ISIS SECOND TARGET STATION PROJECT SANS Beamline Name: SANS2a External Coordinator 1: Coordinator 2: ISIS Contact: Dr Adrian RENNIE Dr Julian EASTOE Dr Richard HEENAN Department of Chemistry School of Chemistry ISIS Facility King’s College London Cantock’s Close Room 1-21, Building R3 The Strand University of Bristol Rutherford Appleton Lab. London, WC2 2LS Bristol, BS8 1TS Didcot, OX11 0QX Tel: 020 7848 2534 Tel: 0117 9289180 Tel: 01235 446744 Fax: 020 7848 2810 Fax: 0117 9250612 Fax: 01235 445720 rennie@colloids.ch.kcl.ac.uk julian.eastoe@bris.ac.uk r.k.heenan@rl.ac.uk Science Case: The main design features of the Second Target Station, of enhanced cold neutron flux, broad spectral range and high resolution, are particularly well suited for exploitation by Small Angle Neutron Scattering, SANS. The gains in performance compared to the existing LOQ instrument will be spectacular, and will provide world leading SANS instrumentation. It will open up new and exciting opportunities in the technologically relevant and emerging areas of Soft Condensed Matter, Advanced Materials, and Biomaterials. SANS is recognised as a powerful technique for determining microstructure in the dimension range of 10 to 5000 Å. The LOQ SANS instrument at ISIS has been highly successful in pioneering the virtues of the white beam time-offlight method, the ability to access a wide range of scattering vectors, Q, (length scales) simultaneously with improved resolution. The experimental programme that has developed on LOQ has resulted in a throughput and publication record that is comparable with world-leading instruments. In the areas of Soft Matter and Advanced Materials there are distinct trends towards the study of complex multicomponent or multi-phase systems, and non-equilibrium systems as well as the use of complex environments. In many such systems the dimension scales of importance range from the molecular to the mesoscale, dictating a need for a wide Q range. Kinetic studies, ranging from chemical reactions to probes of self-assembly, require a broad spectral range and high flux. Multi-component and multi-phase systems are only tractable with the parametric studies possible with high flux and good resolution. Furthermore our ability to study a broad range of biomaterials will be transformed, where complexity, weak scattering from dilute systems, and an emphasis on the longer length scales are all key issues. Hence the main features of SANS on the Second Target Station that are important are: − Using the wide wavelength range available at 10 Hz, the white beam time of flight method will provide a uniquely wide simultaneous Q range. − The optimised grooved coupled cold moderator together with other improvements in neutron optics and detectors will provide overall count rate gains of typically 30 – 40 compared to LOQ, to give world class performance. − Longer flight paths will provide an enhanced low Q performance, and flexibility in the Q range to suit individual experiments. − Good Q resolution compared to reactor instruments. − Improved signal / noise. Given the scientific importance of SANS and the sheer size of the potential user base there is an opportunity for the ISIS second target station, in common with other world class facilities, to provide two SANS instruments from the outset. SANS2a and SANS2b would be optimised for performance at low and intermediate-to-high Q respectively. The SANS2a instrument is optimised to reach smaller Q values, with conventional pinhole collimation, than has previously been achieved at a pulsed neutron source. Its longer collimation length maximises count rates in the SANS2a - Page 1 ISIS SECOND TARGET STATION PROJECT SANS important region between Q = 0.001 and 0.002 Å-1 (d spacings to 5000 Å), but at the expense of a small reduction in neutron wavelength band due to frame overlap and inelastic background considerations with a longer beam line. The general themes of the science case for both instruments are naturally very similar and so will only be given once below. More specific examples are given separately for each instrument. SANS2a is specifically mentioned below in some of the areas of science where it is likely to have a particular advantage over SANS2b. Soft Condensed Matter Soft Condensed Matter encompasses a wide range of molecular materials, from polymers and surfactants, to colloidal suspensions, thermotropic liquid crystals, lyotropic mesophases, and micro-emulsions. All are important in the context of applications in household products, personal care products, lubrication oils, adhesives, protective coatings, drug delivery systems, the development of green solvents, and enhanced oil recovery. The ability to manipulate refractive index, through H / D isotopic substitution, allows complex multi-component or multi-phase systems, to be studied. H/D contrast variation also allows interfacial structures to be highlighted, particularly to obtain polymer density profiles on particles and emulsions droplets, an area which greatly benefits from wide Q range and high flux. The penetrability of cold neutrons makes the use of a range of complex environments (pressure cells, flow cells, rheometers) more tractable. These features coupled with the ability to simultaneously probe a wide range of length scales with substantially enhanced flux will open up exciting new opportunities in the study of Soft Matter and Complex Fluids. Real time experiments on, for example, in-situ synthesis of polymers or colloids requires both high flux and a wide simultaneous Q range. Extensive parametric studies (the need to make measurements over a wide variety of conditions; such as isotopic labelling, temperature, pressure, external force field, pH, concentration, composition and so on) in complex systems will be possible with better resolution, in shorter times, on more dilute samples, and with more detailed isotopic labelling schemes. The ability to measure, say, changes in polymer radii-of-gyration associated with shape changes on short timescales will provide new insights into the mechanisms behind the conversion of chemical energy into mechanical motion in molecular machines. Rapid Screening of the Mesoscale Structure of Novel Multi-Component Materials More complex multi-component systems, often more closely resembling industrially relevant materials, provide the opportunity to investigate the critical role of minor components. SANS2a would be useful for studying phase separation and interactions in colloidal systems involving larger particles and droplets. More complex environments will become possible, more relevant to the conditions faced in different processing routes as, for example, in the extrusion of polymers or the extensional shear of colloidal solutions. The smaller Q on SANS2a is relevant for studies of the conformation and phase separation of high molecular weight polymers and block copolymers. The molecules are large but phase separation can occur on even longer distance scales. Measurements of time-dependent processes will open up poorly explored fields such as the dynamics and kinetics of surfaces, thin films and anisotropic systems of polymeric and surfactant origin including the nanoscale details of complex rheological behaviour. Flow, Deformation and Processing The processing of soft solids in many important industrial processes involves flow, from component mixing in pipe flow, to extensional flow for alignment by extrusion. The flux available on the Second Target Station will make it possible to study the mechanisms controlling these processes at a macromolecular level in realistic conditions and geometries. Measurements of polymer melts with D / H labelling under flow will develop the structure-rheology relationships in polymer processing and address the role of features such as branching, polydispersity and additives. Similar labelling schemes will be exploited in multi-component systems, involving surfactants, polymers and colloidal particles, to provide links between phase behaviour, rheology, and product formulation and stability. The enhanced flux will enable smaller beam sizes to be used, and spatial variations in microstructure and flow properties to be explored. Deformation of the structure of solid polymers during elongation, compression or creep could be studied in real time. SANS2a - Page 2 ISIS SECOND TARGET STATION PROJECT SANS The increased flux would make possible studies of in-plane phase separation within deuterium labelled, spun coated polymer films as thin as 500 Å. Self-Assembly and Colloidal Interactions Self-assembly, involving lyotropic mesophases, micro-emulsions and vesicles of surfactants, lipids or amphiphilic block copolymers, is of interest in a wide variety of fields such as product encapsulation, extraction processes and multi-phase reaction media. The increased flux available on the Second Target Station will enable the dynamic processes associated with structural phase changes, association and dissociation to be studied. For example, the study of coexisting micellar / liposomal phases and their evolution with time, the co-existence of colloidal or micellar / lyotropic phases and their role in phase stability, and polymer-surfactant mixtures. At smaller Q values SANS2a would be very powerful at probing areas such as the dynamics and structure of large self assembled systems like vesicles, and “living polymer” rod-like micelles as well as the interaction of droplets or particles with polymers or surfactants. Many complex systems in these areas are relevant to industrial problems and though light scattering has a role, neutron contrast variation is crucial to obtain the desired information. SANS2a would have a useful overlap in Q with both USANS and light scattering. Aggregation and self-assembly in supercritical fluids, such as CO2, and new ionic liquids, are now important areas for the development of safe alternative solvents in flavour extraction, dry cleaning, controlled polymerisation and potentially many other processes. In-situ pressure cell studies over this broad area will be greatly facilitated as the scattering contrast against CO2 is often weak. Biomolecular Materials The Life Sciences are likely to be the growth area of the future, with many health-related issues underlying this importance. In a broader context of the biological remit, in for example the fields of pharmacy, food science, sensors, biocompatibility and biofunctionality, there is much similarity with problems in Soft Matter, and SANS is poised to make a significant impact. In the last UK Technology Foresight exercise, in the areas of Health and Life Sciences, Food and Drink, Agriculture, and Natural Resources and Environment, priorities in drug preparation and delivery, pharmaceuticals, metabolic pathways, immune manipulation, material changes during processing, and pesticides, are all closely linked to understanding membrane structures, membrane-protein interactions, the structure of macromolecular complexes, aggregation in gels and emulsions, surface adsorption and structure, self-assembly, biocompatibility, and biofunctionality. The enhanced capabilities of SANS, especially SANS2a, on the Second Target Station will greatly expand the contribution that ISIS can make in these areas. Importantly the size and conformation of membrane proteins, which are not easily crystallised, may be examined by SANS, using solvent contrast variation to remove contributions from stabilising surfactants or membrane fragments. A newly emerging area is the development of synthetic artificial biostructured materials (bioengineering) for applications such as bone or tissue replacement. The need to understand the microstructure from molecular to mesoscale and macroscopic dimensions in such materials is central to development in this area, where bioactive and degradable materials are replacing conventional polymers and cements. These complex structures may include drugs, proteins or DNA. Again SANS2a will be useful to probe opaque structures heading towards micron length scales, where other techniques start to become accessible. Macromolecular Assemblies Low resolution studies, using SANS, are already established as an important method for characterising the tertiary conformation and functionality of macromolecular assemblies or complexes, where only the detailed crystalline structure of fragments or components is known, not the overall configuration. The importance of larger complexes, such as the binding of antigens to large antibodies and the link between protein folding / unfolding and the formation of fibrils in amyloid-based disorders (Parkinson’s, HIV, Alzheimer’s) stresses the need for mesoscale structural information. Extending pulsed source SANS to lower Q on SANS2a, whilst maintaining a wide simultaneous Q-range, will provide an exciting opportunity for neutron scattering to probe larger proteins, biomolecular complexes and viral fragments with improved spatial resolution. Deuterium labelling is a key technique in this area which needs the highest possible flux in order to obtain good results in mixed H2O / D2O solvents with small SANS signals and large incoherent backgrounds. SANS2a will also improve understanding of long-range structure or conformation in protein folding problems, materials such as polypeptide tapes, and amyloid plaque formation. Studies of the interaction of macromolecules with smaller entities such as surfactant micelles, lipids, polymers will also benefit from a wide simultaneous Q range and high flux. SANS2a - Page 3 ISIS SECOND TARGET STATION PROJECT SANS Pharmaceuticals In pharmaceutical applications, structural information from SANS may be exploited in the design and optimisation of systems for drug encapsulation and delivery. This again has much in common with many aspects of colloid science and self-assembly. An important area which requires new approaches to drug delivery is that of ‘gene therapy’. SANS has an important role to play in the broader context of drug functionality, in understanding the interactions of drugs or proteins with membrane structures. Many successful drug delivery schemes involve water-soluble polymers. Wide Q range SANS is the ideal method to study not only size but also the density profiles of such systems, with the Second Target Station allowing a wider range of sizes and more extensive parametric studies (changing concentration, temperature, salt etc.). Food Science and Technology Although practical systems in food science are extremely complex, SANS will permit the study of solvent distribution and material changes (denaturing) during processing, and considerable progress can be made using relevant model systems. Understanding kinetics is central to areas such as the control of food colloid stability, foam or gel formation, or protein fouling of filters during processing. The smaller Q’s accessible on SANS2a will help to make a wide range of these in-situ kinetic and structural studies accessible for the first time and help validate existing mesoscale computational models. Advanced Materials The characterisation of Advanced Materials using neutron scattering, including new magnetic materials, ceramics, composites, metal alloys, nanoscale structures, complex materials such as zeolites or templated silicates and novel electrolyte materials, generally requires wide-angle diffraction methods to examine structure at the molecular level. However, many such materials also have hierarchical structures which extend their diffraction patterns into the SANS range. Neutron diffraction from synthetic polymer fibres is currently an under exploited field which will benefit from increases in flux and Q range on SANS2b, as would studies of polymer crystallisation. Several areas would particularly benefit from the smaller Q’s on SANS2a. High flux SANS, coupled in appropriate cases with H / D or other contrast variation, is a powerful technique to study nucleation and growth processes involved in the formation of many advanced materials. These include solid growth from solids (precipitates), solids from liquids (crystallisation), liquids from liquids (phase separation), pre-nucleation, and hydro-thermal synthesis of cements. The ability to accommodate complex sample environments makes it possible to study many of these processes in-situ. There are many questions which SANS2a would help to answer, such as whether in-situ structures are the same as those of dried or calcined samples. In other areas such as ceramic coatings on turbine blades, concretes, metals under creep, and irradiation-induced embrittlement in nuclear plant grade steels, high flux SANS is able to study the formation of cracks and voids. There is potential to greatly expand the use of SANS by the UK engineering materials community, for whom a large sample space to install and position components such as a 200 kg stress rig will be an important feature. Examination of very low concentrations of quite large defects in single crystal silicon, of great importance to the semi-conductor industry, is possible due to the fortuitously low adsorption of neutrons allowing extremely thick samples to be used. Metallurgy SANS has the advantage of averaging over large sample volumes to see low concentrations of small voids, precipitates and defects not easily quantified by conventional methods. Methods of annealing or recrystallisation may be studied insitu. Understanding precipitation hardening in new generations of Ni based superalloys, critical for high temperature turbines, will require both SANS and crystallographic neutron diffraction. Observation of creep cavitation in pressure vessel steels would benefit from any extension of high flux SANS to low Q on SANS2a. When voids are not much larger than micron sized they are strength limiting and the component is no longer reliable. Higher flux would improve spatial resolution and allow in-situ time resolved studies. The interaction of hydrogen with dislocations requires the small Q of SANS2a to see the arrangement of the dislocations coupled with higher Q studies on SANS2b for the trapped hydrogen at dislocation cores, both with very high flux due to weak signals. SANS2a - Page 4 ISIS SECOND TARGET STATION PROJECT SANS Magnetic Materials Neutron scattering has revealed essentially everything that is known about magnetic structures. SANS is able to contribute to investigations of the structures of magnetic domains and the flux line lattice, magnetic colloidal particles, and nano-structured materials with novel magnetic properties. Studies of lateral magnetic correlations and structure in patterned media, proposed as new generation high density recording media, will provide a challenging and significant role for SANS. The increased flux from Target 2 will enable world-class contributions to be made at ISIS. Outline Design Specification: SANS2a Chopper Bender Sample Movable Detector Removable Guides The design for SANS2a proposed here (and summarised in Table 1 below) gives gains of typically 30 – 40 over LOQ, and wider simultaneous Q ranges, coupled with the flexibility of a moving detector to match the overall Q range to a given experiment. The coupled, grooved, cold moderator of the Second Target Station gains a factor of 15-20 in flux per proton at the longer neutron wavelengths crucial for SANS at small Q. The smaller, grooved, moderator is well suited to SANS since it increases count rates at longer sample-detector distances due to the small area of moderator viewed by the pinhole collimation used in SANS. Lower pulse repetition rates and an m=3 (i.e. 3×θcNi ) super-mirror bender will allow a wide band of simultaneous wavelengths, typically 1.5 to 12 Å (cf. 2.2 to 10 Å on LOQ), improving count rates and Q range. Further gains accrue due to larger detectors and improved neutron optics. Combined with a 300 µA proton current these factors more than offset the reduction in repetition rate from 25 to 10Hz to give total count rate gains of from 20 up to several orders of magnitude compared to the present LOQ. To further expand the Q range the movable 1m square multi-wire gas detector may be shifted sideways by at least 300mm. Extended flight paths partly compensate for the longer time spread of a coupled moderator so as to maintain excellent Q resolution compared to reactor sources (except at the smallest Q values where sample size and detector resolution dominate and resolution is always comparable). Typical Q ranges, relative count rates from numerical and Monte-Carlo calculations, and mean resolution data are given in Table 2 and Figure 1. To reach a minimum Q of 0.001 Å-1 a sample-detector distance L2 of around 22m is required. The normal optimisation conditions then suggest a matching incident collimation distance from the neutron source. Situating the sample at 25m from the moderator provides a good compromise with a 48 m long beam line since a possible 100 msec prompt spike does not then overlap important data. With a detector at 47m (L2=22m) SANS2a could with λ=3.0 to 10 Å access Q~0.001 to 0.09 Å-1, and to further enhance the scattering at very small Q, it could use λ=6.75 to 12 Å for Q~0.001 to 0.04 Å-1. The wavelength ranges here are restricted both by frame overlap and the need to avoid inelastic scatter from the previous frame swamping the long wavelength signal. Background suppression by the bender and shielding should be sufficient to avoid having to mask out any prompt spike background at 100 msec, λ=8.4 Å. Experiments requiring a wider Q range and the lowest Qmin on SANS2a could run with a wider λ band at 5 Hz rather than 10 Hz. At a “middle distance” of say L2=12 m SANS2a could use λ=1.5 to 10.7 Å, to access a broader Q range. The sample area will be as large as possible with side access at floor level in order to wheel in sample environment equipment and reduce set up time. An easily adaptable final collimation section and large space will also allow for temporary installation of spectroscopic and other equipment for the application of different techniques simultaneously with SANS. An accurate positioning table to take equipment up to 200 kg in weight will be required. SANS2a - Page 5 ISIS SECOND TARGET STATION PROJECT SANS Table 1 Moderator Coupled cold Incident Wavelengths 1.5 – 14.6 Å at 27 m; 3.0 -10 Å or 6.75 -12 Å at 47 m Q Range -1 Flexible e.g. with detector 300mm offset: ~ 0.001 to 0.09 Å at L2 = 22 m; 0.002 to 0.32 -1 Å at L2 = 12 m L1, L2 Moderator-Sample: 25m, Sample–Detector: 2 to 22 m Flight Path Ni 3θc super-mirror bender; removable, plain Ni guide sections for variable collimation length; evacuated Choppers Variable-phase counter-rotating double-disc (at ~6.5 m, 10 Hz or 5 Hz) Detectors At least 1 m x 1 m position-sensitive, 5 mm pixel size; movable in vacuum tank, with sideways offset Beam Size 6 to 15 mm diameter at sample Detector Tank 22 m long; evacuated; with good shielding Sample Environment Large, side-access sample area to take a wide variety of equipment for in-situ measurements; in air, but with option for evacuated sample position; all standard SE equipment (sample changer, shear cells, magnet, cryostat, etc.) Table 2 Lcoll (m) L2 (m) λmin (Å) λmax (Å) Qmin (Å-1) Qmax1 (Å-1) Qmax2 (Å-1) 2 2 1.5 12 -14.6 0.013 1.38 1.80 4 4 1.5 12 -13.6 0.006 0.71 0.95 12 12 1.5 10.7 0.002 0.24 0.32 18 22 3.0 10.0 0.001 0.066 0.088 Running at 5Hz could increase Qmax at the expense of count rate 18 22 6.75 12.0 0.001 0.029 0.039 More flux at smallest Q from longer λ Comments NOTE: Lcoll is the collimation length. Qmin is calculated for λ=10 Å, though at shorter distances one might routinely use out to 12 Å, and for samples with suitably high transmission, perhaps as far as 14.6 Å (as indicated in the λmax column). Qmax1 is for the diagonal of the square detector when centred on the beam axis, Qmax2 is with the detector offset by 300 mm. At L2=22 m the SANS2a λ range is restricted by both frame overlap and fast inelastic scatter from the previous frame. SANS2a - Page 6 ISIS SECOND TARGET STATION PROJECT SANS Count Rate (arb units) 2m 4m 12m 22m Q(Å-1) Figure 1a Approximate relative count rates for a “flat scatterer” on SANS2a. (upper blue line) L2=2 m, (red line) L2=4 m, (middle blue line) L2=12 m (all with λ=1.5-10.7 Å), (lower blue line) L2=22 m (λ=3-10 Å), and (red +) L2=22 m (λ=6.75-12 Å). Also shown are an approximation to the current LOQ (cyan , using λ=2.2-10 Å), and also a “reactor” instrument with a 10% velocity selector and 10 times time average flux of the ISIS Second Target (i.e. similar to D22 at the ILL); (upper black dashed line) L2=4 m with λ=6 Å and (lower black dashed line) L1=18 m and L2=22 m with λ=10 Å. Except for LOQ all calculations assume a 1m square detector displaced horizontally by 300 mm. Figure 1b The estimated mean Q resolution as the FWHM (full width half maximum) of the distribution. (right blue line) L2=2 m, (red line) L2=4 m, (middle blue line) L2=12 m (all with λ=1.5-10.7 Å), (left blue line) L2=22 m (λ=3–10 Å), and (red +) L2=22 m (λ=6.7512 Å). Also shown are (cyan ) an approximation of the current LOQ (using λ=2.2-10 Å), and a “reactor” instrument with a 10% velocity selector (i.e. an instrument similar to D22 at the ILL); (right black dashed line) L2=4 m with λ=6 Å and (left black dashed line) L1=18 m and L2=22 m with λ=10 Å. The effective resolution on SANS2a at middle Q is actually better than shown due to a sharper peak in the resolution function from long λ values over broad tails from short λ values. SANS2a - Page 7 ISIS SECOND TARGET STATION PROJECT SANS Beamline Name: SANS2b External Coordinator 1: Coordinator 2: ISIS Contact: Dr Adrian RENNIE Dr Julian EASTOE Dr Richard HEENAN Department of Chemistry School of Chemistry ISIS Facility King’s College London Cantock’s Close Room 1-21, Building R3 The Strand University of Bristol Rutherford Appleton Lab. London, WC2 2LS Bristol, BS8 1TS Didcot, OX11 0QX Tel: 020 7848 2534 Tel: 0117 9289180 Tel: 01235 446744 Fax: 020 7848 2810 Fax: 0117 9250612 Fax: 01235 445720 rennie@colloids.ch.kcl.ac.uk julian.eastoe@bris.ac.uk r.k.heenan@rl.ac.uk Science Case: Given the scientific importance of SANS and the sheer size of the potential user base there is an opportunity for the ISIS second target station, in common with other world class facilities, to provide two SANS instruments from the outset. SANS2a and SANS2b would be optimised for performance at low and intermediate-to-high Q respectively. SANS2b represents a new concept for small angle scattering by aiming to maximise either count rate or Q range using two (or more) state-of-the-art 1 m square multi-wire gas detectors. These detectors will be freely movable within a very large rectangular vacuum tank. With a 13 m long vacuum tank an unsurpassed simultaneous Q range of around 0.002 Å-1 to beyond 3 Å-1 could be obtained with one detector at 12 m and the other, offset, at 2 m. The general themes of the scientific case for SANS2b are the same as those for SANS2a and so will not be repeated here. Coupled with the increased flux of Target 2, SANS2b will provide unique opportunities, in a wide range of scientific areas: − With the detectors at different distances, ordering or Bragg peaks may be monitored at high Q simultaneously with SANS at small Q. This is particularly important for polymer, sol / gel and other systems undergoing crystallisation. − Conversely in other systems long-range phase separation may produce ordering or Bragg peaks at small Q whilst changes in more local structure or conformation are followed at high Q. This will be particularly important for studies of polymer processing. − Fibre diffraction requires the widest possible simultaneous Q range to follow all levels of what are often hierarchical structures ranging from intermolecular distances to micron sized voids and phase separation. For example in nylon-6 fibres the lamellar peak is at Q ~ 0.08 Å-1 whilst a series of crystal peaks are between 0.8 and 1.7 Å-1, each of which subtly changes with chemical environment. − Accurately corrected data over a significant Q range will increase confidence in the use of Fourier transform methods of data analysis where direct modelling is not possible, or the use of invariant integrals to estimate phase volumes independent of actual scattering laws. Direct calibration of both high and low angle detectors in the main beam is an important feature of SANS2b which will particularly help these types of analyses. Data inversion is becoming important in detailed studies of copolymers, micelles and other complex solution structures, in addition to its more established use for biomolecular solutions. Invariant integrals have been used for fibre diffraction, particularly by X-rays, but again could be more widely applied. − Higher count rates may be achieved over all but the smallest Q’s by placing the two detectors at similar distances to expand detector solid angle. This would be a particular advantage for anisotropic systems, for example those perturbed by shear or stretch. SANS2b- Page 1 ISIS SECOND TARGET STATION PROJECT SANS Soft Condensed Matter The local interaction of polymers or surfactant micelles with much larger particulate dispersions or emulsion droplets is an important area of study related to many industrial products where neutron contrast variation is used to highlight interfacial structure. A wide Q range and high Q to study local conformation and interactions are essential. Membranes and vesicles require an extremely wide Q range to follow both vesicle diameters at low Q simultaneously with wall thickness, or multi-lamellar Bragg peak scattering, or even details of surfactant molecular conformation at high Q. Studies of clays and vermiculites often require Q ranges spanning the region between conventional small angle and wide-angle diffraction; SANS2b easily provides such a range. Liquid crystals are again anisotropic scatterers that would benefit from an extremely broad Q range: very high Q to study intramolecular conformation simultaneously with intermediate Q for intermolecular ordering and low Q for longrange structure and domain size. Biomolecular Materials Conformational studies of biomolecular complexes require extremely high quality data at high Q to improve the resolution of dummy atom / residue models. Studies of bone and collagen would benefit from simultaneous low and high Q capabilities of SANS2b. Advanced Materials Time resolved studies of the self assembly of surfactant templated inorganic materials, such as zeolites and silicalites, would benefit greatly from being able to see a series of peaks from ordered mesophases at high Q whilst following aggregate size at low Q. The interactions of hydrogen with carbon nanotubes requires high flux, a wide Q range going to high Q and deuterium contrast variation to produce meaningful results, SANS2b would be ideal. Complex systems with hierarchical structures would in general be better resolved, using the advantages of D / H contrast variation, coupled with the high flux and unsurpassed simultaneous Q range of SANS2b. To distinguish between spinodal decomposition or nucleation plus growth mechanisms of real time precipitate formation in say semiconductors or metal matrix composites requires a high flux and extremely wide Q range, since the one process has long range concentration fluctuations whilst the other has short range aggregation. Outline Design Specification: SANS2b Chopper Bender Sample Movable Detectors Removable Guides The initial incident beam for SANS2b is the same as for SANS2a, with a super-mirror bender to reduce fast neutron background, removable guide sections to match the incident collimation length L1 to sample-detector distance L2 but the sample is now at 19 m from the moderator (and not 25 m). The design is summarised in Table 3. Though an arc of fixed “high angle” detectors around the sample position can access an extremely wide Q range out to high Q such approaches are fraught with difficulties. For example such a detector geometry is best suited to cylindrical SANS2b- Page 2 ISIS SECOND TARGET STATION PROJECT SANS and not “flat plate” samples and does not deal easily with anisotropic scattering. More importantly it is extremely difficult, if not impossible, to reliably calibrate fixed detectors close to the sample. SANS2b will have freely movable 1 m square detectors with a closest sample-detector distance of ~2 m, which maintains adequate Q resolution with a ~5mm detector pixel size. The rectangular section vacuum tank would have internal dimensions ~3 m wide x 2.5 m tall x 13 m long. The vacuum tank would be like an inverted ship’s hull, with integral borated wax shielding, bolted to the floor. The two separate 1 m square position-sensitive detectors, with adjustable height, could be positioned anywhere within the vacuum tank to give L2 in the range 2 to 12 m. This could be achieved either by remotely controlled trolleys on a smooth floor or, more likely, with moving “crane gantries” running on rails, which would allow shielding of the floor area. Intermediate baffles will ensure that the detectors only view the sample. Both detectors would be equipped with a movable beam stop to enable each to be calibrated with the same standard sample in the main beam. This would remove most of the calibration difficulties encountered with the fixed “high angle bank” detector approach. The latest large area multi-wire gas detectors have their position encoding electronics in air on the rear of the detector. Thus only simple air flow and return hoses containing a few power and signal leads are required to each detector. For safety in the event of a beam vacuum window breakage fast acting shutters or inflatable air bags (as used in cars) inside the tank will be considered as a means of reducing damage and slowing the air in-rush. A beam vacuum window of about 14 cm diameter at 5 cm from the sample would suffice. Some standard types of sample environment equipment would require modification to enable them to use the full range of scattering angles. Else wise a large area for sample environment access is anticipated, as for SANS2a, with maximum flexibility to adapt to a wide range of sample environment types. The sample would be at 19 m at which the 13º angular separation of adjacent beam lines gives a ~4.4 m space for the front part of the vacuum tank and its shielding, though negotiation with adjacent instruments will be required since the tank is slightly off-centre relative to the beam. SANS2b would allow an even wider simultaneous Q range than SANS2a by combining two different sample-todetector distances (with the collimation length L1 optimised for either count rate at high Q or for Q resolution at small Q to suit the science involved), see Table 4. Alternatively, with the two detectors at similar sample-detector distances, the count rate could be doubled in all but the smaller part of the Q range. With λ=1.5Å the maximum Q in the horizontal direction would be in excess of 2.5 Å-1 and more on the diagonal, allowing good overlap with “wide angle diffraction” and the monitoring of ordering peaks simultaneously with the measurement of the SANS signal. Note that due to using a broad wavelength range a continuous angular coverage is not required for the two detectors to overlap in Q. Although some engineering challenges are presented by this design SANS2b is clearly feasible on the basis of existing technology and would represent a significant step forward from the way SANS is traditionally performed! Table 3 Moderator Coupled cold Incident Wavelengths 1.5 – ~15 Å at 21 m; 1.5 - 12.7 Å at 31 m Q Range Extremely flexible in range 0.002 to >2.5 Å L1, L2 Moderator-Sample: 19 m, Sample–Detector: 2 – 12 m Flight path Ni 3θc super mirror bender; removable, plain Ni guide sections for variable collimation length; evacuated Choppers Variable-phase counter-rotating double-disc (at ~6.5 m, 10 Hz or 5 Hz) Detectors At least two 1 m x 1 m position-sensitive, 5mm pixel size; freely movable horizontally & vertically in vacuum tank Beam size 6 to 15 mm diameter at sample Detector tank Rectangular ~3.5 m (wide) x 2.5 m (high) x 13 m; evacuated; with good shielding Sample environment Large, side-access sample area to take a wide variety of equipment for in-situ measurements; in air, but with option for evacuated sample position; all standard SE equipment (sample changer, shear cells, magnet, cryostat, etc.) -1 SANS2b- Page 3 ISIS SECOND TARGET STATION PROJECT SANS Table 4 Lcoll (m) L2 (m) λmin (Å) λmax (Å) Qmin (Å-1) Qmax1 (Å-1) Qmax2 (Å-1) Comments 2 2 1.5 15 0.013 n/a >2.5 Both detectors at 2 m 12 12 1.5 12 -12.7 0.002 0.24 0.32 Detector 1 at 12 m 2 2 1.5 12 -15 0.15 n/a >2.5 Detector 2 at 2 m Count Rate (arb units) NOTE: Lcoll is the collimation length. Qmin is calculated for λ=10 Å, at 2xRpenumbra, though λ to 12 Å should be routinely used, and for samples with suitably high transmission, perhaps as far as 15 Å (as indicated in the λmax column). Qmax1 is for the corner of a detector centred on the beam axis, Qmax2 is with the centre of Detector 1 offset horizontally (-300,0) mm and Detector 2 offset diagonally (900,900) mm. The maximum offset could be about (1300,900) mm. 2m 12m 2m 12m collim Q(Å-1) Figure 2a Approximate relative count rates on SANS2b, λ= 1.5 12 Å (red lines) one 1 m square detector at L2=12 m, offset 300 mm, the other at L2=2 m, offset (900,900) mm. (Note: this is not the maximum offset, so still higher Q values are possible), and (green {) with both detectors at L2=2 m, dashed lines show the two separate detectors displaced (300,300) and (900,900) mm respectively. Also shown are (cyan ) an approximation of the current LOQ (using λ=2.2-10 Å), and (black dashed line) a “reactor” instrument with a 10% velocity selector and 10 times the time average moderator flux of the ISIS Second Target (i.e. an instrument similar to D22 at the ILL) with L2=4 m and λ=6 Å. Note the improved resolution of the high Q detector at 2 m when using 12m of collimation (but ~36 times less flux). Figure 2b The estimated mean Q resolution as the FWHM (full width half maximum) of the distribution. (red lines) one 1 m square detector at L2=12 m, offset 300 mm, the other at L2=2 m, offset (900,900) mm, and (green {) with both detectors at L2=2 m displaced (300,300) and (900,900) mm respectively. Also shown are (cyan ) an approximation of the current LOQ (using λ=2.210 Å), and (black dashed line) a “reactor” instrument with a 10% velocity selector (i.e. an instrument similar to D22 at the ILL) with L2=4 m and λ=6 Å. The effective resolution on SANS2b at middle Q is better than shown due to a sharper peak in the resolution function from longer λ values over broad tails from short λ values. Sample size here is 10mm, resolution could be improved further at high Q with tighter collimation, at the expense of count rate. SANS2b- Page 4