CALCULATION PROCEDURE for VARIABLE AREA FLOWMETERS ACCORDING TO VDI/VDE 3513 PART 1 Create by : / Order no. : 0.0010 Last modification : Lucia / 29/5/2019 CONE AND FLOAT DATA Type Cone size Cone Float shape Float material Density of float Diameter of float Mass of float Alpha/Ru-List : VA40 : 25 : N 21.09 : A III 21 guided : Stainless steel : 7.94 g/cm³ : 28.27 mm : 130.2 g : KVC_AIIIN_ARU PHYSICAL DATA OF FLUID Origin fluid Density Density(Operation) Viscosity Temperature Pressure Compr. factor : : : : : : : methane (CH4) 1.67 kg/m³ operation 0.00167 kg/l 0.01092 mPa.s 80.0 °C 3.5 bar gauge 1.0 Reference fluid Density Density(Operation) Viscosity Temperature Pressure Compr. factor : Luft / Air : 1.293 kg/m³ standard : 0.001205 kg/l : 0.0181 mPa.s : 20.0 °C : 1.013 bar abs. : 1.0 Ruppel No. 10*E-3 alpha/M : : 0.0074713 0.0067295 Ruppel No. 10*E-3 alpha/M : : FLOWLIST ORIGIN FLOW [m³/h] norm. REFERENCE FLOW [m³/h] norm. ( norm. 1,013 bar abs; 0°C ) 54.0 50.0 45.0 40.0 35.0 30.0 25.0 20.0 15.0 10.0 5.4 16.930 15.680 14.120 12.570 11.000 9.444 7.884 6.320 4.762 3.199 0.000 0.01458 0.0079228