AdvantEDGE series - Watts Water Technologies

advertisement
SERIES
TM
online measurement electrodes
pH
ORP
Conductivity
Turbidity
Optical Dissolved Oxygen
Dissolved Ozone
Free Chlorine
Dissolved Oxygen
Suspended Solids
For three decades HF scientific has developed specialized expertise in measurements for water quality. Our original
involvement in water-quality focused on Turbidity measurement. These products have positioned us as one of the worldwide
leaders in Turbidity measurement and monitoring. Over the years our expertise expanded to include devices for monitoring
other parameters such as Residual Chlorine, Streaming Current and UV% Transmission. We are now excited to expand our
product offering with the addition of our new AdvantEDGE Series™ of measurement electrodes. These electrode systems
have been designed to reduce equipment and installation costs while improving reliability. Measurement parameters include
pH, ORP, Free Chlorine, Raw Water Turbidity, Suspended Solids, Dissolved Oxygen, Optical Dissolved Oxygen, Dissolved
Ozone, Contacting Conductivity and Toroidal Conductivity.
The AdvantEDGE Series™ systems feature factory pre-calibrated electrodes. The electrodes can subsequently be checked
for calibration, diagnosed and configured directly from a PLC or other computer system (e.g., SCADA). Although they are
available for use if desired, a local display/analyzer is not necessary.
Numerous integrated communications protocols are supported, including Modbus RTU, RS-232, RS-485 ASCII, Profibus,
CANOpen, DeviceNet, Ethernet IP, and USB.
Contact us directly or call the HF scientific representative serving your area to learn how the new AdvantEDGE Series™ can
bring your facility into a new generation of water quality monitoring.
HF scientific
3170 Metro Parkway
Ft. Myers, FL 33916
Ph: 239-337-2116 • Fax: 239-332-7643
Toll Free: 888-203-7248
hf.info@wattswater.com
www.hfscientific.com
Table of Contents:
AdvantEDGE series™
AdvantEDGE series™ Overview.............................................................. 1
pH................................................................................................................ 2
ORP............................................................................................................. 2
Dissolved Ozone .................................................................................... 3
Dissolved Oxygen..................................................................................... 3
Optical Dissolved Oxygen........................................................................ 4 Contacting Conductivity........................................................................... 5
Toroidal Conductivity................................................................................ 5
Suspended Solids..................................................................................... 6
Turbidity..................................................................................................... 6
Free Chlorine Panel................................................................................... 7
LD500 Local Display................................................................................. 8
PA500 Profibus Gateway.......................................................................... 8
Mounting Hardware................................................................................... 9
Accessories & Replacement Parts.......................................................... 9
Specifications and Content Information Subject to change without notice
AdvantEDGE series™
The AdvantEDGE sensor assembly consists of three components designed for direct connection to control systems without
the need for intermediate analyzer electronics. A Communications Module plugs into the back of the Interface Module
and provides direct interactive communication of measurement, configuration, calibration and diagnostic information with
computers and networks. A full family of sensor heads plug into the front of the Interface Module and automatically provides
calibrated, high resolution measurements.
Communications Adapter
Provides power and direct interactive communication with control
systems and computers.
A selection of plug and play communication modules provide
power and direct interactive communications with control
systems and computers. The table below shows the available
protocols. If a local display is desired, the LD500 can be used.
Multiple sensor assemblies can be used with the LD500.
Interface Module
The heart of the sensor assembly is the Universal Interface
Module. It’s encapsulated measurement electronics automatically
converts sensor signals into calibrated measurement units.
It works with a variety of Communication Modules to operate
directly with any computer system.
Sensor Head
Pre-calibrated plug and play sensor heads can be replaced at
any time, making AdvantEDGE the most versatile and accurate
analytical measurement system on the market. The table below
shows the parameters available.
PARAMETERS
INTERFACE
pH
Modbus RTU
ORP
Profibus
Free Chlorine
Turbidity
Suspended Solids
Dissolved Oxygen
Direct
Local Display
Dissolved Ozone
CONTROL
ROOM
RS 485 ASCII
SCADA
CANOpen
PLC
Device Net
DCS
Ethernet IP
PC
USB
Conductivity
Optional LD500
Local Display
COMMUNICATIONS
PROTOCOLS
Typical AdvantEDGE System Assembly
1
4-30-2014 Rev 0.9
AdvantEDGE pH
AdvantEDGE ORP
• Differential pH measurement
• Pre-calibrated (no field calibration
required)
• Plug & play sensor heads
• Replaceable quad junction salt bridges
• Electrode protection options
• Direct data reporting
• Plug & play industrial communications
adapters
• Differential ORP measurement
• Pre-calibrated (no field calibration
required)
• Plug & play sensor heads
• Replaceable quad junction salt bridges
• Electrode protection options
• Direct data reporting
• Plug & play industrial communications
adapters
Connect this pH sensor to a PLC for seamless integration with
industrial control systems. Use any computer to display data,
calibrate and customize the measurement without an intermediate
analyzer electronics box. A LD500 local display can be used if
desired. Sensor heads are pre-calibrated and can be replaced or
exchanged with any other type of sensor without taking the system
down. Save space, time and money.
Connect this ORP sensor to a PLC for seamless integration with
industrial control systems. Use any computer to display data,
calibrate and customize the measurement without an intermediate
analyzer electronics box. A LD500 local display can be used if
desired. Sensor heads are pre-calibrated and can be replaced or
exchanged with any other type of sensor without taking the system
down. Save space, time and money.
Measurement System Performance
Range: 0 to 14 pH
Resolution: 0.01 pH
Accuracy: 0.014 pH
Step Response Time: 90% in 30 seconds
Note: Typical at 25°C
Performance unaffected by cable length
Measurement System Performance
Range: -2100mV to +2100mV
Resolution: 0.1mV
Accuracy: 0.1% of reading
Step Response Time: 90% in 30 seconds
Note: Typical at 25°C
Performance unaffected by cable length
Operational Environment
Temperature Range: -5°C to 95°C
Maximum Pressure: 100 psi @ 95°C 85 psi @ 75°C
Maximum Flow Rate: 10 ft/second
Operational Environment
Temperature Range: -5°C to 95°C
Maximum Pressure: 100 psi @ 95°C 85 psi @ 75°C
Maximum Flow Rate: 10 ft/second
Power Requirements
Voltage Range: 10 to 30 VDC
Maximum Power: 200 mW
Typical Power: 120 mW
Note: Class II DC power supply required
Power Requirements
Voltage Range: 10 to 30 VDC
Maximum Power: 200 mW
Typical Power: 120 mW
Note: Class II DC power supply required
Approvals and Ratings
Immunity & Emissions: CE (EN61326 and EN61326:1998)
Safety: CE (EN 61010-1) UL 508
Hazardous Locations: UL Class I, Division 2
Construction
Process Electrodes: Platinum or Gold
Ground Rod: Titanium, 316 Stainless Steel or Hastelloy C
O-Rings: Viton® (other materials available)
Construction
Process Electrodes: “G”, “HF” or Low Temperature Glass
Ground Rod: Titanium, 316 Stainless Steel or Hastelloy C
O-Rings: Viton® (other materials available)
Units of Measure
Measurement Units: mV
Temperature Units: °C, °F
Calibration
Sample: 1 point
Temperature: 1 and 2 point
Note: ORP and temperature are precalibrated at the factory
Units of Measure
Measurement Units: pH, mV
Temperature Units: °C, °F
Temperature Compensation Options
Temperature measurement is provided as an independent
measurement.
Note: Temperature can be entered manually
Calibration
Automatic Buffer: 1 and 2 point
Sample: 1 and 2 point
Temperature: 1 and 2 point
Note: pH and temperature are precalibrated at the factory
Other Configuration Options
Sensor Filter: 0 to 100 seconds
Temperature Filter: 0 to 100 seconds
Temperature Compensation Options
Linear: % per °C
Built-in tables: Ammonia or Morpholine
Note: Temperature can be entered manually
Approvals and Ratings
Immunity & Emissions: CE (EN61326 and
EN61326:1998)
Safety: CE (EN 61010-1) UL 508
Hazardous Locations: UL Class I, Division 2
Other Configuration Options
Sensor Filter: 0 to 100 seconds
Temperature Filter: 0 to 100 seconds
Auto Calibration Buffer
Standards: (4, 7, 10) and DIN 19267
2
4-30-2014 Rev 0.9
AdvantEDGE Ozone
AdvantEDGE DO
• Clark Cell technology
• Pre-calibrated (no field calibration
required)
• Plug & play sensor heads
• Rugged four resistant membrane
• Electrode protection options
• Simple membrane cartridge replacement
• Direct data reporting
• Plug & play industrial communications
adapters
Connect this Dissolved Ozone sensor to a PLC for seamless
integration with industrial control systems. Use any computer to
display data, calibrate and customize the measurement without an
intermediate analyzer electronics box. A LD500 local display can
be used if desired. Sensor heads are pre-calibrated and can be
replaced or exchanged with any other type of sensor without taking
the system down. Save space, time and money.
• Clark Cell technology
• Pre-calibrated (no field calibration
required)
• Plug & play sensor heads
• Rugged four resistant membrane
• Electrode protection options
• Simple membrane cartridge replacement
• Direct data reporting
• Plug & play industrial communications
adapters
Connect this Dissolved Oxygen sensor to a PLC for seamless
integration with industrial control systems. Use any computer to
display data, calibrate and customize the measurement without an
intermediate analyzer electronics box. A LD500 local display can
be used if desired. Sensor heads are pre-calibrated and can be
replaced or exchanged with any other type of sensor without taking
the system down. Save space, time and money.
Measurement System Performance
Range: 0 to 40 ppm
Resolution: 0.01 ppm
Accuracy: 0.04 ppm
Step Response Time: 90% in 90 seconds
Note: Typical at 25°C
Performance unaffected by cable length
Measurement System Performance
Range: 0 to 40 ppm
Resolution: 0.01 ppm
Accuracy: 0.04 ppm
Step Response Time: 90% in 90 seconds
Note: Typical at 25°C
Performance unaffected by cable length
Operational Environment
PEEK®
Sensor Head
Temperature Range: -5°C to 50°C
Maximum Pressure: 150 psi @ 50°C
Maximum Flow Rate: 10 ft/second
Operational Environment
PEEK®
Sensor Head
Temperature Range: -5°C to 50°C
Maximum Pressure: 150 psi @ 50°C
Maximum Flow Rate: 10 ft/second
Power Requirements
Voltage Range: 10 to 30 VDC
Maximum Power: 200 mW
Typical Power: 120 mW
Note: Class II DC power supply required
Power Requirements
Voltage Range: 10 to 30 VDC
Maximum Power: 200 mW
Typical Power: 120 mW
Note: Class II DC power supply required
Construction
Process Electrodes: Gold Cathode, Silver Anode,
Silver Reference
(3 Electrode Polarographic Clark Cell)
Membrane: FEP Teflon
O-Rings: Viton® (other materials available)
Sensor Head Material: PEEK®
Construction
Process Electrodes: Gold Cathode, Silver Anode,
Silver Reference
(3 Electrode Polarographic Clark Cell)
Guard: Platinum
Membrane: PFA Teflon
O-Rings: Viton® (other materials available)
Units of Measure
Measurement Units: ppm, mg/l, %
Temperature Units: °C, °F
Units of Measure
Measurement Units: ppm, mg/l, %
Temperature Units: °C, °F
Calibration
Air: Automatically adjusts for set pressure
Sample: 1 point
Zero: 1 point
Temperature: 1 and 2 point
Note: D.O. and temperature are precalibrated at the factory
Calibration
Air: Automatically adjusts for set pressure
Sample: 1 point
Zero: 1 point
Temperature: 1 and 2 point
Note: D.O. and temperature are precalibrated at the factory
Other Configuration Options
Sensor Filter: 0 to 100 seconds
Temperature Filter: 0 to 100 seconds
Other Configuration Options
Sensor Filter: 0 to 100 seconds
Temperature Filter: 0 to 100 seconds
Approvals and Ratings
Immunity & Emissions: CE (EN61326 and EN61326:1998)
Safety: CE (EN 61010-1) UL 508
Hazardous Locations: UL Class I, Division 2
Approvals and Ratings
Immunity & Emissions: CE (EN61326 and EN61326:1998)
Safety: CE (EN 61010-1) UL 508
Hazardous Locations: UL Class I, Division 2
3
4-30-2014 Rev 0.9
AdvantEDGE Optical Dissolved Oxygen
• Robust
• Cost Effective Options
• No Calibration Hassles
• Low Maintenance
• Accurate
The AdvantEDGE sensor uses the latest
technology to measure dissolved oxygen (DO)
in demanding environments. The AdvantEDGE
series™ Optical Dissolved Oxygen RDO® sensor
o­ffers several advantages for monitoring DO in
process applications:
Specifications
Sensor Type Transmitter/local display Range Accuracy (DO) Response time, cap Resolution Usage life of cap Shelf life of cap Operating temperature IP rating Compliance Storage conditions, cap Storage conditions, sensor
Salinity range pH range Barometric range Internal mounting thread Communications options Maximum cable length Warranty The AdvantEDGE sensor measures dissolved oxygen (DO) using the principle of
“dynamic luminescence quenching.” The sensor (Figure 1) uses lifetime-based
optical fluorescent
technology to provide
an extremely stable,
accurate, lowPhotodiode
maintenance DO sensor.
The sensor measures
the “phase” (or delay)
of the returned signal
for Sensor Diagnostics
compared to the
excitation signal, and
is thus based on the “lifetime”
rather than the “intensity” of the
luminescence. The presence of
oxygen in the foil quenches the
Figure 1: AdvantEDGE sensor design
luminescence and causes a phase
shift in the returned signal, detected
by the photodiode. The phase difference between the blue excitation light and the
return red light is measured, and the result is used to quantify DO.
Luminescent dissolved oxygen sensor
Optional, not required
0 to 20 mg/L concentration
0 to 200% saturation
±0.1 mg/L, 0 to 8 mg/L
±0.2 mg/L, 8 to 20 mg/L
T90: 30 sec T95: 37 sec @ 25°C
0.01 mg/L
1 year from the ­first instrument reading
2 years from date of manufacture
(install within 1 year from date of manufacture)
0° to 50° (32° to 122°F)
IP-67 with cap off, IP-68 with cap installed
Heavy industrial, IEC 61000-6-2:2005
1° to 60° (33° to 140°F), in factory container
-5° to 60° (23° to 140°F)
0 to 42 PSU, fi­xed or real-time capable
2-10 pH
507-1115 mbar, fi­xed or real-time capable
1¼-11½ NPT
Modbus RTU (RS485), 4-20mA, SDI-12
Up to 4000 ft (Modbus and 4-20 mA)
or up to 200 ft (SDI-12)
Sensor: 3 years from date of manufacture
Cap: 2 years from date of manufacture
4
4-30-2014 Rev 0.9
AdvantEDGE Contacting Conductivity
AdvantEDGE Toroidal Conductivity
• High resolution measurement
• Pre-calibrated (no field calibration
required)
• Plug & play sensor heads
• 0 to 2 S/cm range
• Direct data reporting
• Plug & play industrial communications
adapters
• High resolution measurement
• Pre-calibrated (no field calibration
required)
• Plug & play sensor heads
• 0 to 2 S/cm range
• Direct data reporting
• Plug & play industrial communications
adapters
Connect this Conductivity sensor to a PLC for seamless integration
with industrial control systems. Use any computer to display data,
calibrate and customize the measurement without an intermediate
analyzer electronics box. A LD500 local display can be used if
desired. Sensor heads are pre-calibrated and can be replaced or
exchanged with any other type of sensor without taking the system
down. Save space, time and money.
Connect this Toroidal Conductivity sensor to a PLC for seamless
integration with industrial control systems. Use any computer to
display data, calibrate and customize the measurement without an
intermediate analyzer electronics box. A LD500 local display can
be used if desired. Sensor heads are pre-calibrated and can be
replaced or exchanged with any other type of sensor without taking
the system down. Save space, time and money.
Measurement System Performance
Range: 0 to 2,000,000 uS/cm (2 S/cm)
Resolution: 4.5 significant digits
Accuracy: 0.1% of reading
Step Response Time: 90% in 30 seconds
Note: Typical at 25°C
Performance unaffected by cable length
Measurement System Performance
Range: 0 to 2,000,000 uS/cm (2 S/cm)
Resolution: 4.5 significant digits
Accuracy: 0.1% of reading
Step Response Time: 90% in 30 seconds
Note: Typical at 25°C
Performance unaffected by cable length
Operational Environment
CPVC Sensor Head
Temperature Range: -5°C to 50°C
Maximum Pressure: 85 psi @ 50°C
Maximum Flow Rate: 10 ft/second
Temperature Range:
Maximum Pressure:
Maximum Flow Rate:
PEEK®
CPVC
Sensor Head
Sensor Head
-5°C to 50°C
-5°C to 50°C
150 psig @ 50°C 85 psig @ 50°C
10 ft/second
10 ft/second
Power Requirements
Voltage Range: 10 to 30 VDC
Maximum Power: 200 mW
Typical Power: 120 mW
Note: Class II DC power supply required
Power Requirements
Voltage Range: 10 to 30 VDC
Maximum Power: 200 mW
Typical Power: 120 mW
Note: Class II DC power supply required
Construction
Cell Constants: 0.1 for Resistivity
1.0 and 5.0 for Conductivity
O-Rings: Viton® (other materials available)
Electrode Material: Titanium (Standard) or 316 Stainless Steel
Weight: 1.2 lbs
Construction
O-Rings: Viton® (other materials available)
Sensor Head Material: CPVC, Peek®
Weight: 1.2 lbs (Peek®, Ryton® or CPVC)
2.6 lbs (with 316 Stainless Steel interface module)
Units of Measure
Measurement Units: μS/cm, TDS, %, MΩ • cm
Temperature Units: °C, °F
Units of Measure
Measurement Units: μS/cm, TDS, % conc.
Temperature Units: °C, °F
Calibration
Zero: In dry air
Span: 1 point
Temperature: 1 and 2 point
Note: Conductivity and temperature are precalibrated at
the factory
Calibration
Zero: In dry air
Span: 1 point
Temperature: 1 and 2 point
Note: Conductivity and temperature are precalibrated at
the factory
Other Configuration Options
Sensor Filter: 0 to 100 seconds
Temperature Filter: 0 to 100 seconds
Concentration Table: User 10-point
Other Configuration Options
Sensor Filter: 0 to 100 seconds
Temperature Filter: 0 to 100 seconds
Concentration Table: User 10-point
Approvals and Ratings
Immunity & Emissions: CE (EN61326 and EN61326:1998)
Safety: CE (EN 61010-1) UL 508
Hazardous Locations: UL Class I, Division 2
Approvals and Ratings
Immunity & Emissions: CE (EN61326 and EN61326:1998)
Safety: CE (EN 61010-1) UL 508
Hazardous Locations: UL Class I, Division 2
5
4-30-2014 Rev 0.9
AdvantEDGE Suspended Solids
AdvantEDGE Turbidity
• 0 to 20,000 mg/L (diatomaceous earth)
• Fouling correction optics
• Versitile mounting
• Rugged construction
• Local and remote configuration and Connect this Suspended
diagnostics
Solids sensor to a PLC for • Temperature measurement included
seamless integration with • Direct data reporting
industrial control systems.
• Plug & play industrial communications
Use any computer to
adapters
display data, calibrate
• 0 to 4000 NTU measurement range
• Conforms to ISO 7027
• 90 degree and 180 degree light paths
• Rugged, foul resistant construction
• Local and Remote configuration and Connect this Raw
diagnostics
Water Turbidity sensor
• Temperature measurement included
to a PLC for seamless
• Direct data reporting
integration with industrial
• Plug & play industrial communications
control systems. Use any
computer to display data, adapters
calibrate and customize the measurement without an intermediate
analyzer electronics box. A LD500 local display can be used if desired.
Measurement System Performance:
Range: 0 TO 20,000 mg/l (diatomaceous earth)
Resolution: 1 mg/L
Step Response Time: 90% in 15 seconds
Note: Typical at 25 C; Performance unaffected by cable length.
Measurement System Performance:
Range: 0 TO 4000 NTU
Resolution: 0.1 NTU
Accuracy: 0.2 NTU
Step Response Time: 90% in 15 seconds
Note: Typical at 25 C; Performance unaffected by cable length.
and customize the measurement without an intermediate analyzer
electronics box. This versitile system is ideal for monitoring all solids
concentrations throughout a wastewater treatment plant. A LD500
local display can be used if desired.
Operational Environment:
Temperature Range: -5°C to 50°C
Maximum Pressure: 150 psig @ 50°C
Maximum Flow Rate: 10 ft./sec.
Operational Environment:
Temperature Range: -5 to 50 C
Maximum Pressure: 150 psi @ 50 C
Maximum Flow Rate: 10 ft./Sec.
Power Requirements:
Voltage Range: 10 to 30 VDC
Maximum Power: 200 mW
Typical Power: 120 mW
Note: Class II DC power supply required
Power Requirements:
Voltage Range: 10 to 30 VDC
Maximum Power: 200 mW
Typical Power: 120 mW
Note: Class II DC power supply required
Measurement Modes:
Measurement units: detected at 90 degrees and 180 degrees
Temperature: Automatic from -5° to 50° C
Measurement Modes:
Turbidity: 90 degree/ 180 degree, narrow
beam, IR light
Sediment Load: Blue to IR spectrum detected
at 90 degrees and 180 degrees
Temperature: Automatic from -5 to 50 C
Construction:
Windows: Quartz
Optical Topology: 1 cm path lengths for measurement and
reference. Detectors at 90 degrees and 180 degrees to source
O-Rings: Viton® (other materials available)
Sensor Head Material: Polycarbonate
Weight: 1.2 lbs
Construction:
Windows: Quartz or Sapphire
Optical Topology: Multi-Color Source
O-Rings: Viton® (other materials available)
Sensor Head Material: Polycarbonate
Weight: 1.2lbs
Units of Measure:
Measurement Units:
Suspended solids: mg/L, ppm, %
Temperature Units: °C, °F
Units of Measure:
Measurement Units:
Turbidity Mode: NTU Sediment mode: mg/l %
Temperature Units: C, F
Calibration:
Sample: 1 point
Zero: 1 point
Temperature: 1 point
Calibration:
Sample: 1 point
Zero: 1 point
Temperature: 1 and 2 point
Note: Turbidity and temperature are precalibrated at the factory
Other Configuration Options:
Sensor Filter: 0 to 100 seconds
Temperature Filter: 0 to 100 seconds
Other Configuration Options:
Sensor Filter: 0 to 100 seconds
Temperature Filter: 0 to 100 seconds
Approvals and Ratings:
Immunity and Emissions: CE Certified 89/336/EEC:CISPER
11, EN61000 (-4-2,-4-3,-4-4,-4-6, 4-8)
Safety: cULus Listed; 367G E303570
Hazardous Locations: Haz Loc Class 1, Division 2,Groups A,
B, C, D. Max Ambient 50°C
Approvals and Ratings:
Immunity and Emissions: CE Certified 89/336/EEC:CISPER
11, EN61000 (-4-2,-4-3,-4-4,-4-6, 4-8)
Safety: cULus Listed; 367G E303570
Hazardous Locations: Haz Loc Class 1, Division 2,Groups A,
B, C, D. Max Ambient 50 °C
6
4-30-2014 Rev 0.9
AdvantEDGE Free Chlorine Panel
• Free Chlorine
• Amperometric Measurement
• Temperature Compensated
• Optional pH Correction
• Pre-Calibrated
• Plug & Play
• Rugged Teflon Membrane
• Replaceable Membrane Cap
• Remote Measurement, Calibration,
Configuration, and Diagnostics.
This Free Chlorine sensor will measure
accurately in processes that are between
4 and 8 pH. When used with the LD500
Local Display unit and a pH measurement
system, hypochlorous acid (HOCL) and hypochlorite ion (OCL-)
concentration can be used to determine free chlorine present.
Measurement System Preformance:
Ranges: 0 to 5 ppm and 0 to 20 ppm
Resolution: 0.005 ppm
Accuracy: 2% or +/- 10 ppb (whichever is greater)
Step Response Time: 90% in 90 seconds
Operational Environment:
Temperature Range: 0°C to 45°C
(32°F to 113°F)
Maximum Pressure: 15 psig @ 45°C
System Flow in Chamber: 200 to 250ml/min
Power Requirements:
Voltage Range: 10 to 30 VDC
Maximum Power: 200 mW
Typical Power: 120 mW
Note: Class II DC power supply required
Construction:
Process Electrodes: Gold Cathode, Silver Anode.
Membrane: Teflon
O-Rings: Viton
Sensor Head Material: Noryl®
AdvantEDGE Material: CPVC or PEEK®
Weight: 1.2 lbs
Units of Measure:
Measurement Units: ppm, mg/l free chlorine
Temperature Units: °C, °F
Calibration:
Sample: 1 point
Zero: 1 point
Temperature: point
Note: Precalibrated at the factory
Other Configuration Options:
Sensor Filter: 0 to 100 seconds
Temperature Filter: 0 to 100 seconds
Approvals and Ratings:
Immunity and Emissions: CE Certified 89/336/
EEC:CISPER 11, EN61000 (-4-2,-4-3,-4-4,-4-6, 4-8)
Safety: cULus Listed; 367G E303570
Hazardous Locations: Haz Loc Class 1, Division 2,Groups
A, B, C, D. Max Ambient 50 °C
7
4-30-2014 Rev 0.9
AdvantEDGE LD500 Local Display
AdvantEDGE PA500 Profibus Gateway
Functions and Features
•
•
•
•
•
•
•
•
•
•
PA500 Profibus Gateway Adapter connects
up to 30 Modbus Communications Modules
to a Profibus DP (V0 or V1) network with
M12 connector. Full-featured measurement,
calibration, configuration and diagnostics.
24 VDC power.
Connects to any AdvantEDGE system.
Convenient 1/4 DIN enclosure
Display sensor and temperature data
Menu Functions: Calibration, Configuration and Diagnostics
Digital Protocol Options for remote measurement, calibrations,
configurations, and diagnostics.
One 4 to 20 mA loop output (Standard)
Two 4 to 20 mA loop outputs (Optional)
Two form C relays for alarm, control, and wash (Optional)
PID for pH
7 Conduit knockouts for easy cabling
Mounting Options
• Conduit Box, NEMA 4X (Standard)
• 1/2 DIN NEMA 4X Panel Mount (Optional)
Interface
• Display: 2 line by 16 character LCD
• Backlight: high contrast green
• Keypad: 4 directional arrows, MENU, ESCAPE, and ENTER
Enviromental Conditions
•
•
•
•
Ambient Operating Temperature Range: -20 to 65 C
Maximum Relative Humidity: 95% non-condensing
Meets CE requirements for heavy industrial use.
Nema 4X housing.
AdvantEDGE Sensor Connections
Supports 1 to 255 Modbus RTU AdvantEDGE sensors of any type.
• Supported Sensors: pH, ORP, Two Electrode Conductivity, Two electrode resistivity, Toroidal Conductivity, Dissolved Oxygen,
Ozone, Turbidity, and Free Chlorine.
4 to 20 mA Current Loop Options
•
•
•
•
•
One Isolated Current Loop (Standard)
Two Isolated Current Loops (Optional)
Assignable Parameters
PID function when pH parameter selected
Loop power provided by LD500
Relays/Optional
• Two 1 Amp form C relays with NO and NC contacts.
• Assignable Parameters
• Configured for Alarm, Control, & Timer Functions
Digital Communication Options to Host Computer
Network
•
•
•
•
•
•
•
•
MODBUS RTU
Device Net
Profibus DP/V1
RS-232 ASCII
RS-232 DF1
RS485 ACSII
CANOpen
Others: Consult Factory
Power Supply
Input:
10 to 30 VDC; 100 to 240 VAC optional
Power: 2W at 25 C
8
4-30-2014 Rev 0.9
Mounting Options
1.5 Inch Union Mounting
1 Inch Tee Mounting
1.5 Inch Ball Mounting
Material
• CPVC (Low)
• 316 Stainless Steel (Low)
• CPVC (High)
• 316 Stainless Steel (High)
Material
• CPVC
• 316 Stainless Steel
Material
• CPVC
• 316 Stainless Steel
Sanity Mount
Handrail Mounting Assembly
Material
• 316 SS (closed)
• 216 SS (open)
1 Inch Immersion
Mounting with
Junction Box
Material
• Immersion
• PVC
Material: PVC
Accessories & Replacements Parts
pH
Dissolved Ozone
Salt Bridge Replacements
SBS01: PEEK® Protected
SBS02: PEEK® Flat
SBS03: CPVC Protected
SBS04: CPVC Flat
SBC01: Storage cap
with sponge
pH Solutions
RCS03: storage solution, 60 ml
PHSOL4: Buffer, 500 ml
PHSOL7: Buffer, 500 ml
PHSOL10: Buffer, 500 ml
RCSO1-500: Reference Cell
Solution 500 ml
ORP
Salt Bridge Replacements
SBS01: PEEK® Protected
SBS02: PEEK® Flat
SBS03: CPVC Protected
SBS04: CPVC Flat
SBC01: Storage cap
with sponge
SBC01: Storage cap
with sponge
Ozone Solutions
500 ml bottles
RDOK2: Ozone Electrolyte
ZDOKO: Zero Powder
Wash Adapters
MH1142: air purge, CPVC
MH1132: water purge, CPVC
Turbidity
ORP Solutions
RCS04: ORP storage solution, 60 ml
ORPSOL200: 200mV Calibration,
500 ml
ORPSOL600: 600 mv calibration,
500 ml
RCS01: Reference cell solution,
500 ml
Dissolved Oxygen
D.O. Membrane
Replacements
DMR01: Ryton Protected
DMR02: Ryton Flat
DMR03: PEEK Protected
DMR04: PEEK Flat
Ozone Membrane
Replacements
DMR05: Ryton Protected
DMR06: Ryton Flat
DMR07: PEEK Protected
DMR08: PEEK Flat
Wash Adapters
MH1142: air purge, CPVC
MH1132: water purge, CPVC
SBC01: Storage cap
with sponge
9
Turbidity Solutions
500 ml bottles
FOR4K: 4000 NTU Formazin Stock
500 ml
FROO4O: 40 NTU Cal Kit
Wash Adapters
MH1142: air purge, CPVC
MH1132: water purge, CPVC
Conductivity
Conductivity Solutions
500 ml bottles
SOL1000: 1000 μS/cm Calibration Solution
SOL2000: 2000 μS/cm Calibration Solution
SOL5000: 5000 μS/cm Calibration Solution
SOL2C: Conductivity Electrode Cleaner
4-30-2014 Rev 0.9
HF scientific 3170 Metro Parkway, Ft. Myers FL 33916 • ph: (239)-337-2116 • fax: (239) 332-7643 • Toll Free: 888-203-7248 4-30-2014 Rev 0.9