Uploaded by Abdullah Aljbour

XR Basic Service Photo-Tour

advertisement
CLOSE
Select your system from the drop down box below
then click on the GO button
Select System
GO
Navigation
Click on the HOME
icon, located top right
of each page, to
.
return to this page.
Click on the BACK
icon to return to the
product home page.
Click on the
INFORMATION
icon for an overview
of the product.
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Click on the yellow
boxes for more
information
blication
67
Table
Control Area
PRE INSTALLATION MANUAL
FLEXI-DT
System Cabinet
DIGITAL TABLE
Detector
Support
Assembly
(DSA)
Flexi DT Digital Table
Patient Barrier
OTS
Wall Stand
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Jedi Generator:
Generates the
power used by
the x-ray tube
Corona 2 Board
controls the system power
on, off and reset functions
and also serves as a CAN
Hub for the system’s CAN
buses.
Power Distribution
Unit (PDU):
Distributes 120Vac &
230Vac around the
system
Bulkhead
Halo Board
Contains relays which are
used to switch power to the
system.
The relays are controlled by
the Corona 2 Board.
On/Off Switch
System Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Control
Monitor
Patient List/
Image Monitor
Exposure
Switch
Keyboard
Radiographic
Control Interface
Module (RCIM)
Click for details
Mouse
System and
Image Processing
Computer (aka
Magic PC)
Wall Plate
(located under
counter)
CD-RW/DVD
ROM Drive
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Motorised Movements
Vertical
Longitudinal (along overhead
rails)
Lateral (along overhead rails)
Column rotation (rotates at top
of column not at the goose
neck)
Xray Tube Angulation
Carriage
Column
Rotation
Telescopic
Column
Vertical
Tube
Angulation
X-ray Tube
User Interface
Console
Click for details
Collimator
Click for details
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Wallstand
Bucky tilts to
horizontal
Detector Angulation
and SID Display
Vertical Lock
Release
Wall Bucky
Horizontal
Lock
Release
Wall Stand
Handheld Control
Click for details
Detector Support
Assembly (DSA)
contains power supply
and conditioner (chiller)
for wall stand detector
Lon
Wall Stand (Standard
version)
ger
Arm
Extended Arm Version
of Wall Stand
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Movements Enable
Button
Lit= movements disabled
OFF= movements enabled
Floating Table-top
Grid
Detector
handle
Emergency Movements
Off Button
When pressed disables all
room movements
Table Up/Down
Table
Down
Table
Up
Table-top Locks Release
(longitudinal & lateralfloating table-top mode)
X-ray Photo Tour- Rev.2 by E.Singleton
CH
CLOSE
3. Put the template over the tabletop. It has to fit between the two metal frame guides.
Then place the table between the detector and the suspension, as in Figure 13.
5.1. SAFETY
The Flexi DT Digital Table
This portable table is designed to be used
with the OTS and wallstand.
The table has motorised movements and can
be powered from a 120/230Vac outlet or from
its internal 24Vdc batteries.
The table is temporarily secured to the floor
with pins, which are interlocked to prevent FLEXI-DT. DIGITAL TABLE
SERVICE MANUAL
exposures if not in place.
REVISION 5
P/N S0015068
The table also has a collision detection
device to prevent collisions with the detector.
2
Lateral
Drive
The FLEXI-DT
the safety locks.
5.1.1. EMERGE
Flexi DT Table with Wall Stand
Figure 13
9
3
Figure 12
Longitudinal
Drive
8
Figure 1
7
1. Elevating Base
2. Table-top
4
11 of the table (with the template on it) must be aligned with the collimator
4. The centre
3. Control Panel
cross (Figure 14) as with the detector, move the table until getting the collimator
4. Emergency Stop
cross over the center align cross of the template, as in Figure 15.
5. Wheels
6
Electronics
6. Wheel Brakes
1
7. Pins
8. Pin Lodging
9. Safety Locks
10. Pedals
5
10
11. IR Device
Flexi DT Table
Collimator cross
Figure 53
Control Panel
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Patient barrier is placed in front
of the wall-stand.
Its purpose is to prevent the
patient from coming in contact
with the moving wall-stand
detector during certain studies.
Patient stands
on this platform
Barrier attaches to the
floor with steel pins
Patient Barrier
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
LEDs
Longitudinal
Lock Release
Ready
Image
Pasting
Prep
Exposure
Inhibit
Manual
Collimation
Lateral
Detent
SID
Detent
Tube
Angulation
Lock Release
Lateral Lock
Release
Touch-screen
Display
Interlock
Override
Button
Detent Mode
Active LED
Vertical Lock
Release
Column
Rotation Lock
Release
All Transitional
Locks Released
LED
Collimator
Field Light
Detent
Select
OTS User Interface Console
All Locks
Release
Button
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Click on the yellow
boxes for more
information
Image Head
Remote
Touch
Panel
(RTP)
Monitor
Supports
Generator
Positioner
X-ray Tube,
Collimator &
DAP
Collimator
Control Cabinet
Control Area
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Upper Assembly
Area includes CPU
board, Room
Interface board
and Photo-timing
AEC board
Inverter Control
Board
Filament &
Power Supply
boards
Inverter
Assembly
Rotor Driver
Board
High Voltages
(>600Vdc & 480Vac)
are present in this area
Power Input and
Resonant
Inverter Area
High Voltage
Tank
Generator Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Dose Area Product
(DAP) Chamber
Ralco Collimator
with:
Iris
3 Spectral Filters
2 Vascular Contour
Filters
1 Ruler Filter
X-ray Tube & Collimator
Varian G1592 (DSA) or
G1092 (RF) X-ray Tube
X-ray Tube Oil to Air
Cooler (only used with
G1592 x-ray tubes)
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Optics
Assembly
CCD
Camera
Image Intensifier
Power Supply
Image Intensifier
Image Head
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Digital
Monitor
Generator Touchscreen monitor
Radiographic
Exposure Hand
switch
Digital Keyboard
and Mouse
Positioner
Console
Click for details
SVGA Splitter (used
to drive the control
room and in-room
digital monitors)
Fluoro
Footswitch
Isolation
Transformer (used
for PC and monitor
power)
Radiographic
Exposure
Footswitch
Control Area
Digital Image
Processing
Computer
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
In-room Monitor Ceiling
Suspension for two Flat Panel
Monitors
In-room Monitor Ceiling
Suspension for a single Flat
Panel Monitor
In-room Monitor Cart
for Flat Panel Monitor
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Gantry (contains
positioner
electronics)
Positioner Monitor
(displays positioner
movement values
and interlock
information)
Table-top
C-Arc
Positioner
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Touch-screen Monitor
(basic Generator and Digital
functions can be performed
from here)
Positioner
Console
Click for details
Radiographic
Exposure
Footswitch
Fluoro
Footswitch
Remote Touch
Panel (RTP)
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Click on a
selection box
below to see
switch functions
Home &
Preset
Positions
AP/PA ,
Emergency
Off &
Collision
Override
Collimator
& II
Movements
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Click on the yellow
boxes for more
information
Image Head
Intelligent
Digital Device
(IDD)
System
Cabinet
Positioner
Cabinet
Wallstand
Positioner
Overhead Tube
Suspension
Monitor
Supports
Control Area
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Digital System (Saturn):
Responsible for Image Processing
and Storage (hard disk can store up
to 8,000 images).
Windows 95 Operating System
Athena is the applications software
Accessory Rack:
Contains miscellaneous
circuitry used for controlling
the power state of the
system.
High Voltage (>600Vdc)
is present in the
generator even before
the system is turned
on at the console
Jedi Generator: Generates the
power (up to 85kW) used by the
x-ray tube
Power Distribution Unit (PDU):
Distributes 120Vac & 230Vac
around the system
UPS: Provides power to the PCs
(Digital and Magic) during a
hospital power failure
Cable Bulkhead
A25
ON/OFF Switch
System Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
A7
Table Angulation
Panel (Table Tilt Drive)
A22
Longitudinal IDD
Drive CAN Amp
(Power Assist)
A26
Lateral Table-top
Drive Board
A23
Vertical IDD Drive
(SID- Power Assist)
CAN Amp
A21
Relay Panel
Contains relays for
longitudinal table-top
drive
A20
Positioner I/O
Board:
Converts analog signals
from the Positioner to CAN
messages and CAN
message from the
controller PC back into
analog signals
A28
Power Supplies (3)
(+24V) (+12V)
(+5V,+15V,-15V)
Positioner Cabinet
A29
Power
Bulkhead
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
In-room Monitor Ceiling
Suspension for two Flat Panel
Monitors
In-room Monitor Ceiling
Suspension for a single Flat
Panel Monitor
In-room Monitor Cart
for Flat Panel Monitor
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Movements Off
(disables power
to Table, OTS &
Wallstand)
Bucky Tray
Table Controls
Click for details
Patient Step
(blocks table tilt
when pulled out)
Positioner
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Grid (behind the
cone): can be driven
in or out of the field.
Used to improve
image contrast by
attenuating the off
focal radiation
Image
Intensifier
behind cover
IDD Console
Click for
details
IDD in Field
Latch
CCD Camera and
Optics Assembly
behind this cover
Cone: can be driven
in or out of the field.
Used to compress
anatomy
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Integrated User
Interface (IUI)
Touch screen
(Magic PC Display)
RCIM
Click for details
In-room wall
plate for
monitor
connections
Digital (Saturn)
Monitor
Video Switch: switches
video input to monitor
between DVD and Digital PC
System PC (Magic
PC): Functions:
Control of the
generator
Control of
Positioner
movements
using the CAN
Bus
User Interface
OS: Linux
IUI Keyboard
used for Service mode
and for entry of Protocol
name
Exposure
Switch
DVD Recorder
Control Pad
DVD Recorder
Digital
Keyboard and
Mouse
Accessory Box
containing Daxten K/B
Mouse Extender Remote
version 1034-121PL
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Carriage
Assembly
Bridge
Lateral
Movement
X-ray Tube
(MX100)
(behind
cover)
Cable Rail
User
Interface
Click for
details
Collimator
Click for
details
Longitudinal
Movement
Rails
Overhead Tube
Suspension (OTS)
OTS Bridge and rails:
Allows lateral and longitudinal movement of the
Tube Suspension.
Potentiometers used for positional feedback.
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Patient Handles
Wallstands
The Precision 500D has the option of two
types of wallstand, Tilting and Non-tilting.
Early Precisions (approximately May 2002September 2004) had the option of:
SG-60 (non tilting)
SG-100 (tilting)
Later versions of the Precision have the
option of:
SG-80 (non tilting)
SG-120 (tilting)
Attached
to Wall
Differences
SG-60 and SG-100 are attached to the wall
and do not have a patient handle, whereas
SG-80 and SG-120 are floor only mounted
with patient handles.
The Bucky tray can be mounted on either
side (field changeable)
SG-100
SG-120
SG-100 Wallstand
with the bucky
tilted
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Left Screen
Click for
details
Vertical &
Horizontal Image
reverse buttons
Right Screen
Click for
details
Collimator Controls
Fluoro and Record
Exposure Switches
Table Angulation
Knob
Longitudinal tabletop drive Switch
Vertical and
Longitudinal Power
Assist handle
IDD Console
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Tube selected
indicator
Wallstand
receptor select
kV Readout
mAs Readout
Not Used
kV change buttons
Table bucky
receptor select
mAs change buttons
SID Readout
Table-top
receptor select
Tube Angle Readout
Longitudinal Lock
Lateral Lock
Angulation Lock
Detent
Lock or release the detent
magnetic lock. Normally
open self–lock type button.
Vertical Lock
All Locks
Indicators- Lit means:
Ready Exposure
Manual
Hold
Collimation
OTS User Interface
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Click on the yellow
boxes for more
information
Generator
Positioner
Positioner
Cabinet
OTS
Wall Stand
Monitor
Supports
Control Console
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
CPU board and Stepper
Motor control board
(ASSI) (behind the
Interface Board)
Group 25
Interface Board
Group 25
Stepper Motor
Controllers
Group 24
High Power DC Motor
Controls
Group 23
Power Input Rack
Group 22
Optional 277Vac to
230Vac Input Voltage
Transformer (22TR3)
Digital Image Chain
to Generator
Interface Module
(GIM)
Positioner Cabinet
Isolation
Transformer for
Digital PC
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Upper Assembly
Area includes CPU
board, Room
Interface board
and Photo-timing
AEC board
Inverter Control
Board
Filament &
Power Supply
boards
Inverter
Assembly
Rotor Driver
Board
High Voltages
(>600Vdc & 480Vac)
are present in this area
Power Input and
Resonant
Inverter Area
High Voltage
Tank
Generator Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
In-room Monitor Ceiling
Suspension for two Flat Panel
Monitors
In-room Monitor Ceiling
Suspension for a single Flat
Panel Monitor
In-room Monitor Cart
for Flat Panel Monitor
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Generator Touchscreen monitor
Digital
Monitor
Positioner
Console
Click for details
Digital Keyboard
and Roller ball
Digital Image
Processing
Computer
Fluoro Footswitch
Control Console
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Carriage
Telescopic Column
X-ray Tube
OTS Control Panel
Click for details
Collimator
Click for details
OTS
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Tube Stand
X-ray Tube
Table-top
Collimator
Click for
details
Spot Film
Device (SFD)
Cone
(Palpator)
Table Control
Panel
Click for details
Image Intensifier
and CCD Camera
behind this cover
Foot Board
Table Base
Positioner
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Patient
Handle
Bucky
Bucky Tray
Wall Stand
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Click on a
selection box
below to see
button functions
Movements
Display
Collimator &
II
Exposure &
Technique
selection
Tomo
Positioner Console
Nical
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Click on the yellow
boxes for more
information
Table
System
Cabinet
Overhead Tube
Suspension
Wall Stand
Control
Console
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Jedi Generator (behind
Interface Board)
System
Interface Board
Jedi Generator
On/Off Breaker
Power
Distribution
System Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Tomo
Attachment
Movements Enable
Button
Lit= movements disabled
OFF= movements enabled
Tomo Console
Click for details
Floating Table-top
Emergency Movements
Off Button
When pressed disables all
room movements
Bucky
Table-top Locks
Release (longitudinal &
lateral- floating tabletop mode)
Table Up/Down
Table
Table
Down
Table
Up
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Tomo-Link
Drive Unit
LED: On
when OTS is
in lateral
detent
Wallstand
detent
magnetic strip
Lateral
Movement
Longitudinal
Movement
Carriage
Assembly
X-ray Tube
(MX100)
(behind
cover)
Magnetic sensor
electronics behind
cover, used to decode
magnetic detents
User Interface
(OCS)
Click for
details
Collimator
Click for
details
Overhead Tube
Suspension (OTS)
Lateral tabletop detent
magnetic strip
OTS Bridge and rails:
Allows lateral and longitudinal movement of the
Tube Suspension.
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Bucky Tray
Wall Stand
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
OFF and ON
Buttons
Technique
change
buttons
Prep/Expose
Buttons
Prep/Expose
Handswitch
Console
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Tube selected
indicator
Wallstand
receptor select
kV Readout
mAs Readout
Tomo receptor
select
kV change buttons
Table bucky
receptor select
mAs change buttons
SID Readout
Table-top
receptor select
Tube Angle Readout
Longitudinal Lock
Lateral Lock
Angulation Lock
Detent
Lock or release the detent
magnetic lock. Normally
open self–lock type button.
Vertical Lock
All Locks
Indicators- Lit means:
Ready Exposure
Manual
Hold
Collimation
OCS User Interface
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Click on the yellow
boxes for more
information
Control
Console
System
Cabinet
U-Arm
Positioner
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Tube Up/Down
(SID)
User Interface
Click for details
X-ray Tube
Collimator
Click for details
U-Arm
U-Arm pivot point
Grid
Detector Assembly
pivot point
Digital
Detector
U-Arm Control
Panel
Click for details
U-Arm Positioner
Emergency Movements
Off Button
When pressed disables all
movements
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Sedecal
Generator
U-Arm Motor
Controllers
Power
Distribution
System Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
System and Image
Processing
Computer (aka
Magic PC)
Control and
Image Monitor
Mouse
Keyboard
Exposure
Switch
Control Remote
Interface Board
(CRIB)
Click for details
Remote Control
Click for details
Control
Console
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Click on the yellow
boxes for more
information
Table
Control Area
System Cabinet
Detector
Support
Assembly
(DSA)
OTS
Wall Stand
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Image Detection
Controller (IDC) 1
Table
Corona Board
controls the system power
on, off and reset functions
.
IDC 2
Wall Stand
Jedi Generator:
Generates the
power used by
the x-ray tube
Bulkhead
Power Distribution
Unit (PDU):
Distributes 120Vac &
230Vac around the
system
Halo Board
Contains relays which are
used to switch power to the
system.
The relays are controlled by
the Corona Board.
On/Off Switch
System Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Control
Monitor
Patient List/
Image Monitor
Exposure
Switch
Keyboard
Radiographic
Control Interface
Module (RCIM)
Click for details
Mouse
System and
Image Processing
Computer (aka
Magic PC)
Wall Plate
(located under
counter)
CD-RW/DVD
ROM Drive
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Carriage
Assembly
Telescopic
Column
Bridge
Motorized SID
movement
X-ray Tube
(MX100)
(behind
cover)
Longitudinal
Movement
Cable Rail
User Interface
Click for
details
Collimator
Click for
details
Lateral
Movement
Rails
Overhead Tube
Suspension (OTS)
OTS Bridge and rails:
Allows lateral and longitudinal movement of the
Tube Suspension.
Potentiometers used for positional feedback.
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Wallstand
Bucky tilts to
horizontal
Detector Angulation
and SID Display
Wall Bucky
Horizontal
Lock
Release
Wall Stand
Handheld Control
Click for details
Detector Support
Assembly (DSA)
contains power supply
and conditioner (chiller)
for wall stand detector
Vertical Lock
Release
Wall Stand
Rear view of Wall Stand
Bucky
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Movements Enable
Button
Lit= movements disabled
OFF= movements enabled
Floating Table-top
Grid
Detector
handle
Emergency Movements
Off Button
When pressed disables all
room movements
Table Up/Down
Table
Down
Table
Up
Table-top Locks Release
(longitudinal & lateralfloating table-top mode)
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Tube selected
indicator
Wallstand
receptor select
kV Readout
mAs Readout
Not Used
kV change buttons
Table bucky
receptor select
mAs change buttons
SID Readout
Table-top
receptor select
Tube Angle Readout
Longitudinal Lock
Lateral Lock
Angulation Lock
Detent
Lock or release the detent
magnetic lock. Normally
open self–lock type button.
Vertical Lock
All Locks
Indicators- Lit means:
Ready Exposure
Manual
Hold
Collimation
OTS User Interface
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Click on the yellow
boxes for more
information
Image Head
Spot Film
Device (SFD)
System
Cabinet
Positioner
Cabinet
Positioner
Overhead Tube
Suspension
Image Processing
System
Control Area
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
TV Camera
Controller and Video
Processor (VIC)
System Controller
(Titan)
Exposure Control
Module (ECM)
SCPU
Generator
High Voltage
Transformer
System Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Interface Bulkhead
Table and Spot Film
Device Controller
Table Angulation
Controller
Positioner Power
Distribution
Positioner Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Movements Off
(disables power
to Table, OTS &
Wallstand)
Bucky Tray
Table Controls
Click for details
Positioner
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Optics
Assembly
Image
Intensifier
Power Supply
Camera
Image
Intensifier
Spot Film
Console
Click for details
Cassette Tray
Image Head
and SFD
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Integrated User
Interface (IUI)
Touch screen
Power On/Off
Exposure
Switch
Reset
X-ray ON
indicator
Selection
Wheel
Control Console
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Bridge
X-ray Tube
Carriage
Assembly
Longitudinal
Movement
Lateral
Movement
Cable Rail
Collimator
Click for
details
OTS Bridge and Rails
User
Controls
Click for
details
Overhead Tube
Suspension (OTS)
Rails
OTS Bridge and rails:
Allows lateral and longitudinal movement of the
Tube Suspension.
Potentiometers used for positional feedback.
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Image
Monitor
Image Archive
Device (Exabyte
Drive) (optional)
Control Console for
Image Processing
System
Click for details
Digital Image
Processing
Computer Cabinet
Click for details
DRS Gateway
DRS Gateway
The DRS Gateway is an
optional hardware solution to
provide added network
functionality for the DRS.
The DRS Gateway is a stand
alone PC, running Windows
2000. Its functionality is
transparent to the customer
and requires no interaction
with the customer.
Exam Room
Monitor
Image Processing
System
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Hard Disk for Patient
Image Storage
Digital Image
Processing
Computer (DRS)
Cable Storage
Power Supply
Digital Image Processing
System Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Displays number of
Images in Current
Study and remaining
image capacity of
hard disks
Delete Study:
Press twice to
delete selected
study
Delete
Image: Press
twice to delete
current image
Brightness
and Contrast
Adjustment
Specialty Keys:
Click HERE for
details
Patient Directory:
Select Patient
Directory screen on
DRS monitor
New Patient:
Select New Patient
screen on DRS
monitor
Keyboard: used
to enter patient
information
FluoroStore:
Saves the current
fluoro image to the
patient folder
Image Processing
Console (DRS)
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Left Screen
Click for
details
Fluoro Dose
Selection
Vertical &
Horizontal Image
reverse buttons
Right Screen
Click for
details
Collimator Controls
Fluoro and Record
Exposure Switches
Table Angulation
Knob
IDD Console
Longitudinal tabletop drive Switch
Vertical and
Longitudinal Power
Assist handle
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Click on the yellow
boxes for more
information
Image
Head
Spot Film
Device (SFD)
Positioner
Cabinet
System Cabinet
Generator
Cabinet
Positioner
Overhead Tube
Suspension
Image Processing
System
Control Area
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Photospot Controller
(controls Cut Film
Photospot camera)
Exposure &
System Control
(known as Kernel)
TV Camera
Controller and Video
Processor (VIC)
System Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Interface Bulkhead
Table and Spot Film
Device Controller
Table Angulation
Controller
Positioner Power
Distribution
Positioner Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Rotor Controller
(TIRC)
Generator Control
(kV & mA)
On/Off Switch
High Voltage
Transformer
Volt Pack (feeds
primary voltage to the
HV Transformer)
LFX Generator Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Generator Control
(kV & mA)
Rotor Controller
(TIRC)
Generator Power
Distribution
On/Off Switch
Inverter Modules
1 and 2
Input Auto
Transformer
High Voltage
Transformer
MPPU Generator Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Movements Off
(disables power
to Table, OTS &
Wallstand)
Bucky Tray
Table Controls
Click for details
Positioner
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Optics and Image
Gate Assembly
Camera
Cut Film
Camera
(105mm)
Image
Intensifier
Power Supply
Image
Intensifier
Spot Film
Console
Click for
details
Cassette Tray
Image Head and SFD
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Integrated User
Interface (IUI)
Touch screen
X-ray ON
indicator
Reset
Power On/Off
Exposure
Switch
Selection
Wheel
Control Console
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Bridge
X-ray Tube
Carriage
Assembly
Longitudinal
Movement
Lateral
Movement
Cable Rail
Collimator
Click for
details
OTS Bridge and Rails
User
Controls
Click for
details
Overhead Tube
Suspension (OTS)
Rails
OTS Bridge and rails:
Allows lateral and longitudinal movement of the
Tube Suspension.
Potentiometers used for positional feedback.
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Image
Monitor
Image Archive
Device (Exabyte
Drive) (optional)
Control Console for
Image Processing
System
Click for details
DRS Gateway
Digital Image
Processing
Computer Cabinet
Click for details
DRS Gateway
The DRS Gateway is an
optional hardware solution to
provide added network
functionality for the DRS.
The DRS Gateway is a stand
alone PC, running Windows
2000. Its functionality is
transparent to the customer
and requires no interaction
with the customer.
Exam Room
Monitor
Image Processing
System
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Hard Disk for Patient
Image Storage
Digital Image
Processing
Computer (DRS)
Cable Storage
Power Supply
Digital Image Processing
System Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Image Intensifier
Field of View
selection
buttons
Bolus
Lock
Table
Bucky
select
Grid
Drive
Cone
Drive
Fluoro
Timer
Reset
Collimator
Controls
Fluoro
and
Record
Exposure
Switches
Table
Angulation
Knob
SFD Lock
release
buttons
Film Format
Selection
buttons
SFD Console
Photo Spot
Camera
select
Longitudinal &
Lateral table-top
drive Switch
Vertical and
Longitudinal Power
Assist handle
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Displays number of
Images in Current
Study and remaining
image capacity of
hard disks
Delete Study:
Press twice to
delete selected
study
Delete
Image: Press
twice to delete
current image
Brightness
and Contrast
Adjustment
Specialty Keys:
Click HERE for
details
Patient Directory:
Select Patient
Directory screen on
DRS monitor
New Patient:
Select New Patient
screen on DRS
monitor
Keyboard: used
to enter patient
information
FluoroStore:
Saves the current
fluoro image to the
patient folder
Image Processing
Console (DRS)
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Click on the yellow
boxes for more
information
Rotating
Tube Arm
Xray Tube
Locks
release
button
Collimator
On/Off
Switch
Tube Latch
Mechanism &
Switch
Control
Panel
Exposure
switch
Imaging PC
Touchscreen
Power Cord,
On/Off
Breaker and
Network
connection
Drive
Handle
Digital
Detector
AMX 700
Collision Sensor
(blocks forward drive
movement when
activated)
Body of AMX contains
Generator, Digital
Imaging PC, Drive
Batteries, Drive and
Charging Electronics
and Motors
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Click on the yellow
boxes for more
information
Rotating
Tube Arm
Xray Tube
Collimator
Tube Latch
Mechanism &
Switch
Lead Apron
Hanger
On/Off
Switch
Locks
release
button
Control
Panel
Exposure
switch
Power Cord,
On/Off
Breaker
Drive
Handle
Cassette
Storage
Drawer
AMX 4
Collision Sensor
(blocks forward drive
movement when
activated)
Body of AMX contains
Generator, Drive
Batteries, Drive and
Charging Electronics
and Motors
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Click on the yellow
boxes for more
information
Table
Control Area
System Cabinet
OTS
Wall Stand
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Jedi Generator:
Generates the
power used by
the x-ray tube
Corona 2 Board
controls the system power
on, off and reset functions
and also serves as a CAN
Hub for the system’s CAN
buses.
Power Distribution
Unit (PDU):
Distributes 120Vac &
230Vac around the
system
Bulkhead
Halo Board
Contains relays which are
used to switch power to the
system.
The relays are controlled by
the Corona 2 Board.
On/Off Switch
System Cabinet
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Control, Patient List
& Image Monitor
Radiographic
Control Interface
Module (RCIM)
Click for details
Exposure
Switch
CD-RW/DVD
ROM Drive
Mouse
Wall Plate
(located under
counter)
System and
Image Processing
Computer (aka
Magic PC)
Keyboard
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Movements
Vertical
Longitudinal (along overhead
rails)
Lateral (along overhead rails)
Column rotation at the goose
neck
Xray Tube Angulation
Carriage
Telescopic
Column
Vertical
Tube
Angulation
User Interface
Console
Click for details
X-ray Tube
Collimator
Click for details
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Wallstand
Display
Operator
Handle
Wall Stand
(Standard version)
Collimator
Remote Control
Extended Arm
Version of Wall
Stand
Detector Angulation
and SID Display
TRAD Detector
used in Wallstand
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
Movements Enable
Button
Lit= movements disabled
OFF= movements enabled
Floating Table-top
Detector
Emergency Movements
Off Button
When pressed disables all
room movements
Table Up/Down
Table
Down
TRAD Detector
used in Table
Table
Up
Table-top Locks Release
(longitudinal & lateralfloating table-top mode)
X-ray Photo Tour- Rev.2 by E.Singleton
CLOSE
LEDs
Longitudinal
Lock Release
Ready
Exposure
Inhibit
Manual
Collimation
SID
Detent
Lateral
Detent
Long
Detent
Lateral Lock
Release
Tube
Angulation
Lock Release
Interlock Override Switch
Horizontal=override
position
Vertical Lock
Release
Touch-screen
Display
Column
Rotation Lock
Release
All Locks
Release
Indicator
Collimator
Field Light
Detent
Lock
All Locks
Release
Button
Detent Lock
Indicator
OTS User Interface Console
X-ray Photo Tour- Rev.2 by E.Singleton
Download