Product Specification Digital Alarm System Area Line Pressure Alarms With Gas-Specific Transducers Elements • • • • • • One LED on each module indicates normal or abnormal gas line pressure status at a glance Area alarms can monitor up to 6 medical gas or vacuum services Local or remote pressure sensing Alarm module calculates high and low alarm settings automatically Primary power design accepts 115 or 230 VAC (50/60 Hz) UL listed and CSA approved Description Chemetron® area alarm panels shall be designed to meet the requirements of NFPA and CSA standards. Alarms shall be UL listed and CSA approved as an assembly and shall include all necessary gauges, factory wiring, transformers and circuitry requiring only 115 volt or 230 volt primary power. Internal voltage shall be stepped down to 5 volts DC for control circuit power. Voltage to external pressure or vacuum transmitters will be 20 volts DC. Area alarm panels shall be modular in design and have one LED (Light Emitting Diode) to indicate each normal/ abnormal condition. Each gas monitored shall have an LCD (Liquid Crystal Display) to continuously indicate actual pressure. The control module will have a test switch and an alarm silence switch. The test switch shall test all modules one at a time. There will also be an LED on the control module to indicate status of the micro-controller (green is normal, red is for a malfunction). Each line pressure module shall have a normal/abnormal LED, an LCD, and a window for the gas label. The back box shall contain factory installed copper tube extensions, 6" long and 3/8" OD to accept installer furnished lines from the medical gas system. Each inlet tube shall have a 1/8" NPT connection to accept pressure/ vacuum transducers. The signal distribution and power supply board shall have dual transformers and a selector switch to field select either 115 volt or 230 volt primary service. A detachable fuse holder on the primary side shall be included to insure that power is disrupted when the transformer cover is removed. The alarm audio tone shall be continuous. Its sound intensity shall be 90 dBa at 2 meters. The audio signal shall be cancelled only by the alarm silence button or fault correction. The audio alarm condition shall reactivate every half hour until the fault has been corrected. Regardless of the audio alarm, the LED shall indicate abnormal as long as a fault condition exists. The alarm shall automatically reset with the correction of the fault condition. The panel shall be equipped with dry contacts for each pressure/vacuum module. This will enable the alarm to interface with another alarm or central computer system. Gauge model area alarm panels shall be Chemetron® Series 74-14-XXXX. Digital Alarm System – Area Line Pressure Alarms With Gas-Specific Transducers Dimensional Data - Inches (cm) Face Plate Dimensions 3 Gas 1 4 " (35.6) x 8 -7/8" ( 2 2 . 6 ) 6 Gas 2 3 " (58.4) x 8 -7/8" ( 2 2 . 6 ) Product Specification Digital Alarm System – Area Line Pressure Alarms With Gas-Specific Transducers Ordering Information Panels with Dry Contacts for Remote Connection Catalog Number Quantity Description 74-14-001 2 Future, 1 Oxygen 74-14-004 2 Future, 1 Air 74-14-006 2 Future, 1 Nitrogen 74-14-002 2 Future, 1 Vacuum 74-14-014 1 Future, 1 Oxygen, 1 Air 74-14-012 1 Future, 1 Oxygen, 1 Vacuum 74-14-024 1 Future, 1 Vacuum, 1 Air 74-14-124 1 Oxygen, 1 Vacuum, 1 Air 74-14-137 1 Oxygen, 1 Nitrous Oxide, 1 WAGD 74-14-001224 2 Future, 1 Oxygen, 2 Vacuum, 1 Air 74-14-001237 2 Future, 1 Oxygen, 1 Vacuum, 1 Nitrous Oxide, 1 WAGD 74-14-001347 2 Future, 1 Oxygen, 1 Nitrous Oxide, 1 Air, 1 WAGD 74-14-001124 2 Future, 2 Oxygen, 1 Vacuum, 1 Air 74-14-001246 2 Future, 1 Oxygen, 1 Vacuum, 1 Air, 1 Nitrogen 74-14-011246 1 Future, 2 Oxygen, 1 Vacuum, 1 Air, 1 Nitrogen 74-14-013467 1 Future, 1 Oxygen, 1 Nitrous Oxide, 1 Air, 1 Nitrogen, 1 WAGD 74-14-112246 2 Oxygen, 2 Vacuum, 1 Air, 1 Nitrogen 74-14-123467 1 Oxygen, 1 Vacuum, 1 Nitrous Oxide, 1 Air, 1 Nitrogen, 1 WAGD 74-14-011244 1 Future, 2 Oxygen, 1 Vacuum, 2 Air 74-14-011347 1 Future, 2 Oxygen, 1 Nitrous Oxide, 1 Air, 1 WAGD 74-14-112347 2 Oxygen, 1 Vacuum, 1 Nitrous Oxide, 1 Air, 1 WAGD 74-14-112244 2 Oxygen, 2 Vacuum, 2 Air Product Specification Digital Alarm System – Area Line Pressure Alarms With Gas-specific Transducers All specifications are nominal and subject to change without notice. Warranty: See Allied Statement of Warranties for details. © 2003 Allied Healthcare Products, Inc. Printed in Customer Service: USA: (800) 444-3940 FAX: (800) 477-7701 Canada/Mexico: (800) 446-0552 FAX: (800) 246-6201 International: (314) 268-1683 FAX: (314) 771-5183 1720 Sublette Avenue St. Louis, MO 63110 USA (314) 771-2400 www.alliedhpi.com ISO 9001 CERTIFIED Product Specification Form 74-00-3007 June