Fundamental Electrical Relationships: Fundamental Electrical Relationships Ohm’s Law, Strength Duration Curve & Factors Affecting Thresholds Mark W. Sweesy, FHRS,CCDS,CEPS y, , , Arrhythmia Technologies Institute O tline Outline z Ohm’s Law Ohm s Law z Voltage, current and resistance (impedance) z E Energy and power d z Clinical relevance z Strength duration curve z Rheobase z Chronaxie z Clinical relevance Cli i l l V I R Ohm’s La Ohm’s Law V = I x R V I = V / R R = V / I I xR The equation uses current in the form of Amps In pacing current is expressed in mA V I R F Fundamentals of Electronics ndamentals of Electronics z Voltage (V): force moving the current: electromotive push z Current (I): flow of electrons through a circuit Current (I): flow of electrons through a circuit mA mA z Resistance (R) / Impedance: opposition to current flow: unit nit of measure: Ohm ( of measure: Ohm ( Ω Ω) V I R Fundamentals of Electronics Low Resistance = high current flow z PPP{ V g High Resistance = low current flow V PW PW R I I Ohm’s La Ohm’s Law: Pacers Ohm’s Law: Pacers vsICD’s Pacers vsICD’s sICD’s Typical range for pacemaker voltage 1.0‐5.0 V yp g p g Typical range for ICD voltage 50‐800 V Typical range for pacemaker current 2‐16 mA Typical range for ICD current 1 000 75 000 mA Typical range for ICD current 1,000‐75,000 mA Typical range for pacing lead impedance: 300‐1200 Ohms T i l f i l d i d Oh Typical impedance range for a high voltage lead 10‐60 Ohms V I R Ohm’s La Ohm’s Law What is the current in a 2.5 volt system with a 500 Ω resistance? V I R Ohm’s La Ohm’s Law What is the current in a 2.5 volt system with a 500 Ω resistance? I = V R = 2.5 V 500 Ω = .005 A .005 A x1000 mA 1 A = 5 mA 5 When the impedance is 500 ohms the absolute value of the current is 2x s the voltage When the impedance is 500 ohms the absolute value of the current is 2x’s the voltage 18. A lead fracture would result in which of the following? 18 A lead fracture would result in which of the following? a. decreased battery longevity b decreased conductor impedance values b. c. decreased electrode impedance values d decreased current drain d. d d d V I R Energ Energy Formula Form la E = V x E V x E = V V x I I x xPW E = V E = V2x PW R E = I E = I2x R R x x PW If you leave current in mA y and time in Msec your answer will be in uJ 15. A patient’s stimulation threshold was 1.0 Volt & 0.2 msec pulse duration. The pacemaker is programmed p p p g to 2.0 Volt & 0.4 msec duration. Lead impedance measures 1000 Ohms The output energy for this measures 1000 Ohms. The output energy for this patient would be: a 2X threshold a. b. 4X threshold c. 6X threshold h h ld d. 8X threshold E= 12 x .2 = .2 1000 2 V xt R 22 x .4 = 1.6 1000 Doubling the voltage is 4 times the energy and d bli doubling th the PW iis 2 ti times th the energy V I R Lead Impedance and Energ Lead Impedance and Energy Energ Energy = V x I x T V I T Pacemaker Pulse Widths= .1 ‐1.5 ms 4 0 V x 8.0 mAx 4.0 V x 8 0 mAx .5 ms 5 ms = 20 uJ 20 uJ ICD Pulse Widths = 3‐15 ms range1 75 750 V x 14,000 mAx 4, 33 ms = 31.5 J 3 5 or 31,500,000 uJ 1. Thammanomai A, et al. Heart Rhythm 2006; 3:1053-1059 Charge z In In pacing In pacing, charge describes the amount of electricity pacing, charge describes the amount of electricity pacing charge describes the amount of electricity delivered to the delivered to the myocardium during an atrial or myocardium during an atrial or ventricular output. ti l t t z Charge (Q) Charge (Q) is the product of pulse is the product of pulse duration ( duration (tt) ) in in milliseconds & output current (I) milliseconds & output current (I) in in milliamperes milliamperes. In . In pacing it is usually expressed in microcoulombs pacing it is usually expressed in i it i ll d i microcoulombs. i l b . z Q = I x T Power Po er z Power is the rate of doing work or transmitting energy. z Power P = I Power P = I x x V: where P is in watts, I is in amperes and V is in volts and V is in volts. 22. What is the estimated battery service life of a pacemaker with a 1.0 Amp‐hour battery and a 20 k ith A h b tt d microamp current drain? a. 4.0 years 5 y b. 5.0 years c. 5.7 years d 10.0 years d. 10 0 years 1 0 Amp-hour 1.0 Amp hour battery and a 20 microamp current drain Time = Q battery y capacity p y in Ah I current drain in amperes Time = 1,000,000 1 000 000 uA/hr = 50,000 50 000 hr = 5.7 5 7 yrs 20 uA 8,760 , hr/yr y Factors Affecting Current Drain acto s ect g Cu e t a z Lead/tissue interface z Stimulation threshold & safety margins desired z Quiescent current z Impedance z Electrode design & materials z Pacemakers programmed settings z % pacing Quiescent Current z Housekeeping H Housekeeping – k i – current utilized when not pacing; i.e., t tili d h t i i current used for ongoing tasks like monitoring the intrinsic rhythm & logging data. Typically in the 8‐ rhythm & logging data. Typically in the 8 h th & l i d t T i ll i th 8‐12 uA range for A f pacemakers z Reducing the output from 4.0 to 2.0 V may increase the battery longevity from 4 to 6 yrs, still have housekeeping y g y 4 y , p g needs Bipolar Unipolar Fracture / Connector / F t / C Fracture / / Connector / Uni C t / U / U Unii Leads L d Causes Causes of High & Low Impedance o g & o peda ce z High Impedance z Conductor Fracture, Loose Connector or Unipolar leads programmed to Bipolar z Low Impedance z Insulation degradation, Y‐adaptor or parallel pathways, loose connection maintainingcontact with body fluids1 1. In Ellenbogen K, et al. Clinical Cardiac Pacing, Defibrillation & CRT. 2007; pp 834 Measured Data If you could only pick 3 values on which to base your device If ld l i k l hi h b d i evaluation; which 3 would you pick? Measured Data If you could only pick 3 values on which to base your device If ld l i k l hi h b d i evaluation; which 3 would you pick? Stimulation Threshold Stim lation Threshold z The minimal amount of electrical stimulation that consistently produces a cardiac depolarization The threshold can be depolarization. The threshold can be expressed in terms of amplitude (milliamperes or volts), pulse duration (msec or volts), pulse duration (msec), ), charge (microcoulombs charge ( microcoulombs), or energy ), or energy (microjoules). microjoules). Strength Duration Curve z A graphic representation of the pulse amplitude, in voltage or current, required to produce an i lt t i d t d action potential in the cells. The curve is plotted as a function of pulse duration and can be defined f ti f l d ti d b d fi d by 2 points; the Rheobase by 2 points; the Rheobase and the Chronaxie and the Chronaxie. As . As th l d ti i the pulse duration increases the voltage th lt requirements decline (to a point). capture Rheobase z The value derived from a strength‐ The value derived from a strength‐duration curve that indicates the minimum output that will stimulate the myocardium irrespective of the pulse duration The curve myocardium irrespective of the pulse duration. The curve typically levels out around a pulse duration of 1 msec typically levels out around a pulse duration of 1 msec Chronaxie z The pulse duration value that is twice the rheobase. The pulse duration value that is twice the rheobase 26. The term 6 Th or fibrous fib capsule surrounding the electrode: a. is a normal foreign body reaction p b. is an unexcitable capsule of tissue c. decreases the current density at the lead tissue interface d. all of the above ^AfA ^AfA ^AfA Virtual Electrode z is a normal foreign body reaction which results from acute injury to the cell membrane z collagen is formed resulting in anunexcitable collagen is formed resulting in anu g g nexcitable capsule of capsule of p tissue which decreases the current density at the lead tissue interface ^AfA ^AfA ^AfA 17. Which of the above 3 month stimulation threshold graphs would be most characteristic of steroid eluting tined atrial lead? a A a. b. B c. C d. D ^f ^AfA Electrode Threshold Characteristics 1.0 0.8 Polished platinum 0.6 A Porous platinum 0.4 B Activated carbon 02 0.2 C Steroid eluting tined lead D I l t ti ( Implant time (weeks) k ) ^AfA ^ f 14 13 12 11 10 9 8 7 6 5 4 3 2 0.0 1 Pulsse width h (ms) 12 1.2 The difference noted in thresholds when decrementing until loss of capture versus going from sub threshold to until loss of capture versus going from sub‐threshold to capture is known as? a. latency l t b. Wedensky effect c. intermittent threshold p d. polarization Wedensk Wedensky Effect Effect z Also called capture hysteresis z The difference noted in thresholds when decrementing g until loss of capture versus going from sub‐threshold to until loss of capture versus going from sub‐ capture p z Ellenbogen suggests the phenomena is probably explained by varying rates during gain or loss of capture l i d b i t d i i l f t and that effects may be greatest at narrow pulse durations Ellenbogen & Woods. Cardiac Pacing &ICD’s, p 53 Output Programming zRule of thumb R l f h b zDouble Double Voltage Voltage zTriple PW p Doubling Voltage z Chronic Chronic threshold threshold .5ms & 2.5V so voltage 5ms & 2 5V so voltage was doubled z Tripling PW would give an inadequate g q safety margin 63 . If the threshold was determined to be at point C in the above 3 p figure; which of the following would result in an efficient and appropriate safety margin? pp p y g a. Double voltage b. Triple the pulse width c. Triple the voltage d. Double the pulse width Voltage Strength Duration Curve g g best programming O tp t Settings & Longe it Output Settings & Longevity Antiarrhythmics That Increase y Pacing Thresholds z Class IA z Quinidine z Disopyramide z Procainamide z Class IC z Flecainide Flecainide –– notorious z Propafenone p z Class III z Amiodorone z *Sotalol *Sotalol‐‐ decreases ICD thresholds z Prapafenone Electrolyte Abnormalities That y Increase Pacing Thresholds z Acidosis z Alkalosis z Hyperkalemia z Hypocalcemia z Hyperglycemia z Hypercarbia yp z Hypoxemia Temporal Conditions That p Increase Pacing Thresholds z Sleeping z Acute viral illness z Eating z Lead maturation z Excessive alcohol – Excessive alcohol – increases defibrillation increases defibrillation thresholds thresholds1 Papaioannou GI, et al. PACE 2002; 25:1144 Factors That Decrease Pacing g Thresholds z Corticosteroids Corticosteroids ‐‐ dexamethasone z Exercise z Catecholamines z Hypocapnia z Sympathomimetic Sympathomimetic agents agents ‐‐ Isuprel Isuprel, epinephrine, ephedrine, , epinephrine, ephedrine, atropine z Hyperoxia What’s the point? V I R Ohm’s La Ohm’s Law What would be the calculated voltage in a 10 mA system with 500 Ω of impedance? V I R Ohm’s La Ohm’s Law What would be the calculated voltage in a 10 mA system with 500 Ω of impedance? V = I x R 10 mA x 500 Ω = 5000 5000 mV x 1 V 1000 mV 5V Tripling Pulse Width z Chronic threshold 2.5V & .2ms PW, PW programmed d to .6ms providing a 2.1x safety margin. V I R Ohm’s La tattoo Ohm’s Law tattoo V I R V I R Polari ation Polarization Capacitance effect: positively charged ions resist ion flow eee- Na+ e- Na+ N Na+ e- Na+ e- e Electrode ClCl- Na+ Cl- Cl Cl- Electrode-Tissue Interface V I R Polari ation/Afterpotential Polarization/Afterpotential z Voltage that persists for a time after a paced impulse z Could cause oversensing g z Polarization can represent 30‐ Polarization can represent 30‐40% of total pacing i impedance d V I R Garden Hose Analog Garden Hose Analogy Normal Resistance = normal current flow Low Resistance = high current flow High Resistance = low current flow V I R Fundamentals of Electronics z Terminology T i l z Current: flow of electrons through a circuit Current: flow of electrons through a circuit mA mA V I R Current Flow e‐ P Pacemaker k Conductor Wire Wi Heart e- V I R Voltage: Electromotive push e‐ Pacemaker Conductor Wire Heart Provided by The batteryy V = IR Oh Ohm’s ’ L Law V = 55.0 0 mA x .500 500 Kohms V = 22.5 5 Volts V I R Ohm’s La Ohm’s Law What would be the impedance of a 2.5 volt system with a 10 mA current drain? R = V = I 2.5 V 10mA = 2.5 V .01 A 10 mA x 1 A 1000 mA .01 A = 250 Ω V I R F Fundamentals of Electronics ndamentals of Electronics z Energy: Voltage Energy: Voltage x x Current Current x x Time E = V x E = V x I I xt xt Current in mA Current in mA Time in ms Unit = Joule (J) V I R 46. Match the units: watt‐second a. energy b. power c. capacitance d d. current drain d i z z z joule: unit of energy or work henry: unit of inductance (induced by a magnetic field) coulomb: unit of charge l b it f h 34. The rate at which work is done is is: 34 a. joule unit of energy or work b. farad basic unit of capacitance c. power g d coulomb unit of charge d. 62. In the above figure line B represents the: a Chronaxie a. b. Pulse width threshold c Knee of the curve c. d. Rheobase V I R Fundamentals of Electronics Low Resistance = high current flow PW High Resistance = low current flow V I R F Fundamentals of Electronics ndamentals of Electronics z System Impedance z Cathode Electrode Impedance z Anode Electrode Impedance z Polarization Polarization Impedance Impedance z Tissue Impedance z Typical range for pacing lead impedance: 300 to over 1200 Ohms h z Typical impedance range for a high voltage lead 10 Typical impedance range for a high voltage lead 10‐‐60 Ohms Energy, Power, Watts & Current Drain gy, , z Energy: the ability to do work; for batteries usually expressed in 1 watt‐‐hours watt z Energy density: the energy content of a battery or capacitor based on volume or mass. For a battery is expressed in units of watt on volume or mass. For a battery is expressed in units of watt‐ ‐hrs 1 per cubic‐‐centimeter or watt‐ per cubic centimeter or watt‐hrs per gram z Power: the rate of doing work or transmitting energy, a product of 2 voltage & current usually expressed in watts z Watt: international unit of electrical power, watts are equivalent to 2 joules per second z Watt Watt‐‐hour: the amount of energy at a rate of 1 watt per hour, 1 hour the amount of energy at a rate of 1 watt per hour 1 2 joule = 1watt delivered for 1 second, 1 watt‐ joule = 1watt delivered for 1 second, 1 watt‐hour = 3,600 joules z Current drain: the average amount of current drawn from a battery by the external load; typically pacemakers draw from 10 by the external load; typically pacemakers draw from 10 – – 30 2 microamperes 1. In Ellenbogen K, et al. Clinical Cardiac Pacing, Defibrillation & CRT. 2007; pp 236 2. In Hayes &Asirvatham. Dictionary of Cardiac Pacing, Resynchronization, & Arrhythmias. 2007 C rrent I V/R Current I = V/R Pacemaker 4.0 V = (I) 500 ohms P k V (I) h 4.0 V/.500 Kohms 8.0 mA 4.0 V/.500 Kohms = 8.0 mA ICD 600‐830 V i,e. 750 V = (I) 50 ohms ICD 600 830 V i e 750 V = (I) 50 ohms 75 / 5 750 V/.050 Kohms = 15,000 mA 5, V I R Pacemaker Longe Pacemaker Longevity Factors Pacemaker Longevity Factors Longe it Factors z Battery Capacity z Rate of Drain z Circuit Efficiency z Pacing Rate z Programmed output & Pulse Duration z % of time pacing % f ti i 45. The horizontal asymptote of a strength duration curve defines which of the following: g a. charge b. pulse width threshold c. rheobase d. chronaxie V I R