© Google, 2007 X-am 5100 MONITOR TRAINING Marketing, December 2012 DRÄGER X-AM 5100 DISCLAIMER This visual presentation is intended to help the user understand and correctly use the Dräger X-am 5100 instruments. It is not intended to be a substitute for the complete instruction manual supplied with the instrument. IT IS THE USER’S RESPONSIBILITY TO READ AND UNDERSTAND ALL MANUALS AND INFORMATION PROVIDED WITH THE X-am 5100 BEFORE USE! Pictures may not represent appropriate respiratory equipment and PPE for every situation. Proper protective level should always be determined by the specifics of an event. Draeger Safety is happy to assist you with questions on the use of the X-am 5100 instruments: Draeger Safety Customer Service (USA) toll free @ 800-922-5518 Draeger Safety Technical Service (USA) toll free @ 877-DRAEGER (372-3437) 2 | 48 X-am 5100 | Training| 04.2012 General Information DRÄGER X-AM 5100 X-am 5100 DESCRIPTION The Dräger X-am 5100 Monitor is a single gas measuring and alarm device designed for use in personnel monitoring applications. Use to detect: Hydrogen Chloride Hydrogen Fluoride Hydrazine Hydrogen Peroxide DrägerSensor® patented technology 4 | 48 X-am 5100 | Training | 04.2012 DRÄGER X-AM 5100 ENCLOSURE • Size: 1.89 x 5.12 x 1.89 in. (48 x 130 x 61 mm) • Weight: ~7.8 to 8.8oz (220g to 250g) • IP 54 approved - Dust & water resistant • Over-molded rubber boot offers additional protection • RFI Protected: CE mark electromagnetic compatibility (89/336/EEC) • Temp Range - continuous: -4 to 122 F (-20 to 50 C) • Humidity Range: 10 to 95% RH • Pressure Range: 20.7 to 38.4 in. Hg (700 to 1300 hPa) • ≈ -6,000 to +10,000 feet of elevation 5 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 X-am 5100 DESCRIPTION Datalogging is standard in X-am 5100 • >1,000 hours recording at 1 minute intervals • Events are also logged 6 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 BASICS – COMPONENTS ON FRONT Visual alarm Visual alarm XS sensor Horn > 90 db Dot Matrix display +M – Button Navigation 7 | 48 X-am 5100 | Training| 04.2012 OKon – Button Switch on Acknowledge DRÄGER X-AM 5100 INSTRUMENT COMPONENTS - REAR IR Interface Grip Clip Approval Label Battery Pack Battery Capture Screw 8 | 48 X-am 5100 | Training| 04.2012 Battery Charge Contacts - NiMHy DRÄGER X-AM 5100 X-am 5100 BATTERY OPTIONS • X-am 5100 Units come standard with a NiMH battery pack, p/n 8318704 or ‘AA’ alkaline battery pack using 2 cells • Optional high-capacity NiMH pack is available, p/n 8322244 9 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 BATTERY LIFE • X-am 5100 Standard Operating Times - Diffusion Standard NiMH Battery >120 hrs High-Capacity NiMH Battery >180 hrs “AA” Alkaline Batteries >120 hrs • Operating times are shortened by; Excessive Alarms Colder temperatures (less than 50 ºF / 15 ºC) 10 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 CHARGING CRADLE • The Charging Cradle provides a full, managed charge to the NiMHy battery pack in < 4 hours from complete discharge • Up to 20 modules can be ganged together for use with one 8315805 MultiCharger Power supply • Up to 5 modules can be ganged together for use with 8316994 Power Supply • Up to 2 modules can be ganged together for use with 8315635 Power Supply 11 | 48 X-am 5100 | Training| 04.2012 X-am 5100 Operation DRÄGER X-AM 5100 2 BUTTON OPERATION Operation controlled by a 2-button key pad “OK / On”Button Press and hold to turn monitor ON “+M” Button Press and hold “+M” when monitor is off to verify installed sensors and battery charge Press and hold both “OK” and “+M” to turn monitor OFF 13 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 SWITCHING-ON DRÄGER X-AM 5100 • Press and hold the “OK”-key for 3 seconds • Display counts down 3-2-1 • LEDs Flash, the audible alarm sounds and the vibrating alarm pulsates • The self-test begins automatically and shows the following: -The installed software version -The alarm levels -Days to next calibration 14 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 SWITCHING OFF DRÄGER X-AM 5100 • Press and hold the “+” and “OK" keys at the same time for more than 3 sec. • The LEDs will flash and a beep will sound. • The instrument is switched off 15 | 48 X-am 5100 | Training| 04.2012 Display DRÄGER X-AM 5100 DISPLAY / INFO MODE Display • Large, easy to read numbers • Backlight turns on when a key is pressed or when in an alarm condition • Shows gas, units, reading, and battery level • Indicates alarm conditions or faults • Info Mode • Press the “+M” key when the monitor is OFF and the installed sensor and battery condition is displayed 17 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 DISPLAY LAYOUT The X-am 5100 display is divided into 3 sections 1. Gas Type 2. Gas Values and Concentration units 3. Special Symbols 1 18 | 48 X-am 5100 | Training| 04.2012 2 3 DRÄGER X-AM 5100 DISPLAY – SPECIAL SYMBOLS Special Symbols provide a quick message about the instrument condition Error Notice Peak Battery 19 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 DISPLAY ILLUMINATION • Press any key and the display illumination is switched on for about 10 sec. • In an alarm situation, the display illumination is automatically switched on 20 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 DISPLAY – SPECIAL SYMBOLS The Notice symbol indicates there is something that the user should be aware of. The instrument is still usable!! Check / Service instrument at the end of the shift 1 21 | 48 X-am 5100 | Training| 04.2012 2 3 Alarms DRÄGER X-AM 5100 ALARM INDICATIONS Four Alarm indicators are standard with every model to ensure the customer is aware of potential hazards: A loud audible horn - >90 dB @ 1ft. (30 cm) Three bright flashing, red LED’s An Alarm level indication on the LCD Display A powerful vibrating alarm Distinct alarm cadences allow the user to distinguish between A1, A2, and Fault alarms 23 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 A1 GAS CONCENTRATION ALARM • Audible-, visual- and vibrating alarms repeat once per second • The display character “A1” will alternate with the concentration in the display • An A1 alarm, the audible and vibrating alarms can be acknowledged by pushing the “OK ”-key. 24 | 48 X-am 5100 | Training| 04.2012 A1 DRÄGER X-AM 5100 A2 GAS CONCENTRATION ALARM • Double audible-, visual- and vibrating alarms repeat twice per second 20 • The display character “A2” will alternate with the concentration in the display • The audible, visual and vibrating alarms can NOT be acknowledged (silenced) in “A2” alarm Follow the prescribed safety procedures. 25 | 48 X-am 5100 | Training| 04.2012 A2 DRÄGER X-AM 5100 A1 LOW BATTERY ALARM • Audible-, visual- and vibrating alarms repeat once per second • Icon flashes • The alarm can be acknowledged with “OK”-key This is activated when the battery has less than 10 minutes of operation – please change the batteries or charge the instrument. 26 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 A2 LOW BATTERY ALARM • Audible-, visual- and vibrating alarms repeating two times per second • flashes • The instrument will automatically shut down in about 10 seconds • Instrument switches off • The alarm can NOT be acknowledged 27 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 ERROR ALARM • Audible-, visual- and vibrating alarms triple repeat each second • The icon display -- appears in the • These alarms may be silenced with the “OK” - key, but the gas display will still indicate a fault An error is an indication that something needs to be looked at in the instrument immediately. The instrument should be removed from service and the error corrected before further use 28 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 OVER RANGE • Audible-, visual- and vibrating alarms repeat two times a second • and A2 appear alternately • The alarm can NOT be acknowledged • Sensor reading has exceeded the measuring range If the measuring range is exceeded, the following display is shown instead of measured value display. Follow the prescribed safety procedures. 29 | 48 X-am 5100 | Training| 04.2012 Bump Test DRÄGER X-AM 5100 BUMP TEST • • • • • • • • • 31 | 48 Switch on the instrument. Prepare a test gas source (e.g., cylinder, permeation oven); the volume flow must be 0.5 L/min and the gas concentration must be higher than the alarm threshold concentration to be tested. Fit the calibration adapter (68 06 291) to the sensor cap. Connect the test gas source to the calibration adapter. Vent the test gas into a fume hood or into the open air (with a hose connected to the second connector of the calibration adapter). Open the valve on the test gas source to let test gas flow over the sensor. Recommendation: Wait until the instrument displays the test gas concentration with acceptable tolerance. However, wait at least until alarm threshold A1 or A2 has been exceeded. • If the alarm thresholds are exceeded, the instrument displays the gas concentration alternately with » A1 « or » A2 « depending on the test gas concentration. Close the valve on the test gas source. • If the concentration has now fallen below the A1 alarm threshold: • Acknowledge the alarm. If gas concentration display is not of acceptable tolerance: • Calibrate the instrument, see “Calibrating the instrument” on page 22 of the Technical Manual. X-am 5100 | Training| 04.2012 Test Gases: Sensor Hydrazine HF/HCl H2O2 Test Gas Hydrazine HCl SO2 Contact Draeger for assistance. Charging & Maintenance DRÄGER X-AM 5100 CHARGING THE BATTERY PACK • Slide the instrument into the charging module • Connect the charging module with a single- or multi-charger • The instrument is charging • Red LED blinks once per second -> instrument is charging • Red LED lights stay on -> instrument is fully charged • Red LED blinks twice per second -> charging error • 4 hours are needed for a complete charge when battery is completely discharged 33 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 BATTERY CASE • Loosen the screw with the help of a 2.0 mm Allen key, included with the monitor • Remove the battery pack by sliding it down out of the housing • Install the replacement battery and tighten the screw Do not charge the battery in explosion hazard areas. Alkaline-Batteries are part of the Ex-approval. Only the following types should be used: Energizer No. E91 Energizer No. EN91 (Industrial) Varta Type 4106 (power one) The use of alkaline batteries other than those described above invalidates the intrinsic safety approval for the instrument and could result in unsafe operation. 34 | 48 X-am 5100 | Training| 04.2012 Environmental Conditions DRÄGER X-AM 5100 ENVIRONMENTAL CONDITIONS - TEMPERATURE Temperature • The X-am 5100 is designed for continuous operation between -4 to 122 F (-20 to 50 C) with NiMH battery • Use outside of these parameters is possible, consult X-am 5100 and Sensor specifications for details • Rapid or drastic changes in temperature may cause slight fluctuations of the sensors • At temperatures below 32 F (0 C), the sensors may be slower to respond and battery life will be reduced 36 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 ENVIRONMENTAL CONDITIONS - HUMIDITY Humidity • The X-am 5100 is designed for continuous operation between 5 to 95 %RH • Use outside of these parameters is possible, consult X-am 5100 and sensor specifications for details • Rapid or drastic changes in humidity may cause slight fluctuations on the sensors • Condensing water collecting on the sensor area may inhibit gas from entering the sensors 37 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 ENVIRONMENTAL CONDITIONS - PRESSURE Pressure • The X-am 5100 is designed for continuous operation between 20.7 to 38.4 in. Hg (700 to 1300 mbar) • Rapid or drastic changes in pressure (entering pressurized vessels or large changes in elevation) may cause slight fluctuations of the sensors 38 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 ENVIRONMENTAL CONDITIONS – DUST & MIST Dust & Mist • If dust or mist (including splashing water) is allowed to accumulate on the sensors, it can prevent gas from entering the sensors • Sensor position does not affect the ability to detect gas or sensor specifications • If the X-am 5100 is contaminated with dust or liquids, clean and dry the instrument and calibrate before returning to service • CAUTION Abrasive cleaning implements (brushes etc.), cleaning agents and cleaning solvents can destroy the dust and water filters. • If the instrument is very dirty, clean it with a damp cloth. 39 | 48 X-am 5100 | Training| 04.2012 Review of Basics DRÄGER X-AM 5100 QUICK REVIEW • To turn the X-am 5100 “ON” press and hold the “OK” button for a 3-2-1 countdown • If there is an alarm during use, read the display for information icons and gas values • Take the appropriate safety precautions • To turn the X-am 5100 “OFF”, press and hold both the “OK” and “+M” at the same time and hold for a 3-2-1 countdown 41 | 48 X-am 5100 | Training| 04.2012 DRÄGER X-AM 5100 THANK YOU!! 42 | 48 X-am 5100 | Training| 04.2012