P2A - PinPoint Analog Card The Analog card adds a suite of analog tools to the PinPoint. These instruments can be used as part of an integrated test sequence or as individual analog instruments as required. The card includes a: u Frequency Counter u DMM (Digital Multi Meter) u Oscilloscope (option) u Arbitrary waveform generator (1 ARB is included as standard, up to 3 more can be added). Analog Card Specification Analog cards are inserted here Recommended option is one per system u DMM is capable of AC and DC voltage and current measurements, resistance, diode and capacitance measurements. u Frequency counter will perform Frequency, Period, Time AB, Unit Counter and Frequency Ratio measurements. u 16 Channel (8 Differential Channels) Analog Input module. u The Arbitrary Waveform Generator(s) can provide Sine, Square, Triangle, Sawtooth and User Waveforms as well as DC offset levels. u 100MHz Bandwidth Dual - Channel Oscilloscope. u 400Hz 26 Volt AC Avionics Reference Voltage output. The Analog card can be integrated at any time by simply plugging it into the PinPoint system. The software will automatically recognise and configure the card so that the functionality it provides becomes part of the overall system capability. Software is provided to enable the instruments to be used either in Immediate Mode through a Virtual Front Panel or as part of an integrated test procedure. The Analog card provides the functions of a 4 1/2 Digit DMM, Dual Channel Frequency/Timer Counter, Multi-channel Analog Inputs, Arbitrary Waveform Generator(s) and a Dual-Channel 100MHz Bandwidth Oscilloscope. For your local office details please visit our web site: www.diagnosys.com DMM (DIGITAL MULTI METER) TIMER COUNTER Display Digits: 4 1/2 Display digits 8: (32 Bits) Display Count: 20000 Modes: FREQ, PERIOD, UNIT COUNTER TIME AB (100ns Min), FREQ RATIO Conversion Rates: 2,2.5,4,5,10,20 Maximum input voltage: 450V (320V AC) Maximum floating voltage: 1000V (750V AC) Input Channels (CHA, CHB, HIFREQ) Maximum input current: 2A continuous (3A for 30 seconds) Ranges: Maximum open circuit voltage: 450V (320V AC) CHA - 30MHz CHB - 30MHz HIFREQ - 20MHz to 120MHz Input threshold voltage: Programmable ± 10V w.r.t. ground earth (CHA and CHB) 0 - 5V p-p (HIFREQ) Input Impedance: 1MΩ paralleled by 22pF (CHA and CHB) only Accuracy ± 0.001% DMM Measurement Characteristics: Range: 200mV, 2V, 20V, 200V, 1000V DC Voltage: Accuracy: (% + 10 counts) ± 0.1% u Overload protector AC Voltage Range: 200mV, 2V, 20V, 200V, 1000V Accuracy: (% + 40 counts) 50 to 100Hz: ± 0.1% >100Hz to 1Khz: ± 2.5% >1KHz to 10KHz: ± 3.5% >10KHz to 20KHz: ± 5.0% Bandwidth: 20KHz Input Impedance: 1MΩ Paralleled by 100pF Current (AC and DC Ranges) 2A DC Accuracy: (% + 10 counts) ± 0.5% AC Accuracy: (% + 40 counts) 50 to 100Hz: ± 1% >100 to 1KHz: ± 2.5% Bandwidth: <= 1KHz Resistance Ranges: 200Ω, 2KΩ, 20KΩ, 200KΩ, 2MΩ, 20MΩ Ω(% + 10 counts): 0.2% 200Ω: 1.0% 20MΩ: 5.0% Diode Test Test current: 1mA Test voltage: <=10V Capacitance Ranges: 200nF, 2uF, 20uF, 200uF Accuracy: ± 1.0% ARBITARY WAVEFORM / FUNCTION GENERATOR (1 included on Analogue card as standard) 4 independent channels DIGITAL OSCILLOSCOPE Number of channels: 2 Input voltage range: 50mV to 50V u 200 MS/s per channel real-time sampling 12V u 8 bit vertical resolution (1 fitted as standard, up to 3 Optional available by ordering part P2A-ARB and quantity required. These are implemented on daughter boards) u 32K samples per channel u 100 MHz Bandwidth u Internal and external triggering Output voltage range ±: Waveform Generation Types: Sine, Square, Triangle, Sawtooth, Pulse, User Frequency: 0.1Hz to 300KHz 0.1Hz resolution Functions: Offset, Skew/Duty cycle adjustment Waveform Source Resolution: 16Bit Samples/Sec: 30MSPS Frame depth: 32Kbyte Accuracy: ± 0.5% Current Limit Analogue Adaptor Mode: Max current limit 20m Amps. Indefinite per channel In-Circuit Mode: Max current limit 350m Amps. for ≤ 25ms. (Ch 1 only) Analogue Inputs Analogue Inputs: 16 Single Ended or 8 Differential Input Resolution: 16 Bit Sampling range: 1.25 MS/s Voltage range: ± 12V u Internal and External triggering For your local office details please visit our web site: www.diagnosys.com Diagnosys has a policy of continuous product improvement and reserves the right to change technical specifications at any time without prior notice. Diagnosys does not accept liability for errors or misprints in this document.