micromachines Article A Reduced Graphene Oxide Based Radio Frequency Glucose Sensing Device Using Multi-Dimensional Parameters Byeongho Park 1 , Hyung Goo Park 1 , Jae-hoon Ji 1 , Jinsoo Cho 2 and Seong Chan Jun 1, * 1 2 * School of Mechanical Engineering, Yonsei University, 50 Yonsei-ro, Seodaemun-gu, Seoul 120-749, Korea; enggis@naver.com (B.P.); zeustg@naver.com (H.G.P.); jupitersun@naver.com (J.J.) Department of Computer Engineering, Gachon University, Gyeonggi-do 461-701, Korea; jscho@gachon.ac.kr Correspondence: scj@yonsei.ac.kr; Tel: +82-2-2123-5817 Academic Editor: Hiroshi Mizuta Received: 30 May 2016; Accepted: 1 August 2016; Published: 5 August 2016 Abstract: A reduced graphene oxide (RGO) based glucose sensor using a radio frequency (RF) signal is demonstrated. An RGO with outstanding electrical property was employed as the interconnector material between signal electrodes in an RF electric circuit, and it was functionalized with phenylbutyric acid (PBA) as a linker molecule to bind glucoses. By adding glucose solution, the fabricated sensor with RGO and PBA showed detecting characteristics in RF signal transmission and reflection. Frequency dependent electrical parameters such as resistance, inductance, shunt conductance and shunt capacitance were extracted from the RF results under the equivalent circuit model. These parameters also provided sensing characteristics of glucose with different concentrations. Using these multi-dimensional parameters, the RF sensor device detected glucose levels in the range of 1–4 mM, which ordinarily covers the testing range for diabetes or medical examination. The RGO based RF sensor, which fits well to a linear curve with fine stability, holds considerable promise for biomaterials detection, including glucose. Keywords: glucose biosensor; multi-dimensional parameters; radio frequency; reduced graphene oxide 1. Introduction Glucose sensing from blood and urine is the usual diagnosis method for diabetes, and it is also used in the food industry, such as in fermentation for quality checking [1]. Various sensing techniques for glucose have been developed, such as fluorescence detection [2], electrochemistry [3], and surface-enhanced Raman scattering [4]. However, such techniques needed a mediator including enzymes or antibodies, which result in signal loss with electron- or photon-scattering [5,6]. To overcome this limitation, a single wall carbon nanotube with a high electrical property was employed as an interconnector in the sensor, which detected glucose with pico-molar sensitivity [7,8]. Similarly, reduced graphene oxide (RGO) with two-dimensional structure has also provided significant potential for biosensors, owing to its small signal loss under high electrical property and fine selectivity with specific binding affinity [9–11]. RGO has been applied to biosensors in several research areas, such as glucose sensors [12,13], gas sensors [14–16], and electrochemical bacteria sensors [17]. Aromatic ring structure and functional groups in RGO assist in connecting with the target material by π–π* stacking on its basal planes and chemical interaction [18]. Phenylboronic acid (PBA), which also has an aromatic ring structure, is a highly affinitive molecule with RGO, and it also makes the binding with glucose easily because of the diol group [19,20]. On the basis of the above considerations, we report here a development of the glucose sensor composed of RGO and PBA interconnector in a radio frequency (RF) system. In the RF measurement, Micromachines 2016, 7, 136; doi:10.3390/mi7080136 www.mdpi.com/journal/micromachines Micromachines 2016, 7, 136 2 of 8 Micromachines 2016, 7, 136 2 of 9 On the basis of the above considerations, we report here a development of the glucose sensor composed of RGO and PBA interconnector in a radio frequency (RF) system. In the RF measurement, scattering parameter parameter(S-parameter), (S-parameter), ratio of output to for input for RFtransmission signal transmission and scattering thethe ratio of output to input RF signal and reflection reflectionthe through the interconnector, obtained thekHz range of GHz. 500 kHz 4.5 GHz. Such through interconnector, was obtainedwas in the range ofin500 to 4.5 Suchto measurement has measurement has exceptional characteristics forsensing simple by anddropping rapid sensing dropping small volume exceptional characteristics for simple and rapid small by volume of solution sample of solution sample and continuous test with a wireless platform inmicrowave the ultraand continuous real-time test with a real-time wireless examination platformexamination in the ultra-broadband broadbandIn microwave In particular, RF results in highremarkable frequencies and provide remarkable spectrum. particular,spectrum. the RF results in high the frequencies provide sensitive signal and sensitive signal changefeature with the resonance feature [21]. can also be regarded as a change with the resonance [21]. In addition, it can alsoInbeaddition, regardeditas a valuable system since valuable system multi–parameters, since the decomposed multi–parameters, (R), inductance the decomposed resistance (R), inductanceresistance (L), shunt conductance (G) (L), and shunt conductance(C) (G)allow and shunt capacitance (C) allow usprecisely to analyze the results more precisely [22]. capacitance us to analyze the results more [22]. 2. ExperimentalSection Section 2. Experimental D´(+) glucose(C(C purchased from Sigma Aldrich Korea (Yongin, Korea), andused we 6H 126O 6 ) was D−(+) glucose 6H 12O ) was purchased from Sigma Aldrich Korea (Yongin, Korea), and we used several concentrations of glucose solution 100 µM 4 mM since usual reference range several concentrations of glucose solution fromfrom 100 μM to 4to mM since the the usual reference range of of concentration in blood is close to 2.6 mM. The phosphate buffer tablets (10 mM phosphate, 150 mM concentration in blood is close to 2.6 mM. The phosphate buffer tablets (10 mM phosphate, 150 mM sodium sodium chloride, chloride, pH pH 7.3 7.3 to to 7.5) 7.5) were were from from Seoulin Seoulin bioscience bioscience (Seoul, (Seoul, Korea). Korea). Our Our interconnector interconnector material prepared by material composed composed of of graphene graphene oxide oxide (GO) (GO) was was prepared by the the modified modified hummer hummer method method [23], [23], and and it was reduced by annealing, as in a previous work [24–26]. The high resistivity Si substrate purchased it was reduced by annealing, as in a previous work [24–26]. The high resistivity Si substrate purchased from Woori Woori Material Material (Daejeon, (Daejeon, Korea) Korea) with with 500 500 nm nm dielectric dielectric layer layer (SiO (SiO22) for obtaining the stable electrical signal in RF was employed (Figure 1a). Gold electrodes (500 nm) were deposited by electron beam sputtering on the titanium adhesion layer (10 nm). Our Our electrode electrode pattern pattern had had two two signal signal (S) lines with a 4 μm µm gap on the center center of of the the pattern pattern and and ground ground (G) (G) lines, lines, named named ground-signal-ground ground-signal-ground (GSG). To connect the signal line using RGO as an interconnector, the Dielectrophoresis (DEP) method (GSG). To connect the signal line using RGO as an interconnector, the Dielectrophoresis (DEP) by alternating electric fields was used to focus the RGO flakes on the target position and to form method by alternating electric fields was used to focus the RGO flakes on the target position andthe to channel a shortin time [15].time For the RGO suspended solution (3 µL) was form thein channel a short [15].dielectrophoresis, For the dielectrophoresis, RGO suspended solution (3dropped μL) was on the middle of the substrate, an alternating (AC) current through(AC) signalthrough lines was applied under dropped on the middle of theand substrate, and ancurrent alternating signal lines was 10 V of potential KHz of frequency forof2frequency min. applied under 10and V of10potential and 10 KHz for 2 min. Figure 1. (a) Schematic of radio frequency (RF) measurement system with transmission and reflection Figure 1. (a) Schematic of radio frequency (RF) measurement system with transmission and reflection between signal signal electrodes electrodes on on ground-signal-ground ground-signal-ground (GSG) pattern (b) (b) chemical chemical reaction reaction between between between (GSG) pattern phenylbutyric acid (PBA) and glucose. phenylbutyric acid (PBA) and glucose. For planting PBA on the RGO surface, 4-phenylbutyric acid (C6H5(CH2)3COOH, Sigma Aldrich, For planting PBA on the RGO surface, 4-phenylbutyric acid (C6 H5 (CH2 )3 COOH, Sigma Aldrich, St. Louis, MO, USA) and 3-aminophenylboronic acid (C6H8BNO2, Sigma Aldrich, St. Louis, MO, USA) St. Louis, MO, USA) and 3-aminophenylboronic acid (C6 H8 BNO2 , Sigma Aldrich, St. Louis, MO, USA) were dissolved in N,N-dimethylmethanamide (C3H7NO, DMF, Sigma Aldrich, St. Louis, MO, USA), were dissolved in N,N-dimethylmethanamide (C3 H7 NO, DMF, Sigma Aldrich, St. Louis, MO, USA), and then N-(3-Dimethylaminopropyl)-N’-ethylcarbodiimide hydrochloride (C8H17N3·HCl, EDAC, and then N-(3-Dimethylaminopropyl)-N’-ethylcarbodiimide hydrochloride (C8 H17 N3 ¨HCl, EDAC, Sigma Aldrich, St. Louis, MO, USA) was continuously added. We dropped this solution (10 μL) onto Sigma Aldrich, St. Louis, MO, USA) was continuously added. We dropped this solution (10 µL) onto the RGO surface in the sensing device, and kept it for 2 h at room temperature. The functional groups the RGO surface in the sensing device, and kept it for 2 h at room temperature. The functional groups of RGO assisted in attaching PBA on it with chemical interactions. After that, it was rinsed twice with of RGO assisted in attaching PBA on it with chemical interactions. After that, it was rinsed twice pure DMF and distilled water to remove the reagents. Then, using a sensor device with PBA on RGO with pure DMF and distilled water to remove the reagents. Then, using a sensor device with PBA on (PBA/RGO), we measured the S-parameter, which is the ratio of output to input signal between RGO (PBA/RGO), we measured the S-parameter, which is the ratio of output to input signal between Micromachines 2016, 7, 136 Micromachines 2016, 7, 136 3 of 9 3 of 8 connected connected ports, ports, by by GSG GSG probe probe with with network network analyzer analyzer (E5071C, (E5071C, Agilent, Agilent, Santa Santa Clara, Clara, CA, USA) after dropping onon thethe device surface. TheThe network analyzer generated the microwave signal droppingglucose glucosesolution solution device surface. network analyzer generated the microwave and detected the returning signalsignal from the sensor device with with GSG GSG pattern. As shown in Figure 1a, the signal and detected the returning from the sensor device pattern. As shown in Figure 1a, reflected S-parameter (S11)(S11) was was obtained fromfrom port 1port to 11again, and the S-parameter (S21) the reflected S-parameter obtained to 1 again, andtransmitted the transmitted S-parameter was to 2.1 By the between PBAPBA and and glucose molecule described in Figure 1b, (S21)from was port from1port to 2. Byinteraction the interaction between glucose molecule described in Figure glucose molecules werewere caught on the interconnector. 1b, glucose molecules caught on PBA/RGO the PBA/RGO interconnector. For For monitoring monitoring the structure structure of of RGO RGO on on the substrate, we measured measured the field field emission emission scanning scanning electron (SEM) (JSM-6701F, JEOLJEOL Ltd., Tokyo, Japan) and transmission electron microscope electron microscope microscope (SEM) (JSM-6701F, Ltd., Tokyo, Japan) and transmission electron (TEM) (JEM-2010, Ltd., Tokyo, Japan) images, and atomicand force microscope (AFM) (XE-BiO, microscope (TEM) JEOL (JEM-2010, JEOL Ltd., Tokyo, Japan) images, atomic force microscope (AFM) Park Systems, X-ray photoelectron spectroscopy (K-alpha, Thermo Scientific, (XE-BiO, Park Suwon, Systems,Korea). Suwon, Korea). X-ray photoelectron spectroscopy (K-alpha,Fisher Thermo Fisher Waltham, USA) was characterize the graphene sample. Scientific, MA Waltham, MA used USA)towas used to characterize thebased graphene based sample. 3. Resultsand andDiscussion Discussion 3. Results Figure present SEM SEM images images ofofstretched stretchedtwo twosignal signallines linesand and GSG electrode pattern. Figure 2a,b 2a,b present GSG electrode pattern. A A PBA/RGO interconnector connectedbetween betweentwo twosignal signalelectrodes electrodesand andPBA PBA built built up up the the entangled PBA/RGO interconnector connected entangled structure interconnector was structure on on the the RGO RGO sheet. sheet. The The thickness thickness of of the the PBA/RGO PBA/RGO interconnector was 110 110 nm, nm, as as described described in in Figure Figure 2c. 2c. Figure 2. 2. SEM image of of (a) (a) biosensor biosensor device device with with GSG GSG pattern, pattern,(inset) (inset)signal signalline, line,and and(b) (b)PBA/reduced PBA/reduced Figure SEM image graphene oxide (RGO) interconnector part (c) AFM image of PBA/RGO interconnector. graphene oxide (RGO) interconnector part (c) AFM image of PBA/RGO interconnector. With C 1s, N 1s, and B 1s orbitals, Figure 3 represented the X-ray photoelectron spectra (XPS) With C 1s, N 1s, and B 1s orbitals, Figure 3 represented the X-ray photoelectron spectra (XPS) results, which provides the information of chemical interaction, of PBA/RGO and glucose combined results, which provides the information of chemical interaction, of PBA/RGO and glucose combined PBA/RGO. The C 1s results completely match with sp22 C–C bonding corresponding to 284.8 eV, PBA/RGO. The C 1s results completely match with sp C–C 2bonding corresponding to 284.8 eV, C–OH/O–C–O, and COOH near 286 eV [27]; the intensity of sp at 284.8 eV was the most dominant C–OH/O–C–O, and COOH near 286 eV [27]; the intensity of sp2 at 284.8 eV was the most dominant in all features of the carbon based sample. PBA/RGO was reduced by adding glucose solution, and in all features of the carbon based sample. PBA/RGO was reduced by adding glucose solution, its shoulder peak at 286.7 eV decreased owing to new binding between PBA and glucose in the and its shoulder peak at 286.7 eV decreased owing to new binding between PBA and glucose in the replacement of PBA/RGO functional groups [28,29]. The N 1s features consisted of three individual replacement of PBA/RGO functional groups [28,29]. The N 1s features consisted of three+ individual peaks appearing at 398.2 eV for =N–, at 399.7 eV for –NH–, and at 401.1 eV for NH3 . PBA/RGO peaks appearing at 398.2 eV for =N–, at 399.7 eV for –NH–, and at 401.1 eV for NH3 + . PBA/RGO exhibited mainly =N– feature at 398.2 eV from PBA before adding glucose. After glucose treatment, exhibited mainly =N– feature at 398.2 eV from PBA before adding glucose. After glucose treatment, PBA/RGO showed decrement in =N– binding, and small –NH– features remain in the N 1s orbital of PBA/RGO showed decrement in =N– binding, and small –NH– features remain in the N 1s orbital of XPS characterization. In the B 1s, PBA/RGO with glucose presented –C–B(OH)2 feature at 191.6 eV XPS characterization. In the B 1s, PBA/RGO with glucose presented –C–B(OH)2 feature at 191.6 eV [30]. [30]. In PBA/RGO, 3–aminophenylboronic acid that is able to interact chemically with glucose In PBA/RGO, 3–aminophenylboronic acid that is able to interact chemically with glucose molecule molecule has the nitrogen in amino and boron in boronic acid. The XPS features of these two atoms has the nitrogen in amino and boron in boronic acid. The XPS features of these two atoms remarkably remarkably showed the existence of a functional group on PBA/RGO and interaction with glucose showed the existence of a functional group on PBA/RGO and interaction with glucose [31]. [31]. From two signal electrodes or ports on the RF device, a network analyzer can produce the input From two signal electrodes or ports on the RF device, a network analyzer can produce the input electrical power and receive the output power. The magnitude is determined by the ratio of output electrical power and receive the output power. The magnitude is determined by the ratio of output to input signal power. Each port, i and j, can provide four types of S-parameters such as S11 , S12 , S21 , to input signal power. Each port, i and j, can provide four types of S-parameters such as S11, S12, S21, and S22 . These parameters are the reflection (i = j) and the transmission (i ‰ j). S11 and S21 can be and S22. These parameters are the reflection (i = j) and the transmission (i ≠ j). S11 and S21 can be considered as a representative parameter individually to demonstrate all components in S parameters considered as a representative parameter individually to demonstrate all components in S parameters of the RF device. Figure 4 presents S-parameters of RGO and PBA/RGO devices with increasing of the RF device. Figure 4 presents S-parameters of RGO and PBA/RGO devices with increasing glucose solution from 0 to 4 mM. According to the property of interconnector material, S-parameters glucose solution from 0 to 4 mM. According to the property of interconnector material, S-parameters were changed; intrinsic GO had similar characteristics to the open circuit, RGO and PBA/RGO showed were changed; intrinsic GO had similar characteristics to the open circuit, RGO and PBA/RGO showed RGO dominant S11 and S21 results (Figure S1). RGO based devices provided better signal transmission with higher conductivity. Micromachines 2016, 7, 136 4 of 9 RGO dominant S11 and S21 results (Figure S1). RGO based devices provided better signal transmission with higher2016, conductivity. Micromachines 7, 136 4 of 8 Micromachines 2016, 7, 136 4 of 8 Figure Figure 3. 3. XPS XPS results results of of PBA/RGO PBA/RGOinterconnector interconnectorbefore before and and after after adding adding glucose glucose in in (a,b) (a,b) C C 1s, 1s; (c) (c) N N 1s, and (d) B 1s orbital. 1s; and (d) B 1s orbital. Figure 3. XPS results of PBA/RGO interconnector before and after adding glucose in (a,b) C 1s, (c) N 1s, and (d) B 1s orbital. Figure 4. Up to 4.5 GHz, RF measurement result of RGO and PBA/RGO RF device. (a) Signal reflection (signal from port 1 to port 1, S11) and (b)result transmission (signal from port 1 to port 2, S21).reflection Glucose of RGO RGO and and PBA/RGO RF device. (a) Figure 4. Up to 4.5 GHz, RF measurement result of PBA/RGO (a) Signal Signal solution (3 μL) from 1 to 4 mM was dropped on the central position of the PBA/RGO device. (signal from port 1 to to port port 1, 1, S11) S11) and and (b) (b) transmission transmission (signal (signal from from port port 11 to to port port 2, 2, S21). S21). Glucose Glucose solution (3 µL) μL) from from 11 to to 44 mM mM was was dropped dropped on on the the central central position position of ofthe thePBA/RGO PBA/RGO device. After adding glucose solution, the transmission was significantly increased through the whole frequency reflection parameter in Figure 4a showed that theincreased reflectionthrough property not After range. addingThe glucose solution, the transmission was significantly thedid whole After adding glucose solution, the transmission was significantly increased through the whole considerably change under 1 GHz, and over 2 GHz the reflection magnitudes of glucose solutions frequency range. The reflection parameter in Figure 4a showed that the reflection property did not frequency range. The reflection parameter in Figure 4a showed that the reflection property did not were substantially decreased that2ofGHz the bare PBA/RGOmagnitudes line. PBA/RGO device provided considerably change under 1compared GHz, andtoover the reflection of glucose solutions considerably change under 1 GHz, and over 2 GHz the reflection magnitudes of glucose solutions considerable change in the S11 and S21 result after dropping glucose solution. Such S-parameter were substantially decreased compared to that of the bare PBA/RGO line. PBA/RGO device provided were substantially decreased compared to that of the bare PBA/RGO line. PBA/RGO device provided change indicates that the transmission increased decrement of signal loss in the interconnector. considerable change in the S11 and S21 result with after adropping glucose solution. Such S-parameter considerable change in the S11 and S21 result after dropping glucose solution. Such S-parameter The interaction between PBA and glucose assisted in reducing the signal loss and the signal reflection change indicates that the transmission increased with a decrement of signal loss in the interconnector. change indicates that the transmission increased with a decrement of signal loss in the interconnector. in the PBA/RGO interconnector [32,33]. Many hydroxyl groups of the diol, which significantly reduce The interaction between PBA and glucose assisted in reducing the signal loss and the signal reflection The interaction between PBA and glucose assisted in reducing the signal loss and the signal reflection the conductivity of graphene and interfere transmission, in PBA/RGO removed with in the PBA/RGO interconnector [32,33]. Manysignal hydroxyl groups of the diol, whichwere significantly reduce in the PBA/RGO interconnector [32,33]. Many hydroxyl groups of the diol, which significantly reduce condensation reaction of PBA and glucose [34]. As the concentration of glucose increases from 1 to 4 the conductivity of graphene and interfere signal transmission, in PBA/RGO were removed with the conductivity of graphene and interfere signal transmission, in PBA/RGO were removed with mM, S11 magnitude of PBA/RGO device was decreased. Although variations also occurred in the S214 condensation reaction of PBA and glucose [34]. As the concentration of glucose increases from 1 to condensation reaction of PBA and glucose [34]. As the concentration of glucose increases from 1 to result, it ismagnitude too small of to PBA/RGO conceive asdevice a sensor mM, S11 wascharacteristic. decreased. Although variations also occurred in the S21 4 mM, S11 magnitude of PBA/RGO device was decreased. Although variations also occurred in the result, it is too small to conceive as a sensor characteristic. S21 result, it Signal is too small to conceive as a sensor characteristic. 4. Electrical of RLGC 4. Electrical Signal of RLGC Since the S11 and S21 results provided too little variation to be used for sensor application, R, L, G, and C were extracted toresults investigate the specific sensor characteristics PBA/RGO RF device. Since the S11 and S21 provided too little variation to be used of forthe sensor application, R, L, From results, those four parameters were characteristics calculated withofthe G, andthe C S-parameter were extracted to investigate the specific sensor thefollowing PBA/RGOequation RF device. From the S-parameter results, those four parameters the following equation 2 2were calculated2 with 2 2 (1 − S11 + S21 ) Z0 ((1 + S11 ) − S21 A B 1 2 2 = 1 (1 − S + S ) Z0 ((1 + S2 112 )2 − S2 CA D (1 − S11 + S212 ) 21 B 2S121 ((1 − S1111 )2 −21S212 ) C D = 2S Z10 ((1 − S )2 − S 2 ) (1 − S112 + S212 ) 11 21 21 Z (1) (1) Micromachines 2016, 7, 136 5 of 9 4. Electrical Signal of RLGC Since the S11 and S21 results provided too little variation to be used for sensor application, R, L, G, and C were extracted to investigate the specific sensor characteristics of the PBA/RGO RF device. From the S-parameter results, those four parameters were calculated with the following equation « A C Micromachines 2016, 7, 136 B D ff 1 “ 2S21 « 2 ` S2 q p1 ´ S11 21 2 1 2 Z0 pp1 ´ S11 q ´ S21 q 2 2 q ´ S2 Z0 pp1 ` S11 21 2 ` S2 q p1 ´ S11 21 ff (1) 5 of 8 where Then the theimpedance impedance ZZ and where ZZ00 indicate indicatethe theinitial initialload load50 50Ω Ω for for GSG GSG probing probing [35–37]. [35–37]. Then and RF RF propagation constant γ was computed by, γ was computed by, propagation constant c B ´−11 γγ “ (2) = cos co s A, A ,Z Z“= B (2) C C From telegraphermodel, model, the properties of RF electronic were obtained by From telegrapher the properties of RF electronic circuit werecircuit obtained by de-embedding, de-embedding, removing the fixed effect and taking meaningful values from the measured results. removing the fixed effect and taking meaningful avalues from the measured results. The relationship pR ` jωLq { pG ` jωCq and propagation constant, The relationship between impedance, Z “ between a impedance, Z = ( R + jωL) / ( G + jωC) and propagation constant, γ = ( R + jωL)( G + jωC ) is γ “ pR ` jωLq pG ` jωCq is composed of the standard electronic features of R, L, G and C. composed of the standard electronic features of R, L, G and C. These features were extracted as These features were extracted as follows, follows, Im{tγZu RR =“RRe e{tγZu γ Z },,LL =“Im γ Z } /{ω, ω, tγ{Zu , C “ Im tγ{Zu G “ Re G = R e{γ / Z }, C = Im {γ / Z }{ω /ω (3) (3) R considerably related withwith the interconnector’s electrical properties, G and G C values R and andLLvalues valueswere were considerably related the interconnector’s electrical properties, and C were obtained from thefrom interaction between interconnector and substrate. values are represented values were obtained the interaction between interconnector andThese substrate. These values are in Figure 5. in Figure 5. represented Figure 5. 5. (a) (a) Resistance, Resistance, R; R;(b) (b)inductance, inductance,L;L;(c) (c)shunt shuntconductance, conductance,G;G; and shunt capacitance, Figure and (d)(d) shunt capacitance, C, C, of of PBA/RGO device according to adding glucose with different concentrations from 0 to 4 mM. PBA/RGO device according to adding glucose with different concentrations from 0 to 4 mM. According to increasing frequency, the resistances were decreased since the electron According to increasing frequency, the resistances were decreased since the electron transmission transmission was significantly enhanced at high frequency, as shown in Figure 5a. Near the 2.75 GHz was significantly enhanced at high frequency, as shown in Figure 5a. Near the 2.75 GHz measurement measurement frequency, the inductance result of PBA/RGO device showed a specific dip which originated from the resonance characteristic. As the concentration of glucose increased, the resistance (Figure 5a) generally decreased and the conductance (Figure 5c) and capacitance (Figure 5d) increased owing to the enhancement of conductivity with the RGO reduction, which resulted from the condensation reaction (Figure 1b) between PBA and glucose molecules. In particular, in Figure 5b, the inductance provided a resonance Micromachines 2016, 7, 136 6 of 9 frequency, the inductance result of PBA/RGO device showed a specific dip which originated from the resonance characteristic. As the concentration of glucose increased, the resistance (Figure 5a) generally decreased and the conductance (Figure 5c) and capacitance (Figure 5d) increased owing to the enhancement of conductivity with the RGO reduction, which resulted from the condensation reaction (Figure 1b) between PBA and glucose molecules. In particular, in Figure 5b, the inductance provided a resonance dip shifting according to glucose concentration. The dip shift to a lower frequency can be demonstrated through the relation between the resonance features and the viscosity of glucose solution [38,39]. Adding glucose on the PBA/RGO interconnector surface results in the enhancement of viscosity and acts as a damper, especially in the condition with the boronic acid [40]. Under the RF system,6the Micromachines 2016, 7, 136 of 8 conduction transferring the carrier to the opposite port was interfered with by the inductance and took a longer time when inductance of the biosensor device hadwas a high magnitude. Linearity of thethe PBA/RGO glucose investigated with the R-square between the Linearity of the PBA/RGO glucose biosensor was investigated with R-square between the fitted results and experimental data according to glucose concentration. Thethe linearity of the resistance fitted data according to glucose linearity of the resistance resultresults (Figureand 6a)experimental showed the highest linear response forconcentration. glucose with The 95.45% of R-square, and that result (Figure 6a) showed the highest linear response for glucose with 95.45% of R-square, and that of of the inductance result (Figure 6b) corresponded to 93.23%. Shunt conductance shunt the inductance result (Figure 6b) interaction corresponded to 93.23%. Shunt conductance and shunt capacitance, capacitance, determined by the between interconnector and substrate, provided worse determined linearity. by the interaction between interconnector and substrate, provided worse linearity. Figure 6. 6. Linearity Linearity of of (a) (a) resistance, resistance, R; R; (b) (b) resonance resonance feature feature of of inductance, inductance, L; L; (c) (c) shunt shunt conductance, conductance, Figure G; and (d) shunt capacitance, C, of PBA/RGO device according to adding glucose with different G; and (d) shunt capacitance, C, of PBA/RGO device according to adding glucose with different concentrations from 1 to 4 mM. concentrations from 1 to 4 mM. Up to 100 μM level of glucose concentration, our PBA/RGO glucose sensor showed insufficient Up to 100 µM level of glucose concentration, our PBA/RGO glucose sensor showed insufficient linearity with RF results, as shown in Table 1. Over the milli-molar concentration, the linear linearity with RF results, as shown in Table 1. Over the milli-molar concentration, the linear relationship relationship between RF results and glucose concentration was remarkably revealed with high Rbetween RF results and glucose concentration was remarkably revealed with high R-square value in square value in the resistance and the resonance feature of inductance. Generally, linearity variations the resistance and the resonance feature of inductance. Generally, linearity variations were increased were increased under high glucose concentration. While the sensing ability of the resistance was under high glucose concentration. While the sensing ability of the resistance was balanced with a balanced with a small range of error and fine linearity, the resonance feature of inductance exhibited small range of error and fine linearity, the resonance feature of inductance exhibited unstable results unstable results with high fluctuation. The PBA/RGO glucose sensor showed a detection limit of 3 × with high fluctuation. The PBA/RGO glucose sensor showed a detection limit of 3 ˆ 10´5 mol/L, 10−5 mol/L, which is higher than that of the electrochemical glucose sensor composed of a carbon which is higher than that of the electrochemical glucose sensor composed of a carbon nanotube [41]. nanotube [41]. Table 1: R-square of resistance (R), resonance feature of inductance (L), shunt conductance (G), and shunt capacitance (C). Linearity (R-Square, %) Range of Glucose Concentration (mM) R L G C 0.1–0.4 83.15 ± 4.22 69.04 ± 31.41 23.80 ± 2.28 21.51 ± 6.15 Micromachines 2016, 7, 136 7 of 9 Table 1. R-square of resistance (R), resonance feature of inductance (L), shunt conductance (G), and shunt capacitance (C). Linearity (R-Square, %) Range of Glucose Concentration (mM) R L G C 0.1–0.4 1–4 83.15 ˘ 4.22 95.45 ˘ 9.60 69.04 ˘ 31.41 93.23 ˘ 29.37 23.80 ˘ 2.28 69.28 ˘ 42.54 21.51 ˘ 6.15 86.49 ˘ 19.82 5. Conclusions In the present study, we demonstrated a glucose sensor using an RGO based RF system. RGO as an interconnector was functionalized with a PBA linker in order to catch the glucose molecules. The fabricated sensor showed detecting characteristics with fine linearity of the RF signal change as glucose solution was added. From measurement of the RF signal, several parameters such as R, L, G and C can be extracted by the equivalent circuit model. Under a high frequency domain, the RGO based RF sensor could sensitively detect small glucose concentration using multi-dimensional analysis with the four electrical parameters. The resistance was the most effective parameter for monitoring the glucose level with stable linearity and small fluctuation. The developed RF sensor can open up new possibilities in rapid and sensitive detection by applying various interconnector materials with specific binding affinity. Supplementary Materials: The following are available online at http://www.mdpi.com/2072-666X/7/8/136/s1, Figure S1: S-parameter results, (a) S11; (b) S21, of several electrodes with open without interconnector, GO, RGO, PBA/RGO. Acknowledgments: This work was partially supported by the Priority Research Centers Program (2009-0093823), the Korean Government (MSIP) (No. 2015R1A5A1037668) through the National Research Foundation of Korea (NRF) funded by the Ministry of Education, Science and Technology (MEST), and the ICT R&D program of MSIP/IITP (R0101-15-0034). Author Contributions: Byeongho Park wrote and investigated; Hyung Goo Park designed and performed the experiments; Jae-hoon Ji, Jinsoo Cho and Seong Chan Jun discussed results. Conflicts of Interest: The authors declare no conflict of interest. References 1. 2. 3. 4. 5. 6. 7. 8. 9. Harsányi, G. Sensors in Biomedical Applications: Fundamentals, Technology and Applications; CRC Press: Boca Raton, FL, USA, 2000. Khan, F.; Saxl, T.E.; Pickup, J.C. Fluorescence intensity-and lifetime-based glucose sensing using an engineered high-K d mutant of glucose/galactose-binding protein. Anal. Biochem. 2010, 399, 39–43. [PubMed] Wang, S.; Zhang, Q.; Wang, R.; Yoon, S. A novel multi-walled carbon nanotube-based biosensor for glucose detection. Biochem. Biophys. Res. Commun. 2003, 311, 572–576. [CrossRef] [PubMed] Shafer-Peltier, K.E.; Haynes, C.L.; Glucksberg, M.R.; Van Duyne, R.P. Toward a glucose biosensor based on surface-enhanced raman scattering. J. Am. Chem. Soc. 2003, 125, 588–593. [CrossRef] [PubMed] Wang, J. Electrochemical glucose biosensors. Chem. Rev. 2008, 108, 814–825. [CrossRef] [PubMed] Moschou, E.A.; Sharma, B.V.; Deo, S.K.; Daunert, S. Fluorescence glucose detection: Advances toward the ideal in vivo biosensor. J. Fluorescence 2004, 14, 535–547. [CrossRef] Cella, L.N.; Chen, W.; Myung, N.V.; Mulchandani, A. Single-walled carbon nanotube-based chemiresistive affinity biosensors for small molecules: Ultrasensitive glucose detection. J. Am. Chem. Soc. 2010, 132, 5024–5026. [CrossRef] [PubMed] Mu, Y.; Jia, D.; He, Y.; Miao, Y.; Wu, H.-L. Nano nickel oxide modified non-enzymatic glucose sensors with enhanced sensitivity through an electrochemical process strategy at high potential. Biosens. Bioelectron. 2011, 26, 2948–2952. [CrossRef] [PubMed] Gómez-Navarro, C.; Meyer, J.C.; Sundaram, R.S.; Chuvilin, A.; Kurasch, S.; Burghard, M.; Kern, K.; Kaiser, U. Atomic structure of reduced graphene oxide. Nano Lett. 2010, 10, 1144–1148. [CrossRef] [PubMed] Micromachines 2016, 7, 136 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21. 22. 23. 24. 25. 26. 27. 28. 29. 30. 31. 32. 8 of 9 Morozov, S.; Novoselov, K.; Katsnelson, M.; Schedin, F.; Elias, D.; Jaszczak, J.; Geim, A. Giant intrinsic carrier mobilities in graphene and its bilayer. Phys. Rev. Lett. 2008, 100, 016602. [CrossRef] [PubMed] Novoselov, K.; Geim, A.K.; Morozov, S.; Jiang, D.; Katsnelson, M.; Grigorieva, I.; Dubonos, S.; Firsov, A. Two-dimensional gas of massless dirac fermions in graphene. Nature 2005, 438, 197–200. [CrossRef] [PubMed] Lu, L.-M.; Li, H.-B.; Qu, F.; Zhang, X.-B.; Shen, G.-L.; Yu, R.-Q. In situ synthesis of palladium nanoparticle-graphene nanohybrids and their application in nonenzymatic glucose biosensors. Biosens. Bioelectron. 2011, 26, 3500–3504. [CrossRef] [PubMed] Luo, J.; Jiang, S.; Zhang, H.; Jiang, J.; Liu, X. A novel non-enzymatic glucose sensor based on Cu nanoparticle modified graphene sheets electrode. Anal. Chim. Acta 2012, 709, 47–53. [CrossRef] [PubMed] Lu, G.; Ocola, L.E.; Chen, J. Gas detection using low-temperature reduced graphene oxide sheets. Appl. Phys. Lett. 2009, 94, 083111. [CrossRef] Park, H.G.; Hwang, S.; Lim, J.; Kim, D.-H.; Song, I.S.; Kim, J.H.; Woo, D.H.; Lee, S.; Jun, S.C. Comparison of chemical vapor sensing properties between graphene and carbon nanotubes. Jpn. J. Appl. Phys. 2012, 51, 045101. [CrossRef] Lu, G.; Ocola, L.E.; Chen, J. Reduced graphene oxide for room-temperature gas sensors. Nanotechnology 2009, 20, 445502. [CrossRef] [PubMed] Wan, Y.; Wang, Y.; Wu, J.; Zhang, D. Graphene oxide sheet-mediated silver enhancement for application to electrochemical biosensors. Anal. Chem. 2010, 83, 648–653. [CrossRef] [PubMed] Chen, R.J.; Zhang, Y.; Wang, D.; Dai, H. Noncovalent sidewall functionalization of single-walled carbon nanotubes for protein immobilization. J. Am. Chem. Soc. 2001, 123, 3838–3839. [CrossRef] [PubMed] Zayats, M.; Katz, E.; Willner, I. Electrical contacting of flavoenzymes and nad (p)+-dependent enzymes by reconstitution and affinity interactions on phenylboronic acid monolayers associated with au-electrodes. J. Am. Chem. Soc. 2002, 124, 14724–14735. [CrossRef] [PubMed] Li, S.; Davis, E.N.; Anderson, J.; Lin, Q.; Wang, Q. Development of boronic acid grafted random copolymer sensing fluid for continuous glucose monitoring. Biomacromolecules 2008, 10, 113–118. [CrossRef] [PubMed] Lee, H.-J.; Lee, J.-H.; Jung, H.-I. A symmetric metamaterial element-based RF biosensor for rapid and label-free detection. Appl. Phys. Lett. 2011, 99, 163703. [CrossRef] Kim, W.K.; Jung, Y.M.; Cho, J.H.; Kang, J.Y.; Oh, J.Y.; Kang, H.; Lee, H.-J.; Kim, J.H.; Lee, S.; Shin, H. Radio-frequency characteristics of graphene oxide. Appl. Phys. Lett. 2010, 97, 193103. [CrossRef] Tung, V.C.; Allen, M.J.; Yang, Y.; Kaner, R.B. High-throughput solution processing of large-scale graphene. Nat. Nanotechnol. 2009, 4, 25–29. [CrossRef] [PubMed] Wang, X.; Zhi, L.; Müllen, K. Transparent, conductive graphene electrodes for dye-sensitized solar cells. Nano Lett. 2008, 8, 323–327. [CrossRef] [PubMed] Park, B.; Kim, S.J.; Lim, J.; Some, S.; Park, J.-E.; Kim, S.-J.; Kim, C.; Lee, T.J.; Jun, S.C. Tunable wide blue photoluminescence with europium decorated graphene. J. Mater. Chem. C 2015, 3, 4030–4038. [CrossRef] Rani, J.; Oh, J.; Park, J.-E.; Lim, J.; Park, B.; Kim, K.; Kim, S.-J.; Jun, S.C. Controlling the luminescence emission from palladium grafted graphene oxide thin films via reduction. Nanoscale 2013, 5, 5620–5627. [CrossRef] [PubMed] Pei, S.; Cheng, H.-M. The reduction of graphene oxide. Carbon 2012, 50, 3210–3228. [CrossRef] Park, S.; An, J.; Potts, J.R.; Velamakanni, A.; Murali, S.; Ruoff, R.S. Hydrazine-reduction of graphite- and graphene oxide. Carbon 2011, 49, 3019–3023. [CrossRef] Gao, X.; Jang, J.; Nagase, S. Hydrazine and thermal reduction of graphene oxide: Reaction mechanisms, product structures, and reaction design. J. Phys. Chem. C 2010, 114, 832–842. [CrossRef] Sun, X.-Y.; Liu, B.; Jiang, Y.-B. An extremely sensitive monoboronic acid based fluorescent sensor for glucose. Anal. Chim. Acta 2004, 515, 285–290. [CrossRef] Stephenson-Brown, A.; Wang, H.-C.; Iqbal, P.; Preece, J.A.; Long, Y.; Fossey, J.S.; James, T.D.; Mendes, P.M. Glucose selective surface plasmon resonance-based bis-boronic acid sensor. Analyst 2013, 138, 7140–7145. [CrossRef] [PubMed] Zhang, C.; Losego, M.D.; Braun, P.V. Hydrogel-based glucose sensors: Effects of phenylboronic acid chemical structure on response. Chem. Mater. 2013, 25, 3239–3250. [CrossRef] Micromachines 2016, 7, 136 33. 34. 35. 36. 37. 38. 39. 40. 41. 9 of 9 Shiino, D.; Murata, Y.; Kataoka, K.; Koyama, Y.; Yokoyama, M.; Okano, T.; Sakurai, Y. Preparation and characterization of a glucose-responsive insulin-releasing polymer device. Biomaterials 1994, 15, 121–128. [CrossRef] Liu, H.; Zhang, L.; Guo, Y.; Cheng, C.; Yang, L.; Jiang, L.; Yu, G.; Hu, W.; Liu, Y.; Zhu, D. Reduction of graphene oxide to highly conductive graphene by lawesson’s reagent and its electrical applications. J. Mater. Chem. C 2013, 1, 3104–3109. [CrossRef] Dworsky, L.N. Modern Transmission Line Theory and Applications; RE Krieger Pub. Co.: Malabar, FL, USA, 1979. Eisenstadt, W.R.; Eo, Y. S-parameter-based IC interconnect transmission line characterization. IEEE Trans. Compon. Hybrids Manuf. Technol. 1992, 15, 483–490. [CrossRef] Jun, S.C.; Choi, J.; Cha, S.; Baik, C.; Lee, S.; Kim, H.J.; Hone, J.; Kim, J. Radio-frequency transmission characteristics of a multi-walled carbon nanotube. Nanotechnology 2007, 18, 255701. [CrossRef] Craddock, G.; Diamond, P. Theory of shear suppression of edge turbulence by externally driven radio-frequency waves. Phys. Rev. Lett. 1991, 67, 1535. [CrossRef] [PubMed] Jaeger, E.F.; Berry, L.A.; Batchelor, D.B. Second-order radio frequency kinetic theory with applications to flow drive and heating in tokamak plasmas. Phys. Plasmas 2000, 7, 641–656. [CrossRef] Parks, G.S.; Barton, L.E.; Spaght, M.E.; Richardson, J.W. The viscosity of undercooled liquid glucose. J. Appl. Phys. 1934, 5, 193–199. [CrossRef] Harper, A.; Anderson, M.R. Electrochemical glucose sensors—Developments using electrostatic assembly and carbon nanotubes for biosensor construction. Sensors 2010, 10, 8248–8274. [CrossRef] [PubMed] © 2016 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC-BY) license (http://creativecommons.org/licenses/by/4.0/).