Installation/Owner’s Manual
29-5100
Clorox Healthcare™ Optimum-UV™ System
Powered by UVDI Technology
Patent No. D684671
29-6065-15 Rev A
Manual Contents
29-5100
Safety Information ……………………………………………………………………………..3
Product Overview ……………………………………………………………………………...5
Room Treatment Protocols .…………………………………...…………………………….11
Device System Operation …………………………………………………………………....13
Device Remote Operation …………………………………………………………………...15
Storage and Maintenance ……………………………………………………………………17
Replacement Parts ……………………………………………………………………………20
Appendix A: Efficacy Data ……………………………………………………………………21
Appendix B: Touch-Screens …………………………………………………………………22
Equipment Warranty ………………………………………………………………………….34
29-6065-15 Rev A
Safety Information
29-5100
It is mandatory to read and understand this manual before operating this device. This is
for your safety and safe operation of the device.
3
29-6065-15 Rev A
Safety Information
29-5100
WARNING
AVERTISSEMENT
RISK OF ELECTRIC SHOCK! OPERATING
INSTRUCTIONS ARE FOR USE BY
QUALIFIED PERSONNEL ONLY. TO REDUCE
THE RISK OF ELECTRIC SHOCK, DO NOT
PERFORM ANY SERVICING OTHER THAN
THAT CONTAINED IN THE OPERATING
INSTRUCTIONS.
RISQUE DE CHOC ÉLECTRIQUE! LES
INSTRUCTIONS D'INSTALLATION NE
DOIVENT ÊTRE UTILISÉS QUE PAR DU
PERSONNEL QUALIFIÉ. POUR RÉDUIRE LE
RISQUE DE CHOC ÉLECTRIQUE, NE JAMAIS
EFFECTUER DE RÉPARATIONS AUTRES QUE
CELLES CONTENUES DANS LE MANUEL
D'UTILISATION.
WARNING
UV LIGHT AND LAMP BURN HAZARD!
HARMFUL TO BARE SKIN AND EYES. CAN
CAUSE BURNS. CAN CAUSE TEMPORARY
LOSS OF VISION. NEVER LOOK AT THE
LAMPS WHILE ILLUMINATED.
CAUTION
TO REDUCE THE RISK OF ELECTRICAL
SHOCK, THIS EQUIPMENT HAS A
GROUNDING-TYPE PLUG THAT HAS A
THIRD (GROUNDING) PIN. THIS PLUG WILL
ONLY FIT INTO A GROUNDING TYPE POWER
OUTLET. IF THE PLUG DOES NOT FIT INTO
THE OUTLET, CONTACT QUALIFIED
PERSONNEL TO INSTALL THE PROPER
OUTLET. DO NOT ALTER THE PLUG IN ANY
WAY.
CAUTION
THIS DEVICE CONTAINS MERCURY IN THE
SEALED ULTRAVIOLET BULB(S). DISPOSE
OF BULBS IN ACCORDANCE WITH STATE
AND LOCAL LAWS.
CAUTION
USE CAUTION WHEN MOVING THE DEVICE
OVER UNEVEN OR SLOPED GROUND.
AVERTISSEMENT
DANGER, LA LUMIÈRE UV ET LAMPE BRÛLE!
NOCIF POUR LA PEAU NUE ET LES YEUX.
PEUT CAUSER DES BRÛLURES GRAVES.
PEUT ENTRAÎNER UNE PERTE TEMPORAIRE
OU PERMANENTE DE LA VISION. NE JAMAIS
REGARDER LA LAMPE LORSQUE ILLUMINÉ.
ATTENTION
POUR RÉDUIRE LE RISQUE DE CHOC
ÉLECTRIQUE, CET ÉQUIPEMENT DISPOSE
D’UN CONNECTEUR AVEC UNE TROISIÈME
BROCHE DE PRISE DE TERRE. CETTE FICHE
NE PEUT S'INSÉRER QUE DANS UNE PRISE
DE COURANT AYANT UNE PRISE DE TERRE.
SI LA FICHE NE RENTRE PAS DANS LA PRISE,
CONTACTEZ UN PERSONNEL QUALIFIÉ POUR
INSTALLER UNE PRISE APPROPRIÉE. NE
MODIFIEZ PAS LA FICHE EN AUCUNE
CIRCONSTANCE.
ATTENTION
CET APPAREIL CONTIENT DU MERCURE
DANS L'AMPOULE À ULTRAVIOLETS SCELLÉE
(S). DISPOSEZ DES AMPOULES EN
CONFORMITÉ AVEC LES LOIS NATIONALES
ET LOCALES.
ATTENTION
SOYEZ PRUDENT LORSQUE VOUS DÉPLACEZ
L'APPAREIL SUR UN TERRAIN INÉGAL OU EN
PENTE.
4
29-6065-15 Rev A
Product Overview
29-5100
The Clorox Healthcare™ Optimum-UV™ System is designed to treat surfaces
contaminated with pathogens. Typical applications include hospital rooms, operating
suites, emergency rooms, patient waiting rooms, isolation wards, public restrooms,
nurse’s stations, staff break rooms, locker rooms, clean rooms, laboratory rooms, and
food preparation areas.
Please note that the Clorox Healthcare™ Optimum-UV™ System should be used as a
supplement – not as a replacement – to standard terminal cleaning protocols. Please
follow your facility’s manual room cleaning and disinfection protocols before using a UV
device, and consult CDC, SHEA and APIC guidelines for additional steps to take as a
part of a bundled approach.
ATTENTION
PROPER USE OF THIS DEVICE IS CRITICAL IN YIELDING DESIRED
TREATMENT RESULTS. THE UV-C TREATMENT RESULTS ACHIEVED
ARE SUBJECT TO VARIATION IN APPLICATION CONDITIONS.
CLOROX HEALTHCARE RECOMMENDS CULTURE TESTS TO
CONFIRM DESIRED OUTCOMES.
The device emits UV-C energy in a full circle, or 360 degrees, to deliver a programmed
dose of UV light for treating a room’s target surfaces. Using the Clorox Healthcare™
Optimum-UV™ System does not require special room preparation – simply expose
target surfaces to the device prior to starting the room treatment cycle.
It should be noted that because the Clorox Healthcare™ Optimum-UV™ System utilizes
“line-of-sight” UV technology, target surface(s) must be directly exposed to the device’s
UV energy. UV energy will not penetrate materials such as paper, soil, cloth, plastic,
ordinary glass or spills. After completion of the treatment cycle, the room may be
immediately occupied. The Clorox Healthcare™ Optimum-UV™ System does not
generate any harmful byproducts, nor does it leave any residual substances.
Ultraviolet light kills microorganisms via short-wavelength ultraviolet radiation (UV-C).
The Clorox Healthcare™ Optimum-UV™ System produces UV-C at a wavelength of
approximately 254 nm, inactivating the microorganisms’ DNA.
5
29-6065-15 Rev A
Product Overview
29-5100
Figure 1: Clorox Healthcare™ Optimum-UV™ System Line Drawing
6
29-6065-15 Rev A
Product Overview
29-5100
Figure 2: Touch-Screen Operating System
Figure 3: Power Switch Location
Figure 4: Safety Motion Sensors
Figure 5: Locking Wheels
7
29-6065-15 Rev A
Product Overview
29-5100
Specifications
Specification
Description
Lamp Type
Four (4) maximum output UV lamps; 254-nm wavelength
Treatment area
Single room
Ballast (4)
Electronic, solid state
Lamp Length
62 inches
Footprint
36 inch-diameter circular base, 76 inches tall
Lamp life
4,000 hours accumulated “ON” time.
Power Source
120 VAC, 60Hz, 12.5 Amp
Power Connector
15 foot long hospital-grade cord
Delay timer to turn unit ON
30 seconds delay
Power Interruption Memory
Device will resume operation after interruption and
remember condition before power interruption or before
shutdown due to intentional or unintentional interruption.
Motion sensors (safety)
360 degrees
Operating Environment
Indoor 50°F to 80°F, 50% RH @70F, non-condensing
8
29-6065-15 Rev A
Product Overview
29-5100
Device Assembly
Step 1: Install Maximum Output UV-C Lamps
Lamp Information
The Clorox Healthcare™ Optimum-UV™ System is equipped with four (4) maximum
output UV-C lamps. Each lamp is completely encapsulated in a protective polymer
sleeve. The protective polymer sleeve provides safety in the event of lamp breakage by
fully containing the content inside. The protective polymer sleeve is highly transmissible
to UV-C light and does not affect performance of the device.
UV lamps supplied with the Clorox Healthcare™ Optimum-UV™ System are factory
tested to ensure proper UV-C intensity. Each lamp has a serialized label on the lamp
end cap with pins, which verifies the testing process has been completed.
Lamp Orientation
Each lamp has a brown label on the black end caps. When installing the lamp, please
ensure that the brown label faces in towards the mast. Correct orientation is shown in
picture below.
CORRECT ORIENTATION OF LAMP
INCORRECT ORIENTATION OF LAMP
Note: Improper installation of the lamps may affect performance of the Clorox
Healthcare™ Optimum-UV™ System.
9
29-6065-15 Rev A
Product Overview
29-5100
Insert the lamp-end cap into the lamp catch. Refer to Figure 6, below.
Figure 6: Lamp Catch
Step 2: Insert Lamp Pins
Insert lamp pins into white connectors in base stand. Refer to Figure 7, below. Note
that the lamp-connector pins configuration is rectangular and must be oriented
correctly to the receptacle before inserting.
Figure 7: Insert Lamp Pins
10
29-6065-15 Rev A
Room Treatment Protocols
29-5100
The Clorox Healthcare™ Optimum-UV™ System should be used as a supplement – not
as a replacement – to standard cleaning and disinfection protocols. Please follow your
facility’s manual room cleaning and disinfection protocols before using a UV device, and
consult CDC, SHEA and APIC guidelines for additional steps to take as a part of a
bundled approach.
Once you have completed your standard cleaning and disinfection protocols,
please deploy the Clorox Healthcare™ Optimum-UV™ System.
Step 1: Post Warning Sign and Position Clorox Healthcare™ OptimumUV™ System
IMPORTANT
IF OPERATOR DOES NOT EXIT ROOM WITHIN 30 SECONDS, UNIT
WILL FAIL TO BEGIN TREATMENT CYCLE DUE TO MOTION
DETECTOR SHUTDOWN. PUSH START BUTTON AGAIN TO REINITIATE THE CYCLE.
Affix the warning sign on the door to indicate “DO NOT ENTER” during the room
treatment cycle. Position the Clorox Healthcare™ Optimum-UV™ System in an
appropriate location to ensure line-of-sight exposure to the room’s target surfaces.
Ensure that all surfaces requiring treatment are exposed to the unit. Remove papers
and/or other objects that may be shadowing target surfaces. Note that if the room is
equipped with furniture, equipment and/or fixtures, it may be necessary to re-locate the
Clorox Healthcare™ Optimum-UV™ System and run additional cycles to ensure proper
treatment of all target surfaces in the room.
Step 2: Secure Doors
Secure doors leading into the room so that no one can enter and interrupt the treatment
cycle.
CAUTION
HIGH TEMPERATURE! DO NOT TOUCH LAMP OR MAST
Step 3: Start Cycle and Exit Room
Plug the electrical cord into the nearest electrical outlet and power up the Clorox
Healthcare™ Optimum-UV™ System by pressing the power switch located on the left
end of the device handle. The touch-screen display will activate and may be used to
configure and execute all functions. Please refer to Appendix B: Clorox Healthcare™
Optimum-UV™ System Touch-Screens for additional information.
Note: When power is turned on, WELCOME screen displays and device will start a 35
second warm-up period to stabilize motion detector sensors. Touch-screen functionality
begins after the 35 second warm-up period.
11
29-6065-15 Rev A
Room Treatment Protocols
29-5100
Navigate to the ENTER UV CYCLE TIME screen to enter the UV treatment cycle time
and continue to the START/STOP/RESET screen to indicate a Room number, if
desired. Press the touch-screen START button (or initiate using the remote control) to
begin UV treatment. When START is executed, a slow, pulsing, warning beep will be
heard for 25 seconds, followed by a rapid, pulsing beep for the final 5 seconds before
the UV treatment cycle commences. The warning beep’s volume may be adjusted via
the touch-screen display. After the 30-second delay, the Clorox Healthcare™ OptimumUV™ System unit will begin its cycle and a visible blue light will be observed within 2025 seconds.
IMPORTANT
OPERATOR MUST LEAVE THE ROOM DURING ENTIRE CYCLE TIME
TO AVOID UV EXPOSURE
Step 4: Enter Room after Treatment Cycle Completes
After the treatment cycle completes, the operator may immediately enter the room.
When a cycle completes successfully, then the CYCLE COMPLETED screen will be
displayed. The Clorox Healthcare™ Optimum-UV™ System is now ready for the next
treatment cycle and can be moved to a new location. If the cycle is interrupted due to
device sensors detecting movement in the room, then the CYCLE INTERRUPTED screen
will be displayed. The interrupted cycle may be continued to completion or re-started by
following the touch-screen prompts.
After completion of the treatment cycle, the room may be immediately occupied. The
Clorox Healthcare™ Optimum-UV™ System does not generate any harmful byproducts,
nor does it leave any residual substances. Note that a faint, “clean” smell may be
detected in the room after the cycle. UV-C breaks down the molecules associated with
odors and unpleasant smells in the air. This results in the “clean air,” smell after the
programmed cycle. The odor’s chemical composition has been documented and
analyzed, and has been found to pose no risk.
12
29-6065-15 Rev A
Device System Operation
29-5100
Step 1: Power on Device
Turn on the power by pressing the power switch located on the left end of the device
handle. The touch-screen display will activate and may be used to configure and
execute all functions. Please refer to Appendix B: Clorox Healthcare™ Optimum-UV™
System Touch-Screens for additional information. Once the touch-screen has been
activated, navigate to the MAIN MENU screen, which provides the following options:
Menu Options
Function
PROGRAM SET
UP
Sub-screens allow user to enter dose time, select room, start the
cycle and monitor treatment cycle progress.
ALARM SET UP
Enable or disable the cycle start alarm. If enabled, set volume to
high or low.
SYSTEM
INFORMATION
System performance information, including lamp life data, total hours
of use, and UV treatment cycles completed.
CLOCK SET UP
Enables clock set up date and time.
DATA LOG
Records and displays Clorox Healthcare™ Optimum-UV™ System
historical use data.
Step 2: Enter UV Cycle Time
Press the PROGRAM SET UP button and enter the UV CYCLE TIME in hours and
minutes. Note that the message REMOTE ON appears when the remote function is
activated.
Note: The time entered for Minutes field should be a number other than zero.
Step 3: Start the UV Cycle
The unit’s room selection feature is enabled by pressing the ROOM # button. Follow the
on-screen prompts to select the room number for data retention, if desired. Then, from
the START/STOP/RESET screen, push the START button to begin the UV cycle. The
Operator MUST exit the room within 30 seconds of pushing the START button. Note
that before 30 seconds has expired, the operator can push the STOP button to
terminate operation. If the operator does not exit the room, the Clorox Healthcare™
Optimum-UV™ System will fail to start due to the motion detector safety feature.
13
29-6065-15 Rev A
Device System Operation
29-5100
Step 4: Complete the UV Cycle
If the cycle has fully completed, the CYCLE COMPLETED screen will appear. If the cycle
has been interrupted, the CYCLE INTERRUPTED screen will appear. Note that if the
cycle is interrupted, the operator can restart the unit by pressing the START button, OR
reset the timer and start over. If power was interrupted, the timer conditions will return to
settings prior to the power interruption.
CAUTION
ONCE CYCLE STARTS AND THE OPERATOR IS OUTSIDE THE ROOM,
CYCLE CAN ONLY BE INTERRUPTED PRIOR TO CYCLE COMPLETION
BY USING THE REMOTE PROVIDED (SEE DEVICE REMOTE
OPERATION). DO NOT ATTEMPT TO INTERRUPT CYCLE BY
ENTERING THE ROOM AS IT MAY RESULT IN UV EXPOSURE!
14
29-6065-15 Rev A
Device Remote Operation
29-5100
Remote Control
The Clorox Healthcare™ Optimum-UV™ System UV treatment cycle may be started
and stopped remotely with the remote hand-held control. The remote control also
serves as a safety device to cut power to the unit or stop the cycle in case of
emergency.
Figure 8: Clorox Healthcare™ Optimum-UV™ System Remote Hand-held Control
Step 1: Activate Remote Control
First, apply power to the Clorox Healthcare™ Optimum-UV™ System and allow the
device to complete a 35 second warm up. Activate the remote control by pushing the
READY button. Note that the READY button must be depressed at both the top and
bottom of the button to function. When the remote is activated, REMOTE ON will flash
on/off on the Clorox Healthcare™ Optimum-UV™ System unit’s touch-screen.
Note: When power is turned on, WELCOME screen displays and device will start a 35
second warm-up period to stabilize motion detector sensors. Remote control
functionality begins after the 35 second warm-up.
15
29-6065-15 Rev A
Device Remote Operation
29-5100
Step 2: Start Operation
Start operation by pressing the buttons UV + ON simultaneously. The Clorox
Healthcare™ Optimum-UV™ System will commence the 30 second programmed delay
and then perform the programmed UV treatment cycle.
Step 3: Stop Operation
To stop operation, press UV OFF, which requires the UV + OFF buttons to be pressed
simultaneously. This will stop the UV treatment cycle and allow safe entry to the room.
To resume a UV cycle, exit the room and push buttons UV + ON simultaneously. The
UV cycle will resume and complete the remaining cycle time of the current program.
Note that if you have initiated the Start Operation process using the remote, and you
need to interrupt the treatment start cycle during the 30-second delay of the start
sequence, then you can use the UV + OFF feature to do so using the remote or
pressing STOP on the screen display.
For EMERGENCY SHUTDOWN (bypassing the internal controller), the CUT OFF button
may be pressed in the event operator determines that controller has malfunctioned and
treatment cycle must be terminated. Note that the CUT OFF button must be depressed
at the top and the bottom to function.
IMPORTANT
NORMAL CYCLE INTERRUPTION CAN BE ACHIEVED BY PRESSING
THE UV+ OFF BUTTON ON THE REMOTE. CUT-OFF BUTTON SHOULD
ONLY BE USED IF UV+ OFF BUTTON DOES NOT WORK OR ANY
OTHER EVENT CAUSING LAMPS TO NOT SHUT OFF
The Power switch must be cycled OFF and ON to reset the Clorox Healthcare™
Optimum-UV™ System after an EMERGENCY SHUTDOWN.
Note: Activation of remote control by pushing the READY button is not required for
EMERGENCY SHUTDOWN
Step 4: Cycle Again
To start a new cycle after CYCLE COMPLETION or to continue a cycle after CYCLE
INTERRUPTION using the remote, go back to Step 2 in this sequence.
16
29-6065-15 Rev A
Storage and Maintenance
29-5100
Storage
When the Clorox Healthcare™ Optimum-UV™ System is not in use, cover it before
storing. Insert the storage cover into the handle bar, and then carefully lift the cover over
the mast, while ensuring that the receiver antenna fits through the cover hole. Finally,
close the zipper.
CAUTION
LAMPS MUST BE COOLED DOWN BEFORE COVERING THE DEVICE.
COOLING TIME MAY VARY DEPENDING ON ENVIRONMENTAL
CONDITIONS. 10 TO 15 MINUTES ARE RECOMMENDED. PLACE THE
COVER BACK ON THE DEVICE AND MOVE IT TO ITS STORAGE
LOCATION OR THE NEXT AREA TO BE TREATED.
Lamp Information
The Clorox Healthcare™ Optimum-UV™ System is equipped with four (4) maximum
output UV-C lamps. Each lamp is completely encapsulated in a protective polymer
sleeve. The protective polymer sleeve provides safety in the event of lamp breakage by
fully containing the content inside. The protective polymer sleeve is highly transmissible
to UV-C and does not affect performance of the device.
Lamp Replacement
Clorox Healthcare strongly recommends that the lamps be replaced after 4,000 hours of
use. Please contact your local Clorox Healthcare representative to order replacement
lamps. If you do not have your representative’s information, please call 1-800-234-7700.
The use of non-Clorox Healthcare lamps will void the warranty.
Note: Refer to the LAMP STATUS screen to check accumulated lamp hours for each
lamp.
UV-C lamp intensity degrades as a function of time. After 4,000 hours of use, lamp
intensity is 80% of beginning-of-life (BOL) intensity, which is the design basis for
application of proper dose to achieve desired rates. At 4,000 hours of lamp use, a
warning will appear on the START/STOP/RESET screen, advising the operator to replace
the lamp(s).
When a lamp is replaced, the lamp hours of the particular lamp should be RESET using
the screen prompts on the screen display. The LAMP STATUS screen can be accessed
via the MAIN MENU  SYSTEM INFORMATION button (as shown in Appendix B –
screen reference 5-iv). Ensure that you select RESET for the correct lamp within the
touch-screen diagram.
17
29-6065-15 Rev A
Storage and Maintenance
29-5100
Protective Film Over Touch-Screen
The touch-screen is protected from UV exposure by a UV resistant screen. The screen
should be replaced when it exhibits signs of wear. See replacement parts list for
replacement screen ordering information.
Fuse Replacement
The Clorox Healthcare™ Optimum-UV™ System is equipped with a 20A, 250V fuse
located externally at the base of the device or may be located inside the base assembly.
To replace the external fuse, unscrew the fuse cap using a flat head screwdriver and
pull the fuse (see Figure 9). After replacing the fuse, push the fuse all the way into the
receptacle and screw in tightly. If the fuse blows again after replacement, please contact
your local Clorox Healthcare representative or call 1-800-234-7700 for further
instructions.
If an internal fuse is provided, contact your local Clorox Healthcare representative or call
1-800-234-7700 for further instructions.
Figure 9: Clorox Healthcare™ Optimum-UV™ System Fuse Holder
CAUTION
PREVENT ANY LIQUID INGRESS INTO THE BASE OF THE UNIT AS
THIS MAY CAUSE THE UNIT TO MALFUNCTION.
18
29-6065-15 Rev A
Storage and Maintenance
29-5100
Battery Maintenance
Remote Control:
The remote hand-held control is battery operated. The battery life expectancy is 5
years. However, a defective battery may fail earlier and need replacing. If the hand-held
remote is not functioning, remove the back cover and replace the two coin batteries as
per Figure 10 below.
Figure 10: Remote Battery Replacement
Touch-Screen Display:
The touch-screen display has a battery that maintains clock operation and supports
other time based features. The battery life expectancy is 5 to 7 years. However, a
defective battery may fail earlier and need replacing. If the clock on the main screen is
not functioning, the battery needs to be replaced. Contact your local Clorox Healthcare
representative for a service call. If you do not have your representative’s information,
please call 1-800-234-7700 to be directed appropriately.
On-board Computer:
The on-board computer has a rechargeable battery. The battery charges when the
Clorox Healthcare™ Optimum-UV™ System is plugged in, and power is turned on by
pressing the power switch located on the left end of the device handle. A full charge
cycle is 15 hours and the battery should retain charge for many months. However, in the
event the Clorox Healthcare™ Optimum-UV™ System has been unused and not
powered for more than 4 weeks, it is recommended to charge the on-board computer
battery for 15 hours minimum prior to use.
19
29-6065-15 Rev A
Replacement Parts
29-5100
Part Number
(P/N)
Description
Clorox Healthcare™ Optimum-UV™ System includes 4
ea. 62" Maximum Output UV-C Lamps, Remote Control,
Storage Cover, & Warning Sign Kit
Clorox Healthcare™ Optimum-UV™ System
Replacement Lamps: Maximum Output Replacement
Lamps (Protective polymer sleeve encapsulated - 1 pack)
Clorox Healthcare™ Optimum-UV™ System
Replacement Lamps: Maximum Output Replacement
Lamps (Protective polymer sleeve encapsulated - 4 pack)
Clorox Healthcare™ Optimum-UV™ System
Replacement Storage Cover
Clorox Healthcare™ Optimum-UV™ System
Replacement UV Protective Film for Touch-Screen
Clorox Healthcare™ Optimum-UV™ System
Replacement Warning Sign Kit
Clorox Healthcare™ Optimum-UV™ System
Replacement Remote Control (requires service call to
synchronize with parent unit)
Clorox Healthcare™ Optimum-UV™ System
Installation/Operation Manual (IOM)
29-5100
29-4010-15
29-4000-15
29-6041-15
29-6032
29-6063
29-2029-15
29-6065-15
20
29-6065-15 Rev A
Appendix A: Efficacy Data
29-5100
Third-Party Testing Results: Clorox Healthcare™ Optimum-UV™
System against C. difficile Spores and MRSA
Microorganism
C. difficile
(Endospores)
Methicillin
Resistant
Staphylococcus
aureus
(MRSA)
Distance
(ft)
Time
(min)
Percent
Reduction
Compared to
Control
3
8
12
3
8
4
5
8
1
2
>99.995
99.992
>99.995
>99.999
>99.999
>4.32
4.09
>4.32
>5.69
>5.69
12
3
>99.999
>5.69
Log Reduction
Compared to
Control
Note: Clorox Healthcare™ Optimum-UV™ System recommended room dose times
vary based on room size, configuration and other factors. Please consult with your
Clorox Healthcare representative for recommended room dose times.
21
29-6065-15 Rev A
Appendix B: Touch-Screens
29-5100
22
29-6065-15 Rev A
Appendix B: Touch-Screens
29-5100
Note: R = Remote operating device is enabled
Ref No.
1-i
Description
PLC
WELCOME screen, which also displays the unit’s serial
number (SN), current date and time. Press the button in the
lower right when ready to continue.
Note: when power is turned on, WELCOME screen displays
and device will start a 35 second warm-up period to stabilize
motion detector sensors. Touch-screen functionality begins
after the 35 second warm-up.
Same WELCOME screen as 1-i (above), except Remote ON
is indicated when the remote operating device is engaged.
1-i-R
Note: when power is turned on, WELCOME screen displays
and device will start a 35 second warm-up period to stabilize
motion detector sensors. Remote control functionality begins
after the 35 second warm-up.
23
29-6065-15 Rev A
Appendix B: Touch-Screens
2
29-5100
WARNING screen. Click the Next button to continue.
3-i
MAIN MENU screen. Select from one of the five options to
continue.
3-i-R
Same MAIN MENU screen as 3-i (above), except Remote
ON is indicated when the remote operating control is
engaged.
24
29-6065-15 Rev A
Appendix B: Touch-Screens
4-i
29-5100
MAIN MENU  PROGRAM SET UP  ENTER UV CYCLE
TIME. Enter your UV Cycle Time in hours and minutes. For
hours, press the HRS box and enter using the keypad. For
minutes, press the MIN box and enter using the keypad. To
continue to screen 4-ii, press the Next button.
Same ENTER UV CYCLE TIME screen as 4-i (above),
except for the following display features:

4-i-R


Remote ON is indicated when the remote operating
device is engaged.
“ALARM DISABLED!!!” appears when the unit’s
alarm has been disabled (refer to 7-i, Alarm settings).
Change Lamp [1-4]” appears when the respective
lamp(s) reach 4,000 hours. The notification
disappears once the lamp hours are reset (refer to 5iv-c, d, e and f).
25
29-6065-15 Rev A
Appendix B: Touch-Screens
4-ii
4-ii-R
1
29-5100
MAIN MENU  PROGRAM SET UP  ENTER UV CYCLE
TIME  START/STOP/RESET screen. The “CYCLE TIME”
displays the time entered on the previous screen (4-i). Press
the ROOM # button1 to identify the specific room that is
being disinfected (refer to 4-iii for details). After selecting a
room number and pressing the START button, the “EXIT”
timer begins to count down (from 30 seconds to 0 seconds).
Once the countdown reaches 0 seconds, the UV treatment
cycle starts and the “EXPOSURE TIME” counts up to the
entered “CYCLE TIME.” To stop the cycle at any time, press
the STOP button. To reset this screen and/or exit the
countdown, press the RESET button.
Same MAIN MENU  PROGRAM SET UP  ENTER UV
CYCLE TIME  START/STOP/RESET screen as 4-ii
(above), except Remote ON is indicated when the remote
operating device is engaged.
Selecting a room number is optional
26
29-6065-15 Rev A
Appendix B: Touch-Screens
4-iii
MAIN MENU  PROGRAM SET UP  ENTER UV CYCLE
TIME  START/STOP/RESET  ROOM LIST screen.
Select a room list range by pressing the appropriate range
button (which takes you to screen 4-iv). Note that you may
skip room selection by pressing the SKIP button.
4-iv
Select a specific room by pressing the appropriate room
number button. Note that you may skip room selection by
pressing the SKIP button.
4-v
Room confirmation screen. Press the Next button to
continue.
29-5100
27
29-6065-15 Rev A
Appendix B: Touch-Screens
5-i
MAIN MENU  SYSTEM INFORMATION screen. Select
from one of the five options to continue or press the Main
menu button.
5-ii
MAIN MENU  SYSTEM INFORMATION  CYCLE
COMPLETED screen. One of the following status indicators
will appear for each lamp upon cycle completion:
 OK – lamp performed OK
 ON/OFF – lamp malfunctioned
 NEVER ON – lamp did not work
5-iii
29-5100
MAIN MENU  SYSTEM INFORMATION  CYCLE
INTERRUPTED screen. One of the following status
indicators will appear for each lamp upon cycle interruption:
 OK – lamp performed OK
 ON/OFF – lamp malfunctioned
 NEVER ON – lamp did not work.
Press the RESUME button to resume the cycle without
resetting the timer. Alternatively, press the TIMER RESET
button to reset the UV cycle timer and go back to the
START/STOP/RESET screen (4-ii).
28
29-6065-15 Rev A
Appendix B: Touch-Screens
5-iv-a
MAIN MENU  SYSTEM INFORMATION  LAMP
STATUS screen. Press the Lamp Hours button to view the
number of hours that each lamp has been in operation.
5-iv-b
MAIN MENU  SYSTEM INFORMATION  LAMP
STATUS  REMAINING LIFE screen. Note that the
remaining life columns populate with the percent life
(80%/60%/40%/20%/0) left in each lamp. To reset the hours
for a lamp that will be changed out, press the Reset Hrs
button.
5-iv-c
MAIN MENU  SYSTEM INFORMATION  LAMP
STATUS  REMAINING LIFE  Reset Hrs. screen. To
continue, press the Go to Reset button.
29-5100
29
29-6065-15 Rev A
Appendix B: Touch-Screens
5-iv-d
MAIN MENU  SYSTEM INFORMATION  LAMP
STATUS  REMAINING LIFE  Reset Hrs. (continued)
screen. Select an individual lamp position button to reset the
respective lamp’s hours, or the ALL button to reset all the
lamps at once.
5-iv-e
MAIN MENU  SYSTEM INFORMATION  LAMP
STATUS  REMAINING LIFE  Reset Hrs. (continued) 
Confirmation screen. Select YES to proceed or NO to
cancel.
5-iv-f
MAIN MENU  SYSTEM INFORMATION  LAMP
STATUS  REMAINING LIFE  Reset Hrs. (continued) 
Confirmation  Completion screen.
29-5100
30
29-6065-15 Rev A
Appendix B: Touch-Screens
5-v
MAIN MENU  SYSTEM INFORMATION  METERS
screen. Displays the total meters (cumulative device run
time) in hours, minutes and seconds as well as the total
number of UV treatment cycles.
5-vi
MAIN MENU  SYSTEM INFORMATION  CLOROX
HEALTHCARE CONTACT INFORMATION screen.
29-5100
31
29-6065-15 Rev A
Appendix B: Touch-Screens
6-i
MAIN MENU  CLOCK SETUP screen. Shows the current
date and time. To change the date or time, press ENTER
and input the current date/time using the keypad. All fields
must be populated.
7-i
MAIN MENU  ALARM SETUP screen. Enable/disable the
touch sound (the sound emitted by the device when you
touch the screen). Enable/disable the UV Cycle alarm, or
modify the sound level (high/low).
29-5100
32
29-6065-15 Rev A
Appendix B: Touch-Screens
8-i
MAIN MENU  DATA LOG screen. View the activity log for
this device. Fcs. Up scrolls the list up, while Fcs. Dwn scrolls
the list down. Press DEL to clear the device memory. The
Ref. button is currently not in use.
8-ii
Click YES to clear the device’s data log or NO to return to
the previous screen.
9-i
When the emergency CUTOFF button is pressed on the
remote control, this screen appears on the unit’s PLC. To
resume operation, turn the unit’s power off and restart.
Otherwise, contact supervisor.
29-5100
33
29-6065-15 Rev A
Equipment Warranty
29-5100
The Clorox Professional Products Company or its designated affiliate (“Clorox”)
warrants to original buyer for the longer of (i) one year from the date of original
purchase and (ii) the duration of any extended warranty purchased by such buyer from
Clorox (or from another seller under Clorox’s authority, as applicable), that the Clorox
Healthcare™ Optimum-UV™ System (29-5100) (“Product”) will be free from defect in
material and workmanship under normal use and service. Clorox’s obligation under this
warranty shall be limited to the repair or replacement of Product which proves defective,
provided that such Products are installed, maintained, and operated for the purpose and
in the manner intended and for which Clorox instructs or recommends. Neither Clorox
nor Clorox’s supplier shall be liable to the buyer for any special or consequential
damages directly or indirectly arising from the design, construction, installation,
servicing, or operation of the Product. THIS IS THE SOLE WARRANTY PROVIDED TO
BUYER. NEITHER CLOROX NOR CLOROX’S SUPPLIER MAKES ANY OTHER
WARRANTY OF ANY KIND, EXPRESSED OR IMPLIED. ALL IMPLIED WARRANTIES
OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH
EXCEED THE OBLIGATIONS IN THE FIRST SENTENCE ABOVE ARE HEREBY
DISCLAIMED AND EXCLUDED FROM THIS WARRANTY. THE LIABILITY OF
CLOROX AND CLOROX’S SUPPLIER UNDER THIS WARRANTY SHALL IN NO
EVENT EXCEED THE COST OF THE PRODUCT SOLD TO BUYER UNDER THIS
CONTRACT OF SALE. This warranty shall not apply to any Products (a) which have
been subjected to misuse, tampering, negligence, or accidents; or (b) the serial
numbers of which have been altered, defaced, or removed; or (c) which have been used
in a manner contrary to Clorox’s instructions or recommendations. Buyer shall not return
to Clorox any allegedly defective goods without Clorox’s prior written authorization. This
warranty may not be assigned or transferred.
For more information, contact your Clorox Healthcare sales representative
or call 800-234-7700.
email: UV@clorox.com
visit us: www.cloroxhealthcare.com/UV
© 2014 Clorox Professional Products Company, 1221 Broadway, Oakland, CA 94612
Manufactured by:
Specifications subject to change without notice.
34
29-6065-15 Rev A