SIM Bilateral CENAMEP AIP - INTN 50/60 Hz Energy Comparison Protocol 2015 1 Introduction The Mutual Recognition Arrangement (MRA) among National Metrology Institutes (NMIs) places particular importance on comparisons to demonstrate a NMI’s ability to measure certain quantities. Reliable measurements of electric energy are the cornerstone of electric restructuring schemes in economies worldwide. Any opportunity of testing the reliability of measuring methods and reference standards at an international level is very valuable to national metrology institutes. The comparison objective will be to compare the electric energy measurement capabilities of the Centro Nacional de Metrología de Panamá (CENAMEP AIP) and the Instituto Nacional de Tecnología, Normalización y Metrología de Paraguay (INTN). These INMs decided to perform the comparison at 50 and 60 hertz. Three points were selected to test the amplitude and phase measuring capabilities of the NMIs: 120 volts and 5 amperes at power factors 1.0, 0.5 lead (ld), 0.5 lag (lg), where lead (capacitive) indicates that the current leads the voltage and where lag (inductive) indicates that the current lags the voltage. 2 Traveling Standard 2.1 General requirements The traveling standard, the RD-23-433 is designed and built by Radian Research Inc. The RD-23-433 Dytronic Single-Phase Reference Standard aims at providing a worst case accuracy of ± 0.01 % for all measurements functions. It is well suited for making transfer measurements of energy. 2.2 Description of the standard The RD-23-433: Serial number: 20 54 59 Size: 190.5 mm (height); 139.7 mm (wide); 139.7 mm (deep), excluding strap Weight: RD-23-433 = 2.5 kg (total transport container = 10.5 kg) Potential inputs: insulated 4 mm multi-contact brand sockets Current input: 6 mm multi-contact brand sockets. 3 input current channels area available, being channel A which must be used for this comparison. Auxiliary input: insulated 4 mm multi-contact brand sockets Input/gating: port 1, BNC connector Output pulse: port 2; BNC connector, to be used for energy measurements only. Three phase SYNC or analog sense: port 3; BNC connector RS-232 communication: a 8 pin RJ-45 jack Operating rate of the RD-23-433 Input current: 0.2 to 67 A Input voltage: 30 V to 630 V, autoranging Frequency: 45 Hz to 75 Hz (fundamental) Auxiliary power input: 60 V to 630 V, autoranging Phase angle: 0° to 360 ° Power factor: 1 to 0 lead, lag Temperature: 20°C to 30 °C Humidity: 0% to 95% non-condensing. 2.3 Quantities to be measured The calibration error of the traveling standard for energy measurements 1. The result to be reported is the calibration error of the traveling standard, defined as the difference between the measured quantity indicated by the traveling standard and the quantity applied to it, and divided by the applied VAh. The calibration error should be expressed in ± μWh/VAh. The error is positive if the traveling standard's indication is more positive than the applied quantity. 2. The average of at least five sets of measurements should be reported. The traveling standard should be de-energized between each set of measurements for at least one hour, followed by at least one hour warm-up period. 3. The total estimated expanded uncertainty quoted in the report encompasses the Type A uncertainty and the Type B uncertainty of the corresponding NMI calibration system. The expanded uncertainty should be estimated for a level of confidence of 95.45 %. Table 1. Test points of SIM.EM-XX Parameter RMS voltage RMS current Power Factor frequency Energy [Wh] 120 V 5A 1.0 0.5 lead-lag 50 and 60 Hz 2.4 Method of computation of the Bilateral Comparison Reference Value The declared values by CENAMEP AIP will be used as the Bilateral Comparison Reference, for being this NMI the responsible for realize the metrological characterization of the traveling standard. 3 Organization The Centro Nacional de Metrología de Panamá (CENAMEP AIP) will be the pilot laboratory. The traveling standards will begin at CENAMEP AIP. 3.1 Coordinator and members of the review committee group Coordinators and contact persons of the pilot laboratory: Julio González (CENAMEP AIP) jgonzalez@cenamep.org.pa Carlos Espinosa (CENAMEP AIP) cespinosa@cenamep.org.pa Organization: Centro Nacional de Metrologia de Panamá (CENAMEP AIP). Address: Edificio 215, Ciudad del Saber, Clayton, Ciudad de Panamá, Panamá. Phone: +507 517 3131. Fax: +507 517 3200. Members of the review committee group: Julio González (CENAMEP AIP) jgonzalez@cenamep.org.pa Carlos Espinosa (CENAMEP AIP) cespinosa@cenamep.org.pa Jorge Parra (INTN) jparra@intn.gov.py 3.2 Participants Centro Nacional de Metrología de Panamá (CENAMEP AIP) ** Instituto Nacional de Tecnología, Normalización y Metrología de Paraguay (INTN). ** This laboratory participated in SIM.EM- S2 and SIM.EM-S7, both 50/60 Hz Energy Comparisons Criteria for participation in the comparison include the availability of adequate trained staff and high-level measurement apparatus and procedures. It is important that personnel in each participating NMI will be available to complete the comparisons within the scheduled time window. If, for some reason, the measurement facility is not ready or custom clearance should take too much time, the laboratory is requested to contact immediately the pilot laboratory. 3.3 Time Schedule A period of four weeks is allowed for the measurements in each laboratory, including the time necessary for transportation. The CENAMEP AIP, will make measurements before and after the measurements realized by INTN. 3.4 Transportation The traveling standard will be hand carried from one laboratory to another by a technician of CENAMEP AIP. The RD-23-433 will be provided with a rugged plastic container, suitable for transported the standard on airplane. The standard will be packaged along with a temperature/humidity logger. During measurements at the participant’s laboratory, make sure that the logger remains on the top surface of the traveling standard, mainly close to the backlit LCD of the traveling standard, in order to log measurements of ambient temperature and humidity. There is no need for the participating laboratory trying to gain access to the logger; its contents will be downloaded at CENAMEP AIP in order to keep track of the changes of temperature or humidity during transportation or while staying at the participating laboratory. On receipt of the standard, the participant shall inform the pilot laboratory by sending an email notifying the reception. Then it should receive a reply (confirmation) e-mail from the pilot lab. 3.5 Shipping and Handling The traveling standard will be hand carried from one laboratory to another by a technician of the pilot Laboratory. The transport case contains the following items: Parts List 1. (1) The RD-23-433 Dytronic Single-Phase Reference Standard. 2. (1) 120 VAC Auxiliary Power input cable for RD-23-433. 3. (2) Converts the standard 4 mm banana plug to 6 mm current inputs on RD-23-433. 4. Temperature and humidity data logger. 3.6 Failure of the traveling standard In case of failure: 1. Unplug the traveling standard. 2. Write immediately to the pilot laboratory describing the behavior of the traveling standard. Send along any message displayed on the backlit LCD. Take photographs of the measuring system and of the connecting cables to the traveling standard. 3. Wait for instructions from the pilot laboratory on the way to proceed. 3.7 Financial aspects, insurance The traveling standard will be provided by CENAMEP AIP. INTN covers the costs of the shipment, transportation and customs clearance. 4 Measurement instructions 4.1 Tests before measurements 1. There are no tests on the RD-23-433 traveling standards to be performed before measurements at the participant’s laboratory. 2. Energizing the traveling standard. The auxiliary power to the traveling standard should be applied at least 4 hours before starting the tests: Note: In order to achieve the best measurement capability of the traveling standard. The RD-23-433, once the standard has been allowed to warm up at least 4 hours at room temperature, the participant must turn down and on the standards. This action will set the internal references of the standard corrected at their internal temperature. 4.2 Measurement performance 1. Perform the tests within 2 weeks, taking as many independent readings of the traveling standard as stated in the calibration procedures of the laboratory. 2. The traveling standard should be de-energized at least two times during the tests. Allow at least 4 fours of de-energizing time before energizing the traveling standard again. 3. Report the date and time when the traveling standard is de-energized and energized. Particular requirements for connecting the traveling standard. Figure 1. Connecting the RD-23-433. NOTE: for the measurement of current when using the RD-23-433, use Input Terminal A, as shown in Photo 1. Photo 2. RD-23-433 4.3 Method of measurements Recommended method: The comparison of electric pulses. The traveling standard is an energy meter of the energy-to-pulse converting type. The energy constant, KH, of this standard is equal to 100 000 pulses/Wh. At 120 V, 5 A and power factor equal to unit, RD-23-433 are able to generate a train of pulses with a frequency equal to 16 666.66 pulses per second. 1. for the voltage and current sources, make sure that their frequency is set at 50 or 60 Hz, according to the testing points shown in Table 1; 2. at every testing point shown in Table 1, make as many independent measurements of the calibration error of the traveling standard as stated on the calibration procedures of the laboratory; 3. Complete the calibration of the traveling standard by obtaining the mean value of its calibration error obtained at the testing points shown in Table 1. The calibration error is defined as the difference between the measured quantity indicated by the traveling standard and the quantity applied to it, and divided by the applied quantity. The calibration error of the traveling standard should be expressed in μWh/VAh. 4. the number of pulses on the RD-23-433 should be set to 1 000 000 (one Million); 5. for 1 million pulses, the integration time Ti for energy measurements will be approximately equal to: a. Ti = 60 seconds, at 120 V / 5A / pf = unit; b. Ti = 120 seconds, at 120 V / 5 A / pf = 0.5 lead-lag. 6. the average of at least four sets of measurements should be reported; 7. the traveling standard should be de-energized between each set of measurements for at least one hour, followed by at least one hour warm-up period; 8. the total estimated expanded uncertainty quoted in the report encompasses the Type A uncertainty and the Type B uncertainty of the corresponding NMI calibration system. The expanded uncertainty should be estimated for a level of confidence of 95.45 %. 9. report the mean value of the ambient temperature and relative humidity of the laboratory; 10. the measurement report of the participant may be completed according to Annex A2 Layoutof the measurement report 5 Measurement results and uncertainty The uncertainty of measurement must be determined following the Guide for the Expression of Uncertainty in Measurement [3]. Participant laboratories are requested to report the main uncertainty components of their measurement systems, identifying all the pertinent uncertainty sources and quantifying their contribution to the expanded uncertainty. In order to have a comparable uncertainty evaluation, each laboratory is asked to report the following information in the form of an uncertainty budget: 1. The result of the type A method of uncertainty estimation which yields the standard deviation of the mean values of data sets recorded by the participant in order to calculate its final report value. 2. The result of the type B method of uncertainty estimation. 3. The expanded uncertainty estimated at a 95.45 % level of confidence. 4. The degrees of freedom for the estimation of the expanded uncertainty at a 95.45 % level of confidence. A suggested format for the uncertainty budget is given: Main Uncertainty Components (yi) Standard Uncertainty u(yi) Type method A or B of Sensitivity Degrees of Uncertainty evaluation/probability coefficient freedom contribution u(Ri) distribution function Ci ni Desviation of the readings of traveling standard µWh/VAh µWh/VAh Standard calibration uncertainty µWh/VAh µWh/VAh Resolution of Standard µWh/VAh µWh/VAh Drift of Standard µWh/VAh µWh/VAh Standard temperature coefficient µWh/VAh µWh/VAh Root square sum of type A standard uncertainties and effective degrees of freedom µWh/VAh Root square sum of type B standard uncertainties and effective degrees of freedom µWh/VAh Combined standard uncertainty and effective degrees of freedom µWh/VAh Expanded uncertainty (95.45% coverage factor) µWh/VAh 6 Measurement report The participant laboratory should report the following information within 4 (four) weeks from the end of measurements: a) The mean of the calibration error of the traveling standard. The calibration error should be relative and expressed in terms of ± μWh/VAh. b) The expanded uncertainty of the error of calibration of the traveling standard at 95.45 % level of confidence. It must be expressed in terms of ± μWh/VAh. c) The uncertainty components coming out from the Type A and B methods of uncertainty estimation of the calibration error of the traveling standard. d) The expanded uncertainty of the calibration error of the traveling standard should be determined at a 95.45 % level of confidence. e) Measurement uncertainties shall comply with the Guide to the Expression of Uncertainty in Measurement [3]. f) According to the calibration procedures of the participant laboratory, the mean values of ambient temperature and relative humidity should be reported. g) A complete uncertainty budget and any additional information must be reported as shown in Annex A3. The participant must submit their results of the comparison to the pilot laboratory (jgonzalez@cenamep.org.pa; cespinosa@cenamep.org.pa), as soon as possible, in every case, not later than 4 (four) weeks after the measurements are completed. Drafts A and B will be the responsibility of the review committee, following the BIPM guidelines. 7. REFERENCES 1. Rene Carranza, SIM.EM.-S7 Technical Protocol, SIM Supplementary Comparison of 50/60 Hz Energy BIPM 2010-2011. 2. CCEM Guidelines for Planning, Organizing, Conducting and Reporting Key, Supplementary and Pilot Comparisons. CCEM, 21 March 2007. 3. Evaluation of measurement data – Guide to the expression of uncertainty in measurement, BIPM JCGM 100:2008. Annex A1 List of participants Organization Country Contact Person Julio González CENAMEP AIP Panamá INTN Paraguay Carlos Espinosa Jorge Parra E-mail jgonzalez@cenamep.org.pa cespinosa@cenamep.org.pa jparra@intn.gov.py Annex A2 Layout of the measurement report 1. Identification of the traveling standard: RD-23-433 2. Identification of the participant laboratory and its representative 3. Measurement set-up and traceability scheme 4. Measurement procedure 5. Results: a) Mean value of the calibration error of traveling standard at the testing points shown in Table 1, expressed in ± μWh/VAh. b) Expanded uncertainty estimated at a 95.45 % confidence level and the degrees of freedom of the calibration error of the traveling standard. c) Mean date of measurement. d) Ambient conditions: mean value and spread of temperature and humidity measurements. 6. Detailed uncertainty budget (view the point 5 of protocol) 7. Report the date and time when the traveling standard is de-energized and energized. 8. Signature and title of the laboratory representative.