DMM 939 3¾ Digit LCD Autoranging Multimeter This module is a complete multimeter in a panel instrument. It is especially useful in portable equipment and has 12 ranges: D.C. - 400mV, 4V, 40V and 400V (Maximum IP ±375V) D.C. - 4000µA, 40mA, 400mA, 4A, Resistance - 4kΩ, 40kΩ, 400kΩ and 4MΩ. 15mm (0.6“)Digit Height Digital Hold - M + k Autoranging uA mAV Low Battery Indication LED Backlighting D.C. Supply Isolator Continuity Beeper Output Stock Number Standard Meter METER PROTECTION Accuracy D.C. volts 400V D.C. volts all other Ohms 4 kilohms, 400 kilohms Ohms 40 kilohms, 4 megohms D.C. current adjusted Linearity (best straight line) Rollover error Operating temperature range Supply voltage (Vp+ to Vp-) Supply current (excluding backlight) Continuity detect low ohm range Converter isolation (D.C.) Backlight current (5V nom.) RECALIBRATING THE METER Power the meter with its typical working voltage and make connections as follows: - Connect Pin 8 to Pin 1. - Apply -3.700 volts D.C. between V/Ω (Pin 3) and COM (Pin 4). - Adjust R4 until meter displays -3.700V. LOW POWER OPERATION If ultra low power operation is required, then an isolated supply (battery) can be applied across V+ and V- to give this mode of operation. Note - Ensure that the supply does not exceed 11 volts. Low Power Operation Supply voltage (V+ to V-) Supply current Low battery threshold Min. 7 6.5 Typ. 9 1.5 7 Max. 11 2.4 7.5 DMM 939 Specification The meter should be protected against accidental application of high voltages on ohms and current ranges by a fast-blow 5A fuse. Min. Typ. Max. ±1 ±0.5 -1 ±2 ±0.3 ±0.75 ±4 0 5 5 9 7 1.5 50 +1 50 18 13 40 90 Unit % Rdg ± 1 count % Rdg ± 1 count % Rdg ± 8 counts % Rdg ± 9 counts % Rdg ± 2 counts count counts °C V mA kilohms V mA CONNECTOR SOURCING GUIDE Unit V mA V METHOD Cable Mounting IDC Supplied With Product DIMENSIONS All dimensions in mm (inches) 24.0 (0.94) 40.5 (1.59) POL/AC mA/µA K/m Ω/A BAT/V 44.0 (1.73) HΩ-DC/ LΩ-~ Panel cut-out 72 x 40 (2.83 x 1.57) Maximum panel thickness 3mm (0.12). Display in "Overload" 1 19 20 RES b c 1 2 mA I REAR VIEW 1.0 (0.04) 2.0 (0.08) 6.5 (0.26) 11.5 (0.45) 10.0 (0.39) OVERLOAD When the meter displays "OL" this indicates that the meter inputs have been taken beyond their working range. a 72.5 (2.85) 51.0 (2.01) 1 2 VCOM mA/µA 76.0 (2.99) Ω BEEP LED+ Vp- 19 20 Ø3.0 (0.12) a. b. c. d. e. 51 (2.01) REAR VIEW 24.5 (0.96) 24.0 (0.94) e V+ V/Ω mA V/Ω/A mA HOLD LEDVp+ 39.0 (1.54) d Ø5.7 (0.22) 24 (0.94) 61.5 (2.42) 51.0 (2.01) ON BOARD SOLDER LINKS LASCAR ELECTRONICS LTD. MODULE HOUSE, WHITEPARISH, WILTSHIRE SP5 2SJ UK TEL: +44 (0)1794 884567 FAX: +44 (0)1794 884616 E-mail: sales@lascar.co.uk LASCAR ELECTRONICS INC. LASCAR ELECTRONICS (HK) LIMITED 3750 West 26th Street, Erie, PA 16506 USA TEL: +1 (814) 835 0621 FAX: +1 (814) 838 8141 E-mail: us-sales@lascarelectronics.com FLAT C, 5/FL., LUCKY FTY. bldg., 63-65 HUNG TO ROAD KWUN TONG, KOWLOON, HONG KONG TEL: +852 2797 3219 FAX: +852 2343 6187 E-mail: b4lascar@samsongroup.com.hk PIN FUNCTIONS 1. 2. 3. 4. V+ VV/Ω COM Positive power supply connection.Not isolated. Pins 1 and 2 are used for mode selection and low power operation. Negative power supply connection. Positive Volts measuring input. Connect to Ω (Pin 10) to measure resistance. Ground for analogue section of the A/D converter, it is held actively at 2.9V (typ.) below V+. COM must not be allowed to sink excessive current (>100µA) by connecting it directly to a higher voltage. Positive milli-Amperes measuring input. Tri-State Input Connections Tri-state input, connect to V+ to display 'mA' annunciator. Pin No. V+ OPEN or COM VConnect to COM to display 'µA' annunciator. Note - mA/µA 6 mA mA Do not connect to Vannunciators will not be displayed unless Pin 7 is connected to V-. 7 V A Ω Tri-state input, connect to COM to operate voltage range and to 8 HΩ-DC LΩ Do not connect to Vdisplay 'V' annunciator. Connect to V+ to operate Resistance range 11 HOLD AUTO Do not connect to Vand to display relevant 'Ω' annunciator or connect to V- to display either mA or µA, selection between each of these annunicators is determined by Pin 6. Tri-state input, connect to V+ to select High Ohms and D.C. range input. Connect to COM for Low Ohms. All the ranges will show the appropriate annunciator, MΩ, kΩ or ~ symbol. Positive micro-Amperes measuring input. Ohms measuring input. Connect to Pin 3 when measuring resistance. Tri-state input, connect to V+ to hold display reading. The continuity beeper output is designed to drive a piezo-electric transducer at approximately 2KHz with a voltage swing of V+ to V-. Note - The beeper output off state is at the V+ rail. Negative supply to LEDs for backlighting. LED+ to LED- supply voltage of 5V. Positive supply to LEDs for backlighting. Positive isolated power supply connection. It is recommended that the meter is powered using these pins. Negative isolated power supply connection. No pins fitted. 5. mA 6. mA/µA 7. V/Ω/A 8. HΩ-DC/ LΩ 9. µA 10. Ω 11. HOLD 12. BEEP 13. LED14. LED+ 15. Vp+ 16. Vp17-20 VARIOUS OPERATING MODES 1 11 15 Vp+ V+ HOLD 16 VpHΩ-DC/ 8 3 V LΩ 7 10 V/Ω/A Ω 12 BEEP 4 COM SUPPLY (5-18V) Resistance to be placed across these terminals for measurement S2 N.C. S3 N.C. Piezo-electric Transducer + Input Current to be measured - Autoranging Resistance Meter Ranges: Low Ohm - 4kΩ, 40kΩ High Ohm - 400kΩ, 4MΩ 15 SUPPLY (5-18V) 16 V/Ω 3 10 S4A CURRENT S5A COM *RS 0.1Ω 5A Quick Blow Fuse Vp+ Vp- 1 V+ HOLD BEEP 9 µA 5 mA 4 COM 11 HΩ-D.C./ 8 LΩ V Ω V/Ω/A mA/µA V- S5A Rs 0.1Ω * 5A Quick Blow Fuse 5A Quick Blow Fuse S1 8 1 Vp+ V+ HΩ-DC/ 8 16 LΩ Vp11 9 HOLD µA 15 SUPPLY (5-18V) S2 S3 12 7 6 Piezoelectric Transducer + 5 4 mA mA/µA 6 COM V2 V/Ω/A 15 SUPPLY (5-18V) S2 16 3 S5B Input Voltage to be measured 7 Autoranging D.C. Current Meter Ranges: µA - 4000µA, 40mA mA - 400mA, 4000mA *Note - Rs is the current sensing resistor and must be capable of handling as a minimum the full scale current of 4 amps. To allow for overload conditions, it is recommended that a 6W, 0.5% resistor is fitted. It must be fitted when EITHER of the current ranges are being used. 4 8 1 Vp+ V+ Vp- HOLD 11 S2 V V/Ω/A 7 COM Autoranging Volt Meter S4B S5B 2 Autoranging Voltage, Current and Resistance Meter *See note on Current Meter diagram. SWITCHES S1. Switch 1 is an on/off switch. S2. Switch 2 is used to hold the display reading. S3. In Voltage mode, switch 3 is used to select D.C. measurement and the ~ symbol. In Resistance mode, this switch is used to select either high or low ohms ranges. S4. Switch 4 selects either the Voltage, Current or Resistance ranges. S5. Switch 5 selects between Current ranges. HEALTH AND SAFETY 1. 2. 3. 4. When the DMM is used for current or resistance measurements, an external fuse MUST be fitted. When measuring voltages in excess of 60V peak, the system within which the DMM is incorporated, should conform to relevant sections of IEC 1010 (Safety Requirements for Electrical Equipment for Measuring, Control and Laboratory Use). If this product is to be used for measuring mains supply voltages, external transient protection should be adopted. To comply with the Low Voltage Directive (LVD 93/68/EEC), input voltages to the module’s pins must not exceed 60Vdc, then fit scaling resistors externally to the module. The user must ensure that the incorporation of the DPM into the user’s equipment conforms to the relevant sections of BS EN 61010 (Safety Requirements for Electrical Equipment for Measuring, Control and Laboratory Use). Specifications liable to change without prior warning DMM 939 Issue 3 August/1997 M.C. Applies to DMM 939/5