1 Sl.No1. Specification for Water Purification System 1. General: Compact, Wall mountable, sleek water purification system for Ultra – Pure Type-I water for molecular biology application. 2. Quality of water:- Should deliver Type I/Ultra –pure as per International specifications as follows: Resistively: 18.2 MΩcm; TOC (typical):< 5ppb; Bacteria:<1CFU/10ml; Endotoxin: <0.001 EU/ml; RNase: 20pg/ml; DNase:0.002ng/ml Important : Should have an ability to deliver Type III water also 3. Volume: 10-12 litre/day. 4. Feed –water: Potable tap water 5. General Essential Features: (i) Preferably should start giving pure water immediately after start. (ii) Should have built in RO, Dual Wavelength UV lamp, Purification pack or DI cartridge. (iii) Should have built-in integral tank with minimum 5-7 litre capacity. (iv) Facility to attach external point of use Bio filter or any such filter to control endotoxin. 6. Dispenser: Should have a flexible handle dispenser, preferably, with a facility of : a. Variable flow drop by drop up to 2 litre per minute. b. Locked dispense for glassware filling. c. Auto volume dispense from 50ml to 60 liters. 7. Display: TOC TEMPERATURE, conductivity, volume dispensing, alarm and tank water level indicator. 8. Other important features:Sanitization facility to avoid contamination (i) Facility of intermittent water recirculation to avoid contamination build-up. (ii) USB Port to capture data. 9. Accessories:(i) UPS/Stabilizer/CVT as required for functioning of the equipment (ii) All cartridges, filters, pump or any such item which is /are essential for Installation and functioning /operating the equipment. 10. Consumable: Must Quote separately for consumables (cartridges, filters etc.) for ONE YEAR for trouble free working. 11. Warranty: For TWO years guarantee plus warranty for at least three years. 12. List of User: Must include a list of user of the same model of the equipment along with valid email ID and phone no and other contact details. In case the list is NOT provided or incomplete, the quotation will NOT be considered for technical/price comparison. 2 Sl. No. 2 Refrigerated micro centrifuge Technical Specifications: • • It should be table top model with ergonomic design. Must have wide temperature range of -4°C- to + 40°C, which is modifiable during the run operation. • • Centrifuge must have a fast cool function where. Must have a standby cooling and heating mode up to at least 6 hours which enables cooling or heating even when not in operation. Must have automatic temperature start up at a specific time to automatically pre-prepare temperature before centrifugation Must be able to maintain at 4°C at max. speed for the entire period of opeartion Max speed of approximately 17,500rpm, max rcf of 30,000g adjustable from 100rpm upwards in an increment of 10rpm up to 3000rpm and in an increment of 100rpm thereafter. Acceleration time to maximum speed must be less than 15s for all rotors at max load and a deceleration time of less than 15s with the standard 30 x 1.5/2ml rotor, but should have an option to accelerate and decelerate gently for sensitive samples. Timer for run can be set up to 99minutes continuous. It should not produce sound at maximum speed with noise level less than 60 dB. It should be provided with fixed angle rotor accommodating 30x 1.5/2ml tubes, 6x 15/50ml and 8 x 8-PCR strips. • • • • • • • • • It should come with 3 years warranty and on-site repair services as and when required. Technical features described above should be supported by original illustrated catalog, with list of users (with all contact detail), and authentication of vendor’s business record in the form of Income Tax return for, at least, of last 3 years 3 Sl. No. 3. Automated Peptide Synthesizer (01 Unit) 1. A fully automated and programmable solid phase peptide synthesizer with all essential accessories listed at Sl.No. 20 below 2. All operations of solid phase peptide synthesis during the synthesis cycle must be performed with single machine unit 3. The synthesizer must have provisions for performing chemical synthesis of peptides using either FMOC or BOC chemistry without any hardware modification in the system 4. System must have provision for synthesis of at least three peptides simultaneously with completely different peptide sequences in each reaction vessel using three independent programmes 5. Variable scale of synthesis : From 50 micro mol to 0.30 milli mol or more 6. Reaction vessels compatible with different synthesis scales must be provided essentially with the machine. A spare set of reaction vessels (compatible for different synthesis scale) must be provided essentially with the system 7. At least 36 amino acid reservoirs must be provided to store amino acid stock solutions and these reservoirs must be connected to reaction vessels with pressurized tubings under inert atmosphere of nitrogen gas 8. The peptide synthesis vessels should be made up of inert, solvent resistant and corrosion resistant material like teflon or glass 9. The reaction vessels must be connected with inert pressurized tubings to deliver solvents and amino acids as well as to drain excess reactants and washings under vacuum 10. The machine should essentially have software controlled, at least two separate solvent delivery robotic arms: one for delivering the solvents and reactants to reaction vessel and the other to deliver piperidine exclusively during deprotection step 11. Essentially there should be provision for mixing the reactants in the reaction vessel by either 180o rotation or vortex mechanism. The equipments for enabling this mixing mechanism like compatible air-compressor with dryer or vortexer must be provided essentially with the machine 12. The system should automatically take care of complete activation of amino acid derivatives by mixing the reactants under inert nitrogen atmosphere and transfer of desired amino acid into each reaction vessel as per set programme 13. Atleast five solvent reservoirs (made of solvent resistant material) should be available for storing different solvents like DMF/DCM, Methanol/Isopropanol, Diisopropyl ethylamine, Piperidine and activating reagents 14. Waste containers of at least 20L capacity with Teflon inserts should be provided with the machine 15. The machine must allow the user to interrupt synthesis reaction at any desired time to take out a sample of resin from synthesis vessels for checking coupling efficiency and to test the reaction conditions 4 16. Security and mal functioning alarms must be inbuilt with machine and should be displayed in the event of low solvents, amino acids and other reagents in the reservoirs, complete filling of waste container, low inert gas pressure, presence of air bubbles in the delivery tubings, power failure etc. 17. Entire procedure of peptide synthesis should be controlled by a computer equipped with user friendly software with flexibility to custom design different synthesis protocols as per needs of the user 18. The software may have provision to allow the user to change the synthesis steps at any desired time during synthesis. There should be provision to re-start the synthesis protocol from any desired step during synthesis cycle 19. The bidder must essentially agree to perform on-site chemical synthesis of at least two peptides of about 20 to 25 mer length (as per users requirement) satisfactorily at the time of installation of machine in the lab 20. The system must be essentially provided with the following items: a. A dedicated computer with latest configuration to operate the machine and installing operating software of machine b. One branded color laser printer for taking the printouts of the reaction parameters and synthesis data for documentation purpose c. One online 5KVA UPS with at least one hour power backup should be provided essentially for ensuring uninterrupted machine operation during synthesis in the event of power failure d. Three filled nitrogen cylinders with regulators for continuous supply of AR grade nitrogen gas to maintain the inert environment during pressurized delivery of solvents and reactants e. One self-standing fume hood (4 feet width) with provisions for corrosion resistant interiors, front sliding glass door, inert gas port, vacuum port, water tap and electricity socket should be provided essentially for handling the aerosol producing and hazardous solvents used during peptide synthesis. The ducting for connecting the fume hood to exterior of lab should be provided by the bidder f. One split Air Conditioner (1.5/2.0 Tonne capacity) should be provided essentially with the system to maintain the ambient working temperature for the automated peptide synthesizer g. The consumable chemicals and reagents for peptide synthesis must be essentially with the system in quantities listed as Rink Amide Resin (100GM), Wang Resin (100GM), All twenty FMOC-L-Amino Acid Derivatives (100GM each), All twenty FMOC-D-Amino Acid Derivatives (25GM each), Coupling Reagents HBTU (500GM), HOBT (500GM) and TFA (1KG)] h. Along with system, all hardware consumables compatible with the machine like tubings, filters, reagent reservoir vials, reaction vials, septa and membrane caps etc. should be provided in sufficient quantities to enable uninterrupted peptide synthesis 5 i. The cost of all consumables like chemicals, reagents and machine hardware items (listed at Sl. Nos. g and h above) must be quoted separately against each items 21. Warranty on the machine and its essential accessories for a period of at least two years from the date of installation must be provided and AMC charges for additional three years must be given separately 6 Sl. No. 4 Inverted fluorescent microscope with image taking capability supported by software and computer Technical specifications Body of the microscope should be with Infinity optical corrected optical system, Extendable optical free space upto 80mm for attaching others attachment in future, facility for 2 way (100:0, 20/80 left port) or more light distribution of light, up/ down focusing, side port for attaching digital camera upgradable to one additional port for attaching another camera, binocular tube with built-in Bertrand lens & darkslide shutter along with diopter adjustment facility. • Should have Universal turret condenser (suitable for all microscopy techniques) with 5 position • It should have 12V 100W Pre-centred Halogen Illumination system • It should be fitted with Eyepiece: 10X with F.O.V 22 or better and diopter adjustment facility on both eyes, anti fungus type • Microscope should have sextuple revolving nosepiece to accommodate six objectives at a time. • Must be supported with DIC Attachment with polarizer & Analyzer with respective prism • XY Rectangular mechanical stage should be also in the microscope • High performance Long working distance Objectives suitable for Brightfield /Phase Contrast Observation with facility of cover-glass correction. 4X, 10, 20x, & 40X • Specification for Epi fluorescence: With 6 position turret filter block, Noise Terminator mechanism incorporated for high signal to noise ratio images with100W HG Illumination for Epi-fluorescence. Band pass Fluorescent filters for FITC/GFP (Ex 465-495, DM 505, BA515-555), TRITC/Rhoda mine/PI/Alexa Fluor (Ex 540/25, DM 565, BA605/55), DAPI/Hoechst (Ex 340-380, DM 400, BA435-485) applications so that no cross talk is allowed. • Should be attached with a Digital Cooled camera with Specifications:- Digital Cooled camera capable of handling brightfield, fluorescence, DIC, darkfield images with 2/3” high density ccd chip, approx. 5.1 million pixel resolution. Peltier Cooling 20 deg below ambient Live display mode: (5m interlace mode – 5.9 frames/sec.; 23 frame per / sec. With roi & binning); binning modes: 2x2, 4x4, digital zoom: up to 16x (8 steps); interval shooting : 5 sec. – 12 hr intervals. Software should come along with camera for acquiring & capturing of images of reputed make. Should have separate modes for different microscopy techniques i.e., Brightfield, fluorescence, dic, darkfield images. Fire Wire port for attaching camera onto desktop/laptop through single wire. • Software should be with following features: • Acquisition and device control upto three-dimensional acquisition, Image Acquisition, Time Lapse Imaging, Z-Stack, Annotation, ND Viewer, Large Image, Macro, Report 7 • Generator facility, Manual Measurement, Vector layer & Multi-Dimensional File Format (ND format) Microscope, camera & software should be from same manufacturer for better compatibility. • Data Processing Unit: Branded Desktop PC (with a latest processor at least Core I3, TFT monitor at least 19” size, 3 GB RAM and 320 GB hard disk, and preloaded with latest operating system software) as accessories and Color Laser Printer of reputed make. • Microscope should have the provision to be upgraded for motorization. • Technical features described above should be supported by original illustrated catalog, with list of users (with all contact detail), and authentication of vendor’s business record in the form of Income Tax return for, at least, of last 3 years Following Accessories that must be supplied with the equipment: 1) On-line 5 KVA UPS with a minimum of 30 minutes back-up For data back-up one additional system in the form of a laptop PC from a reputed make that has at specifications such as Intel core I5 processor, 3GB RAM, 320 GB hard disk, screen size more than 15” with windows7 and MS office 2010. This computer should also be supported by data analysis software of real-time PCR experiments 8 Sl. No. 5 Specifications for Real Time PCR Technical Specifications: Should be a fast gradient real time PCR system for Nucleic acid and protein analysis Should have thermal block to accommodate 96 well PCR plate, 0.2ml strips or 0.2ml individual tubes of standard format System should support fast PCR protocol with ramp rates of ~ 4°C/sec for heating and 3°C/sec for cooling Block temperature should be maintained at 20°C till the start of PCR cycle to minimize nonspecific amplification System should have Optical Detection Module with minimal mechanical movement to capture sensitive fluorescence emission from reaction volumes as low as 5ul – 100ul System should have a minimum of 6 peltier elements beneath the block to maintain block homogeneity for improved data reproducibility System should have the flexibility to adjust ramp rates to run standard protocol without change in hardware. Should have temperature Gradient range of 20°C in all three steps of PCR to provide additional flexibility while extrapolating established PCR protocols to suit real time chemistries. Excitation Light Source (LED Array) with a life time warranty with the capability to excite commercially available dyes that have emission range max of 520-550 (like SYBR Green, Cy5, Cy3, HEX, FAM, VIC and JOE etc). System should have at least 2 emission filters and channel photomultiplier as detector to provide a 9 log dynamic range Should have particle protection technology to prevent dust accumulation in the optics for reliable fluorescence detection Software should support performing o Absolute quantification, o Relative quantification, o Melting curve analysis, o Allelic Discrimination, o Plus/Minus assays, o End point assays o Thermofluor assay for Protein Characterization. Software should provide raw data of fluorescence as well as advanced options to export configurations compatible with analysis software like qBASE, geNorm, REST and GenEx GenEx and Beacon Designer original version software to be supplied along with the instrument. It should be possible to use the system for regular PCR applications by switching off the optical module It should have Open system to accommodate consumables and reagents from different commercial firms. System should be calibrated / validated and should be traceable to international standards System should be able to support twin-tec, white and transparent plates for ease of their handling for better results. Vendor should have a fully functional Application laboratory and technical support team to support and trouble shoot problems. 9 The instrument should be supplied with 1 Desktop having following configuration: Processor (Intel Core i3-370M), 3GB RAM, 320 GB HDD and at least 19’’ inch screen. It should also be supplied with heavy duty laser printer of reputed make. RS 232 communication port and printer interface It should be supplied with three years Warranty and on demand repair services as and when required. Technical features described above should be supported by original illustrated catalog, with list of users (with all contact detail), and authentication of vendor’s business record in the form of Income Tax return for, at least, of last 3 years Following Accessories that must be supplied with the equipment: 1) On-line 5 KVA UPS with a minimum of 30 minutes back-up 2) Split AC (1.5 tons) of reputed make only 3) For data back-up one additional system in the form of a desktop from a reputed make having following configuration: Processor (at least Intel Core i3-370M), 3GB RAM, 320 GB HDD and at least 19’’ inch screen. 10 Sl. No. 6 Biosensor System (01 Unit) 1. Fully automated Biosensor System working on the principle of Surface Plasmon Resonance (SPR) for real time, label-free monitoring of bio-molecular interactions 2. The system must be capable to carry out real-time data acquisition and analysis of the acquired data 3. The system, software and analysis chemistry must facilitate multi-sample analysis. It must be possible to vary the use from a single sample run to unattended multi-sample runs over a period of time. 4. The system must select samples automatically from vials or microplates for analysis. 5. The system must allow analysis of samples at minimum two analysis spot positions overlaid on a single sensor surface. 6. The system must have possibility of auto-recovery of samples for further analytical investigations 7. The system must have an inbuilt auto-sampler for programmable recovery of samples in multi-well format or vials 8. System must have provision to deliver the solvents or buffers on the sensor surface using at least two separate pumps or dispensers or fluidic channels 9. Reservoirs made of inert material for storage of chemicals and buffers used in bio-molecular interaction analysis and waste container must be provided 10. The sensor surfaces must be compatible with different chemistries used to immobilize the bio-molecules like Amine coupling, Thiol Coupling, Aldehyde coupling, Streptavidin-Biotin interactions hydrophobic interactions etc. 11. The machine must be supplied with a starter kit to demonstrate the real-time biomolecular interactions during installation of machine 12. Analysis Software: Acquisition and evaluation of data through user-friendly, computer operated comprehensive analysis software. The software must have flexibility for developing the desired methods as per requirements. All operations like flow rate, sample volume, auto recovery of samples, analysis time and analysis temperature etc. must be software controlled. Malfunctioning alarms must be displayed for better performance of the machine. There must be option to change sequence of operations using simple and user-friendly commands. The analysis software must have following provisions : a. To calculate concentration of active bio-molecules in solution b. To determine affinity constants for molecular interaction c. To determine kinetic rate constants d. To determine binding affinities e. To determine thermodynamic parameters f. To store and export data files in user-friendly format for documentation 13. Sensor Surfaces: The sensor chips compatible with machine with details as given below must be supplied essentially with the biosensor system mentioning cost of each items separately 11 a. Gold coated sensor chips (Plain) x 2) b. Gold coated CM Dextran sensor chips amino modified c. Gold coated CM Dextran sensor chips carboxyl modified d. Protein A immobilized sensor chips : 100 Nos (50 : 10 Nos : 10 Nos : 10 Nos 14. Technical Information Measuring Principle Type of information/Data Presentation Sample type Sample handling Sample recovery Sample capacity Sample volume Range Injection volume Sample recovery Analysis time per sample Detection spots per sensor surface Surface Plasmon Resonance (SPR) Real time, Label-free monitoring of molecular Interactions Concentration of analytes Kinetics and affinity constants Thermodynamic Parameters Biomolecules like Protein, DNA, RNA, antigen, antibody and Organic molecules like peptides, crude cell extracts, lipid vesicles, drugs and viruses etc. Automated sample loading and injection in desired quantities Automatic sample recovery in 96 vials or 96/384 well plates Facility to store a minimum of 96 samples/reactants in vials or plates In Microlitres 20ul to ≥120ul Fully automatic recovery of samples 1minute to <10 minutes Atleast two spots per sensor surface for detecting interaction between immobilized molecules and test sample against a reference Number of channels At least two channels for delivering samples on the sensor surface Sensor Surface Gold coated glass/solid surface Surface Chemistries Sensor surface must be compatible with chemistries commonly used for site directed immobilization of biomolecules Re-use of sensor The sensor surface must withstand regeneration buffer surfaces pH and conditions for repeated use Refractive Index Range 1.33 to ≥1.37 Analysis temperature 4°C to ≥ 40°C Molecular weight 200 Daltons or more detection Baseline noise ~ < 0.5 RU/min Electric voltage 220-240 15. System must be essentially supplied with a dedicated computer with latest configuration for operating the biosensor system and installing operating and analysis software of machine 12 16. System must be essentially supplied with one branded color laser printer for taking the printouts of biomolecular interaction data for documentation purpose 17. System must be essentially supplied with branded 5KVA Online UPS with at least 1 hour power backup for ensuring uninterrupted functioning of biosensor system 18. One split Air Conditioner (1.5/2.0 Tonne capacity) should be provided essentially with the system to maintain the ambient working temperature for the biosensor system 19. Warranty on the machine and its essential accessories for a period of at least two years from the date of installation must be provided and AMC charges for additional three years must be given separately 20. Letter of consent for providing all software updates/additions and supplying the sensor chips in future for a period of at least five years from the date of installation, as per requirement of the user must be provided essentially. 21. All available sensor surfaces/chips and other accessories compatible with the system like sensor surface preparation unit etc. must also be coated as optional mentioning the price separately against each item 13 Sl. No. 7. Bioanalyzer (01 Unit) 1. Micro-chip based automated system for DNA, RNA, protein and preferably cell analysis 2. The system should be capable of measuring size, integrity and concentration of nucleic acids (DNA and RNA) and proteins 3. The system must provide an automated process of sample delivery, sample injection, electrophoretic separation, staining and de-staining, data analysis and display 4. Simultaneous analysis of multiple samples 5. Highest assay accuracy and precision 6. Capacity to analyze low sample volumes (1 to 5 µl) and multiple sample loading 7. Temperature control during electrophoresis 8. Fluorescence detection 9. Separate kit for bio-molecules. With all required accessories and DNA/RNA/protein/ cell analysis kits. The system should include kits for analysis of at least 100 samples. 10. Works across a wide range of genetic applications 11. Software for one point instrument control, operation and data analysis 12. Various data display option and output 13. Power 220-240V 14. With software and computer system. To be supplied with branded desktop computer with latest configuration along with one color laser printer 15. To be essentially supplied with one branded online UPS with 30 minute power backup 16. The price of all consumables required for running the machine should also be quoted separately 17. Machine should be covered with at least two year warranty 14 Sl. No. 8 Cell culture hood (Biosafety level 2.0) Technical specifications • • • • • • • • • • • • • • • • • • • • Biosafety cabinet should be class II, type A2 Nominal Size should be at least 1.2 meters ( 4 Feet) Internal and external dimension (WxDxH in mm) should be larger than 1200x600x650 mm and 1375x800x1550, respectively Efficiency should be ≥ 99.99% inhibiting particles ≥0.3 µm to provide around 70% down flow air and 30% exhaust Average Airflow Velocity Inflow and down flow should be at least 0.50 m/s (>100 fpm) and 0.35 m/s (70 fpm), respectively. Should be vibration free & low noise working station (<65dB). Should be Fully compliant tested and certified U.S. Standard NSF 49 and European standard EN 12469, UL listed, SEFA member Cabinet should have Fluorescent Lamp ,Germicidal UV lamp with automatic control Cabinet Construction should be of corrosion proof stain resistant stainless steel interior with antimicrobial powder (for example epoxy) coated body offering product, operator and environment protection Should have easy-to-use microprocessor controlled system with alarm. Bio safety Cabinet should have Ergonomic pitched profile and can be decontaminated with common antimicrobials. Installation, testing, validation, demonstration/technical presentation should be conducted. Thick laminated safety glass sash door counter balance should be there. Bio safety Cabinet must have internally electrical socket outlet, universal service fittings for gas, water, air and vacuum. Differential pressure gauge must be there. Should Workable on power supply of 220-240V, 50/60Hz. It should be provided with support stand, castor wheels, fixed foot or foot masters. Should Furnish all the documents like NSF 49 certificate, EN 12469 certificate etc. in the form of hard copy. Technical features described above should be supported by original illustrated catalog, with list of users (with all contact detail), and authentication of vendor’s business record in the form of Income Tax return for, at least, of last 3 years Following Accessories that must be supplied with the equipment: I) Filtration system with Vacuum pump o o o A compatible vacuum pump fitted to an additional filtration unit from any reputed make. This is to be used for aspiration of spent media and also filtering the fresh media. Vacuum pump should have the following specification Die-cast aluminum made body, with Neoprene Diaphragm, Stainless steel Seals Polyurethane Filter Elements Weight should be less than 7.0 kg with occupying minimum space (Upto 8" length x11" height x 7" width) Should be capable of operation in 210-240 V: 50-60 Hz 15 o o o o o o Should be able to generate a vacuum of at least 60 cm Hg with a pressure of 60 or more psig The vacuum should be operated by a motor of 1/8 hp or more capacity Must be able to supply and maintain Oil-free, dust-free delivery system in the laboratory Filtration unit with all accessories like funnel body, stem, Lid support screen and vent plugs should be made up of internally standard materials for using in Biosafety cabinet to filter the media. Should be provided with variable sizes of funnels (150500 ml) with effective corresponding diameters for each funnel, Stem should fit to into standard one-hole stopper, entire system should be able to withstand multiple autoclaves, should be able to hold 47mm or better filter sizes. Vacuum Pump and Filtration Unit should be supplied with a compatible Vacuum Moisture Trap container that could trap aspirated /used media or spent tissue culture liquid materials The entire system should be compatible with the Bio safety cabinet. II) This equipment should be accompanied with one window AC (1.5 tons) for providing controlled environment to ensure proper functioning of the equipment. 16 9. Mini Thermal Cycler Specification • • • • • • • • Capacity-can accommodate 25-48, 0.2ml tubes Ramping speed: 2.5°C/Sec or above Temperature range: 0-99°C Heated lid Alphanumeric representation of protocols Wide range of user-programmable preferences Display: LCD display 1KV online UPS 2 years warranty 10. Analytical Balance Specification • • • • • • • • • Maximum Capacity -220g Readability-0.01mg Repeatability-0.015 to 0.03mg Linearity -0.1mg Taring range -0 to 220g Sensitivity offset -2x10-6 Rnt Setting Time – about 2s Sensitivity Accuracy-2.0x10^-6.Rnt Sensitivity Stability -1x10^-6/a.Rnt 17 S. L. No. 11 SPECIFICATIONS INCUBATOR SHAKER With Refrigeration A. General: Bench top microprocessor controlled Refrigerated Incubator Shaker B. Temperature Range: (i) Temperature control range from 7°C above ambient to 60°C, control accuracy ±0.1°C. (ii) Should be able to maintain 37°C during summer months when ambient temperature is 40-45°C C. Shaking features: i. Electronically controlled shaking mechanism provides quiet operation. Brushless motor is preferable to avoid costly maintenance and noise-less ii. Type of shaking: Orbital. But having any additional feature is advantageous. iii. Speed: 30 – 300 rpm with control accuracy of ±2 rpm iv. It should have acceleration/deceleration circuit prevent sudden starts/stops to eliminate flask closure wetting which would otherwise interfere with oxygen transfer. v. It should have timer of up to 99.9 Hrs. D. Display: (i) Large easy to read display screen clearly indicates speed, running time, alarm condition and temperature character (ii) Display is clearly visible from all angles across the room, and in the dark. E. Safety Features: i. It should have audible and visible alarms alert for over-temperature, over-load, over-speed and leakage ii.There should be over temperature protection iii. Preferably there should be a option to mute the audible alarms iv. The system should have RS-232 port for Data Logging. v. The system should have facility to calibrate the temperature and speed via keypad. vi. Electronic and mechanical components should be enclosed, protected from accidental spills. vii. It should have automatic restart after power interruption with non-volatile memory. viii. It should have open lid cut-out switch stop agitation when chamber is open. F. Essential Accessories: (i) It should come with Universal Platform having capacity to holds assortment of various size of flask sizes upto 2-5 Ltrs and test tube racks. System should be supplied 18 with 250ml clamps (5 Nos.), 500ml clamps (05 Nos), 1000ml (02 Nos) and 2000ml (0102 Nos) (ii) It should be supplied with compatible stabilizer/ servo for smooth operation G: Essential Certification and User List: (i) It should meet CSA, UL & CE standards. Certificates, clearly indicating the make, model and name of the equipment are to be enclosed. (ii) The quote should have the List of the user of the quoted equipment/model (with valid phone number, email ID) of H. Warranty: (i) Should have Two year standard warranty from the date of installation (ii) Should quote for the price of CMC/AMC per year after the expiry of standard warranty 19 Sl. NO. 12 SPECIFICATIONS INCUBATOR SHAKER A. General: Bench top microprocessor controlled Refrigerated Incubator Shaker B. Temperature Range: (i) Temperature control range from 7°C above ambient to 60°C, control accuracy ±0.1°C. (ii) Should be able to maintain 37°C during summer months when ambient temperature is 40-45°C C. Shaking features: vi. Electronically controlled shaking mechanism provides quiet operation. Brushless motor is preferable to avoid costly maintenance and noise-less vii. Type of shaking: Orbital. But having any additional feature is advantageous. viii. Speed: 30 – 300 rpm with control accuracy of ±2 rpm ix. It should have acceleration/deceleration circuit prevent sudden starts/stops to eliminate flask closure wetting which would otherwise interfere with oxygen transfer. x. It should have timer of up to 99.9 Hrs. D. Display: (i) Large easy to read display screen clearly indicates speed, running time, alarm condition and temperature character (ii) Display is clearly visible from all angles across the room, and in the dark. E. Safety Features: ix. It should have audible and visible alarms alert for over-temperature, over-load, over-speed and leakage x.There should be over temperature protection xi. Preferably there should be a option to mute the audible alarms xii. The system should have RS-232 port for Data Logging. xiii. The system should have facility to calibrate the temperature and speed via keypad. xiv. Electronic and mechanical components should be enclosed, protected from accidental spills. xv. It should have automatic restart after power interruption with non-volatile memory. xvi. It should have open lid cut-out switch stop agitation when chamber is open. 20 F. Essential Accessories: (i) It should come with Universal Platform having capacity to holds assortment of various size of flask sizes upto 2-5 Ltrs and test tube racks. System should be supplied with 250ml clamps (5 Nos.), 500ml clamps (05 Nos), 1000ml (02 Nos) and 2000ml (0102 Nos) (ii) It should be supplied with compatible stabilizer/ servo for smooth operation G: Essential Certification and User List: (i) It should meet CSA, UL & CE standards. Certificates, clearly indicating the make, model and name of the equipment are to be enclosed. (ii) The quote should have the List of the user of the quoted equipment/model (with valid phone number, email ID) of H. Warranty: (i) Should have Two year standard warranty from the date of installation (ii) Should quote for the price of CMC/AMC per year after the expiry of standard warranty 21 13. Liquid nitrogen Cell storage system Technical specifications: • • • • • • • • • • In system Storage type should be manual fill in rack and box. Capacity should be at least 60 liters with more than 8 inch neck diameter. Static holding time should be more than 80 days and static evaporation time must be less than 0.9 liter/day System should contain at least four racks for holding of 2 ml vial boxes, each box should provide place for 100 vials. System capacity for storage should be 2000 or more vials each of 1.2/2.0 ml capacity. System exterior should be not more than: - 22.0 x 27.0 inch (Diameter x Height) System should preferably be fitted with liquid nitrogen level monitor, suitable for connecting to a power supply of 240V. System should have 3 year warranty from the date of installation. System should come wheeled accessory cart for easy movement of the storage system Technical features described above should be supported by original illustrated catalog, with list of users (with all contact detail), and authentication of vendor’s business record in the form of Income Tax return for, at least, of last 3 years. 22 S.L NO. 14 Specification for Electroporator 1. Electroporator system with the capability to electroporate eukaryotic cells, prokaryotic cells (bacteria), yeasts and other micro-organisms. 2. Instrument should have features for adjusting voltage & pulse length 3. Pulse voltage rang:10-3,000V 4. Should include all accessories for electroporation. 5. It should also have required software, pre- set optimized programmes and other accessories. 23 S.L. NO 15 SPECIFICATION DETAILS OF CO2 incubator 1. TYPE: Moderate sized (table top with smaller foot-print), microprocessor controlled multi-sided heating with provision for accurate and precise controls for CO2 and relative humidity. Suitable for Mammalian Cell Culture 2. CAPACITY: App. 120-200 litre or more 3. TEMPERATURE RANGE AND UNIFORMITY: ± 2-3 °C ambient temp to 50°C or better; With control of + 0.1°C, Uniformity of + 0.1°C. Should have the external or internal features/ arrangement so that the machine can maintain the 37°C during summer months when ambient temperature is 40-45 °C 4. REGULATION RANGE CO2 range: 0.2 to 20%, with control range + 0.1%; uniformity ± 0.1%; Accurate CO2 sensing using IR or better sensing technology; Auto- start Function 5. HEATING SYSTEM: Muti-sided heating Fan-less convection Should have maximum thermal protection and quick (≤ 4-6 minutes) recovery from swings in ambient temp and door opening. 6. CO2 SENSING Infrared (IR) or better 7. INLET FILTER FOR CO2 HEPA filter: <0.3 micron at 99.999% for optimum filtering, easy access, and simple replacement (no tools needed). 8. DECONTAMINATION: Auto sterilization facility at high temperature of 121°C or using better/improved method 9. 10. DISPLAY & SAFETY :Large display to show set and achieved set parameters of Temp, CO2, RH, Clock, graphical option and preferably with history programmable alarms, diagnostics INTERIOR: 1. Sealed, joint less, Fan-less and polished stainless steel interior with 100% coved corners is easy to clean, saving time and reducing contamination risk. 2. Condensation free interior to avoid growth of contaminants during culture 11. SHELVES: Removable and adjustable stainless shelves 24 12 ESSENTIAL ACCESSORIES: 1. Should be supplied with standard accessories, and all the prerequisites for installation and functioning of the equipment like door lock, air conditioning, calibration kit, filters, Cylinder auto-changer, and alarm system etc 2. Should be supplied with compatible branded online UPS/ Stabilizer with minimum half an hour backup having INTERNAL dry SMF battery which should have warranty/AMC as same as for the incubator 3. Should be supplied with one cylinders and one regulator and all other accessories 13 OTHER MANDATORY FEATURES: 1. MUST have International ISO, CE, UL certified (certificate must be enclosed). 2. Firm MUST provide a compliance statement vis-à-vis specifications in a “tabular form in the prescribed format” clearly stating the compliance and giving justification, if any supported by technical literature 3. A list of users of the model quoted in the present Tender only along with valid email ID and contact numbers and a certificate that they have no objection in case the users are contacted 14 ON-SITE WARRANTY: Two years standard comprehensive warranty and three years AMC with detailed terms & conditions so that equipment is covered at least for the entire project period. Rates may be shown separately. 25 Sl. No. 16 • • • • • • • • • • • • • • • Specifications for Fluorescent Microscope with camera attachment Should have a basic stand with LED illumination, four nosepiece and HCS (harmonic component system) Optics, with 20 mm field of view Should have automatic adjustment of illumination to the contrast methods, auto-off function, LED with service life of 50,000 hours, constant colour temperature Should have bright field, phase contrast, integrated modulation contrast upgradeable to fluorescence Should have trinocular tube with beam splitting position Vis/Phot: 100/0,50/50 & 0/100 Should have four position nosepiece Should have Objective 4×, 10×, 20× FWD 3.7mm & 40× FWD 2.0mm & 10× eyepiece Should have condenser suitable for all the above techniques Should have large stage 248×212×20 mm. (optional-XY adjustment 83×127 mm & holders Should have fluorescence filters for Rhodamine, FITC and DAPI with 100 Watt Hg lamp for fluorescence Should have fluorescence slider with three positions for filter cubes Illumination column should be attachable to both rear & front of microscope stand Should be provided with high resolution cooled digital camera (same make as that of the microscope) with 5MP, 2/3”CCD sensor, pixel size 3.4µm × 3.4µm, exposure time 1 msec-600 sec, fast live image (24 Frame/sec) with Firewire cable, interface to produce 1:1 image on PC Should be provided with a branded PC with latest configuration Should be provided with UPS one kV offline with 30 minutes back up Should be provided with a branded colour printer for making photographic record 26 Sl. No. 17 Specifications for Bio Safety Cabinet Class II, Type B2 (total exhaust) Biosafety cabinet of class II, Type B2 (Total Exhaust) with vertical laminar airflow complying to EN12469:2000 with microprocessor based monitoring system. Automatic control of all functions and all safety alarm systems with double centrifugal fan to provide complete operator, product and environmental protection. Suitable for handling pathogens namely Mycrobacterium paratuberculosis. Internal dimension (WxDxH) should be approximately 850X 700X 550 mm or more The cabinet should be fitted with 1 automatic safety service connection for gas, 1 for vacuum and 1 electrical socket Dual, long life ULPA/HEPA (H-14 grade, according to EN1822) filters for supply and exhaust airflow. The cabinet should consist of ducting facility on the top of the cabinet for direct ducting to facility exhaust system. The Cabinet should be supplied with a UV sterilizing lamp Frameless, shatter-proof sash with automatic UV shut-off on sash opening Machine should have low noise level ( below < 53 dB) The cabinet should have user friendly practical keyboard and display to inform laminar airflow and frontal air barrier velocity , residual lifetime of HEPA filter , UV lamp , total number of hours of operation, saturation level of HEPA filter, inside and outside Temperature Audio and visual alarms required for power failure, out of range or incorrect laminar airflow velocity and frontal air barrier velocity, end of life-cycle of UV lamp , fan-motor malfunction ,saturation of HEPA filters , un-correct position of front sash-window, blockage in the exhaust duct. Interior work area of a single piece of stainless steel and single piece HIGH GRADE stainless steel. Work surface should be consisted of sections easily removable for carrying out routine cleaning and/or require autoclaving sterilization procedures if so desired. Cabinet should be preferably coated with anti bacterial treatment/ solution to prevent microbial contamination. Cabinet should also be supplied with modular stand with castors. Air flow velocity should be at least 90 fpm; efficiency should be > 99.99% at 0.1 micron to 0.3 micron to provide 100% exhaust. Safety device: (i) Dual-wall construction surrounds the work zone with negative pressure plenums for maximum safety. (ii) Fail-safe system ensures that in case of exhaust failure Must meet American (NSF/ANSI) or European standard EN 12469 (type tested) or both. Must submit a copy of EN 12469 or NSF/ANSI certification along with the quote (it is mandatory and without the valid certificate the quote will be considered as non-responsive). 27 Sl.No18. Water purification system Technical Specifications: • The system should be capable to handle potable tap water conditions. • It should support the following applications: Ist stage system for laboratory applications The unit should have the following purification stages: Activated carbon for the removal of free chlorine & organics. Reverse osmosis stage for removal of ionic & inorganic impurities. Conductivity cell should be present before RO stage to measure the RO feed conductivity & system should be capable of measuring percentage of rejection for correct monitoring. Advanced self regenerating electro-deionisation module resistant to scaling should be used. Water quality: Flow rate should be at least 3 L/hr, Ions organics removal upto 99%, TOC < 50 ppb or less. • Additionally, for Molecular Biology applications, the Unit should have a IInd stage system to provide ultrapure water with following features: Resistivity: 18.2 MOhm.cm, Conductivity: 0.055uS/cm, Bacteria: < 0.1cfu/ml, Flow rate: upto 1 liters/min, TOC: < 10 ppb, Endotoxin< 0.001 Eu/ml. Automatic recirculation feature. • It should have a Reservoir Tank of at least 50 liter capacity • Technical features described above should be supported by original illustrated catalog, with list of users (with all contact detail), and authentication of vendor’s business record in the form of Income Tax return for, at least, of last 3 years Following Accessories that must be supplied with the equipment: 1) The equipment should be accompanied with a 2 KV online UPS to back the power supply for at least 30 minutes. 2) 1 portable water purifier for other supporting and general lab work and consumable for media filtration filters. 28 Sl. No. 19 Geldoc for both transillumination and chemiluminescence imaging system with photographic capabilities attached to a CCD camera and computer with software The equipment is expected to be used for applications such as, 2-D protein gels including multiplexing, Stain Free gel and blot imaging, 1-D nucleic acid and protein gels, Proteomics and Genomics studies, Western blots – Chemiluminescence and 3-color multiplex fluorescence and also for Northern and Southern blots studies Technical Specifications required for this equipment is as follow: • • • • • • • • • • • • • • System should be with true 16 bit CCD (0-65,535 levels), with > 4 Megapixel image resolution Cooled CCD with cooling range of -30°C absolute temperature using peltier cooling system (Not air cooled) System with dynamic flat fielding technology for getting accurate Quantitation with CV < 5 %. Multi image color target with multi-channel image display for study of more than 2 dyes at a time for differential expression. Automatic sequential imaging of multi labeled samples. Instrument should have a large transillumination area which is capable of taking sample of minimum 28x36 cm and should have a minimum imaging area of 26 x 35 cm. Must support capability of stain free gel imaging to avoid time consuming gel/sample staining procedure. Instrument should have auto focus technology that is system should automatically take the best focus depending on any zoom level without the movement of sample platform. The sample platform should be fixed at one position to ensure minimal mechanical movement. Pixel size should be 6.45 µm x 6.45 µm or bigger. Instrument should have a minimum 6 position filter wheel for capturing images of various dyes effectively. The filter wheel should be motorized and automatic without the requirement of manually changing or moving the filter wheel. Dynamic range should be >4 orders of magnitude for good quantification. Pixel density should be at least 65,535. System should have at least 8 different modes of illumination like, Trans-UV, epiwhite, no illumination for chemiluminescence, trans-white, epi-red, epi-green, epiblue, and optional trans-blue. All LEDs should be integrated with system for proper imaging of multi target with reproducibility. System should offer various binning options with minimum of 2x2, 3x3 and 4x4 for customized sensitivity. Data Processing Unit: desktop PC (with a latest processor at least Core I3, TFT monitor at least 19” size, 3 GB RAM and 320 GB hard disk, and preloaded with latest operating system software) 29 Software for Image acquisition and Analysis: • • • • • • • • • • • • • • • Should support automatic generation of customizable reports with data organized as desired, including, Lane and band identification, molecular weight or base pair evaluation. Band sizing and quantification should be based on a reference band or quantity standards. Snapshot tool to copy images, lane profiles, and graphs Complete flexibility with automatic and manual detection of lanes and bands, molecular weight determination, purity assessment, and relative and absolute quantification using several algorithms. Should have tri-plex image capture capability and analysis features. Easy copy/paste functionality, crop, zoom, 3D viewer, and colors. Should provide functionality to easily produce image at user-defined dpi and dimension to support publication Should have multiple image optimization tools like unlimited undo and redo functions to easily correct for any missteps Original software should be provided free of cost, and with up-gradation without any extra cost in future. No requirement for license registration software should be Mac and PC compatible 16-bit and 8-bit tiff images with a one-click export option Software should automatically select the appropriate filters, light sources, and camera settings for all applications. It should also ensure that image optimization is specific to a selected gel or blot application or any specific portfolio of detection methods. Applications should support: Chemiluminescent, colorimetric and multiplex fluorescent western blots, nucleic acid and protein detection via colorimetric and fluorescent stains. Software should have automated multi-channel image view and color-coded analysis. Software should have Signal Accumulation Mode (SAM) for easy optimization of exposure time for chemiluminescent detection. Software should have ability for unlimited installations of the software for multiple users with full functionality. Software should offer live update of results with any change of analysis parameters. Accessories: I. Horizontal Mini/Midi gel electrophoresis system • • • • • Complete system with all accessories for running Agarose gels. Should be able to cast gels of size 7 x 7 cm and 7 x 10cm. Should have the sample throughput of 8 – 30 . UV Transparent trays with an integrated fluorescent rulers. Gel caster with Leak proof casting facility, without using tapes. Also option for casting dams should be there. 30 Should come with 8, 12, and 15 wells fixed height comb Should have replaceable electrodes for longer life of equipment. Suitable Power pack to support the system • • • II. Vertical Electrophoresis SDS Page System with Blotting apparatus: Mini Electrophoresis system with 7x8 cm gel size with capacity of 2-4 gels simultaneously ( Preferably 4 gels) Should have integrated spacers with glass plates for ease of casting. Should be a modular system to support blotting and electro-elutions also in the same system. Should come all standard accessories like 10 well, 0.75mm comb, 5 sets of glass plates –both spacer plates, and short plates. Should come with sample loading guide. • • • • • Blotting Apparatus should have the following technical features • • • • • • Gel size 10 x 7.5 cm or any standard specification Buffer requirements:- >400 ml Capacity:- 2 mini protean hand cast gels, 2 mini precast gels Placement of electrode should be 4 cm or more apart for strong electric fields and efficient protein transfer. Color coded cassettes and electrodes to ensure proper orientation Blue cooling unit that absorbs heat generated during rapid transfer Power Supply: for Running and transfer of large gels • • • • • • Programmable power supply should be capable to operate four electrophoresis units simultaneously for four identical runs with graphic 3 digit LED display. The output range should be up to 10- 300 V , 4- 400 mA in I m A steps, 75 W Constant voltage, or constant current Horizontal Electrophoresis system Timer – 1- 999 min No load detection, sudden load change detection, short circuit detection, over voltage protection. Technical features described above should be supported by original illustrated catalog, with list of users (with all contact detail), and authentication of vendor’s business record in the form of Income Tax return for, at least, of last 3 years 31 Following Accessories that must be supplied with the equipment: 1) To support controlled environment one Split AC (1.5 ton) from any reputed make should be provided 2) An additional data back-up system in the format of a laptop PC from a reputed make with a latest processor (at least Core I3) and 3 GB RAM and 320 GB hard disk, screen size more than 15” and preloaded with latest operating system software and 4 lab work stools should be given free of cost. 32 S.L. NO. 20 Automated Blood Gas analyzer for Veterinary Use Specifications 1. 2. 3. 4. 5. Features to measure pH, pCO2, pO2, measure Oximetery parameters, Electrolytes Test Samples: Blood, Plasma, Serum in small amount Analyzer controller with soft ware for analysis of multiple parameters LCD screen with Printer Facility for regular quality assessment of the instrument should be provided by the company 6. Warranty : At least 3 years Provide reagents with the analyzer to run the machine satisfactorily for two years (@1000 samples/year). 33 Sl. No. 21 Blood Flow Meter (To measure quantitative skin blood flow rate in the micro circulatory beds of skin and other tissues of livestock under heat stress, suitable under Indian conditions) Specification: Mode: Continuous laser Doppler flowmetry Primary Measurement: Microvascular blood perfusion (Relative RBC Flux) LDF Units: Relative units (0–7500 Blood Perfusion Units, corresponding to 0–7.5 V output) BSC Units: Relative units (0–100% corresponding to 0–7.5 V output) Laser Type: Semiconductor laser diode (temperature stabilized) Laser Wavelength: 830 nm +10 nm Laser Power: < 0.5 mW from probe Laser Class: Class 1 (as per 21 CFR 1040-10 and 1040-11) Doppler Signal Bandwidth: 16 kHz Linearity Range: up to 0.35% moving scatterers by volume. Flow Response Time: < 0.2 s Reading Stability: 5% (measured with standard motility solution) Zeroing: Automatic, controlled Flux Calibration: Factory or user calibrated using a motility standard (concentration of latex spheres undergoing Brownian Motion) Power Supply Voltage Range: 100–240 V AC (50–60 Hz), 0.8–0.4 A Output Power: +5 V, }12 V DC Fusing: Internal, 2 A 240 V resettable Physical Configuration Power Requirements: 240 V or 120 V (internally set) @ l5 W Operating Temperature: 0–50 °C Operating Humidity: 0–100% (non-condensing 34 Sl. No. 22 Specifications for Real Time PCR Real time PCR with block of 96 x 0.2 ml tubes or plate to run typical 0.2ml tubes, strips, and plates The base thermal cycler should be able to be used for standard PCR Gradient capacity in Real-time Should have a maximum ramping speed of 5 OC per sec Should have the capacity to detect FIVE different fluorescent reporters in the same tube. Should be capable of detecting Cy5, FAM/Sybr Green, VIC/JOE, TAMRA/Cy3, Texas Red, Quasar705 Should have six excitation and six emission channels each filter corresponding to one dye that ensures smooth differentiation of even dyes having high degree of spectral overlap Should have peltier cooling & heating Should have one channel dedicated for FRET experiments Should have excitation –emission range of 450- 730nm No internal reference dye should be required. True 5 Color Multiplexing with use of 5 different flourophores without the need of addition of any internal reference dye Should have LED excitation source with Photodiode detector Should have dynamic range of 10 orders Should be an open system capable of running various chemistries like TaqMan, Molecular Beacon, SYBR green Should have temperature range of 0– 100 OC with accuracy of ±0.2 OC and uniformity of ±0.4 OC within 10 sec of arrival at 90 OC Should be able to run with a minimum sample volume of 1µl Should be able to detect ≤ 10 fmol of fluorescein Should have multiple scan modes with a FAST scan option for reading all wells in 3 seconds Should have facility for automatic allelic discrimination by end point fluorescence or threshold cycle Gene expression analysis by relative quantity (∆Ct) or normalized expression (∆∆Ct) Should have end point analysis for upto 5 fluorophores Should have mode for Melt curve analysis Comparison of upto 5000 Ct values from different data files should be possible Should be supplied with 2KVA online UPS to provide uninterrupted power supply, branded PC/laptop for analysis and cooling system to maintained ambient temperature Analysis of multiple experiments from a single plate should be possible Warranty of 1 year and CMC free of cost for the second year