Applied Radiation and Isotopes ] (]]]]) ]]]–]]] Contents lists available at SciVerse ScienceDirect Applied Radiation and Isotopes journal homepage: www.elsevier.com/locate/apradiso Investigations on TDCR measurements with the HIDEX 300 SL using a free parameter model Carsten Wanke, Karsten Kossert n, Ole J. Nähle Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany a r t i c l e i n f o Keywords: TDCR method Liquid scintillation counting Activity standardization abstract The Triple to Double Coincidence Ratio (TDCR) method requires a special counting system with three photodetectors. The systems currently used by National Metrology Institutes for activity standardizations are custom built, and up to now the HIDEX 300 SL counter1,2 is the only TDCR counter commercially available. At PTB, measurements with a special metrology version of this counter were carried out to investigate its applicability for activity standardizations. The activity results of measurements with the HIDEX counter are compared to those obtained with a PTB–TDCR counter, as such a comparison reduces the model dependence. In addition, a spectrometry method was applied to measure 109Cd samples and a new TDCR-Čerenkov method was tested with 32P samples. & 2012 Elsevier Ltd. All rights reserved. 1. Introduction Liquid scintillation (LS) counting in combination with the TDCR method is a powerful method for activity standardization and can be applied to a number of radionuclides (Broda et al., 2007). The method is based on a free parameter model describing the process of light emission and detection in a scintillation counter. For the application of the method, an LS counter with three photomultiplier tubes (PMT) A, B and C, and suitable coincidence electronics are required. For a large number of detected events, the ratio of the triple and the double coincidence counting rates T/D converges towards the ratio of calculated counting efficiencies eT(l)/eD(l) which depend on the free parameter l. Thus, the free parameter can be derived from experimental information and the counting efficiencies can be calculated. It is to be noted that the TDCR value is neither a quenching indicator nor an efficiency value. The latter assumption is implicitly made in the Hidex control software output, where the result ‘‘disintegrations per minute’’ (DPM) is given assuming that the TDCR value is equal to the double efficiency. This value can be wrong by more than 50% for nuclides like 3H or 55Fe (Cassette, 2010). Taking the TDCR value as a quenching indicator is only applicable if the relation between the TDCR value and the efficiency is unambiguous. This is fulfilled for n Corresponding author. Tel.: þ49 531 592 6110; fax: þ 49 531 592 6305. E-mail address: karsten.kossert@ptb.de (K. Kossert). 1 Manufactured by HIDEX Oy, Finland. 2 Certain commercial equipment, instruments, or materials are identified in this paper to foster understanding. Such identification does not imply recommendation by the Physikalisch-Technische Bundesanstalt, nor does it imply that the materials or equipment identified are necessarily the best available for the purpose. pure beta emitters but not in the case of radionuclides decaying by electron capture (EC) or radionuclides with complex decay schemes, where up to three efficiency values can correspond to the same TDCR value (Cassette, 2010). Up to now, all TDCR counters used by national metrology institutes for activity determinations have been custom built. The HIDEX 300 SL system is the first commercial TDCR counter and is equipped with an automatic sample changer which makes measurements of a large number of samples very convenient. The development and construction of a home-made TDCR counter can be rather challenging, in particular, when an automated samples changer is needed. Thus, it is of great interest to investigate the applicability of the Hidex counter for activity measurements. The first outcome of such an investigation is described in this article. 2. Experimental details and procedures The HIDEX 300 SL installed at PTB is equipped with three Electron Tubes 9102KA photomultipliers. The mounting positions of the multipliers are denoted as R, S and T. The system has a lead shielding and an automatic sample changer. The samples are provided via an aluminum rack with 40 positions for 20 mL-vials; a rack with 96 positions for 7 mL-mini vials can also be installed. The system is computer controlled via serial connection. For all measurements presented in this report, the instrument has been operated using the software ‘‘Commfiler.csv’’ supplied by HIDEX. The system installed at PTB is a modified version of this counter referred to as the metrology version. It offers a special counting mode, denoted as the ‘‘METRO’’ mode, which delivers all 0969-8043/$ - see front matter & 2012 Elsevier Ltd. All rights reserved. doi:10.1016/j.apradiso.2012.02.097 Please cite this article as: Wanke, C., et al., Investigations on TDCR measurements with the HIDEX 300 SL using a free parameter model. Appl. Radiat. Isotopes (2012), doi:10.1016/j.apradiso.2012.02.097 2 C. Wanke et al. / Applied Radiation and Isotopes ] (]]]]) ]]]–]]] information necessary for applying the TDCR method in a metrological sense. This means, in particular, that all events (including single, double (3 pairs) and triple detections, as well as the logical sums of double coincidences and single counts) are recorded. Unless otherwise stated, the measurements presented were performed in the ‘‘METRO’’ mode. The exact mechanism of signal processing is not known to the authors. The system underwent several changes in electronics, until a final setup of the system was reached. Linearity measurements with the first setup showed the need for a better dead-time/live-time determination. Hence, new electronics were designed by Hidex and subsequently improved according to the analysis of the system at PTB. The newer designs include a dead-time determination circuit based on a reference oscillator with a compensation circuit, which can be fine-tuned via a potentiometer. The exact function is unknown to us, the results after proper tuning are, however, reasonable. To assess the applicability of the Hidex 300 SL system for metrological measurements, comprehensive measurements against a reference were performed. The reference was realized by measurements of the same samples with the PTB–TDCR system (Nähle et al., 2010). For processing the analog signals, conventional NIM electronics are used and the digital signal processing including live-time correction is carried out by the MAC3 module developed at LNHB (Bouchard and Cassette, 2000). Currently, a new type of a coincidence module developed by PTB, which is based on FPGAs, is also being tested. The reference solutions used for the preparation of the samples were calibrated by other standardization methods, i.e. LS counting using the CIEMAT/NIST method, 4pb-g coincidence counting, and ionization chamber measurements. The activity concentration of 241Am was determined by liquid scintillation counting (LSC), assuming a counting efficiency of 1 for this alpha emitter (see Kossert et al., 2009a), and confirmed by ionization chamber measurements. The samples used for the measurements consisted of 15 mL LSC cocktail, weighed aliquots of the active solution and an inactive solution (usually H2O, depending on the chemical composition of the active solution) to come to a total volume of 16 mL. The LS cocktail was chosen to be suitable for the chemical composition of the active solution. Ultima GoldTM was used in most cases. The masses of the active solutions were determined gravimetrically using two Mettler balances traceable to the German national mass standard. Low potassium borosilicate vials with a volume of 20 mL were used in most cases. When a higher light output was desirable (e.g. for 55Fe samples), polyethylene (PE) vials were used. All samples were shaken and centrifuged before starting the measurements. For the application of the TDCR method as described above, the raw data were converted to the standard PTB-format for TDCR data and then analyzed with procedures as described by Nähle and Kossert (2011). For some radionuclides, the efficiency was calculated with an extended version of MICELLE2 (Kossert and Grau Carles, 2010). MICELLE2, an enhanced version of MICELLE (Grau Carles, 2007), is a program for the calculation of LS counting efficiencies via Monte Carlo simulations. It uses a stochastic approach for b and b þ branches and a stochastic atomic rearrangement model for the EC branches. Since the individual counting efficiencies of the 3 detector channels can be different (e.g. due to different quantum efficiencies of PMTs) the method actually requires 3 different parameters which can be determined by minimizing the function e T 2 eT T 2 eT T 2 D¼ T þ þ , eAB AB eBC BC eAC AC tables of double and triple efficiencies as a function of two or three parameters, respectively. These tables can then be used for 2D and 3D efficiency interpolation, which allows asymmetries of the PMT efficiencies to be taken into account. The measurement evaluation is performed with another separate, PTB-developed program that relies on the efficiency tables calculated. For all calculations, a constant kB¼0.0075 cm/MeV was used to compute the ionization quenching function. In general, the results of activity measurements should be independent of the free parameter, i.e. measurements with different counting efficiencies should yield the same result. An efficiency variation is realized using a series of samples with different degrees of chemical quenching. The influence of using a different kB value on the activity results should be considered in the uncertainty budget of activity measurements. When ratios of measurement results of two systems are considered, as done in this paper, the influence of the kB value is negligible. Corrections for dead time and decay were applied, using halflives from Schötzig and Schrader (2000). Several uncertainty components were identified as relevant for the results presented in this paper: a statistical component, the uncertainty assigned to the efficiency as well as the uncertainty due to background variations. These components are calculated for the measurement results both with the HIDEX system and the PTB–TDCR system, and are then used to calculate the uncertainty of the deviation between the results of the two systems. Each sample was measured several times, and for each sample, the arithmetic mean and its standard deviation are calculated. From the results of all samples, the weighted mean is used as the final result. The maximum of the inner and outer uncertainty is taken as the corresponding statistical uncertainty. To estimate the uncertainty assigned to the counting efficiency, the mean TDCR value as well as its standard deviation are calculated for each sample. The uncertainty of the efficiency is then calculated and propagated as the uncertainty component for this individual sample. The maximum uncertainty value of all samples is regarded as the uncertainty component assigned to the efficiency for the activity concentration. The third component relevant here is the uncertainty due to background variation. It is calculated as the standard deviation of the mean of 6 or more repeated background measurements. The background not only influences the count rates, but also the TDCR values and, hence, the efficiency values derived from the TDCR values. Both the uncertainty component of the count rates and the uncertainty component of the efficiency due to background variation are included. Correlations between the triple and double background count rates are taken into account. The uncertainty due to dead time and system linearity can be estimated from the linearity checks, Section 4. It is not considered in the results presented here, but should be considered for activity certification statements. The uncertainty due to weighing is calculated according to the calibration certificate of the balance used for mass determinations. It is neglected in the following when ratios of the results of measurements with the HIDEX 300 SL and the PTB– TDCR are considered, as it is common to both results. As the time between the measurements in both counting systems was usually much less that one half-life of the respective radionuclide, the uncertainties for decay corrections can also be neglected, as well as other uncertainties common to both measurement systems like potential radionuclide impurities, adsorption or decay scheme parameters. see e.g. Broda et al. (2007). Here also the coincidence counting rates between each pair of PMTs AB, BC and AC are used. MICELLE2 was therefore extended, so that all electron emission events are written to files. These files are used by a newly developed program to compute 3. Setting up the HIDEX 300 SL Three important parameters can be influenced using the control software of the system: bias voltages, the threshold and Please cite this article as: Wanke, C., et al., Investigations on TDCR measurements with the HIDEX 300 SL using a free parameter model. Appl. Radiat. Isotopes (2012), doi:10.1016/j.apradiso.2012.02.097 C. Wanke et al. / Applied Radiation and Isotopes ] (]]]]) ]]]–]]] Fig. 1. Single count rates in the PMTs over the threshold setting in the measurement of the coincidence time. At the hardware level the dead-time circuitry can be fine tuned provided that a reference for linearity tests is available (cf. section 4). Optimized settings of the coincidence time and the threshold are crucial for the TDCR method. Therefore, measurements have been carried out to find the settings for which the deviations of the activity results to those of the PTB–TDCR are minimized. Setting the bias voltage (HV) for the three multipliers is done with the aid of the counting protocol ‘‘HVset’’. An unquenched 14C sample is necessary for this procedure, which is described in the software reference. The principle of the HV adjustment is to yield the same QPE (quench parameter external) value in all single channels. The QPE value is derived from the spectrum, it is defined as the 99th percentile of all counts in the spectrum (Aalto, 2009). The goal of the HV adjustment is to get the same gain for all three PMTs. This is necessary for a correct evaluation of the measurements by the TDCR method, as the counter only offers one common threshold setting for all three PMTs. In the latest version, an automatic HV adjustment is also possible. Following the manual adjustment instructions leads to practically the same results as using the automatic procedure, which is more convenient to handle. The design of the electronics demands that the bias voltages fulfill the condition (UR-200 V) rUS,T rUR, which means that the PMTs in position S and T cannot have higher bias voltages than that in position R, and they may only be lower by a maximum of 200 V. In the selection of the PMTs, it must be ensured not only that the photon detectors are matched closely, but that the PMT with the lowest intrinsic efficiency is mounted in position R. In the counter installed at PTB, the PMTs at R and S are very similar, while the PMT at T seems to have a higher intrinsic gain and is, therefore, used with a lower bias voltage. Generally, in TDCR measurements the counts in the whole spectrum are integrated, and the threshold must be low enough so that all events in the single electron peak are registered. If, on the other hand, the threshold is set too low, noise from the PMT is counted, which leads to high dead time and might disturb the measurements. In the service manual (Hidex Oy, 2009) and the software reference (Aalto, 2009), a value of 40 is stated as a ‘‘normal value’’. With the EEPROM version 1.56, the value was preset to 100. Subsequent measurements showed that this threshold can 3 63 Ni. be set lower without increasing noise. A value of 45, however, resulted in the counting of noise, making correct measurements impossible. This can be seen in Fig. 1. Therefore, a threshold value of 50 was chosen, to be as low as possible, but without counting noise. Fig. 2 shows the influence of the threshold settings on the measurement results for 63Ni. This nuclide has been chosen, because the beta energies are high enough for the TDCR model to work well, but low enough for the measurement to be sensitive to imperfect settings. For a threshold setting of 50 the deviations from the PTB–TDCR result are reduced to a minimum. The influence of the coincidence time setting has been studied using samples of 99Tc and 63Ni. In contrast to the PTB–TDCR system, which uses a fixed coincidence resolving time of 40 ns, the coincidence time of the HIDEX 300 SL can be selected. Therefore, measurements have been performed with the coincidence time varied, while all other parameters are kept constant. The threshold was set to 50 (see previous section) and not changed for these measurements, because the influence of the coincidence time is expected to be more prominent for lower threshold settings. The results for 63Ni and 99Tc plotted against the coincidence time are shown in Fig. 3. The uncertainty bars plotted comprise contributions due to statistical variation, the uncertainty assigned to the efficiency and to the background count rates both from the HIDEX and the PTB–TDCR system. A rather small deviation for 63 Ni is in the order of 0.3% is found for a setting of 30 ns, which was used for many measurements at the beginning of this study. For 99Tc, the deviations are generally significantly lower. This can be attributed to the higher endpoint energy of 99Tc, which causes a higher number of emitted photons and, hence, a higher detection efficiency, which leads to a better signal definition and a narrower time distribution of the signals. Shortening the coincidence time can lead to a rejection of true coincidences, which in turn might cause false TDCR values and, hence, an incorrect efficiency. There is also some discussion on more subtle effects especially when measuring low energy nuclides and on the proper setting of the coincidence time in these cases. The increasing deviation with a lower coincidence time setting is probably due to a loss of true triple coincidences, which results in a lower TDCR value and, hence, a lower assigned efficiency, which then leads to an activity value that is too high. At the beginning of this study, the coincidence resolving time was adjusted to 30 ns. With this setting and a threshold of 50, a Please cite this article as: Wanke, C., et al., Investigations on TDCR measurements with the HIDEX 300 SL using a free parameter model. Appl. Radiat. Isotopes (2012), doi:10.1016/j.apradiso.2012.02.097 4 C. Wanke et al. / Applied Radiation and Isotopes ] (]]]]) ]]]–]]] Fig. 2. Deviations in the measurement of 63 Fig. 3. Deviations in the measurement of Ni—influence of the threshold setting. The coincidence time was 30 ns. 63 Ni and 99 Tc over the coincidence time setting. The threshold was 50. reasonable agreement with the results of the PTB–TDCR counter is achieved. Later, the usage of a longer coincidence time was investigated, as presented below in this article. 4. Linearity measurements and dead-time correction Linearity tests to check the count-rate dependence of the results were performed using samples of 241Am. The linearity of the system depends on the correct determination of the dead time, so the tests described here are also checks for a correct dead-time measurement. Samples containing 241Am with expected count rates of approx. 106, 5 105, 2 105, 1.4 105, 1 105, 5 104, 2 104, 1.4 104, 1 104, 5 103, 2 103, 1.4 103 and 1 103 counts per minute were prepared as described in Section 2, Ultima GoldTM AB was used as LS cocktail. Measurements with the counter as it was installed revealed significant deviations of the results from the expected values. With the later versions of the analyzer board supplied by Hidex, a tuning of the dead-time correction is provided (Aalto, 2010), however, without further explanations or any tuning instructions. Therefore, the fine-tuning was done using the 241Am samples, so that the deviations in the most prominent count rate region for our measurements (i.e. 1 105–2 105 counts per minute) are less than 70.2%. The results obtained with the current configuration and dead-time compensation setting are shown in Fig. 4. It seems that the deviation is not perfectly constant over the whole range investigated. In our experience, however, other commercially available LS systems show similar behavior. From this it can be concluded that the system is suitable for activity measurements with low uncertainties, but a further improvement of the linearity of the system installed at PTB is desirable for metrological applications. Please cite this article as: Wanke, C., et al., Investigations on TDCR measurements with the HIDEX 300 SL using a free parameter model. Appl. Radiat. Isotopes (2012), doi:10.1016/j.apradiso.2012.02.097 C. Wanke et al. / Applied Radiation and Isotopes ] (]]]]) ]]]–]]] Fig. 4. Linearity measurements—Deviations of the measured count rates from the expected ones in the measurement of uncertainties (see text). 5 241 Am. The bars indicate only statistical Fig. 5. Deviation D of the activity concentration derived from measurement data of the HIDEX counter and the PTB–TDCR system vs. average beta-energy. The coincidence time was 30 ns and the threshold was 50. 5. Measurements of different nuclides and comparison with the PTB–TDCR system Several nuclides have been measured, and the results are compared to those from measurements with the PTB–TDCR system. This reduces the model influence of the TDCR method, as the results are gained by the same evaluation procedure. For correct activity determinations, however, further uncertainty components have to be taken into account. The measurements comprised radionuclides decaying by b , b þ , or EC as well as nuclides with complex decay schemes. All samples are measured with a fixed counting time, usually 600 s or 900 s and several repetitions for each sample. Further count terminators are not set. The ROI comprises channel 0–1023, i.e. the whole spectrum. The measurements were evaluated applying the TDCR method. The evaluations are done with and without taking asymmetries of the PMTs into account. However, the differences between symmetrical and asymmetrical evaluation are low (in the order of 0.1% or less, see below). In the following, results from measurements after March 2010 are presented, i.e. measurements with the two latest analyzer board designs. The results are given as relative deviations D, calculated as D¼ aHIDEX aPTB : aPTB Fig. 5 shows the results from those measurements with optimized threshold and coincidence time settings as deviations from the reference values plotted over the average beta energy. The average beta energy can be seen as a measure for the expectation value of the amount of light produced in a decay event. The average beta energies are calculated from data taken from TdeR (2010) and Nica (2009), respectively. Chlorine-36 and 204 Tl are included here although they are not pure beta emitters, as the respective EC branches have only a low probability. For the Please cite this article as: Wanke, C., et al., Investigations on TDCR measurements with the HIDEX 300 SL using a free parameter model. Appl. Radiat. Isotopes (2012), doi:10.1016/j.apradiso.2012.02.097 6 C. Wanke et al. / Applied Radiation and Isotopes ] (]]]]) ]]]–]]] Fig. 6. Deviation D of the activity concentration derived from measurement data of the HIDEX counter and the PTB–TDCR system vs. average beta-energy. The coincidence time was 35 ns and the threshold was 50. comparison of 177Lu results, the analysis was carried out without a correction for asymmetry. Details about the calculation of the efficiency of 177Lu are described in Kossert et al. (2011a). Also for 32 P the small correction due to PMT asymmetries was neglected. Fluorine-18 is a positron emitter with a weak EC branch and, therefore, does not fit into the diagram of b emitters. The deviation of the HIDEX result to that of the PTB–TDCR system is þ0.21%, and it matches perfectly well with the value obtained by measurements applying the CIEMAT/NIST method. Tritium decays by b emission with an endpoint energy of 18.6 keV and an average beta energy of 5.7 keV. Iron-55 is an EC nuclide with relevant Auger and X-ray energies of about 0.5 keV and 5–6 keV. In the determination of activities of both nuclides, model dependencies have a significant influence, since both nuclides emit electrons of low energy, which do not produce a large number of photons. Measurements of these nuclides can be seen more as a test of the TDCR method and the model used than as a test of a particular detector system. Generally speaking, the TDCR models currently used may become inaccurate when the expected mean number of photons in the PMT is lower than about 1 (Cassette et al., 2000). This is the case for TDCR values of about 0.3 and lower, which we find in the measurement of 55Fe and 3H. Up to now, there is no consensus about the necessary consequences for the models used, see e.g. Broda (2008) and Simpson et al. (2010). The deviations between the Hidex-TDCR and the PTB–TDCR results were found to be about þ0.1% for 3H and þ0.22% for 55Fe. Both results can be regarded as good since both radionuclides create only a few scintillation photons. For some measurements the LS samples with 3H were surrounded with an adhesive tape ‘‘tesas Film matt-unsichtbar’’ (width: 19 mm, tesa AG, Germany) to increase the counting efficiency and to decrease the model dependence. For 55Fe, PE vials were used which even yield a higher counting efficiency. With the above-mentioned settings, the dependence on the beta energy (and hence the amount of light produced in the scintillations), is negligible (Fig. 5). The results are slightly higher than those obtained with the PTB–TDCR system. In most cases the deviations are in the order of about þ0.25%. These deviations might be due the imperfect linearity of the Hidex system and/or due to too short coincidence resolving time. With a longer coincidence time setting of 35 ns, the deviations become indeed lower, see Fig. 6. As noted above, the PTB–TDCR system is equipped with a MAC3 module using a fixed coincidence resolving time of about 40 ns. The lower dependence on the coincidence resolving time in the region above 40 ns, as visible in Fig. 3, supports the usage of a longer resolving time. However, such a change causes a negative deviation for 63Ni. The adjustment of coincidence resolving time and the analysis of low-energy emitters require further investigations. Measurements with the Hidex counter using a coincidence resolving time of 40 ns are under way. As results are comparable with those of the PTB–TDCR system, the HIDEX system can be regarded as suitable for TDCR measurements applying a free parameter model provided the dead time correction and the linearity adjustment is appropriate. A drawback for high-precision measurements is (as in all commercial LS-counters) that the user has no information about the signal processing and has to rely on the proper functioning of the electronics, as no checks in the signal processing chain can be performed. The influence of asymmetries in the PMT efficiencies on the activity result is, in general, very low in the HIDEX system installed at PTB, as the three PMTs are well matched. The deviation of the results assuming identical efficiencies in all PMTs from the results taking asymmetries into account over the average beta energy is shown in Fig. 7 for the measurements presented in Fig. 5. For all results shown here, the deviation is considerably smaller than 0.1%, which makes it negligible for nonmetrologic measurements. There is, however, an influence of the average beta-energy: the deviation of the result assuming identical efficiencies becomes slightly larger with lower energy. This can be easily understood, because a higher energy corresponds to a larger number of photons produced in the scintillator cocktail. If more photons impinge on the photocathode of the PMT, differences in the quantum efficiency of the PMT will become less important. Due to the large amount of light produced by alpha emitters, for example, the detection efficiency equals 1 for all relevant detector configurations, irrespective of the quantum efficiency or other asymmetries of the PMTs. 5.1. LS spectrometry method for activity determination of 109 Cd The applicability of the HIDEX system for the determination of the activity of a solution of 109Cd by conversion-electron counting Please cite this article as: Wanke, C., et al., Investigations on TDCR measurements with the HIDEX 300 SL using a free parameter model. Appl. Radiat. Isotopes (2012), doi:10.1016/j.apradiso.2012.02.097 C. Wanke et al. / Applied Radiation and Isotopes ] (]]]]) ]]]–]]] 7 Fig. 7. Deviation of the result, assuming symmetric efficiencies, to the result, taking asymmetries into account. The coincidence time was 30 ns and the threshold was 50. has also been investigated. The method, which is described in Kossert et al. (2006, 2009b), requires the evaluation of measured spectra, as a separation of events from conversion electrons from other events is necessary. Therefore, the measurements cannot be done in the ‘‘METRO’’ counting mode, as this is not capable of producing spectra. Hence, the measurements were done in the standard ‘‘BETA’’ mode. The 109Cd LS samples were prepared in PE vials containing a mixture of Ultima GoldTM AB and Ultima GoldTM F to achieve a good spectrum resolution. The result of three samples with the 109 Cd solution measured twice each is in excellent agreement with the reference value; the (unweighted) mean shows a deviation of only 0.02%. Further details about the method with the latest improvements were described by Kossert et al. (2009b). 5.2. TDCR-Čerenkov method The TDCR-Čerenkov method (Kossert, 2010) was also applied with the Hidex counter. To this end, weighed portions of a 32P solution were filled into polyethylene vials containing about 12 mL of distilled water. Counting efficiencies were computed with the model as presented by Kossert et al. (2011b) taking into account PMT asymmetries. Also the wave-length dependence of the refractive index of water (dispersion) as well as the PMT response curves were taken into account. The determined activity concentration was found to be about 0.1% lower than the results obtained by LS counting with PTB counters. The result indicates that the TDCR-Čerenkov method can be applied with the Hidex system. This is interesting since a mixture of radioactive material and the organic scintillation cocktail can be avoided. 6. Conclusions and outlook The HIDEX 300 SL is, in general, applicable for activity measurements with low uncertainties. For the nuclides studied here, the results are, in general, comparable to those obtained with the PTB–TDCR system. The prerequisite for correct measurements is, however, that the system linearity is satisfactory. This requires the correct tuning of a dead-time compensation circuit, which requires some amount of work and a series of samples of an alpha emitter like 241 Am, which might not be available to some users. It is desirable to further improve the linearity of the system installed at PTB. Moreover, bias voltages, threshold and coincidence time have to be adequately set. It should also be noted that the model dependencies can be very large when studying nuclides with low photon yield, like 3H and 55Fe. Also the sample composition as well as the choice of vials may have significant influence when measuring such radionuclides. It must be pointed out that the results presented in this work were obtained with a TDCR apparatus which was modified by Hidex Oy. Thus, it would be interesting to check whether other ‘‘METRO’’ versions of the Hidex counter give comparable results. To this end, a comparison between ENEA (Italy) and PTB is planned within the scope of the EMRP MetroFission project (EMRP, 2009). Such a comparison will yield valuable results on the reproducibility of counters. Acknowledgments The authors wish to thank the staff of Hidex Oy for their kind support and for providing and adapting the TDCR counter. The research leading to these results has received funding from the European Union on the basis of Decision No. 912/2009/EC. References Aalto, J., 2009. Hidex 300 SL software reference, valid for embedded software version 1.44 and above, 2009-06-04. Aalto, J., 2010. Logic card, version J and A/B card, version I adjustment procedure. Draft. 2010-06-14. Broda, R., Cassette, P., Kossert, K., 2007. Radionuclide metrology using liquid scintillation counting. Metrologia 44, S36–S52. Broda, R., 2008. Some remarks on photons statistics in the LS-counter. Appl. Radiat. Isot. 66, 1062–1066. Bouchard, J., Cassette, P., 2000. MAC3: an electronic module for the processing of pulses delivered by a three photomultiplier liquid scintillation counting system. Appl. Radiat. Isot. 52, 669–672. Cassette, P., Broda, R., Hainos, D., Terlikowska, T., 2000. Analysis of detectionefficiency variation techniques for the implementation of the TDCR method in liquid scintillation counting. Appl. Radiat. Isot. 52, 643–648. Cassette, P., 2010. TDCR in a nutshell. Presentation at the LSC 2010 conference. /http://www.nucleide.org/LSC2010/presentations/O-60.pdfS. Please cite this article as: Wanke, C., et al., Investigations on TDCR measurements with the HIDEX 300 SL using a free parameter model. Appl. Radiat. Isotopes (2012), doi:10.1016/j.apradiso.2012.02.097 8 C. Wanke et al. / Applied Radiation and Isotopes ] (]]]]) ]]]–]]] EMRP, 2009. Energy—Joint Research Project Summary—Metrology for new generation nuclear power plants MetroFission, 2009. See /http://www.euramet. org/index.php?id=a169jrpsS. Grau Carles, A., 2007. MICELLE, the micelle size effect on the LS counting efficiency for electron-capture and capture-gamma emitters. Comput. Phys. Commun. 174, 35–46. HIDEX Oy (Ed.), 2009. 300 SL Automatic TDCR liquid scintillation counter service manual. Kossert, K., Janßen, H., Klein, R., Schneider, M., Schrader, H., 2006. Standardization and decay data of 109Cd. Appl. Radiat. Isot. 64, 1031–1035. Kossert, K., Jörg, G., Nähle, O., Lierse, v., Gostomski, C., 2009a. High-precision measurement of the half-life of 147Sm. Appl. Radiat. Isot. 67, 1702–1706. Kossert, K., Ott, O., Nähle, O., 2009b. Improved techniques for the activity standardization of 109Cd by means of liquid scintillation spectrometry. In: Eikenberg, J., Jäggi, M., Beer, H., Baehrle, H. (Eds.), LSC 2008, Advances in liquid scintillation spectrometry, pp. 97–107. Kossert, K., 2010. Activity standardization by means of a new TDCR-Čerenkov counting technique. Appl. Radiat. Isot. 68, 1116–1120. Kossert, K., Grau Carles, A., 2010. Improved method for the calculation of the counting efficiency of electron-capture nuclides in liquid scintillation samples. Appl. Radiat. Isot. 68, 1482–1488. Kossert, K., Grau Carles, A., Nähle, O., 2011a. Čerenkov counting and liquid scintillation counting of 36Cl. In: Cassette Philippe (Ed.), Radiocarbon, 2011, LSC2010, Advances in Liquid Scintillation Spectrometry: Proceedings of the 2010 International Conference on Liquid Scintillation Spectrometry, Paris, France, 6–10 September 2010. The University of Arizona, Tucson, AZ, USA, pp. 161–170. ISBN 978-0-9638314-7-7. Kossert, K., Nähle, O.J., Ott, O., Dersch, R., 2011b. Activity determination and nuclear decay data of 177Lu. ICRM 2011, these proceedings. Nähle, O., Kossert, K., Cassette, P., 2010. Activity standardization of 3H with the new TDCR system at PTB. Appl. Radiat. Isot. 68, 1534–1536. Nähle, O., Kossert, K., 2011. Comparison of the TDCR method and the CIEMAT/NIST method for the activity determination of beta emitting nuclides. In: Cassette Philippe (Ed.), Radiocarbon, 2011, LSC2010, Advances in Liquid Scintillation Spectrometry: Proceedings of the 2010 International Conference on Liquid Scintillation Spectrometry, Paris, France, 6–10 September 2010. The University of Arizona, Tucson, AZ, USA, pp. 313–320. ISBN 978-0-9638314-7-7. Nica, N., 2009. Nuclear Data Sheets for A ¼ 147. Nuclear Data Sheets 110 (4), 749–997. Schötzig, U., Schrader, H., 2000. Halbwertszeiten und Emissionswahrscheinlichkeiten von häufig verwendeten Radionukliden. PTB-Report Ra-16 (5th Edition), Braunschweig May 2000, ISBN:3-89701-279-0. Simpson, B.R.S., van Wyngaardt, W.M., Lubbe, J., 2010. Fe-55 activity measurements at the NMISA revisited. Appl. Radiat. Isot. 68, 1529–1533. TdeR, 2010. Table de Radionucléides (Table of Radionuclides). /http://www. nucleide.org/DDEP.htmS. Also in: Nucléide, 2010. CD-ROM version of an interactive database for decay data. Editor: Bé, M.-M., BNM-LNHB, C.E.A. Saclay, Gif-sur-Yvette, France. Please cite this article as: Wanke, C., et al., Investigations on TDCR measurements with the HIDEX 300 SL using a free parameter model. Appl. Radiat. Isotopes (2012), doi:10.1016/j.apradiso.2012.02.097