English Signet 2850 Conductivity/Resistivity Sensor Electronics (PVDF) *3-2850.090-1* 3-2850.090-2 Rev. B 01/15 Operating Instructions 3-2850-51, 3-2850-52 with ¾-in. adapter Description • Signet 2850 Conductivity/Resistivity Sensor Electronics provides either a two-wire 4 to 20 mA output or Digital (S3L) format. • The 4 to 20 mA output models provide eight ranges for each electrode cell constant, plus the ability to invert each range. • The EasyCal feature allows the devices to automatically recognize standard conductivity test solution values for simple field calibration. 3-2850-61, 3-2850-62, 3-2850-63 with Universal adapter • The Conductivity Sensor provided with integral systems will have its custom cell constant information programmed into the electronics at the factory to provide a 2% sensor accuracy. See page 6 for details. Table of Contents 3-2850-51-XXV(D), 3-2850-52-XXV(D) PVDF Integral System Description ............................................................................................ 1 Dimensions ........................................................................................... 1 Warranty Information ............................................................................ 2 Product Registration ............................................................................. 2 Safety Information................................................................................. 2 Specifications........................................................................................ 2 Conductivity Sensor Dimensions .......................................................... 3 Operating Range Chart......................................................................... 3 In-Line Installation................................................................................. 4 Tank Installation .................................................................................... 4 4 to 20 mA Wiring ................................................................................. 5 Digital (S3L) Wiring................................................................................ 5 Dual Input Wiring (Dual Digital (S3L) Output)........................................ 5 Cell Constant Selection ........................................................................ 6 Range Selection for 4 to 20 mA Output ................................................ 7 Calibration............................................................................................. 8 EasyCal ................................................................................................ 8 Dual Input Calibration ........................................................................... 8 Maintenance ......................................................................................... 9 Troubleshooting .................................................................................... 9 Electronic Certification ........................................................................ 10 Ordering Information ........................................................................... 12 Dimensions 2850-51, -52 Integral mount • • • • • English Deutsch Français Español Italiano 95 mm 3.74 in. 2850-61, -62, -63 Universal mount 95 mm 3.74 in. 85 mm 3.34 in. 82 mm 3.24 in. Warranty Information Specifications Refer to your local Georg Fischer Sales office for the most current warranty statement. General Compatible Electrodes ............. Any Signet Conductivity/ Resisitivity electrode (2818-2823, 2839-2842) Temperature Compensation ..... PT-1000 RTD (2% per °C) Pure Water Compensation ....... Auto-switching when using 0.01 cell and raw conductivity value < 0.5 μS EasyCal Automatic recognition of the following conductivity values: • 146.93 μS, 1408.8 μS, 12856 μS (@ 25 °C) (Test solutions per ASTM D1125-95) • 10 μS, 100 μS, 200 μS, 500 μS, 1000 μS, 5000 μS, 10,000 μS, 50,000 μS, 100,000 μS (@ 25 °C) All warranty and non-warranty repairs being returned must include a fully completed Service Form and goods must be returned to your local GF Sales office or distributor. Product returned without a Service Form may not be warranty replaced or repaired. Signet products with limited shelf-life (e.g. pH, ORP, chlorine electrodes, calibration solutions; e.g. pH buffers, turbidity standards or other solutions) are warranted out of box but not warranted against any damage, due to process or application failures (e.g. high temperature, chemical poisoning, dry-out) or mishandling (e.g. broken glass, damaged membrane, freezing and/or extreme temperatures). Product Registration Thank you for purchasing the Signet line of Georg Fischer measurement products. If you would like to register your product(s), you can now register online in one of the following ways: • Visit our website www.gfsignet.com. Under Service and Support click on Product Registration Form • If this is a pdf manual (digital copy), click here Safety Information 1. 2. 3. 4. 5. 6. Depressurize and vent system prior to installation or removal. Confirm chemical compatibility before use. Do not exceed maximum temperature/pressure specifications. Wear safety goggles or faceshield during installation/service. Do not alter product construction. When using chemicals or solvents, care should be taken and appropriate eye, face, hand, body, and/or respiratory protection should be used. Caution / Warning / Danger Indicates a potential hazard. Failure to follow all warnings may lead to equipment damage, injury, or death Electrostatic Discharge (ESD) / Electrocution Danger Alerts user to risk of potential damage to product by ESD, and/or risk of potential of injury or death via electrocution. Personal Protective Equipment (PPE) Always utilize the most appropriate PPE during installation and service of Signet products. Pressurized System Warning Sensor may be under pressure, take caution to vent system prior to installation or removal. Failure to do so may result in equipment damage and/or serious injury. Note / Technical Notes Highlights additional information or detailed procedure. 2 2850 Conductivity/Resistivity Sensor Electronics Power • 12 to 24 VDC (4 to 20 mA output) • 5 VDC ±5% regulated, 3.0 mA max. for Digital (S3L) output (Reverse polarity and short circuit protected) Digital (S3L) Output................. Serial ASCII, TTL level 9600 bps Accuracy • Conductivity .......................... ±2% of reading • Temperature .......................... ±0.5 °C Resolution • Conductivity .......................... 0.1% of reading • Temperature .......................... < 0.2 °C Update Rate • Single Input models .............. < 600 ms • Dual Input models ................. < 1.2 s Current Output Field-selectable ranges. Factory set span: • 0.01 cell (2818, 2819, 2839) .. 4 to 20 mA = 0 to 100 μS • 0.10 cell (2820, 2840) ........... 4 to 20 mA = 0 to 1000 μS • 1.0 cell (2821, 2841) ............. 4 to 20 mA = 0 to 10,000 μS • 10.0 cell (2822, 2842) ........... 4 to 20 mA = 0 to 200,000 μS • 20.0 cell (2823) ..................... 4 to 20 mA = 0 to 400,000 μS Max. Loop Resistance .............. 50 Ω @ 12 VDC 325 Ω @ 18 VDC 600 Ω @ 24 VDC Accuracy ................................... ±2% of output span Resolution................................. 7 μA Update Rate ............................. <600 ms Error Indication ......................... 22 mA Available data via Digital (S3L) Output • Raw conductivity • Calibrated conductivity • Calibrated temperature-compensated conductivity • Temp. Error Indication ........... Open input and out-of-range diagnostics for temperature or internal electronic error. Environmental Material ..................................... PBT Enclosure Rating ...................... NEMA 4X/IP65 Operating Temperature............. -10 to 85 °C (14 to 185 °F) Storage Temperature ................ -20 to 85 °C (-4 to 185 °F) Relative Humidity...................... 0 to 95%, non-condensing Shipping Weight ....................... 0.75 kg (1.75 lb.) Standards and Approvals • CE, RoHS Compliant • Manufactured under ISO 9001 for Quality, ISO 14001 for Environmental Management and OHSAS 18001 for Occupational Health and Safety. China RoHS (Go to www.gfsignet.com for details) This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation. Conductivity Sensor Dimensions 73.7 mm 2.90 in. 59.3 mm 2.33 in. 35.8 mm 1.41 in. 3-2839 41.9 mm 1.65 in. 21.5 mm 0.85 in. 26.8 mm 1.13 in. 3-2841, 3-2842 3-2840 4.6 m (15 ft) cable (std) T1 = 50 mm (2.0 in.) T2 = 64 mm (2.5 in.) 152 mm (6 in.) ¾ in. NPT 73.66 mm (2.9 in.) 107 mm (4.2 in.) 152 mm (6.0 in.) 126 mm (5.0 in.) 2820-x 20.3 mm (0.8 in.) 49.5 mm (1.95 in.) 41.91 mm (1.65 in.) 32.39 mm (1.28 in.) 2821-x 7.6 mm (0.3 in.) 12.7 mm (0.5 in.) 89 mm 3.5 in. 13 mm (0.5 in.) 12.7 mm (0.5 in.) 3-2819, 3-2820, 3-2821 Sanitary 3-2819, 3-2820, 3-2821 3-2818 203 mm 8.0 in. 3-2823 Operating Range Chart Cooling Tower Waste Waters Bottled Water UPW @ 25°C USP (20.0 Cell) (10.0 Cell) (1.0 Cell) (0.10 Cell) Deionization Regen. Chemicals Alkali Cleaners, Acids/Bases 2823 2822/2842 2821/2841 2820/2840 (0.01 Cell) 2818/2819/2839 0.010 μS 0.055 μS 0 0 1 μS 0 0.5 μS 10 μS 50 μS 100 μS 200 μS 1,000 μS . . 20 μS 10,000 μS 100,000 μS 400,000 μS 50,000 μS 200,000 μS 2850 Conductivity/Resistivity Sensor Electronics 3 In-Line Installation Tank Installation While the 2850-5X and 2850-6X electronics cannot be submerged, either model will accommodate tank installation. Select any electrode with a 5 m (15 ft) cable. The cable may be cut to length, but it CANNOT BE EXTENDED. Most of the Conductivity/Resistivity electrodes used with the 2850 can be installed into standard 3/4 in. (NPT or ISO 7-R3/4) fittings. The 2850-5X models are designed for integral systems, where the electronics are attached directly to the sensor. The preferred installation for in-line applications directs flow straight into the electrode. This configuration reduces the probability of entrapped air bubbles, and provides the best continuous sampling of the fluid content. If the electrode is mounted vertically in a tee, do not recess the electrode orifices inside the tee. Mounting upside down may help prevent air entrapment. An oversized tee or flow cell may be helpful for inline installations. At least 4 threads (ANSI B1.20.1) must be engaged to meet the pressure rating as published in the specifications. Use the 2850-6X models for in-line installations where the electronics must be separated from the electrode. Conductivity electrodes with sanitary flange fittings must be installed using this option. 4 2850 Conductivity/Resistivity Sensor Electronics The universal adapter included with 2850-6X models can be attached to the top of a tank or mounted to a surface near the tank. Adapt the 2850-5X models to tank installations by attaching a clamp to the conduit. 4 to 20 mA Wiring 4 2 20 3 V2 4-20 mA Loop Output Sensr Gnd (SHIELD) 4 1 V+ Temp. IN (WHITE) 1 ON 2 3 CTS 9 VDC to 26 VDC 1 ON 4 3 CTS Maximum length of 4 to 20 mA loop is 300 meters (1000 ft) Signal RTN (BLK) SW1 D3 Signal IN (RED) Digital (S3L) Wiring 3-8900.621C I/O Module 3-8900.401-X The Digital (S L) output is compatible with the Signet 8900 Multi-Parameter Controller and the Signet 9900 Transmitter. 3 Refer to the wiring sections of the 8900 or 9900 manual to determine cable length limitations. Frequency Input 1 Frequency Input 2 OR S3L Input 2 S3L Input 1 1 +5VDC (Black) Freq. Input (Red) GND (Shield) +5VDC (Black) Freq. Input 2 (Red) S 3 L (Red) GND (White/Shield) +5VDC (Black) S 3 L (Red) GND (White/Shield) 2 3 4 5 6 7 3 Gnd 8 9 2 Data 10 1 V+ Sensr Gnd (SHIELD) 11 IMPORTANT: Provide Earth Ground as indicated by symbol. Temp. IN (WHITE) 4 CTS + - 12 ON 2 (if applicable) 3 Analog Output 1 Analog Output 2 (if applicable) + - 1 13 Signal RTN (BLK) 14 SW1 D3 SHLD V+ GND DATA Note: The 2819-2823 and 2839-2842 sensors can connect directly to the 9900 via the 9900's optional Direct Conductivity/Resistivity Module 3-9900.394 (159 001 699). Signal IN (RED) 9900 S3L Input er DC Pow ge Loop Volta White 3-9900.395 le Modu H COMM Red Black Dual Digital (S3L) Input Wiring 3-8900.621C 3 4 5 6 3 9 + - 1 12 13 (if applicable) + - 4 1 V+ 3 (if applicable) Analog Output 2 Input 2 2 2 Data 10 11 Analog Output 1 CTS 3 Gnd 8 4 7 2 1 2 Input 1 1 If the Dual Input model is used with a single sensor, connect the sensor to Input 1 and set all of the Input 2 switches to OFF (OPEN). S 3L Input 1 CTS Sensors may be the same or different cell constants. Frequency Input 2 OR S3L Input 2 +5VDC (Black) Freq. Input (Red) GND (Shield) +5VDC (Black) Freq. Input 2 (Red) S 3 L (Red) GND (White/Shield) +5VDC (Black) S 3 L (Red) GND (White/Shield) ON Dual Input is available on Digital (S L) output model 3-2850-63 only, for use with the Signet 8900 Multi-Parameter Controller. Frequency Input 1 ON I/O Module 3-8900.401-X 3 14 SW1 D3 Dual Input applications usually include one sensor with a smaller cell constant and one sensor with a larger cell constant. Place the larger cell constant on Input 1. Sensor 2 Sensor 1 This allows periodic recalibration of the larger cell by simply turning Input 2 OFF (OPEN). IMPORTANT: Provide Earth Ground as indicated by symbol. 2850 Conductivity/Resistivity Sensor Electronics 5 Cell Constant Selection Single Input 2850 Electronics (2850-51, 2850-52, 2850-61, 2850-62) 2 1 CTS ON 1 2 3 4 SW1 SW3 or SW4 CTS ON 1 D3 • Single input models use only SW3. • Use SW3 #1-3 to select the cell constant for the first sensor. • Use SW3 #4 to disable the PT1000 Temp Compensation function in the 2850 (as required for USP applications). 3 NOTE: To work correctly with the 9900, the 2850 must be set for the custom cell constant or the actual electrode cell constant and the 9900 set for a 1.0 cell constant. 2 3 Custom Cell Constant 4 SW3 CTS ON Closed Input 1 • Recommended: Make all switch settings before supplying power. Switch changes made after supplying power will take 15 to 20 seconds before becoming effective. Open 1 2 4 3 0.01/cm-1 Integral 2850 Systems (Single Input) • The 2850-51-XX and 2850-52-XX are shipped from the factory with the custom cell constant and temperature offset programmed into the single-input electronics. • SW3 will be preset to "Custom cell constant". • No additional modifications are required. Replacing Integral System Electrode or Electronics CTS ON Closed Open 1 2 4 3 0.1/cm-1 • When installing a new conductivity electrode or replacing System Electronics, the SW3 switches must be configured to select the appropriate cell constant. • See Single Input section above. • Leaving SW3 in the custom cell constant configuration will effect accuracy. • Use calibration standards to calibrate (see page 8). CTS ON Closed Open 1 2 4 3 1.0/cm-1 Closed Open 1 2 1 CTS ON 1 2 3 10.0/cm-1 4 Closed CTS ON CTS ON Open D3 1 2 3 4 SW3 1 2 3 4 1 SW4 Input 1 Input 2 6 20.0/cm-1 CTS Open 1 #1 C C O C O C #1 C C O C O C 2 #2 C O C C O O #3 C O O O C C #4 O O O O O O Cell Constant Custom 0.01 cm-1 0.1 cm-1 1.0 cm-1 10.0 cm-1 20.0 cm-1 #4: (O) Open = Input 2 INACTIVE (C) Closed = Input 2 ACTIVE 4 3 Examples 1.0 cell constant, temp comp Active CTS ON 2850 Conductivity/Resistivity Sensor Electronics 4 3 Closed SW4 = Cell Constant and Input 2 #4: (O) Open = Temp Comp ACTIVE (C) Closed = Temp Comp INACTIVE 2 ON SW3 = Cell Constant and Temperature Compensation Cell Constant Custom 0.01 cm-1 0.1 cm-1 1.0 cm-1 10.0 cm-1 20.0 cm-1 CTS ON NOTE: Dual Input applications usually include one sensor with a smaller cell constant and one sensor with a larger cell constant. Place the larger cell constant on Input 1. This allows periodic recalibration of the larger cell by simply turning Input 2 OFF (OPEN). #4 O O O O O O 4 3 3 • Recommended: Make all switch settings before supplying power. Switch changes made after supplying power will take 15 to 20 seconds before becoming effective. #3 C O O O C C 2 SW1 • Switch banks SW3 and SW4 are used to select the cell constant of the electrodes. • Use SW3 #1-3 to select the cell constant for the first sensor. • Use SW3 #4 to disable the PT1000 Temp Compensation function in the 2850 (as required for USP applications). This disables the function for BOTH INPUTS. • Use SW4 #1-3 to select the cell constant for the second sensor. • Set SW4 #4 to OPEN to disable Input 2. #2 C O C C O O CTS ON Dual Input 3-2850-63 Electronics Closed 1 2 3 4 Open 1.0 cell constant, temp comp Inactive CTS ON Closed 1 2 3 4 Open Range Selection for 4 to 20 mA Output CTS ON 3 4 3 2 2 1 1 • The Range selection switch bank (SW2) provides eight range selections for each cell constant. • Each range can be inverted, making a total of 16 range options. • Select a range from the table below and set SW2 as indicated. SW2 D3 10.0 Cell 20.0 Cell 0 to 2 μS 0 to 20 μS 0 to 200 μS 0 to 400 μS 2 to 0 μS 20 to 0 μS 200 to 0 μS 400 to 0 μS 2 to 10 MΩ 0 to 5 μS 0 to 50 μS 0 to 500 μS 0 to 1000 μS 10 to 2 MΩ 5 to 0 μS 50 to 0 μS 500 to 0 μS 1 000 to 0 μS O 0 to 2 MΩ 0 to 10 μS 0 to 100 μS 0 to 1 000 μS 0 to 2 000 μS C 2 to 0 MΩ 10 to 0 μS 100 to 0 μS 1 000 to 0 μS 2 000 to 0 μS O 0 to 1 μS 0 to 50 μS 0 to 5 000 μS 0 to 5 000 μS 0 to 10 000 μS C 1 to 0 μS 50 to 0 μS 500 to 0 μS 5 000 to 0 μS 10 000 to 0 μS O 0 to 5 μS 0 to 100 μS 0 to 1 000 μS 0 to 10 000 μS 0 to 20 000 μS C 5 to 0 μS 100 to 0 μS 1 000 to 0 μS 10 000 to 0 μS 20 000 to 0 μS O 0 to 10 μS 0 to 200 μS 0 to 2 000 μS 0 to 50 000 μS 0 to 100 000 μS C 10 to 0 μS 200 to 0 μS 2 000 to 0 μS 50 000 to 0 μS 100 000 to 0 μS O 0 to 50 μS 0 to 500 μS 0 to 5 000 μS 0 to 100 000 μS 0 to 200 000 μS C 50 to 0 μS 500 to 0 μS 5 000 to 0 μS 100 000 to 0 μS 200 000 to 0 μS O O 0 to 100 μS 0 to 1 000 μS 0 to 10 000 μS 0 to 200 000 μS 0 to 400 000 μS O C 100 to 0 μS 1000 to 0 μS 10 000 to 0 μS 200 000 to 0 μS 400 000 to 0 μS 0.01 Cell #2 #3 #4 C C C O 10 to 20 MΩ C C C C 20 to 10 MΩ O C C O O C C C C O C C O C O O C O O C C C O C C O O C O O C O C O O C O O O O O O NOTE: 0.10 Cell 1.0 Cell SW2 Switch Setting #1 3 4 1 2 3 4 SW3 CTS ON The electrode installed is the 3-2840-1V, with a 0.10 cell constant. The required output is 4 to 20 mA = 0 to 100 μS. Set SW2 switch bank to C-C-O-O. To invert range (4 to 20 mA = 100 to 0 μS), close switch #4. 2 SW2 SW1 Example (refer to shaded selections of chart): • • • • CTS ON 1 Resistivity Ranges in BOLD Switch #4 inverts the loop output range (O) Open = 4 to 20 mA (C) Closed = 20 to 4 mA 2850 Conductivity/Resistivity Sensor Electronics 7 Calibration • All 2850 Conductivity/Resistivity Sensor Electronics are factory calibrated for maximum out-of-the-box accuracy. • The EasyCal feature allows calibration to be performed at the sensor. SW2 SW4 Input 2 ON 2 3 2 4 CTS b 1 1 2 c f 4 3 CTS SW3 Input 1 1 d. e. f. 3 ON c. SW2: 4 to 20 mA output range selection switch bank Power and Output terminal block: 4 to 20 mA or Digital (S3L) Outputs SW3 and SW4: Electrode Cell Constant selection switch banks D3: LED indicator SW1: EasyCal Push-button Input 1 & Input 2: Conductivity Electrode connections 1 a. b. ON 2 3 CTS a 4 The 2839-2842 Conductivity sensors come with a certificate of calibration. This certificate provides information regarding the actual cell constant and temperature offset as tested and verified according to NIST standards. This information has been programmed into the 2850 electronics as its default factory value. If required, the sensor can be recertified by contacting the GF Signet repair department. NOTE: When the electrode is changed from its original factory provided electrode or the application requires a cell constant change, reset SW3 or SW4 switches (see page 6). Then perform a wet calibration with calibration standards. e SW1 d D3 EasyCal EasyCal is a single-point calibration system. During this procedure, if the measured value is within ±10% of any of the test values listed below, the 2850 will automatically recognize the test value and calibrate the output to that value. NOTE: The first step (Reset) is recommended each time an electrode is replaced, but is NOT necessary upon initial installation or periodic calibration. 1. Reset the 2850 Sensor to factory calibration: Set all switches on SW3 (and SW4 for Dual Input) to OPEN. Apply power, wait at least 3 seconds, then press and hold SW1 until the LED (D3) comes on steady then goes off again (approx. 10 seconds). When the LED goes off, release SW1; reset is complete. Reset SW3 and SW4 to the proper settings. 2. Place the electrode/sensor assembly into the conductivity test solution appropriate to your operating range. Shake the electrode to dislodge any air bubbles visible on the surface of the electrode. • 146.93 μS, 1408.8 μS, 12856 μS (per ASTM D1125-95) • 10 μS, 100 μS, 200 μS, 500 μS, 1000 μS, 5000 μS, 10,000 μS, 50,000 μS, 100,000 μS. Allow at least 2 minutes for the electrode response to stabilize. 3. Press and hold SW1 for approximately 8 to 10 seconds. During this time, the LED will come on steady then go back off. (If the LED blinks several times rapidly, the calibration was not successful. See the troubleshooting section). Calibration is complete. Return the system to service. Dual Input Calibration Dual Input 3-2850-63 Electronics • The 2850-63 can only be calibrated if both inputs are used, or if Input 2 is disabled. • To disable Input 2: Set SW4, #4 to OPEN. • Place each sensor into a test solution that represents the operating range of that sensor. When SW1 is pressed, the 2850 will calibrate both inputs simultaneously. • Dual Input applications usually include one sensor with a smaller cell constant and one sensor with a larger cell constant. Place the larger cell constant on Input 1. This allows periodic recalibration of the larger cell by simply turning Input 2 OFF (OPEN). 8 2850 Conductivity/Resistivity Sensor Electronics Maintenance The 2850 requires no periodic maintenance. • Coatings on the electrode may cause slow response or drift. • Clean metallic surfaces with a mild detergent and a non-abrasive brush or cotton swab. Troubleshooting LED and Output Condition Possible Causes Suggested Solutions • SW3 and SW4 set to all OPEN (factory setting). • No electrode installed. • SW3 (and SW4) not set correctly. • System not ready. • • • • During EasyCal, the LED blinks rapidly for 4 seconds. • Measured value of the test solution is outside the ±10% tolerance. • The electronics or the electrode is defective. • Use fresh test solutions and restart the calibration. • Replace the electronics or the electrode. After completing calibration procedure, the output values are inaccurate. • Insufficient time allowed for electrode stabilization during calibration procedure. • Test solutions are contaminated. • EasyCal performed with temp comp Inactive, then temp comp Active for measurement. (or vice versa) • Recalibrate and wait at least 2 minutes after placing electrode in solution before pressing SW1. • Use fresh test solutions and restart the calibration. The output of the 2850 indicates a value that is off by a factor of 10 (e.g. 10x, 100x the correct value) • The cell constant selection on the 2850 (SW3 or SW4) does not match the actual input. • Match the 2850 switch settings to agree with the sensor being used. Current Out: LED off, current output is 22 mA (S3L) out: "Check Sensor" Normal for new system. Install electrode. Set SW3 and SW4 correctly. Retry (recycle power). 2850 Conductivity/Resistivity Sensor Electronics 9 Electronic Certification Signet offers conductivity simulators in five different values. These tools enable the user to validate the performance of the electronics independently of the electrode. This requirement is defined by ASTM D 1125-95 (Standard Test Methods for Electrical Conductivity and Resistivity of Water) which is commonly used for USP 24 applications. Cond. - Resist. Simulator @ 25°C For use with Signet 2850 Conductivity NIST Traceable 3-2850.101-1 (159 001 392) 1.0 μS ±0.1% 3-2850.101-2 (159 001 393) 2.5 μS ±0.1% 3-2850.101-3 (159 001 394) 10.0 μS ±0.1% 3-2850.101-4 (159 001 395) 18.2 MΩ ±0.1% 3-2850.101-5 (159 001 396) 10.0 MΩ ±0.1% Before using the Simulator: Before using the simulators, disconnect all electrodes and reset the 2850 Electronics to factory calibration: 1. Set all switches on SW3 (and SW4 for Dual Input) to OPEN. Turn the power on, wait at least 3 seconds 2. Press and hold SW1 until the LED (D3) comes on steady then goes off again (approx. 10 seconds). 3. When the LED goes off, release SW1. Reset is complete. To validate Input 1: 1. Turn the power off. 2. Set SW3 to accept a 0.01 cell constant. Leave SW4 in the all-OPEN position. 3. Install the simulator into the 2850 at the Input 1 conductivity electrode terminals. 4. Connect the 2850 output terminals to the appropriate display instrument. Turn the power on. • The display should indicate a conductivity value within ±2% of the simulator value. • If the output is x10 or x100 of the correct value, recheck step 2. The 2850 is not set to accept a 0.01 cell constant. • If the correct value is displayed, the electronics are calibrated to ASTM D 1125 specifications. Remove the simulator and reset SW3 to the correct cell constant before reinstalling the electrode. • If the correct value is not displayed, contact the factory. To validate Input 2: 1. Turn the power off. 2. Set SW4 to accept a 0.01 cell constant. 3. Install the simulator into the 2850 at the Input 2 conductivity electrode terminals. 4. Set SW3 switches to the all-OPEN position. 5. Turn the power on and repeat step 4. Remember to reset all switches to their correct position before resuming normal operation. 10 2850 Conductivity/Resistivity Sensor Electronics Notes 2850 Conductivity/Resistivity Sensor Electronics 11 Ordering Information Signet 2850 Conductivity/Resistivity Sensor Electronics *All versions include EasyCal Mfr. Part No. 3-2850-51 Code 159 001 398 Description 2850 Sensor Electronics with Digital (S3L) Output and ¾ inch adapter 3-2850-52 159 001 399 2850 Sensor Electronics with 4 to 20 mA Output and ¾ inch adapter 3-2850-61 159 001 400 2850 Sensor Electronics with Digital (S3L) Output and Universal adapter 3-2850-62 3-2850-63 159 001 401 159 001 402 2850 Sensor Electronics with 4 to 20 mA Output and Universal adapter 2850 Sensor Electronics with Dual Inputs, Dual Digital (S3L) Outputs and Universal adapter (for use with 8900 only) 3-2850-51-39V 3-2850-51-40V 3-2850-51-41V 3-2850-51-42V 159 001 818 159 001 819 159 001 820 159 001 821 Integral 2850 system, Digital (S3L) output, 0.01 cell, PVDF NPT threads Integral 2850 system, Digital (S3L) output, 0.1 cell, PVDF NPT threads Integral 2850 system, Digital (S3L) output, 1.0 cell, PVDF NPT threads Integral 2850 system, Digital (S3L) output, 10.0 cell, PVDF NPT threads 3-2850-51-39VD 159 001 822 Integral 2850 system, Digital (S3L) output, 0.01 cell, PVDF ISO threads 3-2850-51-40VD 3-2850-51-41VD 3-2850-51-42VD 159 001 823 159 001 824 159 001 825 Integral 2850 system, Digital (S3L) output, 0.1 cell, PVDF ISO threads Integral 2850 system, Digital (S3L) output, 1.0 cell, PVDF ISO threads Integral 2850 system, Digital (S3L) output, 10.0 cell, PVDF ISO threads 3-2850-52-39V 3-2850-52-40V 3-2850-52-41V 3-2850-52-42V 159 001 826 159 001 827 159 001 828 159 001 829 Integral 2850 system, 4 to 20 mA output, 0.01 cell, PVDF NPT threads Integral 2850 system, 4 to 20 mA output, 0.1 cell, PVDF NPT threads Integral 2850 system, 4 to 20 mA output, 1.0 cell, PVDF NPT threads Integral 2850 system, 4 to 20 mA output, 10.0 cell, PVDF NPT threads 3-2850-52-39VD 3-2850-52-40VD 3-2850-52-41VD 3-2850-52-42VD 159 001 830 159 001 831 159 001 832 159 001 833 Integral 2850 system, 4 to 20 mA output, 0.01 cell, PVDF ISO threads Integral 2850 system, 4 to 20 mA output, 0.1 cell, PVDF ISO threads Integral 2850 system, 4 to 20 mA output, 1.0 cell, PVDF ISO threads Integral 2850 system, 4 to 20 mA output, 10.0 cell, PVDF ISO threads Parts and Accessories 3-9000.392-1 159 000 839 Liquid-tight connector kit, 1 set, 1/2 in. NPT 3-9000.392-2 3-2850.101-1 3-2850.101-2 3-2850.101-3 3-2850.101-4 3-2850.101-5 Liquid-tight connector kit, 1 set, PG 13.5 Plug-in NIST-traceable recertification tool, 1.0 μS Plug-in NIST-traceable recertification tool, 2.5 μS Plug-in NIST-traceable recertification tool, 10.0 μS Plug-in NIST-traceable recertification tool, 18.2 MΩ Plug-in NIST-traceable recertification tool, 10.0 MΩ 159 000 841 159 001 392 159 001 393 159 001 394 159 001 395 159 001 396 Georg Fischer Signet LLC, 3401 Aero Jet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057 For Worldwide Sales and Service, visit our website: www.gfsignet.com • Or call (in the U.S.): (800) 854-4090 For the most up-to-date information, please refer to our website at www.gfsignet.com 3-2850.090-2 Rev. B 01/15 English © Georg Fischer Signet LLC 2015