Optical Microscopy Lab Report Eshita Gurung Micro and Nano System Technology University of South-Eastern Norway Horten, Norway 230738@usn.no Sharaban Tahora Micro and Nano System Technology University of South-Eastern Norway Horten, Norway 230734@usn.no Neda Ghiasi Micro and Nano System Technology University of South-Eastern Norway Horten, Norway 230742@usn.no Abit Babu Micro and Nano System Technology University of South-Eastern Norway Horten, Norway 230736@usn.no Samarbir Sing Micro and Nano System Technology University of South-Eastern Norway Horten, Norway 230739@usn.no Abstract: This essay reports the measurement of several dimensions of a RF MEMS-switch using a Leica microscope. Bright field, dark field mode, depth of focus, depth of field and their relationship with resolution and numerical aperture are discussed. I. INTRODUCTION With the advent of technology, there has been an enormous growth in the application of optical microscopy for micron and submicron level investigations in a wide variety of disciplines. Early microscopes were hindered by optical aberration, blurred images, and poor lens design but advances in digital imaging and analysis have also enabled microscopes to acquire quantitative measurements quickly and efficiently. II. IV. RESULT & DISCUSSION The smallest magnification of objective lens to view the whole structure is 2.5x. Hence, total magnification will be 25x due to 10x magnification of the microscope lens and the field of view is equal to 4574 x 3382 μm. The microscopic image of the test pattern is displayed in the below figure 1. MATERIALS REQUIRED Leica Optical Microscope Sample pattern – RF switch PC with Leica application suite III. METHODES The sample pattern (RF Switch) was kept under the Leica microscope and was adjusted in the correct position and magnification. By adjusting the focus, a clear image of the sample was obtained. After that, the sample pattern was viewed in the ocular and in the Leica application suite in the bright mode initially. In this mode sample roughness could be reviewed. The lateral measurement/dimensions, height differences, areas and angles of the sample pattern were measured and the values were recorded. In order to go over the details of edges, the sample was reviewed in dark filed mode also. Figure 1: Microscopic image of the test patter (RF switch) Figure 2. Transversal cut of the MEMS structure portraying the elements to be measured [2] The values of the L1, L2, L3, V1, W1, W2, Z1 and Z2 which were measured with the help of Leica software are tabulated below. Figure 2b: Measurement of the length (L2) of the bridges in the Microsystem Table 1: Measurement values of the different cross section of the RF switch Sl.No 1 2 3 4 5 6 7 8 Structure L1 L2 L3 V1 W1 W2 Z1 Z2 Measurement (μm) 284.254 μm 2482.498 μm 74.757 μm 46.469 77.701 μm 194.754 μm 1 μm 1 μm In figure 2a and figure 2b measurement of length, withs and angle of the bridges are illustrated. Figure 2a: Measurement of the lengths (L1 & L3), widths (W1 & W2) and angle(V1) of the bridges in the microsystem Figure 2c: Measurement of Z1 and Z2 i.e. height difference between the layers recorded with the depth of focus. Dark Field Mode: It provides a higher contrast due to the scattered light in order to observe defects and edges of the test pattern. It also provides a sense of depth difference not clearly visualized in bright field mode due to the darker tone in the background at least for the magnification observed in figure 3a. Dark filed mode imaging complements the optical microscopy observations when contamination and edge observation is relevant.[3] In comparison to the bright mode, dark field mode is highly suitable to identify the edges, outlines, topography and the surface defects on the sample. In figure 3a sample defects are shown clearly. Damaged parts are viewed as bright points or disturbed edges. Relationship between the depth of focus and the numerical aperture (NA) Depth of focus is inversely proportional to numerical aperture i.e Dfocus= Relationship between the resolution and NA From the equation above, it is evident that “R” (Resolution) is inversely proportional to the numerical aperture, hence smaller features can be resolved with a higher numerical aperture value. Figure 3a: Microscopy image with dark field mode V. CONCLUSION This experiment can be concluded with the following understanding that Leica optical microscope is an efficient device to take the measurements of various samples but the accuracy of the dimensions would be more accurate with the automated process than to the manual process as human error are prone to occur. The depth of Field and depth of Focus for magnification of 100x and 2.5x lenses, relationship between the depth of focus and the numerical aperture and relationship between the resolution and NA were discussed. Figure 3b: Microscopy image with bright field mode References [1] Depth of Field and Depth of Focus for 100x lens and 2.5x lens As lower NA(Numerical aperture) increases the depth of focus and depth of field hence both depth of field and focus is higher for 2.5x lens in comparison to 100x lens as it has lower numerical aperture i.e. NA = 0.07 and NA= 0.09 for 100x lens.[4] where λ is the “Wavelength” and NA is the “Numerical aperture”. s=0.61 [2] [3] [4] ScienceDirect."OpticalMicroscope." https://www.sciencedirect.com/topics/engineering/opticalmicroscope Measurement course lab shit https://www.ruf.rice.edu/~bioslabs/methods/microscopy/dfield.htm l Microstructural Characterization of Materials - 2nd Edition David Brandon and Wayne D. Kaplan © 2008 John Wiley & Sons,Ltd. ISBN: 978-0-470-02784-4