Between Run Precision (Per Shift CV Comparison)
HOSPITAL
DATE
UERMMMC
15-Apr-20
INSTRUMENT
SYSMEX XN-1000
Kindly input the values on the yellow boxes below.
WBC
RBC
HGB
HCT
MCV
MCH
MCHC
PLT
Shift 1
6.09
5.17
15.2
45.8
88.6
29.4
33.2
283
Shift 2
6.03
5.11
15.1
45.3
88.6
29.5
33.3
286
Shift 3
6.06
5.12
15.1
46.7
91.2
29.5
32.3
288
Mean
6.06
5.14
15.2
45.6
88.6
29.5
33.3
284.5
SD
0.02
0.03
0.05
0.61
1.30
0.05
0.48
2.14
CV
0.40
0.51
0.32
1.34
1.47
0.16
1.43
0.75
Acceptable
performance (%)*
<10.0%
<5.0%
<5.0%
< 8.0%
< 8.0%
<2.0%
<2.0%
<10.0%
Performance
(Pass/Fail)
PASS
PASS
PASS
PASS
PASS
PASS
PASS
PASS
* Obtained from XN-1000 Instruction For Use: Chapter 15 - Technical Information
Performed By:
Verified By:
Franz Ehrlick U. Manansala
Kerwin C. Faustino