JACK SKERRITT - STUDENT Chromatography labs Experiment : Introduction to GC. Use if an external standard. Examination of a mixture of cyclohexane, cumene, cymene, ethyl benzene and toluene. Measurement of retention time, adjusted retention time, N values, H(hetp), resolution and the % composition of an unknown mixture of these compounds Date 10/02/20 Lab partner: Abdul Momoh Aims To measure the retention times of the compounds and find the composition of the unknown mixture of these compounds. Getting a knowledge of the GC machine and the different operating temperatures available for use. Procedure As Per manual pg 5-9 Results Table 1 Retention times and adjusted Retention times for all compounds at 95 celsius tr Compund tš mm min 25 1.25 Cyclohexane 198 9.9 Cymene 81 4.05 Cumene 58 2.9 Ethylbenzene 35 1.75 Toluene Calculation of min retention times mm t'r= tr - tš min 0.35 0.35 0.35 0.35 0.35 7 7 7 7 7 mm min 18 191 74 51 28 0.9 9.55 3.7 2.55 1.4 Chart speed = 20mm/min Therefore for cyclohexane 25mm / 20mm/min= 1.25min Table 2 : External standard at 95 degrees and the associated identities tr mm Peak 1 Peak 2 Peak 3 Peak 4 Peak 5 tš min 21 33 54.3 77 143 mm 1.05 1.65 2.665 3.85 7.15 t'r= tr - tš min 7 7 7 7 7 mm 0.35 0.35 0.35 0.35 0.35 1 Identity min 14 26 47.3 70 136 0.7 1.3 2.315 3.5 6.8 Cyclohexane Toluene Ethylbenzene Cumene Cymene JACK SKERRITT - STUDENT Table 3: calculation N and hetp from Program A N N1 N2 70.56 10 28.34 W hetp N3 144 11 13.89 N4 242.47 14 8.25 N5 561.33 13 3.56 Mean of N 484 300.472 26 14.8 4.13 11.634 Calculation of N If we take Peak 1 as an example Tr=21 W= 10 ((4x21)/10)2 = 70.56 Calculation of hetp Where l= the length of the column = 2000mm Taking peak 1 again N = 70.57 L = 2000 2000/70.56 = 28.34 Table 4 : resolution of the compounds at higher temperatures and the temperature program Resolution 95 1. /. 2. 15.60 2. /. 3. 17.63 3. /. 4. 14.46 4. /. 5. 22.31 Calculation of the resolution 105 17.83 17.31 27.00 35.76 115 12.83 13.22 19.18 25.65 T for peak 1 was 21 125 tempprogram 6.00 10.57 8.67 16.67 6.44 13.50 20.22 17.90 (2(33-21))/(10+11) = 15.6 T for peak 2 was 33 W1 = 10 W2 = 11 2 JACK SKERRITT - STUDENT Table 5 : Determination of the % composition in the unknown mixture Peak Area 1 Peak Area 2 Peak Area 3 Peak Area 4 Peak Area 5 External Standard Response factor RF = Area x / Area 1 391961.594 439687.969 433141.549 1 1.12176289 412504.75 389890.813 1.1050612 1.05241115 0.99471688 Unknown Mix 68921.945 119188.523 235054.906 153088.578 Adjusted Area = Area x / RF 68921.945 106251.084 212707.591 145464.611 152029.189 685374.421 151226 Total Composition %v/v 10.0561011 15.5026335 31.0352393 21.2241086 22.1819175 Discussion The differences observed is down to the compounds structure. A stronger structure will lead to a high boiling point which means at the higher temperatures it elutes a lot quicker than the lower boiling points. You can see the retention times getting consistently lower with each increase in temp. The best program to use from looking at the chromatograms is the Temperature program. This gives good separation and a quicker run time for each of the compounds. This is due to each compound being able to elute at its preferred temperature instead of the lower or higher ones. If you look at the lowest temp you can see that the peaks for the last compound are very far down the chromatogram and are not very sharp. For the next temp up we see the resolution is good bu the peaks are getting closer together anf the last compound is still very far down the graph. At 115 we can see the last peaks beginning to sharpen up bu the first peaks getting very close together. At 125 we can see the final peak is closer and still separate but the other peaks are overlapping. Cyclohexane Toluene Ethyl benzene cumene cymene C6H12 C7H8 C8H10 C4H2 C10H14 110 Celsius 136 C 152C 177C 92.14 106.17 120.19 134.21 Boiling points 80 Celsius MM g/mol 84.16 3 100