June 2011 Level 1 Training Manual: KI9600A Series Pocket Power Meters Course Contents 1. 2 2. 3. 4. 5. 6. 7 7. 8. 9. General Features Models Detector Types Keypad Instrument Care Operation Operating Modes Log / Linear display Special Features 2 1 G 1. Generall Features F t ¾ ¾ ¾ ¾ Entry skill level Fits in shirt pocket Test tone detection Industry standard interchangeable connectors – Including SFF ¾ ¾ ¾ ¾ ¾ Battery life – 300 hours Calibration at multiple wavelengths Ruggedised, water & dust resistant 3 year re-calibration interval 3 year warranty – extendable to 5 or 7 years ¾ Special Features: ‘Multi-fiber ID’ , Sl M d and SlowMode d TamperLock T L k 3 2 M 2. Models d l There are two model styles: A. KI9600A: for popular connector types – the most popular. B. KI9600-XL: fitted with a wide area detector - plastic fibre, MT-RJ, ribbon and specialty fibre applications up to 3 mm core diameter. diameter 4 3 D 3. Detector t t T Types Several Detector types are available to suit your application. Choose a detector style y to suit yyour p power and wavelength g requirements. q InGaAs Telco & LAN • Power Levels of up to +5 dBm • Most accurate detector d type at Telco l wavelengths Ge LAN at 850 & 1300nm • Power levels of up to +10 dBm • Not as accurate as InGaAs. • Can be affected by temperature • Not recommended above 1550 nm H series High power – CATV, DWDM • Power level configurations of up to +24 dBm • Filtered InGaAs Si Wide area detector configuration - LAN at 850nm, POF • Power levels of up to 0 dBm • Will not work at 1300 nm • Available XL series meters only 5 3.1 Typical Detector Spectral responses 1.2 Power Meter Wavelength Dependence 1.1 Relative Respons R sivity A/W New InGaAs1 detectors work 09 wellll att 8500.9 nm. E l InGaAs Early I G A M Meters t 0.8 Ge 0 ºC 0.7 0.6 0.5 Si 04 0.4 Ge 25 ºC 0.3 0.2 InGaAs B series 0.1 0 600 700 800 900 1000 1100 1200 Wavelength nm 6 1300 1400 1500 1600 1700 4 K 4. Keypad d Connector housing cap Connector l release DUAL FUNCTION Hold + Min/Max Power On/Off DUAL FUNCTION dB dB dBm, dBr & W + SET REF Wavelength g toggle and special features 7 5 IInstrument 5. t t Care C Keep the instrument in a carry case during storage and transport • Use only high quality batteries. • • • • • • For prolonged storage remove batteries. The instrument is resistant to normal dust and moisture, however it is not waterproof. If moisture gets into the instrument, remove batteries & dry it out carefully before using it again. again Where possible, keep instrument away from strong sunlight. Clean the instrument case using Iso-Propyl-Alcohol (IPA) or other non solvent cleaning agents. agents DO NOT use Acetone or other active solvents as damage may result. 8 6 O 6. Operation ti 1 Fit batteries 1. 2. Turn power meter on 3 Select / determine test cord configuration 3. 4. Fit correct adaptor style to meter 5 Measure 5. 9 6 1 Fit batteries 6.1 b tt i To Fit Batteries. • Press battery cover latch down and push away from case. • Insert 2 ‘AAA’ cells • Replace battery cover Battery life. Alkaline ‘AAA’ batteries: 300 Hrs Low Battery Display. I di t shows Indicator h when h batteries b tt i are low. l Warning. Do not use lithium batteries, batteries or other batteries with a nominal voltage greater than 1 1.8 8 V as the instrument may become damaged. 10 6.2 Power On O To turn on Push the Aqua [On/Off] b tt button -0.59 0 59 1550 nm dB m Hold Pocket Fiber Meter dB/ dBm /mW Set Ref • 10 minutes auto off • Permanent operation • Low battery indicator HOLD Max Min on To Defeat auto time out. Press [On/Off] & [HOLD] for several seconds. • Instrument will beep twice. • ‘on’ on will display. display • ‘Perm’ will display on the upper part of the LCD. Perm Pocket Fiber Meter dB/ dBm /mW Set Ref 11 HOLD Max Min 6.3 Determine type of test cord Identify fibre type required for test cord and configuration required. Connector / boot colours may identify the fibre type:OM1 (62.5 µm) Beige OM2 (50 µm) Black OM3 (50 µm) Aqua SMF Blue or Green Cordage colours may identify the fibre type:SMF: Yellow MMF Blue, Grey/Slate, Aqua or Orange Unlike sources, power meters accept both PC and APC connectors. • EXCEPTION – wavelength selective meters are end face specific! 12 6 4 Fit connector 6.4 t Once the required test cordage configuration is identified, a matching adaptor for the power meter can be selected. Fitting the adaptor depends upon meter style 13 Standard & Wavelength Selective Instruments Supplied with :- SC connector as standard Other connector styles include:ST FC ST, FC, LC, LC E2000, E2000 2 2.5mm 5mm Universal , SMA , MU MU, D4, D4 DIN * Universal not recommended for sources 14 To fit / change g an adaptor p / connector • • • • Locate quick release button on front of instrument at base of connector housing Push and hold button in Pull out existing adaptor Fit new adaptor Sl t tto Slot outside 15 XL Series Connectors • IIndustry d t standard t d d 7/8” 7/8”-28 28 TPI adaptors d t • Select required adaptor style • Screw on adaptor 16 7 O 7. Operating ti Modes M d Connect power meter to DUT. In the presence of light the meter will operate in one of three ways: 1. Measure Power: If receiving light from a ‘simple’ source. – – – Meter M t will ill display di l power. dBm, dBr or W The operator must select correct wavelength. 2 Modulation Frequency: 2. Automatic display of modulation frequency. – Modulation 200 to 2,500 Hz 3. Multi-fibre Identification number: Automatic display of 12 fibre identification numbers. 17 71M 7.1 Measure P Power Power meter receives light from the source. User must select the correct wavelength. To select wavelength. • Toggle [λW] or [Xλ] to select λ -0.59 0 59 – Wavelength toggling is circular. circular 1550 nm dB m Hold Pocket Fiber Meter dB/ dBm /mW Set Ref [<λ] 18 [>λ] HOLD Max Min 7.2 Test Tone When receiving g modulated light. g • Instrument displays modulated frequency & beeps. When e receiving ece g Test est Tone. o e • • 270, 1,000 or 2,000 Hz. Instrument displays modulated frequency & beeps. 270 Tone measurement can be disabled in Slow Mode. 850 Hz Hold Pocket Fiber Meter dB/ dBm /mW Set Ref 19 nm HOLD Max Min 7.3 Fibre ID Number Some Light sources have the ability to send a ‘multi-fibre identification’ number in the range of 1 to 12. 12 When receiving a Multi Multi-fibre fibre identification signal: • Displays Fibre identification number • • Applies to KI9600 Firmware version r0.05 and above. Legacy instruments will display a modulation frequency. Source 0.00 dB m I d04 1550 Perm nm I Pocket Fiber Source MOD d04 1550 nm Hold Pocket Fiber Meter LEVEL dB/ dBm /mW Set Ref 20 HOLD Max Min 8 L 8. Log/linear /li Di Display l The power meter can display power in logarithmic (dB) or Linear (W) units. 1. Absolute Mode Measure actual power level at a particular location – dBm – Typically yp y used during g installation & maintenance activities. 2. Relative Mode Measure power level ‘relative’ to a particular location – dBr – Typically used for loss measurement. measurement 3. Linear Measures power in watts – nW, µW, W – Typically a laboratory measurement. The meter defaults to last setting and wavelength at power on. 21 Toggle - Absolute/Relative/Linear To toggle between Absolute / Relative and Linear mode mode. • Briefly, Press soft button [dB/dBm/mW] to toggle between the three modes. dBm dBr 1625 -24.17 W 1625 -0.56 -24.1 nm dB m Hold Pocket Fiber Meter dB/ dBm /mW Set Ref nm dB R Hold Pocket Fiber Meter dB/ dBm /mW HOLD Max Min Set Ref [dB/dBm/mW] HOLD Max Min [dB/dB / W] [dB/dBm/mW] 22 3.83 1625 nm uW Hold Pocket Fiber Meter dB/ dBm /mW Set Ref HOLD Max Min [dB/dBm/mW] Set S t Reference R f M tb Must be iin R Relative l ti mode d (dBr) (dB ) nott Absolute Ab l t mode d (dBm). (dB ) • Press and hold [Set Ref] for several seconds. – Meter M will ill beep b 2 times i and d zero at indicated i di d wavelength. l h – Reference value remembered at power off. 1625 -0.56 -24.1 dBr dB R Hold dB/ dBm /mW nm dB R Hold Pocket Fiber Meter Pocket Fiber Meter Set Ref 1625 -0.00 -24.7 nm HOLD dB/ dBm /mW Max Min Set Ref [Set Ref] 23 HOLD Max Min 9 S 9. Special i l Features F t 1. Hold Thi feature This f t iis useful f l where h you wantt tto show h a measurementt tto a thi third d party. t 2. Min & Max recording This feature is useful where a level needs to be monitored over a time period. p 3. SlowMode • Tone detection disabled, enables measurement power in the presence of audio level modulation. modulation • Improved low power measurement accuracy. 4. TamperLock Mode • Display settings are locked locked. • Useful for low skilled areas or where only 1 wavelength is in use. 24 9.1 Hold Function T activate To ti t th the H Hold ld function: f ti • Briefly press [HOLD] • The instrument will beep. • The word ‘Hold’ will flash in the lower RHS of the display. display To turn Hold off. • Briefly press [HOLD] -0.59 1550 nm dB m Hold Pocket Fiber Meter dB/ dBm /mW Set Ref 25 HOLD Max Min [Hold] 9.2 Min / Max To retrieve the Maximum and Minimum value: Mi Min Press and hold [Max Min] • Instrument will beep twice and • the word ‘min’ will display on the upper RHS of the LCD. Repeat above to toggle to ‘max’ Repeat above to turn function ‘Off’. • • • -16 55 -16.55 1550 nm min dB m Hold Pocket Fiber Meter dB/ dBm /mW Set Ref HOLD Max Min To reset Max/Min: • • Toggle to another wavelength and back or Turn power Off and then back On. [Max Min] -12.03 1550 Max nm max dB m Hold Pocket Fiber Meter dB/ dBm /mW Set Ref 26 HOLD Max Min [Max Min] 9.3 SlowMode To Activate at Power On • Press P and dH Hold ld [dB/dBm/mW] [dB/dB / W] while hil turning t i meter t On. O ““ tonE” t E” will ill be b displayed. di l d To Activate when meter powered On • Press and Hold [λW], then press [dB/dBm/mW] “- tonE” will be displayed. To Activate when test tone detection is active • Press [dB/dBm/mW] . When SlowMode is On On, alternating “X” X and “B” B symbols will be displayed. To exit SlowMode • Press and hold [λW], then press [dB/dBm/mW] or turn power Off and then back On. 27 -24.17 1625 nm B dB m Hold Pocket Fiber Meter dB/ dBm /mW Set Ref HOLD Max Min 9.4 Tamper Lock Mode To activate • Set meter to required λ & measurement mode mode, e e.g. g 850 and dBm • Press and hold [λW], then press and hold [HOLD] • ‘codE’ will display • Press P any 6 kkeys, other h than h [Power] [P ] • record keystroke sequence for future use. • Instrument will beep 3 times and display ‘Lout’ 850 <<< When active -24.17 24 17 dB m >>> Hold Pocket Fiber Meter • ‘<<<‘ and ‘>>>’ displayed dB/ dBm /mW Set Ref To clear • • • • nm Press and hold [λW] & [HOLD] whilst turning power On ‘codE’ will display , enter 6 key code sequence If successful ‘PASS’ will display. Alternatively, TamperLock will be cleared after 3 unsuccessful attempts. 28 HOLD M Mi Max Min Application Notes Comprehensive selection available at www kingfisher com au/ApplicationNotes htm www.kingfisher.com.au/ApplicationNotes.htm 29