
OFF
ON
16
R139
10K .1%
50PPM
-5VI
1
V
OFF
ON
U39B
LT1112S8
5 +
VSS
200K
R138
10K .1%
50PPM
D13
MMBD1503A
2
LL
ON
OFF
3
X
U38
CD4051BCM
VEE
5K MELF 1W
R137
10K .1%
50PPM
X0
X1
X2
X3
X4
X5
X6
X7
8
5K MELF 1W
LA_ref
6
11
10
9
3
R154
13
14
15
12
1
5
2
4
+5VI
2
RA
ON
II
3
I
LA
I
6
R136
22M
5%
MAX327CSE
U37B
VDD
S2
VSS
14
16
In2
VEE
D2
8
15
CONF
IVCC
7
R341
22M
5%
3
*R153
5K MELF 1W
1
1
LA_Lead
*
R152
5K MELF 1W
2
10
DS2
BA-CMS-90
LL
4
3
2
1
C91
220PF NPO
0805
R146
22M
5%
*R174
-6.5V
TP25
IGND
+6.5V
-5VI
11
R142
22M
5%
*
R173
I
I
R131
953 OHM
.1% 50PPM
I
RA
6
12
1
1
12
DS1
BA-CMS-90
5
-
2
2
3
R135
200K
+
3
3
2
3
*R134
5K MELF 1W
R128
4.75M
Q14
DTC114
2
1
D12
MMBD1503A
R132
22M
5%
*
R133
100K
3
2
5K MELF 1W
7
+
LA
Q13
CMPT3906
1
R353
R130
22M
5%
2
DATE
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
2
R127
100K
R129
22M
5%
CHANGE THE "5K MELF 1W" RESISTORS
TO "499 OHM MELF 1W" No. 449012
10 PLACES.
3
-
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
GP
REVISED PER ECN 1274.
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
+5VI
R340
22M
5%
PATIENT CABLES WITH 1K SERIES
RESISTORS.
+
18
19
20
21
22
23
24
-
*
SPECIAL FOR ACCOMMODATING
1
1K
C90
.1UF CER
0805
+5VI
8
ecg_test
U36B
TL062ACD
7
U42B
LT1058SW
+6.5V
I
I
V_Lead
13
I
Vcc
2
D1
3
S1
MAX327CSE
Vss Gnd U37A
I
In1
4
1
5
V_ref
I
-6.5V
+6.5V
+6.5VI
+6.5V
+5VI
IVCC
+5VI
RL
NOTES:
13
*
R185
*R186
R187
5K MELF 1W
5K MELF 1W
200K
DS5
BA-CMS-90
Vcc
I
D1
3
S1
MAX327CSE
Vss Gnd U44A
In1
4
1
-5VI
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
-6.5VI
5
RL_ref
I
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
-5VI
2
-6.5V
IVY BIOMEDICAL SYSTEMS INC.
-6.5V
I
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
Iso_GND
I
DWN:
Title
Size
D
Date:
MARK TUCCILLO 9/5/01
CHK:
APR:
GP 9/5/01
RAM 9/5/01
ECG Front End
Document Number
Rev
24
5143-00-01 (-03 SCHEMATIC)
Friday, May 27, 2005
Sheet
8
of
16
REVISIONS
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
REVISED PER ECN 1274.
GP
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
REV
F3
500mASB
Grounded
Isolated
+12V
R372
10K
Q36
CMPT3906
C76
.1UF CER
0805
C77
.1UF CER
0805
C78
.1UF CER
0805
C79
.1UF CER
0805
1
R349
10K
3
Q38
CMPT3906
3
1
C75
.1UF CER
0805
DATE
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
R373
806 OHM
2
2
R371
2 OHM 1W
2512 5%
C74
.1UF CER
0805
18
19
20
21
22
23
24
R366
-6.5VI
3
0 OHM
2512
U31
LM2990S-5.0
2
VIN
TP31
+6.5VI
D7
BZT52-C16
16V
4
Feedback
2
1
TP34
IGND
TP23
IGND
7
4
IGND
8
Comp
D10
Gnd
+
C83
100UF TANT
16V 7343
TP32
-6.5VI
I
U33
LM2931M-5.0
+
8
MBRS340T3
VIN
VOUT
2
3
6
7
3
R367
0 OHM
2512
C214
470UF
16V ALUM
1
+5VI
+ C84
6.8UF TANT
35V 7343
TP24
+5VI
I
R121
3.01K
+6.5VI
3
R120
3.01K
1
Q10
FMMT617TA
2
Q11
CMPT3906
TP35
ISO_+5_VREF
3
1
R122
10K
R368
+5_Vref
0 OHM
0805
R123
10K
I
U34
R124
1K
C88
.1UF CER
0805
2
I
5
U32
LM2587S-ADJ
C80 +
6.8UF TANT
35V 7343
TP17
IGND
+
U35
CNY17-3SM
5
Out
4
Comp
C85
.33UF NPO
0805
PS_Synch
IN
1
GND
2
NC
3
3
Switch
C82
100UF TANT
16V 7343
2
1
5
C86
Q12
MMBT3904LT1
1
+
LM3411M5-5.0
C87
6.8UF TANT
20V 3528
47PF NPO
0805
R125
20K
C89
470PF NPO
0805
2
Vin
DGND
1
-5VI
NC
NC
D9
MBRS340T3
MBRS340T3
3
4
5
5
D8
T1
2532-00-10
GND
GND
GND
GND
R119
4.02K
TP33
-5VI
1
+
4
C81
180UF TANT
16V SIZE-R
LOW ESR
VOUT
GND
1
Q39
MMBT3904LT1
1K
2
R350
I
I
NOTES:
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
DWN:
Title
Size
C
Date:
MARK TUCCILLO 9/5/01
CHK:
APR:
GP 9/5/01
RAM 9/5/01
DSP Power Supply
Document Number
Rev
24
5143-00-01 (-03 SCHEMATIC)
Friday, May 27, 2005
Sheet
7
of
16
REVISIONS
REV
ecgX1000
press_1
press_1
press_2
press_2
10K R96
A4
A5
A6
A7
A11
DMS
16MHZ
3
15
23
35
VCC
VCC
VCC
VCC
WR-
10K R98
10K R99
IRQE
10K R100
13
21
20
ls1
ls2
MUX0
MUX1
MUX2
19
18
8
mux0
mux1
mux2
LA_REF
RA_REF
LL_REF
V_REF
RL_REF
25
33
7
11
34
la_ref
ra_ref
ll_ref
v_ref
rl_ref
LEADS_OFF0
LEADS_OFF1
29
16
sel_20
press_test
10
22
30
42
IRQL1
10K R102
IRQ1
GND
GND
GND
GND
IDG
10K R103
DIAG_BW
14
diag_bw
24
PRESS_EN
IRQ1
ECG_TEST
27
ecg_test
TEMP_CAL_SEL
HI_LO_SEL
41
26
temp_cal
sel_40
PCS6
PCS7
PCS8
PCS9
28
9
31
32
pcs6
pcs7
pcs8
pcs9
4MHZ
12
4MHz
RFS0
10K R105
3
15
16
17
18
19
20
21
22
23
24
25
26
27
d11
d12
d13
d14
d15
d16
d17
d18
d19
d20
d21
d22
d23
PRESS_EN
IRQ0
10K R104
ls0
WRRDpcs6
4MHz
IRQL0
10K R101
IVCC
7
11
10
8
12
13
DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7
DB8
DB9
DB10
DB11
DB12
VIN
1
VREF
2
1
AVCC
4
V-
5
+
3
-
2
R91
4.99K
U21A
TLE2062ACD 4
IVCC
-6.5VI
+
-5VI
C225
10UF TANT
10V 3528
I
DVCC
AZ
CAL
28
6
9
ad_az
ad_cal
+
C226
10UF TANT
10V 3528
U23
ADG426BRS
28
D
2
18
3
NC
EN
RS
pcs9
13
WR
D11
D10
D9
D8
14
15
16
17
A3
A2
A1
A0
I
19
20
21
22
23
24
25
26
11
10
9
8
7
6
5
4
ecgX1000
resp_high
base_imp
ecgX10
la_lead
ra_lead
ll_lead
IVCC
I
-6.5VI
rl_ref
press_en
press_test
press_test
v_ref
v_ref
ls0
ls0
diag_bw
ecg_test
sel_40
temp_cal
temp_cal
sclk0
sclk0
sel_20
sel_20
resp_test
C227
resp_test
ls0
ls0
ls1
14
4
5
6
7
8
9
10
11
DB7
DB6
DB5
DB4
DB3
DB2
DB1
DB0
VDD
WR
649K
Isolated
resp_low
3
SOCKET
32 PIN
IDG
O0
O1
O2
O3
O4
O5
O6
O7
13
14
15
17
18
19
20
21
CE
OE
22 BMS24 RD-
IVCC
d8
d9
d10
d11
d12
d13
d14
d15
16
IVCC
R110
402K
1
31
30
VPP
PGM
NC
I
C59
1UF MP
2824
CEXT
7
REXT/CEXT
9
10
11
ecg_test
VCC
VCC
A
B
CLR
Q
5
Q
12
resp_test
IVCC
R112
1
U26
2636-00-04
IDG
8
7
R113
2K
VCC
2
C60
.1UF CER
0805
IDG
Q8
CMPT3906
dsp_txd
dsp_txd
1K
U27
HCNW2601-300
6
3
5
GND
R114
499 OHM
IRQ1
IRQ0
DR1
sclk0
TFS0
RFS0
IRQL1
IRQ2
IRQE
IRQL0
R111
10K
2
IDG
Grounded
U19A
74HCT221D
6
3
R339
10K
15
Vcc
IVCC
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14
A15
A16
2
VREF
GND
d8
12
11
10
9
8
7
6
5
27
26
23
25
4
28
29
3
2
OUT2
ls2
resp_high
8
a0
a1
a2
a3
a4
a5
a6
a7
a8
a9
a10
a11
a12
a13
d16
d17
d18
d11
d10
d9
VCC
32
d15
d14
d13
d12
1
ls1
R109
U24
AD7524JR
75
74
73
72
71
70
69
68
67
66
65
64
63
62
61
60
59
58
57
56
55
54
53
52
51
OUT1
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
IDG
IVCC
GND
100 OHM
U25
ADSP-2184BST-160
13
16
3
2
OUT
16MHz
D15
D14
D13
D12
GND
D11
D10
D9
Vdd
GND
D8
D6/IWR
D6/IRD
D5/IAL
D4/IS
GND
Vdd
D3/IACK
D2/AD15
D1/AD14
D0/AD13
BG
EBG
BR
EBR
WR-
16
RFB
GND
a3
100
a2
99
a1
98
a0
97
96
95
94 PF0
93 PF1
92
91
90
89 PF2
88
87
86
85
d23
84
d22
83
d21
82
d20
81
80
d19
79
d18
78
d17
77
d16
76
Vcc
4
Y3
EMC60S5B-016
A4/AD3
A5/AD4
GND
A6/AD5
A7/AD7
A8/ID7
A9/ID8
A10/AD9
A11/AD10
A12/AD11
A13/AD12
GND
CLKIN
XTAL
Vdd
CLKOUT
GND
Vdd
WR
RD
BMS
DMS
PMS
IOMS
CMS
A3/AD2
A2/AD1
A1/AD0
A0
PWDACK
BGH
PF0[MODE A]
PF1[MODE B]
GND
PWD
Vdd
PF2[MODE C]
PF3
FL0
FL1
FL2
D23
D22
D21
D20
GND
D19
D18
D17
D16
IVCC
IRQE+PF4
IRQL0+PF5
GND
IRQL1+PF6
IRQ2+PF7
DT0
TFS0
RFS0
DR0
SCLK0
Vdd
DT1
TFS1
RFS1
DR1
GND
SCLK1
ERESET
RESET
EMS
EE
ECLK
ELOUT
ELIN
EINT
1
2
3
a6
4
a7
5
a8
6
a9
7
a10
8
a11
9
a12
10
a13
11
12
13
14
15
16
17
18
WR- 19
RD- 20
BMS- 21
DMS- 22
PMS- 23
IOMS-24
CMS- 25
CS
d15
d14
d13
d12
d11
d10
d9
d8
IDG
a4
a5
12
ls2
+
IVCC
pcs8
-
d[23..8]
R364
10K
R363
100PF NPO
0805
U21B
TLE2062ACD
6
7
5
DR1
ll_ref
sel_40
a[13..0]
4
ra_ref
diag_bw
R108
100K
13
Q
la_ref
ll_ref
ecg_test
R107
100K
Q
ls1
ra_ref
IRQ2
R106
100K
A
B
CLR
press_en
la_ref
IVCC
REXT/CEXT
rl_ref
ls1
I
PF0
PF1
PF2
CEXT
v_lead
temp_1
temp_2
press_1
press_2
TFS0
10K R344
1
2
3
pacer_det
S1
S2
S3
S4
S5
S6
S7
S8
S9
S10
S11
S12
S13
S14
S15
S16
IDG
IDG
14
C58
.1UF MP
2824
15
R90
100K
8
IVCC
WR
RD
CS
CLKIN
EOC
INT
U19B
74HCT221D
+6.5VI
TP5
ANALOG IN
U22
ADC1241
LEADS0
LEADS1
LEADS2
IVCC
RD-
ad_az
ad_cal
1
38
39
40
37
1
43
5
IOMS16MHz
BMS-
10K R97
a4
a5
a6
a7
a11
36
17
Vdd
temp_2
temp_2
DMS-
D0
D1
D2
D3
Vss
PMS-
10K R95
2
44
6
4
AGND
temp_1
temp_1
10K R94
DSP_GLUE
AD_AZ
AD_CAL
d16
d17
d18
d19
DGND
pacer_det
pacer_det
-5VI
3
resp_low
IOMS-
1
1
I
14
10K R93
resp_low
2
+6.5VI
C57
.1UF CER
0805
U20
2529-00-04
v_lead
v_lead
CMS-
2
1K
ll_lead
ll_lead
10K R92
+5V_Ref
D6
MMBD1503A
IVCC
R386
ra_lead
ra_lead
C224
10UF TANT
10V 3528
I
la_lead
la_lead
IVCC
+
ecgX10
GND
base_imp
ecgX10
27
resp_high
12
resp_high
DATE
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
3
ecgX1000
base_imp
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
GP
REVISED PER ECN 1274.
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
18
19
20
21
22
23
24
I
IVCC
DGND
VB
1
5
RS
7
TP30
RESET\_DSP
R345
10K
key
RESET
1
3
5
7
9
11
13
2
4
6
8
10
12
14
IDG
IRQ2
pcs7
d15
d14
d13
d12
d11
3
8
VO
PFO
D
IDG
d10
d9
d8
WR-
pcs6
GND
INTRN
CEN
D7
D6
D5
D4
D3
NC
D2
D1
NC
D0
WRN
VCC
RDN
RXD
TXD
MP0
MP1
NC
NC
A2
A1
A0
X1/CLK
X2
RESET
IVCC
IVCC
U29
SCC2691AC1A28
15
16
17
18
19
20
21
22
23
24
25
26
27
28
1
2
3
4
5
6
7
8
9
10
11
12
13
14
RD-
R115
10K
2
WDI
J17
EZ-ICE
IVCC
R116
a2
a1
a0
2
U30
HCNW2601-300
JP5
C62
.1UF CER
0805
-6.5VI
C63
.1UF CER
0805
C64
.1UF CER
0805
C65
.1UF CER
0805
C66
.1UF CER
0805
C67
.1UF CER
0805
C205
.1UF CER
0805
C206
.1UF CER
0805
C207
.1UF CER
0805
IVCC
IVCC
2
4
A
Y
+5V_Ref
IVCC
C70
.1UF CER
0805
C71
.1UF CER
0805
C72
.1UF CER
0805
C73
.1UF CER
0805
IDG
5
dsp_rxd
R118
499 OHM
D
I
OUT
3
Y4
1.8432MHZ SMT
2
C69
.1UF CER
0805
3
C68
.1UF CER
0805
I
R117
2K
C61
.1UF CER
0805
dsp_rxd
NOTES:
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
4
U100
SN74AHC1G04DCK
IGND
+5V_Ref
R365
100 OHM
Vcc
+6.5VI
+5VI
5
IDG
GND
-6.5VI
IRQE
-5VI
Vcc
+6.5VI
GND
GND
-5Vi
8
7
6
3
TP16
IDG
JP4
VCC
Q9
CMPT3906
VCC
RESET
PS_Synch
1
1K
3
PFI
6
VCC
4
GND
2
IDG
U28
MAX690A-CSA
IDG
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
DWN:
Title
+5VI
-5Vi
Size
D
Date:
MARK TUCCILLO 9/5/01
CHK:
APR:
GP 9/5/01
RAM 9/5/01
DSP Processor
Document Number
Rev
24
5143-00-01 (-03 SCHEMATIC)
Friday, May 27, 2005
Sheet
6
of
16
REVISIONS
REV
VCC
VCC
ac
rec_res
sao2_res
nibp_res
etco2_res
spare_out_0
spare_out_1
spare_out_2
spare_out_3
spare_out_4
graph_out
alarm_out
lead_off
rec_res
spo2_res
nibp_res
etco2_res
graph_out
alarm_out
lead_off
graph_out
alarm_out
lead_off
8
d[0..15]
Y2
EMC60S5B-048
D
clock
cs2
ds
siz0
siz1
as
r/w
nv_wr
clear
blue[0..5]
clock
hsynch
vsynch
dot_clock
cs2
ds
siz0
siz1
as
r/w
D
VCC
144
143
142
141
140
139
138
137
136
135
134
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
irq3
D
reset
GND
VCC
DSACK1
DSACK0
A0
A31/IACK7
A30/IACK6
A29/IACK5
A28/IACK4
GND
VCC
A27/IACK3
A26/IACK2
A25/IACK1
A24
D15
D14
D13
GND
D12
GND
VCC
D11
D10
D9
D8
D7
D6
D5
D4
GND
VCC
D3
D2
D1
D0
RMC
R/W
SIZ1
SIZ0
DS
AS
BGACK
BG
BR
BERR
HALT
RESET
GND
CLKOUT
VCC
XFC
VCC
EXTAL
VCCSYN
XTAL
GND
MODCK
VCC
IPIPE
IFETCH
BKPT
FREEZE
TIN1
TOUT1
TGATE1
TCK
TMS
TDI
TDO
VCC
GND
CS0/AVEC
CS1/IRQ2
CS2/IRQ2
IRQ3
CS3/IRQ4
GND
VCC
IRQ5
IRQ6
IRQ7
DONE2
DACK2
DREQ2
DONE1
DACK1
DREQ1
X1
GND
VCC
X2
SCLK
CTSB
RTSB
TXDB
RXDB
TXRDYA
RXRDYA
CTSA
RTSA
GND
VCC
TXDA
RXDA
TIN2
TOUT2
TGATE2
U17
MC68340FE25
POWERED BY MOTOROLA
iack3
108
107
106
105
104
103
102
101
100
99
98
97
96
95
94
93
92
91
90
89
88
87
86
85
84
83
82
81
80
79
78
77
76
75
74
73
cs0
cs1
cs2
irq3
cs3
pcs4_8
irq6
ubre
ubwe
irq5
irq6
irq7
done2
dreq2
done1
dreq1
1_843MHz
1
3
5
7
9
VB
PFO
5
RS
7
Q7
CMPT3906
irq6
3
2
WDI
8
GND VCC
6
1
D
irq7
R84
1K
done1
3
pcs5_8
done2
D
reset
cts_b
rts_b
rxd_b
txd_b
pcs4_8
txd_a
rts_a
rxd_a
cts_a
iack3
irq6
txd_e
rts_e
rxd_e
cts_e
ubre
ubwe
audio_a
audio_b
sinen
souten
d[0..15]
rec_txd
rec_rxd
cts_d
rts_d
txd_d
rxd_d
cts_d
rts_d
txd_d
rxd_d
cts_c
rts_c
txd_c
rxd_c
cts_c
rts_c
txd_c
rxd_c
cts_b
rts_b
rxd_b
txd_b
cts_b
rts_b
rxd_b
txd_b
txd_a
rts_a
rxd_a
cts_a
txd_a
rta_a
rxd_a
cts_a
txd_e
rts_e
rxd_e
cts_e
txd_e
rta_e
rxd_e
cts_e
audio_a
audio_b
filter_clka
filter_clkb
a[0..20]
ecg_gain
ECG
RESP
IBP1
IBP2
PLETH
HR
SpO2
ecg_aout
resp_aout
ibp1_aout
ibp2_aout
pleth_aout
hr_aout
spo2_aout
pcs1-8
ECGX1000
pcs2-8
pcs5-8
speaker+
cs_hires_dac speaker-
ecgX1000
ecg_gain
audio_a
audio_b
D
lbwe
ds
lbwe
ds
Battery
speaker+
speakeretco2_aout
ETCO2
a3
a2
a1
a8
irq3
ether_rxd
ether_txd
3MHz
analog
Battery
a7
a6
a5
a4
a10
a9
a13
a12
a11
a16
a15
a14
a18
a17
a21
a20
2
irq5
1
cts_c
rts_c
txd_c
rxd_c
pcs0-8
ether_rxd
ether_txd
a0
+5VREF
+5VREF
Battery
Battery
analog
VCC
VO
r/w
VCC
irq3
PFI
cts_d
rts_d
txd_d
rxd_d
a[0..21]
TP29
RESET\
4
dsack1
VCC +12V
dsp_txd
dsp_rxd
2
4
6
8
10
U18
MAX690A-CSA
a[0..20]
-12
ds
a19
D
J16
TEST
cs_hires_dac
D
pcs1_8
pcs2_8
pcs5_8
cs_hires_dac
berr
3MHz
VCC
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
VCC
irq5
comm
FC3
FC2
FC1
FC0
VCC
GND
A23
A22
A21
A20
A19
A18
A17
VCC
GND
A16
A15
A14
GND
A13
A12
A11
VCC
GND
A10
A9
A8
A7
A6
A5
A4
VCC
GND
A3
A2
A1
25_MHz
ether_cts
ether_rxd
ether_txd
ether_rts
GND
d3
d2
d1
d0
d11
d10
d9
d8
d7
d6
d5
d4
d12
porta1
porta0
d15
d14
d13
iack3
porta2
iack5
porta4
dsack1
dsack0
a0
porta7
r/w
VCC
pcs0_8
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
irq7
D
d[0..15]
d[0..15]
porta0
porta1
porta2
porta4
porta7
hsynch
vsynch
dot_clock
comm
GND
pcs3_8
blue[0..5]
hsynch
vsynch
dot_clock
video
d[0..15]
D
br
berr
halt
reset
dsack1
red[0..5]
blue[0..5]
audio_a_clk
audio_b_clk
glue
r/w
siz1
siz0
ds
as
bgack
eow_int
A[0..21] cs_hires_dac
hd[0..15]
VCC
intack7
porta0
porta1
porta2
porta4
porta7
sof_int
green[0..5]
red[0..5]
25_MHz
25_MHz
memory
a[0..20]
spare_in_1
spare_in_2
spare_in_3
0 OHM
0805
VCC
VCC
kojak_light
R362
5
green[0..5]
+12
fp3
64KHz
rec_res
sao2_res
nibp_res
etco2_res
OUT
red[0..5]
DGND
fp2
64KHz
1_843MHz
64_KHz
1_8MHz
kojak_light
fp_0
fp_1
fp_2
fp_3
fp1
a[0..20]
green[0..5]
intack7
reset
VCC
fp0
lbwe
ubwe
lbre
ubre
intack5
intack7
AGND
ac
fp0
fp1
fp2
fp3
iack5
GND
cs2
cs3
DATE
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
video
VCC
+5VDC
3_6864MHz
cs2
cs3
ecg_gain
lbwe
ubwe
lbre
ubre
ecg_gain
lbwe
ubwe
lbre
ubre
nv_wr
VCC
cs0
cs1
pcs8-8
GND
3MHz
cs0
cs1
4
25_MHz
r/w
siz0
memory
pcs0_8
pcs1_8
pcs2_8
pcs4_8
pcs5_8
pcs8_8
pcs0_8
pcs1_8
pcs2_8
pcs4_8
pcs5_8
pcs8_8
GND
as
25_MHz
r/w
siz0
ac
VCC
glue
as
18
19
20
21
22
23
24
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
GP
REVISED PER ECN 1274.
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
R72
10K
R75
10K
R78
10K
R81
10K
R85
10K
R87
10K
dreq1
dreq2
bgack
br
berr
halt
R73
10K
R74
10K
R76
10K
R77
10K
R79
10K
R82
10K
R86
10K
R88
10K
reset
dsack0
dsack1
VCC
+12V
-12
-12V
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
DGND
DGND
D
+12
-12V
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
bypass
<Value>
VCC
+12V
R83
10K
NOTES:
A
VCC
R80
10K
D
-12V
AGND
AGND
A
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
DWN:
Title
Size
C
Date:
MARK TUCCILLO 9/5/01
CHK:
APR:
GP 9/5/01
RAM 9/5/01
COLOR CPU MAIN
Document Number
Rev
24
5143-00-01 (-03 SCHEMATIC)
Friday, May 27, 2005
Sheet
5
of
16
REVISIONS
REV
VCC
VCC
U13
SCC2691AC1A28
15
16
17
18
19
20
21
22
23
24
25
26
27
28
d15
d14
d13
d12
d11
VCC
5
d10
d9
d8
Vcc
A
Y
R51
10K
R52
10K
DATE
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
R53
10K
ether_rxd
ether_txd
ether_cts
ether_rts
a3
a2
a1
VCC
2
1
4
GND
3
20
13
14
10
17
8
VCC
T1IN
T2IN
U14
MAX233ACWP
T1OUT 5
T2OUT 18
R1OUT
R2OUT
R55
10K
R56
10K
C1+
C1VVV+
R1IN
R2IN
4
19
C2+
C2+
C2C2-
12
15
11
16
GND
GND
9
6
rts_e
txd_e
txd_e
rts_e
rxd_e
rxd_e
cts_e
txd_a
rts_a
txd_a
rts_a
rxd_a
cts_a
rxd_a
cts_a
cts_e
VCC
R57
10K
2
1
3
20
13
14
10
17
8
VCC
7
R54
10K
T1IN
T2IN
U15
MAX233ACWP
5
T1OUT 18
T2OUT
R1OUT
R2OUT
C1+
C1VVV+
R1IN
R2IN
4
19
C2+
C2+
C2C2-
12
15
11
16
GND
GND
9
6
VCC
d[0..15]
a[0..20]
d[0..15]
U16
82C684CJ/44
a[0..20]
VCC
R356
10K
r/w
pcs0-8
irq3
iack3
d8
d9
d10
d11
d12
d13
d14
d15
7
8
9
10
11
12
13
14
D0
D1
D2
D3
D4
D5
D6
D7
a1
a2
a3
a4
a5
35
34
33
32
31
A1
A2
A3
A4
A5
r/w
pcs0-8
15
16
R/W
CE
irq3
iack3
21
20
19
1
43
5
6
INTRN
IACK
DTACK
SEL
RESET
CLK
X2
28
38
IEI
RDN
VCC
reset
3_MHz
reset
3MHz
R58
10K
R59
10K
R60
10K
R61
10K
R62
10K
R63
10K
R64
10K
R65
10K
R66
10K
R67
10K
R68
10K
R69
10K
R70
10K
R71
10K
RXD A
TXD A
RTS A
CTS A
39
40
41
36
RXD B
TXD B
RTS B
CTS B
4
3
42
37
rxd_b
txd_b
rts_b
cts_b
RXD C
TXD C
RTS C
CTS C
27
26
25
30
rxd_c
txd_c
rts_c
RXD D
TXD D
RTS D
CTS D
17
18
24
29
rxd_d
txd_d
rts_d
cts_d
TIMER 1
TIMER 2
2
22
VCC
DGND
44
23
rxd_b
txd_b
rts_b
cts_b
rxd_c
txd_c
rts_c
cts_c
rxd_d
txd_d
rts_d
cts_d
audio_a
audio_b
audio_a
audio_b
VCC
VCC
5
+5VDC
8
R361
100 OHM
OUT
GND
2
VCC
RDN
RXD
TXD
MP0
MP1
NC
NC
A2
A1
A0
X1/CLK
X2
RESET
R50
10K
3
RESET
lbwe
VCC
VCC
4
U102
SN74AHC1G04DCK
GND
INTRN
CEN
D7
D6
D5
D4
D3
NC
D2
D1
NC
D0
WRN
1
2
3
4
5
6
7
8
9
10
11
12
13
14
7
irq6
pcs4_8
18
19
20
21
22
23
24
VCC
lbre
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
GP
REVISED PER ECN 1274.
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
Y1
3.6864MHZ SMT
NOTES:
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
GND
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
DWN:
MARK TUCCILLO 9/5/01
Title
Size
C
Date:
CHK:
APR:
GP 9/5/01
RAM 9/5/01
COMM
Document Number
Rev
24
5143-00-01 (-03 SCHEMATIC)
Friday, May 27, 2005
Sheet
4
of
16
REVISIONS
REV
VCC
18
19
20
21
22
23
24
DGND
VCC
+12
D
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
GP
REVISED PER ECN 1274.
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
DATE
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
AGND
+12
-12
A
-12
+12
+12
+12
+12
+12
+12
C28 +
C29
4.7UF TANT
.1UF CER
16V 3528
0805
C30
.1UF CER
0805
C31
.1UF CER
0805
C32
.1UF CER
0805
C33
.1UF CER
0805
C36
C35 + .1UF CER
4.7UF TANT
0805
16V 3528
-12
-12
C37
.1UF CER
0805
C38
.1UF CER
0805
C39
.1UF CER
0805
C40
.1UF CER
0805
VCC
VCC
C46 +
C47
4.7UF TANT
.1UF CER
16V 3528
0805
VCC
C52
.1UF CER
0805
A
VCC
VCC
VCC
-12
VCC
VCC
VCC
VCC
C41
.1UF CER
0805
VCC
C48
.1UF CER
0805
VCC
C53
.1UF CER
0805
C42
.1UF CER
0805
VCC
C49
.1UF CER
0805
VCC
C54
.1UF CER
0805
C43
.1UF CER
0805
VCC
C50
.1UF CER
0805
VCC
C55
.1UF CER
0805
+12
C34
.1UF CER
0805
C44
.1UF CER
0805
VCC
NOTES:
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
C45
.1UF CER
0805
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
VCC
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
C51
.1UF CER
0805
VCC
IVY BIOMEDICAL SYSTEMS INC.
C56
.1UF CER
0805
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
DWN:
D
MARK TUCCILLO 9/5/01
Title
Size
A
Date:
CHK:
APR:
GP 9/5/01
RAM 9/5/01
BYPASS
Document Number
Rev
24
5143-00-01 (-03 SCHEMATIC)
Friday, May 27, 2005
Sheet
3
of
16
C13
R14
10K
.1UF CER 0805
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
R15
VCC
100K
8
R16
2
1
+5VREF
U4A
3
TL062ACD
100K
+5VREF
C209
.1UF CER
0805
R19
R20
1M
100K
13
12
11
10
9
8
7
6
DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7
pcs5-8
16
15
14
CS!
RD!
BUSY!
VREF
2
AIN
BOFS
4
3
R21
18
VDD
AGND
U4B
R24
5
Battery
R27
49.9K
17
A
R29
100K
5
filter_clka
AGND
Vout
8
2
V+
V-
7
3
NC
NC
6
4
Vin
CLK
5
filter_clkb
C17
d8
d9
d10
d11
d12
d13
d14
d15
a1
14
13
12
11
10
9
8
7
6
lbwe
pcs1-8
16
15
17
VDD
DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7
DA/DB
2
3
4
OUT A
RFB A
VREF A
13
12
C19
R28
14
C20
.1UF CER
0805
R25
1
AGND
DGND
A
33.2K
6.8UF TANT
35V 7343
20
19
18
OUT B
RFB B
VREF B
1
SHUTDOWN Vo2
8
2
BYPASS
GND
7
3
+IN
Vdd
6
4
-IN
Vo1
-12
5
2
3
4
A
QRS
C23
100PF NPO
0805
A
A
1
A
U8
LM4861M
R30
100K
C212
.1UF CER
0805
R31
100K
U7A
TL034ACD
+12
TP27
SPEAKER (-)
TP28
SPEAKER (+)
C213
.1UF CER
0805
ALARM
speakerspeaker+
A
d[0..15]
R32
a[0..20]
10K
10K
AGND
10
VREF
19
VDD
18
8
U7C
TL034ACD
A
VCC
D
d8
d9
d10
d11
d12
d13
d14
d15
20
19
18
17
16
15
14
13
DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7
a1
a2
a3
24
23
22
A0
A1
A2
pcs2-8
21
WR
VOUT 1
VOUT 2
VOUT 3
VOUT 4
VOUT 5
VOUT 6
VOUT 7
VOUT 8
VDD
VSS
VREF
DGND
ds
pcs5-8
R37
R38
20K
C25
.1UF CER
0805
20K
R41
D
10K
lbwe
lbwe
A
pcs1-8
pcs1-8
9
10
+
pcs5-8
1
10
11
12
+12
-5VREF
+5VREF
R36
D
10K
-12
2
VIN
4
GND
VOUT
U12B
TL034ACD
ecgX1000
8
U12C
TL034ACD
5 +
6 -
7
C210
.1UF CER
0805
6
R390
499 OHM
ecg
R39
12 +
13 -
10K
14
C27
100K
A
R40
10K
R42
A
A
+5VREF
TP1
10K .1%
50PPM
.1UF CER
0805
R44
10K .1%
50PPM
+5
U12D
TL034ACD
C26
.1UF CER
0805
R43
pcs2-8
pcs2-8
ETCO2
IBP1
IBP2
PLETH
HR
SpO2
U11
AD680JR
+5
.033UF MP
2220
VCC
cs_hires_dac
RESP
IBP1
IBP2
PLETH
HR
SpO2
C24
-
ds
cs_hires_dac
RESP
ETCO2
9
8
7
6
5
4
3
2
R35
10K
7
U7B
TL034ACD
11
9
2
5
2
3
A
4
1
6
U10
AD7228ACR
R34
OUT1
10K
+
RFB
20
ds
cs_hires_dac
-
DGND
17
16
+
3
R33
+5VREF
WR
CS
-
DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7
DB8
DB9
DB10
DB11
+
15
14
13
12
11
10
9
8
7
6
5
4
-
U9
AD7545AKSW
d2
d3
d4
d5
d6
d7
d8
d9
d10
d11
d12
d13
C18
.1UF CER
0805
A
40.2K
A
C22
220PF NPO
0805
D
VCC
6.8UF TANT
35V 7343
R26
+
24.9K
A
WR
CS
+
R23
100K
U7D
TL034ACD
Battery
audio_b
1
VCC
D
C21
.1UF CER
0805
D
audio_a
C16
.1UF CER
0805
A
5
A
DGND
Vin
R17
100K
11
1
CLK
5
4
6.8UF TANT
20V 3528
U6
AD7528ACR
100K
VCC
6
A
A
6
VCC
CLK
7
NC
10K
TL062ACD
7
8
V-
NC
+
filter_clkb
filter_clkb
d8
d9
d10
d11
d12
d13
d14
d15
Vout
V+
3
A
R22
4.02K
U5
AD7574JN
filter_clka
AGND
2
R18
1K
-12
A
1
+
A
4
A
C15
.1UF CER
0805
U3
LTC1069-7CS8
VCC
C14
+12
C208
.1UF CER
0805
filter_clka
U2
LTC1069-7CS8
-
18
19
20
21
22
23
24
DATE
+
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
GP
REVISED PER ECN 1274.
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
-
REVISIONS
REV
1
U12A
TL034ACD
+12
C211
.1UF CER
0805
A
A
R45
100K
R46
100K
R47
ecg_gain
MMBD4148
D5
MMBD4148
1
100K
G
+12
+12
1
R48
100K
VCC
TP21
AGND
R49
100K
NOTES:
-12
A
-12
AGND
DGND
-12
A
D
ECG
S
+12
vdd
ECGX1000
Q6
MMBF4393
3
Q5
CMPT3906
VCC +5 VCC
D
2
D4
2
+5VREF
3
+5VREF
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
DWN:
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
Title
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
Size
C
Date:
MARK TUCCILLO 9/5/01
CHK:
APR:
GP 9/5/01
RAM 9/5/01
ANALOG
Document Number
Rev
24
5143-00-01 (-03 SCHEMATIC)
Friday, May 27, 2005
Sheet
2
of
16
pacer_det
cts_c
txd_c
+12V
nibp_gnd
Battery_Pos
Pump_GND
+12V
Vcc
FB3
+12V_HD
Vbuss
Vcc
HD_GND
DGND
dsp
-12V
-5VI
13
4
Switcher_off
12
5
11
6
10
7
9
8
CONNECT TO
MOUNTING HOLE
D
rec_res
sinen
souten
rec_rxd
rec_txd
A
temp1_plus
temp1_neg
temp1_out
temp2_out
IGND
40C_sel
20C_sel
temp_cal#
-5VI
-6.5VI
9
8
10
7
11
6
12
5
13
4
14
3
15
2
J10
COM 1
7
5
3
1
7
5
3
1
D
D
R358
1K
332 OHM
R359
1K
A
301 OHM
C216
100PF
6KV CD
alarm_out
lead_off
+
C5
4.7UF TANT
16V 3528
Vcc
SpO2 Power Supply
VCC
E1
DGND
C8
10UF TANT
10V 3528
+
TP19
DGND
C9
10UF TANT
10V 3528
C12
10UF TANT
10V 3528
+
ether_iso_rxd
ether_iso_+5V
ISOLATED
D1
MMBD4148
R11
10K
+5VREF
NON
ISOLATED
3
1
8
R12
Ethernet Power Supply
1
Q3
MMBT3904LT1
10K
4
3
10
6
11
5
12
4
13
3
14
2
15
1
16
J13
ANALOG OUT
CONNECTOR
1
3
5
7
9
11
13
15
17
19
21
23
25
2
4
6
8
10
12
14
16
18
20
22
24
26
10
6
11
5
12
4
13
3
14
2
15
A
16
graph_com
6
Vbuss
E6
7
A
+12V
Vcc
E5
9
7
SF5
MEA8020T10R0
JP1
CHASSIS_GND
E4
8
1
E3
9
A
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
A
8
+
Vcc
D
SpO2 ISOLATED
+
+
E2
D
ether_txd
ether_rxd
ETHERNET
ISOLATED
C10
4.7UF TANT
16V 3528
etco2
3
NON
ISOLATED
C6
4.7UF TANT
16V 3528
+
A
SaO2_Xmitt
SaO2_Rec
SaO2_Rts
SaO2_Reset
+5VREF
ISOLATED
C11
4.7UF TANT
16V 3528
+12V
E7
D
ecgX1000_out
ecg_out
resp
ibp1
ibp2
pleth
hr
SaO2
D
D3
MMBD4148
5
2
graph
4
+12V
6
K1
390005
G6K-2F-Y-DC12
6
Alarm Out
7
3
1
8
3
D2
MMBD4148
7
5
2
4
3
Iso_SaO2_Xmitt
Iso_SaO2_Rec
Iso_SaO2_Rts
Iso_SaO2_Reset
IGND_SpO2
J24
ECGX1000 OUT
R13
1
Q4
MMBT3904LT1
10K
2
TRANSMIT DATA
RECEIVE DATA
CTSRESET INPUT-
A
graph_out
2
D
-15VI_SpO2
graph
Battery
Ethernet Power Supply
ether_IGND
ether_iso_txd
D
1
2
3
+
J14
SaO2 INTERFACE
DGND
2
4
6
8
10
12
J12
BEDSIDE WIRELESS
INTERFACE (B.W.I.)
Vcc
SF4
MEA8020T10R0
etco2_aout
resp_aout
ibp1_aout
ibp2_aout
pleth_aout
hr_aout
spo2_aout
cpu_main
Vbuss
1
3
5
7
9
11
J9
DISPLAY
191501
C3
.15UF 63V
MP 2824
C4
.15UF 63V
MP 2824
J8
KOJAK LIGHT
D
0 OHM
A
100UH
SMT
R360
C7
4.7UF TANT
16V 3528
+15VI_SpO2
R380
100UH
SMT
R357
1
2
3
D
ECGX1000
SaO2Wform
RespWform
ecg_aout
txd_b
rxd_b
etco2_res
64KHz
Vbuss Vcc
SF10
ACH3218-472
L1
ecgX1000
ether_rxd
ether_txd
Vcc
D
D
NON
ISOLATED
FRONT END ISOLATED
+5VI_SpO2
+ C1
10UF TANT
10V 3528
C2
.1UF CER
0805
2
4
6
8
D
I
1
2
3
4
5
6
7
8
9
10
11
12
13
14
0 OHM
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
line_enable
rxd_d
txd_d
rts_d
spo2_res
J15
ETHERNET
blue0
blue1
blue2
blue3
blue4
blue5
R382
Vcc
J11
COM 2
L2
+12V
green0
green1
green2
green3
green4
green5
D
1
SF3
MEA8020T10R0
D
2
4
6
8
temp_press
SpO2 Power Supply
red0
red1
red2
red3
red4
red5
SF11
ACH3218-101
D
rxd_e
txd_e
DSP Power Supply
-6.5V
IGND
ISOLATED
temp2_plus
temp2_neg
Q2
IRF7205/SO
D
dot_clock
hsynch
vsynch
17
+5_Vref
+5VI
+6.5V
SG2+
SG2-
1
R10
10K
red[0..5]
green[0..5]
blue[0..5]
16
+12V
Q1
MMBT3904LT1
Q42
MMBT3904LT1
1
D
rta_e
cts_e
DGND
R5
10K
J6
DISPLAY BACKLIGHT
D
100K
rxd_a
txd_a
DSP Power Supply
+12V
1
2
3
J21
FAN
D
temp_press
I
1
2
3
D
kojak_light
red[0..5]
green[0..5]
blue[0..5]
dot_clock
hsynch
vsynch
+6.5VI
+5VI
U1
SP720AB
R9
10K
Vbuss
R384
dsp
SF9
ACH3218-472
R7
10K
Vcc
+12V_Rec
J7
SPEAKER
rta_a
cts_a
I
R8
10K
16
15
14
13
12
11
10
9
V+
IN
IN
IN
IN
IN
IN
IN
2
1
2
3
speaker-
AGND
-12V
D
R6
10K
speaker+
DGND
ISOLATED
press_1
press_2
R4
470K
Vcc
5
VCC
IN
IN
IN
IN
IN
IN
IN
-V
1
porta7
porta4
porta2
porta1
porta0
cts_d
rts_b
rts_c
cts_b
SG1+
SG1-
R3
470K
6
-12V
1
2
3
4
5
6
7
8
3
+12V
I
SF7
ACH3218-472
SF2
MEA8020T10R0
2
Vcc
PS_Synch
+5_Vref
Vbuss
470K
Vcc
+12V
PRES_TEST
PRESS_EN
RDBK0
Vcc
R1
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
DATE
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
J3 RECORDER
fp3
fp2
fp1
fp0
SF8
ACH3218-472
I
D
D
Battery
J19
RESP ADJ
Resp
R2
470K
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
7
1
2
3
4
3
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
Vcc
8
Iso_GND
u/d#
offset_cs#
inc#
gain_cs#
2
14
dsp_txd
dsp_rxd
press_test
press_en
press_1
press_2
sel_40
sel_20
temp_cal
15
D
dsp_txd
dsp_rxd
resp_high
resp_low
base_imp
temp_1
temp_2
resp_high
resp_low
base_imp
Common_LED
On_Off
Power_On_Led
Bat_Maintence_LED
Bat_Charging_LED
Battery_Fault_LED
Vcc
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
cts_c
rxd_c
txd_c
nibp_res
+5VI
+5V_Ref
+6.5VI
RA
1
D
VCC
IGND
-6.5VI
-5Vi
PS_Synch
+5VI
+5V_Ref
sclock0
resp_test
D
J4
FRONT PANEL
D
16
ac
+5VREF
sclk0
resp_test
Resp
SF6
ACH3218-472
8
D
ac
DGND
diag_bw
ecg_test
LA
7
9
1
2
Vcc
ecgX1000
ecgX10
ll_lead
la_lead
ra_lead
v_lead
shield_dr
6
10
cpu_main
pacer_det
LL
LA
RA
RL
V
shield_dr
IGND
+5_VREF
sg1+
sg2+
temp1_2
temp2_2
A
NON
ISOLATED
la_ref
ra_ref
ll_ref
v_ref
rl_ref
J5
ISOLATED INPUTS
2
4
6
8
10
12
14
16
18
20
22
24
5
11
18
ls0
ls1
ls2
1
3
5
7
9
11
13
15
17
19
21
23
4
12
-12V
ecg_front_end
LL
LA
RA
RL
V
shield_dr
IGND
+5_VREF
sg1sg2temp1_1
temp2_1
Power Supply
3
13
17
150014
diag_bw
ecg_test
+5VI
+6.5VI
V
-5VI
-6.5VI
Iso_GND
FB5
J20
210062
J23
THERMISTOR
2
14
18
RL
1
2
3
15
17
FB4
150014
Thermistor_Pos
Thermistor_Neg
TP26
GND
1
3
150014
Vbuss +12V
ecgX1000
ECGX10
LL_Lead
LA_Lead
RA_Lead
V_Lead
1
2
3
4
5
6
Battery_Neg
12V_Switcher
Switcher_GND
16
2
RA
J2
POWER
F1
5ASB
18
LA
2
4
6
8
10
12
14
16
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
REVISED PER ECN 1274.
GP
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
18
19
20
21
22
23
24
18
LA_ref
RA_ref
LL_ref
V_ref
RL_ref
LL
FB2
1
3
5
7
9
11
13
15
REV
D
18
FB1
150014
nibp_res
rxd_c
+12V
nibp_gnd
SF1
MEA8020T10R0
Rec_GND
17
J1
NIBP
shield_dr
150014
REVISIONS
Power Supply
leads0
leads1
leads2
17
ecg_front_end
D
K2
390005
G6K-2F-Y-DC12
D
1
8
7
5
2
4
K3
390005
G6K-2F-Y-DC12
Graph Out
NOTES:
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
DWN:
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
Title
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
Size
MARK TUCCILLO 9/5/01
CHK:
GP 9/5/01
APR:
RAM 9/5/01
5143 Main
Document Number
5143-00-01 (-03 SCHEMATIC)
Date: Friday, May 27, 2005
Sheet
1
Rev
24
D
of
16
8
7
6
5
4
3
2
1
REVISIONS
DESCRIPTION
REVISED PER ECN 691.
REVISED PER ECN 709.
REVISED PER ECN 957.
REVISED PER ECN 1201.
REV
01
02
03
04
BY
GP
GP
GP
GP
DATE
8-28-97
11-4-97
6-25-99
8-9-02
D
D
2336-00-01
J202-1
2
VCC
D
VCC
5 RN1 6
1K
C
7 RN1 8
1K
IGET
ET_RES
E4
B++
E7
3
2
R12
100 OHM
B++
D1
L1
25UH
C3
+ C2
100UF
25V
IGET
C4
1500UF
25V
+ .0068UF
1KV CD
2252-00-10
T1
2
.
SB5100
R6
620 OHM
2W 5%
3
4
R14
150 OHM
10W 5%
H.S.
.
1
64KHz
E5
C9
1UF MP
C8
100UF 25V
730MA
D
H.S.
Q4
1G
IRF540
D2
MBR350
2
2N3903
+
Q1
2
2N3906
C10
.1UF MP
R2
10K
.047UF MP
8
7
6
5
REF
R_MID
GND-F
GND-S
IGET
COLLECTOR
COMP
V+
R_TOP
10000PF
NPO
R4
1
2
3
4
+ C6
22UF
TANT
3
Q2
S
R8
6.98K
3
Q
7
DIS
6
THR
R10
10K
U7
ICM7555
TR
R13
1K
C7
2
47PF NPO
R11
10K
CV
5
R9
10K
C5
1000PF
NPO
2
2N3903
B
Q3
3
U6
CNY17-3
C11
D
4
C1
2
IGET
TP2
1
IGET
5
2
B
3
4
U2
BNX002-01
1
1 1
+5VIET
TP1
3
2
4
U5
CNY17-3
1
C
+5VIET
VCC
E3
4
5
IGET
3 RN1 4
1K
9 RN1 10
1K
8
1K
4
2
R3
R5
2K
E2
R
+5VIET
1
2
3
RXD_B
U4
CNY17-3
V+
+5VIET
J204
173011
E1
TXD_B
V-
1
+5VIET
1 RN1 2
1K
IGET
+5VIET
5
IGET
1
5
U3
CNY17-3
R1
2K
1
J202-4
4
J202-3
1
VCC
TTL RECEIVE DATA
TTL TRANSMIT DATA
TTL EXTERNAL RES
J202-2
R7
10K
E6
D_GND
D
DGND
TP3
33.2K
U1
LT1431CN8
IVY BIOMEDICAL SYSTEMS INC.
A
NOTES:
1. ALL RESISTORS ARE 1/4W, 1% U.O.S.
Title
ETCO2 MODULE INTERFACE BOARD ASSEMBLY
2. LAST USED:
Size
3. NOT USED:
B
Date:
8
7
6
5
A
11 BUSINESS PARK DRIVE
BRANFORD, CT. 06405
(203) 481-4183
4
3
Document Number
Rev
04
5108-00-01 (-03 SCHEMATIC)
Thursday, March 10, 2005
2
Sheet
1
of
1
1
Video Add
va[0..16]
Video Data
vd[0..15]
Wave Add
wa[0..14]
Wave Data
wd[0..7]
Host Data
Host Add
sof_int
eow_int
clear
GND
VCCINT
VCCIO
clock
dsack1
green5
green3
blue3
blue2
blue4
green4
GND
VCCINT
VCCIO
blue5
va14
va10
va11
va12
R370
100 OHM
25_MHz
dot_clock
va13
wa6
A1
hsynch
va5
wa12
wa7
wa8
VCCIO
GND
wa9
GND
Data0
wd6
A15
A6
A14
A10
A11
A4
A5
VCCIO
hd11
hd10
hd15
hd14
vd2
vd1
vd6
vd5
vd9
vd12
vd11
vd15
wd3
wd2
dclk
U95
EPF6016QC208-2
vga_fpga
vramadd13
RESERVED
RESERVED
RESERVED
hd0
hd2
hd3
hd6
hd8
GND
VCCIO
hd9
hd12
hd13
video_data0
video_data3
video_data4
video_data7
video_data8
video_data10
video_data13
video_data14
wave_data0
wave_data1
nconfig
GND
VCCIO
nstatus
wave_data4
wave_data5
wave_data7
host_add3
host_add13
host_add12
host_add9
host_add0
host_add8
RESERVED
host_add2
RESERVED
RESERVED
RESERVED
GND
VCCIO
RESERVED
host_add7
audio_a
line_enable
RESERVED
RESERVED
RESERVED
vram_add8
A16
A17
green0
wram_we
green1
green2
host_add20
host_add18
^nCE
GND
VCCINT
VCCIO
host_add16
host_add17
green0
cs_wform_ram
sof_int
eow_int
zone_inc
reserved
reserved
reserved
vert_wave_time
reserved
reserved
reserved
clear
GND
VCCINT
VCCIO
clock
RESERVED
RESERVED
RESERVED
RESERVED
RESERVED
dsack0
RESERVED
RESERVED
green5
green3
blue3
blue2
blue4
green4
GND
VCCINT
VCCIO
MSEL
blue5
vramadd14
vramadd10
25MHz
vramadd11
vramadd12
53
54
55
56
57
58
59
60
61
GND
62
VCCIO
63
hd9
64
hd12
65
hd13
66
vd0
67
vd3
68
vd4
69
vd7
70
vd8
71
vd10
72
vd13
73
vd14
74
wd0
75
wd1
76
77
GND
78
VCCIO
79
nstatus
80
wd4
81
wd5
82
wd7
83
A3
84
A13
85
A12
86
A9
87
A0
88
A8
89
90
A2
91
92
93
94
GND
95
VCCIO
96
97
A7
98
audio_a_clk 99
100
101
102
103
104
GND
VCCINT
VCCIO
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
hd0
hd2
hd3
hd6
hd8
wram_we
green1
green2
A20
A18
wave_add14
wave_add11
wave_add10
vram_add9
hd1
hd5
hd4
hd7
VCCIO
GND
hd11
hd10
hd15
hd14
video_data2
video_data1
video_data6
video_data5
video_data9
wave_data12
wave_data11
wave_data15
wave_data3
wave_data2
dclk
VCCIO
GND
data0
wave_data6
host_add15
host_add6
host_add14
host_add10
host_add11
host_add4
host_add5
wram_cs
wave_add9
RESERVED
RESERVED
RESERVED
RESERVED
VCCIO
GND
RESERVED
wave_add6
host_add1
hsynch
vram_add5
wave_add12
wave_add7
wave_add8
208
207
206
205
204
203
202
201
200
199
198
197
196
195
194
193
192
191
190
189
188
187
186
185
184
183
182
181
180
179
178
177
176
175
174
173
172
171
170
169
168
167
166
165
164
163
162
161
160
159
158
157
VCCIO
GND
wa14
wa11
wa10
va9
hd1
hd5
hd4
hd7
U94
CY7C199-12VI
wa0
wa1
wa2
wa3
wa4
wa5
wa6
wa7
wa8
wa9
wa10
wa11
wa12
wa13
wa14
wave_add13
vram_add3
vram_add4
vram_cs
host_add19
vsync
CONF_DONE
VCCIO
VCCINT
GND
host_add21
vram_add2
cs_hires_dac
vram_add16
vram_add15
vram_add1
wave_add0
wave_add2
wave_add5
wave_add4
wave_add3
wave_add1
RESERVED
RESERVED
intack7
VCCIO
VCCINT
GND
intack5
host_cs3
ds
siz0
as
vram_rd
siz1
r/w
vram_rd
red5
blue1
blue0
red0
red1
red2
red3
VCCIO
VCCINT
GND
red4
audio_b
vram_add0
vram_add6
vram_add7
156
155
154
153
152
151
150
149
148
147
146
145
144
143
142
141
140
139
138
137
136
135
134
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
wa13
va3
va4
vram_cs
A19
vsynch
conf_done
VCCIO
VCCINT
GND
A21
va2
cs_hires_dac
va16
va15
va1
wa0
wa2
wa5
wa4
wa3
wa1
10
9
8
7
6
5
4
3
25
24
21
23
2
26
1
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14
14
DGND
12
11
10
9
8
7
6
5
27
26
23
25
4
28
3
31
2
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14
A15
A16
16
GND
D0
D1
D2
D3
D4
D5
D6
D7
11
12
13
15
16
17
18
19
CE
OE
WE
20
22
27
wd0
wd1
wd2
wd3
wd4
wd5
wd6
wd7
wram_we
VCC
VCC
28
U96
CY7C1009-20VC
va0
va1
va2
va3
va4
va5
va6
va7
va8
va9
va10
va11
va12
va13
va14
va15
va16
intack7
VCCIO
VCCINT
GND
intack5
cs2
ds
siz0
as
siz1
r/w
vram_we
red5
blue1
blue0
red0
red1
red2
red3
D0
D1
D2
D3
D4
D5
D6
D7
13
14
15
17
18
19
20
21
CS1
CS2
22
30
WE
OE
29
24
NC
1
VCC
vd0
vd1
vd2
vd3
vd4
vd5
vd6
vd7
vram_we
vram_cs
VCC
32
VCCIO
VCCINT
GND
red4
audio_b_clk
va0
va6
va7
U97
CY7C1009-20VC
va0
va1
va2
va3
va4
va5
va6
va7
va8
va9
va10
va11
va12
va13
va14
va15
va16
va8
line_enable
line_enable
12
11
10
9
8
7
6
5
27
26
23
25
4
28
3
31
2
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14
A15
A16
16
GND
D0
D1
D2
D3
D4
D5
D6
D7
13
14
15
17
18
19
20
21
CS1
CS2
22
30
WE
OE
29
24
NC
1
VCC
vd8
vd9
vd10
vd11
vd12
vd13
vd14
vd15
vram_we
vram_cs
VCC
32
audio_a_clk
audio_a_clk
audio_b_clk
blue[0..5]
blue[0..5]
green[0..5]
green[0..5]
audio_b_clk
blue[0..5]
green[0..5]
red[0..5]
R330
10K
hd[0..15]
A[0..21]
clear
clock
cs2
intack5
intack7
hd[0..15]
Data0
dclk
nstatus
conf_done
A[0..21]
clear
clock
siz1
as
r/w
ds
1
2
3
4
intack5
VCC
Data
DCLK
OE
nCS
VCC
VCC
nCASC
GND
r/w
ds
dot_clock
25_MHz
8
7
6
5
NOTES:
8 PIN
25_MHz
sof_int
sof_int
eow_int
1. LAST USED: C230, D44, DS6, E7, F4, FB5, J24,
JP8,K3, L15, Q42, R392, RT2, SF11,
T5, TP51,U103, Y4.
eow_int
cs_hires_dac
cs_hires_dac
dsack1
dsack1
2. NOT USED: C221,J18,J22,JP2,JP3,R89,R388.
VCCINT
3. ALL RESISTORS ARE 1/10W, SIZE 0805 UNLESS
OTHERWISE SPECIFIED.
siz0
IVY BIOMEDICAL SYSTEMS INC.
REVISIONS
GND
vsynch
dot_clock
siz1
as
hsynch
vsynch
SOCKET
VCCIO
intack7
red[0..5]
hsynch
VCC
U98
2634-00-04
cs2
VCC
siz0
R331
10K
REV
18
19
20
21
22
23
24
DESCRIPTION
BY
FOR REVISION LEVEL HISTORY REV-00 THRU REV-17
REFER TO ECN 1274 IN THE ENGINEERING RECORD FILES.
GP
REVISED PER ECN 1274.
REVISED PER ECN 1275.
GP
REVISED PER ECN 1289.
GP
REVISED PER ECN 1290.
GP
REVISED PER ECN 1326.
GP
REVISED PER ECN 1354.
GP
REVISED PER ECN 1432.
GP
DATE
11 BUSINESS PARK DRIVE
BRANFORD, CT., 06405 USA
(203) 481-4183
DWN: MARK TUCCILLO 9/5/01
6-11-03
6-11-03
6-25-03
6-25-03
3-4-04
4-13-04
5-27-05
Title
Size
C
Date:
CHK:
GP 9/5/01
APR:
RAM 9/5/01
VIDEO FPGA
Document Number
Rev
24
5143-00-01 (-03 SCHEMATIC)
Friday, May 27, 2005
Sheet
16
of
16
4
3
2
1
J3
PRESSURE 1
J4
PRESSURE 2
J2
ECG
F
E
D
C
B
A
5
A
B
C
D
E
F
J5
TEMP 1
6
A
B
C
D
E
F
J6
TEMP 2
2
1
7
2
1
8
D
D
I
FB6
150013
I
FB5
FB4
FB3
FB1
FB7
FB8
FB10
FB9
FB11
150010
(FB-7 THRU FB-11)
FB2
150013
FB15
150010
150013
23
21
19
17
15
13
11
9
7
5
3
1
C3
.1UF CER
0805
C1
.1UF CER
0805
C4
.1UF CER
0805
I
I
B
I
FB14
150013
FB12
150010
J1
170514
ISOLATED INPUTS
24
22
20
18
16
14
12
10
8
6
4
2
FB18
150010
B
LA
TEMP2_2
150013
150013
FB19
LL
TEMP2_1
FB16
C
FB13
I
150010
RA
TEMP1_2
FB17
RL
TEMP1_1
V
C2
.1UF CER
0805
SHIELD_DR
5V
SG1-
SG2-
SG1+
SG2+
C
I
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVE
BRANFORD, CT. 06405
(203) 481-4183
A
A
Title
ISOLATED INPUT BOARD (ECG/RESP,2P & 2T)
REVISIONS
DESCRIPTION
REVISED PER ECN 1219.
REVISED PER ECN 1220.
REVISED PER ECN 1221.
REV
01
02
03
8
7
BY
GP
GP
GP
6
5
DATE
12-16-02
12-16-02
12-16-02
4
Size
A
Document Number
Date:
Monday, December 16, 2002
Rev
5144-00-01 (-03 SCHEMATIC)
3
Sheet
2
1
03
1
of
1
Download