CHAPTER 2 Voltage and Current Measurement 1 Introduction of Electric Circuit The ultimate goal of the circuit theory is to predict currents and voltages in complex circuits (circuit analysis) and to design electrical circuits with desired properties. The circuits are built with circuit elements. Some of these elements are voltmeters, ammeters, wires, resistors, capacitors, inductors, and switches. 2 Ammeters Electrical currents can be measured with an ammeter. To measure the current in the wire shown in Fig. 1a, the wire should be cut and the ammeter should be inserted. The current will flow through the ammeter (Fig. 1b). 3 Cont’…Ammeters An ideal ammeter should have a negligible effect on the circuit. This means that the voltage difference between its two terminals (A and B) should be zero. In other words, the internal resistance (impedance) of an ideal ammeter is zero. 4 Voltmeter To measure voltage, the two terminals of a voltmeter should be connected to two points in the circuit between which the potential difference is measured. An ideal voltmeter should not affect the circuit. Therefore, current through the voltmeter (this is current in Fig.2) should be zero. In other words, internal resistance (impedance) of an ideal voltmeter is infinity. A real voltmeter is never ideal and its impedance is finite. 5 Kirchhoff Laws Kirchhoff laws are applicable to both the linear and not linear circuits. They provide a universal tool for circuit analysis. Kirchhoff ’s Current Law (KCL) The sum of the currents entering a node is equal to the sum of currents leaving the node. A node is a point where two or more wires are interconnected. Kirchhoff ’s Voltage Law (KVL) An algebraic sum of voltages across all elements along any closed path is zero. Algebraic sum means that we should take “+” sign if the voltage rises after a circuit element and “–“ sign if the voltage drops after a circuit element. 6 Cont’…Kirchhoff Laws Analysis of a circuit. General rules: 1. Identify every loop which does not contain another loop (such a loop is called mesh). Assign a current for every loop. The current direction can be chosen arbitrary. This step ensures that the Kirchhoff’s current law is automatically satisfied. 7 2. Use Ohm’s law (or other relations between voltages and currents if the circuit includes capacitors, inductors, diodes, etc) to calculate the voltage across all elements along every mesh and write equations (for every mesh) using Kirchhoff’s voltage law. Important! If two currents flow through an element, the currents should be added like vectors (their directions are important!). 3. Solve the equations. Example 1 Consider the circuit below. KVL @ mesh 1; I1R1 ( I1 I 2 ) R3 V0 0 (1) KVL @ mesh 2; I 2 R2 I 2 R4 ( I 2 I1 ) R3 0 (2) 8 Example 2 Kirchhoff Law when applied to capacitor and inductor usually produce differential equations. KVL dI L IR VC Vs (t ) 0 dt 9 Electrical Indicating Instruments and Measurement Electrical instruments are classified into two (2) 1. Absolute instruments The value of the electrical quantity to be measured are given by these instruments. The quantity are measured in terms of constants and from deflection of the instruments only. Example : Tangent galvanometer. 10 2. Secondary instruments The value of the electrical quantity to be measured is determined from the deflection of these instruments. With an absolute instrument these instruments are calibrated. There are three categories of secondary instruments 1. Indicating instruments 2. Recording instruments 3. Integrating instruments 11 Categories of Secondary Instruments 1. Indicating instruments The value of the electrical quantity is indicated by these instruments at the time when it is being measured. Pointers moving over the scale give the indication. • Examples: Ammeters, volmeters and wattmeter 2. Recording instruments A continuous record of variations of the electrical quantity over a long period of time is given by these instruments. It has a moving system which carries an inked pen which rest tightly on a graph chart. Examples: Graphic recorders and Galvanometer recorders 12 Cont’… 3. Integrating instruments The total amount of either electricity or electrical energy supplied over a period of time is measured by these instruments. Example : Ampere hour meters, watt hour meter, energy meters 13 1.0 Current Measurement 14 Current Basic analog measurement of current –uses inductive force on the current carrying conductor in magnetic field. This force can be used to measure the needle deflection on a display. Direct Current (DC) Charges flow in one direction commonly found in many low-voltage applications, especially where these are powered by batteries Alternating Current (AC) Flow of electric charge changes direction regularly Example: audio & radio signal Home & school use AC 15 Permanent Magnet Moving Coil Instrument (PMMC) There are TWO(2) types of moving coil instruments; Permanent Magnet Moving Coil Dynamometer Type 16 Construction It consists of a permanent magnet with soft iron pieces. The U-shape magnet is widely used. A coil of many turn is wound on aluminium frame. The coil is can move freely move in the field of a permanent magnet. The soft iron coil is mounted between the poles of permanent magnet giving a very narrow gap. The pointer is carried by the spindle and moving over a graduated scale. 17 Cont’d… The control torque is provide by two phosphor bronze hair springs. Eddy current damping is produced by movement of the aluminium former moving in the magnetic fields of permanent magnets. Phospher – bronze springs, pointer, jewel bearings etc. The current is passed into and out of the coil by means of phospher bronze hair springs provided at both ends. The springs also provide the controlling torque. The aluminium frame supports the coil. It also provides a damping torque by the eddy currents induced in it. 18 Operation When the current to be measured is passed to the moving coil, a deflecting torque, Td is produced on account of reaction of the permanent magnetic field with the coil magnetic field. The direction of deflecting torque can be determined by applying Fleming’ Left Hand Rule. The moving system turns through an angle q at which position the tightened control spring produces a back torque Tc equal to Td. The pointer stabilizes at this stage and gives the reading. 19 Advantages It consumes very small power They have no hysteresis losses They have high torque-weight ratio Their scale is uniform They have very effective and efficient eddy current damping 20 Disadvantages Some errors are set in due to ageing of control springs and the permanent magnet. High cost These instruments cannot be used for AC measurement Friction and temperature causes for error 21 Reason for use on DC only The depletion of a moving coil meter depend on polarity of meter connection. For one polarity of connection, the deflecting torque is acting forward. When connection is reversed, the deflection is also reversed. So it is cannot be used for AC measurement but can be used for DC measurement. 22 Dynamometer Type There are two fixed coils F1 and F2 held parallel to each other. They are electrically connected in series. When a current is passed through them, a uniform magnetic field is produced between the two fixed coils. 23 Cont’d… Within this magnetic field a moving coil is placed and support by a spindle and jewel bearings. The spindle carries two control springs that also serve as current leads to the moving coil. Moving coil can be connected either in series or parallel with fixed coil. Series connection – voltmeter Parallel connection – ammeter 24 Advantages It is free from hysteresis and eddy current losses because there is no iron core. It can be used for AC as well as DC measurement. It has a fairly high degree of accuracy. 25 Disadvantages The power loss is high Torque/weight ratio is small Scale is non-uniform Subjected to errors by stray magnetic fields. Error due to mutual induction of coils while measuring AC. It is comparatively more expensive. Friction and temperature causes for error 26 Ammeter An ammeter is an instrument for measuring the electric current in amperes in a branch of an electric circuit. It must be placed in series with the measured branch, and must have very low resistance to avoid significant alteration of the current it is to measure. connecting an ammeter in parallel can damage the meter 27 Ammeter – Principle of Operation The earliest design is the D'Arsonval galvanometer or moving coil ammeter (respond to ac only) It uses magnetic deflection, where current passing through a coil causes the coil to move in a magnetic field The voltage drop across the coil is kept to a minimum to minimize resistance across the ammeter in any circuit into which the it is inserted. Moving iron ammeters use a piece or pieces of iron which 28 move when acted upon by the electromagnetic force of a fixed coil of (usually heavy gauge) wire (which respond to both dc & ac) Ammeter Design 29 An ammeter is placed in series with a circuit element to measure the electric current flow through it. The meter must be designed offer very little resistance to the current so that it does not appreciably change the circuit it is measuring. To accomplish this, a small resistor is placed in parallel with the galvanometer to shunt most of the current around the galvanometer. Its value is chosen so that when the design current flows through the meter it will deflect to its full-scale reading. A galvanometer full-scale current is very small: on the order of milliamperes. 30 Basic DC Ammeter Circuit In most circuits, Ish >> Im Ammeter Where 31 Rsh = resistance of the shunt Rm = internal resistance of the meter movement (resistance of the moving coil) Ish = current through the shunt Im = full-scale deflection current of the meter movement I = full-scale deflection current for the ammeter D’Ársonval meter movement used in ammeter circuit The voltage drop across the meter movement is Vm I m Rm The shunt resistor is parallel with the meter movement, thus the voltage drop for both is equal Then the current through the shunt is, Vsh Vm I sh I I m By using Ohm’s law, Then we can get shunt resistor as Vsh I m Rm I m Im Rsh Rm Rm I I m I sh I sh I sh 32 Example 1.1 Calculate the value of the shunt resistance required to convert a 1-mA meter movement, with a 100-ohm internal resistance, into a 0- to 10-mA ammeter. 33 The Aryton Shunt The purpose of designing the shunt circuit is to allow to measure current I that is some number n times larger than Im. The number n is called a multiplying factor and relates total current and meter current as I nI m (1) We can get shunt resistance with n times larger than Im is Rm Rsh (2) n 1 34 Examples 1.2 A 100 µA meter movement with an internal resistance of 800 Ω is used in a 0 - to 100 mA ammeter. Find the value of the required shunt resistance. 35 Advantages of the Aryton Eliminates the possibility of the meter movement being in the circuit without any shunt resistance. May be used with a wide range of meter movements. Aryton shunt circuit 36 Cont’d... The individual resistance values of the shunts are calculated by starting with the most sensitive range and working toward the least sensitive range The shunt resistance is Rsh Ra Rb Rc On this range the shunt resistance is equal to Rsh and can be computed by Equation 37 Rm Rsh n 1 Cont’... I m ( Rsh Rm ) Rb Rc I2 I m ( Rsh Rm ) Rc I3 Rb ( Rb Rc ) Rc Ra Rsh ( Rb Rc ) 38 Ammeter Insertion Effects Inserting an ammeter in a circuit always increases the resistance of the circuit and reduces the current in the circuit. This error caused by the meter depends on the relationship between the value of resistance in the original circuit and the value of resistance in the ammeter. ** For high range ammeter, the internal resistance in the ammeter is low. ** For low range ammeter, the internal resistance in the ammeter is high. 39 E Ie R1 E Im R1 Rm Fig 2-4: Series circuit with ammeter 40 Cont’d... hence; Im R1 I e R1 Rm Therefore Insertion error = 41 Im 1 100% Ie Example 1.3 A current meter (ammeter) that has an internal resistance of 78 ohms is used to measure the current through resistor Rc in Figure below. Determine the percentage of error of the reading due to ammeter insertion. 42 AMMETER SHUNT An ammeter may be use to measure greater current than that which the instrument itself can carry with the help of shunts. An ammeter shunt is merely low resistance that is placed in parallel with the coil circuit of the instrument in order to measure fairly large current. The circuit diagram for a shunt and milli-ampere meter for measuring large current is shown. 43 Galvanometer It is an electromechanical transducer that produces a rotary deflection, through a limited arc, in response to electric current flowing through its coil. Galvanometer has been applied to devices used in measuring, recording, and positioning equipment. 44 Galvanometer – Principle of Operation Such devices are constructed with a small pivoting coil of wire in the field of a permanent magnet. The coil is attached to a thin pointer that traverses a calibrated scale. A tiny spring pulls the coil and pointer to the zero position. In some meters, the magnetic field acts on a small piece of iron to perform the same effect as a spring. When a direct current (DC) flows through the coil, the coil generates a magnetic field. This field acts with or against the permanent magnet. The coil pivots, pushing against the spring, and moving the pointer. The hand points at a scale indicating the electric current. A useful meter generally contains some provision for damping the mechanical resonance of the moving coil and pointer so that the pointer position smoothly tracks the current without excess vibration. 45 Galvanometer – Application Are used to position the pens of analog chart (example: electrocardiogram) 46 2.0 VOLTAGE MEASUREMENT Voltmeter 47 A voltmeter is an instrument used for measuring the potential difference between two points in an electric circuit. A voltmeter is placed in parallel with a circuit element to measure the voltage drop across it and must be designed to draw very little current from the circuit so that it does not appreciably change the circuit it is measuring. 48 A galvanometer full-scale current is very small: on the order of milliamperes. To accomplish this, a large resistor is placed in series with the galvanometer. Its value is chosen so that the design voltage placed across the meter will cause the meter to deflect to its full-scale reading. 49 To allow meter movement to measure a greater voltage, we need a voltage divider circuit (series resistances) to proportion the total measured voltage into a lesser fraction across the meter movement’s connection points. A resistor is connected in series with the meter movement and it is called a ”multiplier” resistor because it multiplies the working range of the meter movement as it proportionately divides the measured voltage across it. 50 Principle of Voltmeter Operation The moving coil galvanometer is one example of this type of voltmeter. It employs a small coil of fine wire suspended in a strong magnetic field. When an electrical current is applied, the galvanometer's indicator rotates and compresses a small spring. The angular rotation is proportional to the current that is flowing through the coil. For use as a voltmeter, a series resistance is added so that the angular rotation becomes proportional to the applied voltage. 51 D’Ársonval Meter (DC Voltmeter) The basic d’Ársonval meter movement can be converted to a dc voltmeter by connecting a multiplier Rs in series with the meter movement The purpose of the multiplier: to extend the voltage range of the meter to limit current through the d’Arsonval meter movement to a maximum full-scale deflection current. 52 To find the value of the multiplier resistor, first determine the sensitivity, S, of the meter movement. 1 Sensitivit y (/V) I fs Rs S Voltage Range Internal Resistance V Rm Rs I fsd 53 Example 2.1 Calculate the value of the multiplier resistance on the 50V range of a dc voltmeter that used a 500A meter movement with an internal resistance of 1k. V = 50V; Ifsd = 500A ; Rm = 1k. V 50 R m Rs 1k 99k I 500 fsd 54 Multi-range Voltmeter 55 Example 2.2 A D’Arsonval movement with a full scale deflection of 50µA and having an internal resistance of 500 Ω is to be converted into a multi-range voltmeter. Determine the value of multiplier required for 0-20 V, 0 – 50 V and 0100V using individual multipliers for each range. Calculate the values of the individual resistor. 56 Multiplier in series arrangement R4 is low range multiplier with special manufactured to meet circuit requirement 57 Example 2.3 Convert a basic D’Arsonval movement with a full scale deflection of 10mA and having an internal resistance of 100 Ω into a multirange voltmeter with ranges from 0-1V, 0 – 10 V and 0 - 100V 58 Voltmeter Impact On Measured Circuit A perfect voltmeter has infinite resistance, so that it draws no current from the circuit under test. However, perfect voltmeters only exist in the pages of textbooks, not in real life! 59 Voltmeter Loading Effects When a voltmeter is used to measure the voltage across a circuit component, the voltmeter circuit itself is in parallel with the circuit component. Since the parallel combination of two resistors is less than either resistor alone, the resistance seen by the source is less with the voltmeter connected. Therefore, the voltage across the component is less whenever the voltmeter is connected. The decrease in voltage may be negligible or it may be appreciable, depending on the sensitivity of the voltmeter being used. This effect is called voltmeter loading. The resulting error is called a loading error. 60 Example 2.4 Two different voltmeters are used to measure the voltage across resistor RB in the circuit. The meters are as follows. Meter A: S = 1k/V, Rm = 0.2k, range = 10V Meter B: S = 20k/V, Rm = 1.5k, range =10V Calculate: (a) Voltage across RB without any meter connected across it. (b) Voltage across RB when meter A is used. (c) Voltage across RB when meter B is used (d) Error in voltmeter readings. 61 Solution (a) The voltage across resistor RB without either meter connected is found Using the voltage divider equation: RB VRB E R R B A 5kΩ 30V 5V 25k 5k 62 (b) starting with meter A, the total resistance it presents to the circuit is RTA S Range 1k/V 10V 10kΩ The parallel combination of RB and meter A is RB RTA 5kΩ 10kΩ Re1 3.33kΩ RB RTA 5kΩ 10kΩ Therefore, the voltage reading obtained with meter A, determined by the voltage divider equation, is VRB 63 Re1 3.33kΩ E 30V 3.53V 3.33kΩ 25kΩ Re1 RA (c) The total resistance that meter B presents to the circuit is RTB S Range 20k/V 10V 200kΩ The parallel combination of RB and meter B is RB RTB 5kΩ 200kΩ Re 2 4.88kΩ RB RTB 5kΩ 200kΩ Therefore, the voltage reading obtained with meter B, determined by use of the voltage divider equation, is Re 2 4.88kΩ VRB E 30V 4.9V 4.88kΩ 25kΩ Re 2 RA 64 (d) (Expected value - M easured value) Voltmeter A error 100% Expected value (5 - 3.53) Voltmeter A error 100% 29.4% 5 (5 - 4.9) Voltmeter B error 100% 2% 5 65 Five principal meter movements used in ac instrument 66 1. Electrodynamometer 2. Iron Vane 3. Electrostatic 4. Thermocouple 5. D’Arsonval with rectifier Application of meter movements 67 Meter Movement DC Use AC Use Applications Electrodynamometer YES YES Standards meter, wattmeter, frequency meter “Indicator” applications such as in automobiles Iron Vane YES YES “Indicator” applications such as in automobiles Electrostatic YES YES Measurement of high voltage when very little current can be supplied by the circuit being measured Thermocouple YES YES Measurement of radio frequency ac signal D’Arsonval YES YES with rectifier Most widely used meter movement for measuring direct current or voltage and resistance PMMC Instrument on AC The PMMC instrument is polarized (terminals +ve & -ve) - it must be connected correctly for positive (on scale) deflection to occur. When an AC with a very low frequency is passed through a PMMC, the pointer tends to follow the instantaneous level of the AC. As the current grows positively, the pointer deflection increases to a maximum at the peak of the AC. As the instantaneous current level falls, the pointer deflection decreases toward zero. When the AC goes negative, the pointer deflected (off scale) to the left of zero This kind of pointer movement can occur only with AC having a 68 frequency of perhaps 0.1Hz or lower • At 50Hz or higher supply frequencies - the damping mechanism of the instrument and the inertia of the meter movement prevent the pointer from following the changing instantaneous levels. • The average value of purely sinusoidal AC is zero. • Therefore, a PMMC instrument connected directly to measure 50Hz AC indicates zero average value. • • It is important to note that although a PMMC instrument connected to an ac supply may indicating zero, there can actually be very large rms current flowing in its coils 69 Two types of PMMC meter used in AC measurement 1. Half wave rectification 2. Full wave rectification Voltmeters using half wave and full wave rectification are suitable for measuring only sinusoidal ac voltages. 70 D’Arsonval meter movement used with half wave rectification To convert alternating current (AC) to unidirectional current flow, which produces positive deflection when passed through a PMMC, the diode rectifier is used. Several types of rectifiers are selected such as a copper oxide rectifier, a vacuum diode, or semiconductor or “crystal diode”. VP Vrms 2 0.707Vp Vave V dc 0.318Vp Vave 71 Vp 2 Vrms 0.45Vrms • For example, if the output voltage from a half wave rectifier is 10Vrms so the dc voltmeter will provide an indication of approximately 4.5V dc Therefore, the pointer deflected full scale when 10V dc signal is applied. • When we apply a 10Vrms sinusoidal AC waveform, the pointer will deflect to 4.5V This means that the AC voltmeter is not as sensitive as DC voltmeter. • In fact, an AC voltmeter using half wave rectification is only approximately 45% as sensitive as a dc voltmeter. 72 Actually, the circuit would probably be designed for full- scale deflection with a 10V rms AC applied, which means the multiplier resistor would be only 45% of the value of the multiplier resistor for 10V dc voltmeter. Since we have seen that the equivalent dc voltage is equal to 45% of the rms value of the ac voltage. Edc 0.45Erms Rs Rm Rm I dc I dc Sac = 0.45Sdc 73 Half-wave Rectifier Voltmeter Commercially produced An ac voltmeters that use half wave rectification has an additional diode and a shunt as shown in Figure below: 74 • The additional diode D2 is reverse biased on the positive half cycle and has virtually no effect on the behavior of the circuit. • In the negative half cycle, D2 is forward biased and provides an alternate path for reverse biased leakage current that would normally through the meter movement and diode D1. • The purpose of the shunt resistor Rsh is to increase the current flow through D1 during positive half cycle so that the diode is operating in a more linear portion of its characteristic curve. • Although this shunt resistor improves the linearity of the meter on its low voltage ac ranges, it also further reduces the AC sensitivity. 75 Example 2.5 Compute the value of the multiplier resistor for a 15Vrms ac range on the voltmeter shown in Fig. 1. RS Ifs = 1mA Ein = 15Vrms Fig. 1: AC voltmeter using half wave rectification 76 Rm = 300Ω Solution Method 1 The sensitivity of the meter movement, 1 1 Sdc 1k / V I fs 1m Rs = Sdc × Rangedc – Rm = 1kΩ/V × 0.45Erms - Rm = 1k × 0.45(15) – 300 = 6.45k 77 Cont’d… Method 2 The AC sensitivity for half wave rectifier, Sac = 0.45Sdc = 0.45(1k) = 450/V Rs = Sac × Rangeac – Rm = 450 × 15 – 300 = 6.45k 78 Cont’d… Method 3 Rs = = 0.45E rms Rm I fs 0.45 15 300 1m = 6.45k 79 Example 2.6 Calculate the value of the multiplier resistor required to limit the full scale deflection current in the circuit shown in Fig above and the AC and DC sensitivity. 80 D’Arsonval meter movement used with full wave rectification Fig. 2: Full bridge rectifier used in an ac voltmeter circuit 81 During the positive half cycle, currents flows through diode D2, through the meter movement from positive to negative, and through diode D3. The polarities in circles on the transformer secondary are for the positive half cycle. Since current flows through the meter movement on both half cycles, we can expect the deflection of the pointer to be greater than with the half wave cycle, which allows current to flow only on every other half cycle; if the deflection remains the same, the instrument using full wave rectification will have a greater sensitivity. 82 Cont’d… When the 10Vrms of AC signal is applied to the circuit above, where the peak value of the AC input signal is E p 2xE rms 1.414x(10) 14.14V And the average full wave output signal is E ave E dc 0.636xE p 0.636x14.14 9V Therefore, we can see that a 10Vrms voltage is equivalent to 9Vdc for full-scale deflection. 83 Example 2.7 Compute the value of the multiplier resistor for a 10Vrms ac range on the voltmeter in Figure 1-2. Fig. 1-2: AC voltmeter circuit using full wave rectification 84 Solution The dc sensitivity is 1 1 Sdc 1k / V I fs 1mA The ac sensitivity is Sac = 0.9Sdc = 0.9 (1k) = 900 /V 85 Cont’d… Therefore the multiplier resistor is Rs = Sac x Range – Rm = 900 x 10Vrms – 500 = 8.5k 86 Oscilloscope An oscilloscope is a piece of electronic test equipment that allows 87 signal voltages to be viewed, usually as a two-dimensional graph of one or more electrical potential differences (vertical axis) plotted as a function of time or of some other voltage (horizontal axis) Perform some computations using data taken from the voltage waveform that is displayed such as: • rms value • Average Amplitude • Peak-to-peak Amplitude • Frequency An oscilloscope is easily the most useful instrument available for testing circuits because it allows you to see the signals at different points in the circuit. Using for signal/wave display – Winamp Music Player, electrocardiogram, Potentiometer A potentiometer is a variable resistor that functions as a voltage 88 divider It is a simple electro-mechanical transducer It converts rotary or linear motion from the operator into a change of resistance, and this change is (or can be) used to control any volume. Schematic symbol for a potentiometer. The arrow represents the moving terminal, called the wiper. Usually, this is a three-terminal resistor with a sliding contact in the center (the wiper) - user-adjustable resistance If all three terminals are used, it can act as a variable voltage divider If only two terminals are used (one side and the wiper), it acts as a variable resistor Potentiometer Circuit Any current flow through the Galvanometer, G, would be a result of an imbalance in the measured voltage, Vm and the voltage imposed across points A to B, VAB. If Vm is not equal to VAB, a current will flow through the galvanometer, G. Galvanometer detects current flow due to imbalance in voltage Vm and VAB. When Vm = VAB, there is a balance and no current, means no displacement in Galvanometer. 89 Application of Potentiometer In modern usage, a potentiometer is a potential divider, a three terminal resistor where the position of the sliding connection is user adjustable via a knob or slider. For instance, when attached to a volume control, the knob can also function as an on/off switch at the lowest volume Potentiometers are frequently used to adjust the level 90 of analog signals (e.g. volume controls on audio equipment) and as control inputs for electronic circuits (e.g. a typical domestic light dimmer). 3.0 RESISTANCE MEASUREMENT 91 Resistance Measurement The resistances are classified as ; 1. Low Resistance : All resistances of the order of 1 ohm and below. Example: Machine armature, series field winding shunt etc. 2. Medium Resistance : All resistances of the order of 1 ohm to 100,000 ohms. Example: Winding resistance, multiplier resistance. 3. High Resistance : All resistances of the order of 100,000 ohm and above. Example: Insulation resistance of machines, cables, porcelain insulator etc. 92 Ohmmeter The purpose of an ohmmeter, is to measure the resistance placed between its leads. This resistance reading is indicated through a mechanical meter movement which operates on electric current. The ohmmeter must then have an internal source of voltage to create the necessary current to operate the movement, and also have appropriate ranging resistors to allow just the right amount of current through the movement at any given resistance. 93 The original design of an ohmmeter provided a small battery to apply a voltage to a resistance. It used a galvanometer to measure the electric current through the resistance. The scale of the galvanometer was marked in ohms, because the fixed voltage from the battery assured that as resistance decreased, the current through the meter would increase. A more accurate type of ohmmeter has an electronic circuit that passes a constant current I through the resistance, and another circuit that measures the voltage V across the resistance. 94 The standard way to measure resistance in ohms is to supply a constant voltage to the resistance and measure the current through it. That current is of course inversely proportional to the resistance according to Ohm's law, so that you have a nonlinear scale. The current registered by the current sensing element is proportional to 1/R, so that a large current implies a small resistance. 95 Series Type Ohmmeter It contains galvanometer connected in parallel with shunting resistor R2. This parallel circuit is in series with a resistance R1 and a battery of EMF, E. This series circuit is connected to the terminals A and B of the unknown resistance R R1 current limiting resistor R2 zero adjusting resistor Rm internal resistance of galvanometer 96 Operation Supply is given through the battery B and voltage E is supplied. If terminal A and B are short circuited by shunting the resistance R, maximum power flows through the meter. Under this condition, resistor R2 is adjusted until the pointer indicates full scale current Ifs. The full scale current position of the pointer is marked ‘0’ Ω on the scale. When R is removed from the circuit, the terminals are open, the movement indicates zero, which is marked infinite ‘ ’ for zero current. Intermediate scale markings may be placed on the scale by different known values of resistance ‘R’ in the instrument. The accuracy of these scale markings depends on the movement and the tolerance of the calibrating resistors. 97 Although the series type ohmmeter is a popular design and it is used extensively is portable instrument for general service work. The important disadvantages of ohmmeter is internal battery, whose voltage decreases gradually with time and age. So the full scale drops down and meter does not read zero when the terminals are short circuited. Rectified AC given to some ohmmeter is the power source. In this case, the voltage must be regulated and given. 98 Shunt Type Ohmmeter It consists of a battery in series with an adjustable resistor R1 and a basic d’Arsonval movement (meter) The unknown resistance is connected across terminals A and B, parallel with the meter. In this circuit it is necessary to have an ON-OFF switch to disconnect the battery from the circuit when the instrument is not in use. When the unknown resistor, Rx = 0 (shorted), the meter current is zero. If the unknown resistor Rx = (open), current find path only through the meter and selecting a proper value for resistance R1, the pointer may be made to read full scale. LHS( 0 = no current) and RHS ( = full scale deflection current ) 99 Megger The megger is an instrument used for the measurement high resistance and insulation resistance. 100 Construction There are two pairs of permanent magnets, one set for ohmmeter 101 and the other for the generator. The moving coil consists of three coils namely control coil, deflecting coil and compensating coil. The control coil and deflecting coil are fixed at right angles to one another and free to move on a stationary C shaped iron core. The compensating coil in series with a control coil and protection resistance R is connected across the generator terminals. The coils are connected to the circuits system through flexible leads. These leads do not exert any force on the moving system at anytime which will therefore take up any position when the generator is not driven. A resistor R is connected in series with deflecting coil to protect the deflecting coil under short circuit condition. The guard ring bypasses the leakage current, if any to negative terminal of the generation and prevents leakage current from entering the deflecting coil. Operation The unknown resistance is connected between the terminal L (line) & E (earth). The generator handle is then steadily turned at uniform speed. There is a slip mechanism in the drive which ensures a limited speed. When the resistance value is small, the current through the deflecting coil will be high, its deflecting torque will be very high and hence the pointer will move to the extreme clock wise position ‘0’ or very low resistance value. When the resistance value is high, the current through the current coil will be high, its deflecting torque will be very high and hence the pointer will be taken to the extreme anticlockwise position indicating infinity or very high resistance value. 102 Multimeter A multimeter or a multitester is an electronic measuring instrument that combines several functions in one unit. Multimeter is basically a PMMC meter. The most basic instruments include an ammeter, voltmeter, and ohmmeter with a function switch. 103 DC Ammeter Section DC currents are measured making use a suitably designed shunt resistors. 105 Multi-range Ohm-meter Multi-range ohm-meter is built with the meter movement, battery cells, shunt and series resistors 106 DC Voltmeter Section The meter movement has a resistance of 2000 ohms. Suitable resistor are added as multiplier to get voltage range from 2.5V to 250V. 107 Merits/Demerits of Multimeter Merits It is single meter that performs several measuring functions. It has a small size and portable It can made measurements with reasonable accuracy Demerits It cannot make precise and accurate measurements due to the loading effect. Technical skill is required to handle it 108 AC Voltmeter Section To measure AC voltage the output voltage is rectified before the current passes through the meter using half wave rectifier. 109 Multimeter – Capabilities DC Voltage Measurements AC Voltage RMS Measurements DC and AC Current Measurements Resistance Measurements Capacitance/Inductance Measurements Frequency/Period Measurements Diode Measurements 110 - End – be continue…Chapter 3 111