Users Manual HP 35900E Dual Channel Interface Printing History Manual part number 3590090400 First edition January 1994 Printed in USA Warranty The information contained in this document is subject to change without notice. HewlettPackard makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. HewlettPackard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. © Copyright 1994 by HewlettPackard Company All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws. Safety Information Safety Symbols Caution A Caution calls attention to a condition or possible situation that could damage or destroy the product or the user's work. WARNING A WARNING calls attention to a condition or possible situation that could cause injury to the user. Electrical Shock Hazard WARNING This product is a Safety Class 1 instrument (provided with a protective earth terminal). To avoid electrical shock, the instrument must be connected to the ac power supply mains through a threeconductor power cord, with the third wire firmly connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal, under a fault condition could cause a shock, that could result in personal injury. No user servicable parts inside. Refer servicing to qualified service personnel. Contents Capabilities Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Data Buffering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Independent AnalogtoDigital Converters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Realtime Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Remote Control Bus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Digital (TTL I/O) Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Instrument Control Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Host Computer Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Start/Stop Pushbuttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Run Status Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Hardware Monitoring Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 13 13 14 14 14 15 15 15 15 16 16 Setup Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Checking Parts and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Line Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Configuration Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 21 21 22 Computer Connection Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Handling the MIO Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Configuring an HPIB/RS232 MIO Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Configuration for HPIB Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Configuration for RS232 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Installing the Communication Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Connecting to Host Computer (HPIB/RS232 MIO) . . . . . . . . . . . . . . . . . . . . . . . . . HPIB Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Connecting to Network (TCP/IP MIO) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 31 32 32 33 35 36 37 38 Instrument Connection Analog Input (Signal) Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Remote Control Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Digital I/O Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Instrument Control Port Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 43 45 48 Contents HP 35900E Startup PreRun Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 35900E Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Verify HP 35900E is Ready for Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . In Case of Difficulty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Appendix A Specifications AnalogtoDigital Conversion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Run Control Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Host Communications Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Instrument Control Port Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Appendix B 51 51 52 52 A1 A2 A2 A3 A3 Cabling Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Analog Signal Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Remote Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Remote Control Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Bottle Number Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Digital I/O Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Digital I/O Adaptor Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Instrument Control Port Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HPIB Communications Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B1 B1 B6 B7 B13 B16 B17 B18 B19 1 Capabilities Capabilities Introduction The HP 35900E is a dualchannel run buffered interface that connects analytical instruments to HP ChemStation and ChemServer systems. It converts the analog signal from analytical instruments to digital data and transmits it to the host computer for further processing. The HP 35900E is intended for use with both HP and nonHP manufactured laboratory instruments. Figure 11. HP 35900E. The HP 35900E uses Modular Input/Output (MIO) for host communica tions. MIO is a standard designed to allow for maximum flexibility in peripheral to host communications. The HP 35900E has an MIO slot on the rear panel for plugin MIO cards. The TCP/IP LAN MIO card con nects the HP 35900E to the networked HP ChemServer systems. The HPIB/RS232 MIO card connects the HP 35900E to HP ChemStation systems using the HPIB functionality. 11 Capabilities Introduction The HP 35900E features include the following: D Data buffering D Independent analogtodigital (A/D) converters D Realtime control D Instrument control ports (RS232 ports) D Host communication D Front panel indicators that show run status, interface activity, etc. D Compatibility with HP GC and LC DOS ChemStations and HP Chem Server systems. 12 Capabilities Data Buffering Data Buffering The HP 35900E buffers (stores) digitized data from each run using PC like SIMM RAM modules. Data from the instrument is digitally converted, stored, and passed on to the host computer on request. The HP 35900E comes standard with 1 MB RAM. Through memory accessory kits (G1848A and G1849A) it can be upgraded to a maximum of 8 MB. Two channel data can be buffered for the following lengths of time as deter mined by the sampling rate and amount of RAM installed in the HP 35900E. SIMM RAM installed @ 20 Hz @ 100 Hz 1 MB SIMM RAM 3.4 hours40 minutes 2 MB SIMM RAM 7.0 hours 1.4 hours 4 MB SIMM RAM 14.3 hours 2.8 hours 5 MB SIMM RAM 17.9 hours 3.5 hours 8 MB SIMM RAM 28.8 hours 5.7 hours Independent AnalogtoDigital Converters The analogtodigital channels provide 24bit analogtodigital conversion of up to two independent analog signals. The analog inputs can either be totally asynchronous (have no relationship at all to each other) or syn chronous (separate signals, but logically related to the same START refer ence). To allow asynchronous operation, each channel has separate Start/Stop switches, status indicators, and controllable digital I/O. The analog input range is -18 mV to +1 V. The analog input signal is digi tized by a continuously integrating A/D converter. This means that the input signal is continuously applied to the converter's input, and because of that, no portion of the signal is ever lost due to input multiplexing or the sampling process. 13 Capabilities Realtime Control Realtime Control The HP 35900E provides a set of realtime control functions for a host computer system. These can be separated into two groups: D Remote Control Bus D Digital (TTL I/O) Ports Remote Control Bus The Remote Control Bus is a hardware interface that provides a set of defined signals most frequently encountered in the laboratory environ ment. These signals include START, STOP, READY and START REQUEST. The HP 35900E has one Remote Bus per channel to allow for asynchronous operation. These connectors are 9pin DType" female connectors each having eight bidirectional (input/output) signals. Digital (TTL I/O) Ports Each channel has a dedicated TTL I/O connector with 16 signal lines. Of the 16, eight are bidirectional configurable as inputs or outputs (for such functions as valve control) and the remaining eight lines are inputonly signals (for such functions as bottlenumbers). 14 Capabilities Instrument Control Ports Instrument Control Ports There are two RS232 ports (one for each channel) that may be used to interface with RS232 compatible lab instruments such as certain auto samplers, balances, GC's, etc. Line protocols and operating modes for RS232 communication are config ured via commands from the host computer. Host Computer Communication The HP 35900E has an MIO slot for host communications. There are presently two MIO cards that can be used with the HP 35900E. The RS232/HPIB MIO card, option 011 or G1847A, connects the HP 35900E to HP GC and LC DOSbased ChemStations. The TCP/IP LAN MIO card, option 010 or G1846A, connects the HP 35900E to HP ChemServer systems. Front Panel The front panel of the HP 35900E has START and STOP buttons and three indicator lights for each channel. There are also four indicator lights at the bottom of the panel to convey general information about the state of the HP 35900E interface. Start/Stop Pushbuttons Each channel has START and STOP pushbuttons that may be used to manually start and stop analytical runs. These pushbuttons can be indi vidually enabled or disabled by the host computer software controlling the HP 35900E. 15 Capabilities Front Panel Run Status Indicators Each channel has three indicator LED's (light emitting diodes). READY Indicates there is no current activity on the channel and it is ready to begin a chromatographic run. RUN This light indicates a run is in progress on that channel. NOT READY This light indicates the channel cannot start another run until other events are completed. Hardware Monitoring Indicators There are four indicator LEDs on the bottom of the front panel that dis play the status of the HP 35900E. ACTIVE Indicates that there is data present in the interface that will be lost if the HP 35900E's power is turned off. COMM This light indicates that the HP 35900E is communicating with the host computer. FAULT This light indicates that a nonrecoverable error occurred. POWER This light indicates that power is applied to the instrument and that the LINE switch is turned on. 16 2 Setup Setup Unpacking 1. If the HP 35900E is purchased as part of a system, check the carrier's papers to verify that you received all containers. 2. Inspect the shipping container for damage and watermarks. 3. Remove the cables and documentation from the shipping container and set them aside. 4. Unpack the HP 35900E and inspect it for damage. If you see damage, keep the carton and all packing materials for the carrier's inspection. Contact your nearest HewlettPackard Sales and Service Office immediately. Checking Parts and Accessories 1. Verify that you have all parts on the packing list. If you suspect that an option, accessory, or part is missing or incorrect, call your nearest HewlettPackard Sales and Service Office immediately. 2. Store user information in a safe place. Line Voltage The HP 35900E has a universal input range for line voltage and can be safely connected to both 115V and 230V power sources. There is no line voltage switch to set. 21 Setup Configuration Switch Configuration Switch There is a set of six configuration switches on the rear panel that control HP 35900E operation. The default settings are shown in the insert below. Switch Function 1 Beep/Off 2 HI/LO 3 Channel A TEST/NORM 4 Channel A ADD/IGNORE 5 Channel B TEST/NORM 6 Channel B Add/IGNORE Position Effect Up Causes an audible beep when Start/Stop button is pressed. Down Silent at Start/Stop Up Beep, if any, is loud Down Beep, if any, is soft Up Selects stored test chromatogram. Analog input signal ignored. Down Selects Analog input for Normal Operation Up Add signal input to the stored chromatogram. Switch 3 must be in Up position. Down Do not add (ignore) signal input to stored chromatogram for Ch. A. Up Selects stored test chromatogram. Analog input signal ignored. Down Selects Analog input for Normal Operation Up Add Ch. B. signal input to the stored chromat ogram. Switch 5 must be in Up position. Down Do not add (ignore) signal input to stored chromatogram for Ch. B. 1 2 3 4 5 6 Figure 21. Configuration switch. 22 Setup Configuration Switch Switches 1 and 2 affect the beeper operation. The remaining four switches combine to specify either a test mode or normal operation for each channel. The test mode provides service techni cians with a diagnostic tool to show the effects of electrical noise on a known test signal. Switches 3 and 4 combine to specify either a test mode or normal opera tion mode for Signal A. Switches 5 and 6 combine to specify either a test mode or normal opera tion mode for signal B. 23 3 Computer Connection Computer Connection Overview The HP 35900E and the host computer are connected through a Modular I/O (MIO) card. This type of removable card allows you to configure the HP 35900E with the correct type of link to communicate with your host computer. The HP 35900E can transfer and receive data from any MIO compliant card regardless of the type of connection the card has with the host computer. The HP 35900E presently supports the following MIO cards for host com munications: D HPIB/RS232 D TCP/IP LAN (network) The HPIB/RS232 MIO card requires configuration before use. If this card is used with the HP 35900E, proceed to the next section for instruc tions on configuration. After this card is configured, review the sections Installing the Communication Card" and Connecting to Host Computer." Handling the MIO Card Caution 31 The MIO card is staticsensitive and should be handled with static safe procedures to avoid damage. When handling the board, use the following precautions: D Use antistatic equipment during installation, such as a grounding mat and wrist strap. D Hold the board by its edges. D Do not touch the board's electrical components, or edge connectors. Computer Connection Configuring an HPIB/RS232 MIO Card Configuring an HPIB/RS232 MIO Card The HPIB/RS232 MIO card allows the host computer to communicate with the HP 35900E through either an RS232 or an HPIB connection. By setting the configuration switches on the card you can configure the type of connection desired. Typically, the HP 35900E is used in HPIB mode when connected to HP LC and HP GC ChemStations running on a PC. If necessary, the board switch settings can be inspected or changed by removing the board from the rear of the unit. First, turn the unit off and remove the power cord. Next, loosen the thumb screws on the back of the unit and then pull on the back plate of the MIO card to remove it from the unit. Configuration for HPIB Operations The HPIB/RS232 MIO card contains an 8 position configuration switch. When the position 1 switch is set to OFF, the MIO card uses the HPIB communication mode. In this mode, the ON switch position for switches 2 through 8 are defined in the following table and shown in figure 31. 1 ON 8 OFF REAR Figure 31. Factory default settings for HPIB. 32 Computer Connection Configuring an HPIB/RS232 MIO Card Switch ON Action 2 3 4 5 6 7 8 N/A SRQ Enable Add 16 to HPIB address Add 8 to HPIB address Add 4 to HPIB address Add 2 to HPIB address Add 1 to HPIB address The factory default setting for this switch is shown in figure 31. Because switch 1 is set to OFF, the board is in HPIB mode. Switch 3 indicates that SRQ (Service Request) is enabled. Switches 5, 6 and 8 set to ON and switches 4 and 7 set to OFF combine to form an HPIB instrument address of 13 (0+8+4+0+1). The most common reason for changing this switch setting is to avoid HPIB address conflicts. Conflicts occur when two or more instruments on the bus have identical HPIB addresses. To change the address when a conflict exists, select a new address not used by any other instrument on this HPIB bus. For example, to use 20 as the new address, set switches 4 and 6 ON and switches 5, 7, and 8 OFF (16+0+4+0+0=20). Configuration for RS232 Operation The HPIB/RS232 MIO card contains a single 8 position configuration switch. When the position 1 switch is set to ON, the MIO card uses the RS232 communication mode. In this mode, the ON switch positions for switches 2 through 8 are defined in the following table. 33 Computer Connection Configuring an HPIB/RS232 MIO Card Switch ON Action 2 3 4 5 6, 7, 8 Handshake Enabled Hardware Handshake (otherwise Software Handshake) Parity Enable Even parity (otherwise Odd Parity) Baud rate (see below) An example setting for this switch is shown in figure 32. 1 ON 8 OFF REAR Figure 32. Example settings for RS232. Because switch 1 is set to ON, the board is in RS232 mode. Switch 2 set to ON indicates that handshaking is enabled. With handshake enabled, switch 3 set to OFF indicates software handshaking. Switch 4 set to ON enables parity checking. Switch 5 set to OFF indicates odd parity. 34 Computer Connection Installing the Communication Card Switches 6 through 8 combine to indicate baud rate. See the following table for switch settings and their corresponding baud rates. Switches 6 and 8 set to ON and 7 set to OFF combine to form a RS232 baud rate of 9600 as shown in the table below. Baud Rate 6 Switch 7 8 300 600 1200 2400 4800 9600 19,200 38,400 OFF OFF OFF OFF ON ON ON ON OFF OFF ON ON OFF OFF ON ON OFF ON OFF ON OFF ON OFF ON Installing the Communication Card After you have configured your data communications card as described above, you can install it in your HP 35900E. 1. When handling the board, observe the precautions mentioned on page 31. 2. Remove the power supply cord from the rear of the unit. 3. Slide the MIO card into the back of the HP 35900E so that its two side edges slide into the guide rails. 4. Slide the card in until you feel it seat in the connector at the rear of the slot. 5. Secure the card with the two thumb screws provided on the card. 35 Computer Connection Connecting to Host Computer (HPIB/RS232 MIO) Connecting to Host Computer (HPIB/RS232 MIO) The HP 35900E with a HPIB/RS232 MIO card installed has two host connector types. Only one of these connectors can be used to connect to a host computer at a time. Use the HPIB connector when attaching to a host requiring HPIB communications. Use the RS232 connection when the host uses this protocol. MIO Communication card Thumb screws HPIB Communication Cables HP Part Number Description Page 10833A 1 Meter B-19 10833B 2 Meter B-19 10833C 4 Meter B-19 10833D 0.5 Meter B-19 Figure 33. HPIB cable connection. 36 Computer Connection Connecting to Host Computer (HPIB/RS232 MIO) HPIB Connection 1. Turn off the HP 35900E and the Host Computer. 2. Connect one end of a HP 10833A/B/C/D cable to the MIO card's HPIB connector located on the rear of the HP 35900E. 3. Connect the other end of this cable to the HPIB connector on the host computer or to a HPIB cable attached to another device on the HPIB Bus that eventually terminates at the host computer. HPIB length limits are up to 2 meters per device (the host computer is one device) with a maximum of 20 meters. 37 Computer Connection Connecting to Network (TCP/IP MIO) Connecting to Network (TCP/IP MIO) The HP 35900E with a TCP/IP MIO card installed has two Ethernet IEEE 802.3 connector types: D BNC (thin Ethernet coaxial cable) D RJ45 (twistedpair cable) Before attaching the HP 35900E to the network, the TCP/IP interface card must be installed in the HP 35900E and the network must be opera tional. MIO Communication card Thumb screws RJ45 BNC Figure 34. Network connection. 38 4 Instrument Connection Instrument Connection Analog Input (Signal) Connection The signal must be within the range of -18 millivolts to 1 volt. A simple voltage divider resistor network may be used to attenuate larger signals to this range. Each input has three pins to which the high (+) and low (-) sides of the signal and the shield are connected. Use an instrumentspecific cable whenever possible. Use a general purpose cable in those cases where a custom cable is not available. Appendix B lists pin assignments for all cables. 1. Turn off the power to the instrument supplying the analog output signal. 2. For instrumentspecific cables Connect one end of the cable to the instrument's analog output using the instrument manufacturer's instructions. Caution For correct operation of the HP 35900E, the input voltage must be in the range of -18mv to +1V. Voltages below -10V or above +10V may dam age the instrument. For general purpose cables Use the pin assignments shown on page B1, and the instrument manufacturer's instructions to connect the two signal wires and ground to the instrument. 3. Connect the cable to the appropriate (A or B) analog input on the rear panel of the HP 35900E. 4. If needed, repeat for the other channel. 41 Instrument Connection Analog Input (Signal) Connection Analog Input Channel A Analog Input Channel B Analog Signal Cables 3590060570 HP 5880 Gas Chromatograph Pinouts Page B-2 3590060590 HP 5790A Gas Chromatograph B-2 3590060630 General Purpose Spade Lugs (12 M) B-1 3590060900 General Purpose Spade Lugs (5 M) B-1 3590060910 General Purpose Spade Lugs (15 M) B-1 3590060610 HP 5890A Gas Chromatograph B-4 3590060640 Specialpurpose square pin B-5 3590060620 HP 5710/30 GC and HP 1081B UV B-4 3590060600 HP 1090 Series and HP 1040M DAD B-3 3590060750 HP 1050 and HP 1046A Fluorescence det. B-5 HP Part Number Figure 41 Description Analog signal cables. 42 Instrument Connection Remote Control Connection Remote Control Connection Use an instrumentspecific cable whenever possible. Use a general purpose cable for equipment not listed in the figure below. Appendix B lists pin assignments for all cables. Channel A Channel B Remote Control Cables HP Part Number Description 3590060670 General Purpose Spade Lugs (2 M) B-6 3590060920 General Purpose Spade Lugs (5 M) B-6 3590060930 General Purpose Spade Lugs (15 M) B-6 3590060800 General Purpose Use H" cable B-7 3590060700 HP 5890A Gas Chromatograph HP 7673A Sampler B-8 3590060710 HP 1090A/L Liquid Chromatograph B-9 50613378 HP 1046A Fluorescence Detector; HP 1050A Series B-10 3590060860 HP 19395A/HP 5890/HP 35900 B-11 3590060660 HP 5700 Series Gas Chromatograph B-11 3590060970 HP 5890A GC/HP 7673A Sampler/ HP 35900 System B-12 Figure 42. Remote control cables. 43 Page Instrument Connection Remote Control Connection 1. Turn off the power to both the HP 35900E and the instrument con taining the remote start/stop contacts. 2. For instrumentspecific cables Connect one end of the cable to the instrument's remote control connector using the instrument manufacturer's instructions. For general purpose cables Use the pin assignments shown on page B6, and the instrument manufacturer's instructions to connect the wires to the instrument's remote control contacts. The following notes apply to a general purpose connection: START and STOP lines are used for both input and output. As input, a 10 msec or longer pulse of TTL low(<0.8V) is required. As output, a 120 msec TTL low pulse is sent. READY line is both input and output. TTL low indicates not ready. POWER ON line is shorted to ground when power is off. It is TTL high (>2V) when power is on. 3. Connect the cable to the appropriate (A or B) remote control connector on the rear of the HP 35900E. 4. If needed, repeat for the other channel. 44 Instrument Connection Digital I/O Connection Remote Control Bus Pin Definitions Pin Signal Name Description 1 Digital Ground 2 Prepare* Pulsed low to request prepare for analysis (e.g., calibrate detector, turn on lamp, etc.). 3 Start* Pulsed low to signal that run has started (begin integration, temperature program, etc.). 4 Shut Down Low level indicates that system has a serious problem (e.g., leak: stop pump). 5 Reserved 6 Power On High level indicates that all instruments connected to the system are switched on. 7 Ready HIgh level indicates that system is ready for next analysis (sequence may continue). 8 Stop* Pulsed low to signal end of run; all modules should go to ready state as soon as possible. 9 Start Request* Pulsed low to request start of injection cycle, (auto sampler signals START at time of injection). * Input signals must be TTL low pulses (or contact closures to ground) with a duration of at least 10 msec; output signals are TTL low pulses with a duration of approximately 120 msec. Digital I/O Connection Each channel has a 25 pin female connector with 16 signal lines. Of the 16, eight are bidirectional input/output signals (for such functions as valve control) and the remaining eight lines are inputonly signals (for such functions as reading bottlenumber inputs). A Y" adapter cable (figure 43) is available to route the 16 digital I/O signals to two 9pin connectors. This cable takes the wires from the 8 digi tal I/O signals in the 25 pin connector and routes them to the 9 pin female connector labeled A", and the 8 digital input only signals and routes them to the 9 pin female connector labeled B". 45 Instrument Connection Digital I/O Connection Channel A (setup identical to Channel B) C Channel B 8 lines of digital I/O A Digital I/O Adaptor Cable HP Part No. 3590060960 Digital I/O Cables HP Part No. Description 3590060670 General Purpose Spade Lugs (2 M) General Purpose Spade Lugs (5 M) General Purpose Spade Lugs (15 M) MetraByte 8 channel relay box 3590060920 3590060930 3590060940 8 lines of digital input only B Bottle Number Cables Page HP Part No. Description Page B-6 3590060670 B-13 B-6 3590060920 B-6 3590060930 B-16 3590060840 General Purpose Spade Lugs (2 M) General Purpose Spade Lugs (5 M) General Purpose Spade Lugs (15 M) Sample Number from HP 19395A From HP 1050 Series LC or HP 7673 3590060850 B-13 B-13 B-14 B-15 Figure 43. Digital I/O cables. 46 Instrument Connection Digital I/O Connection Use an instrumentspecific cable whenever possible. Use a general purpose cable for equipment not listed in figure 43. Appendix B lists pin assign ments for all cables. 1. Turn off the power to both the HP 35900E and the instrument con taining the digital I/O connections. 2. Attach the Y" adapter cable's 25 pin connector labeled C" to the appropriate channel's digital I/O connector on the rear of the HP 35900E. 3. For instrumentspecific cables Connect one end of the cable to the instrument's digital I/O connector using the instrument manufac turer's instructions. For general purpose cables Use the pin assignments shown on page B6 for connector A" or B", and follow the instrument manufac turer's instructions to connect to the instrument's digital I/O. 4. Connect each instrument's digital I/O cables to the Y" adapter cable's A" and B" 9 pin female connectors. Note that the Y" cables B" con nector is for bottle number type digital input and its A" connector is for digital I/O used to control relays. 5. If needed, repeat for the other channel. 47 Instrument Connection Instrument Control Port Connection Instrument Control Port Connection Attach one end of the RS232C cable to the channel A or B instrument port and the other end to the instrument. Appendix B lists pin assign ments for all cables. RS232 Connection HP Part Number 3590060980 Figure 44 Description HP 5890 or HP 7673 Page B-18 Instrument control port connection. 48 5 HP 35900E Startup HP 35900E Startup PreRun Checklist Before using the HP 35900E with the operating software on the host com puter, make sure you have completed the following: Rear Panel configuration switches 3 through 6 are DOWN (normal operation) MIO card is correctly configured for host computer communications MIO card connected to host computer or network Analog signal cables connected Remote control cables connected (if needed) Digital I/O cables connected (if needed) Power cord connected HP 35900E Power Up WARNING The power cord must be connected to an outlet with a protective earth contact to ground the unit and minimize shock hazard. Power up the HP 35900E by pressing the line switch (lower left corner of the front panel). 51 D All front panel indicators should turn on for 1 to 2 seconds. D All indicators should turn off, except for POWER (on steadily) and FAULT (blinks while selftests are performed). D After approximately 30 seconds, the FAULT indicator should turn off. Some other indicators may turn on, depending on the instrument con figuration. 35900E Startup Verify HP 35900E is Ready for Operation Verify HP 35900E is Ready for Operation At power up, the HP 35900E automatically executes internal selftests to determine the amount of memory installed and to verify that the main processor, system memory, the MIO communication interface, and other internal circuits are operating properly. These tests take about 30 seconds for a 1MByte system. The FAULT light will blink ON/OFF during this time. When the HP 35900E has completed the internal selftests, verify that: D The FAULT lamp is OFF D The POWER indicator is ON. D No other indicators are blinking. In Case of Difficulty If the unit does not power up correctly, turn power OFF and check that the MIO communication interface is properly installed, then turn power ON to see if the fault is cleared. If the fault condition persists, contact HewlettPackard for service. 52 Appendix A Specifications Appendix A Specifications AnalogtoDigital Conversion General Specifications Processor: Number of channels: Type of A/D conversion: Memory: Electrical Specifications Input signal voltage range: Maximum (no damage): Analog dynamic range: Resolution: Commonmode voltage: DC input impedance: Input noise: Commonmode rejection: Bandwidth: Thermal drift: Integral nonlinearity: (monotonically guaranteed) Conversion factor: Area resolution: A/D sampling rate: RFI immunity: A1 Motorola 68331 at 16.78 MHz Two independent Continuouslyintegrating dualslope, 100% area recovery 1 MB RAM (expandable to 8 MB RAM) -18 mV to +1.0 V "10 V >140 dB (v 2 Hz) typical 24 bits "100 V maximum, relative to instrument chassis Differential: 33 MW, typical 8 MW, minimum Commonmode: 500 MW, minimum 40 nV rms, typical (with input shorted) 150 nV rms, maximum 140 dB minimum, dc to 100 Hz 15 Hz 0.8 mV per _C, typical: 0_C to 35_C 1.7 mV per _C, typical: 35_C to 55_C 3.5 mV per _C, maximum: 0_C to 35_C 10 mV per _C, maximum: 35_C to 55_C 0.004% FS, typical 0.02% FS, maximum 10 nV per count ("3%) 65 nVsecond ("3%) 0.1 to 100 Hz Maximum baseline shift of 50 mV with application of up to 3 V/m of RF field strength from 26 MHz of 1000 MHz. Appendix A Specifications Run Control Specifications Run Control Specifications Remote Control Remote Start/Stop/Ready: Two independent connectors, open collector TTL input/output Digital I/O Number of lines: 16 lines per channel (8 bidirectional, 8 input only) I/O signal levels: TTL level input, open collector outputs Input common mode range: 0-5 V Connectors: Two 25pin female Dsubminiature connectors Notes: 1. Differential ratings apply between the positive and negative input terminals. 2. The common mode ratings apply between both positive and negative input terminal and common GROUND (instrument ground or chassis). Host Communications Specifications Modular Input/output (MIO) cards conform to MIO 5.0 specifications. HP ChemServer Interface LAN MIO Card Network type: Ethernet Protocol: TCP/IP Cable Requirements: Thin Ethernet cable (thin coax [10 Base 2]) Twistedpair cable (unshielded/shielded [10 Base T]) HP GC and LC ChemStation Interface HPIB/RS232C MIO Card Type of interface HPIB Communications protocol: HPIB (HewlettPackard interface bus); conforms to ASNI/IEEE488 standard Type of interface: RS232C Baud rate 300, 600, 1200, 2400, 4800, 9600, 19200, 38400 Transmission: Full duplex Parity: Even, odd, none Start bits: One Stop bits: One Pacing: XON/XOFF both directions, hardware handshake Connectors: One standard 25pin male Dsubmineature connector A2 Appendix A Specifications Instrument Control Port Specifications Instrument Control Port Specifications Type of interface: Baud rates: Transmission: Parity: Start bits: Stop bits: Pacing: Connectors: RS232C 300, 1200, 2400, 4800, 9600 Full duplex Even, odd, none One One, two XON/XOFF both directions, hardware handshake Two standard 9pin male Dsubmineature connectors Physical Specifications Mechanical Size: Weight: Electrical Line voltage range: Line voltage fluctuations: Line frequency range: Line power consumption: Safety classification: Safety installation category: Environmental Heat dissipation: Temperature ranges: Humidity: Altitude: Safety Pollution degree: A3 325 mm (12.8 in.) wide 285 mm (11.2 in.) deep 104 mm (4.1 in.) high 4.1 kg (9 lbs) 115 V setting "10%; for 230 V setting "10% "10% maximum 50 Hz "5%, 60 Hz " 5% Nominal: 35 VA Maximum: 135 VA IEC safety Class 1 IEC overvoltage category II Nominal: 137 Btu/hr Maximum: 205 Btu/hr Operating: 5_C to +40_C 0% to 80% (noncondensing) 2,000 meters maximum IEC pollution degree 2 Appendix B Cabling Appendix B Cabling Overview This section provides cable diagrams for the analog signal, remote control, and bottle number cables that work with the HP 35900E. Analog Signal Cables HP Part Number 3590060630 (12 M) 3590060900 (5 M) 3590060910 (15 M) Instrument Use General Purpose Use-Spade Lugs General Purpose Use-Spade Lugs General Purpose Use-Spade Lugs 1 2 1 2 3 Connector 1 B1 Signal Name Connector 2 1 Shield Orange 2 Signal - Black 3 Signal + Red Appendix B Cabling Analog Signal Cables HP Part Number 3590060570 Instrument Use HP 5880 Gas Chromatograph 1 2 1 2 3 Connector 1 Signal Name Connector 2 1 Shield Orange 2 Signal - Black 3 Signal + Red HP Part Number 3590060590 Instrument Use HP 5790A Gas Chromatograph 1 2 1 2 3 Connector 1 Signal Name Connector 2 1 Shield Orange 2 Signal - Black 3 Signal + Red B2 Appendix B Cabling Analog Signal Cables HP Part Number 3590060600 Instrument Use HP 1090A/L Liquid Chromatograph; HP 1040A Diode Array Detector 1 2 1 1 2 3 8 Connector 1 B3 Signal Name Connector 2 1 Shield 1 2 Signal - 5 3 Signal + 3 Jumper 4, 6 Appendix B Cabling Analog Signal Cables HP Part Number 3590060610 Instrument Use HP 5890A Gas Chromatograph 1 2 5 3 1 1 2 3 6 4 2 No Connections Connector 1 Signal Name Connector 2 1 Shield 5 2 Signal - 1 3 Signal + 3 HP Part Number 3590060620 Instrument Use HP 5710/30 Gas Chromatograph; HP 1081B UV Detector 1 2 1 2 3 B D Connector 1 Signal Name A H E Connector 2 1 Shield E 2 Signal - B 3 Signal + H B4 Appendix B Cabling Analog Signal Cables HP Part Number 3590060640 Instrument Use For Special Purpose Square-Pins 1 2 1 2 3 Connector 1 Signal Name Connector 2 1 Shield Orange 2 Signal - Black 3 Signal + Red HP Part Number 3590060750 Instrument Use HP 1046A Fluorescence Detector; HP 1050 Series 1 2 1 2 3 Connector 1 B5 Signal Name Connector 2 2 Signal - Shield 3 Signal + Center Appendix B Cabling Remote Control Remote Control HP Part Number 3590060670 (2 M) 3590060920 (5 M) 3590060930 (15 M) Instrument Use General Purpose Use-Spade Lugs General Purpose Use-Spade Lugs General Purpose Use-Spade Lugs 1 6 9 2 1 5 Note: When used for remote control, this cable attaches directly to APG remote port on back of HP 35900E. When used for digital I/O or Bottle Numbers, the Digital I/O Adaptor Cable (page B-17) must be used to connect to the HP 35900E. Connector 1 Connector 2 Remote Signal 1 Black GND 2 White Prepare 3 Red Start 4 Green Shut Down 5 Brown Reserved 6 Blue Power On 7 Orange Ready 8 Yellow Stop 9 Violet Start Request - Black Tubing Shield B6 Appendix B Cabling Remote Control Cables Remote Control Cables HP Part Number 3590060800 Instrument Use General Purpose Use 1 6 9 5 male 1 6 1 6 9 5 9 5 male female 1 6 9 5 All Connectors B7 Signal Name 1 Ground 2 Prepare 3 Start 4 Shutdown 5 Reserved 6 Power On 7 Ready 8 Stop 9 Start Request female Appendix B Cabling Remote Control Cables HP Part Number 3590060700 Instrument Use HP 5890A Gas Chromatograph HP 7673A Sampler 1 6 9 2 1 2 1 5 12 11 Connector 1 Signal Name Connector 2 1 Ground 4 3 Start 8 7 Ready 5 - Jumper 1, 10 Jumper 2, 6 - Jumper 7, 11 B8 Appendix B Cabling Remote Control Cables HP Part Number 3590060710 Instrument Use HP 1090A/L Liquid Chromatograph 1 2 1 6 9 1 5 8 Connector 1 B9 Signal Name Connector 2 1 Ground 1, 8 3 Start 4 7 Ready 3 8 Stop 6 Appendix B Cabling Remote Control Cables HP Part Number 50613378 Instrument Use HP 1046A Fluorescence Detector; HP 1050A Series 1 6 9 2 1 1 5 5 Connector 1 Signal Name 6 9 Connector 2 1 Ground 1 2 Prepare 2 3 Start 3 4 Shutdown 4 5 Reserved 5 6 Power On 6 7 Ready 7 8 Stop 8 9 Start Request 9 B10 Appendix B Cabling Remote Control Cables HP Part Number 3590060860 Instrument Use HP 19395A/HP 5890/HP 35900 System 1 6 9 2 1 3 6 5 4 1 Signal Name Connector 1 Connector 2 Ground 1 3, 6 Start 3 5 Start Request 9 4 HP Part Number 3590060660 Instrument Use HP 5710/30, HP 5790 Gas Chromatograph 1 6 9 1 B D 5 Signal Name B11 2 Connector 1 Connector 2 Ground 1 E Start 3 B Ready 7 A A H E Appendix B Cabling Remote Control Cables HP Part Number 3590060970 Instrument Use HP 7673A Sampler HP 5890A Gas Chromatograph HP 35900 System HP 35900 1 6 9 HP 7673 5 2 1 12 11 2 1 12 11 HP 5890 HP 7673 Pin #/Description HP 5890 Pin #/Description 1 - 9 - 2 - 2 Ground 2 Ground 1 Ground 3 - 3 Jumper to Pin 4 6 - 4 - 4 Jumper to Pin 3 4 - 5 Jumper to Pin 8 5 Jumper to Pin 8 5 - 6 Ready Out 6 Ready Out 7 Ready 7 Start Out 1 Start In 3 Start 8 Jumper to Pin 5 8 Jumper to Pin 5 8 - 9 - 7 Start Out 9 Start Rqst 10 Jumper to Pin 11 10 Jumper to Pin 11 11 Jumper to Pin 10 11 Jumper to Pin 10 12 - 12 - HP 35900 Pin #/Description B12 Appendix B Cabling Bottle Number Cables Bottle Number Cables HP Part Number 3590060670 (2 M) 3590060920 (5 M) 3590060930 (15 M) Instrument Use General Purpose Use-Spade Lugs General Purpose Use-Spade Lugs General Purpose Use-Spade Lugs 1 6 9 2 1 5 Note: When used for remote control, this cable attaches directly to APG remote port on back of HP 35900E. When used for digital I/O or Bottle Numbers, the Digital I/O Adaptor Cable (page B-17) must be used to connect to the HP 35900E. Connector 1 B13 Signal Name Connector 2 1 Ground Black 2 BCD 1 White 3 BCD 2 Red 4 BCD 4 Green 5 BCD 8 Brown 9 7 BCD 10 BCD 20 Violet Orange 8 BCD 40 Yellow 6 - BCD 80 Shield Blue Black Tubing Appendix B Cabling Bottle Number Cables HP Part Number 3590060840 Instrument Use Sample Number Cable from HP 19395A 1 2 14 6 9 1 1 5 25 13 Note: Also requires the Digital I/O Adaptor Cable (page B-17). Connector 1 Signal Name Connector 2 2 BCD 1 1 3 BCD 2 2 4 BCD 4 3 5 BCD 8 4 9 BCD 10 5 7 BCD 20 6 8 BCD 40 7 6 BCD 80 7 1 Ground 7 B14 Appendix B Cabling Bottle Number Cables HP Part Number 3590060850 Instrument Use HP 1050 Series Liquid Chromatograph 1 6 9 2 8 1 5 1 Note: Also requires the Digital I/O Adaptor Cable (page B-17). Connector 1 B15 Signal Name Connector 2 2 BCD 1 5 3 BCD 2 8 4 BCD 4 7 5 BCD 8 6 9 BCD 10 4 7 BCD 20 1 8 BCD 40 3 6 BCD 80 2 1 Ground 9 15 9 Appendix B Cabling Digital I/O Cables Digital I/O Cables HP Part Number 3590060940 Instrument Use Special Cable for use with Omega / MetraByte eight channel relay box 1 2 13 6 9 25 1 5 1 14 Note: Also requires the Digital I/O Adaptor Cable (page B-17). Connector 1 Connector 2 1 Black 11 2 White 10 3 Red 9 4 Green 5 Brown 8 7 6 Blue 6 7 Orange 5 8 Yellow 4 9 Violet 3 B16 Appendix B Cabling Digital I/O Adaptor Cable Digital I/O Adaptor Cable HP Part Number 3590060960 Note 3 25 A A 13 1 6 9 5 C C 1 6 B B 9 1 5 Note 2 Connect to HP 35900E Digital I/O Port Digital I/O (Connector A) Connector C 2 Signal Name Ground Bottle Number (Connector B) Connector A Connector C 1 9 Signal Name Ground Connector B 1 15 BCD 100 2 22 BCD 1 2 3 BCD 200 3 10 BCD 2 3 16 BCD 400 4 23 BCD 4 4 4 BCD 800 5 11 BCD 8 5 17 BCD 1000 9 24 BCD 10 9 5 18 BCD 2000 12 BCD 20 7 BCD 4000 7 8 25 BCD 40 8 6 BCD 8000 6 13 BCD 80 6 Notes: 1. Connector C pins 1, 7, 8, 14, 19, 20, 21 are not used. 2. Connection for digital input only such as bottle number. 3. Connection for digital I/O cables. B17 14 Appendix B Cabling Instrument Control Port Cable Instrument Control Port Cable HP Part Number 3590060980 Instrument Use HP 35900E to HP 5890 or HP 7673 13 6 9 25 1 5 1 Female 14 Female HP 35900C Connector Pin Signal Name 1 Carrier Detect 2 Receive Data 3 Transmit Data 4 Data Terminal Ready 5 Signal Ground 6 Data Set Ready 7 Request to Send 8 Clear to Send 9 Ring Detect B18 Appendix B Cabling HPIB Communications Cable HPIB Communications Cable 13 24 1 12 1 12 13 24 10833A 10833B Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 B19 1 Meter 2 Meter 10833C 10833D Signal D101 D102 D103 D104 EOI DAV NRFD NDAC IFC SRQ ATN SHIELD D105 D106 D107 D108 REN P/O TWISTED PAIR P/O TWISTED PAIR P/O TWISTED PAIR P/O TWISTED PAIR P/O TWISTED PAIR P/O TWISTED PAIR SIGNAL GROUND 4 Meter 0.5 Meter Name End or iIdentify Data Valid Not REady for Data Not Data Accepted Interface Clear Service Request Attention WITH WITH WITH WITH WITH WITH 6 7 8 9 10 11 Remote Enable Copyright E 1994 HewlettPackard Printed in USA 1/94 HP Part Number 3590090400 Cut line is ¼ inch below bottom of page.