Technical Note: Response measurement for select radiation detectors in magnetic fields M. Reynoldsa) Department of Oncology, Medical Physics Division, University of Alberta, 11560 University Avenue, Edmonton, Alberta T6G 1Z2, Canada B. G. Fallone Department of Medical Physics, Cross Cancer Institute, 11560 University Avenue, Edmonton, Alberta T6G 1Z2, Canada and Departments of Oncology and Physics, University of Alberta, 11560 University Avenue, Edmonton, Alberta T6G 1Z2, Canada S. Rathee Department of Medical Physics, Cross Cancer Institute, 11560 University Avenue, Edmonton, Alberta T6G 1Z2, Canada and Department of Oncology, Medical Physics Division, University of Alberta, 11560 University Avenue, Edmonton, Alberta T6G 1Z2, Canada (Received 20 December 2014; revised 20 April 2015; accepted for publication 21 April 2015; published 15 May 2015) Purpose: Dose response to applied magnetic fields for ion chambers and solid state detectors has been investigated previously for the anticipated use in linear accelerator–magnetic resonance devices. In this investigation, the authors present the measured response of selected radiation detectors when the magnetic field is applied in the same direction as the radiation beam, i.e., a longitudinal magnetic field, to verify previous simulation only data. Methods: The dose response of a PR06C ion chamber, PTW60003 diamond detector, and IBA PFD diode detector is measured in a longitudinal magnetic field. The detectors are irradiated with buildup caps and their long axes either parallel or perpendicular to the incident photon beam. In each case, the magnetic field dose response is reported as the ratio of detector signals with to that without an applied longitudinal magnetic field. The magnetic field dose response for each unique orientation as a function of magnetic field strength was then compared to the previous simulation only studies. Results: The measured dose response of each detector in longitudinal magnetic fields shows no discernable response up to near 0.21 T. This result was expected and matches the previously published simulation only results, showing no appreciable dose response with magnetic field. Conclusions: Low field longitudinal magnetic fields have been shown to have little or no effect on the dose response of the detectors investigated and further lend credibility to previous simulation only studies. C 2015 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4919681] Key words: linac-MR, solid state detectors, ion chambers, longitudinal magnetic field, dosimetry 1. INTRODUCTION The ultimate goal of the advanced real time adaptive radiotherapy (ART2) team at the Cross Cancer Institute (CCI) in Edmonton, Canada, is to develop a hybrid linear accelerator (Linac) and magnetic resonance imager (MRI) to allow for real time imaging and tumour tracking during radiation treatment.1 The ART2 team has developed a rotating biplanar magnet capable of irradiating either parallel (longitudinal magnetic field) or perpendicular (transverse magnetic field) to the static magnetic field of the imager.2 A Linac–MR integrating a solenoidal magnet with a static field perpendicular to a rotation axis of a Linac is also being developed at UMC Utrecht in the Netherlands.3 Both of these prototypes are capable of real time irradiation and tumour tracking via MR imaging. There are previous studies which characterize the response of various radiation detection devices as a function of magnetic field strength, and the relative orientations of magnetic field, photon beam, and detector4–12 for the expressed use in a 2837 Med. Phys. 42 (6), June 2015 Linac–MR environment. These studies focus primarily on ion chambers and solid state detectors and included measurements made exclusively in transverse magnetic fields. Ion chambers are used for reference dosimetry and for the calibration of various radiotherapy machines. Solid state detectors are typically used for relative dosimetry, such as beam profiles, or percent depth doses (PDDs). Solid state detectors also have the advantage of typically being small enough to accurately measure dose in small fields.13,14 Ion chambers and various solid state detectors can be used in conjunction with one another to offer a full range of dosimetry options. This work will attempt to experimentally validate the previously simulated longitudinal magnetic field dose responses of the PR06C ion chamber, PTW60003 diamond detector, and IBA PFD diode detector.4,5 Experimental verification of these results is of importance in ascertaining both the accuracy and correctness of implementation of the Monte Carlo model employing longitudinal magnetic fields. The ratio of the measured detector signals with and without a 0094-2405/2015/42(6)/2837/4/$30.00 © 2015 Am. Assoc. Phys. Med. 2837 2838 Reynolds, Fallone, and Rathee: Response measurement in magnetic fields 2838 static magnetic field under otherwise identical setup and irradiation is compared to the same ratio simulated previously. Close agreement between measured and previously simulated longitudinal magnetic field dose responses would be indicative of correctly implemented modeling within the Monte Carlo code system. We would also gain confidence in the use of these commercially available detectors as is in a Linac–MR environment. 2. MATERIALS AND METHODS 2.A. Description of geometry It is important to note the different relative orientations of photon beam, magnetic field, and radiation detector when discussing magnetic field dose response of radiation detectors. Simulation and measurement work on the magnetic field dose response has been done in situations where the magnetic field is perpendicular to the photon beam, henceforth called transverse magnetic fields.4,5,7,9 The case where the magnetic field is parallel to the photon beam is henceforth referred to as the longitudinal field case. Within transverse and longitudinal magnetic fields, geometrically cylindrical detectors can be independently oriented either parallel or perpendicular to the photon beam direction. The longitudinal field case has been studied previously in simulations for various detectors;4,5 however, measurements have previously not been performed. This work deals with detectors having cylindrically shaped packaging (PR06C, PTW60003, IBA PFD) in longitudinal field orientations only. To adhere to previous work, the orientation in which the detector long axis, magnetic field, and photon beam are all aligned in parallel will be referred to as orientation III.4,5 The case where the detector long axis is perpendicular to the photon beam and magnetic field (photon beam and magnetic field are parallel) will be referred to as orientation IV.4,5 2.B. Measurement setup 2.B.1. Longitudinal field verification measurements The longitudinal field dose response of the PR06C farmer chamber,4 PTW60003 diamond detector,5 and IBA PFD diode detector5 is verified experimentally in this work, details on the physical properties of these three detectors can be found in Table I. Each detector was irradiated in air with a tight fit buildup cap to ensure electronic equilibrium; the buildup caps used are noted in Table I and were either a 1.3 cm PMMA F. 1. Longitudinal electromagnet cross section. The magnetic field is directed upwards, in the same direction as the blue incident photon beam. The green cylinder represents the long axis of the radiation detector (pictured is orientation III). buildup cap or a 0.2 cm brass buildup cap. The previous simulations of these detectors in orientations III and IV also employed buildup caps to ensure electronic equilibrium. The detectors with buildup were irradiated by the 6 MV photon beam of a Varian 23EX accelerator (Varian Medical Systems, Palo Alto, CA). This is the same photon beam used in the measurement of transverse magnetic field dose response.4,5 The simulations employed a previously published 6 MV beam from a Varian accelerator.15 Spectral differences between the measurement and simulation beams are expected to be comparable to those differences from accelerator to accelerator, and of little practical significance. The longitudinal magnetic field was produced by a pair of individual solenoidal electromagnet coils (2 × GMW 11801364 coil, San Carlos, CA) stacked on top of one another (see Fig. 1). The central bore created in this configuration is 26.5 cm in depth, and 17.7 cm in diameter; the outer diameter of the coils is 39.3 cm. This configuration allows for a relatively homogeneous longitudinal magnetic field in the central bore region; this is the region where the detectors will be placed in either orientation III or IV and irradiated. The magnetic field in the bore was measured with a 3D hall probe (Metro Lab THM1176) as a function of applied current and was 0.003 T/A in the central region. The weight of the electromagnet configuration is too substantial to be placed in the accelerator couch; as a result, the magnets reside on a small stand on the floor beneath the accelerator source. The center of the bore of the electromagnet resultantly sits at 182 cm from the accelerator photon source. This distance differs from the T I. Physical properties of radiation detectors studied previously. Detector PR06C PTW60003 IBA PFD Detector packaging diameter (mm) Detector packaging length (mm) Detection volume Wall material Buildup material Buildup thickness (cm) 7 7.3 7.2 24 20 17 0.65 cm3 1.7 mm3 2.45 mm3 C-552 Polystyrene ABS/epoxy/tungsten PMMA Brass Brass 1.3 0.2 0.2 Medical Physics, Vol. 42, No. 6, June 2015 2839 Reynolds, Fallone, and Rathee: Response measurement in magnetic fields F. 2. Measured and simulated relative dose response of the PR06C ion chamber as a function of longitudinal magnetic field strength. F. 4. Measured and simulated relative dose response of the IBA PFD diode detector as a function of longitudinal magnetic field strength. 100 cm distance employed in the simulation studies; however, this change in distance will only have a significant effect on the dose rate seen by the detector. The relative nature of the dose response measurements precludes any correction for this effect. Each of the three detectors was placed in the center of the bore of the electromagnet (see Fig. 1) and measurements were taken in orientations III and IV as the magnetic field strength was varied from 0 to ∼0.2 T. The field sizes at the plane of the detectors’ active volume centroid were 4 × 7 cm for the PR06C, and 2×4 cm for the PTW60003 and IBA PFD detectors. The longitudinal field measurements were in part completed to verify previously simulated results. Thus, the field sizes used for the various detectors are identical to those that were simulated in the previous studies.4,5 The measurements are presented as the ratio of charge collected with an applied longitudinal magnetic field to that collected without a magnetic field. This ratio will allow for the determination of a correction factor—the inverse of the observed response (0.95 for a response of 1.052)—for the associated radiation detector when used in a particular orientation at a specified magnetic field strength. Three 100 MU measurements for each magnetic field strength were taken, and the average of these three measurements was taken as the data point for the associated magnetic field. The stability of the measurements was verified by remeasuring three field strengths (including zero magnetic field) to ensure they were unchanged. Before use, the detectors were preirradiated according to manufacturer recommendations where applicable. The PR06C chamber was used with a 300 V bias, the PTW60003 detector was used with a 100 V bias, and the IBA PFD detector was used without bias as per manufacturer recommendations. 3. RESULTS AND DISCUSSION F. 3. Measured and simulated relative dose response of the PTW60003 diamond detector as a function of longitudinal magnetic field strength. a)Author Medical Physics, Vol. 42, No. 6, June 2015 2839 3.A. Longitudinal field measurements The results of the longitudinal field measurements of the PR06C farmer chamber, PTW60003 diamond detector, and IBA PFD diode detector are presented in Figs. 2, 3, and 4, respectively. Each figure contains both the orientation III and IV measurements at low field longitudinal magnetic fields and previously published simulation data of these detectors in orientations III and IV.4,5 The standard deviation in the simulation data sets was 0.25%, 0.7%, and 0.9% for the PR06C, PTW60003, and IBA PFD, respectively, while the estimated error in the measurements was ∼0.1%–∼0.2% for all three detectors. Both the measurements as well as the simulated ratio of the detectors’ response with and without magnetic field applied during irradiation are close to 1.0. Moreover, it is likely that any observed deviances in the measurements are merely statistical anomalies, as the difference between measurement and simulation is, in all cases, smaller than the combined standard deviation. The measurements for all three detectors suggest that there is no appreciable magnetic field dose response at field strengths up to near 0.21 T in longitudinal magnetic field configurations. Further, these measurements corroborate the previously published low field orientation III and IV simulations, giving credence to the validity of the higher magnetic field simulation results. 4. CONCLUSION Measurements made at low field longitudinal magnetic fields with the PR06C ion chamber, PTW60003 diamond detector, and IBA PFD diode detector all vary very little with magnetic field strength up to the maximum field strength obtainable (about 0.21 T). These measurements are in the near vicinity of the previously simulated results at the same field strengths. This closeness of measurement and simulation lends further confirmation to the accuracy of the previous simulation only studies. It is clear that at low magnetic fields, a correction factor for these investigated detectors is not required, so long as the magnetic field is longitudinal in nature. to whom correspondence should be addressed. Electronic mail: michaelreynolds@ualberta.net 2840 Reynolds, Fallone, and Rathee: Response measurement in magnetic fields 1B. G. Fallone, B. Murray, S. Rathee, T. Stanescu, S. Steciw, S. Vidakovic, E. Blosser, and D. Tymofichuk, “First MR images obtained during megavoltage photon irradiation from a prototype integrated linac-MR system,” Med. Phys. 36(6), 2084–2088 (2009). 2C. Kirkby, T. Stanescu, S. Rathee, M. Carlone, B. Murray, and B. G. Fallone, “Patient dosimetry for hybrid MRI-radiotherapy systems,” Med. Phys. 35(3), 1019–1027 (2008). 3B. W. Raaymakers, J. J. W. Lagendijk, J. Overweg, J. G. M. Kok, A. J. E. Raaijmakers, E. M. Kerkhof, R. W. van der Put, I. Meijsing, S. P. M. Crijns, F. Benedosso, M. van Vulpen, C. H. W. de Graaff, J. Allen, and K. J. Brown, “Integrating a 1.5 T MRI scanner with a 6 MV accelerator: Proof of concept,” Phys. Med. Biol. 54(12), N229–N237 (2009). 4M. Reynolds, B. Fallone, and S. Rathee, “Dose response of selected ion chambers in applied homogeneous transverse and longitudinal magnetic fields,” Med. Phys. 40(4), 042102 (7pp.) (2013). 5M. Reynolds, B. G. Fallone, and S. Rathee, “Dose response of selected solid state detectors in applied homogeneous transverse and longitudinal magnetic fields,” Med. Phys. 41(9), 092103 (12pp.) (2014). 6K. Smit, B. Asselen, J. G. M. Kok, J. J. W. Lagendijk, and B. Raaymakers, “Reference dosimetry for an MRI-Linac the magnetic field,” Med. Phys. 39(6), 4010–4011 (2012). 7K. Smit, B. van Asselen, J. G. M. Kok, A. H. L. Aalbers, J. J. W. Lagendijk, and B. W. Raaymakers, “Towards reference dosimetry for the MR-linac: Magnetic field correction of the ionization chamber reading,” Phys. Med. Biol. 58(17), 5945–5957 (2013). 8K. Smit, J. Sjöholm, J. G. M. Kok, J. J. W. Lagendijk, and B. W. Raaymakers, “Relative dosimetry in a 1.5 T magnetic field: An MR-linac compatible Medical Physics, Vol. 42, No. 6, June 2015 2840 prototype scanning water phantom,” Phys. Med. Biol. 59(15), 4099–4109 (2014). 9I. Meijsing, B. W. Raaymakers, A. J. E. Raaijmakers, J. G. M. Kok, L. Hogeweg, B. Liu, and J. J. W. Lagendijk, “Dosimetry for the MRI accelerator: The impact of a magnetic field on the response of a farmer NE2571 ionization chamber,” Phys. Med. Biol. 54(10), 2993–3002 (2009). 10S. Goddu, O. Pechenaya Green, and S. Mutic, “TG-51 calibration of first commercial MRI-guided IMRT system in the presence of 0.35 Tesla magnetic field,” Med. Phys. 39, 3968 (2012). 11B. Raaymakers, K. Smit, G. H. Bol, M. R. Van den Bosch, S. P. M. Crijns, J. G. M. Kok, and J. J. W. Lagendijk, “Dosimetry in strong magnetic fields; issues and opportunities,” Radiother. Oncol. 103, S170 (2012). 12M. Gargett, B. Oborn, P. Metcalfe, and A. Rosenfeld, “Monte Carlo simulation of the dose response of a novel 2D silicon diode array for use in hybrid MRI-LINAC systems,” Med. Phys. 42(2), 856–865 (2015). 13M. Bucciolini, F. Banci Buonamici, S. Mazzocchi, C. De Angelis, S. Onori, and G. A. P. Cirrone, “Diamond detector versus silicon diode and ion chamber in photon beams of different energy and field size,” Med. Phys. 30(8), 2149–2154 (2003). 14S. N. Rustgi and D. M. D. Frye, “Dosimetric characterization of radiosurgical beams with a diamond detector,” Med. Phys. 22(12), 2117–2121 (1995). 15D. Sheikh-Bagheri and D. W. O. Rogers, “Monte Carlo calculation of nine megavoltage photon beam spectra using the BEAM code,” Med. Phys. 29(3), 391–402 (2002).