TESTING & MEASUREMENT Timing Considerations Using FFT-based Measuring Receivers for EMI Compliance Measurements Jens Medler Rohde & Schwarz GmbH & Co. KG Abstract The use of FFT-based measuring receivers for EMI compliance measurements is motivated by the desire to reduce the scan time by several orders of magnitude and to gain additional insights by applying longer measurement times.Usage of an appropriate measurement time is the key to comprehensively record the disturbance characteristic of the equipment under test (EUT). Keywords EMI receiver; FFT-based measuring receiver; spectrum analyzer; EMI compliance measurement; CISPR 16-1-1; CISPR 16-2. Introduction signals. This paper is the author's continuation of the topic "Use of FFT- W ith the publication of Amendment 1 to the 3 rd Edition of CISPR 16-1-1 [1] in June 2010, FFT- based EMI test receivers for EMI compliance measurements" which was published in SAFETY & EMC in 2013[3]. based measuring receivers were introduced for EMI compliance measurements. Usage of such receivers is motivated by the desire to reduce the scan time by several orders of magnitude Concept of CISPR 16-2 without entailing any degradation of the accuracy. To achieve this The CISPR 16-2 basic standard series describes methods significant improvement, FFT based receiversmeasure spectral for measuring disturbance and immunity. For disturbance segments that are much wider than the resolution bandwidth measurements, the following parts are relevant: during the measurement time by performing parallel calculations at ■ CISPR 16-2-1 for conducted disturbance measurements multiple frequencies (Figure 1). In contrast, classic EMI receivers ■ CISPR 16-2-2 for measurements of disturbance power measure the signal within the resolution bandwidth in a given ■ CISPR 16-2-3 for radiated disturbance measurements measurement time, thereby requiring a longer scan time for the In all three parts, definition of the minimum measurement entire frequency range. and scan times for continuous disturbances represents a major requirement. In general, it is important to choose a setting that allows measurement of the maximum emission while ensuring that disturbance signals are not overlooked. To achieve this objective, CISPR 16-2 requires usage of minimum measurement times which are equivalent for stepping and FFT-based EMI receivers (see Table 1). In addition, the frequency step size must be equal to half of the resolution bandwidth (band A: 200 Hz, band B: 9 kHz, band Figure 1 FFT-based measurement versus classic stepped scan C/D: 120 kHz) or less. These requirements are also applicable to spectrum analyzers, Related specifications for the measurement methods to be used resulting in the minimum scan times shown in Table2. Note: with FFT-based measuring receivers were published in the relevant The terms sweep and scan time are used interchangeably within parts of CISPR 16-2 (see section II) . This article focuses on the CISPR 16-2. The sweep/scan time Ts is defined as the time between selection of an appropriate measurement time using an FFT-based the start and stop frequency of a sweep or scan. Tables1 and Tables 2 receiver for measuring broadband disturbance and intermittent apply to CW signals. [2] 22 SAFETY & EMC 2015 TESTING & MEASUREMENT Table 1 Minimum measurement times for CW signals acc. to CISPR 16-2-3[2] approach for this purpose. It is essential to use a sufficient number Minimum measurement time Tm Frequency band A 9~150 kHz 1 0 ms B 0.15~30 MHz 0.5 ms C and D 30~1 000 MHz 0.06 ms E 1~18 GHz 0.01 ms of sweep points and a long sweep time (> 15 s). For CISPR band B (150 kHz to 30 MHz), around 6 000 sweep points are sufficient, which corresponds to a step size of about one half of the resolution bandwidth. The thick curve in Figure 3 is a good indication that intermittent signals exist. As long as the spectrum display continues changing, there may be more intermittent signals to discover. Table 2 Minimum scan times for CW signals in the three CISPR bands with peak and quasi-peak detectors acc. to CISPR 16-2-3[2] Frequency band A Scan time Ts for 9~150 kHz Scan time Ts for quasi-peak peak detection detection 10 ms 2 820 s=47 min B 0.15~30 MHz 0.5 ms 5 970 s=99.5 min=1 h 39 min C and D 30~1 000 MHz 0.06 ms 19 400 s=323.3 min=5 h 23 min Depending on the type of disturbance, the measurement time Tm or the scan time Ts may have to be increased - even for quasi-peak measurements. In extreme cases, the measurement time Tm at a certain frequency may have to be increased to 15 s if the level of the observed emission is not steady. Figure 3 Frequency sweep in spectrum analyzer mode; sweep time > 15 around 6 000 sweep points; EUT: touch dimmer of halogen lamp The measurement time used in stepping and FFT-based Next, select a frequency of interest and switch to Zero Span measuring receivers and the sweep time used in spectrum analyzers mode to analyze the signal in the time domain. The example in are the key to ensure that disturbance signals are not missed during Figure 4 shows an interval between pulses of 20 ms. automatic scans or sweeps over frequency spans. Therefore, a timing analysis of the EUT disturbance characteristic has to be performed before executing automated measurements (Figuer 2). Figure 2 Analysis steps for automated or semi-automated measurements Figure 4 Zero span measurement in spectrum analyzer mode shows a pulse interval of 20 ms; EUT: touch dimmer of halogen lamp. For automated measurements, two cases must be considered: To demonstrate such timing analysis, a dimmer of a halogen lamp was measured using the spectrum analyzer's Zero Span single and multiple scans or sweeps. Single scans or sweeps mode. As an alternative, an oscilloscope may be connected to the For a single scan or sweep, the measurement time at each receiver's IF output. First, an overview measurement was performed frequency must be larger than the intervals between pulses for with the peak detector to find critical frequencies for the timing intermittent signals. If the measurement or sweep time is too short, analysis. Performing multiple sweeps with maximum hold is a good erroneous measurement results will be obtained. SAFETY & EMC 2015 23 TESTING & MEASUREMENT Multiple scans or sweeps For performing multiple sweeps with maximum hold, the observation time at each frequency needs to be sufficient to steady. The following equation can be used to calculate the minimum sweep time for multiple sweeps: intercept intermittent signals, i.e. the observation time has to be Ts min=2×(∆f/Bres2) selected according to the pulse repetition interval of the disturbance where signals (see timing analysis above). If the sweep timeper sweep Ts min is the min. sweep time for multiple sweeps is too fast, the probabilityof intercept per sweep is very low and ∆f is the frequency span erroneous measurement results will be obtained (see Figure 5). Bres is the resolution bandwidth (1) Increasing the sweep time per sweep gives a higher probability of intercept (see Figure 6). Timing Considerations using FFTbased Measuring Instruments Today's FFT-based measuring receivers are limited by the currently available analog-to-digital converters(ADC), e.g. for a 1 GHz measuring receiver an ADC is necessary with 2 GS/s sampling rate to meet the Nyquist criterion. Furthermore, preselection filters and high resolution ADCs (e.g. 16 bits for 30 MHz FFT bandwidth) are necessary to meet the CISPR 16-1-1 overload requirements for quasi-peak detection. Therefore, today's FFT-based receivers cannot Figure 5 Visualization of intercepts – low probability of intercept per sweep using very high sweep rate sample and evaluate the complete CISPR bands C/D (30 MHz to 1 000 MHz) and E (1 GHz to 18 GHz) in one shot. Instead, they combine parallel calculations at N frequencies with a stepped scan. For this purpose, the frequency range of interest is subdivided into several segments that are measured sequentially (see Figure 7). Figure 6 Visualization of intercepts – high probability of intercept per sweep using lower sweep rate. In all cases, the fastest possible sweep time per sweep is the multiplicative inverse of the resolution bandwidth based on a step size of 0.5×RBW, e.g. 194 ms for RBW=100 kHz over the frequency range 30 MHz to 1 000 MHz.This fast sweep must be applied continuously with maximum hold for a period of time equal to or greater than the time that would have been spent for Figure 7 FFT scan in sequence;source: CISPR 16-2-3[2] sweeping in line with Table 2, e.g. 0.97 s for peak detection of CW signals corresponding to 5 sweeps of 194 ms each for the frequency The scan time Tscan is calculated as: range 30 MHz to 1 000 MHz. However, depending on the type of Tscan= TmNseg disturbance, the sweep time Ts may have to be increased, e.g. to 15 where scorresponding to 77 sweeps of 194 ms each for the frequency range Tscan Tm is the measurement time for each segment, and 30 MHz to 1 000 MHz if the level of the observed emission is not Nseg is the number of segments. 24 SAFETY & EMC 2015 (2) TESTING & MEASUREMENT FFT-based measuring receivers have to meet the general requirements for the minimum measurement time and step size as defined in CISPR 16-2 (see section II). Therefore, the measurement time Tm must be selected longer than the pulse repetition interval for correct measurement of a broadband spectrum. If the measurement time is too short, enormous measurement errors can result. In a worst case scenario, the FFT-based measuring receiver may not capture the disturbance signal at all. This is problematic if the segment size has a large width, e.g. 30 MHz or more. Fast Sweep versus Fast Time Domain Scan For demonstration purposes, the same EUT was used: a touch dimmer of a halogen lamp. The frequency span was also reduced to 5 MHz for better visualization. First, a single fast sweep without sweep time adjustment(sweep time SWT = 20 ms) was performed. It is clear that the result does not match the spectrum envelope at all (Figure 8). Figure 8 Spectrum analyzer – A single sweep without sweep time adjustment does not matchthe spectrum envelope at all Figure 9 Spectrum analyzer – A single sweep with sweep time adjustment based on timing analysis of the EUT (a touch dimmer of a halogen lamp). 1 s×20 ms pulse interval×1 000 sweep points). However, the result in Figure 10 shows that critical frequencies are still missed! For this EUT, a SWT=2 s for 20 sweeps gives a good match with the envelope. Figure 10 Spectrum analyzer – multiple sweep (20x) with sweep time adjustment based on timing analysis of the EUT. Next, a single sweep with sweep time adjustment based on the Finally, a timedomain scan with measurement time adjustment above timing analysis was performed; this results in a sweep time based on the above timing analysis was performed. An FFT based of 20 s (20 ms pulse interval×1 000 sweep points). The result in measuring receiver with 30 MHz segment size can measure CISPR Figure 9 exhibits a good match with the envelope. band B (150 kHz to 30 MHz) in one shot. According to Equation (1), If the pulse repetition interval is unknown, many users perform this results in a pure scan time of 20 ms; in reality, however, a high- multiple sweeps with fast sweep times using the maximum hold performance receiver can generate the result in about 100 ms. The function to determine the spectrum envelope. For low repetition achieved curve is a full match with the envelope (see Figure 11). An impulsive signals, many sweeps are necessary to fill up the spectrum overall comparison of the different approaches is shown in Table 3. envelope of the broadband component. The correct sweep time can also be determined by increasing it until the difference between More Speed with the Time Domain Scan maximum hold and clear/write displays is below 2 dB, for example. For demonstration purposes, a multiple sweep with sweep time FFT-based measuring receivers can attain measurement speeds adjustment based on the above timing analysis was performed; this a few thousand times faster than can be achieved in conventional, results in a total observation time of 20 s for 20 sweeps (20×SWT = stepped frequency scan mode. As a consequence, frequency scans SAFETY & EMC 2015 25 TESTING & MEASUREMENT Figure 13 Measurement speed of timedomain scan versus stepped frequency scan with the R&S®ESR EMI test receiver Figure 11 FFT-based receiver – time domain measurement with measurement time adjustment based on timing analysis of the EUT Conclusions Table 3 Match with envelope versus scan time Single sweep Pulse repetion Measurement RBW=9 kHz time to get frequency interception 50 Hz 20 ms Match with envelope Span=5 MHz Single sweep RBW=9 kHz Span=5 MHz 20x Multiple 20x Multiple Time domain single sweep single sweep scan receivers can be used for EMI compliance measurements in RBW=9 kHz RBW=9 kHz RBW=9 kHz, Span=5 MHz Span=5 MHz 0.15~5 MHz SWT=1 s SWT=2 s Step=2.25 kHz SWP=1 000 SWP=1 000 MT=20 ms SWT=20 ms SWT=20 s SWP=1 000 SWP=1 000 0.020 s 20 s 20 s 40 s 0.1 s no good not all good good FFT-based measuring accordance with Amendment 1 to the 3rd Edition of CISPR 161-1. The use of FFT-based measuring receiversis motivated by the desire to reduce the in the CISPR bands using the peak detector can be performed scan time by several orders of magnitude without loss of accuracy in only a few milliseconds, and even with the quasi-peak and while gaining deeper insight by applying longer measurement average detector it only takes seconds, thereby rendering preview times.Another benefit is that the disturbance spectrum can be measurements with the peak detector obsolete (see Figure 12). directly measured with the final detector, making measurement of fluctuating disturbances more reliable.For precise and reproducible measurements, the use of an appropriate measurement or sweep time based on timing analysis of the EUT's disturbance characteristic is essential. Figure 12 Analysis steps for automated or semi-automated measurements with an FFT-based measuring receiver References [1] Amendment 1:2010-06 to CISPR 16-1-1:2010-01 (Edition 3) Specification for radio disturbance andimmunity measuring apparatus Rohde & Schwarz has introduced a new generation of FFT- and methods - Part 1-1: Radio disturbance and immunity measuring based EMI test receivers for disturbance measurements compliant with CISPR 16: the R&S®ESR[4]. Its FFT-based time domain scan apparatus - Measuring apparatus. [2] Amendment 1:2010-06 to CISPR 16-2-3:2010-04 (Edition 3) can deliver measurement speeds up to 6 000 times faster than can Specification for radio disturbance andimmunity measuring apparatus be achieved with the traditional single-channel filtering approach and methods - Part 2-3: Methods of measurement of disturbances (see Figure 13). and immunity - Radiated disturbance measurements. The ultra-fast measurement speed is particularly useful if the [3] MEDLER, J., More speed, more insight - Use of FFT-based EMI test equipment under test can be operated only during a short period of receivers for EMI compliance measurements, SAFETY & EMC 2013, time, e.g. a starter motor in cars. Part of the time savings can also 38-42. be exploited to apply longer measurement times in order to reliably detect narrowband intermittent signals or isolated pulses. 26 SAFETY & EMC 2015 [4] ® R&S ESR EMI Test Receiver, Product Brochure Version 02.00, www. rohde-schwarz. com