> REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) < 1 A Standard Protocol for Reliable and TimeSaving Shielding Effectiveness Measurements for MRI Faraday Cages L. Mirarchi, V. Giaquinto, S. Silvestri, R. Massa Abstract— Objective: Aim of this work is to suggest a new procedure for evaluation of shielding performance of Radiofrequency cabins for Magnetic Resonance Imaging (MRI). Methods: On the basis of the only reference document currently available (IEEE-Std 299/2006), some critical aspects on shielding effectiveness measurements are highlighted due to the particular environment and involved frequencies. Results: A strong gap existing between the Std suggested procedure and practical approach towards the shielding performance evaluation came out from our investigations. Conclusion: Taking into account theoretical considerations, measurements carried out in real scenarios, and the consciousness of a lack of suitable normative references in magnetic resonance structures, a novel approach is suggested in order to overcome the inadequacy of the standard in MRI environments. Significance: The simple and practical approach increases the repeatability and reproducibility of measurements allowing quantitative comparisons and performance assessments. Index Terms — Electromagnetic compatibility, Electromagnetic shielding, Interference, MRI Faraday cage, RF signals, Shielding Effectiveness measurements, Standard Protocol confused with pathology. There are multiple causes of artifacts that can be patient-related, signal processing-dependent and hardware (machine)-related [1,2,3]. One of the latter is due to RF noise. RF pulses of MRI instruments and precessional frequencies occupy the same frequency band as common sources such as TV, radio, fluorescent lights and computers. Stray RF signals can cause various artifacts (Fig.1): narrowband noise is projected perpendicular to the frequencyencoding direction, broadband noise disrupts the image over a much larger area due to the Fourier transform needed to reconstruct the image. However, stray RF interferences reduction or elimination can be achieved with appropriate site planning and proper installation but overall with RF shielding realized by means of Faraday cages. I. INTRODUCTION M AGNETIC Resonance Imaging (MRI) is nowadays a powerful diagnostic method; its diffusion is increasing because of its known advantages in terms of high quality images, non-ionizing radiations and relative low risks. The main Healthcare companies are continuously improving their MR devices, developing innovative solutions about MR scanners, Radiofrequency (RF) coils, hardware components and magnets. All MR scanners are recognized as medical devices. In this contest, artifacts are a constant issue in MRI, because they can affect the diagnostic quality or, worst, may be Fig. 1. Examples of RF Artifacts in MRI imaging Manuscript received on Corresponding author: Rita Massa L.M., is with Siemens Healthineers – Via Vipiteno 4 -20124 Milano luciano.mirarchi@siemens.com V.G., is with the Department of Electronic Engineering University of Naples Federico II, Piazzale V.Tecchio 80, 80125, Naples, Italy va.giaquinto@studenti.unina.it S.S., is with Unit of Measurement and Biomedical Instrumentation, University Campus Bio-Medico of Rome, Via Alvaro del Portillo 21, 00128 Rome. s.silvestri@unicampus.it R.M. is with Department of Physics ―Ettore Pancini‖ University of Naples Federico II, CMSA, Via Cintia, 80126 Naples, Italy rmassa@na.infn.it Each MR device, independently from the magnetic static field intensity, needs this particular hosting enclosure, which is strictly dependent on the specific device features. Anyway, differently from the design and installation of MR systems, the design and setup of the shielding enclosures is currently quite simple and it often deals with handed down ―best practice‖; a RF shielding cabin is quite simple from a structural point of view and, after all, the consolidated techniques are in most cases sufficient to guarantee good performances. However, > REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) < this relative simplicity often leads to underestimate some important aspects that instead would deserve more attention, as they can reveal critical consequences. A meaningful example is represented by the shielding effectiveness (SE) measurement methods. Objective of this work is to highlight some critical aspects on SE measurements of MRI Faraday cages due, on the basis of our experience, to a strong gap existing between the reference standard suggested procedure (IEEE-Std 299/2006 [4]) and the practical approach. Our aim is to begin a way into an almost unexplored field, in order to organize current knowledge and to introduce innovation elements. At the end of this work we, thus, suggest a sort of new protocol for SE measurements for MRI shielded rooms. In section II a short description of the electromagnetic scenario is reported, as well as the critical points in designing RF enclosures. Among the quality and safety aspects, the SE measurements are deepened and the IEEE Std 299/2006 is presented in Section III. In Section IV the inadequacy of the standard in the case of MRI Faraday cages is underlined and a novel approach is suggested in order to uniform SE measurements in MRI environments and guarantee the required high accuracy. The conclusions of the study are in Section V. II. MRI ELECTROMAGNETIC SCENARIO Although the complexity of current MRI scanners, the hardware of a MRI system is basically composed of four main elements, which can be easily summarized as follows: 1) main magnet, whose function is to generate a static and homogeneous magnetic induction field (B0) for the nuclei magnetization; 2) gradient coils, responsible for magnetic induction fields linearly varying in space, necessary for images generation; 3) ancillary coils, useful to compensate non-homogeneities and to modify the shapes of main fields; 4) RF coils, that provoke the induction magnetic oscillating field at the desired Larmor frequency that is proportional to B0. Worldwide, more than 60 million clinical MRI scans are performed annually on over 25,000 MRI systems. Most of these systems are used purely for clinical purposes and operate at field strength at or below 3 T. Nevertheless, major neuroimaging centers of hospitals and universities are using or planning to use even higher field strength systems (7 T, 9.4 T, 10.4 T, 11 T) primarily for research applications [5]. In the case of Hydrogen proton, Larmor frequency is 42.57 MHz/T. Thus, for clinical purposes RF transmitting and receiving systems operate at frequencies as 8.514 MHz, 63.85 MHz and 127 MHz for 0.2 T, 1.5 T and 3 T field strengths, respectively. Thus, the widest considerable range of frequencies in MRI applications is estimated approximately around [8 MHz-150 MHz], whereas for very high field MRI scanners (7 T and higher) higher RF signals (i.e. 300 MHz and higher) result. This frequency range is the same as used by common sources present in the environment. Thus, the need of a RF shielded enclosure to prevent radio waves to enter the scanner room, which may cause artifacts on the MRI image as shown in Fig.1. Such enclosure is usually called ―Faraday cage‖, even if, more properly, ―Faraday cages‖ usually indicate enclosures which protect from electrostatic fields. However, in this 2 discussion both the terms, Faraday cage and RF cabin, are used to mean the shielding enclosure surrounding an MRI device. They are realized by means either of aluminum selfsupporting panels, bolted together with steel screws and a conductive scrim, or copper sheets, fixed on wooden panels. The RF cabin is provided with a grounding system, whose resistance should not exceed 1 Ω (connections between RFshield and the equipotential hub), and an electrically insulating carpet (about 2 mm thick). Very high levels of attenuation, about 70-130 dB, are expected according to the specifications required from a given MRI scanner and generally specified at the time cabins are designed, ordered or built. To improve the effectiveness, the cage must have the minimum number of conveniently designed apertures; the electrical continuity has to be guaranteed through the use of specific gaskets, conductive scrims, metallic fingers and so on. The most critical points in designing this RF enclosure are: ● the access door, which must guarantee perfect adherence with the rest of the structure when it is closed. This electrical continuity is achieved through creeping contacts, called fingers (usually made of copper/beryllium) and through the minimization of mechanical stresses; ● the view-window, made in double glass or polycarbonate, with a metal grid inside; ● the apertures for air-vent, realized with particular honeycomb filters or waveguides; ● the filter panel connecting the electronic systems in the adjacent technical area to the Magnet Room; ● the waveguides used to introduce in the MRI room medical gases or liquids. Although the normative overview about medical devices, electromedical equipment and magnetic resonance is abundant, a lack of a specific reference for the SE measurements methods, suitable in MRI locations, results. That is why the manufacturers of MRI Faraday cages, or responsible of the maintenance service have to address their attention to more general documents concerning this issue. A key role is played by the standard IEEE 299-2006 [4], presented in next paragraph. III. THE STANDARD IEEE 299-2006 As in [6] we define SE as follows: “The ratio of the signal received (from a transmitter) without the shield, to the signal received inside the shield; it’s the insertion loss when the shield is placed between the transmitting antenna and the receiving antenna” The IEEE Std 299-2006 is nowadays the only reference for SE measurements of large enclosures (smallest linear dimension greater than or equal to 2.0 m) and it incorporates the basic concepts of MIL-STD 285 [7]. Uniform measurement procedures are suggested for frequencies from 9 kHz up to 18 GHz (extendable down to 50 Hz and up to 100 GHz). As discussed in the next section, although MRI intended rooms fully fall under the domain of applicability of the standard it is often impossible (or redundant) to respect it in all its requirements and this is a relevant issue for producers, testing > REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) < companies, clinical engineers or MRI professionals involved in the acceptance process, installation or maintenance management of MRI cabins. Briefly, following [4], before the SE evaluation preliminary procedures are recommended: preparation of a test plan, calibration of any piece of instrument equipment, evaluation of an appropriate reference level, preliminary shield check, removal of objects not belonging to the usual electromagnetic scenario. For each configuration two measures are required: the first one in absence of the shield (H1 (E1)) indicates the magnetic (electric) field measured with the antennas in absence of the enclosure (reference reading); the second one, with the interposition of the shield between the transmitting and the receiving antennas, (H2 (E2)) is the magnetic (electric) field measured with the receiving antenna inside the enclosure. The whole frequency range (9 kHz-18 GHz) is divided into: low range (9 kHz-20 MHz), resonant range (20 MHz-300 MHz) and high range (300 MHz-18 GHz), for which different procedures, antennas and formulas are required. Specific test frequencies are not declared, but some sub intervals are recommended for each range. In the low range the definition of SE takes into account the magnetic component of the EM field, in the resonant range the electrical one, in the high range power is considered. Measured quantities, related units and SE definitions are resumed in Table I, where they are expressed also in nonlinear units. In the case of MRI scanners (0.2 T-7 T) the frequency range involved is 8 MHz -300 MHz, thus our attention is focused on low and resonant ranges where measurements are carried out. For low range measurements, the use of a small electrostatically shielded loop (0.3 m diameter) is suggested. The measurement chain is composed of a signal generator (plus amplifier) connected by a shielded cable to the transmitting small loop, the receiving small loop and a field detector (preferably a spectrum analyzer although selective receivers are often used). Hence, in the low-range, a magnetic field shall be generated by a current in the small loop antenna, driven by a continuous wave (CW) signal, the transmit loop is placed outside the shield, whereas the receiver one is inside the enclosure and the distance between them shall be 0,60 m (edge to edge) plus the thickness of shielding barrier. The antennas shall be coplanar in a plane perpendicular to the wall, ceiling, or other surface being measured, as sketched in Fig.2. Fig. 2. Coplanar horizontal loop antennas (with or without shield) 3 Around single-panel entry doors, tests should be conducted for 14 loop positions around the door. In areas where shielding enclosure construction is electrically non-uniform (for TABLE I MATHEMATICAL SHIELDING RELATIONSHIPS (ADAPTED FROM IEEE STD 299™-2006) Frequency Range 9 kHz-20 MHz (Extendable Down to 50 Hz) 20 MHz -300 MHz Measured Quantities |H1|, |H2| (or |V1|, |V2|) |E1|, |E2| 300 MHz-18 GHz (up to 100 GHz) P1, P2 Linear Units μA/m, μT (or μV) μV/m W Shielding Effectiveness (dB) Logarithmic Units (All Frequencies) SE=|E1|(dB)-|E2|(dB); SE=|H1|(dB)-|H2|(dB) SE=|V1|(dB)-|V2|(dB); SE=P1 (dB) - P2 (dB) example air vent, or connector panel, panel-to-panel seams), measurements shall be conducted in a similar manner. Regarding the resonant range, in the case of 20 MHz-100 MHz frequency range the use of biconical antennas is suggested, whereas λ/2 dipoles are recommended for frequencies at or above 100 MHz. Again, the basic measurement procedure consists of positioning the transmitting antenna outside the shield and the receiving one inside the shield and measuring the magnitude of the largest received signal. Measurements shall be repeated with antennas in both horizontal and vertical polarization. For both dipole and loop antenna the same method is proposed. The antennas shall be positioned at a distance of 2 m. The reference reading will be the maximum reading among different position readings of both transmitting and receiving antennas. When the shield is inserted, the receiving antenna must be swept in position (throughout the interior shield) and in polarization to obtain the largest detector response that will be recorded for determining the worst SE besides a series of transmit antenna positions shall be selected to cover the overall surfaces. This second measure is schematically represented in figg.3a and 3b, for horizontal and vertical transmitting antenna configuration respectively. It is also suggested the calculation of the lowest resonant frequency of the shielded enclosure. > REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) < 4 7. it does not foresee indications for measurement uncertainty evaluation; 8. in many cases the use of dipole and biconical antenna is impossible/useless to apply in MR environments because of space and logistic limitations1; 9. it never refers to the use of nonmagnetic devices/materials; 10. the resonant frequency calculation should be mandatory in this contest because is very often close to the operating frequency of the machine, and it is currently never calculated. Fig. 3a. Measurement set up, horizontal transmitting antenna configuration resonant range (20-100 MHz), distance = 2 m Fig. 3b. Measurement set up, vertical transmitting antenna configuration resonant range (20-100 MHz), distance = 2 m IV. NEW STANDARD PROCEDURE The inadequacy of the standard currently used came out from our surveys on SE measurements in several MRI units. A strong gap between the approaches suggested in the standard and the practical execution emerged, and in no case the IEEE Std 299-2006 has been rigorously applied. As matter of fact, even if the suggested procedures are detailed and rigorous, a lot of independence is left to the testing organization. In particular, the following issues came out and these are key starting points for our investigations: 1. the standard is too general (wide frequency ranges, since it refers to more complex cages); 2. it suggests very time-consuming procedures going in conflict with the short time machine availability; 3. it requires high dynamic range (which implies high performing instruments); 4. in resonant range test locations for the door, or other crucial points are not specified; 5. the set up showed in figg.3 shall be repeated for each position of the transmitting antenna, with a high waste of time and often not outstanding information; 6. it does not indicate specific set-up parameters (signal level, RBW) nor Pass/Fail limits; In this section we suggest guidelines which borrow from the IEEE Std-299 (2006) but more suitable for MRI RF cabins, highlighting only the procedures that we found inadequate. Thus, if not specified, the details of procedures are assumed to be the same as those suggested in the standard. We here describe a vademecum for a typical measurement at 63 MHz. For measurements at additional frequencies (i.e. 15 MHz, 127 MHz, etc.) one can apply the same proposed procedure. Before starting the test, preliminary activities and checks are required. They can be resumed as follows: 1. preparation of a test plan which shall include: dimensions of the cage, test frequencies, test locations, limits of SE to pass/fail the test, brand and model of the MR device, including the static B0 field value, a detailed description of the instrument set up and other notes; 2. removal of metallic equipment that is not a normal part of the enclosure while ancillary equipment (blower fans, carts, coils) normally present inside the enclosure shall remain in place during the test; 3. restriction of people within the shielded enclosure (if required, a maximum of two (2) persons should be allowed); 4. visual check inside and outside the MR room, in order to identify inaccessible surfaces and areas of leakage and repair them before measurements take place (cleaning of door contacts, replacement of damaged fingers); in addition, an ―electromagnetic check‖ should be done both inside and outside the magnet room in order to evaluate if possible radiofrequency interferences sources are present due to medical devices (RX tomography, Marconi therapy devices), light systems, electronic devices (switching power suppliers, digital thermometers etc.). 5. The range of interest shall be reduced to a restricted RF interval which is strictly linked to the MR apparatus. Thus, for the selection of test frequencies it is necessary to know the operating frequencies of the MR device and the bandwidth of receivers. The widest range that we can currently consider is about 8 MHz - 300 MHz. Given a B0 strength the corresponding Larmor frequency (ftest) is known (e.g. for a 1.5 T device ftest = 63.86 MHz) and measurement at this frequency is mandatory. However, other frequencies, around the operating ftest, can be established. As a matter of fact, we can reasonably believe that the shielding behavior of a surface is quite similar in this very narrow frequency range, so it is up to tester and to testing conditions > REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) < the decision of verifying further test frequencies as in case of a future new machine installation. In addition, variability of data can occur due to the cavity resonance effects. In effect, the operating frequency often belongs to the interval 0.8 fijk < ftest < 3 fijk, where fijk is the fundamental resonant frequency; hence we suggest the calculation of the fundamental resonant frequency (according to the formula presented in [3]) which shall be noted it in the final report: 2 2 1 i j k fijk 2 a b c 2 Where: µ, ɛ are the permeability and permittivity inside the enclosure a, b and c indicate the three dimensions of the enclosure in meters (length, height and width respectively) i, j, k, are positive integers 0,1,2,3… (and not more than one of i,j,k can be zero at the same time). Under ideal conditions, the resonant frequency in MHz is given by 2 2 i j k f ijk 150 a b c 2 and the lowest resonant frequency for the shielding enclosure, where c is the smallest dimension, is 2 i j f r f110 150 a b 2 Once established the frequencies of interest, the test points must be selected. Among the most critical locations in a MR Faraday cage, we suggest, as minimum testing surfaces, the door and the view-window (for each of them specific points are described later); these already give a meaningful idea of shielding behavior of the enclosure, then if further analysis are required, additional testing surfaces could be identified such as filter panel, blind panel, other surfaces containing visible penetrations. Consultation of technical documents of the structure is strongly recommended before selecting test locations since penetrations are often not visible and all surfaces under test shall be noted in the final technical report. Concerning the selection of test points another clarification is needed. Currently, the standard defines an accessible test location as a location that can be reached by a test antenna or probe without modifying a parent structure; in MRI units it often happens that it is not possible to position the transmit and the receiving antennas over a surface, or it is not possible to position them at the same height, or at a given distance. The instrumentation chain suggested by the IEEE standard of course remains valid (signal generator plus amplifier, Tx and Rx antennas, spectrum analyzer plus attenuator), however all instruments and auxiliary equipment (other probes, cables, tripods) must be nonmagnetic to access the magnet room. The instruments, whose operation directly affects the numerical value of the SE, shall be regularly calibrated before any measurement. Since it is a relative measurement it is 5 unnecessary to calibrate every piece of the equipment (each antenna), but it can be sufficient to trust on the spectrum analyzer calibration. The dynamic range (DR) of the spectrum analyzer must be adequate. It is the difference between the reference level and the minimum discernable signal above the noise floor (3 dB or more above the noise floor). We do not require a high DR because, if the spectrum analyzer is calibrated, it is sufficient to reduce the maximum readable level on its display (―Reference Level‖), in combination with an appropriate RBW selection, when executing the closed-door measurement. With the reduction of noise floor the visibility of even a small peak, without requiring a high-performance instrument is guaranteed. The measurement procedure requires two different stages: in the first one, the reference level shall be evaluated (i.e. the measure acquired with transmitting and receiving antennas one in front of the other, without the shielding surface), then the measure of the signal received with TX and RX over a shielding surface follows. The measure of reference level should be done (almost) in the same conditions of the other measurement, to guarantee a more accurate SE evaluation. Besides, the reference level should not be less than -10 dBm. Notice that the setting parameters do not affect significantly the peak value, but can only have influence on noise level, that is why a small RBW is suitable especially for low level signals. For measurements in the radiofrequency range we strongly encourage to prefer dipole antennas to biconical ones, since they are more appropriate for this kind of measurements (test frequencies are fixed, a broadband analysis is not required). It is currently on study the possibility of using custom magnetic loop antennas as probes for SE measurements purpose also in the resonant range. The potential substitution of the biconical (or dipoles) antenna would produce numerous advantages for MRI SE measurements, object of this work. The EM fields should be generated by power applied to Tx antenna, and received from a similar Rx antenna. A continuous wave signal shall be used to drive the Tx antenna with a power adequate to maintain a suitable DR. Over the door, we retain that 5 points are sufficient (instead of 14 as indicated by the standard, fig. 4a and 4b). These positions allow the evaluation of shielding behavior at the center of the door and along the weakest lines. The same configuration could be adopted for the view-window. About the distance between Tx and Rx antennas we suggest to get the measurements at two distances: 0.60 m and 2.50 m (instead of 2 m only) plus the thickness of the shield, for 1.5 T MRI for a better characterization of the enclosure when different field sources are considered. For each selected distance (and at each frequency), the antennas shall be in the same configuration (i.e. the same polarization and position) over the shielding surface. For the calculation of the SE, we can use the following equation SE=Pref-Pshield [dB] > REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) < being Pref and Pshield the values obtained in absence and in presence of the screen respectively. However, we also suggest to note the maximum value found on the spectrum analyzer, which corresponds to minimum SE (SEMIN), this one can be obtained by varying the receiver antenna position and polarization SEMIN=Pref-Pshield,MAX [dB] In this way, for each point we shall have two Shielding Effectiveness values which shall be noted it in the final report. 6 3. Set up of instruments. Then, for each selected frequency: 4. measurement of the reference level (in each antenna configuration and for each selected distance). For each point on the surface: 5. Measurement of the minimum SE (by varying the position and the orientation of the receiving antenna); 6. Measurement of effective SE (with antennas aligned and in the same polarization). Then: 7. Analysis of collected data (to establish the test results (passed/failed)); 8. Calculus of average SE as a qualitative indicator of the global shield performance; 9. Test Report. We hope that our contribution can help in adopting a simple and practical approach when facing SE assessments of MRI cabin which could allow to increase the repeatability and reproducibility of measurements thus, allowing quantitative comparisons and performance assessments. These are nowadays unavoidable tasks, even more so in public or private health structure. a b Fig. 4 Single panel entry door (view-window). Measurement points and antenna orientations suggested in reference [4] (a), in this study (b). Finally, once all data have been collected, one can mediate all the results obtained at a specific frequency and distance if a unique value of SE is desired for each shielding surface. At the end of each measurement a test report shall be fullyformed. The information to be reported are almost the same of those indicated in the standard [4]. Conclusions about the test data (pass/fail) shall also be included in the final report, according to specific attenuation performances required by the MR producers. Typical required values are > 80 dB. V CONCLUSIONS Although the design and building of MRI shielding enclosures undergo affirmed best-practice, there are still many aspects which cannot be taken for granted. The rapid and increasing development of MR devices calls for modernization of hosting environments too. About Shielding Effectiveness measurements methods, we believe that important goals have been achieved in this work: first of all, we highlighted the inadequacy of current normative references and attempted a possible path towards an effective standardization that we retain an urgent task because this problem deals with quality, safety and safeguard. In conclusion, the present proposal suggests a novel approach to SE measurements in MRI environments in a narrow range of frequencies of interest; we can summarize it with the following steps: 1. Preliminary check procedures; 2. Test plan; ACKNOWLEDGMENT The authors gratefully acknowledge the research support from INES S.r.l. Company. REFERENCES [1] L. J. Erasmus, D. Hurter, M. Naudé, H. G. Kritzinger, S. Acho ―A short overview of MRI artefacts‖ SA J of Radiology, pp 13-17, August, 2004 [2] E. Pusey, D. D. Stark, R.B. Lufkin, R. W. Tarr, R.K.J. Brown, W. N. Hanafee, M.A. Solomon, ―Magnetic resonance imaging artifacts: Mechanism and clinical significance‖, RadioGraphics, vol. 6, n.149, pp.891-911 September, 1986 [3] M.J.Lee, S. Kim, S.A. Lee, HT Song, Y.M. Huh, DH Kim, SH Han, JS Suh, ―Overcoming Artifacts from Metallic Orthopedic Implants at High-Field Strenght MR Imaging and Multidetector CT‖ , RadioGraphics 27, pp. 791803, 2007 [4] IEEE Std-299-2006. ―IEEE Standard method for measuring the effectiveness of electromagnetic shielding enclosures‖. Institute of electrical and Electronics Engineers (IEEE), Piscataway, NJ, Feb.28, 2007 [5] J. H. Duyn: ―The future of ultra-high field MRI and fMRI for study of the human brain‖, NeuroImage 62, pp. 1241–1248,2012,doi: 10.1016/j.neuroimage.2011.10.065 [6] IEEE100. The Authoritative Dictionary of IEEE Standards Terms, 7th ed.New York:IEEE Press.2000 [7] MIL-STD-285, Method of Attenuation Measurements for Enclosures, Electromagnetic Shielding, for Electronic Test Purposes, U.S. Department of Defense (June 1956) Luciano Mirarchi Chartered Engineer Member of Institution of Engineering and Technology, has been MR Support Specialist in Siemens Healthcare . He joined the University Campus BioMedico in Rome as contract professor and he is teaching equipments for Nuclear Medicine, CT and MR. Author of many publications and 2 books with focus on MR, CT and HTA. > REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) < Valentina Giaquinto received the Master Degree in Biomedical Engineering (summa cum laude) from University of Naples, Federico II, on July, 2017. Her thesis project concerned quality and safety aspects in MRI environments (with particular focus to shielding effectiveness measurements). It was supervised by Professor Rita Massa and Luciano Mirarchi and received the Italian Electrotechnical Committee award ―Best Thesis Degree 2017‖. During the internship c/o INES Srl, her main activities were ordinary maintenances on Faraday Cages for MRI and related technical reports. She currently works for Altran Italy, a multinational company operating in different engineering fields and she is involved in electromagnetic compatibility and human exposure to EM fields analysis in railway sector. Sergio Silvestri is full professor of Measurement for the Departmental Faculty of Engineering at University Campus Bio-Medico of Rome. The scientific activity is mainly focused on measurements and measurement systems for clinical diagnostics and monitoring, in particular: on non-invasive biomedical instrumentation, contactless measurement of respiratory parameters, novel methodologies for obtaining temperature maps by imaging instrumentation such as TC, US and RM. He is Head of the Clinical Engineering Service of University Hospital Campus Bio-Medico of Rome and carries out researches on performance assessment and quality evaluation of medical instruments. The scientific activity scientific activity is resumed in a monograph, a book chapter, more than 50 print publications, more than 70 contributions to national and international conferences, one international patent pending, one Italian granted patent on a novel non-invasive cardiac output measurement method, two collaborations in books ―Fondamenti di Ingegneria Clinica‖ Vol 1 and 2. Rita Massa received the Physics degree (summa cum laude) and Ph.D. degree in Electronic and Computer Science from the University of Naples Federico II, Naples, Italy. In 1992, she became a Researcher, and in 2001, a Professor of Electromagnetic Fields of the University of Naples Federico II, with the Department of Electronic and Telecommunication Engineering (DIET), and, actually, the Department of Physics (DP) ―Ettore Pancini‖. She has been the chief of the Microwave Interaction Division (MIND) Laboratory (DIET), responsible for the Biophysics Laboratory (DP) and actually she is the chief of Non Ionizing Radiation Laboratory (DP). She is currently the Director of the ―Interuniversity Center for the Study of Interactions between Electromagnetic Fields and Biosystems (ICEmB)‖. Her research interests are is in the framework of the interactions of electromagnetic fields and materials, dealing with the biological effects of electromagnetic fields, electromagnetic dosimetry/exposure assessment, therapeutic and industrial applications of electromagnetic fields, and nondestructive testing of materials. She was and is currently involved in research projects, in cooperation with small, medium and/or great enterprises, for which she played and is currently playing the role of scientific coordinator/WP Leader. She has coauthored more than 200 papers published in peer-reviewed journals, proceedings, and conference papers. 7