I. Introduction to Biomedical Engineering Electrical Safety • Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 1 • Industry Regulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 3 • BMET Responsibilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 4 • History of Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . .page 4 Made Easy II. Fundamental Concepts of Electrical Safety . . . .page 6 III. Physiological Effects of Electricity Table Of Contents • Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 7 • Macroshock and Microshock . . . . . . . . . . . . . . . . . . . . . . . . .page 7 • High Frequency Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 8 • Potential Equipment Safety Hazards . . . . . . . . . . . . . . . . . . .page 8 • Equipment Safety Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 8 • Electrical Shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 8 • The Electrical Power System . . . . . . . . . . . . . . . . . . . . . . . . . .page 8 • Contact with a Ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 9 • Skin Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 9 • Leakage Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 10 • Electrical safety Power System Devices . . . . . . . . . . . . . . . .page 10 • Ground Fault Circuit Interrupter . . . . . . . . . . . . . . . . . . . . .page 10 • Isolation Transformer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 10 • Equipotential Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . .page 11 • Codes and Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 11-12 IV. The AAMI Standard Bibliography Glossary www.bapcoinfo.com . . . . . . . . . . . . . . . . . . . . . . . . . .page 13-32 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 32 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 33 T he field of medical instrumentation is by no means new. Many instruments were developed as early as the nineteenth century – for example, the electrocardiograph, first used by Einthoven at the end of that century. Progress was rather slow, however, until after World War II when a surplus of electronic equipment such as amplifiers and recorders became available. At that time, many technicians and engineers, both within industry and on their own, started to experiment with and modify existing equipment for medical use. This process occurred during the 1950’s and the results were often disappointing. For the experimenters soon learned that physiological parameters are not measured in quite the same way as physical parameters. They also encountered a severe communication problem with the medical profession. During the next decade many instrument manufacturers entered the field of medical instrumentation, but development costs were high, and the medical profession and hospital staffs were suspicious of new equipment and often uncooperative. Many developments with excellent potential seemed to have become lost causes. It was during this period that some progressive companies decided that rather than modify existing hardware, they would design medical instrumentation specifically designed for medical use. Although it is true that many of the same components were used, the philosophy was changed; equipment analysis and design were applied directly to medical problems. A large measure of help was provided by the U.S. government, in particular NASA (National Aeronautics and Space Administration). The Mercury, Gemini and Apollo programs needed accurate physiological monitoring for the astronauts; consequently, much research and development money went into this area. The aerospace medicine programs were expanded considerably, both within NASA facilities and through grants to Universities and Hospital research units. Some of the concepts ad features of patient monitoring systems presently in use in hospitals throughout the world evolved from the base of astronaut monitoring. The use of adjunct fields, such as biotelemetry, also found some basis in the NASA programs. Also, in the 1960’s, an awareness of the need for engineers and technicians to work with the medical profession developed. All the major engineering technical societies recognized this need by forming "Engineering Medicine and Biology" subgroups and new societies were organized, such as the Biomedical Engineering Society. Along with the medical research programs at the universities, a need developed for courses and curricula in biomedical engineering and today almost every major university has some type of biomedical engineering program. However, much of this effort is not concerned with biomedical instrumentation per se. One of the problems of "biomedical engineering" is defining it. The prefix bio- of course, denotes something connected with life. Biophysics and biochemistry are relatively old disciplines in which basic sciences have been applied to living things. One school of thought subdivides bioengineering into different engineering areas – for example, biomechanics, and bioelectronics. These categories usually indicate the use of that area of engineering applied to living rather than to physical components. Bioinstrumentation implies measurement of biological variables, and this field of measurement is often referred to as biometrics, although the latter Electrical Safety Made Easy I. Introduction to Biomedical Engineering -1- term is also used for mathematical and statistical methods applied to biology. Electrical Safety Made Easy I. Naturally, committees have been formed to define these terms; the professional societies have become involved. The latter includes the IEEE Engineering in medicine and biology group, the ASME Biomechanical and Human Factors division, the Instrument society of America and the American Institute of aeronautics and Astronautics. Many new cross-disciplinary societies have also been formed. Several years ago an engineering committee was formed to define bioengineering. This was subcommittee B (Instrumentation) of the Engineers Joint Council Committee on Engineering on engineering Interactions with biology and Medicine. Their recommendation was that bioengineering be defined as application of the knowledge gained by a cross fertilization of engineering and the biological sciences so that both will be more fully utilized for the benefit of man. More recently, as new applications have emerged, the field has produced definitions describing the personnel who work in it. A tendency has risen to define the biomedical engineer as a person working in research or development in the interface area of medicine and engineering, whereas the practitioner working with the physicians and patients is called a clinical engineer. One of the societies that has emerged in this interface area is the Association for the Advancement of medical Instrumentation (AAMI). This association consists of both engineers and physicians. In late 1974, they developed a definition that is widely accepted: -2- "A clinical engineer is a professional who brings to health care facilities a level of education, experience and accomplishment which will enable him to responsibly, effectively, and safely manage and interface with medical devices, instruments and systems and the use thereof during patient care, and who can, because of this level of competence, responsibly and directly serve the patient and physician, nurse and other health care professionals relative to their use of and other contact with medical instrumentation." Most engineers go into the profession through the engineering degree route, but many start out as physicists or physiologists. They must have at least a B.S. degree and many of them have M.S. or Ph.D. degrees. Another popular term, also coined in recent years, the biomedical equipment maintenance technician (BMET) is defined as follows: "A biomedical equipment maintenance technician (BMET) is an individual who is knowledgeable about the theory of operation, the underlying physiologic principles and the practical, safe clinical application of biomedical equipment. His capabilities may include installation, calibration, inspection, preventive maintenance and repair of general biomedical and related technical equipment as well as operation or supervision of equipment control, safety and maintenance programs and systems." This was also an AAMI definition. Typically, the BMET has two years of training at community college. This person is not to be confused with a medical technologist. The latter is usually used in an operative sense, for example in blood chemistry and in the taking of electrocardiograms. The level of sophistication of the BMET is usually higher than that of the technologist in terms of equipment, but possibly lower in terms of the life sciences. In addition, other titles have been used, such as hospital engineer and medical engineer. In one hospital the title biophysicist is preferred for their biomedical engineers, for reasons best known to themselves. These definitions are all noteworthy, but whatever the name, this age of the marriage of engineering to medicine and biology is destined to benefit all concerned. Improved communication among engineers, technicians and doctors, better and more accurate instrumentation to measure vital physiological parameters and the development of interdisciplinary tools to help fight the effects of body malfunctions and diseases are all a part of this field. The name itself is actually not all too important; however, what the field can accomplish is important. With this point in mind, we will be using the term biomedical engineering for describing the field in general and biomedical instrumentation for the methods of measurement within the field. INDUSTRY REGULATION In it’s infancy, biomedical equipment was best serviced and maintained by the original equipment manufacturer as there were few, if any, qualified technicians who could do so adequately. A lack of standards linking the technology, an absence of service literature, and the general feeling from hospital personnel that, unless trained by these OEMs, no one would be able to provide a suitable level of services. Thus the early BMET was relegated to maintain and repair only the simplest of clinical equipment such as centrifuges, suction pumps and other such equipment. Several factors soon arose to alter this conception. First, agencies such as the Food and Drug Administration (FDA) developed subdivisions under which fell the area of medical instrumentation. This brought about a set of standards and practices for manufacturing of hospital equipment, as well as the maintenance and repair of such. Secondly, other agencies, under which hospital equipment was monitored and controlled, invoked requirements such as the release of service literature on any new equipment purchased or manufactured to the hospital. This literature, the law stated, must contain the procedures for calibration and alignment of said equipment. Thirdly, the training of biomedical equipment technicians began to catch up with the advancing technology so that the level of competence was significantly improved. Finally, through the feeling that they provided the only service available, the OEMs began to charge significantly more for their service contracts each year. Surprisingly, this was not a major issue to the hospital community until the middle of 1980 when the government, in an effort to control the rising cost of medical treatment, and the subsequent rise in medical insurance claims, imposed major restrictions upon hospitals. These restrictions, known as Diagnostic related Groups (DRGs) limited the amount that medical insurance groups would reimburse both hospitals and physicians for each type of illness for which a patient might enter the hospital. Without going into great detail about DRGs, suffice it to say that their introduction brought with it a need for hospitals to lower their operating costs. One area in which these costs could be lowered was in the area of biomedical instrumetation. Where once a hospital would make a major equipment expenditure, such as buying a new chemistry analyzer, every five or so years, they began keeping sophisticated equipment such as this for longer periods. So doing, the rising OEM service contract prices suddenly became an issue and the responsibility for these equipment fell upon the BMET as a lower cost alternative. As more and more responsibility fell upon the BMET within the Electrical Safety Made Easy I. -3- Electrical Safety Made Easy I. hospital for larger and more sophisticated instrumentation, new problems arose for the hospitals. In order for the equipment to be adequately maintained, more BMETs needed to be employed, each having sufficient training in the sophisticated equipment involved. Hospitals were faced with weighing the additional cost of salaries, benefits and the costs of attending additional manufacturers schools and seminars with the cost of the OEM contracts. In very large hospital facilities, this was not such a significant problem as it was with smaller facilities. In many smaller facilities, typically those under 200 patient beds, it was simply not feasible to consider an in-house BMET program and, thus, these facilities were forced to pay for the OEM contracts. be employed on sensitive, lifesaving medical equipment. Again, the government became involved. Federal and state agencies began to require all hospital facilities to comply with stringent safety guidelines for patient related equipment. This included a routine electrical safety program under which all hospital equipment would be tested, evaluated and forced to comply with. Typically, the BMET provides one or more of the following services to hospital facilities: In an effort to satisfy applicable federal and state requirements as well as to insure that clinical equipment are utilized both safely and to their optimum, some BMET shops provide periodic in-service and education services. This often ranges from simple electrical safety lectures to the proper use of a newly acquired piece of equipment for which hospital personnel might not yet be familiar. BMET RESPONSIBILITIES This brings us to the main purpose of this document. Periodic testing of patient and non-patient hospital equipment to insure that they meet the safety requirements and guidelines as set fourth by applicable federal and state regulations. This in no way measures the equipment ability to perform the task for which is was designed. Corrective Maintenance -4- Non-scheduled repair of hospital equipment other than during Preventive Maintenance inspections. These repairs range from replacement of minor parts or components to total equipment overhauls as may be required. In an effort to further provide a cost savings to the hospital facility, some BMETs attempt to repair equipment to the smallest component level, rather than simply replacing circuit boards or sub-assemblies which may be very costly. Other BMETs follow a philosophy that component-level repair entails far too many risks of repeated failure or improper component soldering techniques to Preventative Maintenance Inspections Periodic testing and evaluation of patient and non-patient related equipment to insure that they operate within the guidelines set fourth by federal, state and manufacturer guidelines. This includes calibration and alignment of said equipment along with the repair and/or replacement of component parts in order to bring the equipment into compliance. It is the belief of most BMETs that good, quality preventive maintenance inspections further reduce hospital costs by reducing equipment down-time for repairs. Inservice & Education Electrical Safety Inspections HISTORY OF ELECTRICAL SAFETY The introduction of electricity into commerce at the close of the 19th century carried with it the need to know how to deal with it safely. For the hospital environment, much of the electrical safety programs started after World war II. Shortly thereafter, federal funds via the Hill-Burton Act increased the number of hospital facilities. Grounding for electrical safety was implemented because most electrical accidents in the home and industry occurred because exposed metal was energized. The National Electrical Code stated that "non-current carrying" metal parts of the electrical apparatus could prevent such accidents. Unfortunately, by the end of the 1950’s, electrical power in the hospitals was supplied in a haphazard, eclectic fashion. In the hospital environment of the sixties, electricity came to be used more often on, in and around patients to a degree beyond conception only a few years previously. It was early in 1961 that there appeared the first news that "microshock" (small electric currents applied to a conductor near the heart) was happening in the medical field. In 1969, Carl Walter, M.D., who was at the time a well-known surgeon, stated that, "1,200 patients were being accidentally electrocuted in U.S. hospitals each year." Although many engineers and health care professionals believed that Dr. Walter’s estimates were unrealistically high, the concept of microshock suddenly became publicized. Then, on June 16th, 1970, Ralph Nader (a household word because of his 1965 book, Unsafe at Any Speed) gave a speech in Detroit. In this speech, he stated that, "1200 annual electrocutions in US hospitals was a very least figure" and quoted other experts indicating that the real number might be significantly higher. The event was picked up by a wire service and run under an arresting headline the next day: Hospital electrocutions cited Detroit (UPI) – "Accidental electrocutions claim 5,000 lives in American Hospitals every year but seldom get reported due to the "close nature of hospitals" consumer critic Ralph Nader said yesterday. During the 1970’s, several proposals and regulations were introduced to manage this suspected problem in hospitals. In 1971, the National Fire Protection Association published a recommended standard (76BM) to help hospital engineers understand the principles of electrical safety and coordinate a program of medical equipment electrical testing in their facilities. In the spring of 1972, Underwriters laboratories issued document UL544 Medical and Dental Equipment, which was intended to serve as a guideline for medical equipment manufacturers. In 1970, AAMI (the Association for the Advancement of medical Instrumentation) published a firstdraft standard for electrical leakage current standards that was adopted as an American national standard in 1978. Electrical Safety Made Easy I. Probably the most dramatic proposal was the Joint Commission on the Accreditation of Hospitals (now the Joint Commission on the Accreditation of Healthcare Organizations – JCAHO) 1976 recommendation that hospitals maintain equipment control programs to provide for electrical safety training, create a documented preventive maintenance program, and perform semi-annual safety and performance equipment inspections and annual inspections of electrical receptacles. Today, health care institutions in the United States support clinical engineering programs that provide ongoing electrical safety and performance testing as well as preventive maintenance and repair of medical equipment. Typically, these programs use the most recent editions of NFPA 70, NFPA 99, NFPA 101, AAMI Recommended Standards, and Joint Commission accreditation manuals for their reference standards. -5- S Electrical Safety ome of the most basic concepts of electricity must be understood in order to grasp the potential hazards of electrical devices as they are used in a clinical setting such as a hospital. Some of these important ideas concerning electrical safety in a medical environment are explained below. Made Easy The Nature of Electricity II. Fundamental Concepts Of Electrical Safety Electrical current flowing through a conductor is the result of electrons moving from the outer shell of atoms induced by an electric field that is imposed on a conductor. This field can be caused by any voltage generating source, such as a local utility company, a battery, or a chemical reaction. In the hospital, this voltage source is provided by the local electrical company and is redirected through a series of transformers to 240, 208, or 120 volts alternating current (AC). The amount of current that flows through an electrical device is determined by the resistance that the device is designed to provide to the applied field. This relationship is called "Ohm’s Law" and is written as: I=V R BAPCO’S SA2115 “Safety Certifier” is the most complete safety analyzer available that provides everything you need to safety test: • Patient to nonpatient, medical to commercial devices • 110 V @ 200amp to 240V @ 15 amp • Automatically to AAMI standards -6- Where "I" is the current in amperes that flows through a device or conductor, "V" is the magnitude in volts of the applied electric field, and "R" is the resistance that the device or conductor has, measured in Ohms. Thus, if an electromedical device such as an aspirator is plugged into a 120 volt receptacle and the unit requires 3.0 amperes, then the resistance that the aspirator provides to the voltage is: 120 volts = 40 Ohms 3 amperes a measure of the amount of power that an electrical device consumes during operation is defined as the product of voltage and current and is recorded in Watts. From the previous example: Power (watts) = (V) (I) = (120) (3) = 360 Watts B ecause the amount of electricity that will flow through any medium depends on the resistance that it encounters, current can flow through the human body and cause various effects. If direct current (polarized, nonchanging) or high frequency alternating current passes through the body, heating effects and ultimately burns will occur. It is this effect that is intentionally created when electrosurgical generators are used to cut tissue and coagulate fluids. If low frequency alternating current is applied to the body, muscular polarization and depolarization take place that can ultimately create a "circus movement" in the heart muscle, resulting in fibrillation and death. It is this effect that normally accounts for death due to electric shock. Unfortunately, the typical resistance of the human body in combination with the frequency of commercially generated electricity (60 hertz) can create a potentially hazardous situation in the hospital environment. MACROSHOCK AND MICROSHOCK The effect of electric shock on the human body can be anything from barely perceptible tinges, to muscle spasms, to death. Each can occur from small or large currents, depending on how the currents are introduced into the body. Large currents (milliamperes or larger) that are introduced into the body from one external point to another (arm to leg, for example) can result in macroshock. If small currents (as low as 10 microamperes) are introduced into the body from an external source such as a catheter or cardiac pacing wires, the resistance to the heart muscle can be very low, and electrocution can occur from microshock. The chart below describes the effect of different levels of current that are introduced into the human body by creating a voltage across each arm. For microshock, the same effect can be produced with current levels that are only 1/10,000 as great as those listed. CURRENT EFFECT .001 Ampere (1 Milliampere) "Tingling", threshold of perception .020 Amps (20Ma) Muscle Spasms, hard to release grip .050 amps (50 Ma) Pain, possible fainting, transient interruption of respiration .100 Amps (100 Ma) Ventricular fibrillation >5 Amps Sustained myocardial contraction, possible burns, temporary respiratory paralysis Electrical Safety Made Easy III. Physiological Effects Of Electricity -7- Electrical Safety Made Easy HIGH FREQUENCY EFFECTS EQUIPMENT SAFETY TESTING As mentioned previously, the primary effect of high frequency current (500 kilohertz to 2 megahertz) is to heat tissue as it is concentrated in a certain area. The amount of heat that is generated depends on the amount of current applied and the area that the current passes through. Tests to determine the electrical safety of medical devices include a measurement of the continuity as well as the leakage current between the chassis and the cord grounding conductor. Equipment that has been designed with patient leads or contact points is also measured for current leakage from these points. Figures 1 and 2 show typical circuits that may be used to measure this leakage current. This relationship is : Heat = I x I A III. Where "I" is the applied current and "A" is the area that the current flows through. POTENTIAL EQUIPMENT SAFETY HAZARDS Because current flow through the body can be hazardous if it is of a certain amplitude and frequency, stray currents must be eliminated from medical equipment. The best method to prevent leakage currents is to ensure that all conductive parts of the equipment are connected to the hospital grounding system through the power plug. Also, to minimize patient contact with current leakage on ECG signal leads, electrically isolated amplifiers should be used. -8- In theory, the use of line isolated power systems helps to minimize safety hazards by isolating the neutral power line from earth ground. A line isolation monitor is also installed with these systems to identify visually and aurally the presence of leakage current between the isolated power line and the grounded conductor. Other power system devices that help to identify or disconnect power sources when leakage currents are present include ground fault detectors (GFDs) and ground fault interruptors (GFIs). ELECTRICAL SHOCK The three ingredients in the scenario of electrocution are : 1. contact with the live conductor of a grounded electrical system 2. contact with a ground 3. diminished skin resistance THE ELECTRICAL POWER SYSTEM Domestic voltages in the United States are from 120V for lighting and small appliances, and 240V for electrical ranges and dryers. In the hospital, outlets are 120V while 277V and 208V are often used for fixed lighting and special receptacles. Typically, a voltage of 240V is provided by the service drop to the power meter of the hospital. This line consists of a bare cable plus two wires having black insulation. The bare cable is connected to "earth" ground via a water pipe or grounding rod. It is this connection that makes the electrical service a grounded system. NOTE : The national electric Code defines ground as "a conducting connection, whether intentional or accidental, between an electrical circuit or equipment and the earth, or to some conducting body which serves in place of the earth." There are three reasons to ground electrical systems: 1. Grounding of the system protects against introduction into the structure, via the wiring, of high voltages (with respect to ground) such as might arise from lightning or insulation failure in a high voltage pole transformer. 2. Tying the circuit to ground obviates the multiple problems that might ensue were the domestic circuits "floating" at some undefined but high voltage relative to ground. 3. To facilitate operation of over current protective devices (fuses) which are located in the "live" (ungrounded) side of each circuit. CONTACT WITH A GROUND For purpose of illustration, we will use an electrocardiograph (ECG) as our example of a medical device. Back in the 1950s, the standard ECG was fitted with a two-prong plug. The patient was connected to the chassis ground via the right-leg electrode, and one wire of the power cord was connected to the chassis through a 200,000-Ohm resistor. This connected the chassis to ground via the natural conductor in the power system. If the plug was inserted with polarity reversed, then the patient was connected to the "hot" side of the power system. However, the 200KOhm built-in resistor limited the current to 0.6 milliamperes if the patient or bystander provided a pathway to ground. Added safety to the machine included a fivemilliamp fuse in series with the right-leg electrode. As it turns out this approach was not practical because the fuse kept blowing after the patient was defibrillated. The codes changed this practice and today’s ECG machines do not ground any of the electrodes to earth ground. In fact, the "front end" electronics on most medical devices have an isolated power supply and grounding system. Plus, the manufacturers use plastic cases minimizing the exposed metal. SKIN RESISTANCE Water affects skin resistance, and for a given voltage, resistance determines current. When dry, skin has a resistance of upwards 100,000 ohms. If there is an accidental application of 120 volts between the two hands, only 1.2 milliamperes will flow. In a wet environment or on a hot and humid day, that same current path may come to have a resistance as low as 1,000 ohms, resulting in a current flow of 120 milliamperes. Electrical Safety Made Easy III. If a patient is exposed to electrical current, it takes about 1 milliampere of 60 Hz AC for a threshold of sensation. The sensation becomes uncomfortable if 5 milliamperes is attained. Strong muscle spasms appear at currents of 10 to 20 milliamperes ("let-go" current before sustained muscular contraction). A current flow of 100 milliamperes or greater may induce ventricular fibrillation and death. These values are for currents introduced at the body surface. They are termed macroshocks and require two points of external body contact. There is a possibility of a direct electrical path to the heart via a needle or catheter in an artery or vain. This directly reduces the resistance and current threshold. Small amounts of current (100 microamperes) can be potentially lethal. Electrical shock in these circumstances is termed "microshock". The frequency of the current is also important when considering electrical shock. If the frequency is raised above 1 KHz, these current levels no longer produce such sensations or life-threatening phenomenon. High frequencies in the megahertz region will not cause shock at all. -9- LEAKAGE CURRENT Electrical Safety Made Easy III. All electrically operated devices have some current that flows from the energized electrical portions of the device to the metal chassis. This current is referred to as leakage current and has two components, capacitive and resistive. Capacitive leakage current results from distributive capacitance between two wires or a wire and a metal chassis case. Components that cause capacitive leakage currents are RF filters, power transformers, power wires, and any device that has stray capacitance. Resistive leakage current arises from the resistance of the insulation surrounding the power wires and transformer primary windings. ELECTRICAL SAFETY POWER SYSTEM DEVICES Several techniques are available to protect clinicians and patients from electrical shock. The most common ones are the ground fault circuit interrupter (GFCI), the isolation transformer and equipotential grounding. GROUND FAULT CIRCUIT INTERRUPTER The GFCI acts like a circuit breaker when it senses an inequality of as little as 6-ma between the "hot" and neutral wires of the circuit. It is mandated by the national electric code where electrical outlets are so situated that simultaneous contact with a grounded surface is especially likely. "Wet locations" such as a whirlpool bath or bathroom are examples of GFCI usage. -10- In the GFCI, current in the "hot" wire passes through one transformer coil on the same transformer core. From the design, the net flux is zero, when the currents are equal. If the current is unbalanced (i.e. current flowing through ground) a net flux is induced across the third coil. This current will trigger a switch opening the "hot" line. The GFCI affords economical protection against electrical shock. However, the GFCI depends upon an active system, and the integrity of it’s mechanical operation is crucial. ISOLATION TRANSFORMER The Isolation transformer offers electrical safety by converting grounded power into ungrounded power. This is accomplished by grounding the primary winding of the transformer and not grounding the secondary winding. Isolation is not perfect for two reasons. First, the isolation transformer has some stray capacitance to ground. Second, every medical device that is attached to the transformer possesses stray capacitance which causes some degree of coupling between its power-carrying wires and the grounded frame. To monitor the system isolation, a line isolation monitor (LIM) is employed. Its function is to analyze the entire isolated circuit and quantify its degree of isolation from ground. The LIM provides visual and audible alarm signals when the predicted ground-seeking current exceeds a specified magnitude. The LIM does not indicate an existing current flow, rather it predicts the current that would flow if a short circuit were to develop between isolated wire and ground. In the operating room, isolated power systems were first installed as a measure directed against sources of ignition rather than electrical shock. Flammable anesthetics such as diethyl ether and cyclo-propane were used. Today, most anesthesiologists use non-flammable anesthetics and isolated power is not required. EQUIPOTENTIAL GROUNDING Another technique that reduces electrical shock is equipotential grounding. This is accomplished by adding another grounding wire from each chassis to a central point that is in parallel with the third wire in the power cord. If the chassis of all equipment is at the same potential there will be no current leakage to the heart. This technique has its advantages and disadvantages and is not typically used in today’s health care environment. CODES AND STANDARDS In the 1960s, all aspects of hospital activities involving fire and explosion hazards (including electrical shock and emergency electrical power) were seen to be in need of some guidelines and/or standards. As it turns out, in the standards arena of the 1970s, patient safety was twice corrupted. First, standards were generated in ignorance (solutions were imposed before the problems were defined) and second, safety was blatantly exploited for ego-serving, bureaucratic, and commercial gains. Today, there are two standards that specify electrical safety: NFPA 99 and the AAMI Standard for Safe Current Limits. The 1996 edition of NFPA 99 includes numerous technical changes that relate to leakage current and it now more closely correlates to the international standards. Originally, NFPA 76B covered the electrical wiring system and electrical appliances. The summary of the standard as it relates to electrical safety is in the Table on the following page. Electrical Safety Made Easy III. One further comment about standards. On may 30, 1972, Underwriter’s laboratories (UL) released the first edition of UL-544, Standard for Safety for Medical and Dental Equipment. UL-544 deals with details regarding enclosure safety, mechanical stability, and integrity of insulation. It’s leakage current, isolation requirements, and test methods are similar to but not identical with those of AAMI and NFPA. -11- NFPA - 99 1996 Electrical Safety Made Easy III. Chassis Source Current, Cord Connected (Portable) Ground Open Ground Intact With Isolated Patient Connection 300µa 300µa 300µa 100µa With Nonisolated patient Connection 300µa 300µa 300µa 100µa Likely to Contact Patient 300µa 300µa 300µa 100µa No patient Contact 300µa 300µa 300µa 100µa Chassis Source Current, Permanently Connected Ground Open Ground Intact Ground Ground Open Intact With Isolated Patient Connection 5000µa 5000µa 5000µa 100µa With Nonisolated patient Connection 5000µa 5000µa 5000µa 100µa Likely to Contact Patient 5000µa 5000µa 5000µa 100µa No patient Contact 5000µa 5000µa 5000µa 100µa Lead to Ground Current Ground Open Ground Intact Ground Ground Open Intact With Isolated Patient Connection 50µa 10µa 50µa 10µa With Nonisolated patient Connection 100µa 100µa 100µa 50µa Ground Open Ground Intact With Isolated Patient Connection 50µa 10µa n/a n/a With Nonisolated patient Connection 50µa 50µa n/a n/a Sink Current (Isolated Test) Ground Open Ground Intact With Isolated Patient Connection n/a 20µa n/a n/a With Nonisolated patient Connection 50µa 50µa n/a n/a Ground Open Ground Intact Ground Ground Open Intact Ground Intact Ground Ground Open Intact Ground Open 4 Oz Ground Intact Ground Ground Open Intact Ground Open Ground Intact Ground Ground Open Intact Lead to Lead Current Ground Impedance 0.5 Ohms Existing System 0.2 Ohms New Construction 0.1 Ohms Receptacle Indicators Ground Open Hospital Grade Emergency System Isolated Ground Retention Force, Ground Blade -12- AAMI/AAMI ESI 1993 GFCI Trip Current Ground Ground Open Intact Ground Ground Open Intact Ground Ground Open Intact Listed Red Orange 6mA T he AAMI Electrical Safety Committee began deliberating the issue of safe risk current limits for electromedical apparatus in 1967 and first published recommended limits and test methods in 1971. The first edition of this American National Standard was approved by the American National Standards Institute (ANSI) in 1978, and a second edition was approved and published in 1985. Work on this third edition involved the dedicated efforts of concerned health care professionals, industrial scientists, and government representatives. AAMI expresses its gratitude for the service of all persons involved in the development of this standard. fault conditions and provides risk current levels for each of these states. Also included in this standard are test methods for equipment with nonconductive enclosures and for doubleinsulated equipment as well as limits for earth leakage or earth risk currents. In this third edition of the standard, the risk current limits have been raised to be compatible with, although not identical to, the limits set forth by the International Electrotechnical Commission (IEC) in its standard, Medical electrical equipment—Part 1: General requirements for safety (IEC 601-1-1988). As in the IEC standard, this edition of Safe current limits for electromedical apparatus (ANSI/AAMI ES1—1993) introduces the concepts of normal operating conditions and single Suggestions for improving this standard are invited. Comments and suggested revisions should be sent to Technical Programs, AAMI, 3330 Washington Boulevard, Suite 400, Arlington, VA 22201-4598. This standard is intended primarily for the testing of electromedical apparatus intended for use in or near the patient care area. This standard should be considered flexible and dynamic. As technology advances and new data are brought forward, the standard will be reviewed and, if necessary, revised. NOTE — This foreword does not contain provisions of the American National Standard, Safe current limits for electromedical apparatus (ANSI/AAMI ES1—1993), but does provide important information about its development and intended use. Electrical Safety Made Easy IV. The AAMI Standard For Safe Current Limits For Electromedical Apparatus -13- Electrical Safety Made Easy IV. SAFE CURRENT LIMITS FOR ELECTROMEDICAL APPARATUS 1 SCOPE 1.1 Inclusions This standard sets risk current limits and referee test methods for electromedical apparatus intended for use in the patient care vicinity and also sets limits for nonpatient-contact electromedical apparatus. The standard applies to line- and battery-powered apparatus and to apparatus used singly or with properly connected accessory equipment. When more than one electromedical apparatus is powered by a single power cord, the equipment assembly acts like a single apparatus in terms of risk current limit requirements, and shall be considered as such for the purposes of this standard. The safety and performance criteria defined in this standard are intended for use in design qualification by the device manufacturer. NOTE — The referee test methods of Section 5 are intended to provide means by which conformance with the standard can be established. These tests are not intended for use in verifying the performance of individual devices in routine quality assurance inspections. Also, referee tests allow for the use of alternative methods for design qualification, provided that devices so qualified will also meet the requirements of this standard when tested in accordance with the referee methods. -14- 1.2 Exclusions This standard does not set limits for the composite risk current when several devices are performing different functions for the same patient and are independently connected to the utility power system. This standard does not apply to therapeutic currents. In addition, this standard does not apply to apparatus designed primarily for nonmedical applications and used in conjunction with electromedical apparatus, but located outside of the patient care vicinity. NOTE — As indicated above, all devices cannot be readily covered by this standard. Equipment such as video cassette recorders or computing devices are now being used in health care facilities. Such equipment is designed to other standards. If such equipment is not located in the patient care vicinity, such devices are not considered to pose a risk as they are not likely to contact the patient. 2 DEFINITIONS For the purposes of this standard, the following definitions apply. 2.1 accessory Device produced or recommended by the manufacturer of an electromedical apparatus, and intended to be electrically connected to that apparatus in order to make the apparatus useful or to improve its efficacy or versatility, and not a modular part of that apparatus. 2.2 auxiliary apparatus Electromedical apparatus used in conjunction with other electromedical apparatus to achieve a common purpose. NOTE — Auxiliary apparatus includes both interconnected apparatus and noninterconnected apparatus. 2.3 composite risk current Total risk current derived from the risk currents of all the apparatus associated with the patient that can flow through the patient, medical staff, or bystander. NOTE — This definition is included for reference only. A method of derivation and limits for composite risk current are not covered in this standard. 2.4 electromedical apparatus Instrument, equipment, system, or device that directly or indirectly uses electricity for any medical purpose. NOTE — Also included are all parts that are connected to such equipment and are required for the normal use of the equipment, including associated patient wiring or cables. 2.5 enclosure Exterior surface of the electromedical apparatus, including all accessible parts, knobs, grips, and shafts. 2.6 exposed electrically conductive surface External metal or otherwise electrically conductive surface that is connected to the internal circuits, mechanisms, or enclosure of an electromedical apparatus. 2.7 input part Part of the electromedical apparatus, other than a patientapplied part, that is intended to receive input signal voltages or currents from other equipment. 2.8 isolated patient connection Connection between the patient and the electromedical apparatus that is isolated from power ground (earth)1), the utility power system, and other supporting circuitry to such a degree that the risk current flowing through the connection does not exceed the limits given in Table 1—Summary of risk current requirements in rms microamperes (mA), provided in Section 4.2. 2.9 modular apparatus Electromedical apparatus that includes modules in its construction. 2.10 module Self-contained assembly that performs a function or class of functions in support of the major function of an electromedical apparatus. NOTE — Modules can generally be removed or replaced without affecting the operation of other assemblies in the apparatus. 2.11 nonoperational environmental conditions Temperature, humidity, altitude, or acceleration limits specified by the manufacturer for storage or shipment. 2.12 normal condition (NC) Condition in which all means provided against safety hazards are intact and the device is operating as desired. 2.13 output part: Part of the electromedical apparatus, other than a patientapplied part, that is intended to deliver output signal voltages or currents to other equipment. Electrical Safety Made Easy IV. 2.14 patient-applied part Entirety of any part of the equipment that comes intentionally into contact with the patient via a patient connection. 2.15 patient-applied risk current Current flowing from the electromedical apparatus through the patient to power ground (earth) or between patient-applied parts. 2.16 patient care vicinity Space, within a location intended for the examination or treatment of patients, extending 6 feet (ft) (1.8 meters [m]) beyond normal location of the bed, chair, table, treadmill or other device that supports the patient during examination and treatment. The patient care vicinity extends vertically to 7 ft, 6 inches (in) (2.3 m) above the floor. 2.17 patient connection Deliberate connection that can carry current between an electromedical apparatus and a patient. This can be a surface contact (e.g., an ECG electrode), an invasive connection (e.g., an implanted wire or catheter), or an incidental longterm connection (e.g., connective tubing). -15- Electrical Safety Made Easy IV. NOTE — As used in this standard, "patient connection" is not intended to include adventitious or casual contacts, such as push buttons, bed surfaces, lamps, and hand-held appliances. 2.18 patient isolation risk current Current flowing from the patient to power ground (earth) through a part applied to the patient due to the unintended introduction of a voltage from an external source on the patient. 2.19 risk current: Nontherapeutic current that can flow through the patient, medical staff, or bystander as a result of the use of electromedical apparatus. 2.20 single fault condition (SFC) Condition in which a single means of protection against a safety hazard in equipment is defective, a component failure could increase the risk current, or a single external abnormal condition exists. 2.21 sink current: Current that flows into a device or any part thereof, when an external voltage is applied to it. 2.22 source current Electrical current that flows from any part of an electromedical apparatus to any other part or to power ground (earth), when no external voltages are applied. 2.23 therapeutic current: Current that is intentionally applied to the patient for treatment of disease or disorder. 3 CLASSIFICATION OF ELECTROMEDICAL APPARATUS AND MEASUREMENT CONDITIONS -16- The following Sections define specific classes of electromedical apparatus and measurement conditions and detail how these classifications are applied in this standard. 3.1 Classification of electromedical apparatus For purposes of this standard, four categories of electromedical apparatus have been defined. For each category, risk currents are established. These four categories are listed below : a) Electromedical apparatus with isolated patient connection: Electromedical apparatus intended to be connected to the patient with the patient circuit isolated from power ground (earth), utility power systems, and other circuitry. b) Electromedical apparatus with nonisolated patient connection: Electromedical apparatus intended to be connected to the patient. c) Electromedical apparatus likely to contact the patient: Electromedical apparatus that does not have a patientapplied part, but that is intended for use in the patient care vicinity. NOTE — See Section 2 for definition of patient care vicinity. d) Electromedical apparatus with no patient contact: Electromedical apparatus that is intended for use outside the patient care vicinity and that has no patient connections. 3.2 Classification of measurement conditions The following Sections define normal and fault conditions. 3.2.1 Normal operating conditions Under normal operating conditions, a device is operating as designed with all means provided for protection against safety hazards intact, connected properly and securely to an approved power source and, if the device includes patient-applied parts, with such parts applied according to the manufacturer's instructions. The following are considered normal operating conditions: a) power switch on/power switch off; b) power polarity normal/power polarity reversed (cordconnected apparatus only); c) patient grounded; patient not grounded. 3.2.2 Single fault condition A single failure of a device's protection mechanism against a safety hazard or the failure of a single device component can introduce a hazard condition or lead to the existence of an external hazardous condition. The following are considered to be single fault conditions : a) power ground (earth) conductor open; b) short circuit of either barrier of double insulation; c) failure of a single component that can produce a hazardous current; d) (for equipment that is not intended to be grounded) the application of line voltage to an input or output part or to accessible conductive hardware of the enclosure; e) (for electromedical apparatus with isolated patient connections) the application of line voltage on a patientapplied part. 4 REQUIREMENTS 4.1 Labeling and documentation requirements 4.1.1 Isolated patient connections (labeling) Patient connections that meet the requirements of this standard for isolated patient connections shall be identified as being isolated at the connector of the apparatus. NOTE — Labeling of isolated patient connections shall comply with symbol number 3 of table DII, page 329, in the International Electrotechnical Commission (IEC) standard 601-1, second edition (IEC, 1988). 4.1.2 Information manuals The manufacturer shall supply the user with operating and maintenance instructions specifying how the electromedical apparatus should be operated and maintained to prevent the device's risk current from increasing beyond the limits set by this standard for its particular category (refer to Section 3.1). In addition, the manufacturer shall disclose the risk current category for which the apparatus is designed and shall identify the specific limits defined by this standard for that category. Electrical Safety Made Easy IV. 4.2 Risk current requirements (general) Electromedical apparatus shall meet the applicable risk current limits of this standard under normal conditions and under the single fault conditions specified in the test methods of Section 5. Table 1 (see next page) summarizes these requirements. 4.2.1 Apparatus interconnection Electromedical apparatus shall meet the risk current limits of this standard when manufacturerdesignated auxiliary apparatus, modular apparatus, or accessories are attached in the quantity and combinations stipulated by the manufacturer. The manufacturer shall supply the user (and shall label the apparatus) with limitations and with directions for the interconnection of modular apparatus, accessories, and auxiliary apparatus, and with directions for the use of convenience receptacles. 4.2.2 Cleaning and sterilization Electromedical apparatus shall meet the risk current limits of this standard after exposure to any disinfection or sterilization process specified by the manufacturer. -17- Electrical Safety Made Easy 4.2.3 Environmental conditions Electromedical apparatus shall meet the risk current limits of this standard after exposure to the nonoperational environmental conditions (e.g., storage, transportation, etc.) and under the worst-case environmental operating conditions specified by the manufacturer. 4.3 Enclosure risk current 4.3.1 General Enclosure risk current, when measured with the AAMI standard test load, is that current that flows between power ground (earth) and IV. a) exposed chassis conductive surfaces or hardware; or b) a 200 cm2 (centimeters squared) foil in contact with a nonconducting enclosure. NOTES : 1) The frequency-weighted network compensates for the allowable increase in risk current limits with increasing frequency. For measurement purposes with a voltmeter as shown, the limit remains constant at 1 mA/mV, independent of frequency. With the meter connected, the entire circuit is called the "risk current tester." 2) Refer to Section 5.7.2 for component requirements and tolerances. 4.3.2 Risk current limits Limits for enclosure risk current for all categories of electromedical apparatus, whether batterypowered, cord-connected, or permanently connected, and under both normal and single fault conditions, are shown in Table 1 on next page. 10,000 Ω 0.015 µF INPUT MILLIVOLTMETER 1,000 Ω 100 Ω Figure 1 AAMI standard test load . -18- Category Normal Condition PatientPatient CordApplied Risk Isolation General Other Connected/ PermaCurrent Risk Current CordCordPermaBatterynent (source (sink connected connected)1 nent2) current) current) 100 mA 100 mA 10 mA N/A 500 mA 2,500 mA 5,000 mA ISOLATED Single Fault Condition 300 mA 5,000 mA 50 mA 50 mA 1,000 mA 5,000 mA 10,000 mA Normal Condition 100 mA 100 mA 10 mA N/A 500 mA 2,500 mA 5,000 mA NON-ISOLATED Single Fault Condition 300 mA 5,000 mA 100 mA N/A 1,000 mA 5,000 mA 10,000 mA 100 mA N/A N/A 500 mA 2,500 mA 5,000 mA LIKELY TO CONTACT PATIENT Single Fault Condition 300 mA 5,000 mA N/A N/A 1,000 mA 5,000 mA 10,000 mA Normal Condition Normal Condition 100 mA 100 mA 100 mA N/A N/A 500 mA 2,500 mA 5,000 mA NO PATIENT CONTACT Single Fault Condition 500 mA 5,000 mA N/A N/A 1,000 mA 5,000 mA 10,000 mA Electrical Safety Made Easy IV. Table 1 Summary of risk current requirements in rms microamperes (dc to 1 kHz) 1) Equipment that has no protectively grounded (earthed) accessible parts and no means for protective grounding (earthing) of other medical equipment and which complies with the applicable requirements for enclosure leakage current and patient leakage current; also mobile x-ray equipment and mobile equipment with mineral insulation. 2) Equipment specified to be permanently installed with a protective power ground (earth) that is electrically connected and secured at a specific location so that the connection can only be loosened or moved with the aid of a tool. -19- 4.4 Patient-applied risk current (source current)2) Electrical Safety Made Easy IV. 4.4.1 General Patient-applied risk current, when measured with the AAMI standard test load, is that current that flows between any patient-applied part and : a) power ground (earth); b) exposed chassis conductive surfaces or hardware; c) a 200 cm2 foil in contact with a nonconducting enclosure; or d) any other patient-applied parts. Patient-applied risk current is also that current that flows between all patient connections tied together and (a), (b), (c), and (d) listed above when measured with the AAMI standard test load. NOTE — The current shall be measured at the patient end of the cable when connected to the device. 4.6 Earth risk current 4.6.1 General Earth risk current, when measured with the AAMI standard test load, is that current that flows in the protective earth conductor (ground conductor). NOTE — Not applicable to double insulated devices using a two-wire power cord. 4.4.2 Risk current limits Limits for patient-applied risk current for all categories of electromedical apparatus, whether battery-powered, cord-connected, or permanently connected, under both normal and single fault conditions, are shown in Table 1. 4.6.2 Risk current limits Limits for earth risk current for all categories of electromedical apparatus, whether cordconnected or permanently connected, and under both normal and single fault conditions, are shown in Table 1. NOTES : 4.7 Risk current limits versus frequency The risk current limits specified in Table 1 are for frequencies from dc to 1 kilohertz (kHz). Above 1 kHz, the limit is increased proportionally to a maximum value 100 times the limit at 1 kHz. Above 100 kHz, the limit is that which is determined for 100 kHz (see Figure 2). The use of the AAMI test load automatically compensates for frequency. 1) These limits are not applicable to electromedical apparatus without direct patient-applied connections. 2) The current shall be measured at the patient end of the cable when it is attached to the device. The cable is specified by the manufacturer. 4.5 Patient isolation risk current (sink current) 4.5.1 General Patient isolation risk current, when measured with the AAMI standard test load, is that current that would flow into a patient-applied part if the patient came into direct contact with a potential of 120 volts (V), 60 hertz (Hz) with respect to power ground (earth). -20- 4.5.2 Risk current limits Limits for patient isolation risk current for electromedical apparatus with isolated patientapplied part(s), whether cordconnected or permanently connected, are shown in Table 1. 5 Tests This Section contains referee tests and procedures by which compliance with the requirements of Section 4 and Table 1 can be determined. WARNING — These tests can expose personnel to hazardous electric shock and must be carried out with caution. Electrical Safety 100 mA NORMALIZED CURRENT Made Easy 1 mA 500 Hz 1000 Hz 10 KHZ 100 KHZ 1 MHZ 10 MHZ FREQUENCY (HZ) Figure 2 — Normalized current limits versus frequency 5.1 Compliance with the labeling requirements Compliance with the labeling and documentation requirements of Section 4.1 shall be verified by inspection. 5.2 Compliance with the risk current requirements (general test procedures) The risk currents of electromedical apparatus shall be measured by the methods described in this Section. 5.2.1 Test equipment and power system 5.2.1.1 Measuring instruments The risk current tester consists of the AAMI standard test load and a millivoltmeter as shown in Figure 1. The millivoltmeter shall measure true rms volts; however, it may be calibrated to true rms microamperes (mA) by employing a conversion factor of one microampere per millivolt (mV). The millivoltmeter shall have an input impedance of at least 1 megohm and have a bandwidth of dc to at least 1 megahertz (MHz) (–3 decibels). In the band from dc to 100 kilohertz (kHz), the indicated measurement shall not display an error of greater than 5 percent of reading, and shall resolve a signal as small as 1 mV. IV. Instruments that indicate true rms microamperes and have internal frequency compensation identical to that shown in Figure 1 meet the requirements of this Section if the measurement indicated does not display an error of greater than 5 percent of reading and resolves a signal as small as 1 mA in the band from dc to 100 kHz. 5.2.1.2 Power source 5.2.1.2.1 For line voltage powered equipment, the tests shall be performed on a grounded power system at the rated line voltage plus 10 percent. In the grounded system, the potential between the neutral and grounding conductors at the receptacle selected for the test shall not exceed 3 V. 5.2.1.2.2 The power ground (earth) terminal used in these tests shall be the grounding terminal of the specific receptacle powering the instrument under test. 5.2.1.2.3 Battery-powered apparatus shall be tested while powered by the type of battery recommended by the manufacturer and, if applicable, while connected to line power. -21- 5.2.2 Test conditions Electrical Safety Made Easy IV. 5.2.2.1 General First, the apparatus shall be disconnected from all other apparatus except auxiliary apparatus, modular apparatus, or accessories, as defined in the normative definitions (see Section 2). A single- or multi-function apparatus in a cabinet or in multiple cabinets with a single power cord connection is tested as a single apparatus. Each individual apparatus shall also be tested independently if described by the manufacturer as a stand-alone apparatus. Tests shall be conducted at the rated line voltage plus 10 percent. 5.2.2.2 Nonconducting enclosure The risk current shall be measured from an electrically conductive foil the size of the enclosure, but not to exceed 200 cm2, in immediate contact with the enclosure. The foil shall be placed at a location— determined by experimentation— such that the current measured to power ground (earth) is a maximum. If exposed chassis hardware is likely to be touched by personnel, then the hardware shall be treated as an exposed electrically conductive surface. Testing of nonconductive exposed surfaces of patient wiring and cables is not required. delivers therapeutic energy to the patient (e.g., a pacemaker), the therapeutic energy shall be zero during the test. Otherwise, the instrument shall be in the active or operable mode; i.e., with output switches closed, with electrodes properly connected to dummy loads, and with final circuit stages properly functioning but without a physiological drive signal. 5.2.2.4 Operation During the test, the apparatus shall run through a normal cycle and activate all accessories and/or auxiliary apparatus. 5.3 Enclosure risk current 5.3.1 Application The enclosure risk current tests shall apply to cord-connected, linepowered apparatus, to batterypowered apparatus with the charger connected, and to permanently connected apparatus. 5.3.2 Cord-connected, normally grounded apparatus 5.3.2.1 Using the test circuit of Figure 3, the enclosure risk current shall be measured: 5.2.2.3 Controls During risk current tests, all operator-accessible controls shall be adjusted to yield the largest risk current found by experiment. If the electromedical apparatus normally Reversing Switch S1 a) between enclosure and power ground (earth); b) between electrically conductive surfaces and power ground (earth); c) between a 200 cm2 foil in contact with the nonconducting enclosure and power ground (earth). 200 cm2 fail Rated Line Voltage +10% Apparatus Under Test Select Per 5.3.2.1 Ground M Ground Switch S2 -22- Ground Wire Exposed Conductive Surface Risk Current Tester Figure 3 — Enclosure risk current test circuit (normally grounded) NOTE — The 200 cm2 foil line leading from the "select" box to the apparatus under test refers to the connective mode with insulated apparatus. The line with an arrow leading from the "select" box to the apparatus under test refers to connections made with conductive enclosure. 5.3.2.2 Each measurement is performed when: a) the utility electricity supply polarity is normal and when the utility electricity supply polarity is reversed (by reversing S1). These are normal conditions; b) the apparatus power switch is on; the apparatus power switch is off. These are normal conditions; c) the ground switch (S2) is open; the ground switch is closed. The first is a single fault condition; the second is a normal condition; d) each barrier of double insulation is short circuited. These are single fault conditions. NOTE — The test methods for all measurement conditions are not supplied in this standard because they are device- and circuitspecific. 5.3.2.3 The power on/power off test also applies to apparatus with nonrechargeable batteries. Rated Line Voltage +10% Reversing Switch S1 5.3.3 Cord-connected, normally ungrounded apparatus 5.3.3.1 Using the test circuit of Figure 4 (see next page), the enclosure risk current shall be measured: a) between enclosure and power ground (earth); b) between electrically conductive surfaces and power ground (earth); c) between a 200 cm2 foil in contact with the nonconducting enclosure and power ground (earth). 5.3.3.2 Each measurement is performed when: Electrical Safety Made Easy IV. a) the utility electricity supply polarity is normal; the utility electricity supply polarity is reversed (by reversing S1). These are normal conditions; b) the apparatus power switch is on; the apparatus power switch is off. These are normal conditions; c) each barrier of double insulation is short-circuited. These are single fault conditions. NOTE — The test methods for all measurement conditions are not supplied in this standard because they are device- and circuitspecific. 200 cm2 fail Apparatus Under Test Select Per 5.3.2.1 Risk Current Tester Exposed Conductive Surface Ground Figure 4 — Enclosure risk current test circuit (normally ungrounded) -23- 5.3.3.3 The power on/power off test also applies to apparatus with nonrechargeable batteries. Electrical Safety Made Easy e) all patient connections shorted together and power ground (earth); f) all patient connections shorted together and any exposed, electrically conductive surface; g) all patient connections shorted together and a 200 cm2 foil in contact with the nonconducting enclosure; h) any patient connection and all other patient connections connected together. 5.3.4 Permanently connected apparatus Before line-powered apparatus is permanently installed, the enclosure risk current shall be measured according to the procedures described in 5.3.2. 5.4 Patient-applied risk current (source current) IV. 5.4.1 Application The patient-applied risk current tests of this Section shall apply to line-powered and battery-powered electromedical apparatus that has a patient connection(s). NOTE — The test methods for all measurement conditions are not supplied in this standard because they are device- and circuitspecific. 5.4.2.2 Each measurement is performed when: 5.4.2 Apparatus with isolated patient connection 5.4.2.1 Using the test circuit of Figure 5 (see next page), the patient-applied risk current (source current) shall be measured between: a) any patient connection and power ground (earth); b) any patient connection and any exposed, electrically conductive surface; c) any patient connection and a 200 cm2 foil in contact with the nonconducting enclosure; d) any patient connection and any other patient connection; a) the utility switch (S1) is normal/the utility switch is reversed. These are normal conditions; b) the apparatus power switch is on/the apparatus power switch is off. These are normal conditions; c) the ground switch (S2) is open/the ground switch is closed. The first is a single fault condition; the second is a normal condition; 120K Select Per 6.4.2.2.d Reversing Switch S1 Rated Line Voltage +10% Ground Switch S2 120V / 60Hz Patient Connection Apparatus Under Test 200 cm2 fail Select Per 5.3.2.1 Select Per 5.3.2.1 Exposed Conductive Surface -24- Figure 5 — Patient-applied risk current test circuit Risk Current Tester d) line voltage is applied to an input or output part or to accessible conductive hardware of the enclosure, if not grounded under normal conditions. These are single fault conditions; e) each barrier of double insulation is short circuited. These are single fault conditions. NOTE — The 120 K resistance is intended to protect the test operator. 5.4.3 Apparatus with nonisolated patient connection The patient-applied risk current shall be measured by the procedures described in 5.4.2. 5.5 Patient isolation risk current (sink current) The patient isolation risk current shall be measured in each individual patient connection that is labeled "isolated" when a potential of 120 V rms, 60 Hz, is applied through a series 120 kilohm resistance to the labeled patient connection, as shown in Figure 6. The patient isolation risk current is measured with respect to power ground (earth). For solely batterypowered apparatus, the patient isolation risk current is measured with respect to an electrically conductive surface on which the apparatus is positioned, and with an exposed conductive surface or other external electrical connection on the apparatus grounded. This test shall be performed with the apparatus both on and off and properly connected to its electrical supply. The patient cable shall be placed 20 cm away from a grounded surface. NOTE — The 120 K resistance is intended to protect the test operator. Electrical Safety Made Easy 5.6 Earth risk current 5.6.1 Application The earth risk current test shall apply to cord-connected apparatus, to battery-powered apparatus with the charger connected, and to permanently connected apparatus. IV. 5.6.2 Cord-connected, normally grounded apparatus 5.6.2.1 Using the test circuit of Figure 7 (see next page), the earth risk current shall be measured in the protective power ground (earth). 5.6.2.2 Measurement shall be performed when: a) the utility electrical supply is normal/when the utility electrical supply is reversed (by reversing S1). These are normal conditions; 1:1 Rated Line +10% Apparatus Under Test 120V 60Hz Select Per 5.5 120K Ground Risk Current Tester Figure 6 — Patient isolation risk current test circuit -25- b) the apparatus power switch is on/the apparatus power switch is off. These are normal conditions; c) each supply conductor is interrupted, one at a time (opening S2 and S3 in turn). This is a single fault condition; d) each barrier of double insulation is short-circuited. This is a single fault condition. Electrical Safety Made Easy IV. 5.6.3 Permanently connected apparatus Before the line-powered apparatus is permanently installed, the earth risk current shall be measured according to the procedures described in 5.6.2. 5.7 Risk current limits versus frequency 5.7.1 General When multiple risk currents of various frequency and phase relationships are present during a single test, the resultant risk current is related to the voltage across the AAMI standard test load. The risk current of an apparatus shall be the largest current measured during any of the required tests and conditions. The apparatus must meet all applicable limits of Table 1. 5.7.2 AAMI standard test load As shown in Figure 1, the test load shall be constructed using metalfilm resistors with a tolerance of 1 percent or better, and a mica- or plastic-dielectric (extended foil) capacitor with a tolerance of 5 percent or better. The AAMI standard test load has an impedance frequency characteristic (Figure 8) which is the approximate inverse of the curve of Figure 2, which shows risk current versus frequency. 5.7.3 Risk current calculation Using the AAMI standard test load of Figure 1 and a voltmeter calibrated to indicate rms millivolts, the weighted risk current is read directly from the meter, because: V(mV rms) Z(k ohms) I(mA rms) = Line Interruption Switches S2• Reversing Switch S1 Rated Line Voltage +10% S3• Apparatus Under Test Protective Earth (Ground Conductor) Risk Current Tester -26- Relative Magnitude (dB) Figure 7 — Earth risk current test circuit Figure 8 — Relative frequency characteristics of millivoltmeter reading in AAMI standard test load of Figure 1 0 -20 -40 10 101 102 103 104 105 A.1 General The rationale discusses the need for the standard and describes the basic underlying principles, empirical data, assumptions, and sources that support the requirements and test methods adopted in the standard. A.2 Need for the standard This standard seeks to reduce the risk of inadvertent electric shock from medical devices. In particular, it concerns itself with the risk of injury from the small currents that inevitably flow from or to electromedical apparatus. The intent of the AAMI Electrical Safety Committee was to develop a general baseline standard. The extent to which the standard should be applied is to be determined by individual institutions, standards groups, and other authorities. A.3 Classification of electromedical apparatus and measurement conditions Changes from the second edition of the standard (AAMI, 1985) have been made in keeping with changes to the requirements given in Section 4. A.4 Rationale for the specific provisions of the standard A.4.1 Labeling and documentation requirements A.4.1.1 Isolated patient connection (labeling) Fault conditions can contribute to patient risk due to source and sink currents. The greatest risk is with direct cardiac applications; such applications should utilize isolated patient connections. In order to better manage patients requiring direct cardiac connection, the user should be able to readily identify electromedical apparatus with isolated patient connections. Therefore, labels should appear on the electromedical apparatus itself. The standard does not require nonisolated patient connections to carry any special labeling. A.4.1.2 Information manuals By identifying the risk current classification and risk current limits, the manufacturer is informing the user of the device's intended purpose as that purpose relates to the risk of electric shock. Any special user actions required to ensure that the risk current limits are maintained throughout the life of the equipment should be described in the operating instructions or maintenance manuals. Electrical Safety Made Easy IV. A.4.2 Risk current requirements (general) The committee felt that grounding should not be the primary approach to limiting the risk of electric shock because it is possible to have a single fault failure in the grounding system. Redundant means of grounding are possible but are controlled by the user and not the manufacturer of the electromedical apparatus. If the grounding is lost or if other safety means fail, the risk current available from the enclosure should not represent a substantial hazard to the patient. The risk current limits were changed for certain categories as compared to the previous version of this standard (ANSI/AAMI ES1— 1985). These changes have been made in order to bring this document into closer harmonization with risk current limits specified in the International Electrotechnical Commission standard, Medical electrical equipment—Part 1: General requirements for safety (IEC 601-1, 1988). -27- Electrical Safety Made Easy IV. -28- In its review of the allowable risk current levels, the committee considered the following: a) Likelihood of stimulation of excitable tissue. The likelihood of stimulation of excitable tissue depends upon: 1) the location of the sites at which current enters and leaves the body; 2) the area of contact; 3) the amount of current flowing; 4) the susceptibility to mechanical stimulation; 5) the presence of a fault condition; 6) the probability of the current having a given value. Medical devices have been classified into the risk categories described in Section 3 because of the different magnitudes of risk associated with these categories. b) Survey of published data. During the several years since the publication of the previous version of this standard (ANSI/AAMI ES1—1985), experience has been gained with respect to the incidence of problems related to risk current and the probability of occurrence of the potential hazard. The available published data on currents causing ventricular fibrillation in humans have also been reexamined. The following have been noted: 1) The combination of an open power ground (earth) wire and a person touching both a conductive part of the enclosure and the patient is a low probability event. 2) The combination of an open power ground (earth) wire and a person touching both a conductive part of the chassis and the distal end of an invasive cardiac connection is a low probability event. 3) In one study, Raftery (1975) found that the smallest current that produced a disturbance in rhythm in humans was 80 mA. In a second study, Watson (1973) found that the smallest current that produced ventricular fibrillation in humans was 15 mA. 4) Mechanically induced ventricular fibrillation has been observed during cardiac catheterization at zero current. 5) The human data obtained by Starmer (1973) and Watson (1973) follow, reasonably well, a normal distribution for currents to 300 mA. All patients are not equally susceptible to current-induced ventricular fibrillation. According to Figure A.1, there is approximately a 1 percent probability of fibrillation at 30 mA. 6) Current perception is a function of contact location, contact pressure, skin condition, moisture, and contact area. Experiments report a wide range of current perception. Dalziel (1968) reports that only approximately 1 percent of the population can perceive 500 mA passing from the fingers of one hand to the fingers of the other hand. Tan and Johnson (1990) report that 300 mA produces a strong sensation for electrodes placed 10 cm apart on the upper arm. Levin (1991) reports that nearly all the population will perceive 500 mA without any reaction for current passing from the finger on one hand to the underside of the wrist on the other hand. Levin further reports that, on the underside of the wrist, the stratum corneum (layer of dead material on the skin surface) is not as thick as that on the forefinger and, therefore, the sensitivity to current perception might be higher. Startle current is that level of perception current that, when first perceived, might result in a nurse or other clinician's involuntary reaction to the sudden sensation of perception current. This uncontrolled reaction is of great concern. 7) Worldwide, since the advent of risk current standards, concern about grounding, and use of better practices in handling catheters and invasive cardiac connections, there have been no reports of incidents involving risk current passing through the patient. NOTE — A minority of the AAMI Electrical Safety Committee were opposed to increasing risk current limits unless scientific studies supported higher limits. Also, in order to harmonize with IEC 601-1 and to allow for additional fault conditions as compared to the current standard, the test measurement classifications of "normal condition" (NC) and "single fault condition" (SFC) were introduced. Measuring risk current under SFC is important because: — components can fail; — single faults exist that are not now considered in the standard; — faults can occur in accessory equipment. A.4.2.1 Apparatus interconnection The total risk current associated with a device can be a function of the modules, accessories, and interconnections used with the device. Voltage differences can occur between different parts of a device, particularly if current flows in the grounding circuit. Thus, a remote accessory powered from the device or from a separate source is, for purposes of the standard, considered part of the device. Auxiliary power outlets may be provided for powering additional devices. The labeling requirements provide some assurance that the user has appropriate guidance about the limitations applying to equipment or accessories connected to an auxiliary power receptacle. A.4.2.2 Cleaning and sterilization The long-term effects of repeated disinfection or sterilization of a device on risk currents must be considered because of possible degradation of insulating materials. A.4.2.3 Environmental conditions Temperature, humidity, atmospheric pressure, mechanical shock, and similar environmental constraints can have an effect upon the risk currents. To protect the patient, the risk current limits must also be met in the intended environment. Electrical Safety Made Easy IV. A.4.3 Enclosure risk current The 100 mA NC values were selected for cord-connected and permanently connected apparatus on theoretical grounds. For a typical ground resistance of 0.2 ohms, 100 mA of enclosure risk current, measured as in Figure 3, requires that 500 mA flow in the power ground (earth) wire. This would be a major fault condition. The 300 mA SFC values for isolated, nonisolated, or likely-to-contactpatient, cord-connected apparatus were selected on the basis of reaction current measurements by Levin and the low probability of risk current reaching the distal end of an invasive heart connection via another person. Levin reported that currents of 300 mA will not produce sensations leading to a "startle" reaction (Levin, 1991). The 500 mA limit for SFC for cordconnected, no-patient-contact equipment was selected because there is no concern about this current reaching the patient. The 5,000 mA limit for SFC for permanently connected equipment is basically a power ground (earth) -29- Electrical Safety Made Easy IV. wire current, because the apparatus is, by definition, permanently grounded. This is the current allowed from the enclosure of permanently connected equipment if the power ground (earth) wire were opened. The permanently connected power ground (earth) wire is not expected to open. that produced ventricular fibrillation in humans are considered, the probability is approximately zero. The 50 mA SFC limit for isolated equipment was selected because the probability of causing ventricular fibrillation is low if the data of Figure A.1 are extrapolated and is approximately zero if only observed values of currents producing ventricular fibrillation in humans are considered. A.4.4 Patient-applied risk current (source current) The 10 mA NC limit was selected as the current that may flow directly into the heart continuously. Figure A.1 shows that the probability of inducing ventricular fibrillation is very small if the data on which the chart is based are extrapolated. If only observed values of currents The 100 mA SFC limit was selected for nonisolated equipment, because if this current enters and exits the surface of the body, then only a fraction will reach the heart. Starmer 60 Hz D = 1.25mm Watson 60 Hz D = 2.0mm 99.9 99 Cumulative Percent 95 + + 90 50 + + + 20 + + + + + 5 1 0.1 0 50 100 150 200 250 300 350 400 450 500 Microamperes Figure A.1 — Normal probability plot -30- A.4.5 Patient isolation risk current (sink current) The 50 mA SFC value for isolated equipment was allowed because of the low probability of line voltage appearing on a patient, and because of the low probability of 50 mA inducing ventricular fibrillation. For line voltage to appear on a patient, a power ground (earth) wire must be open and there must also be a fault in basic insulation. If the data in Figure A.1 are extrapolated, then 50 mA has a low probability of inducing ventricular fibrillation. If only recorded human data are considered, the probability of inducing fibrillation at 50 mA is approximately zero. A.4.6 Earth risk (ground risk) current The original standard of 1978 and the revised edition of 1985 did not include earth current as a potential risk current. This was not considered to be an issue as most medical devices of the era utilized conductive enclosures requiring grounding. Thus, earth current was effectively measured as enclosure (chassis) current under the open ground condition. In the last 25 years, however, the change to nonconductive enclosures negated this equality. The enclosure current is now measured as the capacitivecoupled current to a 200 cm2 foil in contact with the enclosure. This current bears little resemblance to the earth risk current due to the current-limiting characteristics of the capacitive coupling of the nonconductive enclosure. The earth risk current does not pose a direct risk to the patient or medical personnel. However, excessive earth risk current, either by design or internal breakdown, will raise the potential of the device's ground with respect to true power ground (earth) as represented by structural elements, modular wall units, cold water pipes, and other installed piping, or by an adjacent receptacle power ground (earth). Contact with such elements and a second device under test will result in current flow. Thus, the committee felt that leaving the earth risk current unmeasured and unlimited constituted a potential hazard that should be avoided. The allowable values for earth risk current detailed in this standard are not thought to pose a direct hazard as the current is safely returned to earth. The values selected were chosen to avoid any significant increase in the current flowing through the protective grounding system of the installation and to be consistent with the limits of IEC 601-1 for power ground (earth) and nonconducting enclosures. Further, the 2.5 mA limit in normal mode is within the limit for isolation monitors set by the National Fire Protection Association (NFPA, 1993, Section 3.4.3.3), which specifies that isolation monitors should not alarm at 3.75 mA. Electrical Safety Made Easy IV. A.4.7 Risk current limits versus frequency Figure 2 of the standard was derived from strength/frequency data for perceptible and lethal currents (Geddes and Baker, 1971). The flat portion between 100 kHz and 1 MHz does not reflect physiological data obtained with purely sinusoidal currents. Stimulation has been observed with complex waveforms at high frequencies, but little data are available. In the absence of data, it was deemed prudent not to extrapolate beyond 100 kHz. A.5 Tests The test procedures documented in Section 5 of the standard provide referee test methods for verifying compliance with the requirements of Section 4. These referee tests are not necessarily intended for purposes of manufacturing or quality control (although these applications are not precluded), as equivalent measurements may be obtainable by other means. -31- ASSOCIATION FOR THE ADVANCEMENT OF MEDICAL INSTRUMENTATION. Safe current limits for electromedical apparatus. ANSI/AAMI ES1—1985. Arlington (Vir.): AAMI, 1985. ISBN 0-910275-50-5. Electrical Safety Made Easy DALZIEL, CF. Reevaluation of lethal electric currents. IEEE Trans Indus Gen Appl, GA-4, 1968, vol. 1, no. 5, p. 467-476. GEDDES, LA. and BAKER, LE. Response to the passage of electric currents through the body. J Assn Adv Med Instrum, 1971, vol. 5, p. 13-18. INTERNATIONAL ELECTROTECHNICAL COMMISSION. Medical electrical equipment—Part 1: General requirements for safety, 2nd ed. IEC 601-1. Geneva: IEC, 1988. LEVIN, M. Perception of chassis leakage current. Biomed Instrumentation and Technology, 1991, vol. 25, no. 2, p.135-140. IV. NATIONAL FIRE PROTECTION ASSOCIATION. Standard for health care facilities. NFPA 99-1993. Quincy (Mass.): NFPA, 1993. RAFTERY, EB., GREEN, HL., and YACOUB, MH. Disturbances of heart rhythm produced by 50 Hz leakage currents in human subjects. Cardiovascular Research, March 1975, vol. 9, no. 2, p. 263-265. STARMER, CF. and WHALEN, RE. Current density and electrically induced ventricular fibrillation. J Assn Adv Med Instrum, 1973, vol. 7, no. 1, p. 3-6. TAN, KS. and JOHNSON, DL. Threshold of sensation for 60 Hz leakage current: Results of a survey. Biomed Instrumentation and Technology, 1990, vol. 24, no. 3, p. 207-211. Bibliography WATSON, AB., WRIGHT, JS., and LAUGHMAN, J. Electrical thresholds for ventricular fibrillation in man. Med J Australia, 1973, vol. 1, p.1179-1182. You can contact BAPCO: BAPCO 3200 Sencore Drive Sioux Falls, SD 57107 1-800-419-4000 http://www.bapcoinfo.com email: bapco@bapcoinfo.com -32- Glossary Of Terms AMPACITY: Current-carrying capacity of electrical conductors expressed in amperes. ANESTHETIZING LOCATION: Any area of the facility that has been designated for the administration of any flammable or nonflammable inhalation anesthetic agents in the course of examination or treatment, including the use of such agents for relative analgesia. CONDUCTIVE: Materials, such as metals, that are commonly considered electrically conductive, and materials that, when tested, have a resistance not exceeding 1,000,000 ohms. Such materials are required where electrostatic interconnection is necessary. CRITICAL EQUIPMENT: Equipment that is essential to the safety of the occupants in the facility. CRITICAL SYSTEM: A system of feeders and branch circuits in nursing homes and custodial care facilities arranged for connection to the alternate power source to restore service to critical receptacles, task illumination, and equipment. DIRECT ELECTRICAL PATHWAY TO THE HEART: An external conductive pathway, insulated except at its ends, one end of which is in direct contact with the heart muscle and the other outside the body, that is accessible for inadvertent or intentional contact with grounded objects or energized, ground-referenced sources. Catheters filled with conductive fluids and electrodes, such as may be used for pacing of the heart, are examples of direct electrical pathways to the heart. DOUBLE-INSULATED APPLIANCES: Appliances having an insulation system comprising both basic insulation necessary for the functioning of the appliance and for basic protection against electrical shock and supplementary insulation. The supplementary insulation is independent insulation provided in addition to the basic insulation to ensure protection against electric shock in case of failure to the basic insulation. ELECTRODE: A device intended to probe an electrically conductive connection through a cable to a patient. There are several types: Active Electrode: An electrode intended to generate a surgical effect at its point of application to the patient. Dispersive Electrode: An electrode intended to complete the electrical path between patient and appliance, and at which no surgical effect is intended. It is often called the "indifferent electrode", the "return electrode", the "patient plate", or the "neutral electrode". EXPOSED CONDUCTIVE SURFACES: Those surfaces that are capable of carrying electric current and that are unprotected, uninsulated, unenclosed, or unguarded, permitting personal contact. FAULT CURRENT: A current in an accidental connection between an energized and a grounded or other conductive element resulting from a failure of insulation, spacing, or containment of conductors. FREQUENCE: The number of oscillations, per unit of time, of a particular current or voltage waveform. The unit of frequency is the Hertz (Hz). (The unit of frequency used to be "cycles per second", a term no longer preferred). Note : The waveform may consist of components having many different frequencies , in which case it is called a complex or nonsinusoidal waveform. GROUND-FAULT CIRCUIT INTERRUPTER: A device whose function is to interrupt the electric circuit to the load when a fault current to ground exceeds some predetermined value that is less than that required to operate the overcurrent protective device of the supply circuit. GROUNDING SYSTEM: a system of conductors that provides a low-impedance return path for leakage and fault currents. It coordinates with, but may be locally more extensive than, the grounding system described in Article 250 of NFPA 70, National Electric Code. HAZARD CURRENT: for a given set of connections in an isolated power system, the total current that would flow through a low-impedance if it were connected between either isolated conductor and ground. The various hazard currents are: Fault Hazard Current: The hazard current of a given isolated power system with all devices connected except the line isolation monitor. Monitor Hazard Current: The hazard current of the line isolation monitor alone. Total Hazard Current: The hazard current of a given isolated system with all devices, including the line isolation monitor, connected. IMPEDANCE: Impedance is the ratio of the voltage drop across a circuit element to the current flowing through the same circuit element. The circuit element may consist of any combination of resistance, capacitance, or inductance. The unit of impedance is the Ohm. INTRINSICALLY SAFE: As applied to equipment and wiring, equipment and wiring that are incapable of releasing sufficient electrical energy under normal or abnormal conditions to cause ignition of a specific hazardous atmospheric mixture. Abnormal conditions may include accidental damage to any part of the equipment or wiring, insulation or other failure of electrical components, application of overvoltage, adjustment and maintenance operations, and other similar conditions. ISOLATED PATIENT LEAD: A patient lead whose impedance to ground or to a power line is sufficiently high that connecting the lead to ground or to either conductor of the power line results in a current flow below a hazardous limit in the lead. ISOLATED POWER SYSTEM: A system comprising an isolating transformer or its equivalent, a line isolation monitor, and ungrounded circuit conductors. ISOLATION TRANSFORMER: A transformer of the multiple-winding type, with the primary and secondary windings physically separated, that inductively couples its ungrounded secondary winding to the grounded feeder system that energizes its primary winding. LEAKAGE CURRENT: Any current, including capacitively coupled current, not intended to be applied to a patient, but may be conveyed from exposed metal parts of an appliance to ground or to other accessible parts of an appliance. LINE ISOLATION MONITOR: An instrument that continually checks the hazard current from an isolated surface to ground. MACROSHOCK: The effect of large electric current (milliamperes or larger) on the body. MICROSHOCK: The effect of small electric currents (as low as 10 microamperes) on the body. To be hazardous, such currents must be applied to a conductor inside or very near the heart. mV: Millivolt. mA: Milliampere. PATIENT CARE AREA: Any portion of a health care facility where patients are examined or treated. Note : Business offices, corridors, lounges, day rooms, dining rooms, or similar areas are not classified as patient care areas. PATIENT-CARE-RELATED ELECTRICAL APPLIANCE: An electrical appliance that is intended to be used for diagnostic, therapeutic, or monitoring purposes in a patient care area. PATIENT EQUIPMENT GROUNDING POINT: A jack or terminal that serves as a collection point for redundant grounding of electrical appliances serving a patient vicinity or for grounding other items in order to eliminate electromagnetic problems. PATIENT LEAD: Any deliberate electrical connection that may carry between an appliance and a patient. This may be a surface contact (such as an ECG electrode), an invasive connection (such as an implanted wire or catheter), or an incidental long-term connection (such as conductive tubing). Adventitious or casual contacts such as a push button, bed surface, lamp, hand-held appliance, and so fourth, are not considered patient leads. PATIENT VICINITY: In an area in which patients are normally cared for, the patient vicinity is the space with surfaces likely to be touched by the patient or an attendant who can touch the patient. Typically in a patient room, this is a space within the room 6 ft. (1.8m) beyond the perimeter of the bed in its normal location and extending vertically within 7 ft. 6 in. (2.3m) of the floor. REACTANCE: The component of impedance contributed by inductance or capacitance. The unit of reactance is the Ohm. REFERENCE GROUNDING POINT: A terminal bus that is the equipment grounding bus, or an extension of the equipment grounding bus, and is a convenient collection point for installed grounding wires or other bonding wires where used. WET LOCATIONS: Those patient care areas that are normally subject to wet conditions, including standing water on the floor, or routine dousing or drenching of the work area. Routine housekeeping procedures and incidental spillage of liquids do not define a wet location. 1-800-419-4000 • www.bapcoinfo.com