Count rate of 30035 and 30020 SiPM Count rate/dynamic range of 30035 - - Fig 1 shows the dynamic range of 30035 when a bias voltage of 28 V was applied at the SiPM. Wavelength = 405 nm (L405P150). dc light (circular guassian beam ≥ 20 cm at 0 degree AOI using diffuser) was transmitted at the SiPM Intensity of the light was varied with wire-grid polariser and monitored by Newports’ calibrated PD At each intensity, the SiPM bias current/photocurrent was measured using Keithley digital ammeter (195A) The measured SiPM photocurrent against a range of irradiance is shown in Fig. 1 (dashed line) The measured photocurrent can be converted into average counts (dotted line in Fig. 1) using the following realtion (1.1) where Idark equals to 0.53 µA at 28 V Ndark is the average number of dark counts equal to approximately 1 Gcps measured using peak counting method ISiPM is the average photocurrent at a given irradiance and Nph_det is the average number of detected counts per second At higher values of irradiance the SiPM bias current or photocurrent starts to saturate, which means that the count rate will also saturate accordingly Fig. 1 SiPM bias current and count rate against irradiance at the SiPM of 30035 - The count rate (dotted line) in Fig. 1 saturates at around 54 Gcps. Model for 30035 1. SensL[1] - Equation 1.2 gives the number of fired microcell (Nfired) against the number incident photons, which can be used to predict the output of a SiPM ππππππ = ππππ (1 − π - −ππ·πΈ×ππβ ππππππππ‘ ππππ ) (1.2) where PDE is the photon detection efficiency of the SiPM, Nph_incident is the number of incident photon per unit of time over activer area of a SiPM and Meff is the effective number of microcells, which depends upon the microcell recovery time (Trecovery), the pulse duration (Tpulse), and the total number of microcells. The effective number of microcell is given as ππππ πππ΄ππ·π , = { πππ’ππ π , πππ’ππ π < ππππππ£πππ¦ πππ’ππ π > ππππππ£πππ¦ ππππππ£πππ¦ - The microcell recovery time (Trecovery) is given as ππππππ£πππ¦ = πΆµππππ (π π + π π × ππππ΄π·π ) - - (1.3) (1.4) where Cµcell is the effective capacitance of the microcell, RS is the series resistance, and RQ is the quenching resistance To predict the output of the 30035 SiPM, the SiPM has NSPADs = 5676, PDE = 46% at 28 V, Trecovery = 85 ns with Rs = 50 β¦ (Evaluation Board), calculated from Equation 1.4 with Cµcell = 164 fF at 40 degree celcius [N.otte.A et al], RQ = 220 kβ¦, calculated from Equation 1.4 with known Trecovery = 37 ns with Rs = 1 β¦. Fig. 2 shows the output of the 30035 SiPM in terms of the number of fired microcell estimated with Equation 1.2 Fig. 2 The number of fired microcell against a range of irradiance for the 30035 SiPM. The dashed line shows the number of fired microcell when the Trecovery = 85 ns and solid line when Trecovery = 104 ns - The number of fired microcell saturates at approx: at 66 Gc/s when Trecovery = 85 ns and saturates at approx: 54 Gc/s when Trecovery = 104 ns If the number of fired microcell per second is equal to the number of detected photons per second then the effective recovery time is equal to 104 ns (i.e. 1.2×RC or 70% of max amplitude), as the Equation 1.2 shows the number of fired microcell equal to approx: 54 Gc/s, when the recovery time is 104 ns. Particularly, the max count rate calculated using equation 1.1 is also approximately equal to 54 Gcps 2. Max count rate of a passively quenched SiPM[2],[3] - For a passive quenching SPAD array, the maximum count rate can be calculated [2],[3] π ×ππππ΄π·π πΆπππ₯ = expππ’ππ π (1)×π πππππ£πππ¦ - (1.5) Table 1 shows the number of fired microcell based on Equation 1.2 and the maximum count rate based on Equation 1.5 for various recovery times, and the maximum count rate calculated using Equation 1.1 Maximum number of fired microcell Gc/s (Equation 1.2) ON Semi SiPM Trecovery = 85 ns RS = 50 β¦ T’recovery = 104 ns Micro30035 66 54 Maximum count rate Gcps (Equation 1.1) Maximum count rate Gcps (Equation 1.5) Trecovery = 37 ns (Datahseet RS = 1 β¦) Trecovery = 45 ns (Datasheet RS = 1 β¦) Trecovery = 85 ns (Equation 1.4 RS = 50 β¦) 56.4 46.4 24.5 54.5 Trecovery = 15 ns Trecovery = 59 ns (Datahseet (Equation 1.4 137.8 RS = 1 β¦) RS = 50 β¦) 244.2 138.6 353 89.8 Table 1: Number of maximum fired pixel (Equation 1.2), max count rate (Equation 1.5), and the max count rate (Equation 1.1) Micro30020 Trecovery = 59 ns RS = 50 β¦ T’recovery = 104 ns Count rate of 30020 - Fig. 2 shows the dynamic range of 30020 when a bias voltage of 28 V was applied at the SiPM. Wavelength = 405 nm (L405P150). dc light (circular guassian beam ≥ 20 cm at 0 degree AOI using diffuser) was transmitted at the SiPM Intensity of the light was varied with wire-grid polariser and monitored by Newports’ calibrated PD At each intensity, the SiPM bias current/photocurrent was measured using Keithley digital ammeter (195A) The measured SiPM photocurrent against a range of irradiance is shown in Fig. 2 (triangular marker) The measured photocurrent can be converted into average counts (solid line in Fig. 1) using Equation 1.1 where Idark equals to 0.15 µA at 28 V Ndark is the average number of dark counts equal to approximately 0.6 Gcps measured using peak counting method ISiPM is the average photocurrent at a given irradiance and Nph_det is the average number of detected counts per second Since 300320 has a large number of microcells compared to the 30035 (2.5 times the 30035), so ideally, the 30020 should be linear at higher irradiance As expected, the 30020 saturates at higher values of irradiances compared to the 30035 and accordint to Equation 1.1, the number of counts saturates at around 137 Gcps as shown in Fig. 2. Fig. 2 SiPM bias current and count rate against irradiance at the SiPM of 30020 - Fig. 3 shows the count rate of 30020 and 30035 measured using Equation 1.1 Fig. 3 Comparison of count rate of 30020 and 30035 Model for 30035 1. SensL[1] - - To predict the output of the 30020 SiPM, the SiPM has NSPADs = 14410, PDE = 30% at 28 V, Trecovery = 59 ns with Rs = 50 β¦ (Evaluation Board), calculated from Equation 1.4 where Cµcell = 62 fF can be calculated using the following relation[1] π πΆπ’ππππ = πΊ × πππ£ o e is the elementary charge and G is the gain and Vov is the overvoltage applied at the SiPM RQ = ~220 kβ¦ can be calculated from Equation 1.4 with known Trecovery = 15 ns with Rs = 1 β¦. Fig. 4 shows the output of the 30020 SiPM in terms of the number of fired microcell estimated with Eq: 1.2 Fig. 4 The number of fired microcell against a range of irradiance for the 30035 SiPM. The dashed line shows the number of fired microcell when the Trecovery = 59 ns and solid line when Trecovery = 104 ns - The number of fired microcell saturates at approx: at 244 Gc/s when Trecovery = 85 ns and saturates at approx: 138 Gc/s when Trecovery = 104 ns If Nfired per second is equal to the number of detected photons per second then the effective recovery time is equal to 104 ns (i.e. 1.2×RC or 70% of max amplitude), as the Equation 1.2 shows the number of fired microcell equal to approx: 138 Gc/s, when the recovery time is 104 ns. Moreover, the max count rate calculated using equation 1.1 is also approximately equal to 138 Gcps Tentative conclusion/summary - Summary of the number of fired microcells per second and the count rate of 30035 and 30020 SiPM has been shown in Table 1. Ratio of the total number of microcells of 30020 to the total number of microcells 30035 ≈ 2.5 Ratio of measured count rate of 30020 to the measured max count rate of 30035 using Equation 1.1 ≈ 2.5 Fig. 5. Dark count results by Will