ARVO 2016 Annual Meeting Abstracts 535 Refractive Error and Aberration Thursday, May 05, 2016 11:00 AM–12:45 PM Exhibit/Poster Hall Poster Session Program #/Board # Range: 6235–6257/A0172–A0194 Organizing Section: Visual Psychophysics/Physiological Optics Contributing Section(s): Anatomy/Pathology Program Number: 6235 Poster Board Number: A0172 Presentation Time: 11:00 AM–12:45 PM Refraction change after prolonged accommodation with different visual tasks Jason Shen, Frank Spors, Lance McNaughton, Chunming Liu. Extramural Res, Western Univ of Hlth Sciences, Pomona, CA. Purpose: Hyperopic defocus, either on- or off-axis, has been hypothesized to be a stimulus for the development of refractive error. Prolonged accommodation has been identified as a risk associated with myopia progression, but it is still unclear if the accommodation stimulated by the various prolonged daily visual tasks is in the comparable ammount. The purpose of our study was to observe the refraction change after prolonged accommodation with different visual tasks. Methods: A commercial Shack-Hartmann wavefront aberrometer was used to take measurements. Thirty subjects were recruited in the study and they were instructed to work on different visual tasks (read on computer, game on smartphone, watch movie on computer and read on paper, all at a distance of 50 cm) for 30 minutes. The measurements of central and peripheral refraction (30 degrees temporal visual field of right eye) were taken at start, 15 minutes and 30 minutes. Results: Slightly increased accommodation (0.08 ± 0.013 D) was observed when subjects were given reading tasks printed on paper from 15 minutes to 30 minutes. All the other three visual tasks, displayed either on computer monitor or smart phone screen, leaded a decreased accommodation over the measurements time frame (0.19 ± 0.063 D). Conclusions: With prolonged near work, accommodation lag increased for visual tasks displayed on computer or smart phone screen. Increased accommodation lag indicated hyperopic shift of the retinal image and this shift of hyperopic defocus was symmetric across the visual field. The formed hyperopic image shell after prolonged accommodation is associated with further myopia progression. Commercial Relationships: Jason Shen, None; Frank Spors, None; Lance McNaughton, None; Chunming Liu, None Program Number: 6236 Poster Board Number: A0173 Presentation Time: 11:00 AM–12:45 PM Effects of different multifocal contact lenses on accommodation and ocular aberrations Irene Siso-Fuertes, Hema Radhakrishnan. Optometry, The University of Manchester, Manchester, United Kingdom. Purpose: At present, simultaneous vision multifocal contact lenses (MFCLs) are being used not only for presbyopia correction, but for other purposes such as myopia control. We investigate the influence of two different designs of multifocal contact lenses on the accommodative response (AR) function and higher order aberrations (HOAs). Methods: Measurements of the ocular aberrations and AR were taken for 21 eyes of young healthy subjects aged between 18 and 41 years with astigmatism less than 1.25 D (mean spherical equivalent was: -1.58 D). We used a Hartman-Shack aberrometer (IRX3, Imagine Eyes) which incorporates a Badal system that was programmed to present fixed stimuli from 0 D to 4 D of accommodation in 0.50 D steps. Accommodation measurements were recorded for distance vision correction of the spherical equivalent with the Badal system and with two MFCLs. Participants were randomly fitted with two different MFCLs: PureVision 2 For Presbyopia (centre-near design) and Proclear Multifocal (centre-distance design). The later has been previously used in myopia control studies. Aberrations data was extracted for a 4mm fixed pupil diameter, so the effects of the contact lenses could be assumed to be the same for every subject. Results: The slope of the best linear trend fit for the AR function and the change of equivalent spherical aberration (ESA) with accommodation was calculated. No statistically significant differences were found either for the AR (F = 0.415, p = 0.663) or for the ESA (F = 1.862, p = 0.169) between the three conditions tested. However, the high standard deviation values show that the effective AR as well as the ESA during accommodation differs across subjects. Conclusions: Young subjects fitted with the MFCLs used in this study preserved their accommodative function. Therefore, good visual performance and tolerance of these lenses can be expected by early-presbyopes and young patients undergoing myopia control treatments. Presbyopes may still be benefited by the use of these MFCLs since our results are highly subject dependent and they might develop different neural strategies to discriminate the two foci generated by these lenses. Commercial Relationships: Irene Siso-Fuertes, None; Hema Radhakrishnan, None Support: AGEYE 608049-FP7-People Program Number: 6237 Poster Board Number: A0174 Presentation Time: 11:00 AM–12:45 PM Accommodative behavior of young eyes fit with multifocal contact lenses Basal Altoaimi, Meznah S. Almutairi, Arthur Bradley. School of Optometry, Indiana University, Bloomington, IN. Purpose: Hyperopic defocus is implicated in the development of myopia, and multifocal contact lenses (CLs) have been proposed for removing this hyperopic growth stimulus. We examined the effectiveness of such CLs for removing hyperopic defocus while accommodating. Methods: Ocular wavefront errors and pupil sizes were measured with a high resolution Shack-Hartmann aberrometer as 20/40 characters were viewed at distances from 2m to 20 cm by young adult subjects (n=8), who were instructed to keep the target as clear as possible and tested both monocularly and binocularly with single vision (SV), center-distance (CD) and center near (CN) +2.00D add multifocal CLs. Accommodation stimulus/response curves were sampled at 0.25 diopter increments, and dynamic accommodative responses were measured as stimuli jumped from the distance to the near optics’ retinal conjugate planes. Results: Average pupil size (unaccommodated) and near pupil miosis were 3.9 mm and -0.14 mm/diopter, respectively. Elevated levels for spherical aberration contributed by concentric bifocals (e.g. mean/SD C40 and C60 are 0.24±0.18μm and -0.08±0.1μm, for CD lenses with 5mm pupils) resulted in paraxial Rx being 2.1D less myopic than minRMS refractive state. As target distance was reduced in 0.25D steps, four subjects accommodated monocularly and binocularly in order to focus the distance optic, while four others failed to accommodate until the target distance was closer than the near optic far point. At -2.5 D, two subjects switched from focusing the distant optic to the near optic (they relaxed their accommodation to a nearer target) when viewing monocularly. However, these subjects continued to focus the distant optic when viewing binocularly. Mean/SD accommodative gain for paraxial Rx was 1.06±0.11, 0.99±0.09, 1.04±0.11 with SV, CD, and CN lenses, respectively. These abstracts are licensed under a Creative Commons Attribution-NonCommercial-No Derivatives 4.0 International License. Go to http://iovs.arvojournals.org/ to access the versions of record. ARVO 2016 Annual Meeting Abstracts When stepping from the distant to the near retinal conjugate planes, most subjects did not accommodate when viewing monocularly, but they accommodated with typical latency and dynamics when viewing binocularly. Conclusions: By introducing a bifocal optic into the visual path, one image plane (created by the near optic) will exist in front of the retina as long as the subject accommodates to focus the distant optic. However, some young subjects focus with the near optic to reduce accommodative effort, and thus create hyperopic defocus with the distance optic while viewing at near. Commercial Relationships: Basal Altoaimi, None; Meznah S. Almutairi, None; Arthur Bradley, None Program Number: 6238 Poster Board Number: A0175 Presentation Time: 11:00 AM–12:45 PM Age, eye dominance, and phoria cause errors in measurement of objective refractions Yukari Tsuneyoshi, Hidemasa Torii, Yasuyo Nishi, Yuki Hidaka, Sachiko Masui, Kazuo Tsubota, Kazuno Negishi. Ophthalmology, Keio University School of medicine, Tokyo, Japan. Purpose: Some binocular open-field autorefractors have been developed to avoid errors resulting from measurement using conventional monocular autorefractors under unnatural conditions. There may be smaller errors in older subjects with less accommodation amplitude (AA) because the instrumental myopia is smaller in older subjects. This study tested the hypothesis that the difference between binocular and monocular objective refractions is less in older subjects with less AA than in younger subjects. Methods: Fifty-eight healthy eyes of 29 subjects aged 25 to 60 years (mean, 38.4 ± 10.0 [standard deviation] years) were enrolled prospectively. Objective monocular refractions (MR) were measured with the Nidek Auto Ref/Keratometer ARK-730A (-2.20 ± 2.09 diopters [D]). Objective binocular open-field refractions (BR) (−1.69 ± 2.07 D) and objective AA were measured with the Grand Seiko Auto Ref/Keratometer WAM-5500. Ocular dominance was determined by the hole-in-the-card test. The presence and magnitude of far and near (30 cm) phoria were evaluated using the cover test and alternating cover test using a prism bar. Results: The BR was significantly more hyperopic than the MR by 0.51 ± 0.33 D (P < 0.001). The results of subtracting the MR from the BR were significantly negatively correlated with age (r = -0.231, P = 0.04) and positively correlated with AA (r = 0.223, P = 0.046). When dominant eyes (DE) and non-dominant eyes (NDE) were assessed separately, the correlation between the BR minus the MR and age remained significant in the DE (r = −0.372, P = 0.02) but not in the NDE (r = −0.102, P = 0.30). Far and near phoria were present in one and 10 subjects, respectively. The results of the BR minus MR were significantly correlated with the amount of near phoria (r =0.403, P = 0.02) in the NDE, although there was no correlation in the DE (r =0.110, P = 0.29). Conclusions: Our results are consistent with our hypothesis that the difference between binocular and monocular objective refractions is less in older subjects with less AA. The ocular dominance and position in addition to the AA should be considered when dealing with data measured by monocular instruments. Commercial Relationships: Yukari Tsuneyoshi, None; Hidemasa Torii, None; Yasuyo Nishi, None; Yuki Hidaka, None; Sachiko Masui; Kazuo Tsubota, None; Kazuno Negishi, None Program Number: 6239 Poster Board Number: A0176 Presentation Time: 11:00 AM–12:45 PM Sex-related differences in axial length, anterior chamber depth and lens thickness in elementary school students Takehiro Yamashita, Naoya Yoshihara, Taiji Sakamoto. Kagoshima University, Kagoshima-Shi, Japan. Purpose: To investigate the sex-related differences and the association in axial length, anterior chamber depth and lens thickness in elementary school students’ eyes. Methods: Prospective cross-sectional observational study of 108 right eyes in healthy Japanese young 54 boys and 54 girls (age 8 or 9 years). Axial length, anterior chamber depth and lens thickness were measured with OA-2000 (TOMEY, Japan). The sex-related differences and the association in axial length, anterior chamber depth and lens thickness were investigated using Welch’s t-test and linear regression analysis. Results: The axial length was significantly longer in boys (23.71 ± 0.81 mm) than in girls (23.17 ± 0.92 mm) (p=0.001). The anterior chamber depth was significantly shallower in girls (3.53 ± 0.25 mm) than in boys (3.71 ± 0.22 mm) (p=0.001). There was no significantly difference in lens thickness between boys (3.53 ± 0.81 mm) and girls (3.56 ± 0.19 mm) (p=0.63). The anterior camber depth (R=0.63, 0,56, p<0.001) and lens thickness (R=-0.54, -0.59, p<0.001) were significantly associated with axial length in boys and girls. Conclusions: In elementary school students, there was significantly difference in axial length and anterior chamber depth between boys and girls. The anterior chamber depth was significantly positively associated with axial length and the lens thickness was significantly negatively associated with axial length in boys and girls. Commercial Relationships: Takehiro Yamashita, None; Naoya Yoshihara, None; Taiji Sakamoto, None Support: JSPS KAKENHI grant number 26462643 Clinical Trial: http://www.umin.ac.jp/, UMIN000015239 Program Number: 6240 Poster Board Number: A0177 Presentation Time: 11:00 AM–12:45 PM Relationship between lifestyle and axial length in Japanese elementary school students Kazuki Fujiwara, Takehiro Yamashita, Naoya Yoshihara, Taiji Sakamoto. Kagoshima University, Kagoshima, Japan. Purpose: To investigate the relationship between lifestyle and axial length in Japanese elementary school students’ eyes. Methods: Prospective cross-sectional observational study of 122 right eyes in healthy Japanese young 61 male and 61 female (age 8 or 9 years). Axial length were measured with OA-2000 (TOMEY, Japan). Questionnaires about the children’s daily lifestyles (indoors studying, television viewing, screen time on a computer/smart phone, outdoor activities, bedtime, dietary habits (Japanese or Western style)), parental myopia were completed. The relationship between the axial length and lifestyle or family members’ myopia were investigated Spearman’s correlation analysis. Results: The mean axial length was 23.1 ± 0.9 mm. The indoor studying, television viewing, outdoor activities, bedtime were not correlated with axial length. The screen time (R=0.24, p=0.008), Westernization of dietary habits (R=0.24, p=0.01) and parental myopia (R=0.39, p<0.001) were significantly correlated with axial length. Conclusions: In Japanese elementary school students, long screen time, Westernization of dietary habits and parental myopia were associated with longer axial length. Commercial Relationships: Kazuki Fujiwara, None; Takehiro Yamashita, None; Naoya Yoshihara, None; Taiji Sakamoto, None These abstracts are licensed under a Creative Commons Attribution-NonCommercial-No Derivatives 4.0 International License. Go to http://iovs.arvojournals.org/ to access the versions of record. ARVO 2016 Annual Meeting Abstracts Support: JSPS KAKENHI grant number 26462643 Clinical Trial: http://www.umin.ac.jp/, UMIN000015239 Program Number: 6241 Poster Board Number: A0178 Presentation Time: 11:00 AM–12:45 PM Relationship between retinal artery trajectory and axial length in Japanese elementary and junior high school students Aiko Hayashi, Takehiro Yamashita, Naoya Yoshihara, Taiji Sakamoto. Kagoshima University, Kagoshima city, Japan. Purpose: Trajectory of the supra and infra temporal retinal artery is associated with the position of the nerve fiber layer defects in glaucomatous eyes. However, there is no report about the changes of retinal artery trajectory (RAT) along with growth. Therefore, the purpose of this study was to investigate the difference of the RAT between elementary and junior high school students’ eyes and its association with axial length. Methods: Prospective cross-sectional observational study of 122 right eyes in healthy elementary school students, 61 male and 61 female (age 8 or 9 years) and 170 right eyes in healthy junior high school students, 83 male and 87 female (age 12 or 13 years). Axial length was measured with OA-2000 (TOMEY, Japan). Color fundus photograph was taken by 3D OCT-1 Maestro (TOPCON, Japan). The RAT was plotted in the color fundus photographs and fitted to a second degree polynomial equation (ax2/100+bx+c) by ImageJ. The coefficient “a” represented the steepness of the trajectories. The RAT and axial length differences between elementary and junior high school students were investigated using Mann-Whitney U test. The association between RAT and axial length was investigated using Spearman’s correlation analysis. Results: The axial length and RAT of junior high school students were significantly greater than that of elementary school students (p<0.001). The RAT was significantly associated with axial length in elementary (R=0.26, p=0.005) and junior high school students (R=0.32, p<0.001). Conclusions: Junior high school students have longer axial length and narrower RAT than elementary school students. A longer axial length is associated with narrower RAT in elementary and junior high school students. Commercial Relationships: Aiko Hayashi, None; Takehiro Yamashita, None; Naoya Yoshihara, None; Taiji Sakamoto, None Support: JSPS KAKENHI grant number 26462643 Clinical Trial: http://www.umin.ac.jp/, UMIN000015239 Program Number: 6242 Poster Board Number: A0179 Presentation Time: 11:00 AM–12:45 PM Mirror symmetry of peripheral aberrations for the eyes of isoand aniso-myopes Uchechukwu L. Osuagwu, David A. Atchison, Marwan Suheimat. Optometry & Vision Sciences, Institute of Health & Biomedical Innovation, Queensland University of Technology, Brisbane, QLD, Australia. Purpose: Peripheral optical quality is evaluated usually in only one eye of a person with the assumption that peripheral aberrations change similarly across visual fields of both eyes. Only one study has tested this assumption off-axis, and then only for a few angles along the horizontal field of isometropes. We investigate the assumption by measuring peripheral aberrations in both eyes of iso- and anisomyopes across the field. Methods: Cycloplegic peripheral aberration for 5 mm pupils was measured at 39 locations across 42°x32° of right and left eye fields with a COAS-HD Hartmann-Shack aberrometer in 19 isomyopes [mean age 30±3 yrs; spherical equivalent refraction M (right/ left): –2.2±1.9D/–2.4±1.9D] and 10 anisomyopes [29±6 yrs; M: –4.0±1.8D/–4.3±2.9D]. Isomyopes had interocular refraction differences <1.0D. Anisomyopes had interocular refraction differences between 1.0D and 2.6D. Pearson correlations of 2nd–4th order Zernike coefficients between the two eyes were determined at corresponding field positions e.g. temporal positions were compared. Orthogonal regression determined relationships between coefficients of the two eyes across the visual field. Results: Aberration coefficient patterns across the visual field changed similarly in both refractive groups as follows: quadratic rates of change for C(2,-2) & C(2,2), linear rates for C(3,1) and C(3,-1), and little change in C(4,0) and other 4th-order aberrations. Ignoring C(2,0), there were significant right-left correlations at >50% of field locations for C(2,2), C(3,-1), C(2,-2), C(4,0), C(3,1), C(4,-2) and C(4,2) in isomyopes and for C(2,2), C(3,-1), C(4,0), C(3,1), C(3,-3), C(4,-2) and C(4,2) in anisomyopes. The slopes of the correlation between right and left eyes in the orthogonal regression were not significantly different from either +1 for C(2,2), C(3,-1), C(3,-3), C(4,2) and C(4,4) or from –1 for C(2,-2), C(3,1), C(3,3) and C(4,-4) in isomyopes. For anisomyopes, the slopes were not significantly different from +1 for C(2,2), C(3,-1), C(4,0), C(4,2) and C(4,4) or from –1 for C(2,-2), C(3,1), C(3,3), C(4,-2) and C(4,-4). Conclusions: Aberrations were similar between eyes of iso- and aniso-myopes across the visual field for 2nd – 4th order coefficients. In a pooled data set, coefficients require sign changes for one eye. The slopes of the correlation show that most ocular aberrations are mirror symmetric between eyes about the vertical axis. Commercial Relationships: Uchechukwu L. Osuagwu, None; David A. Atchison, None; Marwan Suheimat, None Support: Australian Research Council DP140101480 Program Number: 6243 Poster Board Number: A0180 Presentation Time: 11:00 AM–12:45 PMOcular Biometry at Different Times of the Day in Human Subjects Anne Bertolet, Debora L. Nickla, Frances J. Rucker, Fuensanta A. Vera-Diaz. New England College of Optometry, Boston, MA. Purpose: Previous studies show that axial length (AL) and choroidal thickness (CT) fluctuate diurnally. Work in animal models suggests the time of day defocus is introduced has an effect on the diurnal changes seen in AL and CT (Nickla et al 2014). We investigated how CT, retinal thickness (RT) and AL change at two different times of the day and after blur. Methods: AL, RT and CT were measured at the fovea using a Lenstar LS900 before and after a period of myopic defocus at two times of the day (around 12pm and around 6pm, randomized order, two visits). The sequence of tests for each visit was: 20min wash out period, baseline measures, 35min blur [+2.00D over best correction, movie @4m], post-blur measures. Subjects were 21-28yrs old with normal vision in each eye and were classified into myopes (-3.23±2.34D; n=24) or emmetropes (0.24±0.25D; n=18). Criteria for CT and RT followed the guidelines by Read et al (2010). AL, RT and CT were compared between baseline measures at noon and 6pm, and between the pre- and postblur conditions. Results: As expected, baseline data showed that CT was significantly larger at 6pm (348.2±62.3μm) than at noon (333.9±60.5μm) in emmetropes (p=0.01), but not in myopes. There were no correlations with CT and the amount of myopia. RT was larger at noon (190.6±12.7μm) than at 6pm (188.2±11.9μm) for the overall group (p=0.03). In the myopic group, RT was significantly greater for higher amounts of myopia at noon (ρ -0.54, p=0.01), but not at 6pm. No correlation was found for emmetropes. As expected, there These abstracts are licensed under a Creative Commons Attribution-NonCommercial-No Derivatives 4.0 International License. Go to http://iovs.arvojournals.org/ to access the versions of record. ARVO 2016 Annual Meeting Abstracts was a significant correlation for AL and the amount of myopia (ρ=0.81, p<0.01). No differences in AL were found between the noon (24.07±0.34mm) and 6pm (24.11±0.32mm) measures (p=0.35) in either group. While exposure to positive blur caused a decrease in RT at noon that was significant for emmetropes (-1.8±10.1μm; p=0.03), not in myopes, these changes are very small. No significant effect of blur was found on CT for any condition, maybe due to higher variability of the measures. Exposure to blur significantly decreased AL at 6pm (Emmetropes: -5.2±33.0μm, p=0.01; Myopes: -6.9±55.3μm, p=0.02), but not at noon. This may be due to the effect of positive blur. Conclusions: While CT and AL have been shown to follow diurnal rhythms, this is the first report of thicker retinas at noon compared to evening in emmetropes. While AL decreased post blur in the evening, further studies are needed to explore this relationship. Commercial Relationships: Anne Bertolet, None; Debora L. Nickla; Frances J. Rucker, None; Fuensanta A. VeraDiaz, None Program Number: 6244 Poster Board Number: A0181 Presentation Time: 11:00 AM–12:45 PM Relationship between peripapillary choroidal thickness and peripapillary retinal tilt, axial length, papillo-macular position in young healthy eyes Yohei Matsushita, Takehiro Yamashita, Naoya Yoshihara, Yuya Kii, Minoru Tanaka, Kumiko Nakao, Taiji Sakamoto. Kagoshima University, Kagoshima, Japan. Purpose: To investigate the relationship between the peripapillary choroidal thickness (PPCT) and peripapillary retinal tilt (PRT), axial length, papillo-macular position (PMP) in young healthy eyes. Methods: Prospective observational cross-sectional study comprised 119 right eyes of 119 healthy young Japanese participants. PPCT was manually measured at eight locations around the optic disc using the B-scan image of the TOPCON 3D OCT-1000 MARK II RNFL 3.4 mm circle scan. The trajectory of the retinal pigment epithelium of the B-scan image was fitted to sine curve using Image J and the amplitude of the sine curve was used for the degree of the PRT. PMP was assessed using color fundus photograph. The relationship between the PPCT and the PRT, the axial length or PMP were investigated using the Spearman and multiple correlation analysis. Results: The mean age was 25.8 ± 3.9 years and the mean axial length was 25.5 ± 1.4 mm. The PPCT was significantly and negatively associated with the axial length (R=-0.43~-0.24, p<0.01) and positively associated with the PMP (R=0.28~0.37, p<0.01) in all eight locations. The temporal and infra-temporal PPCT were significantly and negatively associated with the PRT (R=-0.31, -0.20, p<0.05). The results of multiple regression analysis were similar to that of Spearman correlation analysis. Conclusions: The PPCT decreased as the axial length increased and PMP decreased. The temporal and infra-temporal PPCT decreased as the PRT increased. Commercial Relationships: Yohei Matsushita, None; Takehiro Yamashita, None; Naoya Yoshihara, None; Yuya Kii, None; Minoru Tanaka, None; Kumiko Nakao, None; Taiji Sakamoto Clinical Trial: http://www.umin.ac.jp/, UMIN000006040 Program Number: 6245 Poster Board Number: A0182 Presentation Time: 11:00 AM–12:45 PM Comparing Topographic Metrics of Disease Detection in Individuals with and without Down Syndrome Jason D. Marsack1, Pete S. Kollbaum2, Heather A. Anderson1. 1 College of Optometry, University of Houston, Houston, TX; 2School of Optometry, Indiana University, Bloomington, IN. Purpose: A high incidence of refractive error and an association with keratoconus (KC) have been reported for the Down syndrome (DS) population. Mathematical metrics that utilize corneal topographic data have been developed to distill complex corneal surface data into a single number in order to detect the presence of KC. Two such metrics are 1) maximum corneal power and 2) KISA%, a combination of 4 independent topographic metrics: central corneal steepening (central K), inferior-superior dioptric asymmetry (I-S), degree of regular corneal astigmatism (AST) and skewed radial axis index (SRAX). The purpose of this study was to compare topographic disease detection metrics in both eyes of individuals with and without DS. Methods: Zeiss Atlas corneal topography was attempted on 140 subjects with DS (age range: 8 to 55, mean: 25±9 yrs) and 138 control subjects with self-reported unremarkable ocular history (age range: 7 to 59, mean: 25±10 yrs). Subjects where at least 1 measure of topography was not recorded in both eyes (DS: 23 and control: 0) were excluded from further analysis. Maximum corneal power was determined from the topographic data. Indices central K, I-S, AST, SRAX were calculated and combined to calculate KISA%. Results: Maximum corneal power in the right eyes (mean±std: 48.87D±4.47D) and left eyes (48.56D±4.20D) of subjects with DS were not statistically different. Likewise, in control eyes, maximum corneal power in the right eyes (44.66D±1.75D) and left eyes (44.59D±1.69D) were not statistically different. In subjects with DS, the KISA% values in the right eyes (1Q: 9.60, M (median):23.33, 3Q: 50.13) and left eyes (1Q: 6.81, M: 12.38, 3Q: 33.56) were significantly different (p = 0.0251). In control eyes, KISA% values in the right eyes (1Q: 3.33, M:11.04, 3Q: 16.00) and left eyes (1Q: 3.74, M: 10.25, 3Q: 12.32) were not significantly different. For the DS sample, 22 subjects (19%) exhibited KISA% values indicative of KC unilaterally, with 7 subjects (6%) exhibiting these characteristics bilaterally. None of the control eyes exhibited KISA% values indicative of KC. Conclusions: Between-eye differences in KISA% values were observed in the DS sample, and not seen in the control sample, suggesting more variability in disease severity in the DS group. Further analyses are needed to better characterize the inter-ocular variability for subjects with DS. Commercial Relationships: Jason D. Marsack, None; Pete S. Kollbaum, None; Heather A. Anderson, None Support: NIH R01 EY024590 to HAA NIH T35 EY7088-28 to UHCO Program Number: 6246 Poster Board Number: A0183 Presentation Time: 11:00 AM–12:45 PM Repeatability and Between-Instrument Agreement of the Discovery System Mylan Nguyen, David A. Berntsen. The Ocular Surface Institute, University of Houston College of Optometry, Houston, TX. Purpose: To assess the between-visit repeatability of refractive error (lower-order) and higher-order aberration measurements made with a new commercially-available aberrometer, the Discovery System, and to examine the between-instrument agreement of refractive error measurements between the Discovery System and the Grand Seiko WAM-5500 open-field autorefractor. These abstracts are licensed under a Creative Commons Attribution-NonCommercial-No Derivatives 4.0 International License. Go to http://iovs.arvojournals.org/ to access the versions of record. ARVO 2016 Annual Meeting Abstracts Methods: Five cycloplegic aberrometry measurements of the right eye were made on 25 young adults using the Discovery at two separate visits by the same examiner. Ten cycloplegic measurements were made at visit 2 with the Grand Seiko. Refractive error values (sphere, cylinder, and axis) and Zernike coefficients calculated over a 6-mm pupil through the 6th order were obtained from the Discovery System. Refractive error measurements from both instruments were converted to vector space (M, J0, and J45) and averaged. Higher-order RMS (HORMS) was calculated (3rd through 6th order). Between-visit repeatability and between-instrument agreement of refractive error was assessed using Bland-Altman difference versus mean plots. HORMS and spherical aberration (C4,0) repeatability was also evaluated. A t-test was used to compare the mean difference to zero (bias), and the 95% limits of agreement (LoA) were calculated. Results: Mean (±SD) age and spherical-equivalent refractive error at visit 1 were 23.4 ± 1.7 years and -2.91 ± 1.85 D, respectively. The median interval between the two visits was 5 days (range: 1-14 days). When evaluating between-visit repeatability of the Discovery System, there was no between-visit bias for M, J0, J45, HORMS, or spherical aberration (all p>0.70), and the 95% LoA were ±0.34 D, ±0.14 D, ±0.15 D, ±0.095 µm, and ±0.068 µm, respectively. There was no bias between instruments for defocus (p>0.23), but for J0 and J45, the Discovery on average measured 0.19 D and 0.12 D more positive than the Grand Seiko, respectively (both p<0.01). The 95% LoA for M, J0, and J45 between the two instruments were ±0.82 D, ±0.32 D, and ±0.27 D, respectively. Conclusions: The Discovery System is repeatable and appropriate for measuring lower- and higher-order aberrations longitudinally. Minimal differences in astigmatic measurements between the Discovery and Grand Seiko indicate that refractive error measurements between the instruments should not be used interchangeably. Commercial Relationships: Mylan Nguyen, None; David A. Berntsen Support: NIH/NEI: T35-EY007088 Program Number: 6247 Poster Board Number: A0184 Presentation Time: 11:00 AM–12:45 PM Corneal shape and optical properties: principal component analysis of corneal Zernike coefficients and comparison with other wavefront error representations Jens Buehren1, 2, Krishna P. Vunnava2, Mehdi Shajari2, Thomas Kohnen2. 1Augenpraxisklinik Triangulum, Hanau, Germany; 2 Dept of Ophthalmology, Goethe University Frankfurt, Frankfurt am Main, Germany. Purpose: To describe the corneal wavefront error with principal components obtained from corneal Zernike coefficients and to compare the ability of the novel wavefront error representaion to describe corneal optical properties with other wavefront error represenations. Methods: From 792 normal eyes of 495 patients corneal tomography scans were taken with a commercial Scheimpflug system (Pentacam HR, Oculus, Germany). Based on a ray tracing model, total corneal wavefront aberrations were calculated using a Zernike decomposition up to the 6th order over a pupil diameter of 6 mm. From 27 Zernike coefficients, a principal component analysis (PCA) based on the correlation matrix was performed (SPSS 11.0, Varimax rotation). Coefficient loads of less than |0.25| were ignored. For component selection, an eigenvalue of >1 was applied as threshold. Wavefront errors were built up in a stepwise fashion using the novel components, of which each contained a subset of Zernike coefficients. Similarly, wavefront errors were described with single Zernike coefficients, beginning with C3-3, by the root-mean square (RMS) of Zernike orders 3-6 and by the RMS of all coma, spherical and residual aberrations. For each wavefront error, the optical quality metric BCVSOTF (visual Strehl ratio based on the optical transfer function, simulated for best spectacle correction) was computed (VOL-Pro 7.14, Sarver and Ass.). The number of components to explain 95% of the variance of BCVSOTF was compared. Results: PCA produced 11 components with an eigenvalue >1, accounting for 72% of the total variance. The first 4 components accounted for 95% of the BCVSOTF variance. Using individual Zernike modes, 9 coefficients were necessary to describe 95% of the BCVSOTF variance. For wavefront description with RMS values three components were needed each to describe at least 95% of BCVSOTF variance. Conclusions: Novel wavefront components obtained by PCA were able to describe corneal optical properties as comprehensively as coarser representations such as RMS of Zernike orders. Commercial Relationships: Jens Buehren, Oculus (C); Krishna P. Vunnava, None; Mehdi Shajari, Oculus (C); Thomas Kohnen, Oculus (C) Program Number: 6248 Poster Board Number: A0185 Presentation Time: 11:00 AM–12:45 PM Clinical Evaluation of Handheld Wavefront Aberrometer to Measure Refractive Error in Children Jinu Han1, Nicolas S. Brown2, Sangchul Yoon3, Geunyoung Yoon4. 1 Department of Opthamology, Yonsei University College of Medicine, Seoul, Korea (the Republic of); 2Ovitz Corporation, Rochester, NY; 3Department of Opthamology, National Medical Center, Seoul, Korea (the Republic of); 4Flaum Eye Institute, University of Rochester, Rochester, NY. Purpose: Performing vision screenings and refraction on small children often takes significant amounts of time and requires visiting clinics. The EyeProfiler (Ovitz Corp.) is a portable Shack-Hartmann wavefront sensor used to perform objective refractions. In this study, we demonstrate the feasibility of the device to measure refractive errors in young children. A clinical study is performed on pediatric subjects to compare refraction measurements of the wavefront sensor with established objective refraction techniques. Methods: 17 visually normal subjects (33 eyes) were measured with cycloplegia using the EyeProfiler, cycloplegic retinoscopy, and a commercial autorefractor (Canon RK-1). Subjects ranged in age from 3 to 11 years old (mean ± standard deviation = 6.6 ± 2.1) and were primarily of Asian descent. The wavefront sensor requires the subject to look at a red dot for about 6 seconds. The device automatically determines when it is aligned and records 5 consecutive wavefront images. Measurements were recorded with the other techniques and compared with the wavefront sensor. Results: Initial analysis of the wavefront images found that measurements made on the two youngest subjects (3 years old) were misaligned due to the subject not looking at the red target. These data were discarded for the following correlations. Subjects older than 4 years old were measured properly. Significant correlations were observed between the portable wavefront sensor and cycloplegic retinoscopy for both sphere (R = 0.98, p < 0.001) and cylinder (R = 0.84, p < 0.001). Significant correlations were also observed between the portable wavefront sensor and the commercial autorefractor for both sphere (R = 0.98, p < 0.001) and cylinder (R = 0.84, p < 0.001). The wavefront sensor cylinder measurement tended to differ significantly from the other techniques when their cylinder value was close to zero diopters, reducing the correlation. Conclusions: The portable wavefront sensor was demonstrated to produce accurate refraction measurements in young children based on These abstracts are licensed under a Creative Commons Attribution-NonCommercial-No Derivatives 4.0 International License. Go to http://iovs.arvojournals.org/ to access the versions of record. ARVO 2016 Annual Meeting Abstracts clinical comparisons to existing objective refraction measurements. Measurements can be made rapidly without subjective and complicated input from the patient. This is valuable for examinations in office clinics and can be performed outside office settings. Additional measurements in children should be continued to ensure the statistical significance of the results. Commercial Relationships: Jinu Han, None; Nicolas S. Brown, Ovitz Corporation; Sangchul Yoon, Ovitz Corporation (C); Geunyoung Yoon, Ovitz Corporation (C) Program Number: 6249 Poster Board Number: A0186 Presentation Time: 11:00 AM–12:45 PM Higher order statistical eye model for keratoconic eyes Jos J. Rozema2, 1, Pablo Rodriguez3, Rafael Navarro3, Marie-José Tassignon2, 1. 1Medicine and Helath Sciences, University of Antwerp, Antwerp, Belgium; 2Ophthalmology, Antwerp University Hospital, Edegem, Belgium; 3University of Zaragoza, Zaragoza, Spain. Purpose: To present a stochastic model of the corneal and ocular biometry in keratoconic eyes, based on previous work. This model is capable of generating an unlimited number of random, but realistic biometry sets, including the corneal elevation, intraocular distances and wavefronts, with the same statistical and epidemiological properties as the original keratoconic data it is based on. Methods: The data of 145 keratoconic eyes of 145 patients (aged 18 – 60 years) was recorded with an autorefractometer, Scheimpflug imaging (Oculus Pentacam), optical biometer (Haag–Streit Lenstar) and an aberrometer (Tracey iTrace), which lead to a set of 97 biometric parameters. In order to reduce this number to 18 parameters, the Zernike coefficients of corneal elevation were compressed using Principal Component Analysis. These data were subsequently fitted with a linear combination of multivariate Gaussians through an Expectation Maximization algorithm, from which it is possible to generate an unlimited number of random biometry sets with the same distributions as the original data. These biometry sets can then be used to calculate the associated wavefronts and other ocular parameters. Equality between the original keratoconic data and the synthetic data was assessed using “two onesided” tests. Results: In order to verify the accuracy of the wavefront calculations, the wavefronts derived from the measured biometry were compared to the originally measured wavefronts and found significantly equal (two one-sided t test, p < 0.05). Next the biometry of 1000 synthetic eyes were generated by the stochastic model, followed by ray tracing to obtain the associated wavefronts. These synthetic data were found significantly equal to the originally measured data (two one-sided t test, p < 0.05), thus making them statistically indistinguishable. Conclusions: To the best of our knowledge this is the first eye model dedicated specifically to keratoconus. It produces synthetic biometry data of eyes that is indistinguishable from actual measurements. This model may be interesting for visual optics researchers that do not have access to actual biometry data. Commercial Relationships: Jos J. Rozema, None; Pablo Rodriguez, None; Rafael Navarro, None; MarieJosé Tassignon, None Support: IWT/110684, FIS2014-58303-P Program Number: 6250 Poster Board Number: A0187 Presentation Time: 11:00 AM–12:45 PM Simulation of commercial versus theoretically optimized contact lenses David Rio, Kelly WOOG, Richard Legras. Department of Optometry, Laboratoire Aimé Cotton, CNRS, ENS Cachan, Université Paris-Sud, Univ. Paris-Saclay, Orsay, France. Purpose: Based on image simulation, we aimed to compare throughfocus image qualities (TFIQ) with optical profiles measured on various multifocal contact lenses and theoretically optimized bifocal profiles. Methods: The subjective image quality was assessed using a continuous 5-items grading scale. Twenty young normally sighted subjects judged 3 times the quality of computationally blurred images through a 3-mm artificial pupil limiting the impact of their aberrations. The simulated images were calculated for a 4.5-mm pupil diameter from -5 to +2 D each 0.25 D and with 10 optical profiles. We based our simulation on published measurements of the optical profiles of 4 highest addition multifocal contact lenses : Oasys for presbyopia®, Biofinity Multifocal®, Air Optix Aqua Multifocal® and Purevision Multifocal®. These optics were compared to previously published optimized bifocal optical profiles with 2, 5 and 8 concentric zones and their variants including spherical aberrations (SA), and finally a combination of SA. To quantify the efficiency of an optical profile, we calculated, based on the TFIQ score, Depth-of-Focus (i.e. DoF, range of proximities over which an acceptable level of vision is obtained), and the benefit defined as the area under the throughfocus subjective quality of vision curve higher than 2 (i.e. level from which the quality of vision becomes acceptable). This criterion was normalized by the naked eye condition. Results: TFIQ score showed large inter-individual variation, but the average curve was similar to previously published TFIQ score. Except with the Oasys profile, the other commercial profiles provided TFIQ curves with only one peak of quality. Based on DoF and benefit criteria, commercial profiles did not provide acceptable distance and near qualities of vision. The eight concentric zone profile was found to be the most efficient optic. Adding combinations of SA to a bifocal profile degraded the benefit. We could split the population into 3 groups as a function of the advantage they could obtain whatever the tested optics (i.e. a degradation, no difference or a large benefit), explaining why some subjects could never be satisfied with multifocal contact lenses. Conclusions: Image simulation allows to efficiently evaluate optical profiles and permits to determine that optimized profiles could be better than tested commercial optics. Benefit as a function of Depth-of-Focus for the 10 tested profiles These abstracts are licensed under a Creative Commons Attribution-NonCommercial-No Derivatives 4.0 International License. Go to http://iovs.arvojournals.org/ to access the versions of record. ARVO 2016 Annual Meeting Abstracts Commercial Relationships: David Rio, None; Kelly WOOG; Richard Legras, None Program Number: 6251 Poster Board Number: A0188 Presentation Time: 11:00 AM–12:45 PM Current trends of power and spherical aberration profiles in commercial multifocal soft contact lenses Eon Kim1, Ravi C. Bakaraju1, 2, Klaus Ehrmann1, 2. 1Brien Holden Vision Institute, Sydney, NSW, Australia; 2School of Optometry and Vision Science, UNSW, Sydney, NSW, Australia. Purpose: To evaluate the optical power and spherical aberration (SA) profiles of commercially-available soft multifocal contact lenses (MFCL) across a range of powers and compare their optical designs. Methods: The power profiles of thirty-eight types of the most commonly prescribed MFCLs from the 4 major commercial CL manufacturers were evaluated. Three lenses each were measured in powers +6 D, +3 D, +1 D, -1 D, -3 D and -6 D using NIMO TR1504 (Lambda-X, Belgium). All lenses were measured across 8 mm optic zone diameter in multifocal mode of operation. The amount of SA was calculated between 1.0 and 3.5 mm half chord for all types. Results: Three classes of power profiles were identified: center-near, center-distance and concentric-zone ring-type designs. Most of the lenses were designed with noticeable amounts of spherical aberration. The lens types were categorized into two distinct groups: lens types where SA was power dependent, usually more negative for higher minus powers and more positive for higher plus powers, which on average varied from -1.91 D to +1.20 D across the power range with slope of 0.258 (95% confidence interval (CI) 0.234 to 0.282), and lens types where SA was consistent across the power range, which on average varied from -1.92 D to -1.60 D across the power range with slope of 0.020 (95% CI 0.003 to 0.038) (see Figure 1). In terms of generational changes, the relative plus power for PureVision Hi add MF of approximately 2.50 D was reduced to 2.00 D for the PureVision2 lens. The relative plus for the high add Acuvue lenses dropped from 3.00 D for the Acuvue Bifocal to 1.70 D for the ACUVUE® OASYS® for PRESBYOPIA and further to 1.35 D for the 1-DAY ACUVUE® MOIST MF. Conclusions: Power profiles vary widely between the different lens types, however certain similarities were noticed between some of the center-near designs. For the more recently released lens types, there appears to be a trend emerging to reduce the relative plus with respect to prescription power, to include negative spherical aberration, and to keep the power profiles consistent across the power range. Average of spherical aberration across six power range for lens types with non-uniform SA and lens types with uniform SA as a function of power. Error bars indicate standard deviations between lens types in each group. Commercial Relationships: Eon Kim; Ravi C. Bakaraju, Brien Holden Vision Institute; Klaus Ehrmann, Brien Holden Vision Institute Program Number: 6252 Poster Board Number: A0189 Presentation Time: 11:00 AM–12:45 PM Global refraction and aberration profiles with single-vision and multifocal contact lenses Ravi C. Bakaraju2, 1, Cathleen Fedtke2, Jiyoon Chung2, Darrin Falk2, Klaus Ehrmann2, 1. 1School of Optometry and Vision science, Sydney, NSW, Australia; 2Brien Holden Vision Insititue, Sydney, NSW, Australia. Purpose: To evaluate the effects of single-vision (SV) & multifocal (MF) contact lenses (CL) on the horizontal (H), oblique (O) & vertical (V) peripheral refraction (PR) & aberration profiles. Methods: Forty myopic participants [24.2 yr ± 2.4, 65% F, SE:-0.5 to -4.5D] were fitted randomly with the following lenses: 6 SV’s [AirOptix, Biofinity, Clariti, Proclear, Night & Day and Oasys] and 9 MF’s [Acuvue bifocal, AirOptix, Purevision in low & high adds and MiSight]. The PR profiles across H, O and V meridians, spanning visual field angles from -50° to +50° in 10° steps, were measured with BHVI-EyeMapper, for all lens types at +1D and -3D target vergences. Four repeats were performed with fellow eye occluded. Analysis was done at 4mm pupil. Relative PR (RPR) are reported. Results: At +1D vergence, with exception of Night & Day SV (M=0.31D), all SVCLs produced a hyperopic shift in H-RPR; Biofinity SV produced the steepest drift (M=0.27D). The RPRs remained myopic for all SVCLs in the O & V meridians, where Night & Day SV produced greatest (M=-0.52D) and Biofinity produced the least myopic shift (M=-0.16D). All MFCLs, except MiSight (M=-0.89D), produced hyperopic shifts in the H, O and V meridians (M range 0.06 to 0.44D). In all meridians, the RPR patterns observed at +1D vergence were maintained at -3D target vergence. Regardless of the lens type or accommodative state, J0 and J45 became increasingly more negative for the H & O meridians, respectively. Along the V meridian, J0 became increasingly more positive as a function of field eccentricity. In H & V meridians, magnitude of J45 was relatively insignificant. All lenses measured negative spherical aberration (range: -0.004 to -0.098µm), on-axis, except MiSight and Night & Day which measured positive spherical aberration (range: 0.02 to 0.086µm). These abstracts are licensed under a Creative Commons Attribution-NonCommercial-No Derivatives 4.0 International License. Go to http://iovs.arvojournals.org/ to access the versions of record. ARVO 2016 Annual Meeting Abstracts Conclusions: In the H meridian, only Night & Day SV and MiSight produced peripheral myopic shifts. Remaining lenses produced low to moderate levels of peripheral hyperopia (PH). In the O & V meridians, as the baseline RPR profiles were already in the myopic direction, none of the test lenses produced significant levels of PH. In the light of PH theory, we speculate that some lenses that reduced PH in the H meridian may offer a therapeutic benefit. However, whether myopic defocus in the O & V meridians (at baseline & with most test lenses) can offer a stop signal to myopia progression needs further investigation. Commercial Relationships: Ravi C. Bakaraju; Cathleen Fedtke, Brien Holden Vision Insititue (P), Brien Holden Vision Insititue; Jiyoon Chung, Brien Holden Vision Insititue; Darrin Falk, Brien Holden Vision Insititue (P), Brien Holden Vision Insititue; Klaus Ehrmann, Brien Holden Vision Insititue (P), Brien Holden Vision Insititue Clinical Trial: ACTRN12612000370808 Program Number: 6253 Poster Board Number: A0190 Presentation Time: 11:00 AM–12:45 PM Validity and precision of a novel instrument that combines wavefront aberrometry, autofraction and corneal topography with a stationary Scheimpflug camera Ariela Gordon-Shaag, Cyril Kahloun, Liat Gantz, David Markov, Tzadok Parnas, Tal Ben Yaacov, Rebecca Cohen Levy, Einat Shneor. Optometry and Vision Science, Hadassah Academic College, Jerusalem, Israel. Purpose: To evaluate the validity and precision of a novel instrument, the VX120 (Visionix Luneau, Chartres, France), that combines wavefront based autorefraction with stationary Scheimpflug imaging of the anterior chamber. Methods: In this prospective study, subjects were recruited from healthy first year students at Hadassah Academic College, ages 18-38. Subjects were measured 3 times with the Sirius rotating Scheimpflug camera (CSO, Italy) and the VX120, by different technicians. Subjective refraction was carried out by one qualified optometrist. The optometrist and technicians were masked to one another’s results and exams were performed in a random order. A subset of subjects was tested one week later for repeatability evaluation. Bland and Altmann analysis was used to assess agreement and precision. Only the right eye was included for analyses. Results: 61 subjects (42 women) participated in the refraction validation study (24.36±7.29 years old). The mean difference between subjective refraction and the VX120 autorefraction for sphere, spherical equivalent (SE) and astigmatic vectors J0 and J45 was, 0.14±0.47D, 0.01±0.34D, 0.10±0.18D and 0.047±0.17D, respectively. Repeatability was assessed on 37 patients. Intra-test repeatability showed small within-subjects standard deviation (Sw=0.38 to 0.39) and inter-test repeatability showed no statistically significant difference between the first and the second session for all parameters (p>0.33). 71 subjects (49 women) participated in the Scheimpflug imaging validation study (24.30±6.61 years old). The difference between the VX120 and Sirius for central corneal thickness (CCT), iridocorneal angle and anterior chamber depth (AD) were 26.12±11.46µm (-3.51±11.46µm with calibration offset), 0.93 ± 3.83° and -0.005 ± 0.118mm, respectively. Repeatability was assessed on 35 subjects. Intra-test repeatability showed small Sw of 2.85, 1.08 and 0.21 for CCT and iridocorneal angle and AD, respectively. Inter-test repeatability showed no statistically significant difference for any of the anterior chamber parameters measured by the VX120 (p>0.25). Conclusions: The VX120 shows good agreement and repeatability values. Commercial Relationships: Ariela Gordon-Shaag, Visionix (F); Cyril Kahloun; Liat Gantz, None; David Markov, None; Tzadok Parnas, Visionix; Tal Ben Yaacov, None; Rebecca Cohen Levy, None; Einat Shneor, None Support: Visionix provided compensation to subjects for travel and time as well as the VX120 for the duration of the experiment Program Number: 6254 Poster Board Number: A0191 Presentation Time: 11:00 AM–12:45 PM Intra-measurement variability of a commercially available portable autorefractor based on Shack-Hartmann sensing Kaccie Y. Li, Vickram Jain, Yaopeng Zhou. Smart Vision Labs, New York, NY. Purpose: A portable Shack-Hartmann based autorefractor (SVOne by Smart Vision Labs) has recently been shown to produce accuracy levels similar to that of other subjective and objective procedures [1]. The SVOne captures 5 consecutive images about 1 second apart and processes them individually before combining the result to produce the final measurement output. In this study, we assess the variability across the 5 single-image results that make up a particular measurement. Methods: Raw data for 25 eyes from 25 individuals were selected from a database of SVOne measurements. Each data set contains 5 Shack-Hartmann spot-pattern images of a particular eye, and the only selection criterion was that no artifacts due to corneal reflection was visible. Pupil sizes represented ranged from 2 mm to beyond 6 mm, but only a 6-mm subpupil is analyzed for pupils larger than that. An improved image processing/analysis procedure was developed for this study. The new procedure automatically isolates the spot pattern, estimates spot locations via an iteratively weighted centroiding algorithm, and matches each found spot to its appropriate reference spot in order to determine the wavefront slope. A wavefront reconstructor comprised of only low-order Zernike modes was used to reconstruct the wavefront. Refractions were calculated individually for each of the 5 images and decomposed into its spherical equivalent (M) and crossed-cylinders (J0 and J45). Results: Spherical equivalent error ranged from -5.83 D to 4.60 D in the 25 eyes analyzed, and intra-measurement variability was within 0.25 D standard error (SE) for 80 percent of the cases. For the J0 and J45 terms, intra-measurement variability was well within 0.25 D SE in all cases. In fact, variability in J0 and J45 was below 0.125 D in the majority of the cases. A myopic shift due to accommodation during the measurement may explain some of the higher variabilities seen in M. For example, the measurement with the highest variability in M (0.44 D SE) contained 4 single-image measurements all within 0.11 D of each other and an outlier that is more than 1 D off from the rest. Conclusions: Intra-measurement variability for SVOne is generally low. Higher variations were only observed in the spherical equivalent term likely due to accommodation. 1. KJ Ciufreda and M Rosenfield, OVS, 92(12): 1133-1139 These abstracts are licensed under a Creative Commons Attribution-NonCommercial-No Derivatives 4.0 International License. Go to http://iovs.arvojournals.org/ to access the versions of record. ARVO 2016 Annual Meeting Abstracts and LC logMAR VA is > ±0.05, the Rxs provide clinically equivalent VA. Paired t-tests also showed no statistically significant difference between the Rxs. Mean (±SD) differences (objective – subjective) in terms of M, J0, and J45 respectively were –0.62 D (±0.39), +0.08 D (±0.17), –0.03 D (±0.13) undilated, and –0.32 D (±0.23), +0.02 D (±0.16), –0.05 D (±0.12) dilated. The Euclidean separation of the Rxs in power vector dioptric space was 0.68 (±0.35) undilated and 0.40 (±0.21) dilated. Preferences (objective better: no difference: subjective better) were 19:0:9 for undilated eyes and 15:6:7 for dilated eyes. Conclusions: Optimizing objective refraction using VSX provided an Rx equivalent to subjective refraction in VA and generally preferred in a monocular comparison. The two Rxs were virtually identical in astigmatic components and similar in equivalent sphere. Commercial Relationships: Gareth D. Hastings, None; Jason D. Marsack, None; Han Cheng, None; Lan C. Nguyen, None; Raymond A. Applegate, University of Houston (P) Support: NIH/NEI R01EY019105 (RAA and JDM) Commercial Relationships: Kaccie Y. Li, Smart Vision Labs; Vickram Jain; Yaopeng Zhou, Smart Vision Labs Program Number: 6255 Poster Board Number: A0192 Presentation Time: 11:00 AM–12:45 PM Evaluating Objective Refraction Optimized with the Visual Strehl Image Quality Metric Gareth D. Hastings, Jason D. Marsack, Han Cheng, Lan C. Nguyen, Raymond A. Applegate. College of Optometry, University of Houston, Houston, TX. Purpose: To determine whether using the image quality (IQ) metric VSX (visual Strehl ratio) to optimize objective refraction from wavefront error (WFE) measures can provide a refraction as good as, or better than, subjective refraction while undilated or dilated. Methods: 14 subjects, mean (±SD) age 27.6 (±5.3) years participated; all had refractive error >1D (range +1.25 to –7.00D sphere and 0 to –1.25D cylinder), correctable to 20/20, and no ocular pathology. Uncorrected WFE was measured undilated; autorefraction and subjective refraction were performed. For each eye >15 000 spherocylindrical combinations (S-C Rx) (from +3 to –2D sphere in 0.25D steps, 0 to –2D cylinder in 0.25D steps, and 2° axis steps, around the second order defocus term) were converted to Zernike terms and added to the measured WFE, each generating a residual WFE. VSX was calculated for each resultant WFE and the S-C Rx with best VSX was identified as optimizing visual IQ. High (HC) and low contrast (LC) visual acuity (VA) were recorded (3 times and averaged) through each Rx. Subjects also evaluated their distance vision preference between the two Rxs on a seven-point Likert scale. Pupils were dilated with 1% tropicamide and the entire procedure repeated. Results: Both eyes of each subject were analyzed. Mean (±SD) difference (objective – subjective) in logMAR VA between the Rxs was –0.003 (±0.037) and –0.026 (±0.081) for HC and LC respectively for the undilated condition, and +0.003 (±0.070) and –0.003 (±0.088) for the dilated condition. Given test-retest variability (SD) of both HC Program Number: 6256 Poster Board Number: A0193 Presentation Time: 11:00 AM–12:45 PM Validating Optical Predictions of Sensitivity in Vertebrate Eyes Robert F. Rosencrans1, Keith Perkins1, 2, William C. Gordon1, 4, Corinne Richards-Zawacki3, Nicolas G. Bazan1, 4, Hamilton E. Farris1. 1Neuroscience Center of Excellence, Louisiana State University Health Sciences Center, New Orleans, LA; 2Southern University at New Orleans, New Orleans, LA; 3Biological Sciences, University of Pittsburgh, Pittsburgh, PA; 4Ophthalmology, Louisiana State University Health Sciences Center, New Orleans, LA. Purpose: The Land sensitivity equation (1981) has the potential to describe the optical sensitivity of any eye. In practice, however, this simple and useful tool has primarily been used to examine invertebrate optics. In the current study, using a comparative approach we empirically demonstrate that the Land sensitivity equation can predict physiologically measured sensitivity in vertebrate eyes. Methods: Electroretinograms (ERGs) were conducted in two species of diurnal frogs and two species of nocturnal frogs using the Espion Ganzfeld Dome (Diagnosys LLC). Following ERGs, optical measures were obtained. Maximal pupillary diameter (aperture) was measured using infrared photography on dark-adapted, atropinedilated animals. Following euthanasia, photoreceptor outer segment dimensions (diameter and length) were obtained from both plastic embedded semithin sections and formalin-fixed frozen sections. Focal lengths were obtained from fresh, flash-frozen retinal sections. Results: As predicted by visual ecology, ERGs in diurnal frogs required significantly higher light intensities to achieve 10% response thresholds, as compared to nocturnal frogs. Diurnal and nocturnal frogs were significantly different with respect to all optical variables with the exception of photoreceptor outer segment diameter. Nocturnal frogs exhibit 3- to 4-fold larger aperture diameters as compared to diurnal frogs. Similarly, focal lengths are between 2 and 3 times larger in nocturnal frogs. Finally, nocturnal frogs have between 1- and 2-fold longer photoreceptor outer segments. Assuming constant densities of rhodopsin across the four species, solutions to the Land sensitivity equation match sensitivity differences measured using ERGs: the equation predicts 1 order of magnitude greater sensitivity in nocturnal frogs, similar to the 1-2 orders of magnitude differences observed in ERG thresholds. Conclusions: Using the comparative method, four species from different light environments were selected. As predicted, these species differed with respect to absolute sensitivity to light. Because optical measurements predicting sensitivity are consistent These abstracts are licensed under a Creative Commons Attribution-NonCommercial-No Derivatives 4.0 International License. Go to http://iovs.arvojournals.org/ to access the versions of record. ARVO 2016 Annual Meeting Abstracts with independent physiological measures of sensitivity using ERGs, the data represent a validation of the Land sensitivity equation in vertebrates and provide insight into how visual systems may evolve in different visual ecologies. Commercial Relationships: Robert F. Rosencrans; Keith Perkins, None; William C. Gordon, None; Corinne Richards-Zawacki, None; Nicolas G. Bazan, None; Hamilton E. Farris, None Support: NIH, NIGMS grant P30 GM103340 Program Number: 6257 Poster Board Number: A0194 Presentation Time: 11:00 AM–12:45 PM Assessing the dynamic postblink changes in tear film with ageing and contact lens wear Aikaterini Moulakaki1, Irene Siso-Fuertes2, Robert Montés-Micó1, Hema Radhakrishnan2. 1Optics and Optometry and Vision Science, University of Valencia, Valencia, Spain; 2Optometry, The University of Manchester, Manchester, United Kingdom. Purpose: The stability of the precorneal tear film is essential for maintaining the optical quality of the eye between blinks. Changes in ocular aberrations over time occur, during the postblink period, in healthy individuals of all ages. However, tear film function decreases with age causing tear film irregularities in older individuals, and thus poorer image quality of the retina during the postblink period. Contact lens wear also changes the retinal image quality after a blink, as it alters the tear film characteristics and along with the drying of the contact lens surface can lead to further changes in aberrations. To understand the issue, we conducted a preliminary study on the changes in higher ocular aberrations (HOAs) during the postblink period in different age groups with and without multifocal contact lens wear. Methods: Three different age groups (i.e. group I: 18-29 years, group II: 30-39 years and group III: > 40 years) comprised of 8 eyes each group were included in this study. Eyes with astigmatism less than 1D, corrected visual acuity 20/20 (feet) or better and normal non-invasive tear break-up time (NIBUT) findings were enrolled. Total ocular aberrations were measured in the left eye sequentially for 12 seconds (sec) after a blink, using a Hartmann-Shack wavefront aberrometer (irx-3, Imagine Eyes). Dynamic tear meniscus measurements were obtained during 12 sec postblink period, employing optical coherence tomography (OCT). All measurements were performed initially without contact lenses and then with contact lenses. Results: Systematic changes in the HOAs significantly increased with time after a blink for the young subjects tested and included in group I (p< 0.002). In this group, a positive relationship was identified between time and spherical aberration (p<0.002) during the postblink period, with the curves to be coalesced 3 sec after a blink, due to tear film consolidation. The tear meniscus height and its area did not change significantly (p>0.79) over the postblink period. A large amount of inter-subject variability was also observed within any of the groups. While, contact lens wear significantly changed the aberration profile during the post blink period (p<0.05). Conclusions: The optical quality of the human eye varies considerably over time after a blink, due to changes in the tear film characteristics. Age and contact lens wear also influence on the dynamics postblink changes in aberrations. Commercial Relationships: Aikaterini Moulakaki, None; Irene Siso-Fuertes, None; Robert Montés-Micó, None; Hema Radhakrishnan, None Support: MC Grant FP7-LIFE-ITN-2013-608049 Clinical Trial: The project protocol was approved by the Senate Committee on the Ethics of Research on Human Beings of the University of Manchester., No:608049 These abstracts are licensed under a Creative Commons Attribution-NonCommercial-No Derivatives 4.0 International License. Go to http://iovs.arvojournals.org/ to access the versions of record.