METRAVI XB-32 DIGITAL MULTIMETER GENERAL SPECIFICATIONS Confirms to Pollution degree Altitude Operating temperature : : : : l Storage temperature : l Temperature Coefficient : MAX. Voltage between terminals and earth ground Fuse Protection : : Display Over Range indication Low battery indication Polarity indication Power source Battery type Dimensions Weight : : : : : : : : l l l l l l l l l l l l l l 600V CAT.III and 1000V CAT.II 2 < 2000 m. 0~40ºC (32ºF to 104ºF), (<80% RH, non-condensing) -10~50 ºC (14ºF to 122ºF), (<70% RH, battery removed) 0.1X (specified accuracy) / ºC (<18ºC or >28ºC) 750V AC rms or 1000V DC mA :F 200mA/250V Æ5X20, 10A: unfused. LCD, 1999counts, updates 2-3/sec. LCD will display "1". The " " is displayed on the LCD. "-" displayed automatically. 4.5V AAA 1.5V 158(L)´74(W)´31(H) mm. 220g. Approx. (battery included). TECHNICAL SPECIFICATIONS Resistance Range 200W 2kW 20kW 200kW 2MW 20MW Resolution 0.1W 1W 10W 100W 1kW 10kW Accuracy ±(0.8% of rdg+3 digits) Accuracy : ±(% of reading + number of digits) at 18ºC DC Current Resolution Accuracy to 28ºC (64ºF to 82ºF) with relative humidity Range ±(0.8% of rdg+1 digit) 80%. 200mA 0.1mA (Accuracy is specified for a period one year after calibration). 2mA 1mA ±(0.8% of rdg+1 digit) ±(1.0% of rdg+2 digits) DC Voltage 20mA 10mA Overload protection: 250V DC or 250V AC rms. Range Resolution Accuracy 200mA 0.1mA ±(1.2% of rdg+1 digit) 200mV 0.1mV 10A 10mA ±(2.0% of rdg+5 digits) Diode and Audible continuity Test 2V 1mV ±(0.5% of rdg + 1 digits) l Overload protection Range description Test Condition : F200mA/250V fuse 20V 10mV (10A range unfused). Built-in buzzer Open Circuit Voltage l Max. input current sounds if resistance approx. 2.8Volts 200V 100mV : mA: 200mA DC or 200mA AC rms is less than l 10A 1000V 1V ±(0.8% of rdg + 2 digits) : 10A continuous, 20A 15 sec. MAX. approx. 50W l Input Impedance : 1MW Display read approx. Forward DC current AC Current l Max. input voltage forward voltage approx. 1mA. : 250VDC or AC rms for 200mV Range Resolution Accuracy of diode Reversed DC voltage range and 1000VDC or peak AC 2mA 1mA ±(1.2% of rdg+3 digits) approx. 2.8Volts. for other ranges. 20mA 10mA Overload protection: 250V DC or 250V AC rms. AC Voltage 200mA 0.1mA ±(2.0% of rdg+3 digits) Transistor Range Resolution Accuracy 10A 10mA ±(3.0% of rdg+7 digits) Range Description Test Condition 2V 1mV l Overload protection : F200mA/250V fuse hFE Display read approx. Base Current approx. 20V 10mV ±(0.8% of rdg +3 digits) (10A range unfused). HFE value (0-1000) 10µA, Vce approx. 200V 100mV of transistor under 2.8V. l Max. input current : mA: 200mA DC or 200mA AC rms test (all type). 750V 1V ±(1.2% of rdg +3 digits) l 10A : 10A continuous, 20A 15 sec. MAX l Frequency Range l Input impedance : 40Hz-1kHz; : 10MW OPTIONAL ACESSIORIES: l Indication l Max. input voltage : Average (rms of sine wave) : 250VDC or AC rms for 200mV Infra-red non contact temperature adaptor to measure temperature range and 1000VDC or 750V AC from -30°C to 550°C with fixed emissivity of 0.95 and distance to rms for other ranges, sighting ratio of 8:1. Model TS-03 l Frequency Range Clamp on Current Adaptor to measure current up to 1000A. Model : 40Hz-1kHz; CA-1000D l Indication : Average (rms of sine wave) l l l l Clamp on Current Adaptor to measure upto 200A AC. Model CA-01 *Technical Specifications & Appearance are subject to change without prior notice THE QUALITY LEADER www.metravi.com