ELEC 103 LABORATORY EXERCISE 1 USE OF THE MULTIMETER PURPOSE OF EXPERIMENT The purpose of this experiment is to determine the resistance of two circuit components, light bulb #1 and light bulb #2. In this experiment, you will determine the resistance of the light bulb's by measuring the current flowing through each bulb and the voltage drop across each bulb. Then Ohm's Law will be used to calculate the resistance of each bulb by dividing the voltage measured across the light bulb by the current measured flowing through the light bulb. EQUIPMENT LIST 1 − DC Power Supply 2 − 6V Light Bulb's 1 − Simpson Model 260 VOM PROCEDURE − PART 1 1. Carefully examine the Simpson Model 260 multimeter and note the location and configuration of the following controls and scales: RANGE/FUNCTION SWITCH is located in the center of the control panel. It has 12 positions including five positions for voltage. They are oriented to the left. Four positions for current are located across the top and upper right. And three positions for resistance located to the right. STATUS/POLARITY SWITCH is located on the left side of the control panel. This switch usually has an off or transport position, a +DC position, a −DC position, and an AC position. ZERO OHMS ADJUST KNOB is located on the right side of the control panel. This knob is used to adjust the meter pointer to zero ohms on the resistance ranges when the meter leads are connected together. DC SCALE is the second scale down from the top under the glassed portion of the multimeter. It is sometimes located directly under the parallax mirror. This scale is used for all DC voltage and current ranges. The scale is divided into 50 divisions and is similar to the scale used in the figures during your Unit One learning experiences. When using the 2.5 volt range switch position and when using the 500 volt range/function switch position, you must mentally move the decimal point to the appropriate position. On all other range switch positions the numbers on the scale are appropriate. COMMON (−) AND PLUS (+) RECEPTACLES are located in the lower left corner of the control panel. Always make certain the black lead is connected to the minus or negative receptacle and the red lead is connected to the plus or positive receptacle. Page 1 ELEC 103 LABORATORY EXERCISE 1 USE OF THE MULTIMETER PROCEDURE − PART 2 NOTE: It is not always possible to determine the approximate value of the current flowing in a circuit before you measure it with the milliammeter. Therefore, when the milliammeter is connected in series with the element you should start with the range switch adjusted to the highest current measuring range. After power has been applied to the circuit, adjust the range switch to lower current ranges until you obtain an appropriate current measurement. If the meter pointer moves to the left, reverse the meter leads and continue with the experiment. 1. Connect the circuit as shown in Figure 1−1. Be certain to connect the milliammeter to the circuit so it is connected in series with the circuit. Additionally, connect the meter to the circuit observing correct polarity. The negative lead of the milliammeter is connected toward the negative or ground side of the power source, and the positive lead is connected toward the positive side of the power source. Use the highest current range possible. 2. Close the switch. (Turn the switch on.) You should see no current flow indicated by the meter pointer since you are on a very high current range. Adjust the range/function switch to the next lower current measuring range position. Continue doing this until the meter pointer moves up scale when you have adjusted the range/function switch to a lower range. Your reading on the meter will tell you if it is safe to go to the next lower scale. For example, if the meter pointer shows a reading of 15 milliamperes when it is Figure 1−1 on the 100 milliampere range, then it would NOT be safe to go to the 10 milliampere range. CONCLUSION The resistance of a component can be determined by calculation when the voltage across the component and the current through the component has been measured. NOTE: The resistance of the bulb is different when the bulb is HOT than when it is cold. The resistance of the bulb cannot be measured directly when it is hot and has current flowing through it. However, the resistance of the bulb's filament can be determined indirectly using the voltage and current measurements, and Ohm's Law as was done in this experiment. Page 2 ELEC 103 LABORATORY EXERCISE 1 USE OF THE MULTIMETER DATA TABLES 4. 5. 6. Record the current you measured in step 3 in the box to the right. It = ____________ mA Measure the voltage across light bulb #1 and record it as V1. The correct polarity must be observed when connecting the voltmeter to the circuit. V1 = ___________ V Measure the voltage across light bulb #2 and record it as V2 The correct polarity must be observed when connecting the voltmeter to the circuit. V2 = ___________ V 7. Calculations using Ohms Law: R1 = Volts Amps R2 = Volts Amps = 8. Calculations using Ohms Law: Page 3 = R1 = ____________ Ω R2 = ____________ Ω ELEC 103 LABORATORY EXERCISE 2 VOLTAGE MEASUREMENTS PURPOSE OF EXPERIMENT To familiarize the user with the DC voltmeter scales and ranges of a general purpose multimeter such as the Simpson, Model 260 multimeter. EQUIPMENT LIST 2 − 6 Volt Dry cell batteries 2 − 1.5 Volt Dry cell batteries 1 − Simpson, Model 260 multimeter 1 − 1000Ω, 1 watt resistor PROCEDURE 1. Move the meter STATUS/POLARITY switch to the +DC position. Move the RANGE SWITCH to the 10 Volt range position. Check the meter leads to be certain they are plugged into the correct receptacles. The black lead should be plugged into the minus/common receptacle and the red lead should be plugged into the plus receptacle. 2. Connect the negative lead (black lead) of the voltmeter to the negative (minus) post on one of the 6 Volt batteries. Connect the positive lead (red lead) of the voltmeter to the positive (plus) post on the same 6 Volt battery. The meter pointer should have moved up the scale (to the right). If the meter pointer moved to the left, you must immediately disconnect the meter and return to step 1 and start the experiment again. 3. Read the meter and record the voltage in the appropriate place on the datasheet supplied with this experiment. 4. Without disconnecting the meter from the circuit, move the RANGE SWITCH to the 50 Volt range. Read the meter and record the voltage in the appropriate place on the datasheet. 5. Without disconnecting the meter from the circuit, move the RANGE SWITCH to the 250 Volt range. Read the meter and record the voltage in the appropriate place on the datasheet. 6. Repeat step 2 through step 5 for the other 6 Volt battery and record the voltage measurements in the appropriate place on the datasheet. 7. Using one of the test leads to connect the batteries together, connect the plus (+) terminal of one of the 6 Volt batteries to the negative (−) terminal of the other 6 Volt battery. This is a SERIES AIDING CONNECTION. Draw the fully labeled schematic, with appropriate electrical symbols, including the voltmeter, Figure 2−1 on the datasheet. Refer to the schematic in Figure 2−1 for some hints in drawing the schematic. Page 1 ELEC 103 LABORATORY EXERCISE 2 VOLTAGE MEASUREMENTS 8. 9. 10. 11. 12. 13. 14. Move the RANGE/FUNCTION SWITCH to an appropriate range to measure the batteries when they are connected as described in step 7. Observe correct polarity when connecting the Voltmeter to the unused terminals of the 6 Volt batteries. Record the measured voltage on the datasheet. Measure the voltage of a 1.5 Volt battery using the correct RANGE SWITCH position. Record the voltage on the datasheet. Connect the negative terminal of the 1.5 Volt battery to the positive terminal of one of the 6 Volt batteries. Again, as in step seven, this is called a SERIES AIDING CONNECTION. The battery voltages add in a series aiding connection. Select the appropriate meter range and measure the voltage across the two batteries. Draw a schematic of the circuit, including the voltmeter, in the space provided on the datasheet. Also, record the measured voltage. Connect the positive terminal of one 6 Volt battery to the positive terminal of a 1.5 Volt battery. Draw the fully labeled schematic, including the voltmeter, in the space provided on the data sheet. This is called a SERIES OPPOSING CONNECTION. Measure the voltage across the two batteries using the correct RANGE SWITCH position. Record the measured voltage on the data sheet. Using test leads, connect a series combination of a 1.5 Volt battery, a 6 Volt battery and a 1000 Ω resistor with the voltmeter across the resistor, as shown in Figure 2−1. Draw a schematic of the circuit. Record the measured voltage on the data sheet. Reverse the test leads to the 6 Volt battery in the circuit shown in Figure 2−1. Connect the meter to the circuit with the same polarity shown in Figure 2−1. However, before making the measurement, move the STATUS/POLARITY SWITCH to minus DC. Draw a schematic of the circuit. Record the measured voltage on the data sheet. Page 2 ELEC 103 LABORATORY EXERCISE 2 VOLTAGE MEASUREMENTS DATASHEET FOR EXPERIMENT 2 STEPS 3 − 6: Six Volt Battery Voltage Readings. Battery #1 Battery #2 6 Volt battery using the 10 Volt Range _________ _________ 6 Volt battery using the 50 Volt Range _________ _________ 6 Volt battery using the 250 Volt Range _________ _________ STEP 7: Step 8: Schematic − Two Batteries Connected in SERIES AIDING. Series Aiding Voltage Measurement. Voltage = _____________ Range = _____________ Step 9: 1.5 Volt Battery Measurement. Voltage = _____________ Range = _____________ STEP 10: Schematic − Two Batteries Connected in SERIES AIDING. Voltage = __________ Range Page 3 = __________ ELEC 103 LABORATORY EXERCISE 2 VOLTAGE MEASUREMENTS STEP 11: Schematic − Two Batteries Connected in SERIES OPPOSING. Step 12: Series Opposing Voltage Measurement. Voltage = _____________ Range = _____________ STEP 13: Schematic − Series Aiding Sources series circuit. Voltage = _________ Range = _________ STEP 14: Schematic − Series Opposing Sources series circuit. Voltage = __________ Range = _________ Page 4 ELEC 103 LABORATORY EXERCISE 3 CURRENT MEASUREMENTS PURPOSE OF EXPERIMENT To familiarize the user with the DC milliammeter scales and ranges of a general purpose multimeter such as the Simpson, Model 260 multimeter, and acquaint the user with current measuring techniques. EQUIPMENT LIST 2 − 6 Volt Dry cell batteries 1 − 1.5 Volt Dry cell battery 1 − Simpson, Model 260 multimeter 1 − 1000Ω, 10,000Ω, 1 Watt Resistor PROCEDURE I − CURRENT MEASUREMENT IN SERIES CIRCUITS 1. Using test leads, connect the series circuit as shown in Figure 3−1. Do not connect the meter into the circuit. On the meter, place the STATUS/POLARITY SWITCH in the +DC position and the RANGE/FUNCTION SWITCH in the 10 mA range position. 2. NOTE Connecting the milliammeter into the circuit Figure 3−1 requires you to place the meter in SERIES WITH THE ELEMENT. To measure current one must first break the circuit at the place the measurement is desired. Then, complete the circuit using the milliammeter leads. Connect them with the negative lead toward the negative side of the source and the positive lead toward the positive side of the source. In the circuit of Step #1, where two sources are used, you will use the larger of the two sources to determine the polarity of the source. Connecting the milliammeter as a voltmeter or reversing meter polarity, can cause serious damage to the multimeter. Also, current measurements must be made with a resistor in the circuit. A current meter must never be connected directly to the source. 3. Insert the milliammeter into the circuit. Draw a schematic of the circuit with polarities indicated and the direction of current flow shown. Then, read and record the measured current in the space provided on the datasheet. 4. Remove the meter from the circuit and place a test lead in the break. Now, break the circuit between the negative terminal of the 1.5 Volt battery and the 1000Ω resistor. Insert the meter into the break. Use the larger of the two batteries to determine the source polarity. Observe correct polarity when inserting the milliammeter into the break. 5. Draw a schematic of the circuit as it is presently connected. Read the milliammeter and record the measured current in the space provided on the datasheet. 6. In the previous circuits, the two batteries were connected in series opposing. In as much as current in a series circuit increases with an increase in voltage, reconnect the circuit so the batteries are connected in series aiding. This will increase the source voltage and, therefore, the circuit current. Insert the milliammeter into the circuit in one of the two places it was located previously. It makes no difference where you place the meter. Since this is a series circuit, the current will be the same no matter where you insert the meter into the circuit. Page 1 ELEC 103 LABORATORY EXERCISE 3 CURRENT MEASUREMENTS 7. 8. 9. Draw a schematic of the circuit as it is presently connected. Read the milliammeter and record the measured current in the space provided on the datasheet. Connect the two 6 Volt batteries in a series aiding configuration and connect them with the 1000 Ω resistor in a series circuit. Before connecting the milliammeter into the circuit, calculate the current flowing in the circuit. Adjust the RANGE SWITCH to the appropriate current range. If you are unable to determine a suitable range, start at the highest range available. Then, reduce the range, to give you a reading with maximum pointer deflection without going beyond full scale. Insert the milliammeter into the circuit. Draw a schematic of the circuit as it is presently connected. Then, read and record the measured current in the space provided on the datasheet. PROCEDURE II − CURRENT MEASUREMENT IN PARALLEL CIRCUITS 10. All the measurements in PROCEDURE I were performed using series circuits. That is, circuits in which components were connected end−to−end so only a single path for current flow exists. A series circuit schematic is shown in Figure 3−1. Measurement techniques in parallel circuits, however, are not exactly the same as series circuits. The milliammeter is still placed in SERIES WITH THE ELEMENT, or component, through which the desired current to be measured flows. However, measured currents are different in different parts of the circuit. Figure 3−2 shows a parallel circuit with a milliammeter placed to measure the current flowing through the 1000 Ω resistor and another to measure the total current flowing in the circuit. The current through the other resistor can be measured by connecting the milliammeter in the same way it is connected to measure the current through the 1000 Ω resistor. Since you were issued only one milliammeter, you will be inserting the milliammeter Figure 3−2 in one place and then moving it to the other points in the circuit until you have measured the three individual currents. You should be aware that the total circuit current in a parallel circuit is equal to the sum of the individual branch currents. In this parallel circuit, the circuit shown in Figure 3−2, the sum of the currents through the two resistors will equal the total circuit current, labeled IT in the figure. 11. Using test leads. connect the 6 Volt battery, the 1000 Ω resistor and the 10,000 Ω resistor as shown in Figure 2−Lab 3. This is a parallel circuit configuration. Since you have only one milliammeter, the milliammeter should be connected to measure the total circuit current. 12. Using a safe current range on the milliammeter, and observing correct polarity, measure the total current flowing in the circuit (the current flowing through the source). Draw a schematic on the datasheet including the milliammeter placement used in the measurement. Also, record the measured current in the same space. 13. Using a safe current range on the milliammeter, and observing correct polarity, measure the current flowing through the 1000 Ω resistor. Draw a schematic of the circuit on the datasheet including the milliammeter placement used in the measurement. Also, record the measured current in the same space. Page 2 ELEC 103 LABORATORY EXERCISE 3 CURRENT MEASUREMENTS 14. Using a safe current range on the milliammeter, and observing correct polarity, measure the current flowing through the 10,000 Ω resistor. Draw a schematic of the circuit on the datasheet including the milliammeter placement used in the measurement. Also, record the measured current in the same space. 15. Add the current measured in step 13 to the current measured in step 14. Record the sum in the appropriate space on the datasheet. This current should be the same as the current measured in step 12. 16. Answer the questions on the Data Sheet in the space provided. Page 3 ELEC 103 LABORATORY EXERCISE 3 CURRENT MEASUREMENTS STEP 3: − Schematic − Current = __________ Range = __________ STEP 5: − Schematic − Current = __________ Range = __________ STEP 7: − Schematic − Current = __________ Range = __________ Page 4 ELEC 103 LABORATORY EXERCISE 3 CURRENT MEASUREMENTS STEP 9: − Schematic − Current = __________ Range = __________ STEP 12: − Schematic − Current = __________ Range = __________ STEP 13: − Schematic − Current = __________ Range = __________ Page 5 ELEC 103 LABORATORY EXERCISE 3 CURRENT MEASUREMENTS STEP 14: − Schematic − Current = __________ Range = __________ STEP 15: Current measured in STEP 13. = Current measured in STEP 14. = Total current flow in Figure 2 = QUESTIONS: 1. State the conclusions you draw from the results of step 3 through step 9. 2. What happens to the current in a series circuit if you increase the source voltage (Refer to steps 7 and 9)? 3. What conclusions do you draw from STEP 13? 4. What conclusions do you draw from STEP 15? Page 6 ELEC 103 LABORATORY EXERCISE 4 RESISTANCE MEASUREMENTS PURPOSE OF EXPERIMENT To familiarize the user with the DC Ohmmeter scales and ranges of a general purpose multimeter such as the Simpson, Model 260 multimeter. And, to acquaint the user with safety precautions, equipment precautions, and measurement methods. EQUIPMENT LIST 1 − Simpson, Model 260 multimeter 1 − 1000Ω, 10,000Ω, and 1,000,000Ω 1 Watt Resistor 1 − 10kΩ potentiometer. PROCEDURE 1. In addition to being used as a voltmeter, and a milliammeter, the Simpson, Model 260, multimeter can be used as an Ohmmeter. Observe the top scale line on you meter. Notice the following significant differences between the Ohms scale and the DC scale used previously: a.) Zero is at the right end of the scale and the scale markings increase in value going to the left. b.) The divisions on the Ohms scale are not linear (equal in size). In fact, note that the physical length of the segment between 2 and 5 on the Ohms scale is roughly the same as the segment between 5 and 10. Yet 5 minus 2 = 3 units, whereas 10 minus 5 = 5 units, different values for roughly equal length scale segments. This type of scale is generally referred to as a nonlinear scale. Nonlinear scales are used for Ohmmeters while linear scales are used for voltage and current meters. Now turn your attention to the RANGE/FUNCTION SWITCH. There are three resistance or Ohmmeter ranges, R X 1, R X 100, and R X 10000. Depending upon the size of the resistance to be measured, one of these ranges is selected and used with the Ohms scale. 2. On the R X 1 range, any scale reading is interpreted using the numbers printed on the scale. That is, all meter readings are multiplied by 1. On the R X 100 range or the R X 10,000 range, the meter readings are multiplied by 100 or by 10,000 respectively. Consequently, if the needle is pointing at 5 on the Ohms scale, it means the reading is 5 Ω if you are using the R X 1 range. It would mean 500 Ω if using the R X 100 range or 50,000 Ω if using the R X 10,000 range. 3. Before proceeding with the experiment, the following Ohmmeter precautions must be reviewed since serious damage to the multimeter can result if they are not carefully followed. OHMMETER PRECAUTIONS A. Move the status switch to the +DC position. B. Move the range switch to one of the three resistor ranges. Unlike the voltmeter and the Page 1 ELEC 103 LABORATORY EXERCISE 4 RESISTANCE MEASUREMENTS milliammeter, the Ohmmeter does not need to be adjusted to the highest range. The Ohmmeter can be adjusted to any range if it is connected to the circuit properly. C. Remove all power from the circuit. D. Remove at least one end of the component being measured from the circuit. E. Connect the ohmmeter leads together and use the zero ohms adjust knob to make the pointer read zero ohms before each use. 4. Adjust the STATUS/POLARITY switch to the +DC position and make certain the multimeter test leads are in the proper receptacles. Turn the RANGE switch to the R X 1 range and short (connect together) the two meter leads. Observe that the meter pointer moves up the scale and comes to rest somewhere around zero ohms. With your test leads shorted, you want your meter to "think" it is reading ZERO OHMS of resistance since there is no resistance between the meter leads. To accomplish this strategy leave the meter test leads shorted. Turn the ZERO OHMS adjust knob slowly to the extreme counterclockwise position and then to the extreme clockwise position. Observe the pointer movement. You should see the pointer move from below full scale to greater than full scale. If this does not occur, contact you lab instructor immediately. 5. Adjust the pointer, using the ZERO OHMS ADJUST knob, until the pointer is over the zero on the Ohms scale. Once this has been accomplished, you have calibrated your Ohmmeter for this measurement and you can UNSHORT the meter test leads. NOTE: It is mandatory that you ZERO OHMS ADJUST the Ohmmeter before each initial measurement on a specific range. Therefore, each time you change ranges you must zero ohms adjust the meter. 6. Using appropriate ranges, measure the resistance of the 1000Ω resistor, the 10,000Ω resistor, and the 1,000,000Ω resistor. Record the measured values on the datasheet. MAKE CERTAIN YOU RE−ZERO THE METER WHENEVER CHANGING RESISTANCE RANGES. 7. Re−measure the 1,000,000Ω resistor using the appropriate range. However, this time place one hand on each end of the resistor so your hands are in contact with the resistor and the meter test leads. Record the results of the measurement on the datasheet. 8. Repeat Step 7, only wet your fingers before touching the resistor and the meter leads. Record the results of the measurement on the datasheet. 9. The potentiometer is a variable resistor that has three terminals. One pair of terminals is fixed and you will measure the total potentiometer resistance between these terminals. The other lead connected to a wiper. When you connect the ohmmeter between the wiper and one of the fixed terminals, you will measure a resistance that may be any value between 0 Ω and the maximum resistance value of the potentiometer. The resistance value that you measure will depend on the position of the wiper. Connect the ohmmeter across the fixed leads of a potentiometer. Adjust the meter to make an appropriate resistance reading. Slowly adjust the FIGURE 4−1 wiper and observe the reading on the meter but be careful not to touch the leads with your fingers. Record your observation on the Data Sheet as either remains constant or varies. Page 2 ELEC 103 LABORATORY EXERCISE 4 RESISTANCE MEASUREMENTS DATASHEET FOR EXPERIMENT 4 STEP 6: Color Code Value Measured Value Meter Range Used 1,000Ω 10,000Ω 1,000,000Ω STEP 7: Measured resistance = _____________ Range = ___________ STEP 8: Measured resistance = _____________ Range = ___________ STEP 9: As you move the wiper, the resistance between the two fixed terminals ___________________. As you move the wiper, the resistance between the wiper and the fixed terminal ____________. QUESTIONS: 1. When you read the Ohmmeter in STEP 6, you probably noticed a difference between the meter reading and the resistance value predicted by the color code. Give a reason for the difference. 2. What was the purpose of STEP 7 ? 3. Using the measured values of the resistors in STEP 6, calculate the value you would measure if you placed the three resistors in parallel. Show the formula and the calculations used to arrive at the answer. Page 4 ELEC 103 Unit 5 The Oscilloscope PURPOSE To determine whether a circuit is operating properly, or malfunctioning, it is often necessary to observe the electronic signals in the circuit. A VOM or DMM will give the magnitude of an ac current or voltage, but is unable to display the signal or provide information to allow the user to determine if there is distortion on the signal, or if there is an improper phase shift in the circuit. The oscilloscope allows the technician to view these signals and determine the circuit operating characteristics. The purpose of this exercise, is to use the ac meter and the oscilloscope to make voltage and current measurements in an ac circuit. The student will be able to describe the procedure to measure voltage, period, frequency, and phase relationships for ac waveforms with the oscilloscope. EQUIPMENT AND MATERIALS REQUIRED Oscilloscope with 2 probes 2 each DMM’S and VOM’s Powered Protoboard Audio Signal Generator Frequency Counter Resistor, ½W, 5%, 1.5kΩ Capacitor, 35V, 10%, 0.1µF INTRODUCTION There is a wide variety of test equipment available to the technician, which provides information on circuit operation. All test equipment, used to measure voltage in a circuit, must have a high input impedance to prevent the test equipment from loading the circuit under test and thereby provide inaccurate measurements. The VOM usually has a moderately high input impedance while the DMM and oscilloscope have a high input impedance of 1MΩ or higher. This high input impedance assures the measurements will be accurate since loading does not occur. The VOM and DMM are used to measure ac currents and voltages in the same way they were used in the ELEC 111 course experiments. Since the polarity of the waveform is constantly changing, there is no polarity to consider when making ac voltage and current measurements. The oscilloscope provides a visual representation of the voltage at any point in the circuit on a CRT. As a result, the technician may view the display and determine the shape of the waveform, the amplitude of the waveform, the frequency of the waveform, the phase relationships between voltage waveforms at various points in the circuit, or the duration of some event lasting one or more cycles. An ac voltage is one that is continuously changing in magnitude and direction, with respect to time. These ac signals have a frequency of repetition measured in Hertz (cycles per second), which is the number of complete waveforms created in one second. The symbol for frequency is f, and the unit of measurement is the Hertz (Hz). Reference may also be made to the amount of time required for two alternations, which make up one cycle, of the signal, which is called the period of the waveform. The period of the waveform uses the symbol T, and has the second as the unit of measurement. There is an inverse relationship between the frequency and period of the waveform. This may be expressed as: 1 f= T <5 − 1> The analog waveform used to carry information in an ac circuit is the sinewave. These sinusoidal signals may be symmetrical above and below a zero volt reference, or they may be riding on a dc level. Page 1 ELEC 103 Unit 5 The Oscilloscope The amplitude of a signal is the height of the signal above or below the reference. In most cases, the reference used is zero volts or ground. The polarity of a signal refers to the waveform’s value, with respect to the common level in the system, which is usually ground or zero volts. All parts of the signal that rise above the zero volt level have a positive voltage polarity, and all parts of the signal that fall below the zero volt level have a negative voltage polarity. The sinewave in Figure 5 − 1 has equal excursions above and below the zero volt reference. The maximum positive level from the zero volt reference is called the positive peak voltage, VPP, and the maximum negative level from the zero volt reference is called the negative peak voltage, −VP. Since the variation above and below the reference is the same, the magnitude of VP is equal to −VP. While this is the case for Figure 5 − 1, it is not required. An ac information signal may also ride on a dc level. When the dc level is such that there is no excursion about the zero volt reference, but is always either above or below zero volts, this signal is no longer an ac waveform but a pulsating dc voltage. The amplitude is defined as the voltage level between the positive and negative peaks, FIGURE 5 − 1 and is called the peak−to−peak voltage, or VPP. When the magnitude of VP is equal to the magnitude of −VP, we may determine VPP from the equation: VPP = 2VP The amplitude of the ac signal may be measured and defined in different formats. The instantaneous voltage is the value of the voltage at a particular time. The symbol v is used to denote an instantaneous voltage, and the unit of measurement is the Volt. The time at which we wish to determine the instantaneous voltage may be θ (theta) and is measured in either degrees or radians, or time ( t ) measured in seconds. Both θ and t must be measured with respect to some defined reference of the signal. This is normally when the signal is at zero volts and rising, or zero degrees. When the instantaneous voltage is determined using t, the time must first be multiplied by 2 π f, which is the angular velocity (also known as ω) of the voltage. The units of measurement for ω is rad/s (radians per second). Since the voltage is sinusoidal, the instantaneous voltage may be determined from the equations: v = VP Sin θ v = VP Sin (2 π f t) Another measurement to define an ac voltage or current is the effective or Root−Mean−Square (rms) voltage. This measurement is important, since the rms voltage or current has the same heating value as the corresponding dc voltage or current. The symbol used to denote an rms voltage measurement is Vrms, and is measured in Volts. Current measurements use the symbol Irms, and have the Ampere as the unit of measurement. The VOM and DMM provide rms voltage and current measurements. The rms voltage may be calculated from the peak voltage using the equation: Vrms = VP 2 which equals Vrms = 0.707 Vrms Page 2 ELEC 103 Unit 5 The Oscilloscope The final measurement, used to refer to an ac voltage or current, is the average value. The average value for an ac voltage is the average of all the voltage levels for one−half of a cycle of an ac voltage. This measurement is important since it corresponds to the deflection of the pointer for an analog ac meter, even though the ac meter scale is calibrated to provide rms readings. The relationship between the peak and average voltage of a sinusoidal ac voltage or current is: 2VP VPP = = 0.6366 VP π π Remember that VPP, VP, Vrms, and Vav values are equivalent methods to define the same voltage. The type of measurement made is determined by the equipment used to make the measurement. A VOM would indicate 120Vrms supplied to the home, while an oscilloscope would indicate the same voltage as 169.71VP, and 339.41VPP. To find the relationship between the rms and average voltages recall: Vav = VP = 2 Vrms and ⎛ π⎞ VP = Vav ⎜ ⎟ ⎝ 2⎠ Therefore, if we equate these two readings we obtain: ⎛ π ⎞ ⎛π⎞ 2 Vrms = Vav ⎜ ⎟ and Vrms = Vav ⎜ ⎟ 2 ⎝ ⎠ ⎝ 2 2 ⎠ Since π, 2, and 2 2 are constants, we may solve for the constants in the previous equation and obtain: Vrms = 1.1107Vav and Vav = 0.9 Vrms The oscilloscope is used to measure the amplitude of a voltage, displayed on the vertical axis, against time, which is displayed on the horizontal axis. The vertical and horizontal axis have marked graduated scales, in each direction, calibrated into major and minor divisions. A range switch, in conjunction with the vertical amplifier, controls the vertical deflection of the electron beam in the oscilloscope, while the time base and associated circuitry control the horizontal sweep speed. The vertical gain of the scope is calibrated in volts per division. The signal is applied to the vertical deflection circuitry of the scope that deflects an electron beam vertically on the face of the oscilloscope according to the polarity and magnitude of the applied voltage. To determine the amplitude of the voltage, count the number of divisions over which the beam is deflected, and multiply this value by the gain setting. If the measurement is made between the zero volt reference, and the peak value, a peak voltage is being measured, while a measurement between the positive and negative peaks is a peak−to−peak voltage measurement. For Figure 5 − 1, if the beam is deflected 7.2 divisions between the positive and negative peaks, and 3.6 major divisions between the zero volt reference and the positive peak, with the vertical gain set to 5 V/div, the voltage represented by this deflection is: VPP = 7.2 div x 5 V/div = 36VPP and VP = 3.6 div x 5 V/div = 18VP The horizontal deflection on the oscilloscope is controlled by a sweep generator. The sweep generator is used to move the beam from the left side of the CRT to the right side of the CRT. When the beam reaches the right side of the CRT, it is swept back very rapidly and the process is repeated. The control for the sweep speed is calibrated in seconds/division, and the horizontal axis is used to measure time. Therefore, on the horizontal axis of the scope, you will be able to determine the period of the waveform. To determine the frequency of a waveform, with an oscilloscope, the period of one cycle is usually determined. The sweep speed is adjusted until more than one complete cycle occupies the horizontal axis of the oscilloscope. Then the number of divisions for one cycle is counted and multiplied by the Page 3 ELEC 103 Unit 5 The Oscilloscope sweep speed. For Figure 5 − 1, if the sweep speed is set to 0.5 ms/div and one cycle occupies 6.6 major divisions, the period for one cycle is: T = 0.5 ms/div x 6.6 div = 3.30 ms Since the frequency of a waveform is the inverse of the period, the frequency of the signal in Figure 5 − 1 is: 1 1 f = T = 3.3ms = 303 Hz. When two or more waveforms of the same frequency are present at the same time in a circuit, it is possible to define the time relationship between corresponding values of the waveform. This relationship is called the phase or phase angle. The phase angle uses θ as the symbol of measurement, and has either degrees or radians as the units of measurement. Two waveforms are in phase when both waveforms begin at zero volts and rise or fall together. If both waveforms are zero volts at the same time, but one waveform rises as the other falls, the waveforms are antiphase (θ is 180o or π rads out of phase). Waveforms may also be 90o out of phase, which is referred to as phase quadrature. The phase angle of one waveform in a circuit may lead or lag the other. Assume waveform A in Figure 5 − 2 is the reference voltage for the circuit, usually the supply voltage. Since waveform B is at zero volts and rising, after waveform A is at zero volts and rising, waveform B lags waveform A by some phase angle θ. Using the same analogy, we may also say that waveform A leads waveform B by the phase angle θ. A leading phase angle is considered positive, while a lagging phase angle is negative. The phase angle may be any value between 0o and 180o (0 and π rads). Figure 5 − 2 contains two signals of the same frequency applied to the inputs of an oscilloscope. FIGURE 5 − 2 The amplitude of each channel may be determined, and each channel may have the vertical gain set to different levels. To determine the phase relationship between the two waveforms. First, select the channel that contains the reference waveform. Assume waveform A is the reference signal for Figure 5 − 2. The vertical gain is adjusted to provide maximum deflection about the zero volt reference, and both signals must have the same zero volt reference on the CRT. The sweep speed is adjusted to provide a minimum of 180o for each signal. This oscilloscope has ten major divisions across the CRT. That breaks down to 18o per division. Notice, in Figure 5 − 2, the horizontal divisions are usually calibrated in milliseconds per division or microseconds per division and this scope display is no different. However, we will convert the horizontal directly into degrees, using a ratio. Page 4 ELEC 103 Unit 5 The Oscilloscope Usually, the number of divisions across the horizontal are divided into 180o 180o, the number of degrees in the one alternation being displayed. = α = 18o/ div 10div That provides the number of degrees per division. If you measure the number of divisions along the zero volt reference that waveform A and ⎛ 2div ⎞ waveform B are displaced from one another, you can calculate the θ = 180o ⎜10div⎟ ⎝ ⎠ phase angle. The phase angle between waveform A and waveform B may be calculated using a proportion, as shown to the left. Since one θ = 36o alternation is 180o and there are 10 divisions, a displacement of two divisions between the waveforms equals a phase shift of 36o. The oscilloscope may be used to measure the phase angle between two waveforms of the same frequency by displaying a pattern known as the Lissajous Figure. Under this method, the time base or sweep speed is disconnected from the horizontal deflection circuitry. The reference signal is connected to the horizontal (X) deflection circuitry, an unknown signal is connected to the vertical (Y) deflection circuitry, and the sweep speed selector is set to the X−Y mode. With no signals applied, a dot will appear on the CRT. Do not leave the CRT in this mode for too long at a high intensity or damage to the phosphor coating on the CRT will occur. The dot is adjusted so the dot appears at the intersection of the horizontal and vertical scales. The input signals are applied, and a waveform similar to that of Figure 5 − 3 will appear on the CRT. Make certain the pattern is centered on the zero axis and the amount of deflection for A and B is measured. The phase angle, θ, may be determined from the equation: A θ = sin−1 B FIGURE 5 − 3 NOTE When connecting line−powered equipment to a circuit, make sure the ground terminals from all devices are connected to the circuit ground as indicated on the schematic diagram. PROCEDURE 1. Measure the values for R and C and record the values in Table 5 − 1. 2. Calculate and record the values for Table 5 − 2 and Table 5 − 3 using information in the tables. Page 5 ELEC 103 Unit 5 The Oscilloscope 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. Build the circuit of Figure 5 − 4. Make certain the ground leads of the frequency counter (FC) and the audio generator (VS) are connected to the same point in the circuit. Connect one DMM across the output terminals of the audio generator, and insert the second DMM to measure the current. Adjust the settings of the DMM’s to their proper positions. Connect a probe to the oscilloscope Channel 1 Figure 5 − 4 terminal. The other probe will be connected to Channel 2 later in the experiment. Connect the ground lead, of the oscilloscope probe connected to CH 1, to the same point in the circuit as the ground lead of the audio generator. Connect the other lead of the oscilloscope probe to measure the voltage across the resistor. Set the oscilloscope to trigger internally on Channel 1. Make sure all controls are set to their calibrated position. Check all circuit connections. Apply power to all the equipment. Adjust the frequency of the audio generator until the frequency counter reads 500 Hz. Adjust the voltage output of the audio generator until the DMM reads 3V. Adjust the sweep speed to obtain a display similar to Figure 5 − 1. It would probably be better to view two or three cycles instead of the single cycle displayed in Figure 5 − 1. Slide the selector under the CH 1 vertical gain control, marked AC−GND−DC, to the GND position. Adjust the CH 1 position control to place the trace in the center of the CRT. Move the CH 1 selector to AC, and adjust the vertical gain until the amplitude of the signal looks similar to Figure 5 − 1. Use the CH 1 selector to make sure the vertical position of the trace is centered on the CRT. Measure the period of the waveform with the oscilloscope and record the data in Table 5 − 1. Using the data from step 9, calculate the frequency of the waveform and record the results as a measured value in Table 5 − 1. Measure the applied voltage (Vrms) with the DMM and record the results in Table 5 − 2 on the appropriate line. Measure the total circuit current (Irms) with the current DMM and record the results in Table 5 − 3 on the appropriate line. Measure VP and VPP with the oscilloscope, and record the data in Table 5 − 2. From the data in step 13, calculate IP by dividing VP by R and record the result as the measured value in Table 5 − 3. Repeat for IPP. Adjust the voltage output of the audio generator until VP on the oscilloscope reads 7V. Then, repeat steps 11 to 14. Page 6 ELEC 103 Unit 5 The Oscilloscope 16. 17. 18. 19. 20. 21. 22. 23. Adjust the voltage output of the audio generator until VPP on the oscilloscope reads 20V. Then, repeat steps 11 to 14. Remove power. Build the circuit in Figure 5 − 5. Connect CH 1 of the oscilloscope across the audio generator. Connect the second probe to the CH 2 input of the oscilloscope and then connect CH 2 across the resistor as shown in Figure 5 − 5. FIGURE 5 − 5 Apply power to all equipment. Adjust the frequency of the audio generator to 1000Hz, and the voltage output to 5Vrms. Place the CH 1 and CH 2 AC−GND−DC selectors to GND and center the traces on the CRT. Return the CH 1 and CH 2 selectors to the AC position. Adjust the oscilloscope, as necessary, to obtain a display similar to Figure 5 − 2. Measure and record the readings for the Number of Divisions for 180o and the number of divisions the two waveforms are displaced from one another along the zero axis, where they cross the zero axis going positive, and enter the data in Table 5 − 4. Calculate the Phase Angle, and record the result in Table 5 − 4. Set the sweep speed to the X−Y position. Place the CH 1 and CH 2 AC−GND−DC selectors to GND and center the dot on the CRT. Return the CH 1 and CH 2 selectors to the AC position. Adjust the oscilloscope as necessary to obtain a display similar to Figure 5 − 3.This is best accomplished by adjusting the amplitude of both channels for the same deflection, one channel at a time. Disconnect channel one by moving the AC−GND−DC selector to GND and adjust channel two for a specific amplitude. Then disconnect channel two and adjust channel one to the same amplitude. After both channels are adjusted for the same number of divisions deflection, activate both channels and record the readings for the Number of Divisions for distance A and for distance B in Table 5 − 5. Calculate the Sine θ and the Phase Angle. Record the results in Table 5 − 5. Page 7 ELEC 103 Unit 5 The Oscilloscope DATA TABLES Value R C f T Rated 1.5kΩ 0.1µF 500 Hz 2 ms Measured TABLE 5 − 1 Voltage Vrms Calculated 3V VP VPP Measured Calculated 5V Measured Calculated 15V Measured TABLE 5 − 2 Current Calculated Irms IP IPP 2mA Measured Calculated 4.67mA Measured Calculated 13.33mA Measured TABLE 5 − 3 Number of Divisions from 0o to 180o Waveform A to B Phase Angle TABLE 5 − 4 Number of Divisions from A B Sine θ TABLE 5 − 5 Page 8 Phase Angle