Installation/Owner’s Manual 29-5100 Clorox Healthcare™ Optimum-UV™ System Powered by UVDI Technology Patent No. D684671 29-6065-15 Rev A Manual Contents 29-5100 Safety Information ……………………………………………………………………………..3 Product Overview ……………………………………………………………………………...5 Room Treatment Protocols .…………………………………...…………………………….11 Device System Operation …………………………………………………………………....13 Device Remote Operation …………………………………………………………………...15 Storage and Maintenance ……………………………………………………………………17 Replacement Parts ……………………………………………………………………………20 Appendix A: Efficacy Data ……………………………………………………………………21 Appendix B: Touch-Screens …………………………………………………………………22 Equipment Warranty ………………………………………………………………………….34 29-6065-15 Rev A Safety Information 29-5100 It is mandatory to read and understand this manual before operating this device. This is for your safety and safe operation of the device. 3 29-6065-15 Rev A Safety Information 29-5100 WARNING AVERTISSEMENT RISK OF ELECTRIC SHOCK! OPERATING INSTRUCTIONS ARE FOR USE BY QUALIFIED PERSONNEL ONLY. TO REDUCE THE RISK OF ELECTRIC SHOCK, DO NOT PERFORM ANY SERVICING OTHER THAN THAT CONTAINED IN THE OPERATING INSTRUCTIONS. RISQUE DE CHOC ÉLECTRIQUE! LES INSTRUCTIONS D'INSTALLATION NE DOIVENT ÊTRE UTILISÉS QUE PAR DU PERSONNEL QUALIFIÉ. POUR RÉDUIRE LE RISQUE DE CHOC ÉLECTRIQUE, NE JAMAIS EFFECTUER DE RÉPARATIONS AUTRES QUE CELLES CONTENUES DANS LE MANUEL D'UTILISATION. WARNING UV LIGHT AND LAMP BURN HAZARD! HARMFUL TO BARE SKIN AND EYES. CAN CAUSE BURNS. CAN CAUSE TEMPORARY LOSS OF VISION. NEVER LOOK AT THE LAMPS WHILE ILLUMINATED. CAUTION TO REDUCE THE RISK OF ELECTRICAL SHOCK, THIS EQUIPMENT HAS A GROUNDING-TYPE PLUG THAT HAS A THIRD (GROUNDING) PIN. THIS PLUG WILL ONLY FIT INTO A GROUNDING TYPE POWER OUTLET. IF THE PLUG DOES NOT FIT INTO THE OUTLET, CONTACT QUALIFIED PERSONNEL TO INSTALL THE PROPER OUTLET. DO NOT ALTER THE PLUG IN ANY WAY. CAUTION THIS DEVICE CONTAINS MERCURY IN THE SEALED ULTRAVIOLET BULB(S). DISPOSE OF BULBS IN ACCORDANCE WITH STATE AND LOCAL LAWS. CAUTION USE CAUTION WHEN MOVING THE DEVICE OVER UNEVEN OR SLOPED GROUND. AVERTISSEMENT DANGER, LA LUMIÈRE UV ET LAMPE BRÛLE! NOCIF POUR LA PEAU NUE ET LES YEUX. PEUT CAUSER DES BRÛLURES GRAVES. PEUT ENTRAÎNER UNE PERTE TEMPORAIRE OU PERMANENTE DE LA VISION. NE JAMAIS REGARDER LA LAMPE LORSQUE ILLUMINÉ. ATTENTION POUR RÉDUIRE LE RISQUE DE CHOC ÉLECTRIQUE, CET ÉQUIPEMENT DISPOSE D’UN CONNECTEUR AVEC UNE TROISIÈME BROCHE DE PRISE DE TERRE. CETTE FICHE NE PEUT S'INSÉRER QUE DANS UNE PRISE DE COURANT AYANT UNE PRISE DE TERRE. SI LA FICHE NE RENTRE PAS DANS LA PRISE, CONTACTEZ UN PERSONNEL QUALIFIÉ POUR INSTALLER UNE PRISE APPROPRIÉE. NE MODIFIEZ PAS LA FICHE EN AUCUNE CIRCONSTANCE. ATTENTION CET APPAREIL CONTIENT DU MERCURE DANS L'AMPOULE À ULTRAVIOLETS SCELLÉE (S). DISPOSEZ DES AMPOULES EN CONFORMITÉ AVEC LES LOIS NATIONALES ET LOCALES. ATTENTION SOYEZ PRUDENT LORSQUE VOUS DÉPLACEZ L'APPAREIL SUR UN TERRAIN INÉGAL OU EN PENTE. 4 29-6065-15 Rev A Product Overview 29-5100 The Clorox Healthcare™ Optimum-UV™ System is designed to treat surfaces contaminated with pathogens. Typical applications include hospital rooms, operating suites, emergency rooms, patient waiting rooms, isolation wards, public restrooms, nurse’s stations, staff break rooms, locker rooms, clean rooms, laboratory rooms, and food preparation areas. Please note that the Clorox Healthcare™ Optimum-UV™ System should be used as a supplement – not as a replacement – to standard terminal cleaning protocols. Please follow your facility’s manual room cleaning and disinfection protocols before using a UV device, and consult CDC, SHEA and APIC guidelines for additional steps to take as a part of a bundled approach. ATTENTION PROPER USE OF THIS DEVICE IS CRITICAL IN YIELDING DESIRED TREATMENT RESULTS. THE UV-C TREATMENT RESULTS ACHIEVED ARE SUBJECT TO VARIATION IN APPLICATION CONDITIONS. CLOROX HEALTHCARE RECOMMENDS CULTURE TESTS TO CONFIRM DESIRED OUTCOMES. The device emits UV-C energy in a full circle, or 360 degrees, to deliver a programmed dose of UV light for treating a room’s target surfaces. Using the Clorox Healthcare™ Optimum-UV™ System does not require special room preparation – simply expose target surfaces to the device prior to starting the room treatment cycle. It should be noted that because the Clorox Healthcare™ Optimum-UV™ System utilizes “line-of-sight” UV technology, target surface(s) must be directly exposed to the device’s UV energy. UV energy will not penetrate materials such as paper, soil, cloth, plastic, ordinary glass or spills. After completion of the treatment cycle, the room may be immediately occupied. The Clorox Healthcare™ Optimum-UV™ System does not generate any harmful byproducts, nor does it leave any residual substances. Ultraviolet light kills microorganisms via short-wavelength ultraviolet radiation (UV-C). The Clorox Healthcare™ Optimum-UV™ System produces UV-C at a wavelength of approximately 254 nm, inactivating the microorganisms’ DNA. 5 29-6065-15 Rev A Product Overview 29-5100 Figure 1: Clorox Healthcare™ Optimum-UV™ System Line Drawing 6 29-6065-15 Rev A Product Overview 29-5100 Figure 2: Touch-Screen Operating System Figure 3: Power Switch Location Figure 4: Safety Motion Sensors Figure 5: Locking Wheels 7 29-6065-15 Rev A Product Overview 29-5100 Specifications Specification Description Lamp Type Four (4) maximum output UV lamps; 254-nm wavelength Treatment area Single room Ballast (4) Electronic, solid state Lamp Length 62 inches Footprint 36 inch-diameter circular base, 76 inches tall Lamp life 4,000 hours accumulated “ON” time. Power Source 120 VAC, 60Hz, 12.5 Amp Power Connector 15 foot long hospital-grade cord Delay timer to turn unit ON 30 seconds delay Power Interruption Memory Device will resume operation after interruption and remember condition before power interruption or before shutdown due to intentional or unintentional interruption. Motion sensors (safety) 360 degrees Operating Environment Indoor 50°F to 80°F, 50% RH @70F, non-condensing 8 29-6065-15 Rev A Product Overview 29-5100 Device Assembly Step 1: Install Maximum Output UV-C Lamps Lamp Information The Clorox Healthcare™ Optimum-UV™ System is equipped with four (4) maximum output UV-C lamps. Each lamp is completely encapsulated in a protective polymer sleeve. The protective polymer sleeve provides safety in the event of lamp breakage by fully containing the content inside. The protective polymer sleeve is highly transmissible to UV-C light and does not affect performance of the device. UV lamps supplied with the Clorox Healthcare™ Optimum-UV™ System are factory tested to ensure proper UV-C intensity. Each lamp has a serialized label on the lamp end cap with pins, which verifies the testing process has been completed. Lamp Orientation Each lamp has a brown label on the black end caps. When installing the lamp, please ensure that the brown label faces in towards the mast. Correct orientation is shown in picture below. CORRECT ORIENTATION OF LAMP INCORRECT ORIENTATION OF LAMP Note: Improper installation of the lamps may affect performance of the Clorox Healthcare™ Optimum-UV™ System. 9 29-6065-15 Rev A Product Overview 29-5100 Insert the lamp-end cap into the lamp catch. Refer to Figure 6, below. Figure 6: Lamp Catch Step 2: Insert Lamp Pins Insert lamp pins into white connectors in base stand. Refer to Figure 7, below. Note that the lamp-connector pins configuration is rectangular and must be oriented correctly to the receptacle before inserting. Figure 7: Insert Lamp Pins 10 29-6065-15 Rev A Room Treatment Protocols 29-5100 The Clorox Healthcare™ Optimum-UV™ System should be used as a supplement – not as a replacement – to standard cleaning and disinfection protocols. Please follow your facility’s manual room cleaning and disinfection protocols before using a UV device, and consult CDC, SHEA and APIC guidelines for additional steps to take as a part of a bundled approach. Once you have completed your standard cleaning and disinfection protocols, please deploy the Clorox Healthcare™ Optimum-UV™ System. Step 1: Post Warning Sign and Position Clorox Healthcare™ OptimumUV™ System IMPORTANT IF OPERATOR DOES NOT EXIT ROOM WITHIN 30 SECONDS, UNIT WILL FAIL TO BEGIN TREATMENT CYCLE DUE TO MOTION DETECTOR SHUTDOWN. PUSH START BUTTON AGAIN TO REINITIATE THE CYCLE. Affix the warning sign on the door to indicate “DO NOT ENTER” during the room treatment cycle. Position the Clorox Healthcare™ Optimum-UV™ System in an appropriate location to ensure line-of-sight exposure to the room’s target surfaces. Ensure that all surfaces requiring treatment are exposed to the unit. Remove papers and/or other objects that may be shadowing target surfaces. Note that if the room is equipped with furniture, equipment and/or fixtures, it may be necessary to re-locate the Clorox Healthcare™ Optimum-UV™ System and run additional cycles to ensure proper treatment of all target surfaces in the room. Step 2: Secure Doors Secure doors leading into the room so that no one can enter and interrupt the treatment cycle. CAUTION HIGH TEMPERATURE! DO NOT TOUCH LAMP OR MAST Step 3: Start Cycle and Exit Room Plug the electrical cord into the nearest electrical outlet and power up the Clorox Healthcare™ Optimum-UV™ System by pressing the power switch located on the left end of the device handle. The touch-screen display will activate and may be used to configure and execute all functions. Please refer to Appendix B: Clorox Healthcare™ Optimum-UV™ System Touch-Screens for additional information. Note: When power is turned on, WELCOME screen displays and device will start a 35 second warm-up period to stabilize motion detector sensors. Touch-screen functionality begins after the 35 second warm-up period. 11 29-6065-15 Rev A Room Treatment Protocols 29-5100 Navigate to the ENTER UV CYCLE TIME screen to enter the UV treatment cycle time and continue to the START/STOP/RESET screen to indicate a Room number, if desired. Press the touch-screen START button (or initiate using the remote control) to begin UV treatment. When START is executed, a slow, pulsing, warning beep will be heard for 25 seconds, followed by a rapid, pulsing beep for the final 5 seconds before the UV treatment cycle commences. The warning beep’s volume may be adjusted via the touch-screen display. After the 30-second delay, the Clorox Healthcare™ OptimumUV™ System unit will begin its cycle and a visible blue light will be observed within 2025 seconds. IMPORTANT OPERATOR MUST LEAVE THE ROOM DURING ENTIRE CYCLE TIME TO AVOID UV EXPOSURE Step 4: Enter Room after Treatment Cycle Completes After the treatment cycle completes, the operator may immediately enter the room. When a cycle completes successfully, then the CYCLE COMPLETED screen will be displayed. The Clorox Healthcare™ Optimum-UV™ System is now ready for the next treatment cycle and can be moved to a new location. If the cycle is interrupted due to device sensors detecting movement in the room, then the CYCLE INTERRUPTED screen will be displayed. The interrupted cycle may be continued to completion or re-started by following the touch-screen prompts. After completion of the treatment cycle, the room may be immediately occupied. The Clorox Healthcare™ Optimum-UV™ System does not generate any harmful byproducts, nor does it leave any residual substances. Note that a faint, “clean” smell may be detected in the room after the cycle. UV-C breaks down the molecules associated with odors and unpleasant smells in the air. This results in the “clean air,” smell after the programmed cycle. The odor’s chemical composition has been documented and analyzed, and has been found to pose no risk. 12 29-6065-15 Rev A Device System Operation 29-5100 Step 1: Power on Device Turn on the power by pressing the power switch located on the left end of the device handle. The touch-screen display will activate and may be used to configure and execute all functions. Please refer to Appendix B: Clorox Healthcare™ Optimum-UV™ System Touch-Screens for additional information. Once the touch-screen has been activated, navigate to the MAIN MENU screen, which provides the following options: Menu Options Function PROGRAM SET UP Sub-screens allow user to enter dose time, select room, start the cycle and monitor treatment cycle progress. ALARM SET UP Enable or disable the cycle start alarm. If enabled, set volume to high or low. SYSTEM INFORMATION System performance information, including lamp life data, total hours of use, and UV treatment cycles completed. CLOCK SET UP Enables clock set up date and time. DATA LOG Records and displays Clorox Healthcare™ Optimum-UV™ System historical use data. Step 2: Enter UV Cycle Time Press the PROGRAM SET UP button and enter the UV CYCLE TIME in hours and minutes. Note that the message REMOTE ON appears when the remote function is activated. Note: The time entered for Minutes field should be a number other than zero. Step 3: Start the UV Cycle The unit’s room selection feature is enabled by pressing the ROOM # button. Follow the on-screen prompts to select the room number for data retention, if desired. Then, from the START/STOP/RESET screen, push the START button to begin the UV cycle. The Operator MUST exit the room within 30 seconds of pushing the START button. Note that before 30 seconds has expired, the operator can push the STOP button to terminate operation. If the operator does not exit the room, the Clorox Healthcare™ Optimum-UV™ System will fail to start due to the motion detector safety feature. 13 29-6065-15 Rev A Device System Operation 29-5100 Step 4: Complete the UV Cycle If the cycle has fully completed, the CYCLE COMPLETED screen will appear. If the cycle has been interrupted, the CYCLE INTERRUPTED screen will appear. Note that if the cycle is interrupted, the operator can restart the unit by pressing the START button, OR reset the timer and start over. If power was interrupted, the timer conditions will return to settings prior to the power interruption. CAUTION ONCE CYCLE STARTS AND THE OPERATOR IS OUTSIDE THE ROOM, CYCLE CAN ONLY BE INTERRUPTED PRIOR TO CYCLE COMPLETION BY USING THE REMOTE PROVIDED (SEE DEVICE REMOTE OPERATION). DO NOT ATTEMPT TO INTERRUPT CYCLE BY ENTERING THE ROOM AS IT MAY RESULT IN UV EXPOSURE! 14 29-6065-15 Rev A Device Remote Operation 29-5100 Remote Control The Clorox Healthcare™ Optimum-UV™ System UV treatment cycle may be started and stopped remotely with the remote hand-held control. The remote control also serves as a safety device to cut power to the unit or stop the cycle in case of emergency. Figure 8: Clorox Healthcare™ Optimum-UV™ System Remote Hand-held Control Step 1: Activate Remote Control First, apply power to the Clorox Healthcare™ Optimum-UV™ System and allow the device to complete a 35 second warm up. Activate the remote control by pushing the READY button. Note that the READY button must be depressed at both the top and bottom of the button to function. When the remote is activated, REMOTE ON will flash on/off on the Clorox Healthcare™ Optimum-UV™ System unit’s touch-screen. Note: When power is turned on, WELCOME screen displays and device will start a 35 second warm-up period to stabilize motion detector sensors. Remote control functionality begins after the 35 second warm-up. 15 29-6065-15 Rev A Device Remote Operation 29-5100 Step 2: Start Operation Start operation by pressing the buttons UV + ON simultaneously. The Clorox Healthcare™ Optimum-UV™ System will commence the 30 second programmed delay and then perform the programmed UV treatment cycle. Step 3: Stop Operation To stop operation, press UV OFF, which requires the UV + OFF buttons to be pressed simultaneously. This will stop the UV treatment cycle and allow safe entry to the room. To resume a UV cycle, exit the room and push buttons UV + ON simultaneously. The UV cycle will resume and complete the remaining cycle time of the current program. Note that if you have initiated the Start Operation process using the remote, and you need to interrupt the treatment start cycle during the 30-second delay of the start sequence, then you can use the UV + OFF feature to do so using the remote or pressing STOP on the screen display. For EMERGENCY SHUTDOWN (bypassing the internal controller), the CUT OFF button may be pressed in the event operator determines that controller has malfunctioned and treatment cycle must be terminated. Note that the CUT OFF button must be depressed at the top and the bottom to function. IMPORTANT NORMAL CYCLE INTERRUPTION CAN BE ACHIEVED BY PRESSING THE UV+ OFF BUTTON ON THE REMOTE. CUT-OFF BUTTON SHOULD ONLY BE USED IF UV+ OFF BUTTON DOES NOT WORK OR ANY OTHER EVENT CAUSING LAMPS TO NOT SHUT OFF The Power switch must be cycled OFF and ON to reset the Clorox Healthcare™ Optimum-UV™ System after an EMERGENCY SHUTDOWN. Note: Activation of remote control by pushing the READY button is not required for EMERGENCY SHUTDOWN Step 4: Cycle Again To start a new cycle after CYCLE COMPLETION or to continue a cycle after CYCLE INTERRUPTION using the remote, go back to Step 2 in this sequence. 16 29-6065-15 Rev A Storage and Maintenance 29-5100 Storage When the Clorox Healthcare™ Optimum-UV™ System is not in use, cover it before storing. Insert the storage cover into the handle bar, and then carefully lift the cover over the mast, while ensuring that the receiver antenna fits through the cover hole. Finally, close the zipper. CAUTION LAMPS MUST BE COOLED DOWN BEFORE COVERING THE DEVICE. COOLING TIME MAY VARY DEPENDING ON ENVIRONMENTAL CONDITIONS. 10 TO 15 MINUTES ARE RECOMMENDED. PLACE THE COVER BACK ON THE DEVICE AND MOVE IT TO ITS STORAGE LOCATION OR THE NEXT AREA TO BE TREATED. Lamp Information The Clorox Healthcare™ Optimum-UV™ System is equipped with four (4) maximum output UV-C lamps. Each lamp is completely encapsulated in a protective polymer sleeve. The protective polymer sleeve provides safety in the event of lamp breakage by fully containing the content inside. The protective polymer sleeve is highly transmissible to UV-C and does not affect performance of the device. Lamp Replacement Clorox Healthcare strongly recommends that the lamps be replaced after 4,000 hours of use. Please contact your local Clorox Healthcare representative to order replacement lamps. If you do not have your representative’s information, please call 1-800-234-7700. The use of non-Clorox Healthcare lamps will void the warranty. Note: Refer to the LAMP STATUS screen to check accumulated lamp hours for each lamp. UV-C lamp intensity degrades as a function of time. After 4,000 hours of use, lamp intensity is 80% of beginning-of-life (BOL) intensity, which is the design basis for application of proper dose to achieve desired rates. At 4,000 hours of lamp use, a warning will appear on the START/STOP/RESET screen, advising the operator to replace the lamp(s). When a lamp is replaced, the lamp hours of the particular lamp should be RESET using the screen prompts on the screen display. The LAMP STATUS screen can be accessed via the MAIN MENU SYSTEM INFORMATION button (as shown in Appendix B – screen reference 5-iv). Ensure that you select RESET for the correct lamp within the touch-screen diagram. 17 29-6065-15 Rev A Storage and Maintenance 29-5100 Protective Film Over Touch-Screen The touch-screen is protected from UV exposure by a UV resistant screen. The screen should be replaced when it exhibits signs of wear. See replacement parts list for replacement screen ordering information. Fuse Replacement The Clorox Healthcare™ Optimum-UV™ System is equipped with a 20A, 250V fuse located externally at the base of the device or may be located inside the base assembly. To replace the external fuse, unscrew the fuse cap using a flat head screwdriver and pull the fuse (see Figure 9). After replacing the fuse, push the fuse all the way into the receptacle and screw in tightly. If the fuse blows again after replacement, please contact your local Clorox Healthcare representative or call 1-800-234-7700 for further instructions. If an internal fuse is provided, contact your local Clorox Healthcare representative or call 1-800-234-7700 for further instructions. Figure 9: Clorox Healthcare™ Optimum-UV™ System Fuse Holder CAUTION PREVENT ANY LIQUID INGRESS INTO THE BASE OF THE UNIT AS THIS MAY CAUSE THE UNIT TO MALFUNCTION. 18 29-6065-15 Rev A Storage and Maintenance 29-5100 Battery Maintenance Remote Control: The remote hand-held control is battery operated. The battery life expectancy is 5 years. However, a defective battery may fail earlier and need replacing. If the hand-held remote is not functioning, remove the back cover and replace the two coin batteries as per Figure 10 below. Figure 10: Remote Battery Replacement Touch-Screen Display: The touch-screen display has a battery that maintains clock operation and supports other time based features. The battery life expectancy is 5 to 7 years. However, a defective battery may fail earlier and need replacing. If the clock on the main screen is not functioning, the battery needs to be replaced. Contact your local Clorox Healthcare representative for a service call. If you do not have your representative’s information, please call 1-800-234-7700 to be directed appropriately. On-board Computer: The on-board computer has a rechargeable battery. The battery charges when the Clorox Healthcare™ Optimum-UV™ System is plugged in, and power is turned on by pressing the power switch located on the left end of the device handle. A full charge cycle is 15 hours and the battery should retain charge for many months. However, in the event the Clorox Healthcare™ Optimum-UV™ System has been unused and not powered for more than 4 weeks, it is recommended to charge the on-board computer battery for 15 hours minimum prior to use. 19 29-6065-15 Rev A Replacement Parts 29-5100 Part Number (P/N) Description Clorox Healthcare™ Optimum-UV™ System includes 4 ea. 62" Maximum Output UV-C Lamps, Remote Control, Storage Cover, & Warning Sign Kit Clorox Healthcare™ Optimum-UV™ System Replacement Lamps: Maximum Output Replacement Lamps (Protective polymer sleeve encapsulated - 1 pack) Clorox Healthcare™ Optimum-UV™ System Replacement Lamps: Maximum Output Replacement Lamps (Protective polymer sleeve encapsulated - 4 pack) Clorox Healthcare™ Optimum-UV™ System Replacement Storage Cover Clorox Healthcare™ Optimum-UV™ System Replacement UV Protective Film for Touch-Screen Clorox Healthcare™ Optimum-UV™ System Replacement Warning Sign Kit Clorox Healthcare™ Optimum-UV™ System Replacement Remote Control (requires service call to synchronize with parent unit) Clorox Healthcare™ Optimum-UV™ System Installation/Operation Manual (IOM) 29-5100 29-4010-15 29-4000-15 29-6041-15 29-6032 29-6063 29-2029-15 29-6065-15 20 29-6065-15 Rev A Appendix A: Efficacy Data 29-5100 Third-Party Testing Results: Clorox Healthcare™ Optimum-UV™ System against C. difficile Spores and MRSA Microorganism C. difficile (Endospores) Methicillin Resistant Staphylococcus aureus (MRSA) Distance (ft) Time (min) Percent Reduction Compared to Control 3 8 12 3 8 4 5 8 1 2 >99.995 99.992 >99.995 >99.999 >99.999 >4.32 4.09 >4.32 >5.69 >5.69 12 3 >99.999 >5.69 Log Reduction Compared to Control Note: Clorox Healthcare™ Optimum-UV™ System recommended room dose times vary based on room size, configuration and other factors. Please consult with your Clorox Healthcare representative for recommended room dose times. 21 29-6065-15 Rev A Appendix B: Touch-Screens 29-5100 22 29-6065-15 Rev A Appendix B: Touch-Screens 29-5100 Note: R = Remote operating device is enabled Ref No. 1-i Description PLC WELCOME screen, which also displays the unit’s serial number (SN), current date and time. Press the button in the lower right when ready to continue. Note: when power is turned on, WELCOME screen displays and device will start a 35 second warm-up period to stabilize motion detector sensors. Touch-screen functionality begins after the 35 second warm-up. Same WELCOME screen as 1-i (above), except Remote ON is indicated when the remote operating device is engaged. 1-i-R Note: when power is turned on, WELCOME screen displays and device will start a 35 second warm-up period to stabilize motion detector sensors. Remote control functionality begins after the 35 second warm-up. 23 29-6065-15 Rev A Appendix B: Touch-Screens 2 29-5100 WARNING screen. Click the Next button to continue. 3-i MAIN MENU screen. Select from one of the five options to continue. 3-i-R Same MAIN MENU screen as 3-i (above), except Remote ON is indicated when the remote operating control is engaged. 24 29-6065-15 Rev A Appendix B: Touch-Screens 4-i 29-5100 MAIN MENU PROGRAM SET UP ENTER UV CYCLE TIME. Enter your UV Cycle Time in hours and minutes. For hours, press the HRS box and enter using the keypad. For minutes, press the MIN box and enter using the keypad. To continue to screen 4-ii, press the Next button. Same ENTER UV CYCLE TIME screen as 4-i (above), except for the following display features: 4-i-R Remote ON is indicated when the remote operating device is engaged. “ALARM DISABLED!!!” appears when the unit’s alarm has been disabled (refer to 7-i, Alarm settings). Change Lamp [1-4]” appears when the respective lamp(s) reach 4,000 hours. The notification disappears once the lamp hours are reset (refer to 5iv-c, d, e and f). 25 29-6065-15 Rev A Appendix B: Touch-Screens 4-ii 4-ii-R 1 29-5100 MAIN MENU PROGRAM SET UP ENTER UV CYCLE TIME START/STOP/RESET screen. The “CYCLE TIME” displays the time entered on the previous screen (4-i). Press the ROOM # button1 to identify the specific room that is being disinfected (refer to 4-iii for details). After selecting a room number and pressing the START button, the “EXIT” timer begins to count down (from 30 seconds to 0 seconds). Once the countdown reaches 0 seconds, the UV treatment cycle starts and the “EXPOSURE TIME” counts up to the entered “CYCLE TIME.” To stop the cycle at any time, press the STOP button. To reset this screen and/or exit the countdown, press the RESET button. Same MAIN MENU PROGRAM SET UP ENTER UV CYCLE TIME START/STOP/RESET screen as 4-ii (above), except Remote ON is indicated when the remote operating device is engaged. Selecting a room number is optional 26 29-6065-15 Rev A Appendix B: Touch-Screens 4-iii MAIN MENU PROGRAM SET UP ENTER UV CYCLE TIME START/STOP/RESET ROOM LIST screen. Select a room list range by pressing the appropriate range button (which takes you to screen 4-iv). Note that you may skip room selection by pressing the SKIP button. 4-iv Select a specific room by pressing the appropriate room number button. Note that you may skip room selection by pressing the SKIP button. 4-v Room confirmation screen. Press the Next button to continue. 29-5100 27 29-6065-15 Rev A Appendix B: Touch-Screens 5-i MAIN MENU SYSTEM INFORMATION screen. Select from one of the five options to continue or press the Main menu button. 5-ii MAIN MENU SYSTEM INFORMATION CYCLE COMPLETED screen. One of the following status indicators will appear for each lamp upon cycle completion: OK – lamp performed OK ON/OFF – lamp malfunctioned NEVER ON – lamp did not work 5-iii 29-5100 MAIN MENU SYSTEM INFORMATION CYCLE INTERRUPTED screen. One of the following status indicators will appear for each lamp upon cycle interruption: OK – lamp performed OK ON/OFF – lamp malfunctioned NEVER ON – lamp did not work. Press the RESUME button to resume the cycle without resetting the timer. Alternatively, press the TIMER RESET button to reset the UV cycle timer and go back to the START/STOP/RESET screen (4-ii). 28 29-6065-15 Rev A Appendix B: Touch-Screens 5-iv-a MAIN MENU SYSTEM INFORMATION LAMP STATUS screen. Press the Lamp Hours button to view the number of hours that each lamp has been in operation. 5-iv-b MAIN MENU SYSTEM INFORMATION LAMP STATUS REMAINING LIFE screen. Note that the remaining life columns populate with the percent life (80%/60%/40%/20%/0) left in each lamp. To reset the hours for a lamp that will be changed out, press the Reset Hrs button. 5-iv-c MAIN MENU SYSTEM INFORMATION LAMP STATUS REMAINING LIFE Reset Hrs. screen. To continue, press the Go to Reset button. 29-5100 29 29-6065-15 Rev A Appendix B: Touch-Screens 5-iv-d MAIN MENU SYSTEM INFORMATION LAMP STATUS REMAINING LIFE Reset Hrs. (continued) screen. Select an individual lamp position button to reset the respective lamp’s hours, or the ALL button to reset all the lamps at once. 5-iv-e MAIN MENU SYSTEM INFORMATION LAMP STATUS REMAINING LIFE Reset Hrs. (continued) Confirmation screen. Select YES to proceed or NO to cancel. 5-iv-f MAIN MENU SYSTEM INFORMATION LAMP STATUS REMAINING LIFE Reset Hrs. (continued) Confirmation Completion screen. 29-5100 30 29-6065-15 Rev A Appendix B: Touch-Screens 5-v MAIN MENU SYSTEM INFORMATION METERS screen. Displays the total meters (cumulative device run time) in hours, minutes and seconds as well as the total number of UV treatment cycles. 5-vi MAIN MENU SYSTEM INFORMATION CLOROX HEALTHCARE CONTACT INFORMATION screen. 29-5100 31 29-6065-15 Rev A Appendix B: Touch-Screens 6-i MAIN MENU CLOCK SETUP screen. Shows the current date and time. To change the date or time, press ENTER and input the current date/time using the keypad. All fields must be populated. 7-i MAIN MENU ALARM SETUP screen. Enable/disable the touch sound (the sound emitted by the device when you touch the screen). Enable/disable the UV Cycle alarm, or modify the sound level (high/low). 29-5100 32 29-6065-15 Rev A Appendix B: Touch-Screens 8-i MAIN MENU DATA LOG screen. View the activity log for this device. Fcs. Up scrolls the list up, while Fcs. Dwn scrolls the list down. Press DEL to clear the device memory. The Ref. button is currently not in use. 8-ii Click YES to clear the device’s data log or NO to return to the previous screen. 9-i When the emergency CUTOFF button is pressed on the remote control, this screen appears on the unit’s PLC. To resume operation, turn the unit’s power off and restart. Otherwise, contact supervisor. 29-5100 33 29-6065-15 Rev A Equipment Warranty 29-5100 The Clorox Professional Products Company or its designated affiliate (“Clorox”) warrants to original buyer for the longer of (i) one year from the date of original purchase and (ii) the duration of any extended warranty purchased by such buyer from Clorox (or from another seller under Clorox’s authority, as applicable), that the Clorox Healthcare™ Optimum-UV™ System (29-5100) (“Product”) will be free from defect in material and workmanship under normal use and service. Clorox’s obligation under this warranty shall be limited to the repair or replacement of Product which proves defective, provided that such Products are installed, maintained, and operated for the purpose and in the manner intended and for which Clorox instructs or recommends. Neither Clorox nor Clorox’s supplier shall be liable to the buyer for any special or consequential damages directly or indirectly arising from the design, construction, installation, servicing, or operation of the Product. THIS IS THE SOLE WARRANTY PROVIDED TO BUYER. NEITHER CLOROX NOR CLOROX’S SUPPLIER MAKES ANY OTHER WARRANTY OF ANY KIND, EXPRESSED OR IMPLIED. ALL IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH EXCEED THE OBLIGATIONS IN THE FIRST SENTENCE ABOVE ARE HEREBY DISCLAIMED AND EXCLUDED FROM THIS WARRANTY. THE LIABILITY OF CLOROX AND CLOROX’S SUPPLIER UNDER THIS WARRANTY SHALL IN NO EVENT EXCEED THE COST OF THE PRODUCT SOLD TO BUYER UNDER THIS CONTRACT OF SALE. This warranty shall not apply to any Products (a) which have been subjected to misuse, tampering, negligence, or accidents; or (b) the serial numbers of which have been altered, defaced, or removed; or (c) which have been used in a manner contrary to Clorox’s instructions or recommendations. Buyer shall not return to Clorox any allegedly defective goods without Clorox’s prior written authorization. This warranty may not be assigned or transferred. For more information, contact your Clorox Healthcare sales representative or call 800-234-7700. email: UV@clorox.com visit us: www.cloroxhealthcare.com/UV © 2014 Clorox Professional Products Company, 1221 Broadway, Oakland, CA 94612 Manufactured by: Specifications subject to change without notice. 34 29-6065-15 Rev A