BJUI Safer urethral catheters: how study of catheter balloon pressure and force can guide design BJU INTERNATIONAL Alex K. Wu*, Sarah D. Blaschko*, Maurice Garcia*, Jack W. McAninch* and David S. Aaronson*† *Departments of Urology, University of California, San Francisco, CA, and †Kaiser Permanente, Oakland, CA, USA Accepted for publication 21 April 2011 OBJECTIVES • To better define urethral catheter balloon pressures and extraction forces during traumatic placement and removal of urethral catheters. • To help guide design for safer urethral catheters. MATERIALS AND METHODS • Measurements of balloon pressure were made upon filling within the urethra vs the bladder. • Extraction forces were measured upon removal of a catheter with a filled balloon from the bladder. • Models for the bladder and urethra included an ex vivo model (funnel, ‘bladder’, attached to a 30 F tube, ‘urethra’) and fresh human male cadavers. • The mean (SEM) balloon pressures and extraction forces were calculated. RESULTS • In the ex vivo model, the mean (SEM) pressures upon filling the balloon with INTRODUCTION The urethral catheter is ubiquitous in medical care. A urethral catheter is inserted in nearly a quarter of all hospitalized patients [1]. Unlike other medical devices, all levels of healthcare professionals from doctors to nursing trainees place urinary catheters [2,3]. National healthcare efforts have focused on reducing the incidence of catheter-related UTIs, deemed a ‘never event’ by the Centers for Medicare and Medicaid Services [4]. However, catheter-related iatrogenic injuries receive little attention despite the fact that an estimated 0.3% of 111 0 What’s known on the subject? and What does the study add? The urethal catheter is a ubiquitous device that has not been modified or improved for safety in the last 20 years, although it can be associated with significant patient harm. This study utilizes force and pressure measurements of the urethral catheter in order to aid in future safety modifications. 10 mL were on average three-times higher within the ex vivo ‘urethra’ (177 [6] kPa) vs ‘bladder’ (59 [2] kPa) across multiple catheter types. • In the human cadaver, the mean balloon pressure was 1.9-times higher within the urethra (139 [11] kPa) vs bladder (68 [4] kPa). • Balloon pressure increased non-linearly during intraurethral filling of both models, resulting in either balloon rupture (silicone catheters) or ‘ballooning’ of the neck of the balloon filling port (latex catheters). • Removal of a filled balloon per the ex vivo model ‘urethra’ and cadaveric urethra, similarly required increasing force with greater balloon fill volumes (e.g. 9.34 [0.44] N for 5 mL vs 41.37 [8.01] N for 10 mL balloon volume). CONCLUSIONS hospital patients suffer catheter placement injuries [5]. Additionally, accidental removal of an urethral catheter with a filled balloon is estimated to occur in 5% of intensive care unit (ICU) patients [6]. Iatrogenic urethral catheter injury represents a significant safety issue that warrants further investigation as it can cause pain, severe bleeding, urethral stricture, and need for surgical intervention. catheter placement and removal. Significant urethral injury can occur when the catheter balloon is inappropriately filled in the relatively non-distensible urethra [7,8] or when the filled catheter balloon is accidentally pulled out of the bladder [6]. Iatrogenic catheter-related urethral injury occurs by various mechanisms. To identify ways to prevent catheter-related injuries, we focused on the two main mechanisms: • Iatrogenic complications from improper urethral catheter use is common. • Catheter balloon pressures and manual extraction forces associated with urethral injury are significantly greater than those found with normal use. • The differences in pressure and force may be incorporated into a safer urethral catheter design, which may significantly reduce iatrogenic urethral injury associated with catheterization. KEYWORDS urethral catheter, urethral trauma, safety, iatrogenic injury, biomedical device design The force used to place urethral catheters has been studied in an ex vivo model [9], but, to our knowledge, no study has shown force and pressure measurements associated with urethral catheter injury. We hypothesize that urethral catheter balloon pressures are greatly increased when the balloon is filled within the urethra vs the © BJU INTERNATIONAL © 2 0 11 T H E A U T H O R S 2 0 11 B J U I N T E R N A T I O N A L | 1 0 9 , 111 0 – 111 4 | doi:10.1111/j.1464-410X.2011.10510.x URETHRAL CATHETER BALLOON PRESSURE AND FORCE FIG. 1 The 5 mL balloon of this latex catheter is filled with 10 mL of water while placed in a non-distensible 30 F tube. Note how the balloon port has expanded due to the high pressures within the retention balloon system. bladder. Also, we hypothesize that withdrawal of a filled catheter balloon from the bladder into the urethra requires significant force. Our overarching aim is to design a safer urethral catheter to prevent unnecessary patient harm. MATERIALS AND METHODS INTRA-URETHRAL FILLING STUDIES USING AN EX VIVO URETHRAL MODEL Studies were conducted using a mock urinary bladder neck and urethra, using a rigid plastic funnel connected to a well lubricated 30 F semi-rigid tube. We used 16 F latex catheters with 5 mL retention balloons (Allegiance, USA; recommended fill volume: 10 mL), 16 F silicone catheters with 5 mL retention balloons (Rochester Medical, USA; recommended fill volume 10 mL), and 22 F latex catheters with 30 mL retention balloons (Allegiance, USA; recommended fill volume 35 mL). Pressure within the balloon was measured using a pressure gauge (Marsh Pressure Gauge, Marshall Instruments, Inc., Anaheim, CA, USA) attached to the catheter balloon filling-port. Pressure measurements were recorded initially and at 10 s upon filling the balloon within the ‘urethra’ and the ‘bladder’ using 0, 5, 7, and 10 mL of water. The 10-s measurement was added after the catheter system was noted to require time to equilibrate. Care was taken to exclude air from the retention balloon system. All studies were repeated in triplicate; result are reported as the mean (SEM). © INTRA-URETHRAL FILLING STUDIES: CADAVER MODEL Lubricated 16 F latex catheters with 5 mL balloons were inserted per urethra of three fresh male cadavers in standard fashion [10]. Sterile water was used to fill the catheter balloon with 0, 5, and 10 mL in the bladder and at defined locations within the urethra, including the prostatic urethra, the bulbar urethra, and the pendulous urethra. The location was estimated as the catheter was withdrawn past the bladder neck. Pressure within the balloon was recorded when the balloon was filled with each amount of water as described above. The same catheter was used sequentially with deflation of the balloon before movement to the next segment of the urethra. If balloon port expansion occurred from high balloon port pressures, then a new catheter was used. Additional measurements were taken with the catheter balloon filled with 0–5 mL of water in the fossa navicularis to replicate the pressure effects during retrograde urethrography. Pressure was measured upon initial filling of the balloon and 10 s afterward to account for tissue relaxation and system equilibration. At 10 s after balloon filling, no further changes in balloon pressure were noted (data not shown). The experiments were repeated in a total of three cadavers. Individual cadavers were not used for multiple experiments due to concern that the urethral tissue would be significantly altered by each experiment. FORCED REMOVAL OF A URETHRAL CATHETER (BALLOON FILLED) PER URETHRA: EX-VIVO MODEL Studies using our ex vivo model (described above) were performed. Latex and silicone 16 F catheters (5 mL balloon) were filled with 5, 7, 10 and 20 mL of sterile water. Latex 22 F (30 mL balloon) catheters were filled with 5, 10, 35, and 55 mL of sterile water. The largest volumes instilled represent the common practice of balloon over-filling to reduce postoperative haematuria after surgical procedures such as TURP. A force meter (Accu-Scale & Systems Inc, Melrose, MA, USA) was used to measure the force required to remove the catheter from the ‘bladder’ with the balloon filled at varying amounts. All studies were repeated in triplicate with mean (SEM) reported. FORCED REMOVAL OF A URETHRAL CATHETER (BALLOON FILLED) PER URETHRA: CADAVER MODEL Lubricated 16 F latex catheters were placed per urethra in three fresh male cadavers using standard technique. The catheter balloon was filled with 0, 5, 7, and 10 mL of sterile water. After filling to the designated amount, the catheter was forcibly removed. Experimental measurements were performed as describe above for studies using the ex vivo model. After force measurements were obtained for removal with 10 mL of fill, the cadaver was not used again out of concern that the bladder neck would be significantly altered by catheter balloon removal. Experimentation was repeated in three cadavers. All cadavers were obtained by the University of California, San Francisco, Orthopedic Surgery department by voluntary donation for the purpose of medical training and research. RESULTS PRESSURE ASSOCIATED WITH FILLING THE BALLOON IN THE URETHRA The catheter balloons, free of external constriction, had good compliance. For all catheters (varying by material type and balloon volume), the mean (SEM) balloon pressure within the ‘bladder’ was 59 (2) kPa. Balloon pressures in the ‘bladder’ model, using a 16 F latex catheter, increased from 63 (2) kPa to 67 (3) kPa with 5 mL to 20 mL of fluid, respectively. In the ex vivo model, silicone catheters and the larger latex catheters also showed relatively stable balloon pressures with increasing fill volumes within the model bladder (Table 1). The mean (SEM) balloon pressure within the ‘urethra’ was 177 (6) kPa, about three-times higher, than in the mock bladder model across catheter types with 10 mL of fill. At balloon fill volumes of 7 mL, within the ‘urethra’, the neck of the balloon filling port expanded preferentially (Fig. 1). Furthermore, balloon pressure did not increase in a linear fashion with filling volumes >7 mL because the neck of the balloon port shunted the excess pressure. Using the 16 F latex catheter, maximal pressure was achieved when the balloon was filled with 5 mL [mean (SEM) 182 (21) kPa]; when the balloon 2 0 11 T H E A U T H O R S BJU INTERNATIONAL © 2 0 11 B J U I N T E R N A T I O N A L 1111 WU ET AL. TABLE 1 Balloon pressure in the ex vivo model at various states of filling and using different catheter types. The percentage change in pressure from the bladder to the urethra model is shown in the rightmost column Balloon filling volume, mL 0 5 7 10* 20 0 5 7 10* 20 0 5 10 35* 55 16 F latex 5 mL balloon 16 F silicone 5 mL balloon 22 F latex 30 mL balloon Mean (SEM) pressure in bladder model, kPa 0 (0) 63 (2) 62 (2) 63 (3) 67 (3) 0 (0) 47 (9) 48 (9) 50 (10) 53 (6) 0 (0) 73 (7) 68 (3) 60 (0) 57 (3) Mean (SEM) pressure in urethra model, kPa 0 (0) 182 (21) 183 (17) 173 (15) 167 (9) 0 (0) 207 (30) 210 (24) Balloon burst Balloon burst 0 (0) 212 (9) 177 (12) 137 (9) 137 (9) % Change 0 287 297 274 250 0 443 435 NA NA 0 289 265 228 241 *Manufacturer-recommended balloon filling volume; NA, not applicable; 1 kPa = 0.0099 atm = 0.15 psi = 7.5 mmHg. TABLE 2 Balloon pressures upon filling a 5 mL balloon of a latex catheter in different locations (prostatic, bulbar, and penile urethra) of the cadaver model. The balloon was filled with either 5 or 10 mL sterile water. Pressure was measured both immediately after filling and after 10 s, to allow system equilibration. The percentage change in pressure from the bladder after 10 s is shown in the third and rightmost column Bladder Prostatic urethra Bulbar urethra Penile urethra 5 mL Mean (SEM) immediate pressure, kPa 95 (3) 130 (38) 147 (15) 165 (9) Mean (SEM) pressure after 10 s, kPa 65 (8) 110 (21) 120 (25) 132 (2) % change from bladder after 10 s NA 169 185 204 10 mL Mean (SEM) immediate pressure, kPa 97 (3) 178 (38) 180 (15) 183 (9) Mean (SEM) pressure after 10 s, kPa 68 (4) 112 (19) 140 (21) 150 (16) % change from bladder after 10 s NA 164 205 220 NA, not applicable. was filled with 20 mL of fluid, the mean maximal pressure was only 167 (9) kPa. Balloon filling with greater volumes than the threshold pressure level resulted in continued balloon port expansion, not increasing intra-balloon pressure. Similar results were found with 22 F latex catheters (30 mL balloon), although the threshold fill volume at which we noted balloon port expansion was 10 mL. Silicone catheters did not perform in a similar fashion: as the balloon was filled within the urethra, the balloon port did not expand to serve as a ‘pop-off’ valve. Rather, when the balloon was filled with >7–10 mL of water, the balloon ruptured within the model ‘urethra’ (Table 1). 111 2 Identical experiments in the male cadavers yielded similar results. Balloon pressures within the urethra were nearly twice as high as within the bladder (Table 2), and, in all measurements (within the bladder and in various locations within the urethra), balloon pressure was always highest immediately upon filling, and slightly lower 10 s after filling, presumably due to relaxation of the cadaveric tissue and balloon port neck expansion. The mean balloon filling pressure was highest within the penile urethra (≈2.2-times that within the bladder), followed by the bulbar (≈2.0-times that within the bladder) and prostatic (≈1.6-times that within the bladder) urethra with 10 mL (Table 2). When the balloon was filled within the fossa navicularis, filling with only 3 mL of fluid resulted in a balloon pressure increase of 145 kPa, which is comparable with that achieved with three-fold greater filling volume with the balloon in the bulbar urethra, 140 kPa. Expansion of the balloon port was seen in all trials using a latex catheter with balloon filling within the urethra. FORCE REQUIRED TO WITHDRAW A CATHETER FROM THE BLADDER Experiments using both ex vivo models and fresh male cadavers confirmed that at higher balloon fill volumes, greater manual © BJU INTERNATIONAL © 2 0 11 T H E A U T H O R S 2 0 11 B J U I N T E R N A T I O N A L URETHRAL CATHETER BALLOON PRESSURE AND FORCE TABLE 3 Traction force required to dislodge catheter from ex vivo ‘bladder’ and cadaver. Data are shown for various degrees of balloon filling volume, and, using catheters made of different materials (silicone, latex) and of different diameters (16 F and 22 F) 16 F latex catheter with 5 mL balloon 16 F silicone catheter with 5 mL balloon 22 F latex catheter with 30 mL balloon 16 F latex catheter with 5 mL balloon in cadaver model Balloon filling volume, mL 0 5 7 10* 20 0 5 7 10* 20 0 5 10 35* 55 0 5 7 10* Mean (SEM) force, N 0 9.34 (0.44) 25.80 (4.45) 41.37 (8.01) >89† 0 4.45 15.57 (1.78) 33.36 (1.78) >44.5‡ 0 6.67 (1.33) 16.46 (1.78) >89† >89† 0 20.02 (2.22) 24.46 (2.22) 45.82 (4.00) *Manufacturer-recommended balloon filling volume; †Exceeded experimental limit, catheter could not be removed; ‡Balloon burst at this force. traction force is required to withdraw the catheter from the bladder (Table 3). In the ex vivo ‘bladder/urethra’ model removal of a 16 Fr latex catheter required an average of 9.34 (0.44) N and 41.37 (8.01) N traction force when the balloon was filled with 5 mL and 10 mL fluid volume, respectively. Over-filling of the 5 mL balloon with 20 mL fluid required significantly greater (>89 N) force, exceeding the measurement threshold of our measurement apparatus. Similar results were noted using a 22 F latex catheter (30 mL balloon) filled with >35 mL fluid as a manual traction force >89 N was required to withdraw the filled balloon per urethra. Silicone catheters also required greater manual traction force to withdraw the balloon per urethra at higher filling volumes. At balloon volumes >20 mL water, the silicone balloon invariably burst with a force >44.5 N. Incorrect insertion or removal can result in significant patient harm. While incorrect insertion or partial removal can be readily identified [11], the trauma at this point has already occurred. Unfortunately, incidents of catheter balloon filling within the urethra and accidental catheter removal of an indwelling catheter with the balloon filled continue to be reported in the patient care setting. Although the exact incidence of these problems has not been well studied, two key factors suggest the incidence of these safety-related issues will continue: i. Catheters are used by such a diverse group of health professionals that rigorous training is a practical challenge; ii. Few modifications to catheter design have occurred within the last 20 years, and such efforts have been primarily limited to mechanisms to reduce catheter-related UTIs [12]. DISCUSSION Surprisingly, no attention has been placed on how to make catheters safer to insert and use. We sought to define the physical parameters whereby iatrogenic injuries from urethral catheters occur to help guide design of a safer device. We found that Indwelling urethral catheters are commonplace in healthcare settings, and they are inserted into patients by health professionals at all levels of training. © balloon pressures, when the catheter balloon was filled appropriately within the bladder, varied little and were less than half that seen when filled in the urethra in both the ex vivo and cadaveric model. Interestingly, we discovered that a threshold balloon pressure exists for latex, but not silicone catheters, at which point the neck of the balloon port served as an unrecognised pressure-release valve mechanism (albeit set at a level that does not prevent patient harm). The less compliant silicone catheters do not demonstrate this pressure release mechanism, as they burst beyond the threshold balloon pressure level. Finally, we found balloon pressures vary in different portions of the urethra with higher balloon pressures as the balloon was filled more distally. For instance, at the fossa navicularis, balloon pressures were similar to those found within the prostatic urethra with much lower balloon filling volumes (3 vs 10 mL). This may help explain the frequent expression of patient discomfort during retrograde urethrography where the urethral catheter is filled in the fossa navicularis. The second portion of the present study showed in both models that increasing force was required to dislodge a urethral catheter from the bladder into the urethra as the balloon fill volume increased. This trend was noted in all catheter sizes and types tested. Interestingly, catheters composed of silicone were the only type to burst their balloon upon forceful removal from the bladder, which reflects the less compliant properties of silicone compared with latex. It is imperative that healthcare professionals performing urethral catheterization receive appropriate training [2]. A recent study of a training intervention on urethral catheter placement at a single institution significantly reduced the incidence of male urethral trauma [5]. Yet, despite this intervention iatrogenic injuries still occurred. Designing a safer urethral catheter that facilitates healthcare professionals in the placement of the catheter within the bladder would reduce unnecessary patient harm and warrants further exploration. We hypothesize that a catheter that has a thinner wall at the neck of the balloon port, serving as a designed ‘pop-off’ pressure valve, might provide manufacturers the unique opportunity to inexpensively add a catheter safety mechanism. This mechanism 2 0 11 T H E A U T H O R S BJU INTERNATIONAL © 2 0 11 B J U I N T E R N A T I O N A L 111 3 WU ET AL. would serve to notify the health professional placing the catheter of the incorrect location of the catheter balloon within the urethra through visual and tactile feedback by expansion of the balloon port. This ‘second balloon’ would also serve to prevent harmful high intra-urethral pressures thereby avoiding urethral rupture, bleeding, pain, and increased duration of hospitalization. The present study also addresses the common occurrence of accidental dislodgement of a catheter with a filled balloon from the bladder. An estimated one in twenty ICU patients suffers this event during their hospitalization [6]. The present study raises the question of whether or not to over-fill the catheter balloon to make it more difficult for a patient at higher risk for urethral self-injury to apply sufficient force to dislodge their catheter. However, one caveat is that a larger balloon may cause more urethral damage if a patient was still able to apply sufficient extraction force to remove their catheter. Additionally, the present findings suggest that given the propensity of silicone balloons to burst at high pressures or forces, these catheters might not be ideal for patients with altered mental status who are at higher risk of pulling out their catheters. Notwithstanding, intravesical bursting of the catheter balloon may lead to retained catheter fragments with bleeding and trauma from the burst itself necessitating cystoscopic intervention [13]. due to the wider and shortened female urethra. The number of repetitions performed in ex vivo experimentation may be considered a limitation, but we feel that the use of highly standardized, high-grade medical catheters negates the need for additional testing. Finally, balloon pressures serve as a surrogate measure of urethral trauma. Certainly, as evidenced by the non-zero filling pressures within the bladder, balloon pressures are related to both the baseline compliance of the catheter balloon material and the amount of pressure exerted on and from surrounding tissue. Filling of urethral catheter retention balloons within the urethra results in two or more times greater balloon pressures than when filled within the bladder. Catheters with larger balloon filling volumes require greater force to remove from the bladder when filled completely. The differences in balloon pressure found with filling within the urethra vs the bladder, and the force required to remove a catheter with a filled balloon from the bladder, could potentially be exploited to design a safer urethral catheter thereby preventing unnecessary patient harm. 111 4 5 6 7 8 9 10 11 CONFLICT OF INTEREST None declared. REFERENCES 1 Study limitations include the inability to test all catheter types and sizes in the cadaveric model, and the lack of in vivo data, which would not be ethical. However, ex vivo studies provided us with inferential data as they trended with our findings in the fresh cadavers. We acknowledge that actual pressure and force measurements will change in a living human with perfused tissue, but feel that the data trends and catheter deformation findings would remain constant. We did not design the present study for the female population because catheter related injuries are far less common 4 2 3 Haley RW, Hooton TM, Culver DH et al. Nosocomial infections in U.S. hospitals, 1975–1976: estimated frequency by selected characteristics of patients. Am J Med 1981; 70: 947–59 Thomas AZ, Giri SK, Meagher D, Creagh T. Avoidable iatrogenic complications of urethral catheterization and inadequate intern training in a tertiary-care teaching hospital. BJU Int 2009; 104: 1109–12 Saint S, Lipsky BA. Preventing catheter-related bacteriuria: should we? Can we? How? Arch Intern Med 1999; 159: 800–8 12 13 Wald HL, Ma A, Bratzler DW, Kramer AM. Indwelling urinary catheter use in the postoperative period: analysis of the national surgical infection prevention project data. Arch Surg 2008; 143: 551–7 Kashefi C, Messer K, Barden R, Sexton C, Parsons JK. Incidence and prevention of iatrogenic urethral injuries. J Urol 2008; 179: 2254–7; discussion 2257– 8 Lorente L, Huidobro MS, Martin MM, Jimenez A, Mora ML. Accidental catheter removal in critically ill patients: a prospective and observational study. Crit Care 2004; 8: R229–33 Sellett T. Iatrogenic urethral injury due to preinflation of a Foley catheter. JAMA 1971; 217: 1548–9 Buddha S. Complication of urethral catheterisation. Lancet 2005; 365: 909 Canales BK, Weiland D, Reardon S, Monga M. Urethral catheter insertion forces: a comparison of experience and training. Int Braz J Urol 2009; 35: 84– 9 Thomsen TW, Setnik GS. Videos in clinical medicine. Male urethral catheterization. N Engl J Med 2006; 354: e22 Vaidyanathan S, Hughes PL, Oo T, Soni BM. Long catheter sign: a reliable bedside sign of incorrect positioning of foley catheter in male spinal cord injury patients. Cases J 2008; 1: 43, doi:10.1186/1757-1626-1-43 Schumm K, Lam TB. Types of urethral catheters for management of short-term voiding problems in hospitalized adults: a short version Cochrane review. Neurourol Urodyn 2008; 27: 738–46 Chrisp JM, Nacey JN. Foley catheter balloon puncture and the risk of free fragment formation. Br J Urol 1990; 66: 500–2 Correspondence: Alex K. Wu, Department of Urology, University of California, San Francisco, San Francisco, CA 94143-0320, USA. e-mail: wuak@urology.ucsf.edu Abbreviation: ICU, intensive care unit. © BJU INTERNATIONAL © 2 0 11 T H E A U T H O R S 2 0 11 B J U I N T E R N A T I O N A L