IEC 6060 1- 1-2,2001:New EMC requirements for Medical Equipment Sanjay Baisakhiya, R.Ganeasn, Sisir K.Das SAMEER - Centre for Electromagnetics Znd Cross, ClT Campus, Taramani, Chennai - 600 113, India Phone: 044- 22541817,22541424 Fax: 044 - 22541424 e-mail: sameercem@vsnl.com ABSTRACT-International Eleclrotechnical Commission (IEC) brought out comprehensive standard IEC 60601-12 , 1 9 9 3 ~ Iediiion) to d r e s s EMC problems in Medical Equipmenl. Technological developments in medical equipment and changes in Electromagnetic environment rendered jirst edition inadequate in d r e s s i n g all issues concerning to Electromagnetic CompatibiliryfEMC).lEC 60601-12,2Wl(second edition) was brought out to oddress several of such issues, which were not the part of first edilian. This paper intends to analysis two issues Firstly major changes from first edition and reasons for these changes. This will helps in befferunderstanding of requirements of second edilion. Secondly redesigning products f m EMC perspective to achieve complinnce to new standard I. INTRODUCTION In modem days, medical electricaVelectronics equipments are widely used by medical professionals to diagnosis & monitor the patients and administer treatment to them. Electrocardiogram (ECG), Ultrasound scanner, pulse oximeter, infant incubator, X-ray machines are some of the example used for these purposes. Any malfunction or erroneous performance of these equipment may lead to wrong diagnosis and could even endanger the life of patient. Due to criticality of their operation, several standards making organization have brought out guidelines and regulations to achieve EMC in health care area. Federal Communication Commission (FCC), Center for Devices and Radiological Health (CDRH) of Food and Drug Administration (FDA), International Electrotechnical Commission (IEC), Committee European Electrotechnique (CENELEC) are few important organizations, who promogulated standards / regulations in their respective countries. IEC has mandate of many countries to prepare and publish intemational standards in electrical and electronics field. First edition of IEC 60601-1-2 was Proceedings of INCEMIC published in April 1993.0flate. it was realized that though equipment tested for IEC 60601-1-2.1993 standard, still sometimes it may exhibit malfunction during their use. Authors came across one such case, where ultrasound scanner, which is used to scan soft tissues was malfunctioning, inspite of being tested for the old standard. It was found that conducted emission from PC, which was used to register patient name, was interfering with ultrasound scanner. Onset of wireless communication revolution has added another dimension to electromagnetic environment faced by medical equipment. In this paper issues concerning the changes in the standard and redesigning of products &e address. 11. IEC 60601 STANDARD The Intemational Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising all national electrotechnical committee. IEC publishes the international standard. The work of preparation of standard is entrusted to technical committee. Standards prepared by this committee are in the form of recommendation for the intemational use. . . IEC 60601-series is product standard for medical electrical equipment and its scope is the safety and essential performance of them. This standard is grouped in two parts, those numbered “-1-X are considered relevant to general requirement for safety and are labeled collateral standard. Those numbered “2-X’ contain requirement for specific kind of medical equipment. IEC 60601-1-2 is “Medical electronics equipment-part 1-2:general requirement . for safety,collateral standard-EMC requirement and test. The IEC 60601-2-X standard known as particular or part. two standard are product standard> e.g. IEC. 60601-2-25, particular ‘requirement for safety of - 2003 Authorized licensed use limited to: UNIVERSIDADE FEDERAL DE MINAS GERAIS. Downloaded on June 18, 2009 at 16:20 from IEEE Xplore. Restrictions apply. 409 electrocardiograph. They supplement and in some cases substitute for requirement of product family standard. and IEC 60601-1-4 standards. For immunity failure criteria is defined very elaborately. Failure criterions are given in clause 36.202.13 of the standard. Ill. SALIENTFEATURES OF IEC 60601-1-2,1993, FIRST Second edition envisages that achieving EMC in medical equipmenthystem is shared responsibility. Manufacturer is required to test equipmendsystem as given in the standard, also it is expected from manufacturer to inform end user for the type of environment equipmenthystem is designed. Manufacturers is also expected to guide end user to suggest the ways to achieve EM environment for which equipment is designed. It is responsibility of end user to follow guidelines provided by manufacturer to achieve intended EM environment. EDITION This standard is based on the standard prepared by IEC technical committees SC62 A, TC 65 (Industrial process measurement and control), TC 77(Electromagnetic compatibility including network) and CISPR (International special committee on radio interference). This standard specifies general requirement and test for Electromagnetic Compatibility of medical equipment and/or system. Radiated emission test, conducted emission test, ESD, Radiated immunity test, Electrical Fast Transient and Surge immunity test are included. IV.NEED FOR CHANGES IN FIRST EDITION Biggest change that has taken place since publication of the first edition is the wide spread use of wireless communication in health care Facilities. Cellular phones are omnipresent and hospitals are no exception. Wireless technology is also being used to transmit and store data related to patient. Use of wireless technology has brought potential source of interference i.e. transmitter close to the sensitive medical electricallelectronics equipment. It became necessary for these equipment to withstand RF field from wireless transmitter. Apart from these changes, many standard, which were in draft form have been published and are available now. So these standards are included in the second edition. Also, as a matter of policy IEC reviews its standard at certain interval or before. This exercise is required to take note of all changes, which have taken place in course of time. After review IEC issues revised standard. V. IEC 60601-1-2,2001 SECOND EDITION IEC 60601-1-2,2001 i.e. second edition is completely revised standard. Many new requirements have heen included. For emission, equipments are to be properly categorize depending on their intended use. Second edition requires manufacturer to establish minimum performance of the equipment, which shall be ensured during immunity test. This minimum performance is to be established by doing risk analysis. Risk analysis may be done using I S 0 14971 410 Higher immunity levels have been prescribed in second edition for life supporting equipment to achieve better safety margins. However lower immunity levels are acceptable like first edition for certain equipmenthystems provided proper justifications are given in accompanying document. VI. COMPARISON O F EMISSION REQUIREMENTS Table-1 gives comparison of important requirement in the first and second edition of IEC 60601-1-2 TABLE - I Emission Requirements - IEC 60601-1-2:1993 & 2001 Emission requirements Referred Standard for high frequency tmissions (0.15-loo0 IEC 60601-1- 2. 1993 1 IEC 60601-1-2. 2001 CISPR 14 CISPR 14 CISPR 15 CISPR 22 Referred Standard for low frequency emission No requirement Relaxation limits Relaxation in field strength limits (RE),if equipment I system installed in shielded locatio", relaxation has to be justified in accompanying document IEC 6 l W 3 - 2 for harmonics IEC 6 1 W 3 - 3 for voltage fluctuation and flicker Relaxation in RE limit, if equipment I systems installed in shielded location by an amount upto applicable specified value of minimum shielding effectiveness MHI) in Relaxation in conducted emission limit by an amount upto applicable specified value of minimum R€ filter altC""2.tiO" Proceedings of INCEMIC Authorized licensed use limited to: UNIVERSIDADE FEDERAL DE MINAS GERAIS. Downloaded on June 18, 2009 at 16:20 from IEEE Xplore. Restrictions apply. - 2003 First edition requires equipment or system to be tested as per CISPRII, and for medical radiological equipments/systems are to be tested for click as per the provision of CISPR 14. Second edition requires classification of medical equipmentkystem in appropriate category and it gives guidelines by way of flowchart and tables. These flowchart and guideline tables helps in identifying appropriate applicable standard to them. In second edition ,Equipment or systems are to be tested as per CISPR14,if they do not contain any electronic circuit. Dental drills used by dentist is one typical example to be tested as per CISPR14.Lighting equipment used in medical arena are to be tested as per CISPR 15.Lightning equipment used for illuminating X-ray films and in operation theatre are typical example of medical equipment to qualify for CISPR 15.Information Technology equipment used with medical equipment are to be tested for CISPR 22 standard. All other medical equipmenthystems are to be tested as per CISPR 11.They are to be classified in group1 or group2 and class A or class B depending on their intended use. Second edition requires all electricaUelectronics equipment which are to be connected to low voltage public network and draws current less than 16A to be tested for harmonics emission and voltage variation and flicker test. VII. COMPARISON OF IMMUNITY REQUIREMENTS Comparison of first and second edition immunity requirements are given in Table-2.Conducted disturbance test, voltage dips and interruption, power frequency magnetic field are the new test added in the second edition, which were not the part of fxst edition. First edition referred IEC 801 series for immunity test, many of them were in draft form at the time of publication of its publication. According to compliance criteria given in the fxst edition equipment must continue to perform as per specifications of the equipment given by manufacturer and if it fails, it must remain safe, whereas the compliance criteria is more elaborate in the second edition. This requires that all those specifications / functions, which are critical in patients's diagnosis / Proceedings of INCEMIC treatment, are to be identified after risk analysis. These specifications /function must not change or deteriorate as given in clause. Risk analysis is not required, if all specificatiodfunctions are monitored for above degradation during the test. . VIII ACCOMPANYING DOCUMENTS IEC 60601-1-2 has stated objective to achieve electromagnetic compatibility by coordinated effort among manufacturer ,customer and users. Requirement of accompanying documents is a way to implement this objective. In first edition, clause 6.8.201 gives the information required to be supplied to customer/users, if applicable, by mentioning them in accompanying document. For equipments using RF energy for its operation, it is required to give guidelines for identifying , avoiding and resolving EM1 caused by this RF energy. All those medical equipments, which are sensitive for the purpose of their operation ,it is required to give appropriate guidelines to ensure their proper operation. Accompanying document must include justification, if different immunity level other than specified in the standard are used. Accompanying documents requirements in second edition are very elaborate compared to first edition. More areas are identified ,which affect EMC of medical equipment during their use. Information relevant to these area is required to be mentioned in accompanying document. Some of the important requirement of accompanying documents in second edition are as follows. (a) For all medical equipments, it should cany statement that portable and mobile RF equipment can affect medical equipment. (b) If any connector of medical equipment is exempted from ESD testing, then ESD warning symbol is to be put on medical equipment. In case these ESD sensitive connectors are to be touched, necessary ESD precautions are to be given in accompanying documents. All users of such medical equipment must be informed of this ESD warning symbol and precautions to be observed in handling such equipments. - 2003 Authorized licensed use limited to: UNIVERSIDADE FEDERAL DE MINAS GERAIS. Downloaded on June 18, 2009 at 16:20 from IEEE Xplore. Restrictions apply. 411 Details of accessories used during the testing and warning that if accessories other than mentioned in this report are used. It may affect emission and immunity levels of the equipment. If any lower or higher immunity levels than specified in the standard are used,proper justification shall be given. Guidelines have to be given to achieve electromagnetic environment, for which medical equipment is designed. Radiated RF Electromagnetic Field: In second edition, the upper frequency. range for this test is extended from lGHz to 2.5 GHz.This may require following, changes, if medical equipmend system has metallic enclosure. (a) Size of openings in the metallic'enclosure may be required to reduce to offer shielding at higher frequency. (b) Gasket may be required to prevent field leaking Guidelines on recommended distance between portable and mobile RF communication equipment and medical equipment VIII. ADDITIONAL DESIGN REQUIREMENTS: through the seam at high frequency. For life support equipment field strength requirement is changed from 3V/m to 10 V/m, this may require additional shielding to achieve compliance to new requirement. Emission: Conductdd RF Immunity Test: Harmonics emission is a new requirement in second edition. Though this requirement is normally met by low power medical equipmendsystem but for high power medical equipment /system this may he little tough to qualify. This requirement may necessitate use of filter and change in power supply circuit for compliance. Immunity. Electrostatic Discharge: In second edition the contact discharge level is increased to +I- 6KV from +/- 3KV specified in previous edition. Increased contact discharge may require thicker insulation on push buttons, touch screen to prevent contact discharge. Increased contact discharge level will also result in higher level of radiation during indirect discharge. This may require steps to reduce the radiated coupling to equipment. For grounded enclosure to maintain same level of ESD protection from secondary arc discharge due to increased contact level may require increase in creepage distance. It may also require to increase the distance of electronics component kept close to opening of the enclosure to prevent exposure of them to intense electric field created when current flows during contact discharge 412 This is new requirement in second edition. In this test coupling is predominantly through conducting cables connected to EUT i.e. main supply, signal line or earth connection. Mostly analog circuits are the victims of this type of interference and those operating at lower signal level are more prone. Though medical equipment operates at low frequency but they are susceptible to high level of W energy due to failure mode audio rectification. ECG and EEG patient connecting cables are to be treated specially as these cables are connected to analog circuit for further processing and very poor decoupling of signal lines is done to keep low leakage current. Following techniques may be used with external cables connected to medical equipment to harden them against this interference. (a) Use of shielded cable and properly terminating them. (b) Filter the cable, whenever possible. (cj Use of common mode choke on external cable Surge: In second dition,lower voltage levels are also defined, while in first edition only higher levels were Proceedings of INCEMIC Authorized licensed use limited to: UNIVERSIDADE FEDERAL DE MINAS GERAIS. Downloaded on June 18, 2009 at 16:20 from IEEE Xplore. Restrictions apply. - 2003 defined. Hence suppression should be taken care at lower level also, if equipment has to comply to second edition.. Voltage Dips and interruption: This is a new requirement in second edition. Following changes may be useful in complying to this requirement. (a) Medical equipmenthystems using bridge rectifier in their power supply may require bigger value capacitor on its DC bus. (b) Medical equipmentlsystem using AC main may require backup supply l i e U P S or batteries to tide over these dips and interruption on power lines. Power Frequency Magnetic Field This is new requirement and it is relevant for medical equipment using CRT display or Hall effect sensors. Proper magnetic shielding of them may be useful in achieving compliance to this requirement. X. CONCLUSION . Second edition of IEC 60601-1-2 is a completely revised standard. Which is substantially different from first edition on many aspect. This paper illustrate the major difference between the two .Efforts is made to highlight the area in equipment design, which will need attention to ensure smooth transition to second edition. REFERENCES: (i) IEC 60601-1-2,1993:Medical electrical equipment, Collateral Standard Electromagnetic Compatibility requirement and test (ii)IEC 60601-1-2,2001:Second Edition, Medical electrical equipment, collateral standard, Electromagnetic Compatibility: Requirement and test - ,-roceedings of INCEMIC - 2003 Authorized licensed use limited to: UNIVERSIDADE FEDERAL DE MINAS GERAIS. Downloaded on June 18, 2009 at 16:20 from IEEE Xplore. Restrictions apply. 41 3 Table -2 Immunity Requirements - IEC 60601-1-2:1993 Br 2001 Immunity requirements Elechostatic Discharge Radiated radio frequency elechomagnetic field test IEC 60601-1-2, 2001 IEC 60601-1- 2,1993 EC 801-2 :antact:+/- 3KV k + I - 8KV EC 801-3 ?eq:26-1 000MHz %Id strength:lVlm (for shielded ocations), 3VIm for other locations EC 610004-2 :ontact:+/-2KV,+/-4KV,+/-SKV 4ir:+I?;KV,+l4KV.+I-6KV E C 61000-4-3 'req:80-2500 MHz 'ield strength:3 V/m(for non-life supporting :quipmentlsystem).lOV/m(for life supporting equipment) ?or shielded locations the levels are to be reduced prapodonal to ihielding effectiveness of the Locations Wodulation frequency . . :2 Hz(lntended use is to control or meaure I physiological parameter and operating frequency is 4 Hz or 5 3HZ Electrical Fast Transient EC 8014 IKV-Plug connected equipment !KV-Permanently installed :quipmenVsystem 1.5 kV for interconnecting lines lanrer than E C 610004-4 +I-2 KV AC/DC lines +I-IKV Signal and interconnecting cable for length>3m )m Surge Conducted Disturbance, Induced by RF field EC 801.5 IKV-differenud made 1KV-Common made br power line only E C 61000-4-5 Vo requirement IEC 61000-4-6 +I-0.5KV.+/-IKV and +/- 2KV for ac power line(s) to ground +I-OSKV,+/- IKV for ac power line(s) to line(s) Frequency: I50 KHr-80 MHz (For most of the life suppolting equipment), 3 V m . Freq:I50 KHz-80 MHz (for most of life supporting equipment. 3Vms for frequency band other than ISM lOVms for ISM frequency band dwell time = min 3sec for equipmentlsystems tested with 2HZ and lsec for all other erruioment Voltage Dips and voltage variation on power supply input lines No Reauirement 60% for 5 period 30% for 25 period 295 for 5 sec (For equipment with rated input power of IKVA or less and all life supporting equipment) For not life supporting equipment and for which rated input power is greater than IKVMrated current is less than I6Mphase deviation from 36.202.2 j is allowed provided the equipment remain safe experience no component failure and is restorable to pre-test state with operator intervention. For life supporting are .. . equipment >16Mphase equipment .~ exempted from testing IEC 610004-8 Power frequency.3AJm ~~ Power Frequency magnetic Field 41 4 No requirement Proceedings of INCEMIC Authorized licensed use limited to: UNIVERSIDADE FEDERAL DE MINAS GERAIS. Downloaded on June 18, 2009 at 16:20 from IEEE Xplore. Restrictions apply. - 2003