List of commands - Metrohm Autolab

advertisement
Insert new module for the Autolab 8-Series
INSERT NEW MODULE FOR THE
AUTOLAB 8-SERIES
Editorial
Metrohm Autolab B.V.
Kanaalweg 29 G, 3526 KM Utrecht
The Netherlands
Tel: +31 30 2893154
Fax: +31 30 2880715
E-mail: autolab@metrohm-autolab.com
Website: www.metrohm-autolab.com
© Copyright 2011 Metrohm Autolab
1|Page
Insert new module for the Autolab 8-Series
Table of Contents
1 – Introduction ............................................................................................................ 3
2 – Overview table loose modules ................................................................................ 4
3 – Opening the Autolab cabinet .................................................................................. 5
4 – How to insert the new module(s) ............................................................................ 8
5 – Analog bus overview ............................................................................................ 10
6 – Analog bus adjustments for the FRA2 or FRA32M module .................................... 10
6.1 – Connections for the FRA2 module ................................................................ 10
6.2 – Connections for the FRA32M module ........................................................... 11
7 – Analog bus adjustments for the ECD module ........................................................ 12
8 – Analog bus adjustments for the ADC10M module ................................................ 13
9 – Trigger cable for SCAN250/ADC10M combination ................................................ 14
10 – Analog bus adjustments for the BA module ........................................................ 16
10.1 – If the FRA2 module is present ..................................................................... 16
10.2 – If the FRA32M module is present ................................................................ 17
10.3 – If the FRA2 or FRA32M module is not present ............................................ 18
11 – Connections to the ECN module ......................................................................... 18
12 – Connections to the pX1000 module ................................................................... 19
13 – Calibration of the FRA32M and FRA2 module with Nova .................................... 20
13.1 – Determination of the C1 and C2 parameter in Nova ................................... 22
13.1.1 – Determination of C1 ....................................................................... 22
13.1.2 – Determination of C2 ....................................................................... 26
13.2 – No calibration possible................................................................................ 30
13.3 – Verification of the calibration of the FRA module ........................................ 31
14 – Module test ........................................................................................................ 31
2|Page
Insert new module for the Autolab 8-Series
1 – Introduction
The Autolab instrument has 3 different casings, which can be recognized as follows:
1. Old types with serial numbers starting with AUTyynnnn where yy stands for the year
of production. These were instruments produced before 1999 (s/n ≤ AUT990789).
2. Autolab cabinets with the blue or green corner pieces with serial number starting with
AUT7, produced from 1999 (s/n ≥ AUT70790).
3. Autolab cabinets with serial number starting with AUT8, produced since May 2007
(s/n ≥ AUT83100).
This document pertains to instruments with serial number beginning with AUT8 (point 3
above).
3|Page
Insert new module for the Autolab 8-Series
2 – Overview table loose modules
Module
BA
FRA2
FRA32M
pX1000
EQCM
ECD
MUX
SCAN250
ADC10M
ECN
FI20
Items
Quantity
Module label
SMB-Coax cable
SMB-Coax cable copper
SMB-Coax –Banana cable
Alligator clip
Warning on PGSTAT100N
Module label
SMB-Coax cable
CD with calibration file
1
3
1
1
1
1
2
3
1
Module label
SMB-Coax cable
Module label
Cover for front connector
Lemo to banana cable
Analog bus connector
Warning on PGSTAT100N
1
3
1
1
2
1
1
Module label
Oscillator box
DB9 Oscillator connection cable
Crystal 6 MHz Au/TiO2 polished
EQCM cell
Au counter electrode
Reference electrode Ag/AgCl, 3M KCl
Driving force adjustment tool
Spare O-ring Reference electrode
Spare O-ring Counter electrode
Spare O-ring Viton Crystal
Spare O-ring Cell cover
2.5 mm hex driver
Spare EQCM cell screw
Warning on PGSTAT100N
Module label
SMB-Coax cable
1
1
1
2
1
1
1
1
1
1
4
1
1
3
1
1
1
Module label
Digital MUX cable
Warning on PGSTAT100N
Module label
SCAN250-ADC trigger cable old
SCAN250-ADC trigger cable new
1
1
1
1
1
1
Module label
ADC10M-SCAN trigger cable old
ADC10M-SCAN trigger cable new
Ferrite bead
Module label
Analog bus connector
Special ECN cable
Warning on PGSTAT100N
1
1
1
1
1
1
1
1
Module label
1
Required software
PGSTAT100N
compatible?
GPES 4.9.7 or higher
Nova 1.2 or higher
YES, with adjusted DIFFAMP
FRA 2.4 or higher
Nova 1.2 of higher
YES
Not supported in FRA
Nova 1.8 or higher
YES
Not supported in GPES
Nova 1.4 or higher
NO
Not supported in GPES
Nova 1.4 or higher
YES
GPES 4.5 or higher
Nova 1.3 or higher
YES
GPES 4.7 or higher
FRA 4.7 or higher
Nova 1.5 or higher
NO
GPES 4.9.7 or higher
Nova 1.2 or higher
YES
Not supported in GPES
Nova 1.2 or higher
YES
GPES 4.8 or higher
FRA 4.9.5 or higher
Nova 1.5 or higher
NO
GPES 4.5 or higher
Nova 1.5 or higher
YES
Table 1 – Module compatibility table
Table 1 provides the overview for loose modules available for 8-Series Autolab
instruments, the part list for each module and the software and hardware requirements
or restrictions.
4|Page
Insert new module for the Autolab 8-Series
3 – Opening the Autolab cabinet
The 8-series instruments are fitted in the N form factor cabinet shown in Figure 1.
Figure 1 - The 8-series Autolab instruments
Four instruments fit in this group:
•
•
•
•
PGSTAT128N
PGSTAT302N
PGSTAT302F
PGSTAT100N
Follow these steps to open the instrument cabinet:
1. Switch off the instrument and unplug from the mains
2. Unscrew the four screws on the rear of the Autolab (indicated in Figure 2). Leave
the screws in the top cover.
Figure 2 – Locate the four indicated screws on the back plane of the cabinet
3. Remove the top cover by sliding it from the front to the rear. The best way to do
this is to stand in front of the instrument, put your both hands at the rear part of
the top cover and press with your fingers on the rear panel, while pressing on the
top cover (see Figure 3). The top cover should come off easily.
5|Page
Insert new module for the Autolab 8-Series
Figure 3 – Open the cabinet by sliding the cover of the cabinet
4. Lift the top cover and remove it from the instrument and disconnect the earth
cable (see Figure 4).
Figure 4 – Lift the cover
Now the analog bus can be reached from the top. If you do not have to install a module,
it is unnecessary to remove the front panel of the instrument. If a new module needs to
be installed in the instrument, it is necessary to remove the front panel.
Follow these steps to remove the front panel:
1. Unscrew the four screws, indicated in Figure 5, holding the front panel to the
frame from the mounting tabs.
6|Page
Insert new module for the Autolab 8-Series
Figure 5 – Locate the four indicated screws holding the front panel
2. At the bottom of the front panel the same construction is used. The screws can be
reached through holes in the bottom cover. These screws must be removed
without turning the Autolab upside down. In order to do this, slide the
instrument slightly over the edge of the table and remove the four screws
indicated in Figure 6 from below.
Figure 6 – Remove the four indicated screws from below the instrument
3. Now the front panel can be shifted smoothly to the front and taken off. Take care
not to damage the LCD display (see Figure 7).
7|Page
Insert new module for the Autolab 8-Series
Figure 7 – Remove the front panel
4 – How to insert the new module(s)
The optional modules for the 8-Series Autolab instrument must be installed in the predefined positions indicated in Figure 8.
Figure 8 – Each module must be installed in a pre-defined position in the instrument
Follow these steps to install a module:
1. Locate the module slot for the installed module (see Figure 8).
2. Remove the black front panel and insert the module carefully in the slot. Insert the
module carefully and do not force the module in the slot.
8|Page
Insert new module for the Autolab 8-Series
3. Connect the inserted module to the analog bus, using the pre-installed green
connectors 1. Depending on the module, additional steps may be necessary:
a. For the ECD module additional connections are required (please refer to
Section 7).
b. For the FRA2 or FRA32M module additional connections are required
(please refer to Section 6.1 and 6.2). Furthermore, the FRA2 or FRA32M
module needs to be calibrated (C1 and C2 determination, please refer to
Section 13).
c. For the BA module additional connections are required (please refer to
Section 10).
d. For the ECN or pX1000 a dedicated analog bus extension, supplied with
the module, must be installed (please refer to Section 11 for the ECN and
to Section 12 for the pX1000 module). The combination of the ECN and
the pX1000 is not possible.
e. For the ADC10M a modification of the installed analog bus may be
required (see Section 8).
f. For the SCAN250 and the ADC10M a triggering cable must be installed if
both modules are installed in the same instrument (see Section 9).
4. When the module is installed properly, tighten the front plate of the module to
the frame of the instrument with the provided screws.
5. Reattach the front panel of the Autolab and close the instrument by following the
steps detailed in Section 3 in reverse order.
6. Replace the blank module label with the label supplied with the module.
Note: the leftmost slot of the Autolab is shared by several modules (FI20, MUX, ECN and
pX1000). When the FI20 module is installed together with one or more modules assigned
to the same slot, the priority should be given to the FI20 module and the other modules
should be installed in any other free slot.
1
This is not required for the MUX and the EQCM.
9|Page
Insert new module for the Autolab 8-Series
5 – Analog bus overview
Figure 9 shows a complete overview of the analog bus of the Autolab.
Figure 9 – Detailed overview of the default analog bus wiring diagram
6 – Analog bus adjustments for the FRA2 or FRA32M module
The FRA2 and FRA32M modules are supplied with three additional SMB cables. These
cables must be placed between the module and the analog board of the PGSTAT. The
connections are specified in Section 6.1 for the FRA2 module and in Section 6.2 for the
FRA32M. After installation, FRA2 or FRA32M module needs to be calibrated (C1 and C2
determination, please refer to Section 13).
6.1 – Connections for the FRA2 module
Place the SMB cables between the FRA2 module and the PGSTAT analog board, as shown
in Figure 10.
1. Connect DSG signal of the FRA2 to DSG of the PGSTAT.
2. Connect Ein of the FRA2 to FRA E of the PGSTAT.
3. Connect Iin of the FRA2 to FRA I of the PGSTAT.
10 | P a g e
Insert new module for the Autolab 8-Series
Figure 10 – Overview of the connections between the FRA2 module and the PGSTAT
6.2 – Connections for the FRA32M module
Place the SMB cables between the FRA32M module and the PGSTAT analog board, as
shown in Figure 11.
1. Connect PGSTAT DSG of the FRA32M to DSG of the PGSTAT.
2. Connect PGSTAT U of the FRA32M to FRA E of the PGSTAT.
3. Connect PGSTAT I of the FRA32M to FRA I of the PGSTAT.
11 | P a g e
Insert new module for the Autolab 8-Series
Figure 11 – Overview of the connections between the FRA32M and the PGSTAT
7 – Analog bus adjustments for the ECD module
An additional SMB cable is provided with the ECD module. Locate CONN3 on the ECD
module and CONN4 on the control amplifier of the Autolab. Place the SMB cable
between these two connectors (see Figure 12).
12 | P a g e
Insert new module for the Autolab 8-Series
Figure 12 – An additional SMB cable must be placed between the ECD and the control amplifier
The ECD installation can be tested using Nova (version 1.3 or higher), as explained in
Section 14 of this document.
The ECD module installed can also be tested with the GPES software following the
instructions provided in the Autolab 4.9 Installation and Diagnostics manual.
8 – Analog bus adjustments for the ADC10M module
For the ADC10M module, a ferrite bead should be placed in the Analog bus (see Figure
13). For instruments with serial numbers ≥ AUT83340 the ferrite bead is already installed
and this modification is not required.
13 | P a g e
Insert new module for the Autolab 8-Series
Figure 13 – A ferrite bead must be installed on the analog bus connection to the ADC10M module
Follow these steps to install the ferrite bead, supplied with the module:
1. Locate the ADC10M wire 3 (see Figure 13).
2. Unscrew the wire at the ADC10M end of the Analog bus and slide the wire
through the ferrite bead. Reattach the wire end to the module connector.
3. Repeat step 2 for the two other wires.
The ADC10M installation can be tested using Nova (version 1.2 or higher), as explained in
Section 14 of this document.
Note: the ADC10M is NOT supported in the GPES software.
9 – Trigger cable for SCAN250/ADC10M combination
When the ADC10M and the SCAN250 module are installed in the same instrument, a
dedicated trigger cable must be installed (see Figure 14). This cable is supplied with the
modules.
14 | P a g e
Insert new module for the Autolab 8-Series
Figure 14 – The trigger cable must be installed between the SCAN250 and the ADC10M
Locate CONN5 on the ADC10M and CONN4 on the SCAN250 and connect both with the
supplied trigger cable 2.
The ADC10M/SCAN250 combination can be tested using Nova (version 1.2 or higher), as
explained in Section 14 of this document.
Note: the ADC10M/SCAN250 combination is NOT supported in the GPES software.
2
Other combinations are possible: ADC10M/SCANGEN or ADC750/SCAN250.
15 | P a g e
Insert new module for the Autolab 8-Series
10 – Analog bus adjustments for the BA module
For the BA module the SMB coaxial cables should be connected between the BA module
and the PGSTAT and the FRA2 or FRA32M, if applicable. Note: The supplied jumper cable
is NOT used for the BA module in an 8-Series instrument!
10.1 – If the FRA2 module is present
For the BA module the SMB coaxial cables should be connected as shown in Figure 15.
Figure 15 – Overview of the SMB cables required between the BA module and the PGSTAT and
optional FRA2 module
4. Disconnect the three SMB cables between the FRA2 and the PGSTAT and
reconnect these cable to the BA module:
a. Connect DSG signal of the FRA2 to FRA DSG of the BA.
b. Connect Ein of the FRA2 to E-FRA of the BA.
c. Connect Iin of the FRA2 to I-FRA of the BA.
5. Connect the special copper labelled SMB cable between the WEx of the BA
module and the BIPOT WE2 of the PGSTAT. Please note that the location of BIPOT
WE2 on the PGSTAT differs slightly depending on the model (see Figure 15).
6. Connect the three additional SMB cables:
a. Connect To PGstat of the BA to DSG of the PGSTAT.
b. Connect E-PGSTAT of the BA to FRA E of the PGSTAT.
16 | P a g e
Insert new module for the Autolab 8-Series
c. Connect I-PGSTAT of the BA to FRA I of the PGSTAT.
7. Connect the SMB to 4 mm banana cable, labelled WE2, to the dedicated plug on
the Autolab differential amplifier.
10.2 – If the FRA32M module is present
For the BA module the SMB coaxial cables should be connected as shown in Figure 16.
Figure 16 – Overview of the SMB cables required between the BA module and the PGSTAT and
optional FRA32M module
8. Disconnect the three SMB cables between the FRA32M and the PGSTAT and
reconnect these cable to the BA module:
a. Connect PGSTAT U of the FRA32M to E-FRA of the BA.
b. Connect PGSTAT I of the FRA32M to I-FRA of the BA.
c. Connect PGSTAT DSG of the FRA32M to FRA-DSG of the BA.
9. Connect the special copper labelled SMB cable between the WEx of the BA
module and the BIPOT WE2 of the PGSTAT. Please note that the location of BIPOT
WE2 on the PGSTAT differs slightly depending on the model (see Figure 16).
10. Connect the three additional SMB cables:
a. Connect To PGstat of the BA to DSG of the PGSTAT.
b. Connect E-PGSTAT of the BA to FRA E of the PGSTAT.
c. Connect I-PGSTAT of the BA to FRA I of the PGSTAT.
17 | P a g e
Insert new module for the Autolab 8-Series
11. Connect the SMB to 4 mm banana cable, labelled WE2, to the dedicated plug on
the Autolab differential amplifier.
10.3 – If the FRA2 or FRA32M module is not present
Follow the same steps detailed above, starting at point 2 (skip point 1).
Note: the BA module acts as a feed through for the cables from the FRA2 module.
The BA installation can be tested using Nova (version 1.2 or higher), as explained in
Section 14 of this document.
The BA module installed can also be tested with the GPES software following the
instructions provided in the Autolab 4.9 Installation and Diagnostics manual.
11 – Connections to the ECN module
The ECN module must be connected to the ADC164 module using the provided extra
analog bus connector as shown in Figure 17.
Figure 17 – Overview of the connections of the ECN module to the ADC164
The following connections are necessary for the ECN module:
•
•
Pin 1 of the ECN connector must be connected to one of the 3 pins of the ground
connector of the ADC164.
Pin 2 must be connected to ADC channel 8.
18 | P a g e
Insert new module for the Autolab 8-Series
The ECN installation can be tested using Nova (version 1.5 or higher), as explained in
Section 14 of this document.
The ECN module installed can also be tested with the GPES software following the
instructions provided in the Autolab 4.9 Installation and Diagnostics manual.
Important: for GPES 4.9, a modification of the SYSDEF40.INP file is required. Open the
file (located in the C:\Autolab folder) with Notepad and verify that line 17 of the file has
the following format:
17: 7,8,,,
Adjust the SYSDEF40.INP file if necessary and save the file.
Important: the combination of the ECN with the pX1000 is not possible.
12 – Connections to the pX1000 module
The pX1000 module must be connected to the ADC164 module using the provided extra
analog bus connector as shown in Figure 18.
Figure 18 – Overview of the connections of the pX1000 module to the ADC164
The following connections are necessary:
•
•
Pin 1 of the pX1000 connector must be connected to one of the 3 pins of the
ground connector of the ADC164.
Pin 2 must be connected to ADC channel 7.
19 | P a g e
Insert new module for the Autolab 8-Series
•
Pin 3 must be connected to ADC channel 8.
The pX1000 installation can be tested using Nova (version 1.4 or higher), as explained in
Section 14 of this document.
Note: the pX1000 is NOT supported in the GPES software.
Important: the combination of the pX1000 with the ECN is not possible.
13 – Calibration of the FRA32M and FRA2 module with Nova
The procedure described below is explained for the Nova software (version 1.8 or higher).
Both the FRA32M and the FRA2 module need to be adjusted to the Autolab
potentiostat/galvanostat.
Note: the FRA2 module is supplied with a specific calibration file (FRA2CAL.INI) which
contains the calibration data for the module itself. For the FRA32M, the calibration data is
stored on the on-board memory of the module.
In the rest of this section, both the FRA32M and the FRA2 modules will be referred to as
FRA module, except when specifically mentioned as FRA32M or FRA2.
Follow the steps described in this section to calibrate the FRA module after installation in
the Autolab:
1. Install the Nova software.
2. The FRA2CAL.INI calibration file is delivered with the new FRA2-module. The
Autolab number must be changed to the serial number of the connected Autolab
(i.e. AUT83185, see rear side of Autolab). Open the FRA2CAL.INI with Notepad
and locate the [System] header in the file (see Figure 19). Change the
highlighted AUT70000 number to the serial number of the instrument. Save the
FRA2CAL.INI file and exit Notepad.
Figure 19 – Adjust the serial number of the instrument in the FRA2CAL.INI file
3. Connect the instrument and switch it on.
4. Open the Hardware setup of Nova (Tools – Hardware setup). Adjust the hardware
setup if necessary. For the FRA2 module, click the
button to add the
20 | P a g e
Insert new module for the Autolab 8-Series
FRA2CAL.INI file to the hardware setup (see Figure 20). When prompted, select
the type of instrument in the C1/C2 calibration factors dialog window, as show in
Figure 20.
Figure 20 – Importing the FRA2CAL.INI file in NOVA
5. Select the instrument type in the Main Module frame in the hardware setup
window and verify that the values of C1 and C2 values in the frame on the righthand side are set to 0.0 (see Figure 21). Adjust these values, if necessary.
21 | P a g e
Insert new module for the Autolab 8-Series
Figure 21 – Verify that the value of C1 and C2 are set to 0.0 before starting the calibration
13.1 – Determination of the C1 and C2 parameter in Nova
The value of the C1 and C2 parameters must be measured and modified in the
FRA2CAL.INI file (for the FRA2 module only). The determination of C1 and C2 requires
the following items:
•
•
Autolab Dummy cell
Faraday cage
13.1.1 – Determination of C1
Follow the steps described in this section to determine the value of the C1 parameter.
1. Start the Nova software.
2. Be sure C1 and C2 are set initially to 0.0 (see previous section and Figure 21).
3. Import the procedure PGSTAT C1 calibration.nox from the NOVA installation
folder 3 (File – Import procedure).
3
The default installation folder is C:\Program Files\Metrohm Autolab\SharedDatabases\Module test.
22 | P a g e
Insert new module for the Autolab 8-Series
4. Connect the Autolab Dummy cell as shown in Figure 22. Connect the ground lead
from the PGSTAT to the Faraday cage. Do not connect the ground lead from the
PGSTAT to the Dummy cell. Place the dummy cell in the Faraday cage.
Figure 22 – Overview of the connections required for the determination of C1
5. Start the measurement and wait until it has been finished. Ignore the warning
displayed during the procedure validation (see Figure 23).
Figure 23 – Ignore the warning shown during the validation of the procedure
6. A reminder message is shown at the beginning of the measurement (see Figure
24).
23 | P a g e
Insert new module for the Autolab 8-Series
Figure 24 – A reminder message is shown at the beginning of the measurement
7. During the measurement, the measured data will be plotted as a Bode plot similar
to the example shown in Figure 25.
Figure 25 – Typical Bode plot obtained during the C1 calibration
8. The data is automatically fitted and the results of the fitting are reported in a
Message box at the end of the measurement (see Figure 26).
24 | P a g e
Insert new module for the Autolab 8-Series
Figure 26 – The experimentally determined value of C1 is reported in a Message box at the end of
the measurement
9. Open the Hardware setup of Nova (Tools – Hardware setup). Select the instrument
type in the Main Module frame in the hardware setup window and adjust the
value of C1 to the value reported in the Message box 4 (see Figure 27).
Figure 27 – Change the value of C1 to the value reported in the Message box
4
It is possible to copy and paste the contents of the Message box.
25 | P a g e
Insert new module for the Autolab 8-Series
10. Click OK to save the changes and wait for the Autolab to be reinitialized using the
updated Hardware setup.
11. For the FRA2 module, open the source FRA2CAL.INI with Notepad and locate the
[CFCalib] header in the file (see Figure 28). Change the C1=0 value to the
value reported in the Message box, in pF (1.61E-11 F corresponds to 16.1 pF).
Save the changes to the FRA2CAL.INI file.
Figure 28 – Adjust the C1 value in the FRA2CAL.INI file
13.1.2 – Determination of C2
Follow the steps described in this section to determine the value of the C1 parameter.
1. Import the procedure PGSTAT C2 calibration.nox from the NOVA installation
folder 5 (File – Import procedure).
2. Disconnect the Dummy cell and leave the leads open in the Faraday cage. CE and
RE must be connected together as well as WE and S (as shown in Figure 29). Make
sure RE/CE and WE/S are not connected together. Connect the ground lead from
the PGSTAT to the Faraday cage.
Figure 29 – Overview of the connection required for the determination of C2
5
The default installation folder is C:\Program Files\Metrohm Autolab\SharedDatabases\Module test.
26 | P a g e
Insert new module for the Autolab 8-Series
3. Start the measurement and wait until it has been finished. Ignore the warning
displayed during the procedure validation (see Figure 30).
Figure 30 – Ignore the warning shown during the validation of the procedure
4. A reminder message is shown at the beginning of the measurement (see Figure
31).
Figure 31 – A reminder message is shown at the beginning of the measurement
5. During the measurement, the measured data will be plotted as a Bode plot similar
to the example shown in Figure 32.
27 | P a g e
Insert new module for the Autolab 8-Series
Figure 32 – Typical Bode plot obtained during the C2 calibration
6. The data is automatically fitted and the results of the fitting are reported in a
Message box at the end of the measurement (see Figure 33).
Figure 33 – The experimentally determined value of C2 is reported in a Message box at the end of
the measurement
7. Open the Hardware setup of Nova (Tools – Hardware setup). Select the instrument
type in the Main Module frame in the hardware setup window and adjust the
value of C2 to the value reported in the Message box 6 (see Figure 34).
6
It is possible to copy and paste the contents of the Message box.
28 | P a g e
Insert new module for the Autolab 8-Series
Figure 34 – Change the value of C2 to the value reported in the Message box
8. Click OK to save the changes and wait for the Autolab to be reinitialized using the
updated Hardware setup.
9. For the FRA2 module, open the source FRA2CAL.INI with Notepad and locate the
[CFCalib] header in the file (see Figure 35). Change the C2=0 value to the
value reported in the Message box, in pF, rounded to the first decimal place
(5.43E-11 F corresponds to 0.5 pF). Save the changes to the FRA2CAL.INI file.
Figure 35 – Adjust the C2 value in the FRA2CAL.INI file
29 | P a g e
Insert new module for the Autolab 8-Series
Important: always send the final FRA2CAL.INI (for the FRA2 module) or the determined
C1 and C2 values (for the FRA32M module), together with the serial number of the
instrument to Metrohm Autolab by email (service@metrohm-autolab.com). This is
important for future service. If not sent back, Metrohm Autolab will not be able to service
this instrument properly.
13.2 – No calibration possible
If you do not have a Faraday cage or if you are not able to calibrate, you can fill in the
typical values for C1 and C2 given in Table 2. This will however affect the accuracy of the
measurement.
Instrument type C1
C2
PGSTAT302N
PGSTAT302F
1.6E-11
1.6E-11
3.0E-13
1.0E-12
PGSTAT128N
2.6E-11
1.0E-12
PGSTAT128N
1.6E-11
1.0E-12
PGSTAT100N
1.6E-11
5.0E-13
For instruments with serial number >
AUT84179
Table 2 – Typical values for C1 and C2
To manually specify the value of C1 and C2, open the hardware setup in Nova (Tools –
Hardware setup). Select the instrument type in the Main module frame in the hardware
setup window and manually enter the C1 and C2 values in the frame on the right-hand
side (see Figure 36).
30 | P a g e
Insert new module for the Autolab 8-Series
Figure 36 – Editing the values of C1 and C2 in Nova
13.3 – Verification of the calibration of the FRA module
A dedicated test for the FRA module is provided in NOVA. Please refer to section 14 for
more information on the testing of modules in NOVA.
The calibrated module must pass the described test. If the values are out of range,
something is wrong with the installation or with the calibration. Please check the
complete installation procedure. If re-calibration is necessary, please remember to set C1
and C2 values back to zero in the Hardware setup of Nova.
14 – Module test
After installation of the module, the following tests should be taken to verify that the
module is working as expected:
1. Connect mains power.
2. Switch Autolab on.
3. Install the delivered software if necessary (please refer to Table 1 for the software
version requirements for each module).
4. Select Hardware setup and activate inserted module.
31 | P a g e
Insert new module for the Autolab 8-Series
5. Refer to the module tests described in the Nova Getting Started manual or in the
Module test in Nova, available from the Help menu (see Figure 37).
Figure 37 – The Module test with Nova document described how to test the modules in Nova
32 | P a g e
Download