NT® Pressure Transducer Models 4100, 4210, 4130, 4230 User Guide

P/N 01-1027289 (Rev. A 04/14)
NT® PRESSURE TRANSDUCER
MODELS 4100, 4210, 4130, 4230
User Guide
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
Table of Contents
Safety Alert Symbol ............... 2
Certifications .................... 12
CE Compliance . . ................. 12
Introduction ........................ 3
Class I, Division 2
Hazardous Environments ... 12
Installation ......................... 3
Mechanical Installation .........
Electrical Installation ...........
Power Supply Requirements ....
Warnings ..........................
3
4
4
4
Ordering Information ........... 13
Model 4100 Single-port ........
Model 4210 Flow-through ......
Model 4130 Single-Port ........
Model 4230 Flow-through .......
13
14
15
15
Unit Operation ..................... 6
Operating Environment .........
Process Connections .............
Pressure Transducer
Cover Assembly ................
Pressure and Temperature
Requirements ..................
Pressure Limits ...................
Pressure Reference
Accuracy ........................
Linear Output Signal ............
6
6
6
6
7
Installation Drawings .......... 16
Model 4100 Single-port ........
Model 4210 Flow-through ......
Model 4130 Single-port ........
Model 4230 Flow-through .......
16
18
20
21
For More Information . . ........... 22
7
8
Terms and Conditions
of Sale ............................... 22
Troubleshooting ................... 9
Product Warranties ................ 22
Maintenance ....................... 9
Zero and Span Adjustment ...... 9
Reference ......................... 11
Physical Specifications ......... 11
Electrical Specifications ....... 11
Performance Specifications .. ... 11
ENTEGRIS, INC.
USER GUIDE
1
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
Safety Alert Symbol
!
Indicates a hazardous situation which, if not avoided,
could result in serious injury or death.
WARNING!
Introduction
WARNING!
Attempting to install or operate
NT standard 4000 Series Pressure
Transducers without reviewing the
instructions contained in this
manual could result in personal
injury or equipment damage.
This manual is for use with the
following standard NT® Pressure
Transducers 4000 Series Models
4100, 4130, 4210 and 4230. These
instruments have been designed for
use in high purity applications in the
semiconductor industry.
These products feature no moving
parts and no filled cavities, which
reduces the possibility of a contaminated process. The wetted parts of
these nonmetallic transducers are
constructed with PTFE, PFA or other
similar high purity inert materials.
!
Cut-away Example: NT Single-port Pressure Transducer
Cut-away Example: NT Flow-through Pressure Transducer
Polypropylene electrical
connector option
Signal conditioning circuit
Polypropylene electrical
connector option
Signal conditioning circuit
Electronic pressure sensor
Electronic pressure sensor
PTFE Body
PTFE Body
Process connection
Process connection
2
USER GUIDE
ENTEGRIS, INC.
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
Installation
!
WARNING!
The pressure transducer has
been factory sealed. Do not
attempt to remove the cover
of the pressure transducer.
Any attempt at removal of
the pressure transducer cover
will void the warranty and
damage the unit.
NOTE: NT Pressure Transducers
for DI water applications, Models
4130 and 4230, are built with
a removable grommet to allow
adjustment of the zero and span
potentiometers.
!
WARNING!
Do not tighten the nuts that
protect the process connections
during shipment. Do not tighten
the nuts unless the proper tubing
has been installed. Tightening
these nuts may result in damage
to the pressure transducer process
connections.
ENTEGRIS, INC.
Mechanical Installation
Standard NT Pressure Transducers
utilize flared tube connections and
must be used with the proper tubing
size and coupling nut.
When installing flared tubing to
the pressure transducer, the flared
tube is pushed over the transducer’s
fitting until the fitting reaches the
smaller tube diameter. The amount
of torque required to tighten the
coupling nut is dependent upon
the size of the fitting.
Fitting Size
1⁄4”
3⁄8”
1⁄2”
3⁄4”
1”
Torque (in•lbs)
5
8
11
14
30
Care should be taken when installing flared tubing onto the pressure
transducer. Do not use excessive
torque or subject the transducer
to high heat during installation.
The transducer should be firmly
mounted to a solid surface to ensure stability. Verify that the pressure transducer and the sig­nal cable
are free from mechanical stress
or excessive bending from the
surrounding equipment.
USER GUIDE
3
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
Electrical Installation
The pressure transducer provides
an analog (0–5 VDC, 0–10 VDC or
4–20 mA) electrical output proportional to the pressure measured.
NOTE: The white wire is not
required for the 4–20 mA output
configuration, please refer to the
wiring diagrams below.
+24 VDC
GND
Pressure transducer
4–20 mA output version
WHITE
RED
BLACK
No connection
4–20 mA pressure signal
Display
or PLC
Pressure Transducer
0–5 VDC, 0–10 VDC output version
Voltage pressure signal
WHITE
RED
Display
or PLC
BLACK
Power Supply Requirements
The pressure transducer requires
a 12–30 volt DC power supply with
less than 2% ripple at 100 or 120 Hz.
The required power supply voltage
varies with the load resistance
(RLoad), please refer to the formulas
on page 6. The power supply must
provide clean power and must be
used only to power similar measurement-type devices. The power supply
must not be used to power inductive
loads, such as motors, relays or solenoids. These devices may produce
transients that may affect the pressure transducer measurements when
the inductive device is powered up or
powered down.
NOTE: Be sure to ground the shield
of the cable to local ground.
4
USER GUIDE
Warnings
!
!
!
!
WARNING!
DO NOT REMOVE OR REPLACE
WHILE CIRCUIT IS LIVE UNLESS
THE AREA IS KNOWN TO BE FREE
OF IGNITIBLE CONCENTRATIONS
OF FLAMMABLE SUBSTANCES.
WARNING!
DO NOT REPLACE COMPONENTS
UNLESS POWER HAS BEEN DISCONNECTED OR THE AREA IS
KNOWN TO BE FREE OF IGNITIBLE
CONCENTRATIONS OF FLAMMABLE
GASES OR VAPORS.
WARNING!
EXPLOSION HAZARD. DO NOT
CONNECT WHILE THE CIRCUIT
IS LIVE OR UNLESS THE AREA IS
KNOWN TO BE FREE OF IGNITIBLE
CONCENTRATIONS.
WARNING:
SUBSTITUTION OF COMPONENTS
MAY IMPAIR SUITABILITY FOR
DIVISION 2
ENTEGRIS, INC.
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
Figure 1. Power Required for a 4–20 mA Loop
RLoad (Ohms)
850
Operating
region
0
Power supply voltage
(VPS)
12
24
29
Load Resistance: Current Output
(4–20 mA Configuration)
Load Resistance: Voltage Output
(0–5, 0–10 VDC)
If a load resistor, R Load, is used in
series with the current output, the
value of RLoad is dependent on the
supply voltage and the meter resistance and is calculated from the
following formula:
The output impedance is 1 kOhm.
The input impedance should
be ≥1 megohm for ±0.1% load
impedance error.
R Load =
VPS – 12 V
20 mA
– Rmeter
where:
RLoad
= maximum load
resistance (ohms)
VPS
= power supply
voltage (volts)
Rmeter
= meter resistance (ohms)
(theoretically = 0)
ENTEGRIS, INC.
USER GUIDE
5
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
Unit Operation
Operating Environment
The pressure transducer is to
be mounted in a well vented and
controlled environment. Refer to
the Reference section on page 12
for additional specifications.
Process Connections
To avoid possible pressure leaks,
make sure all process connections
have been performed in accordance
with the Mechanical Installation
guidelines on page 4.
Pressure Transducer Cover
Assembly
Standard NT Pressure Transducer
covers are factory sealed and should
not be tampered with or opened.
Opening the cover shall void the
product warranty. The NT Pressure
Transducers for DI water applications, Models 4130 and 4230, are built
with a removable grommet to allow
adjustment of the zero and span
potentiometers.
!
6
Pressure and Temperature
Requirements
The minimum pressure required
is atmospheric pressure. The pressure transducer may be damaged if
subjected to vacuum pressure (pressure that is less than atmospheric
pressure).
The pressure transducer is rated for
use with fluids at process temperatures between 10–65°C (50–149°F)
under normal operating conditions.
!
WARNING!
NT Pressure Transducers, Models
4100, 4130, 4210 and 4230,
may be damaged if the sensor is
subjected to any level of vacuum
pressure (pressure less than
atmospheric pressure).
WARNING!
Any attempt to remove or
tamper with the transducer
cover will void the warranty
and damage the unit.
USER GUIDE
ENTEGRIS, INC.
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
Pressure Limits
Transducer Range
Maximum Pressure Limit
@ 23°C (73°F)
Maximum Pressure Limit
@ 65°C (149°F)
0–30 psig
690 kPa (100 psig) 3
690 kPa (100 psig) 2, 4
0–60 psig
1034 kPa (150 psig)1, 3
690 kPa (100 psig) 2, 4
psig)1, 3
690 kPa (100 psig) 2, 4
0–100 psig
1034 kPa (150
0–150 psig
1310 kPa (190 psig)1, 3
N/A
1. Maximum pressure limit for 3⁄4” Flaretek® fitting connections at 23°C (73°F) is 758 kPa (110 psig).
2. Maximum pressure limit for 3⁄4” Flaretek fitting connections at 65°C (149°F) is 620 kPa (90 psig).
3. Maximum pressure limit for 1” Flaretek fitting connections at 23°C (73°F) is 517 kPa (75 psig).
4. Maximum pressure limit for 1” Flaretek fitting connections at 65°C (149°F) is 372 kPa (54 psig).
For detailed pressure limit information on Flaretek tube fitting connections, refer to the Maximum
Pressure Capabilities chart in the Flaretek fittings product section on Entegris’ website at
http://www.entegrisfluidhandling.com.
!
WARNING!
The pressure limits for standard
NT Pressure Transducers (4000
Series) decrease significantly
for temperatures above 65°C
(149°F). Exceeding these limits
may result in personal injury
or equipment damage.
Pressure Reference Accuracy
The accuracy of the pressure transducer output is ±1% of full scale.
This accuracy includes the effects
of linearity, hysteresis and repeatability, measured at room temperature.
Accuracy specifications for nonstandard product configurations
might vary.
ENTEGRIS, INC.
USER GUIDE
7
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
Linear Output Signal
The output signal of the pressure
transducer is a linear function proportional to the applied pressure.
Current Output
(mA)
20 mA
4 mA
0 psig
Pressure Measurement
Full scale
Voltage Output
(V)
10 V
or 5 V
0V
0 psig
8
Pressure Measurement
USER GUIDE
Full scale
ENTEGRIS, INC.
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
Troubleshooting
Troubleshooting the NT Pressure
Transducers (4000 Series) may be
accomplished by measuring the
out­put signal of the pressure transducer with a battery powered
current/voltage meter. The meter
may be placed in series with the
pressure transducer to measure the
current output or it may be used to
directly measure the voltage output.
Using the battery powered current/
volt meter is an effective method to
determine whether the device or the
on-site data acquisition system is not
functioning properly.
Maintenance: (NT Pressure Transducers for DI
Water Applications, Models 4130 and 4230 Only)
Zero and Span Adjustment
Unit Calibration Steps:
1.Remove pressure transducer
from service.
2.Connect the pressure transducer
inlet outlet port connector to a
calibrated pressure source.
3.Connect the pressure transducer
electrical connection to a power
source.
4.Connect the pressure transducer
electrical connection to a display
to monitor the mA or voltage
output.
5.Remove the access grommet
(see Diagram A).
6.Locate the zero and span adjustment screws (see Diagram B).
ENTEGRIS, INC.
Diagram A
USER GUIDE
9
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
7.At ambient pressure (no pressure applied), rotate the zero adjustment screw (see Diagram C)
clockwise or counterclockwise
until the display reads 4.0 mA or
0 Volts.
8.Apply full scale pressure to the
unit. Rotate the span adjustment
screw clockwise or counterclockwise until the display reads the
full scale value (i.e., depending
upon configuration: 20.0 mA, 10
Volts or 5 Volts).
9.To obtain best accuracy, repeat
steps 7 and 8.
10. The calibration procedure is
complete. Reinstall the access
grommet.
11. Safely vent any applied pressure.
Disconnect the display and
disconnect the power.
12. Return the pressure transducer
to service.
10
USER GUIDE
Diagram B
Span
Pot
Zero
Pot
Diagram C
ENTEGRIS, INC.
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
Reference
The following lists the specifications
for the NT Pressure Transducer
product line. Please consult the
factory for product specifications
manufactured for nonstandard
applications.
NOTE: Specifications are subject to
change without notice.
Physical Specifications
Materials:
Wetted parts
Body
PTFE
Sensor interface
CTFE or PFA (Models 4100 and 4210)
O-ring
PFA (Models 4130 and 4230)
Kalrez® (Models 4100 and 4210)
EPDM(Models 4130 and 4230)
Nonwetted parts
Connection type:
Polypropylene, polyethylene, PVDF (in addition to
materials listed above)
Flared tube, NPT
Electrical Specifications
Input voltage:
24 VDC (12–30 VDC)
Input current:
20 mA maximum
Pressure signal output ranges:
4–20 mA, 0–5 VDC, 0–10 VDC
Electrical connection:
6’, 12’, 30’ FEP-jacketed pigtail or polypropylene 3-pin connector
Models 4100 and 4210)
12’ FEP-jacketed pigtail (Models 4130 and 4230)A
Electrical enclosure:
IP54
Performance Specifications
Reference accuracy:
±1% of full scale (includes linearity, hysteresis and repeatability)
at 23°C (73°F)
Process temperature:
10–65°C (50–149°F)
ENTEGRIS, INC.
USER GUIDE
11
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
Certifications
CE Compliance
Entegris products have been tested
to various test standards required
by the EMC 89/336/EEC directive.
The results of this testing are on
file at Entegris and are available
upon request.
Please contact the factory for
the latest information. The most
current specifications may be
found on Entegris’ Web site at:
http://www.entegrisfluidhandling.com/
Default.asp?G=1613&1n=en
Class I Division 2
Hazardous Environments
Entegris has tested our standard
products to the UL 61010-1, ANSI/ISA
12.12.01 standards for use in Class I,
Division 2 Group A–D, T6 10°C
≤ Ta ≤65°C hazardous environments.
The results of this testing are on file
at Entegris and are available upon
request.
All spec­i­fications are subject to
change. Please contact the factory
for the latest information.
12
USER GUIDE
ENTEGRIS, INC.
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
Ordering Information
Refer to the following for the desired configuration.
Model 4100 Single-port Pressure Transducer
Example model number: 4100-100G-F02-D06-A-P1-U1
Code
Pressure Range
Code
Sensor Interface
030G
060G
100G
150G
207 kPa (30 psig)
414 kPa (60 psig)
690 kPa (100 psig)
1034 kPa (150 psig)
P1
P2
CTFE sensor interface (default)
PFA sensor interface
Code
Inlet/Outlet Port Connection
F02
F03
F04
W02
W03
W04
N02
N04
M02
1⁄4”
Flaretek tube fitting
Flaretek tube fitting
1⁄2” Flaretek tube fitting
1⁄4” Super 300 Type Pillar®
tube fitting
3⁄8” Super 300 Type Pillar
tube fitting
1⁄2” Super 300 Type Pillar
tube fitting
1⁄4” FNPT
1⁄2” FNPT
1⁄4” MNPT
3⁄8”
Code
Electrical Connector Type
B06
B12
B30
D00
D06
D12
D30
FEP-jacketed 6’ pigtail
electrical cable
FEP-jacketed 12’ pigtail
electrical cable
FEP-jacketed 30’ pigtail
electrical cable
Polypropylene connector
(cable not included)
Polypropylene connector
and 6’ PVC cable
Polypropylene connector
and 12’ PVC cable
Polypropylene connector
and 30’ PVC cable
Code
Electrical Outputs
A
B
C
4–20 mA (12–30 VDC input)
0–10 VDC (12–30 VDC input)
0–5 VDC (12–30 VDC input)
ENTEGRIS, INC.
Code
Primary/Secondary Seal
U1
U2
U3
S1
S2
S3
Kalrez 4079 / Viton® (default)
Kalrez 1050 / Viton
Kalrez 6375 UP / Viton
Kalrez 4079 / Kalrez 4079
Kalrez 1050 / Kalrez 1050
Kalrez 6375 UP / Kalrez 6375 UP
USER GUIDE
13
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
Model 4210 Flow-through Pressure Transducer
Example model number: 4210-100G-F02-D06-A-P1-U1
Code
Pressure Range
Code
Sensor Interface
030G
060G
100G
207 kPa (30 psig)
414 kPa (60 psig)
690 kPa (100 psig)
P1
P2
CTFE sensor interface (default)
PFA sensor interface
Code
Inlet/Outlet Port Connection
F02
F03
F04
F06
F08
W02
W03
W04
W06
W08 1⁄4”
Flaretek tube fitting
Flaretek tube fitting
1⁄2” Flaretek tube fitting
3⁄4” Flaretek* tube fitting
1” Flaretek* tube fitting
1⁄4” Super 300 Type Pillar
tube fitting
3⁄8” Super 300 Type Pillar
tube fitting
1⁄2” Super 300 Type Pillar
tube fitting
3⁄4” Super 300 Type Pillar
tube fitting
1” Super 300 Type Pillar
tube fitting
3⁄8”
Code
Electrical Connector Type
B06
B12
B30
D00
D06
D12
D30
FEP-jacketed 6’ pigtail
electrical cable
FEP-jacketed 12’ pigtail
electrical cable
FEP-jacketed 30’ pigtail
electrical cable
Polypropylene connector
(cable not included)
Polypropylene connector
and 6’ PVC cable
Polypropylene connector
and 12’ PVC cable
Polypropylene connector
and 30’ PVC cable
Primary/Secondary Seal
U1
U2
U3
S1
S2
S3
Kalrez 4079 / Viton (default)
Kalrez 1050 / Viton
Kalrez 6375 UP / Viton
Kalrez 4079 / Kalrez 4079
Kalrez 1050 / Kalrez 1050
Kalrez 6375 UP / Kalrez 6375 UP
* For detailed pressure limit information on Flaretek
tube fitting connections, refer to the Maximum
Pressure Capabilities chart in the Flaretek
fittings product section on Entegris’ website
at http://www.entegrisfluidhandling.com.
Code
Electrical Outputs
A
B
C
4–20 mA (12–30 VDC input)
0–10 VDC (12–30 VDC input)
0–5 VDC (12–30 VDC input)
14
Code
USER GUIDE
ENTEGRIS, INC.
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
Model 4130 Single-port Pressure Transducer
Example model number: 4130-100G-F02-A
Code
Pressure Range
Code
Electrical Outputs
030G
060G
100G
150G
207 kPa (30 psig)
414 kPa (60 psig)
690 kPa (100 psig)
1034 kPa (150 psig)
A
B
C
4–20 mA (12–30 VDC input) 0–10 VDC (12–30 VDC input)
0–5 VDC (12–30 VDC input)
Code
Inlet/Outlet Port Connection
F02
F03
F04
W02
W03
W04
N02
N04
M02
1⁄4”
Flaretek tube fitting
Flaretek tube fitting
1⁄2” Flaretek tube fitting
1⁄4” Super 300 Type Pillar tube fitting
3⁄8” Super 300 Type Pillar tube fitting
1⁄2” Super 300 Type Pillar tube fitting
1⁄4” FNPT
1⁄2” FNPT
1⁄4” MNPT
3⁄8”
Model 4230 Flow-through Pressure Transducer
Example model number: 4230-100G-F02-A
Code
Pressure Range
Code
Electrical Outputs
030G
060G
100G
207 kPa (30 psig)
416 kPa (60 psig)
690 kPa (100 psig)
A
B
C
4–20 mA (12–30 VDC input) **
0–10 VDC (12–30 VDC input)
0–5 VDC (12–30 VDC input)
Code
Inlet/Outlet Port Connection
F02
F03
F04
F06
F08
W02
W03
W04
W06
W08
1⁄4”
Flaretek tube fitting
Flaretek tube fitting
1⁄2” Flaretek tube fitting
3⁄4” Flaretek tube fitting*
1” Flaretek tube fitting*
1⁄4” Super 300 Type Pillar tube fitting
3⁄8” Super 300 Type Pillar tube fitting
1⁄2” Super 300 Type Pillar tube fitting
3⁄4” Super 300 Type Pillar tube fitting
1” Super 300 Type Pillar tube fitting
3⁄8”
ENTEGRIS, INC.
* For detailed pressure limit information on Flaretek
tube fitting connections, refer to the Maximum
Pressure Capabilities chart in the Flaretek fittings
product section on Entegris’ website at
http://www.entegrisfluidhandling.com.
USER GUIDE
15
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
­­Installation Drawings
4100 Single-port Pressure Transducer with Flared Connection
3 conductor FEP-jacketed cable,
pigtail connection (BXX)
(Cable length up to 30 feet)
ELECTRICAL CONNECTOR INFORMATION
25.4 mm (1.00”)
Ø 47.8 mm
(1.88”)
A
Wire
4–20 mA
Output Version
Voltage
Output Version
Red
VDC+
24 VDC
(12–30 VDC)
24 VDC
(12–30 VDC)
Black
VDC–
Ground
Ground
White
Not used
0–5 VDC signal or
0–10 VDC signal
Inlet/Outlet
Port Connection
B
A
B
Flaretek
76.2 mm (3.00”)
25.2 mm (0.99”)
3⁄8”
Flaretek
76.2 mm (3.00”)
26.9 mm (1.06”)
1⁄2”
Flaretek
76.2 mm (3.00”)
29.0 mm (1.14”)
1⁄4”
3 conductor connector/
cable (DXX) (Cable length
up to 30 feet)
ELECTRICAL CONNECTOR INFORMATION
76.2 mm (3.00”)
Ø 47.8 mm
(1.88”)
4–20 mA
Output Version
Voltage
Output Version
Black
VDC–
Ground
Ground
11
Red
VDC+
24 VDC
(12–30 VDC)
24 VDC
(12–30 VDC)
12
White
Not used
0–5 VDC signal or
0–10 VDC signal
Pin
Wire
10
A
Inlet/Outlet
Port Connection
Receptacle with
plug removed
A
B
Flaretek
76.2 mm (3.00”)
25.2 mm (0.99”)
3⁄8”
Flaretek
76.2 mm (3.00”)
26.9 mm (1.06”)
1⁄2”
Flaretek
76.2 mm (3.00”)
29.0 mm (1.14”)
1⁄4”
B
11.4 mm (0.45”)
16
USER GUIDE
ENTEGRIS, INC.
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
4100 Single-port Pressure Transducer with NPT Connection
3 conductor Teflon®-jacketed
cable, pigtail connection (BXX)
(Cable length up to 30 feet)
ELECTRICAL CONNECTOR INFORMATION
25.4 mm (1.00”)
Ø 47.8 mm
(1.88”)
A
Wire
4–20 mA
Output Version
Voltage
Output Version
Red
VDC+
24 VDC
(12–30 VDC)
24 VDC
(12–30 VDC)
Black
VDC–
Ground
Ground
White
Not used
0–5 VDC signal or
0–10 VDC signal
Inlet/Outlet
Port Connection
B
A
B
1⁄4”
MNPT
76.2 mm (3.00”)
15.0 mm (0.59”)
1⁄4”
FNPT
82.3 mm (3.24”)
—
1⁄2”
FNPT
88.9 mm (3.50”)
—
3 conductor connector/cable (DXX)
(Cable length up to 30 feet)
ELECTRICAL CONNECTOR INFORMATION
76.2 mm (3.00”)
Ø 47.8 mm
(1.88”)
A
Receptacle
with plug
removed
4–20 mA
Output Version
Voltage
Output Version
Black
VDC–
Ground
Ground
11
Red
VDC+
24 VDC
(12–30 VDC)
24 VDC
(12–30 VDC)
12
White
Not used
0–5 VDC signal or
0–10 VDC signal
Pin
Wire
10
Inlet/Outlet
Port Connection
A
B
1⁄4”
MNPT
76.2 mm (3.00”)
15.0 mm (0.59”)
1⁄4”
FNPT
82.3 mm (3.24”)
—
1⁄2”
FNPT
88.9 mm (3.50”)
—
B
ENTEGRIS, INC.
11.4 mm (0.45”)
USER GUIDE
17
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
4210 Flow-through Pressure Transducer with Flared Connection
44.5 mm
(1.75”)
59.9 mm (2.36”)
44.5 mm” 47.2 mm
(1.75)
(1.86”)
5.1 mm (0.20”) (2)
3 conductor FEP-jacketed cable,
pigtail connection (BXX)
(Cable length up to 30 feet)
25.4 mm (1.00”)
A
6.4 mm (0.25”)
C
B
ELECTRICAL CONNECTOR INFORMATION
Wire
4–20 mA
Output Version
Voltage
Output Version
Red
VDC+
24 VDC
(12–30 VDC)
24 VDC
(12–30 VDC)
Black
VDC–
Ground
Ground
White
Not used
0–5 VDC signal or
0–10 VDC signal
Inlet/Outlet
Port Connection
1⁄4”
A
Flaretek
89.7 mm (3.53”)
3⁄8”
Flaretek
1⁄2”
Flaretek
3⁄4”
Flaretek
1” Flaretek
18
B
C
94.5 mm (3.72”)
18.5 mm (0.73”)
89.7 mm (3.53”)
98.3 mm (3.87”)
17.0 mm (0.67”)
96.0 mm (3.78”)
102.4 mm (4.03”)
21.6 mm (0.85”)
104.4 mm (4.11”)
108.5 mm (4.27”)
25.4 mm (1.00”)
112.5 mm (4.43”)
120.7 mm (4.75”)
30.5 mm (1.20”)
USER GUIDE
ENTEGRIS, INC.
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
44.5 mm
(1.75”)
44.5 mm”
(1.75)
59.9 mm (2.36”)
47.2 mm
(1.86”)
5.1 mm (0.20”) (2)
3 conductor connector/cable (DXX)
(Cable length up to 30 feet)
Receptacle with
plug removed
76.2 mm (3.00”)
11.4 mm (0.45”)
A
C
6.4 mm (0.25”)
B
ELECTRICAL CONNECTOR INFORMATION
4–20 mA
Output Version
Voltage
Output Version
Black
VDC–
Ground
Ground
11
Red
VDC+
24 VDC
(12–30 VDC)
24 VDC
(12–30 VDC)
12
White
Not used
0–5 VDC signal or
0–10 VDC signal
Pin
Wire
10
Inlet/Outlet
Port Connection
1⁄4”
A
Flaretek
89.7 mm (3.53”)
3⁄8”
Flaretek
1⁄2”
Flaretek
3⁄4”
Flaretek
1” Flaretek
ENTEGRIS, INC.
B
C
94.5 mm (3.72”)
18.5 mm (0.73”)
89.7 mm (3.53”)
98.3 mm (3.87”)
17.0 mm (0.67”)
96.0 mm (3.78”)
102.4 mm (4.03”)
21.6 mm (0.85”)
104.4 mm (4.11”)
108.5 mm (4.27”)
25.4 mm (1.00”)
112.5 mm (4.43”)
120.7 mm (4.75”)
30.5 mm (1.20”)
USER GUIDE
19
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
4130 Single-port Pressure Transducer
3 conductor
FEP-jacketed cable
Pigtail connection
(Cable length 12 feet)
25.4 mm (1.00”)
S
Z
Ø 47.5 mm (1.87”)
A
B
ELECTRICAL CONNECTOR INFORMATION
Wire
4–20 mA
Output Version
Voltage
Output Version
Red
VDC+
24 VDC
(12–30 VDC)
24 VDC
(12–30 VDC)
Black
VDC–
Ground
Ground
White
Not used
0–5 VDC signal or
0–10 VDC signal
Inlet/Outlet
Port Connection
1⁄4”
A
B
Flaretek
79.2 mm (3.12”)
25.2 mm (0.99”)
3⁄8”
Flaretek
79.2 mm (3.12”)
26.9 mm (1.06”)
1⁄2”
Flaretek
79.2 mm (3.12”)
29.0 mm (1.14”)
1⁄4”
MNPT
79.2 mm (3.12”)
15.0 mm (0.59”)
1⁄4”
FNPT
87.9 mm (3.46”)
—
1⁄2”
FNPT
94.5 mm (3.72”)
—
20
USER GUIDE
ENTEGRIS, INC.
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
4230 Flow-through Pressure Transducer
44.5 mm
(1.75”)
59.9 mm
(2.36”)
44.5 mm
(1.75”)
5.1 mm (0.20”)
3 conductor
FEP-jacketed cable
Pigtail connection
(cable length 12 feet)
25.4 mm (1.00”)
S
Z
A
C
B
ELECTRICAL CONNECTOR INFORMATION
Wire
4–20 mA
Output Version
Voltage
Output Version
Red
VDC+
24 VDC
(12–30 VDC)
24 VDC
(12–30 VDC)
Black
VDC–
Ground
Ground
White
Not used
0–5 VDC signal or
0–10 VDC signal
Inlet/Outlet
Port Connection
1⁄4”
Flaretek
A
95.3 mm (3.75”)
B
C
94.5 mm (3.72”)
18.5 mm (0.73”)
3⁄8”
Flaretek
95.3 mm (3.75”)
98.3 mm (3.87”)
17.0 mm (0.67”)
1⁄2”
Flaretek
101.6 mm (4.00”)
102.4 mm (4.03”)
21.6 mm (0.85”)
3⁄4”
Flaretek
110.0 mm (4.33”)
108.5 mm (4.27”)
25.4 mm (1.00”)
118.1 mm (4.65”)
120.7 mm (4.75”)
30.5 mm (1.20”)
1” Flaretek
ENTEGRIS, INC.
USER GUIDE
21
NT PRESSURE TRANSDUCER, MODELS 4100, 4210, 4130, 4230
For More Information
Please call your Regional Customer
Service Center today to learn
what Entegris can do for you.
Visit www.entegris.com and select
the Customer Service link for the
center nearest you.
Terms and Conditions
of Sale
All purchases are subject to Entegris’
Terms and Conditions of Sale. To
view and print this information, visit
www.entegris.com and select the
Legal Notices link from the footer.
Product Warranties
For Product Warranties, visit
www.entegris.com and select the
Legal Notices link from the footer.
22
USER GUIDE
ENTEGRIS, INC.
Entegris®, the Entegris Rings Design®, Creating a Material Advantage®, NT ® and Flaretek® are registered trademarks of Entegris, Inc.
Kalrez® and Viton® are registered trademarks of DuPont Dow Elastomers, L.L.C.
Teflon® is a registered trademark of E.I. du Pont de Nemours and Company.
Pillar® is a registered trademark of Nippon Pillar Packaging Company, Ltd.
ENTEGRIS, INC.
Corporate Headquarters | 129 Concord Road | Billerica, MA 01821 USA
Customer Service Tel. +1 952 556 4181
Customer Service Fax +1 952 556 8022
In North America 800 394 4083 | www.entegris.com
©2004–2014 Entegris, Inc.
P/N 01-1027289 (Rev. A 04/14)
All rights reserved
Printed in USA
3960-5426ENT-0414