Brandenburg University of Technology Faculty 4 Environmental Sciences and Process Engineering Institute Mechanische Verfahrenstechnik Bachelor Thesis Nr.42 Development and optimization of an electrostatic corona charger BO YANG 3519809 Supervised by: Prof. Dr. -Ing. Ulrich Riebel Dipl. -Ing. Andreas Groll 09.05.2016 Contents Declaration of Academic Integrity Acknowledgement List of symbols 1. Introduction ...................................................................................................... 1 2. Theory Background ......................................................................................... 3 3. 4. 2.1 Conception of the corona discharge and corona wind ................................... 3 2.2 Charging mechanism ................................................................................... 4 2.3 Onset voltage ............................................................................................... 6 2.4 The voltage limit and spark .......................................................................... 8 2.5 Positive and negative discharge .................................................................... 8 2.6 The relationship of current and voltage ...................................................... 10 2.7 Speed of ionic wind ................................................................................... 12 2.8 Particle concentration ................................................................................. 14 2.9 Coagulation ............................................................................................... 15 Experiment (1) Set-up .................................................................................... 17 3.1 Current and voltage characteristics ............................................................. 19 3.2 Voltage and velocity characteristic ............................................................. 22 3.3 The influence of polarity ............................................................................ 25 3.4 Influence of the diameter of needle ............................................................ 27 3.5 Influence of distance between electrodes .................................................... 30 3.6 Influence of mesh shape ............................................................................. 34 3.7 Influence of amount of needle .................................................................... 38 3.8 Conclusion ................................................................................................. 41 Experiment (2) Set-up .................................................................................... 43 4.1 Corona discharge texting............................................................................ 44 4.2 Different positions of setting corona discharge ........................................... 48 4.3 Conclusion ................................................................................................. 59 5. Future Study ................................................................................................... 61 6. Summary ........................................................................................................ 62 Reference ........................................................................................................ 64 Appendix………………………………………………………………………...66 List of Tables Table 1 Different diameters of needle(Negative) .............................................. 28 Table 2 Different distance between electrodes(Negative) ................................. 31 Table 3 Different distance between electrodes(Positive) .................................. 31 Table 4 Different mesh shape........................................................................... 34 Table 5 Comparison of total number concentration in different conditions ....... 45 Table 6 Comparison of total number concentration in different positions ......... 51 Table 7 Comparison of total number concentration in faster velocity of aerosol 57 List of Figures Figure 1 Positive corona discharge device .......................................................... 4 Figure 2 Particle charge limits ........................................................................... 6 Figure 3 The process of positive corona discharge before starting with onset voltage ....................................................................................................... 9 Figure 4 The process of positive streamer .......................................................... 9 Figure 5 The process of negative corona discharge before starting with onset voltage ..................................................................................................... 10 Figure 6 The process of negative streamer ....................................................... 10 Figure 7 positive corona discharge current-voltage characteristic curve ........... 11 Figure 8 Characteristic curve of velocity-voltage ............................................. 14 Figure 9 Experiment set-up design ................................................................... 19 Figure 10 The comparison between Theoretical and Experimental value in I-U 21 Figure 11 The comparison between Theoretical and Experimental value in V-U curve ........................................................................................................ 23 Figure 12 The Current-Velocity curve ............................................................. 24 Figure 13 The comparison between different polarities in I-U curve ................ 25 Figure 14 The comparison between different polarities in V-U curve ............... 26 Figure 15 The comparison between different polarities in V-I curve ................ 26 Figure 16 The comparison between different diameter of wire in I-U ............... 29 Figure 17 The comparison between different diameter of wire in V-U ............. 29 Figure 18 The comparison between different distance between electrodes in VU(Negative) ............................................................................................. 32 Figure 19 The comparison between different distance between electrodes in VU(Negative) ............................................................................................. 32 Figure 20 The comparison between different distance between electrodes in VU(Positive)............................................................................................... 33 Figure 21 The comparison between different distance between electrodes in VU(Positive)............................................................................................... 33 Figure 22 Corona discharge with fan................................................................ 34 Figure 23 Different shape of mesh: Mesh 1(1), Mesh 2(2), Mesh 3(3), Mesh 4(4) ................................................................................................................ 35 Figure 24 The comparison between different shape of mesh in V-U(Negative) 37 Figure 25 The comparison between different shape of mesh in V-U with initial velocity(Negative) ................................................................................... 37 Figure 26 The comparison between different shape of mesh in I-U(Negative) .. 38 Figure 27 The comparison between different amount of needle in I-U ............. 40 Figure 28 The comparison between different amount of needle in V-U ............ 40 Figure 29 The comparison between different amount of needle in V-I ............. 41 Figure 30 The corona discharge designing ....................................................... 42 Figure 31 Experiment set-up with FMPS ......................................................... 44 Figure 32 The comparison of different amount of needle in number concentration with 25kv discharge ................................................................................. 45 Figure 33 Number concentration under different conditions ............................. 46 Figure 34 Mass concentration under different conditions ................................. 47 Figure 35 Position (a) (b) (c) ............................................................................ 48 Figure 36 Total number concentration in position(a) (b) (c) with 0kv,15kv,25kv discharge.................................................................................................. 50 Figure 37 Number concentration with 15kv discharge under different positions. ................................................................................................................ 52 Figure 38 Mass concentration with 15kv discharge under different positions ... 53 Figure 39 Number concentration with 25kv discharge under different positions. ................................................................................................................ 53 Figure 40 Mass concentration with 25kv discharge under different positions ... 54 Figure 41 Increasing the velocity of aerosol in (d) and (e) ................................ 55 Figure 42 Total number concentration in position (d) (e) with 0kv,15kv,25kv discharge.................................................................................................. 56 Figure 43 Number concentration with high-speed aerosol ................................ 58 Figure 44 Mass concentration with high-speed aerosol..................................... 59 Figure 45 Two stages corona discharge device ................................................. 61 Declaration of Academic Integrity I promise that the thesis contains no material which has been accepted for the award of any other degree or diploma in any institutes of higher learning and that, to the best of my knowledge and belief, the thesis contains no material previously published or written by another person, except when due reference is made in the text of the thesis. I understand that to do so would mean that I had committed plagiarism and that it is my responsibility to be aware of the College’s regulations on plagiarism and their importance. Signed: Date: Acknowledgement I would like to express my appreciation to my home university (the University of Shanghai for Science and Technology) for giving me the chance to study in Germany. During this year, I have learned many things, not only the courses I studied here but also the different cultures and the way to face difficulties. I would like to express my acknowledgment to Prof. Dr. Riebel for supporting this topic for me. I am an exchange student from China, and this is the first time I study in Germany. Everything was difficult for me at beginning especially when I was searching for the Bachelor thesis, but Prof. Dr. Riebel kindly gave me the chance to do my thesis in his project. The topic is absorbing for me, and I can do the research as a real German student, which is beyond my expectation before. From Prof. Dr. Riebel, I learned what was the true “Spirit of Germany,” for his spirit of rigid, squareness, patience and preciseness, which will have a huge influence in my future. I would like to thank my supervisor Dipl.Groll, who helps me all the time, though I cannot speak German. He tried his best to communicate with me in English, which gives me a lot helps during my thesis. He is always humorous and cheerful, which gave me much comforts when I first entered the laboratory. I feel very relax to communicate him, and also, he is very patient for resolving all my questions even some of them are very simple. I would also like to thank Students in Lab: Zhe Lei, Yan Ma, Shouhui Li, Jiawei Lu, who gave me many supports of my study and life in Germany. Finally, I would like to thank my parents in China, who always support me and whatever difficulties I may meet, they can always give me the strength to go through it. List of symbols ππ : Mean thermal speed of ions[m/s] D: Diffusion coefficient of particles d:Diameter of needle[m] ππ : Diameter of particle[m] e: Coefficient of restitution E: Electrical field strength[V/m] πΈπ€ : Onset corona field on surface of needle[V/m] πΈπ : Corona field strength on the surface of needle[V/m] πΉπ : field strength[π] G: Distance between needle and plate[m] I: Current[A] J: Current density[A/m²] K: Geometric factor πΎ0 : Coagulation coefficient[m³/s] πΎπΈ : Electrostatic constant of proportionality ππ : The concentration of ions. P: Pressure of cross section[N/m²] q: Charge on single particle[C] r: radius of curvature of needle[m] S: Area of sectional surface[m²] ππ : Area of wire[m²] v: Flow velocity[v/m] V: Power supply[V] π0 : Onset voltage[V] ππ : Mobility of the ions[m²/sV] μ: Electron mobility[m²/sV] ρ: field density [C/m] ρπππ : Air density[kg/m³] π : Relative permittivity 1. Introduction The system of solid particles or liquid droplets suspended in a gas is called aerosol [1], the sizes of particles are range from 0.001μm~100μm. The resources of an aerosol can be classified into two types: natural resource and man-made resource. Natural resources are usually generated by the eruption of a volcano, dust carried by the wind, etc. [2]. The manmade resource is mainly relief of gasses from industries or burning fossil fuels, etc. [2]. Aerosol can cause harm to human beings and be threaten to health. Aerosol can be stably suspended in the air, sometimes could last for several years. The elimination of aerosol in a natural way is mainly relied on coagulation [2]. However, it is not far more enough compared to the much faster generation rate of an aerosol of human activities. Thus, cleaning devices are necessary to be applied, which can be achieved by particles charging, particles which are charged will move under electrical forces if there is a collector, such as a tube or a plate, connecting to ground, charged particles will move to the collector and eventually be collected. The way to charge particles are flame charging, static electrification, diffusion charging and field charging, but to have a highly charged particle, only the corona discharge fulfilled it and is usually applied [1]. Corona discharge can be achieved by a corona discharge device connected to a high voltage, this method can be widely used in air cleaning or cooling system. There is a point-to-plane type of corona discharge device which is designed as a metal needle connected to the high voltage and a plate with mesh used for collecting particles. Several factors can have influences on the process of generating the corona wind. This paper aims to set up the experiment to explore how these factors, such as 1 polarity of discharge, the diameter of the needle, the amount of needle, etc. For further practice, the optimized device will be installed in a container, and aerosol will be injected to create a range of particle sizes. By using the FMPS device, the data of number concentration, mass and volume will be collected and analyzed, the result will test how the device work. 2 2. Theory Background 2.1 Conception of the corona discharge and corona wind There are several ways to have charging particles. However, only corona discharge can satisfy the condition that high concentration unipolar ions will be produced, thus to charge aerosols [2]. During the process of corona discharge, the wind will be generated without any fans, this can be called corona wind. Corona wind is a phenomenon which is conducted by the collisions between ionized particles and neutral particles, and it can also be called ionic wind [3]. To give the information about this process, an example of the point-and-plate type of positive corona discharge is described in Fig 1. Corona electrode side is connected to high voltage apply, this is usually can be achieved as a needle. Collector electrode will be linked to ground, usually is a flat plate. Thus, an electrical field is established. As the voltage rising gradually and reaching the onset voltage, the electrons around corona electrode will be ionized [4], because electrons and positive ion will be separated from air molecule due to the extremely uneven electrical field. The electron will move to needle, and positive will move towards field as the effect of electrical force, theses ionized molecules will eventually be collected by the collector electrode. During the motions of positive ions, some natural air molecule will be collided and move with ionized molecules along the field together. As a result, there will be a constant flow between the electrodes [5]. 3 Figure 1 Positive corona discharge device Recent researches showed that corona wind can be mainly utilized in flow cleaning process due to its higher efficiency [1], other researchers intended to design a cooling device by corona wind, which is more silent and superior speed because of no machine parts being existed [6]. Alternatively, some researchers were focused on dry technic [7]. 2.2 Charging mechanism Field charging is that in a strong electrical field, neutral particle in the field will be collided by ions and be charged until it reaches the saturation charge and no ions can collide it. The rate of charge can be given by: 3π πΈπ 2 ππΎ πππ ππ π‘ n(t) = (π+2)(4πΎ ππ)(1+ππΎπΈ πΈ πππ ππ π‘ πΈ 4 ) (2.1) Where π is relative permittivity, E is electrical field strength, ππ is particle diameter, πΎπΈ is electrostatic constant of proportionality, e is coefficient of restitution, ππ is the mobility of the ions, usually is about 0.00015m²/Vβs [450cm²/stVβ s], ππ is the concentration of ions. Usually when ππ > 1013 /π3 , the sufficient time is about 3s to have the saturation charge 3π πΈπ 2 ππ = (π+2)(4πΎππ ) (2.2) πΈ Another charging is diffusion charging, which does not need any external field, because of Brownian motion of the ions and particles, collisions will happen and charge the neutral particles. This can be described by: π πΎπ ππΎπΈ ππ ciΜ π 2 ππ π‘ πΈ 2ππ n(t) = 2πΎπ ln(1 + π2 ) (2.3) Where ππ is the mean thermal speed of ions (ππ = 240π/π at standard conditions) . In a corona discharge, a field charging can be established by the needle and plate, both of the corona and corona wind are generated by field charging. Particles which are larger than 1.0μm are dominated by field charging while for particles less than 0.1μm are charged by diffusion charging. Thus, corona discharge usually combines two types of charging which is called combined charging. The limit of combined charging can be shown in Fig 2. 5 Figure 2 Particle charge limits [1]. 2.3 Onset voltage As the voltage supply growing up, gas particles around corona electrode will be ionized, and the corona discharge will be realized. The initial voltage that corona discharge starts to work is called onset voltage. This field strength around the tip of needle can be described by Peek’s law which is for wire-to-plate system [5]: πΈπ€ = πΈ0 (1 + 2.62β10−2 √π ) (2.4) Where πΈπ€ [π/π] is the onset corona field on surface of needle, r[m] is the radius of curvature of needle, πΈ0 = 3.31 β 106 [π/π]. The formula of onset voltage can be given as[8]: πΈπ = 2π π ln( 6 4πΊ ) π (2.5) Where V[π] is the voltage supply, πΈπ [π/π] is the corona field strength on the surface of needle,πΊ [π] is the distance between needle and plate. CombineοΌ2.4οΌwithοΌ2.5οΌwhen the field strength reaches to onset field strengthοΌ πΈπ = πΈπ€ , the onset voltage is given asοΌ π0 = πΈπ€β π ln( 2 4πΊ ) π οΌ2.6οΌ Where π0 [π] is the onset voltage. The high voltage applied to the needle causes an intensive active electrical field around the tip of the needle, the field is so strong that is stronger than the field of air around it and causes air to be ionized. This onset field is the maximum field to have corona discharge, of formula (2.6), this onset field largely depends on the size of needle, the minimum value is πΈ0 = 3.31 β 106 [π/π] for a needle which has an infinitely large diameter and is impossible to happen. Whether the size of wire is large or small, the onset electrical field must exceed more than πΈ0 ,but as the diameter increasing, the field needed to start corona is getting smaller, so for formula (2.6), the onset voltage is related to onset field πΈπ€ , a small onset field can reduce the onset voltage, what is more, the onset voltage is also linked to needle size and the distance between electrodes, a smaller needle and closer distance can reduce the onset voltage. But latter in Chapter 3, the actually results are completely opposite to formula (2.6), which shows that a sharper needle have a quick discharge. In conclusion, to start corona discharge quickly, π0 should be as small as possible. The shape of needle and distance between electrodes will play the main role. 7 2.4 The Voltage limit and spark Due to the mechanism of corona discharge, when positive particles or electrons move to the needle, they can be soaked into the needles and generate a pulse current. As power supply growing up higher enough, a current called streamer discharge current is produced, as this current reaching the collector electrode, space will cause the breakdown, and corona discharge will be covert to spark discharge. This dielectric will shift to the conductor, causing a huge jump of current. If the power is not high enough, a stable voltage was difficult to be sustained and causes the interruption of spark discharge, so only a small voltage is generated between electrodes, this will be a disadvantage to charge particles, which should be avoided. Normally the breakdown field of air is 1.3 × 106 . 2.5 Positive and negative discharge The mechanism of charging particles under positive and negative discharge are entirely different, also, will have influences on the process of charging particles. This is mainly due to the different mobility, where the mobility of electron is much higher than the mobility of positive ion [9]. For positive corona, electrons will move around needle while positive ions move towards to the collector electrode, however, due to the slower mobility, many positive ions are still around the needle. As a result of the interaction of electrical field between positive and negative particles, the electrical field will be reduced around the needle, so it will be harder to create corona discharge and thus need a higher onset voltage. This shows in Fig 3. 8 Figure 3 The process of positive corona discharge before starting with onset voltage While in the movement of positive streamer, because electrons move quickly to the needle, the electrical field in front of the stream is stronger, which is easier to get to the collector electrode. Thus, the stream is getting stronger and longer, which can reach negative electrode faster. As a result, the breakdown between electrodes will happen, which shows in Fig 4. So the breakdown voltage of positive discharge is lower. Figure 4 The process of positive streamer For negative discharge, positive ions are around needle while electrons are moving towards the electrical field, due to the higher speed of electrons, the intensive field around needle is much stronger than the field where the interaction is happening in the cross section. The process is given in Fig 5. This causes corona discharges easily to be established and lower onset voltage. 9 Figure 5 The process of negative corona discharge before starting with onset voltage In the process of negative streamer, when electrons move in front of positive ions, new positive ions will be attracted, and new electrons will go to the positive electrode, so a new negative streamer is generated. However, due to the weak electrical field in the head of the streamer, the process of producing a new streamer is slower, so the negative stream is harder to reach the positive side, which can be seen in Fig 6. Thus, the breakdown voltage is higher than positive discharge. Figure 6 The process of negative streamer 2.6 The relationship of current and voltage Due to the theory of Townsend [10], the current during the process of corona discharge can be described as: 10 I=KμVΔV=KμVΔV οΌ2.7οΌ Where I[A] is the current, K is the geometric factor, V[V] is the power supply, μ[m²/Vs] is the ion mobility, and ΔV[V] can be described as: ΔV=π − π0 (2.8) Where π0 [π] is the onset voltage. Combine with (2.7) and (2.8), the current can be given by: I = KμV(π − π0 ) (2.9) Thus, the current-voltage characteristic curve can be described by Fig 7, after the Current(μA) onset voltage, the current has a relationship with voltage as I~π 2 . Onset voltage Voltage(kv) Figure 7 positive corona discharge current-voltage characteristic curve 11 2.7 Speed of ionic wind From the introduction in Chapter 2.1, the ionic wind is generated by the collision of charged particles and neutral air particles, the experiments in this essay show that the velocity always increases with growing voltage, to find this relationship of speed and voltage, some theoretical calculations can be given. We assume that in the electric field, the field is even so field strength is always the same, the force that particles get from the field can be described by the formula below [11]. πΉπ = qE (2.10) Where πΉπ [π] is field strength, q[C] is the charge on single particle, E[V/m] is field strength. If the amount of ionic particles in a volume unit is n, thus the total field fore is F = nqE (2.11) Also, no means the full amount of ionic particles in a volume unit, which can be described as charge density ρ [8]. Thus, the force is F=ρE (2.12) Where ρ [C/m] is field density. It should be noticed that Current can be described as I = Jππ = ρμππ E 12 (2.13) Where J[A/m²] is current density, μ[m²/sV] is ionic mobilityοΌππ [m²] is area of wire, so by dividing (2.12) and (2.13), the relation between force and current can be described as πΌ F = ππ π (2.14) We assume that the wind speed is ultimately generated by the field force if the field is in an insulating tube and there are no friction losses on the wall of tube, the relationship between strength and flow speed is described by Bernoulli Equation: πΉ 1 P = π = 2 ρπππ π£ 2 (2.15) Where P[N/m²] is the pressure of cross sectionοΌS[m²] is area of sectional ππ surface, ρπππ [π3 ] is air density, v[m/s] is the flow velocity. Thus by connecting (2.9), (2.14) and (2.15)οΌthe velocity can be described as: v = √(2 · ρ 1 πππ μππ S ·) V(π − π0 ) (2.16) Thus, from the formula (2.16), the characteristic curve of velocity-voltage has a linear relationship v~V, which is given by Fig 8. 13 Velocity(m/s) Oneset voltage Voltage(kv) Figure 8 Characteristic curve of velocity-voltage 2.8 Particle concentration The properties of corona discharge have been introduced before, the purpose of corona discharge in this paper is aim to clean aerosols in the air. Therefore, to find out how corona discharge can reduce aerosols in experiments, there are several properties of the particle should be known. Mass concentration is the most valuable goods and also commonly measured by measurement, which is the mass of particulate matter in a unit volume of aerosol [1]. The unit is usually g/m³, mg/m³, orμg/m³. Number concentration is another important property, which is the number of particles in a unit volume of aerosol. The unit is usually number/cm³ or number/m³. 14 2.9 Coagulation The result of the experiment shows that the number concentration of particles will decrease as time going by when injecting aerosol into an airtight container having a particular volume. This is due to coagulation of aerosols. Coagulation is a process of collision where particles collide with each other and coagulate into larger size particles. As a result, the number concentration of particles will decrease while larger particles are getting more. There are two types of coagulation, one is thermal coagulation caused by Brownian motion, and another is kinematic coagulation which is caused by external forces such as gravity or electrical force [1]. Due to coagulation, the number of particles in a unit volume will decrease as time passing, to describe this change, the formula is given as ππ ππ‘ Where ππ ππ‘ = −πΎ0 π 2 (2.17οΌ is change rate of number, πΎ0 [π3 /π ] is the coagulation coefficient, which is: πΎ0 = 4πππ π· (2.18) Where D is the diffusion coefficient of particles, ππ (m) is the diameter of particle. In a standard condition, πΎ0 can be simply quale to πΎ0 = 3.0 × 10−16 πΆπ [π3 /π ]. To have the initial number of particles (π0 ), a relationship of initial number and the number at time t is given by: π N(t) = 1+π 0πΎ 0 0π‘ 15 (2.19) Therefore, with an initial number, the calculation of collect efficiency is available. Also, if the condition is not standard, a corrected coagulation coefficient K is K = πΎ0 π½ (2.20) Whereβcan be found on references. Thus, as a consequence, the particles number centration will reduce because of coagulation while the mass concentration remained the same. 16 3. Experiment (1) Set-up Firstly, to test the properties of corona discharge, an experimental stage has established the outline of the operation can be shown in Fig 9. A corona electrode is set inside a metal tube, where the total length of the tube is 44.50cm, out diameter is 11.30cm, inner diameter is 10.25cm. The distance between the corona electrode and mesh of one side of the tube is 9.34 cm, the position of wire in corona discharge can be adjusted, thus changing different diameter, amount of needle, or removing the needle to change the gap distance between electrodes, these can be conducted. It is important to cover the needle with pipe so that only a tip of needle of one side can generate corona wind. The mesh is installed on the side of the tube where is near the needle as the collector electrode, there are several types of mesh can be changed, so comparing effects by changing different mesh is possible. Both of the tube and mesh are linked to the ground as the collector electrode. A power supply (PHYWE HV-POWER SUPPLY) is connected to corona discharge, which has a range from 0KV to 25KV, the cables can be shifted to different polarities so that the experiment can be conducted in the negative and positive side. A voltammeter (PeakTeach 070 DMM) is linked to measuring the current, the measurement range will be chosen as the range from 0~2000 μA. To gauge the velocity of corona wind, a little hole that can allow a LAMBRECHT KLIMATOLOGISCHE MESSTECHNIK (flow measurement) to stretch in and be set inside the middle of the tube to have the maximum speed. The precision range is 0.01m/s, there are options to by reading average values in 1s, 10s, the 30s, and 60s, to have the accurate velocity and faster measure speed, we at this moment close to record average velocity in 30s. In the field of corona discharge, there will be lots of charged particles passing through the electrodes. Thus, the electrical field could have an 17 influence on recording results or even cause damages to measurement. Also, if the measurement is set behind the mesh, the uneven surface of the mesh may cause the loss of speed, so it is necessary to avoid them by establishing measurement on the back side of corona discharge for 36.50 cm from the mesh. 18 Figure 9 Experiment set-up design 3.1 Current and voltage characteristics From chapter 2.6, we assumed that the current follows the Townsend ’s theory by having a quadratic relationship with voltage, to verify this prosumer, an experiment is established. The diameter of the needle is chosen to be 0.50mm, the distance between electrodes is adjusted for 5.12cm, the mesh has square units with the length of 2.25mm. The needle is connected voltage within positive polarity, power supplies are improved from 0kv to 25kv, increasing 1KV every time, the values of current and velocity are recorded after every 30s when the voltage is changed. After finishing the recording, the result will be compared with the theoretical calculations in which the formula is given by (2.4). The value of the geometric factor k and electronic mobilityμis hard to obtain because of the restraints of measurement equipment, but by reading another reference who made their experiments at ambient conditions set the mobilityμto 0.0002[γ‘ /Sv][12 ], and k could be 0.000112. The result can be shown in Fig 10. Also, the theoretical onset voltage is also calculated to verify it which can be found in Chapter 2.3. According to formula (2.4), the onset electrical field can be calculated by: πΈπ€ = πΈ0 (1 + 2.62β10−2 √0.00025 the onset voltage is: 19 ) = 8.79 × 106 [V/m] π0 = 8.79×106 β0.00025 ln( 2 4×0.0512 ) 0.00025 = 7.37[ππ] While as the records in this experiment, the onset voltage is about 7 kV, which is slightly less than the theoretical value, there are no precise explanations to verify the reasons, but the needle does not have an ideal shape and surface, this may cause the electrical field intensity to be uneven and thus influence the onset voltage, in addition, the power losses in the circuit may also play a role, the residences of current measurement or the wire will separate the voltage supply. Thus, the actual voltage provided on the needle will be reduced and make the onset voltage delay. However, the tip of the needle is not a regular shape, this causes a sharper tip, and the real radius of the needle may be less, which will reduce the onset voltage. Last but not least, the defect of recording method may also cause the onset voltage too large because the voltage is increased 1kv every time so the actual onset voltage may range from 6kv~7kv, which is neglected by observation. 20 140 120 Current(uA) 100 80 60 40 20 0 0 5 10 15 20 25 Voltage(kv) Experimental vlaue Theoretical value With theotrical onset voltage Figure 10 The comparison between Theoretical and Experimental value in I-U However, if using the same onset voltage as being recorded in the experiment, the difference between theoretical calculation and experimental result is not very big. However, it is still not acute because geometric factor k and electronic mobilityμare still uncertain, and only a guess that both the theatrical calculations and experimental result have a very similar increasing trend that the current shows a square relationship with voltage. Despite the dramatic increasing on current when increasing voltage supply from 19kv~23kv, the observed phenomenon including noise and slight lighting, this may be caused by spark discharge. When the gas of the space between electrodes has breakdown, the conductor will become good driver. Thus, residence primarily decreases and causes current having a sharp growth. In conclusion, both of the theoretical onset voltage and I-U curve are not far from the actual values, which can verify the formulas can be suitable for the corona discharge. 21 3.2 Voltage and velocity characteristic Follow by the result in Chapter 3.1, the speed of corona wind is also recorded as the voltage supply increasing, and the mode of recording is selected by using 30s mode to record the average values in 30s when the power increases by 1kv, the result shows in Fig 11. To verify the theoretical calculation in Chapter 2.7, the result of calculations is also given by graph below. The density of air is valued as 1.2kg/m³, the diameter of wire and channels are available to be measured as 0.0005m and 0.1025m, the onset voltage is assumed the same as the experimental value, the result will be compared with experimental records. Although both lines show an upwards trend, the Theoretical value is much higher than the value we measured. Even though they have the same onset voltage, the growing rate of theoretical line is much faster than the experimental line and the maximum velocity (23kv, 0.97m/s) in theoretical calculation almost double the value in laboratory records (23kv, 0.46m/s). 22 Velocity-Voltage 1 0.9 Velocity(m/s) 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 0 5 10 Voltage(KV) 15 Experimental value 20 25 Theoritical value Figure 11 The comparison between Theoretical and Experimental value in V-U curve There are several factors can cause this result. First of all, the air going through the tube is not an ideal flow and tube is not smooth which means there is always frictions on the tube, and this will have dynamic losses when the air is going through. Thus, the method of measuring has a significant influence on the little experimental value because of the position where the speedometer is set on the back of corona discharge for avoiding damages caused by charged particles. The long distance between speedometer and corona discharge will get the rapid loss. In addition, the theatrical calculation is based on the even field, which assumes that the whole field has an even field strength so that the velocities of flow passing the field are remained same While the real field strength is not even and the field strength around needle is much stronger than other position, so the highest speed wind will be generated near the needle, as the particles moving forward, the collisions of other particles may change their directions and if the diffusion discharge appeared, the field will become more complicated and 23 uneven. So the actual speed will bias more. Also, because the tube is metal, some charged particles will move to all sides and collected by the wall of the tube, which will also cause the velocity of straight line decreasing. Although there is a significant difference between theatrical and experimental values, the growing trends are the same as the velocity showing a linear relationship with voltage, which fits the conclusion in Chapter 2.7. 0.45 0.4 Velocity(m/s) 0.35 0.3 0.25 0.2 0.15 0.1 0.05 0 0 20 40 60 80 100 120 140 Current(uA) Experimental value Figure 12 The Current-Velocity curve By establishing the relationship between velocity and current, the velocity is also increasing with current growing up. However, the trend shows that there is a decrease in the growth rate, the velocity will firstly grow up sharply and then get more smooth. Finally, the velocity will be stable of particular value where spark discharge occurs at the same time, the spark limits the velocity so the corona wind cannot increase unlimitedly as voltage growing up. Thus, the experiment 3.1 and 3.2 show that both of current and velocity will increase as well as the voltage increasing while due to spark discharge when the voltage is high enough, there is always a limitation of corona wind speed. 24 3.3 The influence of polarity The mechanisms of negative and positive discharge are quietly different, by the established experiment, the same condition in Chapter 3.2 and Chapter 3.2, but recording results under both positive and negative discharge. The result is given in Fig 13 and Fig 14. Although the onset voltage under negative should be more advanced to happen, the difference is quite small which is recorded in the experiment. The onset voltage under negative discharge is 5kv while 7kv under positive condition, the negative discharge will have less onset voltage due to the faster mobility of electron, the more separation of electrons and positive particles around the tip of the needle will have more intensive electrical field strength and easier to occur corona discharge. As a result, negative discharge with lower onset voltage will have higher current and velocity which can be seen in Fig 15. 160 140 Current(uA) 120 100 80 60 40 20 0 0 5 10 15 20 25 30 Voltage(kv) Negative Positive Figure 13 The comparison between different polarities in I-U curve 25 0.6 Velocity(m/s) 0.5 0.4 0.3 0.2 0.1 0 0 5 10 15 20 25 30 Voltage(kv) Netagive Positive Figure 14 The comparison between different polarities in V-U curve 0.6 0.5 0.4 0.3 0.2 0.1 0 0 20 40 60 80 Negative 100 120 140 160 Positive Figure 15 The comparison between different polarities in V-I curve To summarize, the velocity within negative supply is slightly higher than positive supply, and the current and ionic wind are more stable, generally the negative discharge are more widely used in industries. However, it should be pointed out that spark is more frequently and easily generated in positive supply, so it is harder for corona wind to increase to a particular value in positive supply. On the other hand, during the process of negative corona discharge, there is ozone being generated in the negative supply, this is due to the negative electrons move from needle to everywhere where the electrons 26 will collide with an oxygen atom to become ozone, which is harmful to human health. So the positive discharge is better to be used under indoor condition. 3.4 Influence of the diameter of needle The Chapter 2.3 points out that the onset field and voltage are mainly depending on the diameter of the needle. To compare the influence of different shapes of needles, we tested three separate sizes of needles, which are separately d1=0.15mm, d2=0.50mm, d3=1.35mm. The electrodes gap remains 5.12cm, the result can be given in Fig 16 and Fig 17, which is conducted within negative voltage supply. With the onset voltage provided by Chapter 2.3, the theoretical onset voltage is U1=3.95kv, U2=7.37kv, U3=12.27kv respectively. While the onset voltage in the experiment is Ue1=5kv, Ue2=7kv, Ue3=8kv, which is given in Table 1. Despite U2 and Ue2 where there is a small difference between them, the needle with 0.15mm diameter shows an onset voltage delay, about 1kv. This may be due to the loss in the circuit, also. Due to the sheer size of needle, it is hard to keep straight and easy to have some curves on the wire, this sharp curves may also sharp enough to have an electrical field which will influence and reduce the field strength on the tip of the needle. However, the needle with 1.25mm diameter has a significant advance in onset voltage, up to 4kv. Despite the shape of the tip of the needle, the granulated surface may also play an important part, it has a larger surface, and there some protuberances generated by the oxidation over an extended period can be treated as another way to make the tip sharper, which enhances the intensity of field around the tip. As a result, there is the quite small difference in current when the voltage is given from 10kv~15kv, as voltage growing up, the difference is becoming more apparent, on the other hand, velocity is keeping stable because there is a linear relation between voltage and velocity. 27 Table 1 Different diameters of needle(Negative) Diameter(mm) Theoretical onset voltage (kv) 3.95 0.15 7.37 0.50 12.27 1.25 Onset Voltage (kv) Maximum Voltage (kv) Maximum Current (μA) Maximum velocity (m/s) 5 7 8 25 25 25 134 114 100 0.59 0.53 0.48 Nevertheless, though the experimental records do not completely submit to the formula, they have the same trend that smaller size of the needle can have an early onset voltage. As a result, an advanced onset voltage will soon start to generate ionic wind, and reach a higher speed. So the velocity shows a negative relation with a diameter of wired, due to the smaller diameter, the field could be more inventive, a faster process to have ionization. However, it should be pointed out that from formula πΈπ€ = πΈ0 (1 + 2.62β10−2 √π ), a larger diameter should reach the onset field early and easier to have corona discharge, but from experiment result and calculation of onset voltage, the result is completely opposite. There are no exactly explanations yet, but due to the mechanism of corona discharge, the intensive field around tip will cause a part strong field strength which can ionize the particles, a smaller diameter needle will have more intensive field to cause corona discharge. 28 160 140 Current(uA) 120 100 80 60 40 20 0 0 5 10 15 20 25 30 Voltage(kv) d1=0.15mm d2=0.50mm d3=1.25mm Figure 16 The comparison between different diameter of wire in I-U 0.7 0.6 Velocity(m/s) 0.5 0.4 0.3 0.2 0.1 0 0 5 10 15 20 25 Voltage(kv) d1=0.15mm d2=0.50mm d3=1.25mm Figure 17 The comparison between different diameter of wire in V-U 29 30 3.5 Influence of distance between electrodes According to Chapter 2.3, the onset voltage not only depends on the diameter of the needle but also decided by the distance between needle and collector electrode. So this experiment is based on the different distance G with the needle having 0.50mm diameter and mesh with 0.225mm length of square units. The distance G can be adjusted as moving the needle to change in corona discharge device. The experiment is conducted under both positive and negative discharge. The results are demonstrated by Fig 18 to Fig 19. In the negative side, the records show a trend that the onset voltage is smaller as distance reducing. When G=3.12cm, it is early to have corona discharge, and higher current and velocity, due to the shorter distance, the field strength is stronger, and there will be more particles being charged and enhance the intensity around the tip. So an advanced onset is possible, however, a short distance is much easier to occur breakdown because steamer can reach another electrode quickly with a short distance. As experiment demonstrating, even under negative condition, which shows in Table 2, the voltage cannot reach 25kv, and the current assumption is much higher, while the velocity is as same as it when distance is 4.12cm, but have lower current and more smooth curve. On the other hand, a longer length will cause the onset voltage delay and lower speed because far from the collector electrode, the field strength between electrodes is weak so it needs more field strength to have corona discharge. Also, a low electrical field is not beneficial for ionized particles to move towards the field, which also causes the velocity decreasing. However, the streamer travel in a longer distance will need more field strength to be robust enough to reach the collector electrode so that the breakdown will delay. 30 It is more obvious in positive supply, which is given in Table 3, when G=3.12cm, it only can reach 22kv and then break-down happened, the limitation of speed is much lower than it when G=4.12cm. Nevertheless, the difference is still not so significant when the distance ranges from 4.12cm~6.12cm. Especially when the voltage is lower than 15kv, the lines demonstrated in graphs almost converge to one line. It may due to the field is weak when the distance is longer than 3.12cm, which will cause the result not to be very obvious compared with each other. In collusion, as the distance decreasing, there is a trend that the onset voltage will be lower and can get a higher velocity, while if the distance was too close, it is much easier to have voltage break-down and reach to the limitation. In this experiment, the best distance with high efficiency may range from 3.12cm~4.12cm. Table 2 Different distance between electrodes(Negative) Distance G (cm) 3.12 4.12 5.12 6.12 Onset Voltage (kv) 4 5 6 6 Maximum Voltage (kv) 23 25 25 25 Maximum Maximum Current Velocity (μA ) (m/s) 195 0.54 151 0.54 130 0.51 124 0.47 Table 3 Different distance between electrodes(Positive) Distance G (cm) 3.12 4.12 5.12 6.12 Onset voltage (kv) 5 6 7 7 Maximum Voltage (kv) 22 25 25 25 31 Maximum Maximum Current Velocity (μA ) (m/s) 173 0.48 129 0.52 103 0.46 108 0.43 Negative 250 Current(uA) 200 150 100 50 0 0 5 10 15 20 25 30 Voltage(kv) g=3.12cm g=4.12cm G=5.12cm G=6.12cm Figure 18 The comparison between different distance between electrodes in VU(Negative) Negative 0.6 Velocity(m/s) 0.5 0.4 0.3 0.2 0.1 0 0 5 10 15 20 25 30 Voltage(kv) G=3.12cm G=4.12cm g=5.12cm g=6.12cm Figure 19 The comparison between different distance between electrodes in VU(Negative) 32 Positive 200 180 160 Current(uA) 140 120 100 80 60 40 20 0 0 5 10 15 20 25 30 Voltage(kv) g=3.12cm g=4.12cm G=5.12cm G=6.12cm Figure 20 The comparison between different distance between electrodes in VU(Positive) Positive 0.6 Current(uA) 0.5 0.4 0.3 0.2 0.1 0 0 5 10 15 20 25 30 Voltage(kv) G=3.12cm g=4.12cm g=5.12cm G=6.12cm Figure 21 The comparison between different distance between electrodes in VU(Positive) 33 3.6 Influence of mesh shape The collector electrode is designed to a metal mesh, that can allow the air to pass the mesh, and the ionic particles will be collected by the mesh, though there is no exact theory about how the shape of the mesh may have influences on corona discharge which is not mentioned before. The result of the experiment shows that mesh can play a major role in the velocity of the wind. In this experiment, there is four mesh with different shapes, which is given in Table 4 and Fig 23, the minimum unit size is measured as L1=1.00cm, L2=0.23cm, L3=0.17cm, R4=0.55cm. The diameter of needle and distance between electrodes remain 0.5mm and 5.12cm, the experiment is conducted under the negative and positive condition, also, a fan has been used in front of mesh. So there will be an initial velocity for discharge. The results are given by Fig 24 to Fig 25. Figure 22 Corona discharge with fan Table 4 Different mesh shape Mesh 1 1.00 Length(Radius) of unit(cm) Mesh 2 0.23 34 Mesh 3 0.17 Mesh 4 0.55 Figure 23 Different shape of mesh: Mesh 1(1), Mesh 2(2), Mesh 3(3), Mesh 4(4) The result shows that though there is a huge difference between the shapes of mesh. Whether the negative or positive side has almost the same value of current, thus the mesh has nearly non-influences on current because the current is mainly determined by the needle and the distance between electrodes, also the media filled with an electrical field, but the designing of collector electrode will not change these factors. However, the shape of mesh can change velocity a lot, when there is no fan working, the Fig 24 shows that the Mesh 4 has a large loss of speed. Mesh 4 is designed as a plate with big holes, but also has a larger area of the metal plate, which will hinder the particles going through the mesh, and the velocity will primarily reduce. For Mesh 2 35 and Mesh 3, they are very similarly designed, but the sizes of grid unit are not the same, smaller group size will also cause the speed to drop, and there will be lots of pressure, that is the reason why velocity with Mesh 3 can reach a higher speed. Nevertheless, both Mesh 2 and Mesh 3 have much better than Mesh 4 for particles to go through. While Mesh 1 is designed to have a much bigger size of the unit compared to Mesh 2 but has an even distribution of grids as Mesh 2, the result shows that Mesh 1 and Mesh 2 have similar velocity increasing trends, despite the velocity with Mesh 2 increases more after 20kv under negative side. If the experiment is conducted with fan, there will be an initial velocity of the wind. So the difference between them will be more obvious to see. Mesh 1 has the highest initial voltage because little pressure loss and also can accelerate to 7.6m/s. Follow by Mesh 1 is Mesh2, but there is still a significant difference between them while Mesh 3 Mesh 4 have a large loss in pressure because the grid restricts air to go through it. However, Mesh 3 can have more velocity when the corona discharge is working, may due to the more even in the distribution of grid units. Considering that in pollution cleaning process, the corona discharge device should have a good condition to have flow recycle, Mesh 1 has the highest initial velocity and also the highest after particles being charged. So to have the most efficient grid, Mesh 1 is suitable for it. 36 Negative 0.6 0.5 Velocity(m/s) 0.4 0.3 0.2 0.1 0 0 5 10 15 20 25 30 Voltage(kv) Mesh 1 Mesh 2 Mesh 3 Mesh 4 Figure 24 The comparison between different shape of mesh in V-U(Negative) Negative 0.8 0.7 Velocity(m/s) 0.6 0.5 0.4 0.3 0.2 0.1 0 0 5 10 15 20 25 30 Voltage(kv) Mesh 1 Mesh 2 Mesh 3 Mesh 4 Figure 25 The comparison between different shape of mesh in V-U with initial velocity(Negative) 37 Negative 140 120 Current(uA) 100 80 60 40 20 0 0 5 10 15 20 25 30 VoltageοΌkv) Mesh 1 Mesh 2 Mesh 3 Mesh 4 Figure 26 The comparison between different shape of mesh in I-U(Negative) 3.7 Influence of amount of needle Some articles say that a multi-needle can enhance the velocity of ionic wind because more needles will create a high field density and more stable discharge [13]. We design three multi-needles by twisting 2,3 and four needles into one respectively. The diameter of all the needle is 0.50mm, each needle in 2-needles design arranges into a line for 180°angle, the angle is 120°in 3-needles designing and 90°in 4-needles designing. All of the needles have the same distance(1.77cm) to the axes. The distance between electrodes is 5.12cm. The result is given by Fig 27 and Fig 28, which is conducted under negative discharge. From the comparison, there is a vast improvement when the amount increases from one to two, both of the velocity and current are much higher than them in 1-needle design. That mat due to that two needles can have a wider area of discharge, and more particles can be ionized, thus, the field density will be stronger and more ionized particle can collide with air molecules so that the ironical wind will be faster. Also, it 38 should be noticed that the current and velocity do not have double values of them when the amount increases from one to two, so the relationship cannot be simplified as πΌ2 = 2πΌ1 or π2 = 2π1 . There should be a more complicated relationship between them. Although there is a significant difference between 1-needle and 2-needle design, when the amount raises to 3 and 4, the improvement is not so obvious, especially when it comes to 2-needles and 3-needles design, the lines of velocity nearly converge together. While 4-needles has the highest line of speed, but the difference is still not so remarkable. On the other hand, current-voltage characteristic curves in these designs have more stable rising compared to each other. It should be mentioned that the distances between each tip of the needle are not the same, but to the axes, a close range will make the fields interactional, and that will weaken the field strength, so the improvement will not as high as we expected. If we compare them with current velocity curve which can be seen in Fig 29, the development on velocity is quiet little between 2,3,4-needles designs, but the current consumption will get much more. Also, the change of some needles seems that has no influences on the onset voltage, theses four designs have nearly the same onset voltage because the diameter of each needle and the distance between electrodes remain no changes. To sum up, a multi-needles design can enhance velocity compared to 1-neelde design, but more needles will cause a mutual influence on each other, and the improvement will be not as much as we wanted. 39 250 150 100 50 0 0 5 10 15 20 25 30 Voltage(kv) 1 Neeld 2 Needles 3 Needles 4 Needles Figure 27 The comparison between different amount of needle in I-U 0.9 0.8 0.7 0.6 Velocity(m/s) Current(uA) 200 0.5 0.4 0.3 0.2 0.1 0 0 5 10 15 20 25 30 Voltage(kv) 1 Needle 2 Needles 3 Needles 4 Needles Figure 28 The comparison between different amount of needle in V-U 40 0.9 0.8 Velocity(m/s) 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 0 50 100 150 200 250 Current(uA) 1 Needle 2 Needles 3 Needles 4 Needles Figure 29 The comparison between different amount of needle in V-I 3.8 Conclusion From experiment 3.1 to 3.7, we obtained how several factors could affect the corona discharge. As a result, we designed a corona discharge device as it can have smallest onset voltage, bigger breakdown voltage, high wind speed. So due to the comparisons from these experiments before, we at this moment use a needle with 0.15mm diameter as corona electrode, because a thin wire can have lower onset voltage and higher speed, the needle is connected to high voltage cable and it can be changed easily. Mesh 1 is used as collector electrode because the experiment shows it can have not only more top speed but also the highest initial velocity so that the low-pressure loss can allow the air cycling more smoothly. A plastic stick connects and fixes the mesh and needle, both of their positions can be adjusted, so the distance G can be changed flexibly, G=4.12cm because it will have a comparably lower onset voltage and higher speed, but also can make the break-down delay. The wires connected to needle and mesh will go through a plastic tube and link to the power supply and ground, the whole 41 length of the device is 20.50cm. This device will be tested in a container filled with aerosol, because the device needs to go inside the container, the diameter of the mesh has to be reduced to 10.25cm. Figure 30 The corona discharge designing 42 4. Experiment (2) Set-up A test stage will be set up to text the corona discharge device, There is a metal barrel which has a diameter of 62cm, the height is 95cm, and the volume is 0.29m³. Aerosol will be injected through a hole with a diameter of 8cm in the middle of the top. Two holes on the both sides of the barrel, one has the diameter of 15cm and corona discharge will be set in the barrel through it. Another is set 86cm from the top and has the same diameter, where a fan will connect to and pump the air inside the barrel. The whole barrel will be linked to ground. Aerosol will be generated by a Laskin-nozzle aerosol generator, which container the liquid named AEROSIL OX50, including 99.8% SiO2 and particle size is about 40nm. When putting air into the aerosol container, due to The Venturi Effect, aerosol will be pulverized into tiny liquid particles and be injected from the top into the barrel through a tube with a diameter of 2.54cm. The pressure of air pumped into the aerosol container is keeping at 2bar. A FAST MOBILITY PARTICLE SIZER™ SPECTROMETER MODEL 3091 (FMPS) is connected to the tube of existing, allows mixed air from container entering it with a flux of 10L/min, or 0.6m³/h. This FMPS is used for measuring the distribution of different particle sizes from 5.6~560nm, and the data can be demonstrated as total number concentration, total mass concentration, and total volume concentration. The mixed flow in the container will always be pumped out of it, and a fan realizes this, a Testo 520D0E10 MODO2 digital manometer is linked to the tube of existing and can measure the pressure range from 0~200mbar. All of the experiments are conducted in the room with a temperature of 22β and a standard pressure of 100Kpa. The whole operation device shows in Fig 31. 43 Figure 31 Experiment set-up with FMPS 4.1 Corona discharge texting To test the practical function of corona discharge device, the device will be set on the side of the container, the position can be shown in Fig 31. The volume flow rate of the outlet remains 8m³/h. Air will be injected into aerosol generator under the pressure of 2bar. 44 The fan will keep running while FMPS starts to record. As the total number concentration stable, insert aerosol for 5 minutes, then stop to wait for 5 minutes. Again pump aerosol for minutes but open corona discharge under negative discharge within 20kv. This operation will be conducted for three times while using a different amount of needle, 1-needle, 2-needles, 3-needles design. Figure 32 The comparison of different amount of needle in number concentration with 25kv discharge The data is collected, and we select data during the 5 minutes of injecting aerosol and running corona discharge to have the average values, which is given in Table 5. Then the result will be compared in Fig 33 and Fig 34. Table 5 Comparison of total number concentration in different amount of neeldes Total number concentration (#/cm³) Elimination efficiency Without Corona 6.9 × 106 45 1-Needle 2-Needles 3-Needles 4.5 × 106 4.1 × 106 3.9 × 106 34.19% 40.55% 42.21% From the experiment result, without corona discharge, the total number concentration is about 6.9 × 106 , while after using 1-Needle corona discharge, this value can be reduced to 4.6 × 106 , decreasing by 34.19% if compared with non-corona discharge condition. 2-Needles design can reduce the total number concentration by about 40.55%, which improves 6% while 3-Neeldes design only increases by 2% compared 2-Needles design. This result can also match with Chapter 3.7 as the amount of needle increasing, the efficiency does not show a significant growing trend. Number concnetration Number concentration(#/cm³) 800000 700000 600000 500000 400000 300000 200000 100000 6.04 6.98 8.06 9.31 10.8 12.4 14.3 16.5 19.1 22.1 25.5 29.4 34 39.2 45.3 52.3 60.4 69.8 80.6 93.1 107.5 124.1 143.3 165.5 191.1 220.7 254.8 294.3 339.8 392.4 453.2 523.3 0 Particle diameter(nm) Without corona 1-Needle 2-Needles 3-Needles Figure 33 Number concentration under different amount of needles From the graph of the number concentration distribution, particle sizes range from 6.04~453.2nm, while particles range from 34~60.4nm take the largest par. Under corona discharge, the separation of particles causes the bigger size particles this field will shift to 25.5~45.3nm because the aerosol particles are ionized and collected by the mesh. Thus from Chapter 2.9, where the rate of coagulation can be described as ππ ππ‘ = −πΎ0 π 2 . The decreasing of the number of particles N will drop which will reduce the rate of coagulation. In addition, πΎ0 can be described as πΎ0 = 4πππ π· ,so that when 46 the rate of coagulation is decreasing, there will be less big size particles to be generated, which will also reduce the rate. The result also shows that there is a drop of number concentration between three types of corona discharge, but the difference between 2-needles and 3-needles is not noticeable if compared to 1-Needle design. The number concentration of particles ranges from 25.5~45.3nm has a slight reduce, but it is nearly the same compared with 2-Needles and 3-Needles design. Mass concentration Mass concnetration(μg/lm³οΌ 250 200 150 100 50 6.04 6.98 8.06 9.31 10.8 12.4 14.3 16.5 19.1 22.1 25.5 29.4 34 39.2 45.3 52.3 60.4 69.8 80.6 93.1 107.5 124.1 143.3 165.5 191.1 220.7 254.8 294.3 339.8 392.4 453.2 523.3 0 Particle diameter(nm) Without corona 1-Needle 2-Needles 3-Needles Figure 34 Mass concentration under different amount of neeldes For the mass concentration, thought particles range from 80.6~453.2nm have only a small number concentration, they take a large part of mass concentration because normally the large size particles have much heavier mass. While with corona discharge, there is a significant decrease in mass concentration, which shows that the separation impedes the coagulation. It is evident that there is almost non-difference between 2Needls and 3-Needles design, the improvement is very limited. 47 4.2 Different positions of setting corona discharge This experiment aims to find how the different positions can influence the result of corona discharge. For each test, firstly inject aerosol with the pressure of 2bar until the number concentration keeping stable, then start the corona discharge under 15kv, as the number concentration stable, adjust power to 25kv. The volume flow rate of outlet remains 13.6m³/h, and the diameter of the tube which injects aerosol is 2.54cm. The positions of discharge can be seen in Fig 35. Position(a) is the same in Chapter 4.1, the corona discharge is set on the side of the container. While position(b) is established under the nozzle, so aerosol particles go through discharge field vertically. Position(c) is also placed under the nozzle, but it has the same direction as aerosol particles Figure 35 Position (a) (b) (c) 48 Position(a) Position(b) 49 Position(c) Figure 36 Total number concentration in position(a) (b) (c) with 0kv,15kv,25kv discharge The total concentration can be seen in Fig 36, the difference between them is very obvious. The initial total number concentration is nearly the same, around3.0 × 106 . When to start to have corona discharge under 15kv, concentration shows a reduced trend. Position(a) under 15kv reduces less than position(b), and when power supply rises to 25kv, only a slight drop in position(a) while decreasing in position(b) is obvious. Nevertheless, position(c) has a significant reduce of total number concentration. Concentration goes sharply to 6.5 × 105 , nearly one-third compared with position(b), as power rising to 25kv, concentration drops again and it has the lowest values between three different positions. 50 Table 6 Comparison of total number concentration in different positions Position Total number concentration under 0kv (#/cm³) a 3163255 b 3058151 c 2887848 Total number Total number concentration concentration under15kv under 25kv (#/cm³ ) (#/cm³) (Elimination (Elimination efficiency) efficiency) 2130997 1333182 (32.63%) (57.85%) 1727934 711424 (43.50%) (76.74%) 656291 299668 (77.27%) (89.62%) The values of total number concentration are selected during the time when the concentration is stable, then they are calculated to have the average values, which are given in Table 6. Total number concentrations in three positions are nearly the same when the discharge is not working. As power rising up, concentration reduces by 32.36% under 15kv and 57.85% under 25kv in position(a), nearly increases by 25%. It reduces nearly the half by 43.50% under 15kv in position(b), when power is growing to 25kv, it increases to 76.74%. The result in position(c) is the best of them, it can reduce by 77.27% under 15kv, the efficiency is much higher compared to others, and it can even reach to 89.62% under 25kv. The result in position(a) is not as well as others because discharge device is set on the side of the barrel so only particles in a particular area can be ionized by discharge, even when power turns to the maximum, the flux of corona wind is not sufficient to ionize much more particles. While both position(b) and position(c) are very close to the nozzle, where has the highest concentration of particles. Thus, it is much easier to ionize more particles for them. However, in position(b), the direction of corona wind is vertical to the direction of aerosol flow, so some particles may cross electric field 51 directly which will reduce the number of particles charged. As power growing up, corona wind gets faster, and this will have a larger influence on the motions of particles which can change the direction of particles. So that there are more chances to ionize particles. In position(c), corona wind and aerosol flow have the same direction, more particles can be charged, and corona wind can also get a higher speed, so it is much easier to collect particles. Also, the distance between electrodes is the only 4.12cm, so the space of electrical field is extremely limited. So position(c) has an advantage because most of the particles can pass the needle and get ionized, while for the position(b), not all of the particles can go exactly through the needle and be ionized. Number concentration Number concentration(#/cm³) 400000 350000 300000 250000 200000 150000 100000 50000 6.04 6.98 8.06 9.31 10.8 12.4 14.3 16.5 19.1 22.1 25.5 29.4 34 39.2 45.3 52.3 60.4 69.8 80.6 93.1 107.5 124.1 143.3 165.5 191.1 220.7 254.8 294.3 339.8 392.4 453.2 523.3 0 Particle diameter(nm) Position(a):Without discharge Position(b):Without discharge Position(c): Without discharge Position(a):With discharge under 15kv Position(b):With discharge under 15kv Position(c):With discharge under 15kv Figure 37 Number concentration with 15kv discharge under different positions. 52 Mass concentration 90 Mass concentration(μg/m³) 80 70 60 50 40 30 20 10 6.04 6.98 8.06 9.31 10.8 12.4 14.3 16.5 19.1 22.1 25.5 29.4 34 39.2 45.3 52.3 60.4 69.8 80.6 93.1 107.5 124.1 143.3 165.5 191.1 220.7 254.8 294.3 339.8 392.4 453.2 523.3 0 Particle diameter(nm) Position(a):Without discharge Position(b):Without discharge Position(c):Without discharge Position(a):With discharge under 15kv Position(b):With discharge under 15kv Position(c):With discharge under 15kv Figure 38 Mass concentration with 15kv discharge under different positions Number concentration Number concentration(#/cm³) 400000 350000 300000 250000 200000 150000 100000 50000 6.04 6.98 8.06 9.31 10.8 12.4 14.3 16.5 19.1 22.1 25.5 29.4 34 39.2 45.3 52.3 60.4 69.8 80.6 93.1 107.5 124.1 143.3 165.5 191.1 220.7 254.8 294.3 339.8 392.4 453.2 523.3 0 Particle diameter(nm) Position(a):Without discharge Position(b):Without discharge Position(c): Without discharge Position(a):With discharge under 25kv Position(b):With discharge under 25kv Position(c):With discharge under 25kv Figure 39 Number concentration with 25kv discharge under different positions. 53 Mass concentration Mass concentration(μg/m³) 90 80 70 60 50 40 30 20 10 6.04 6.98 8.06 9.31 10.8 12.4 14.3 16.5 19.1 22.1 25.5 29.4 34 39.2 45.3 52.3 60.4 69.8 80.6 93.1 107.5 124.1 143.3 165.5 191.1 220.7 254.8 294.3 339.8 392.4 453.2 523.3 0 Particle diameter(nm) Position(a):Without discharge Position(b):Without discharge Position(c):Without discharge Position(a):With discharge under 25kv Position(b):With discharge under 25kv Position(c):With discharge under 25kv Figure 40 Mass concentration with 25kv discharge under different positions Fig 37 to Fig 40 show how the number concentration and mass concentration change during the experiment. It is evident that there is a huge drop in the number and mass concentration when power turns to from 0kv to 25kv, especially for the position(c), the number and mass concentration reduce nearly to 90%. Due to the separation, the line of number concentration shifts to left side slightly. Position (c) has the best collect efficiency even when it supplies with 15kv voltage, both of number and mass concentration is much lower than it in position(a) and (b) with the 25kv voltage supply. The number concentration in position(c) is lower than others, this mainly depends on the hinder of mesh. When aerosol goes through discharge device, they will collide to the mesh where can cause the pressure loss and give particles more time to have coagulation. Which can also show in Fig 38, the line of mass concentration is slightly higher compared to position(b). However, there are more big size particles in position(a) when the discharge is not working, somehow coagulation occurs in advance, so there are more big size particles though the total number concentration is nearly the same. 54 Nevertheless, it is still obvious that discharge set in position(c) can have the highest collect efficiency. Figure 41 Increasing the velocity of aerosol in (d) and (e) Another idea aims to increase the velocity of aerosol by adding a muzzle with a smaller diameter, reducing from 2.54cm to 1.32cm, which shows in Fig 4.10. Position(d) 55 Position(e) Figure 42 Total number concentration in position (d) (e) with 0kv,15kv,25kv discharge The result is given in Fig 42, when the velocity of the aerosol is improved, there is less number concentration in the barrel, especially in position(e). This result is mainly caused by the good condition for coagulation. When the aerosol stream is coming inside the barrel with high speed, the mixture is better than before and then, there will be more chances for particles to collide with each other and coagulation will happen more frequently. As a result, particles change to big size particles and the number will reduce. For position(e), the aerosol stream will go through the mesh, which may have a hinder function for particles. Thus, there is more resident time for particles to mix, and coagulation occurs even more. 56 Table 7 Comparison of total number concentration in faster velocity of aerosol Position Total number concentration under 0kv (#/cm³channel) d 2447086 e 2127784 Total number concentration under15kv (#/cm³channel) (Elimination efficiency) 1862913 (23.87%) 756734 (64.44%) Total number concentration under 25kv (#/cm³channel) (Elimination efficiency) 1483171 (39.40%) 412834 (80.60%) By calculating the average values, the average total number concentration during the time when total number concentration is stable can be seen in Table 7. Due to adding a smaller diameter of the nozzle, there is a mechanical loss when aerosol goes through it. Thus, the initial number concentration is lower than the result given in Table 6. When the speed of aerosol streams increases, the collect efficiency is not getting higher. By contraries, efficiency shows a trend of decrease, especially when the position is set as position(d), almost drops a half compared to position(b). The reason may be caused as when high-speed aerosol passes the electrical field, there is not sufficient time for corona discharge to ionize the particles which travel in a vertical direction of corona wind and across the electrical field directly. So a part of particles cannot be ionized, and efficiency reduces. Although there is also a drop of collect efficiency when the device is set as position(e), the influence is not as much as position(d). The collect efficiency is still the highest and also can reach to 80% when the voltage increases to 25kv. Because aerosol particles and corona wind have the same direction, it is much easier for the needle to catch and ionize particles. Thus, position(e) can still work well with high-speed aerosol. 57 From Fig 43 and 44, it shows that there are more big size particles in position(e). Because aerosol stream must pass the complete discharge device, so there are more speed loss and more resident time to have coagulation. The line of number concentration shifts to the right side and higher mass concentration to a great size particle. Number concentration 300000 250000 200000 150000 100000 50000 0 6.04 6.98 8.06 9.31 10.8 12.4 14.3 16.5 19.1 22.1 25.5 29.4 34 39.2 45.3 52.3 60.4 69.8 80.6 93.1 107.5 124.1 143.3 165.5 191.1 220.7 254.8 294.3 339.8 392.4 453.2 523.3 Number concentration(#/cm³) 350000 Particle diameter(nm) Position(d):Without discharge Position(e):Without discharge Position(d):With discharge under 15kv Position(e):With discharge under 15kv Position(d):With discharge under 25kv Position(e):With discharge under 25kv Figure 43 Number concentration with high-speed aerosol 58 Mass concentration Mass concentration(μg/m³) 60 50 40 30 20 10 6.04 6.98 8.06 9.31 10.8 12.4 14.3 16.5 19.1 22.1 25.5 29.4 34 39.2 45.3 52.3 60.4 69.8 80.6 93.1 107.5 124.1 143.3 165.5 191.1 220.7 254.8 294.3 339.8 392.4 453.2 523.3 0 Particle diameter(nm) Position(d):Without discharge Position(e):Without discharge Position(d):With discharge under 15kv Position(e):With discharge under 15kv Position(d):Wit discharge under 25kv Position(e):With discharge under 25kv Figure 44 Mass concentration with high-speed aerosol 4.3 Conclusion Chapter 4 conducts several experiments to test the corona discharge designing in a barrel with aerosol stream. From results, the corona discharge can have the effect on collecting aerosol particles, which causes both of number and mass concentration decreasing. Also due to the function of separation, more small particles are being generated which hinder coagulation. As voltage growing up, this collect effect gets to more obvious. Also, the collect efficiency relates to the velocity of corona wind. As we see in Chapter 3.7, more needles can have a higher speed. Thus, greater collect efficiency can be realized. However, the improvement of collect efficiency is limited when needles get more. Also, the experiment shows that the position of corona discharge has a huge influence on collect ability. When corona discharge device has the same direction of 59 the aerosol stream, because particles will move towards the electrical field, it is much easier for particles to get ionized and be collected. While if the corona discharge device was set on the side of the barrel, the efficiency is not very high because the speed of corona wind is very limited and cannot have sufficient time to ionize more particles in the barrel. When the direction of corona wind is vertical to the direction of aerosol, the result is not as well as that when corona wind and aerosol have the same direction but also much higher than it when the device is set on the side of the barrel. Thus, it is evident that when a corona discharge is set to close to the entrance of aerosol, where gathers high concentration of particles, the collect efficiency can always be high. However, when the velocity of aerosol gets higher, collect efficiency seems to have a drop. Fast particles have the disadvantage to be ionized by the needle. 60 5. Future study The previous research is focusing on corona discharge with only one high voltage supply. Although there is a significant increasing of corona wind speed after several improvements, this development is limited because voltage supply can only reach to a certain value. Otherwise, there will be breakdown and spark. So the idea is to set two or more voltage supplies, which can be seen in Fig 43. The mechanism of this two-stages corona discharge is the same as the previous study, but when natural molecules are collided by moving ionized particles and pass through the mesh, another corona discharge needle will ionize them. Thus, these natural molecules with initial dynamic energy will be accelerated again in the next stage, which will have a higher speed and larger ionizing area. While it is still cannot be sure whether this two-stages discharge device can largely improve or not, and also how long the distance between two stages should be set is remained unclear, which should be investigated in the further study. Figure 45 Two stages corona discharge device 61 6. Summary This project aims to optimize a corona discharge device. First of all, from the theoretical knowledge we know that the corona discharge happens when there is an extremely uneven electrical field, frequently this can be realized by a needle which is connected to high voltage. Thus, this partial strong electrical field can ionize particles around the needle. Also, as ionized particles move to the other electrode, some natural air molecules will be collided and obtain the dynamic energy to travel towards the electrical field, which can generate corona wind. In a process of corona discharge, the on-set voltage is an imperative factor for a corona discharge. From the experiment (1) set-up, we verify that the both of current and velocity of corona wind can increase as voltage growing up, and an early onset voltage discharge can reach to higher current and higher speed. While as the voltage is high enough, the breakdown will happen, and corona wind cannot be faster anymore, corona discharge has lost the effect. Because the different mechanism of corona discharge in the positive and negative side, negative discharge can always have a lower onset voltage and higher breakdown voltage. A thinner needle with a sharp tip can have an absolute field strength, so it is much easier to have corona discharge. Close distance between electrodes can also reduce the onset voltage, but it will also make corona discharge to have break-down in advance because the electrical stream can quickly reach to another electrode. The shape of mesh does not have an influence on the onset voltage, so current is nearly the same. However, it can significantly affect the speed of corona wind because of the pressure drop when airflow goes through it. We also text a multi-discharge with two, three and four needles. Some needles seem to have no influence on the onset voltage, but current and velocity can largely increase when more needles are being set. This may due to a larger ionize area so that more particles can be ionized. While as amount increasing 62 more, the improvement seems to be very limited because each needle can influence other needles, so it is very complex to explain it. Based on experiment (1) set-up, we designed a corona wind discharge which can be set inside a barrel. Thus, experiment (2) set-up will test this corona discharge with aerosol flows. By using the discharge device, aerosol particles can be ionized and collected by the electrode. The result shows that there is a noticeable drop in number and mass concentration. Also, due to the separation which hinders coagulation, so there are less big size particles existing in the barrel. The position of setting the discharge device also has a significant influence. It is vital to set discharge device nearby the entrance of aerosol so that a much higher collect efficiency can be obtained, especially when the corona wind has the same direction of aerosol flows, collect efficiency can even reach to 90%. For further study, based on the previous conclusion, a two-stages corona discharge can be tested with experiment set-up, which aims to have a higher collect efficiency. 63 Reference [1] W.C.Hinds, Aerosol Technology, 2nd edition, Wiley Interscience, pp.3-4,1999. [2] E.Y.Zhu, B. Mao, The Present Situation of Aerosol Research in China, Journal of Environmental science and management, Vol. 33, NO.12, pp.57-58, 2008. [3] Y. M. Kang, M. L. Rong, H. G. Shen, M. X. Wang, Corona Discharge and Ionic Wind in Air Cleaning Process, Ziran Zazhi, Vol.24, pp.125-129,2002. [4] G.D. Conanan, F.C. Lai, Performance Enhancement of Two Stage Corona Wind Generator in a Circular Pipe, University of Oklahoma, 2012. [5] E. Jewell-Larsen, S.V. Karpov, A. Krichtafovitch, V. Jayanty, C.P. Hsu, A.V. Mamishev, Modeling of corona-induced electrohydrodynamic flow with COMSOL Multiphysics, Department of Electrical Engineering, University of Washington, 2008. [6] S. Tapuchi, S. Makarenko, M. Malinkovski, R. Schwartz, O. Amsalem, Method for Improved Efficiency and Reduced Energy Consumption of an Ion Motor, 2012. [7] F.D.Li, L.T.Li, S.J.Sun, Effect of electrohydrodynamic (EHD) technique on drying process and appearance of okara cake,Journal of Food Engineering, Vol.77, pp.275-280,2006. [8] K. Adamiak, V. Atrazhev, P. Atten, Corona discharge in the hyperbolic point-plane configuration: Direct ionization criterion 64 versus approximate formulations, Ieee Transactions on Dielectrics and Electrical Insulation, vol. 12, pp.1025-1034, 2005. [9] N.Ding,Z.Q.Xie, Research on Corona Discharge Plasma Properties, Journal of Guangdong Chemical, Vol. 214, pp.119-121, 2011. [10] H. Ait said, H. Nouri, Y. Zebboudj, Effect of air flow on corona discharge in wire-to-plate electrostatic precipitator, Journal of Electrostatics, Vol. 73, pp.19-25, 2015. [11] W.C. Hinds, Aerosol Technology, UCLA School of public Health, Los Angeles, California, pp.318-319, 1999. [12] L.N.Li,S.J.Lee,W.J.Kim,D,Kim, An empirical model for ionic wind generation by a needle-to-cylinder dc corona discharge, Journal of Electrostatics, Vol.73,pp. 125-130,2015. [13]H.F.Chen,Y.M.Zhu,P.H.Su,J.Y.Yang,Current—Voltage Characteristics of Bipolar Corona Discharge in Multi-needle Electrode Configuration,Journal Engineering,Vol.33,No.10,pp.92-95,2007. 65 of High Voltage Appendix Experiment(1)set-up Corona wind speed measurement 66 Corona discharge device Collect electrode 67 Experiment(2)set-up Laskin-nozzle Aerosol generator 68 Fan in experiment(2) FMPS Device 69 Corona discharge device Position(a) 70 Position(b) Position(c) 71 Aerosol with higher speed 72