Introduction to TDS 1012 Digital Oscilloscope

advertisement
Introduction to TDS 1012 Digital Oscilloscope
This laboratory exercise is designed as a brief overview of the features of the TDS 1012
digital oscilloscope. The oscilloscope has long been one of the most important tools in
any physics or electronics laboratory.
The main purpose of an oscilloscope is to display a graph of input voltage on the vertical
axis as a function of time on the horizontals axis. As with any graph you need to have the
ability to adjust the scale of the vertical and horizontal axis. The VERTICAL and
HORIZONTAL sections of the controls allow you to make these adjustments.
The oscilloscope can display graphs of signals that vary rapidly. This requires a special
trigger circuit and the TRIGGER controls form a major part of the controls. The trigger
has to decide where to "start" the plot. This is usually set to the point when the voltage
reaches a certain value when it is increasing (positive slope) or decreasing (negative
slope). The trigger level is shown on the screen as a small triangle.
The digital oscilloscope is a recent development, which has added many new features to
this important laboratory device. Most of the buttons above the VERTICAL and
HORIZONTAL controls and to the right of the screen allow the user to make use of these
digital features.
1
Introduction: TDS 1012 Digital Oscilloscope
Read through each section, perform the suggested operations and answer the questions.
You can’t break anything so feel free to explore the controls. The instructions are
sketchy, so if you can’t figure something out, ask your instructor.
Procedure:
1. Connect the MAIN OUT of the Wavetek Function Generator to channel 1 (CH 1)
of the oscilloscope and connect the AUX OUT to channel 2 (CH 2). Set the
Wavetek to frequency = 1.0 kHz, Vpp = 4.0 volts, function = sine wave,
amplitude modulation = OFF and DC OFFSET = 0.
2. Turn on the oscilloscope and press the AUTOSET button. This makes the
necessary adjustments to the oscilloscope so you should be able to see the
waveforms of a sine wave in channel 1 and a square wave in channel 2. These
waveforms are a plot of voltage vs. time with scale information at the bottom of
the screen. The scale information is given in units of volts or seconds per division
where division refers to the lines on the screen that are about a cm apart.
3. Please note that the AUTOSET button is very useful to get you started, but it
changes a number of settings, so it should not be used all of the time.
4. Experiment to determine the effect of the POSITION and VOLTS/DIV knobs for
both channel 1 and channel 2. Note that the volts per division for each channel
appear at the bottom left of the screen.
5. Press the CH1 MENU button several times and note that the CH1 waveform is
toggled on and off. Pressing the CH1 MENU also gives you some input options
on the right side of the screen, which can be activated by pressing the buttons
beside them. Try each of these options.
Assignment 1. – Draw a sketch of the oscilloscope screen with the two signals
described above. Sketch in the grid lines. Record the volt/div for each signal and
time/div at the bottom of the screen. Measure the number of vertical and horizontal
divisions (including fractional parts) and determine the amplitude and period for each
of the signals. Show the calculations.
Coupling – DC coupling passes both AC and DC components of the signal; AC blocks
the DC component and GND (ground) disconnects the input signal. Under
"TRIGGER" press the menu button and change the Coupling to AC.
Assignment 2 – Adjust the DC OFFSET on the Wavetek to 1.0 volt and look at channel 1
with AC and DC coupling. Describe what happens and why. Return the DC OFFSET on
the function generator to 0.
BW Limit – limits the bandwidth to reduce display noise.
Volts/DIV – selects the resolution (Coarse or Fine) of the VOLTS/DIV knob.
Note the effect of turning the knob when this is set for Coarse and Fine (notice the
volts/DIV in the lower left corners of the screen).
2
Probe – A probe is a special connector that reduces a high voltage to a value that the
oscilloscope can handle. The Probe option allows you to tell the oscilloscope that a
special probe is attached.
The Probe should be set for 1x in this laboratory since we will not use special probes.
The channel 2 menu works the same as for channel 1.
Trigger Control - The function of the trigger control is to set the initial point in time for
the waveform to be displayed. Usually this is a certain rising or falling voltage level on
channel 1 or channel 2. If the trigger is not set properly then the display will be unstable.
Experiment with adjustments to the trigger level and notice how the display can
move and become unstable.
Press the TRIGGER MENU button and experiment with the different menu options.
Assignment 3 – Describe what happens when the TRIGGER slope is switched from
“rising” to “falling” and explain why.
Horizontal Controls This section controls the horizontal axis, which is time. Notice the
effect of turning the POSITION knob and the SEC/DIV knob. It is important to see that
the SEC/DIV scale information appears at the bottom-center of the screen.
The HORIZONTAL MENU allows you to adjust the window zone (leave it on main)
and also selects whether the Trigger Level knob adjusts the trigger level (volts) or hold
off time (seconds). The hold off time delays the triggering of the waveform for a fixed
amount of time. It is generally better to leave this option in the “level” mode.
Special Menus
ACQUIRE – The acquire menu gives various options for smoothing the data. Keep in
mind that the Wavetek continuously produces sine and square signals. These signals are
converted to a digital value and up to 2500 values are stored and displayed.
In the sample mode, the oscilloscope records one sweep, displays it and then records
another. In the Peak Detect mode it overlaps successive sweeps. In the Average mode it
averages a number of sweeps and displays the average. This is usually slower. It gives a
cleaner waveform, but you can be tricked into thinking the waveform is very stable when
it is not.
CURSOR – The cursor menu allows you to make measurements of time or voltage from
the waveforms using two cursors, which are controlled by the ch1 and ch2 vertical
3
position knobs. Push the CURSOR button, set Type to Voltage and Source to CH 1. Two
horizontal cursors appear. Use the cursor controls (CH 1 and CH 2 POSITION knobs) to
place the cursors on the top and bottom of the sine wave. Note that the voltage levels of
cursor 1, cursor 2 and the difference between them (Delta) appear to the right of the
screen.
Assignment 4 – Use the voltage cursors to determine the Vpp of the sine wave and
compare this result with the Vpp that the Wavetek display indicates.
Assignment 5 – Use the time cursors to determine the period of the sine wave. From this
value, calculate the frequency of the sine wave and compare with the frequency that the
Wavetek display indicates. Show your calculations.
MEASURE – The measure menu instructs the oscilloscope to make various direct
measurements of the signals. When you select “SOURCE”, you can choose between ch1
and ch2. When you select “TYPE”, you can select the type of measurement (Vpp,
Period, Frequency, etc…)
Assignment 6 – Use the MEASURE menu to measure the period, frequency, and Vpp
for the sine wave. Compare your results with the answers of Assignment 4 and 5.
DISPLAY – skip
SAVE/RECALL – This menu allows you to save or recall instrument waveforms or
instrument setups (which we will skip). The saved waveforms are referred to as
Reference waveform A or B.
To save the waveform of channel 1 as Reference waveform A: select Waveforms, Source
= CH1, REF = A, and press Save. To display a reference waveform, use the lowest button
to turn Ref A on.
Assignment 7 Save the waveform of the sine wave in Ref A. Then change the frequency
range on the Wavetek to 200 HZ, which reduces the frequency of the sine wave by a
factor of four. Display the saved waveform (Ref A). Draw a sketch of the active
waveform and the stored waveform. Return the frequency range of the Wavetek to 1 kHz.
UTILITY – skip
HARDCOPY – skip. This button can be used to print display if a suitable printer is
attached.
RUN/STOP – Can be used to start or stop the waveform acquisition.
4
Reading Data into the Computer
The digital scope at each workstation is connected to the workstation computer. The
program Open Choice Desktop controls the reading in of the waveform data from the
oscilloscope. Follow the steps below to read in the waveform from the Power response
procedure. Make a plot of the data (voltage vs. time) in Excel. Print out the graph to turn
in with your experimental data.
1. Click on the icon to open the program.
2. Select (click on) Waveform Data Capture
3. Select Instrument:
a. Select ASCRL1::INSTR
b. Click ok
4. Select channels if needed
5. Click on Get Data
6. Click on Copy to Clipboard
7. Paste into spreadsheet
8. Make a plot of the data: voltage vs. time
9. Label graph properly and print out the plot only
5
Measuring Unknown Signals
Name:____________________
Teachscope Model #____________
The Teachscope is a device that generates twelve unknown voltage signals. Try to locate
each of the patterns listed below. Start with the easy ones and then look for the more
difficult ones. Record in each table the channel you found the signal and then report on
the measurements requested for each signal. Be sure to include units. Please note that
every Teachscope box is different so it is important to record the model number of
the box you use.
square wave
channel number
maximum voltage
minimum voltage
frequency (Hz)
sine wave (no offset, 5.4 volts peak-peak)
channel number
maximum voltage
minimum voltage
frequency (Hz)
sine wave (one half-cycle clipped, no offset)
channel number
maximum voltage
minimum voltage
frequency (Hz)
cycle clipped (positive or
negative)
6
+5 VDC
channel number
ground (with low level noise spikes)
channel number
sawtooth (long repetition time, positive offset)
channel number
maximum voltage
minimum voltage
repetition time (secs)
Triangle wave (positive offset)
channel number
maximum voltage
minimum voltage
frequency (Hz)
TTL pulse (short duration, long repetition time, ground to positive)
channel number
amplitude (peak volts)
repetition time
pulse width (nanoseconds)
7
TTL pulse (short duration, long repetition time, negative transition) differentiated square
wave (no offset)
channel number
amplitude (peak volts)
repetition time
pulse width (nanoseconds)
differentiated square wave (no offset)
channel number
maximum voltage
minimum voltage
frequency (Hz)
sine wave
channel number
maximum voltage
minimum voltage
frequency (Hz)
TTL pulse (short duty cycle)
channel number
maximum voltage
minimum voltage
frequency (Hz)
8
Download