THERMOFID

advertisement
FID analyzer system for Total hydrocarbons
from ppb to LEL and VOC to TOC:
THERMOFID
EN 12619
Determination of the mass concentration of total gaseous organic carbon at low
concentrations in flue gases - Continuous flame ionisation detector method
Explosion protection according EN50054 en EN50057, thermal and catalytic oxidizers,
regenerative thermal oxidizers (RTO), vapor recovery units (VRU) scrubbers, industrial
furnaces. SIL approval. EN-IEC 61779, EN-IEC 60079
Alternative for Flammability FTA Flame Temperature analyzer Prevex SNR van Control
Instruments Inc.
Emission measurement of total organic carbon according EN-IEC 13526, EN-IEC 12619,
17. BImSchV, TA-Luft, MCerts (Guideline 2000/76/EG), , VDI 3481/3, CvGM-VKL-013.
• Measurement of total organic C. according MCerts 2000/76/EG
• Emission measurement on incinerators (EN 12619 en EN-13526)
• Detection and measurement of flammable gases according
EN-IEC 50054 + EN-IEC 50057
• ppb to Vol % CxHy in air, nitrogen, CO2 , steam and condensate.
• T90 time < 2 seconds
• Interference 0 - 25 Vol % O2: < 3 % Span
• Option for methane / non methane analysis
• Analyse of TOC + CO&CO2 in N2, Argon, Helium, H2 en
• Designed to measure chlorinated hydrocarbons
THERMOFID
Analysis of hydrocarbons at all levels
TOC, VOC, VOS, TC, TC - CH4, TC + CO + CO2,
3
mg C/m , ppm, Vol %, mg CxHy, L.E.L.
Multifunctional
The THERMOFID is used in a variety of industrial applications, environmental measurement systems and R&D
projects. Typical use of the Thermo-FID are LEL control, flue/exhaust gas at waste incinerators or the petrochemical
industry, ambient air analysis, work place control MAK/TRK-limits and process control optimization.
Microprocessor for optimal control and operation
The microprocessor controlled Thermo-FID provides menu driven ‘clear text’ user friendly interface information on an
alphanumeric display. All operational and self diagnostic data are also available on RS232 and/or analog/digital
outputs (status alarm PCB, optional). A fully automated and continuous self diagnostic and log book entries for all
operation and calibration data guaranties an optimum analyzer up time.
The unique close couple detector design (no cables, etc.) provides a low drift
operation and a wide dynamic range with always optimum signal noise ratio. All
digital input feature are designed according NAMUR guidelines, all digital outputs
are potential free contacts. Analogue output signals are offered as standard 1x 0/420mA, 600 Ohm load (not galvanic isolated), optional 4 x 0/4-20mA optical galvanic
isolation is available on the status & alarm card. Instrument operation, configuration
and remote control is also available on 2 x RS232.
The measurement principle of the Thermo-FID is based on the ionisation of
hydrocarbons in a hydrogen flame. An exceptional constant sample gas flow and a
constant hydrogen stream are drawn to the burner nozzle which being on a negative
potential measured by the close coupled detector. In the temperature controlled
burning chamber, the mixture is burned by adding hydrocarbon free air in relation of
approx.10 : 1. The resulting ions are collected on a polarized electrode, amplified
and put to display
Best measurement performance is provided by offering a very constant
sample mass flow and pressure conditions
the Ionisation
chamber.Detector
Due to the
THERMOFID to
Flame
Maximum reliability minimal maintenance
optimum design of the instrument, pressure variation at the sample point are
The corrosion proof detector is heated
to 200 °C. 800mbar
Twin air-jet abs. • up
allowedupbetween
to 1600mbar
without lack of
Complete
diagnose of allabs.
utilities
aspirators for the sample flow in a heated detector never wear
performance.
out. Flow controllers, sensors and valves get in contact only with
clean (compressed) air, hydrogen and span gas.
This in combination with the dilution of the sample gas with
compressed air as exhaust gas, results in an analyser with low
service costs and high availability..
Service
The modular design of the THERMOFID en the accessibility of
essential parts in the detector guarantee a simple service and
maintenance on-site. Specialists with decades experience on
analyser and sample-handling with FID´s will assist you with
advice and engineering, coordination of the system integration
and assistance before during and after start up.
• Automatic control & adjustment of flame temperature
• Automatic start up
• Automatic calibration including shut off valve
• Heated corrosion proof detector
• Maintenance free air-jet pumps
• A/D converter direct at the measuring electrode
• Sample gas pressure 800 to 1600 mbar absolute
• Integrated catalyser for Zero and burner air
THERMOFID
Technical data
Supply
Consumption
Analogue output
Serial output
Input
Output
Compressed air
Hydrogen Q. 5.0
Spangas
:
:
:
:
:
:
:
:
:
230/115 V~, 50..60Hz
250 VA
1 x 0(4)-20 mA (600 Ω)
RS 232, Epson / HP compatible
1 x PT 100 for heated line or probe, adjustable: 60 … 210 °C
930 VA for heated line or probe
3
5 bar Abs. / 2Nm /h
2 bar Abs. / ca 40 ml/min
3 bar Abs. / 130 ml/h during calibration
Analytical data
Range
Detection limit
Response T90
Response T90
Sample Flow
Inlet pressure
Detector temp.
3
: 0 … 1 mg org. C/m to 100 % LEL
3
: < 0,1 mg org. C/m
3
: < 0,5s at a measured value > 20 mg org. C/m , excl. sample line
3
type MK with 0,5m probe < 3s at a measured value > 20 mg org. C/m
3
< 2 s with optional dilution for MK range > 20 mg org. C/m
: according 17. BimSchV: 5s
type MK with 0,5m probe < 7,5s
: aspiration by air jet ; 10, 25 or 60 l/h at 1013 mBar
: 800 – 1600 mBar.
: 110 … 200 °C
Housings
19” Rack 84TE/3HE type ES
Dimensions
: 132,5 x483 x 271 mm (HxWxD)
Protection
: IP 20
Table top 84TE/3HE type TS
Dimensions
: 135 x463 x 275 mm (HxWxD)
Protection
: IP 20
Wall mount type FE
Dimensions
: 450 x 440 x 400 mm (HxWxD)
Protection
: IP 54
Direct extraction type MK
Dimensions
: 450 x 440 x 400 mm (HxWxD)
Pretection
: IP 54
Probe
: Flange DIN 2527 PN16 Form B DN80
I/O card status & alarm








3 x D/O
4 x A/O
4 x D/O
2 x A/I
2 x D/O
2 x D/I
2 x D/I
1 x D/O
: Fault, Service, Maintenance (potential free contact)
: 4, 0(4)-20 mA, galvanic isolation, 600 Ω
: Alarm relay, free programmable (potential free contact)
: Pressure transmitter for gas utilities (H2 , span), included 24Vdc supply
: 24VDC/ 500mA, for H2 switchover
: for Auto-Cal, Start & Cal Stop
: free selectable status parameter
: 24VDC/ 500mA, free selectable timer
••••••••••••••••••••••••• •
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
THERMOFID accessories
MSU 2 x 8 points
Sample flow
Flow
Flowalarm
Tubing
Sample tube
Selection
Output
: Parallel with air jet, for area measurement of TOC, VOC, VOS.
: 35 Nl/h
: each sample point
: ø 4mm id / 6mm od
: max 70 mtr per point
: Programmable by ThermoFID
extern by optional I/O card
: Analogue 0(4)-20 mA per point (option)
Potential free contact per point (option)
.
TOC/VOC in steam
Pressure
Points
Housing
Heating
Selection
Output
: 2 – 16 bar
: 1, 4 of 8
: Stainless steel
: Electric up to 180 °C
alternative by steam or condensate (6bar)
: Programmable by ThermoFID
extern by optional I/O card
: Analogue 0(4)-20 mA per point (option)
Methane / Non methane
THC
CH4
Linearity
Zero drift
T90
O2 influence
SPAN gas
: 10 to 1.000 ppm C3H8 with zero air via catalyser
2 tot 1.000 ppm C3H8 with zero air N2
: 10 tot 1.000 ppm CH4 with zero air via catalyser
2 tot 1.000 ppm CH4 with zero air N2
at THC < 100 ppm C3H8 methane failure is < 1 ppm
: < 1 % of measured value
: < 1 % of measured value /24h
: THC< 15 sec, CH4 < 30 sec
: < 4% of measuring range at 8 to 25 Vol% O2
: CH4 in air
THC + CO + CO2
Principle
Range
Sample gas
Catalyser
Maximum
: catalyser for CO and CO2 with H2 to CH4
: 0,01 ppm to 300 ppm THC
: 0,2 to 4 bar
: 350- 370 °C, controlled at 1 °C
: 300 ppm CO/CO2 as sum
Explosion proof:
Versions
Zone
Approval
: FE: Wallmount
MK: Direct mounting to stack
: Zone 1 and Zone 2 versions
: BVS 09 ATEX E 119 X (ATEX 94/9/EG) IIC 2G Ex px II T2-T4
BVS 09 ATEX E 144 X (ATEX 94/9/EG) II 2/3G Ex pz II T2-T4
EEx p versions have limited pressure and temperature specifications.
Compur Monitors SARL
155 Avenue du General de Gaulle
F-92140 Clamart
Phone: +33 145378951
Fax : +33 145378856
E-Mail: compur@compur.fr
Compur Monitors BV (Benelux)
De Veldoven 49
3342 GR Hendrik Ido Ambacht (NL)
Telefoon : +31(0)78 – 682 05 37
Fax
: +31(0)78 – 682 19 36
E-mail
: info@compur.nl
Internet : www.compur.nl / be
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Download