Education & Training Playing Dice with Steam Sterilization Process Monitoring by Martha Young, BS, MS, CSPDT by David Jagrosse, CSPDT Objectives After completion of this self-study activity, the learner will be able to: 1. Choose an internal chemical indicator that provides the information you desire about the sterilization process. 2. Correctly use Class 5 integrating indicator process challenge devices to release non implant loads. 70 MANAGING INFECTION CONTROL 3. 4. Develop a recall policy and procedure based on the results of biological and chemical indicator process challenge devices. Discuss the risks involved with using a chemical indicator process challenge device instead of a biological indicator process challenge device for routine sterilizer monitoring and release of implants. December 2009 Education & Training Test Questions 6. 2. 7. 1. 3. 4. 5. Sterilizer qualification testing for dynamic-air-removal sterilizers includes running a BI PCD in three consecutive empty cycles followed by running a Bowie-Dick test pack in three consecutive empty cycles. A. True B. False Class 5 integrating indicators respond to all the critical parameters of the sterilization process and their performance is designed to be equivalent to or exceed the performance requirements of BIs under ideal steam sterilization conditions A. True B. False A Class 5 integrating indicator process challenge device may be used to release implants. A. True B. False Running a biological indicator in each load will reduce the total number of items affected by a positive biological indicator and, thereby, reduce overall patient risk. A. True B. False Class 6 emulating indicators are cycle verification indicators designed to react to all critical variables of specified sterilization cycles. A. True B. False Many thanks to the team at 3M Health Care for working with Managing Infection Control to provide the following accredited course. IAHCSMM has awarded 1 contact point for completion of this continuing education lesson toward IAHCSMM recertification. The CBSPD has preapproved this inservice for 1 contact hour for a period of five (5) years from the date of publication, and to be used only once in a recertification period. This inservice is 3M Health Care Provider approved by the California Board of Registered Nurses, CEP 5770 for 1 contact hour. This form is valid up to five (5) years from the date of publication. Instructions for submitting results are on page 86. Managing Infection Control and 3M Health Care will be working collaboratively to provide continuing education courses in monthly editions of Managing Infection Control. 72 MANAGING INFECTION CONTROL 8. 9. The usage of Class 6 emulating indicators is not discussed in the Association for the Advancement of Medical Instrumentation’s, Comprehensive guide to steam sterilization and sterility assurance in health care facilities, ANSI/AAMI ST79:2006, A1:2008, and A2:2009. A. True B. False The sterility assurance level (SAL) of a steam sterilization process can be measured by a chemical indicator. A. True B. False The Association for the Advancement of Medical Instrumentation’s, Comprehensive guide to steam sterilization and sterility assurance in health care facilities, ANSI/AAMI ST79:2006, A1:2008, and A2:2009 recommends the use of biological indicators for routine sterilizer efficacy testing, monitoring loads containing implants, and sterilizer qualification testing. A. True B. False If the Class 5 integrating indicator or Class 6 emulating indicator in a process challenge device indicates a sterilization process failure for a load and no operator error is identified then you do not have to recall loads back to the last negative biological indicator. A. True B. False 10. Use recommended practices and standards developed by the Association for the Advancement of Medical Instrumentation, the Association of periOperative Registered Nurses, and the Centers for Disease Control and Prevention to guide best practices and protect patients. A. True B. False Introduction There is nothing worse than a recall. Okay, maybe there is, such as a recall that does not collect 100 percent of the contents of the affected products. A few years ago I had a brand new steam sterilizer that was producing random positive biological indicators. It was the darnedest thing. Each time a sterilization process failure occurred, we would immediately re-verify the results of what I refer to as the “holy trinity”: the physical, chemical (CI) and biological indicator (BI) results. All were performing as expected. There were no signs of a sterilization process failure other than the occasional positive BI. It came to the point that the staff did not “trust” the new $100,000 steam sterilizer anymore. After an exhaustive step-by-step process involving facility engineering, the sterilizer manufacturer, and December 2009 Education & Training the BI manufacturer it was determined that the culprit was a manufacturing defect involving a steam trap. It wasn’t up to specifications which caused inconsistencies in pulling or achieving the required deep vacuums needed to effectively remove air. The part was replaced (not only on ours but on the entire line of the manufacturer’s steam sterilizers) and the intermittent process failures ceased. As a side note, 100 percent of my recalled goods were captured during these events due to my awesome staff following our recall policy. We also re-qualified the steam sterilizer each time it failed following the guidance of the Association for the Advancement of Medical Instrumentation’s Comprehensive guide to steam sterilization and sterility assurance in health care facilities, ANSI/AAMI ST79:2006.1 In our dynamic-air-removal steam sterilizer we ran a BI PCD in three consecutive empty cycles followed by a BowieDick test pack in three consecutive empty cycles. The sterilizer was put back into use if the BIs were negative and the BowieDick test sheets showed a uniform color change. (ANSI/AAMI ST79 section 10.7.5.1) 2 We turned this negative experience into a positive chance to review the entire steam sterilization process and identify areas of opportunity for improvement. The important takeaways were that we needed to set up a more challenging and frequent monitoring regimen thereby increasing our chances of not only detecting sterilization process failures but doing so as soon as possible. The review identified the following improvements. Upgrade Internal Chemical Indicators We were using a Class 4 multi-variable indicator for internal pack monitoring. For biological indicator monitoring our frequency of testing was once daily and with every implant load. The exception being during the time we had problems with the “bad” steam sterilizer, when BIs were used on every cycle. We chose to upgrade to Class 5 integrating indicators because they provide more information about the performance of the steam sterilization process than Class 4 CIs by monitoring all the critical parameters of the steam sterilization process. In addition their performance is designed to be equivalent to or exceed the performance requirements of BIs under ideal steam sterilization conditions. The ANSI/AAMI ST79 guidance for choosing the class of CI is: “An internal CI should be used with each package, tray, or rigid sterilization container system to be sterilized. This internal CI may be a singlevariable indicator (Class 3 CI), multi-variable indicator (Class 4 CI), or integrating indicator (Class 5 CI). The class of CI chosen will depend upon how many critical process variables are to be monitored and how much information is desired about the sterilization process.” (ANSI/AAMI ST79 section 10.5.2.2.2) 2 Today, we use a Class 5 CI inside each package and do not use that package if it has not reached its endpoint. The Class 5 CI has the 3 stated values required to meet the Association for the Advancement of Medical Instrumentation Sterilization of health care products-Chemical indicators-Part 1: General requirements, ANSI/AAMI/ISO 11140-1:2005 CI standard. This standard upgraded the performance requirements of Class 5 CIs by requiring their performance correlate with a BI at three time/temperature relationships (e.g., 121°C/250°F, 132°C/270°F and at least one temperature in between such as 128°C /263°F). The standard also requires that for the stated value at 121ºC/ 250ºF be greater than 16.5 minutes to ensure that chemical indicators labeled for use in a 132°C/270°F cycle do not change too quickly or inappropriately at lower temperatures.3 74 MANAGING INFECTION CONTROL December 2009 Education & Training In our facility a BI/Class 5 CI PCD is used in every load and implants are quarantined until the BI is negative. When the internal chemical indicator detects a problem we follow the guidance of ANSI/AAMI ST79: “If the interpretation of the CI suggests inadequate steam processing, the contents of the package should not be used. The department head or designee in the sterilizing department should then decide whether to recall that sterilized load. The decision should be based on the results of physical monitoring (time and temperature recordings), the results of internal CIs elsewhere in the load, and, if applicable, the results of any PCDs in the load (a PCD containing a BI, a PCD containing a BI and a Class 5 integrating indicator, or a PCD containing a Class 5 integrating indicator). If the results of a PCD containing a BI are not yet available, the remaining packages from the same load should be quarantined and not used until the BI results are obtained.” (ANSI/AAMI ST79 section 10.5.2.2.2) 2 We felt it was important to have the most accurate internal chemical indicator on the market. Since the Class 5 integrating indicator mimics the BI, we have greater assurance that if it passes then the BI will be negative. We have detected mechanical issues by using the Class 5 integrating indicator. Typically we have needed a steam trap replaced because air was not being completely removed from the chamber on all cycles. ANSI/AAMI ST79 guidance says a “PCD containing a Class 5 integrating CI may be used to routinely release loads containing nonimplantable items to assure the adequacy of the sterilization cycle.” (ANSI/AAMI ST79 section 10.5.4) 2 A Class 5 CI PCD cannot be used to release implants. (ANSI/AAMI section 10.5.4) 2 We adopted the use of Class 5 Integrating Indicator PCDs when AAMI provided guidance for use. We decided to run these Class 5 CI PCDs in every load and a BI daily plus implants because we thought it would be more cost effective. When we realized that we would have to recall back to the last negative BI when the Class 5 PCD showed a failure, the savings was not there so we went to every load monitoring with a BI PCD (see Policy and procedure update). Policy and Procedure Update The staff realized that it only made sense to run BIs on every load to reduce the total number of items affected by a positive BI 76 MANAGING INFECTION CONTROL and thereby reduce overall patient risk. The cost was measured against the possibility of one potential lawsuit with our risk management department and it was determined that the every load BI testing was potentially less expensive than one law suit. We also wanted to deliver the same standard of care for our patients. Wouldn’t you want the instrument sets for your surgery procedure tested with a BI? We hold BIs to be the gold standard of testing as they contain living microorganisms. To bridge the gap between the 3-hour rapid readout BI result we would also use a standalone Class 5 CI PCD on every load. Shortly after the rapid readout BI manufacturer introduced the rapid readout BI PCD containing a Class 5 integrating indicator, we had the best of both worlds. We reduced cost and storage space with the new bundled BI/ Class 5 CI PCD. In addition the new ANSI/AAMI ST79:2006 guidance stated that a Class 5 integrating CI should be included in the BI PCD which should be used to monitor every sterilization load containing implants. “A Class 5 CI should be included with the BI in the PCD so that if an implant must be released on an emergency basis, additional information about the critical parameters of the sterilization process will be available and documented.” (ANSI/ AAMI ST79 section 10.6.1) 2 The implant should be quarantined until the BI result is available. (ANSI/AAMI ST79 section 10.6.3) 2 In cases of documented medical exceptions (see ANSI/AAMI Annex L for examples of an implant log and exception form) the implant could be released based on the Class 5 CI result. This event needs to be documented and the information fully traceable to the patient. (ANSI/AAMI ST79 10.6.3) 2 In our facility a BI/Class 5 CI PCD is used in every load and implants are quarantined until the BI is negative. Loads are not released on the results of a Class 5 CI except in defined emergency situations. Nonimplant loads are released based on the Class 5 CI result. This also meets the ANSI/AAMI guidance which states a BI should be used with PCDs for routine sterilizer efficacy monitoring at least weekly, but preferably ever day that the sterilizer is in use. (ANSI/AAMI ST79 section 10.5.3.2) 2 December 2009 Education & Training That was then. This is now. Today Class 6 emulating indicators are available from at least three manufacturers in the U.S. market. Currently there are no ANSI/AAMI recommended practices covering the use of these devices. The new A2:2009 amendments in ANSI/AAMI ST79 only include a definition or acknowledgement of their existence. “Emulating indicators (Class 6) are cycle verification indicators designed to react to all critical variables of specified sterilization cycles, with the stated values having been generated from the critical variables of the specific sterilization process. NOTE 1-This edition of ANSI/AAMI ST79 does not cover the use and application of Class 6 emulating indicators. NOTE 2-See ANSI/AAMI/ISO 15882:2008 for information on the selection, use, and interpretation of chemical indicators.” (ANSI/AAMI ST79 section 10.5.2.1) 2 The Class 6 emulating indicator products have caused a good deal of debate since their introduction, in my opinion primarily due to poor and confusing marketing and a lack of data and education. I have always had a philosophy that I describe as the “Rainbow.” In nature why would we ever want to remove a color from the rainbow? We would much rather add one. The same philosophy holds for monitoring the sterilization process. When there is a sterilization process failure, I want a plethora of products to accurately represent what was occurring in that cycle. Class 6 CIs were marketed as replacements for all other CIs and BIs. The new Holy Grail had been found. I am among many who have not quite determined where these products will play a role (if any) in my arsenal of monitoring the effectiveness of the sterilization process. I am currently doing my own product evaluations and testing. In the testing performed at my facility, my dynamicair-removal sterilizers were capable of killing BIs in a PCD at a one minute exposure at 135ºC with a three phase vacuum cycle. In the same cycle, the Class 5 CIs in a PCD did not reach their end point until three minutes. The current Class 5 CIs and PCDs are correlated to mimic a BI death kill under defined ideal steam sterilization conditions while Class 6 CIs are not. Class 6 CIs in a PCD changed at four minutes. Even though it is a small window of time, the Class 6 CIs took longer to reach their endpoint but is that important since we are not inactivating chemicals on our surgical instruments, we are killing living microorganisms. The ANSI/AAMI definition of an SAL is: “sterility assurance level (SAL):probability of a single viable microorganism occurring on an item after sterilization.” (ANSI/AAMI section 2, definition 2.121) 2 78 MANAGING INFECTION CONTROL ANSI/AAMI ST79 defines sterilization as: “Validated process used to render a product free from viable microorganisms.” (ANSI/AAMI section 2, definition 2.122) 2 ANSI/AAMI ST79 also states: “Biological indicators are the only sterilization process monitoring device that provides a direct measure of the lethality of the process.” (ANSI/ AAMI section 10.5.3.1) 2 Chemical indicators cannot determine or confirm the SAL achieved for a sterilization load. The SAL can only be estimated for completely validated processes. Biological indicators provide the only direct confirmation that a process killed microorganisms. You have to ask yourself how much further do we want to go beyond dead spores. Aren’t dead spores dead already? ANSI/AAMI ST79 has very clear guidance on the usage of BIs. A BI PCD should be used for: Routine sterilizer efficacy testing weekly, preferably each day the sterilizer is in use (section 10.5.3.2); Every implant load, which should be quarantined until the BI testing is available (section 10.5.3.2 and 10.6.3); and Sterilizer qualification testing after the sterilizer installation, relocation, malfunctions, and major repairs and after sterilization process failures, in each type of cycle used on a sterilizer (section 10.5.3.2).2 “Rationale: The use of BIs provides evidence of efficacy by challenging the sterilizer with a large number of highly resistant bacterial spores. Biological monitoring provides the only direct measure of the lethality of a sterilization cycle.” (ANSI/AAMI ST79 section 10.7.1) 2 ANSI/AAMI is very clear that CIs do not replace the use of BIs but are used as external and internal CIs. In addition Class 5 CI PCDs may not be used to monitor implant loads. The Association of periOperative Room Nurses (AORN) have clear recommendations for both the use of chemical and biological indicators in Recommendation XVI of the Recommended Practices for Sterilization in the Perioperative Practice Setting as do the Centers for Disease Control and Prevention (CDC) in Recommendation 16 in the 2008 Guideline for Disinfection and Sterilization in Healthcare Facilities.4,5 Dr. Sheila Murphey of the FDA also is clear about the fact that CIs do not replace the use of BIs. This is stated in an email from her that is posted on the CBSPD Web site on April 11, 2009. December 2009 Education & Training “The Steris Verify SixCess CI is a Chemical Indicator. It may be used as you would use any other Chemical Indicator. However, it is NOT a Biological Indicator and should not be used in place of a BI.”6 What about recalls? If the Class 6 CI PCD shows a failure do you recall back to the last Class 6 CI PCD that showed a pass? The answer is “No” based on the A2:2009 amendments for ANSI/AAMI ST79. All recalls are back to the last negative BI. An updated section addresses what actions to take when a BI PCD or CI PCD (Class 5 CI, Class 6 not included) indicates a sterilization process failure. (ANSI/AAMI ST79 section 10.7.5) 2 This section now includes a decision tree for conducting investigations of steam sterilization continued on page 82 Figure 12. Decision tree for conducting investigations of steam sterilization process failures. Reprinted from ANSI/AAMI ST79:2006 and A1 & A2 with permission of Association for the Advancement of Medical Instrumentation, Inc.© 2009 AAMI www.aami.org. All rights reserved. Further reproduction or distribution prohibited. December 2009 MANAGING INFECTION CONTROL 79 Education & Training Table 8 Checklist for identifying reasons for steam sterilization process failures OPERATOR ERRORS Incorrect use and interpretation of monitoring tools • Incorrect physical monitors for the load • Incorrect use of BI or BI PCD Incorrect selection of BI or BI PCD for the load Incorrect placement of BI PCD in the load (e.g., another pack was placed on top of the PCD) Incorrect incubation of BI Misinterpretation of BI result Incorrect documentation of BI result • Incorrect use of Class 5 integrating CI PCD Incorrect selection of CI PCD for the load Incorrect placement of CI PCD in the load (e.g., another pack was placed on top of the PCD) Misinterpretation of Class 5 integrating CI result Incorrect documentation of Class 5 integrating CI result • Incorrect use of internal CI Incorrect selection of internal CI for the load Misinterpretation of internal CI result Incorrect documentation of internal CI results • Incorrect storage of any CIs or BIs • Failure to check physical monitors for functionality before running cycle • Use of broken media ampoule or ampoule with missing spore strip • Use of BI PCD or CI PCD that is missing the BI or CI • Use of defective CI (e.g., a CI that is expired, faded, shows a partial color change because of incorrect storage, or has been previously exposed to the sterilant) Selection of incorrect cycle for load contents (containment device or medical device manufacturer’s instructions for use not followed) Use of inappropriate packaging materials or packaging technique • Incorrect packaging or containment device for the cycle parameters • Incorrect preparation of containment device for use (e.g., incorrect filters, valves, or bottom tray) • Use of a paper–plastic pouch, woven or nonwoven • • • • • • • • wrapper, or towel in a 270ºF to 275ºF (132ºC to 135ºC) gravity-displacement cycle Use of a tray that does not allow air removal and steam penetration Use of a wrapper that is too large for the application Placement of a folded paper–plastic pouch inside another paper–plastic pouch Placement of a paper–plastic pouch inside a wrapped set or containment device without verification of adequate air removal and steam penetration by product testing Incorrect placement of basins in set (i.e., basins are not aligned in the same direction) Failure to use nonlinting absorbent material between nested basins Preparation of textile packs that are too dense to sterilize with the cycle parameters chosen Inadequate preconditioning of packaging materials (i.e., not holding package materials at 68ºF to 73ºF (20ºC to 23ºC) for 2 hours before use) Incorrect loading of sterilizer • Stacking of containment devices if not recommended by manufacturer • Stacking of perforated instrument trays • Incorrect placement of instrument trays (i.e., not laying instrument trays flat or parallel to the shelf) • Incorrect placement of paper–plastic pouches (e.g., placing pouches flat instead of on edge; not allowing sufficient space between pouches; not placing pouches with plastic sides facing one direction) • Incorrect placement of basins (i.e., not placing basins on their sides so that water can drain) • Incorrect placement of textile packs (i.e., not placing them on edge) • Placement of packages too close together, impeding air removal and sterilant penetration in the load (continued on page 82) 80 MANAGING INFECTION CONTROL December 2009 Education & Training (continued from page 80) Table 8 Checklist for identifying reasons for steam sterilization process failures STERILIZER OR UTILITIES MALFUNCTIONS Poor steam quality or quantity • Wet steam Improper insulation of steam lines Malfunction of trap in steam line or no trap in steam line Malfunction of drain check valve or no drain check valve Steam contact with a cold load Too much water in steam produced at boiler • Superheated steam Improper heatup of chamber Desiccated packaging materials (e.g., towels) Steam pressure too low for the temperature Excessive reduction of steam pressure too close to sterilizer Faulty steam control valve or pressure reducer control valve • Other steam problems Variations in steam pressure because of clogged filter, poorly engineered piping, or excessive demands Out-of-calibration pressure gauges and controllers Clogged steam lines Clogged steam supply strainer Clogged chamber drain line, strainer, or chamber drain screen Malfunction of valves Incomplete air removal • Inadequate vacuum or vacuum depth or other air removal system • Clogged chamber drain line, strainer, or chamber drain screen • Clogged vent lines • Leak caused by faulty door gasket • Leak in other areas of chamber • Plugged, faulty or incorrectly adjusted control valves • Low steam pressure • High water temperature • Inadequate water supply pressure • Clogged water supply strainer • Trapping of air by the load • Incorrect cycle parameters for the load Inadequate cycle temperature • Out-of-calibration temperature gauge • Long heatup time for large loads (i.e., heat lag) • Clogged chamber drain line, strainer, or chamber drain screen • Variations in steam pressure because of clogged filter, poorly engineered piping, or excessive demands on steam supply • Presence of noncondensable gases in steam line and load • Inadequate steam supply pressure • Clogged steam supply strainer Insufficient time at temperature • Out-of-calibration control timer • Inappropriate cycle parameters for the load being processed • Come-up time of less than 1.5 minutes in a 270ºF to 275ºF (132ºC to 135ºC) gravity-displacement cycle • Oversized load Reprinted from ANSI/AAMI ST79:2006 and A1 & A2 with permission of Association for the Advancement of Medical Instrumentation, Inc.© 2009 AAMI www.aami.org. All rights reserved. Further reproduction or distribution prohibited. continued from page 79 process failures (see Figure 12) and a trouble shooting checklist for identifying reasons for steam sterilization process failures (see Table 8).2 Figure 12 says that if the CI PCD fails (see CI failures on first line of decision tree), “Quarantine load, remove sterilizer from service, and investigate cause of failure.”2 Figure 12 goes on to say: “If cause of failure is immediately identified (usually operator error) and confined to one load or one item within the load (internal CI), correct the cause and reprocess the load. If cause of the failure 82 MANAGING INFECTION CONTROL is not immediately identified, quarantine the load and recall all loads back to the last negative BI.”2 This means that if the Class 5 or any other CI PCD (i.e., Class 6 emulating indicator) in the load indicates a failure and the reason for the failure is not immediately identified (usually operator error) then all loads processed since the last negative BI should be recalled. A recall back to the last passed CI PCD is not a recall. Recalls are based on the BI results.2 A CI PCD may assist in identifying a sterilization process failure for a specific load but: December 2009 Education & Training “While the performance of Class 5 integrating CIs has been correlated to the performance of BIs, these sterilization monitoring devices do not contain spores and thus do not directly measure the lethality of a sterilization cycle; however, they provide additional information about the attainment of the critical parameters of the sterilization process.” (ANSI/AAMI ST79 10.5.2.1) 2 This recall requirement would hold true for Class 6 CIs because they also do not contain spores and their performance is not required to correlate to a BI. Do we ignore these guidelines if we are using a Class 6 CI and not recall back to the last negative BI if this CI PCD shows a failure? Class 6 CIs are cycle specific. This is good and bad for Class 6. It is good for monitoring extended cycles if FDA ever chooses to clear those products for use but it is a concern for end users. My staff is very intelligent and very capable It is a slippery legal path to deviate from established practice and guidelines by cherry picking which ones to follow and which ones to ignore. It is a legal homerun for the plaintiff. at Middlesex hospital, the best of the best, but they are apprehensive about having to select specific monitoring products for specific cycles with the large variety of specific cycles we run these days in the department. Although ANSI/AAMI ST79 states “As technology progresses, new sterilization process monitoring devices may be cleared by FDA and become available for use in health care facilities. Health care facilities should rely on the knowledge and expertise of their infection prevention and control, central service and surgical services professionals in the selection and use of process monitoring devices.” (ANSI /AAMI ST79 section 10.4) 2 AAMI cannot simultaneously provide guidelines for every product as soon as it comes to market. They are giving us latitude but that does not mean you can ignore recommended practices such as using a BI to release implants which should be quarantined until the BI result is known. These recommendations are developed to improve patient care and should not be ignored. Don’t ever make decisions in a vacuum or be the lone wolf within what The Joint Commission considers to be your local standard of practice (i.e., facilities within your geographic area). Healthcare facilities must consider the legal ramifications of being a lone wolf when these decisions could potentially be part of a jury trial. The question is “Why are you the only one doing this outside the standard of practice. Who made this decision and why? I can hear the plaintiff asking you why you decided to follow some of the current practices and guidelines but then ignored other portions of it. It is a slippery legal path to deviate from established practice and guidelines by cherry picking which ones to follow and which ones to ignore. It is a legal homerun for the plaintiff. Summary Albert Einstein once said “God does not play dice with the universe” meaning there was a purpose and order for things we see in it. I believe this applies to our field of work. As a myriad of new monitoring products along with their stated uses enter the market they will find a place, a meaning, and a purpose. Recommended practices and standards will be developed by AAMI, AORN, and the CDC to guide us to best practice thereby protecting the people for whom we are all doing our best every day, our patients. References 1. 2. 3. 84 MANAGING INFECTION CONTROL Association for the Advancement of Medical Instrumentation. Comprehensive guide to steam sterilization and sterility assurance in health care facilities. ANSI/AAMI ST79:2006. Association for the Advancement of Medical Instrumentation. Comprehensive guide to steam sterilization and sterility assurance in health care facilities. ANSI/AAMI ST79:2006, A1:2008 and A2:2009. Association for the Advancement of Medical Instrumentation. Sterilization of health care products-Chemical indicators-Part 1: General requirements, ANSI/AAMI/ISO 11140-1:2005. December 2009 Education & Training Sterile Process and Distribution CEU Information CEU Applicant Name _________________________________________________ Address___________________________________________________________ City____________________________ State________ Zip Code ______________ 4. 5. 6. Association of periOperative Registered Nurses. Recommended Practices for Sterilization in the Perioperative Practice Setting. AORN:2009. Centers for Disease Control. Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008. FDA clarification of Class 6 emulating indicators clearance from April 11, 2009. CBSPD website: http://www.sterileprocessing. org/ info.htm#ci David Jagrosse, CSPDT, is the current manager of central sterile service at Middlesex hospital in Middletown, Conn. He has more than 20 years of CSS/SPD experience in the capacities of technician, supervisor and manager. Mr. Jagrosse has served as both a member (06) and chair (07) of the former ASHCSP’s recognition committee and is now an active member of IAHCSMM. He has served IAHCSMM’s Connecticut Central Service Association chapter (www.ctsterile.org) as communications officer (98-00), vice president (06-08) and current president. He is also currently an active voting member of the AAMI standards workgroup that publishes AAMI ST79 (STWG40) that develops guidelines within the CSS/SPD field. Mr. Jagrosse is president of David Jagrosse Consulting LLC, 203.465.8502, which offers public speakers, educational seminars, and CSS/SPD audits. Mr. Jagrosse is a consultant for 3M. ANSWERS 1. 2. 3. 4. 5. A A B A A 6. 7. 8. 9. 10. A B A B A Reprint with permission from Workhorse Publishing L.L.C. 86 MANAGING INFECTION CONTROL The CBSPD (Certification Board for Sterile Processing and Distribution) has preapproved this inservice for 1 contact hour for a period of five (5) years from the date of publication. Successful completion of the lesson and post test must be documented by facility management and those records maintained by the individuals until recertification is required. DO NOT SEND LESSON OR TEST TO CBSPD. For additional information regarding Certification contact: CBSPD, 148 Main St., Lebanon, NJ, 08833 or call 908-236-0530 or 800-555-9765 or visit the Web site at www.sterileprocessing.org. IAHCSMM has awarded 1 Contact Point for completion of this continuing education lesson toward IAHCSMM recertification. Nursing CE Application Form This inservice is approved by the California Board of Registered Nurses, CEP 5770 for one (1) contact hour. This form is valid up to five (5) years from the date of publication. 1. Make a photocopy of this form. 2. Print your name, address and daytime phone number and position/title. 3. Add the last 4 digits of your social security number or your nursing license number. 4. Date the application and sign. 5. Answer the true/false CE questions. KEEP A COPY FOR YOUR RECORDS. 6. Submit this form and the answer sheet to: 3M Sterilization Assurance, Attn HC4160 RR Donnelly Fulfillment Services 585 Hale Ave N., Oakdale, MN 55128-9935 7. For questions, contact craig@manageinfection.com. 8. Participants who score at least 70% will receive a certificate of completion within 30 days of Managing Infection Control’s receipt of the application. Application Please print or type. Name______________________________________________________________ Mailing Address______________________________________________________ City, State, Country, Zip _______________________________________________ Daytime phone ( )__________________________________________ Position/Title_______________________________________________________ Social Security or Nursing License Number ________________________________ Date application submitted _____________________________________________ Signature __________________________________________________________ Offer expires November 2014 On a scale of 1-5, 5 being Excellent and 1 being Poor, please rate this program for the following: 1) Overall content ___________________ 3) Usability of content ________________ 2) Met written objectives ______________ <12/09> Copyright©2009-2010/Workhorse Publishing L.L.C./All Rights Reseved. December 2009