7.73 1.15 SENSOR BODY 1" NPT CONVERTIBLE STYLE SENSOR

advertisement
8
6
7
1" NPT
5
4
1
REVISION
pH/ORP ELECTRODE
(GLASS, PLATINUM,
OR GOLD)
SENSOR BODY
2
3
DESCRIPTION
REV
.91
APPROVED
D
D
1.15
1.38
7.73
CONVERTIBLE STYLE SENSOR
pH/ORP ELECTRODE
(GLASS, PLATINUM,
OR GOLD)
.91
C
C
1.38
1.15
INSERTION STYLE SENSOR
pH/ORP ELECTRODE
(GLASS, PLATINUM,
OR GOLD)
2.50
1.15
B
SANITARY STYLE SENSOR
B
1.89
NOTE:
1.
DIMENSIONS ARE IN INCHES
NOTICE--HACH COMPANY CLAIMS PROPIETARY RIGHTS IN THE INFORMATION DISCLOSED ON THIS DRAWING. IT IS ISSUED
IN CONFIDENCE FOR ENGINEERING INFORMATION ONLY AND MAY NOT, IN WHOLE OR IN PART, BE USED TO MANFACTURE
ANYTHING, WHETHER OR NOT SHOWN HEREON, REPRODUCED OR DISCLOSED TO ANYONE WITHOUT DIRECT PERMISSION
FROM HACH COMPANY.
UNLESS OTHERWISE SPECIFIED:
MATERIAL
Material <not specified>
A
DIMENSIONS ARE IN INCHES
TOLERANCES:
.X
= .03
.XX
= .01
.XXX
= .005
ANGLES = .25
NAME
DRAWN
5600 LINDBERGH DR.
LOVELAND, CO. 80539
ENGINEER
THIRD ANGLE PROJECTION TITLE:
INTERPRET GEOMETRIC
TOLERANCING PER:
ANSI Y14.5M 1994
SIZE DWG. NO.
B
7
6
5
4
3
differential_pHandORP
SCALE: 1:2 WEIGHT: 0.360
DO NOT SCALE DRAWING
8
HACH COMPANY
DATE
2
REV
SHEET 1 OF 10
1
A
8
5
6
7
4
2
3
1
NOTE:
1.
DIMENSIONS ARE IN INCHES
C
C
11.40
10.90
7.69
8.59
7.42
CLEARANCE REQUIRED
FOR PROBE REMOVAL
D
CLEARANCE REQUIRED
FOR PROBE REMOVAL
D
B
B
8.33
1.85
1.40
1" NPT SCHEDULE 80
CPVC TEE
1" NPT SS316 TEE
3.81
A
A
3.00
MH334N4NZ (CPVC TEE)
FLOW THROUGH MOUNT
MH314N4MZ (SS316 TEE)
FLOW THROUGH MOUNT
SIZE DWG. NO.
B
REV
differential_pHandORP
SHEET 2 OF 10
SCALE: 1:2 WEIGHT:
8
7
6
5
4
3
2
1
8
7
5
6
4
2
3
1
NOTE:
1.
DIMENSIONS ARE IN INCHES
C
C
CLEARANCE REQUIRED
FOR PROBE REMOVAL
D
CLEARANCE REQUIRED
FOR PROBE REMOVAL
D
13.85
pHD sensor
13.85
pHD sensor
7.22
7.75
B
10.10
9.57
B
8.47
9.00
2.72
2.72
A
A
1-1/2" CPVC TEE
3.88
4.94
MH536N4NZ (CPVC TEE)
UNION MOUNT
8
7
6
SIZE DWG. NO.
B
MH516N9Z (SS316 TEE)
UNION MOUNT
5
4
3
REV
differential_pHandORP
SHEET 3 OF 10
SCALE: 1:2 WEIGHT:
2
1
8
5
6
7
4
3
2
1
NOTE:
1.
DIMENSIONS ARE IN INCHES
D
D
NYLON STRAIN RELIEF
PVC FITTING
PVC JUNCTION BOX
C
C
47.00
4.87
1/2 CPVC SHEDULE 80 PIPE
B
B
57.00
60.60
CLEARANCE REQUIRED
FOR PROBE REMOVAL
IMMERSION MOUNT - STANDARD
A
A
SIZE DWG. NO.
B
differential_pHandORP
SHEET 4 OF 10
SCALE: 1:2 WEIGHT:
8
7
6
5
4
3
2
REV
1
8
6
7
5
4
2
3
1
NOTE:
1.
DIMENSIONS ARE IN INCHES
D
D
1 1/2" SCHEDULE 80 CPVC PIPE
5.20
91.60
CENTER OF HAND RAIL TO
CENTER OF POLE
POLE & PROBE SWIVEL INTO
PROCESS OR ONTO SERVICE
SUPPORT BRACKET
CPVC SUPPORT
FOR SERVICING
PROBE
C
CPVC PROBE
MOUNTING ADAPTER
SCHEDULE 80
1 1/2" NPT TO 1" NPT
93.10
CLEARANCE REQUIRED FOR
DIGITAL PROBE REMOVAL
PROBE CABLE EXISTS THRU
BUNA S RUBBER GROMMET
PVC END CAP
C
CONVERTIBLE
STYLE SENSOR
STAINLESS STEEL
BRACKET & U-BOLT
PHENOLIC KNOB (2X)
CUSTOMER HAND RAIL REF.
B
B
PIPE IS ADJUSTABLE IN
CLAMP FOR FLUID LEVEL
5.20
REF
IMMERSION MOUNT - HANDRAIL
A
A
STAINLESS STEEL BRACKET & U-BOLTS
SIZE DWG. NO.
CPVC SWIVEL & CLAMP
B
REV
differential_pHandORP
SHEET 5 OF 10
SCALE: 1:2 WEIGHT:
8
7
6
5
4
3
2
1
8
7
6
5
4
3
2
1
NOTE:
1.
DIMENSIONS ARE IN INCHES
MAXIMUM RATED PRESSURE: 50 PSI (3.5 BAR); MAXIMUM RATED TEMPERATURE: 122 F (50 C)
2.
3-WAY BRASS AIR ASSIST
VALVE WITH FEMALE 1/4" NPT
2 PLACES
D
NYLON STRAIN RELIEF
WITH NEOPRENE BUSHING
D
1-1/2" NPT
CLOSE NIPPLE
4.50 MAX.
SENSOR INSERTION
DEPTH
25.00 MAX.
INSERTED SENSOR
C
C
35.50 MAX.
EXTRACTED SENSOR
TO CLOSE BALL VALVE
56.00 MIN.
CLEARANCE REQUIRED TO REMOVE
PROBE FROM OR INSTALL PROBE IN DEVICE
B
B
1.40
INSERTION MOUNT MH736M4MZ (CPVC)
A
A
SIZE DWG. NO.
B
REV
differential_pHandORP
SHEET 6 OF 10
SCALE: 1:4 WEIGHT:
8
7
6
5
4
3
2
1
8
7
5
6
NOTE:
1.
DIMENSIONS ARE IN INCHES
4
3
2
1
3-WAY BRASS AIR ASSIST
VALVE WITH FEMALE 1/4" NPT
2 PLACES
D
1-1/2" NPT
CLOSE NIPPLE
D
3.51LOCKING
COLLAR
1-1/2" STAINLESS STEEL
BALL VALVE
5.91
VALVE HANDLE PROTRUDES
PAST VALVE BODY
WHEN IN THE CLOSED POSITION
NYLON STRAIN RELIEF
WITH NEOPRENE BUSHING
C
C
3.20
SQUARE
25.00 MAX.
INSERTED SENSOR
35.50 MAX.
EXTRACTED SENSOR
TO CLOSE BALL VALVE
56.00 MIN.
4.50 MAX.
SENSOR INSERTION
DEPTH
CLEARANCE REQUIRED TO REMOVE
PROBE FROM OR INSTALL PROBE IN DEVICE
B
B
2.25
INSERTION MOUNT MH716M4MZ (SS316)
A
A
SIZE DWG. NO.
B
differential_pHandORP
SHEET 7 OF 10
SCALE: 1:4 WEIGHT:
8
7
6
5
4
3
2
REV
1
8
5
6
7
4
3
2
1
NOTES:
1.
MATERIAL:
TEE - 316 STAINLESS STEEL
CLAMP - 304 STAINLESS STEEL
D
2.
NO GASKETS PROVIDED WITH THIS KIT.
PROBE GASKET COMES WITH PROBE.
3.
CLAMP & TEE CONFORM TO PROVISIONS OF
3-A SANITARY STANDARDS.
4.
DIMENSIONS ARE IN INCHES
D
2.00
CLEARANCE REQUIRED
FOR SENSOR REMOVAL
C
C
HEAVY DUTY CLAMP
8.99
2" SANITARY SST TEE
B
B
2.75
2.52
3.50
7.00
SANITARY MOUNT MH018S8SZ (SS316 TEE)
A
A
SIZE DWG. NO.
B
differential_pHandORP
SHEET 8 OF 10
SCALE: 1:4 WEIGHT:
8
7
6
5
4
3
2
REV
1
8
7
6
5
4
3
2
1
Specifications
pH Sensors
D
ORP (Redox) Sensors
Wetted Materials
PEEK® or Ryton® body, salt bridge of matching material with
Kynar® (PVDF) junction, glass process electrode, titanium ground
electrode, and Viton® O-ring seals
Wetted Materials
PEEK® or Ryton® body, salt bridge of matching material with Kynar® (PVDF)
junction, glass supported platinum (or gold) process electrode, titanium
ground electrode, and Viton® O-ring seals
(pH sensor with optional HF-resistant glass process electrode
has 316 stainless steel ground electrode, and perfluoroelastomer
wetted O-rings; consult factory for other available wetted O-ring materials)
Operating Temperature Range
23 to 203ºF (-5 to +95ºC)
Operating Temperature Range
23 to 203ºF (-5 to +95ºC)
Pressure/Temperature Limits
(Without mounting hardware)
100 psi at 221ºF (6.9 bar at 105ºC)
Pressure/Temperature Limits
(Without mounting hardware)
100 psi at 221ºF (6.9 bar at 105ºC)
C
Maximum Flow Rate
10 ft. (3m) per second
Maximum Flow Rate
10 ft. (3m) per second
C
Built-in Temperature Element
NTC 300 ohm thermistor for analyzer temperature readout only-no
automatic temperature compensation necessary for ORP measurement
Built-in Temperature Element
NTC 300 ohm thermistor for automatic temperature compensation
and analyzer temperature readout
Measuring Range
-1500 to +1500 mV
Measuring Range
0-14 pH
Sensitivity
Less than 0.5 mV
Sensitivity
Less than 0.005 pH
Stability
2mV per 24 hours, non-cumulative
Stability
0.03 pH per 24 hours, non-cumulative
B
D
Maximum Transmission Distance
3000 ft. (914 m)
Maximum Transmission Distance
3000 ft. (914 m)
Sensor Cable (integral)
5 conductor (plus two isolated shields) cable with XLPE (cross-linked polyethylene)
jacket; rated to 302ºF (150ºC); 15ft. (4.5 m) standard length
B
Sensor Cable (integral)
5 conductor (plus two isolated shields) cable with XLPE (cross-linked polyethylene)
jacket; rated to 302ºF (150ºC); 15ft. (4.5 m) standard length
A
A
SIZE DWG. NO.
B
differential_pHandORP
SHEET 9 OF 10
SCALE: 1:4 WEIGHT:
8
7
6
5
4
3
2
REV
1
8
7
6
5
4
3
2
1
Engineering Specifications
Ryton Sensor
PEEK Sensor
D
1. The pH or ORP sensor shall be of Differential Electrode Technique design using two
D
electrodes to compare the process value to a stable internal reference standard buffer solution.
The standard electrode shall have non-flowing and fouling-resistant characteristics.
1. The pH or ORP sensor shall be of Differential Electrode Technique design
using two electrodes to compare the process value to a stable internal
reference standard buffer solution. The standard electrode shall have
non-flowing and fouling-resistant characteristics.
2. The sensor shall have a hex shaped body to facilitate mounting, and shall
be constructed of PEEK material for exceptional chemical resistance and
mechanical strength. This material shall enable the sensor to be installed in
metal fittings without leakage usually caused by heating and cooling cycles
when dissimilar materials are threaded together.
3. The sensor shall have a:
C
a) Convertible body style featuring 1-inch NPT threads on both ends to mount
into a standard 1-inch pipe tee, into a GLI adapter pipe for union mounting
with a standard 1-1/2 inch tee, or onto the end of a pipe for immersion into a vessel.
b) Insertion body style featuring 1-inch NPT threads only on the cable end to mount
into a GLI ball valve hardware assembly, enabling the sensor to be inserted into or
retracted from the process without stopping the process flow.
c) Sanitary body style featuring an integral 2-inch flange to mount into a GLI 2-inch
sanitary tee. The sanitary body style sensor shall include a special cap and EDPM
compound gasket for use with GLI sanitary hardware.
4. The built-in electronics of the sensor shall be completely encapsulated for protection
from moisture and humidity.
2. The sensor shall have a hex shaped body to facilitate mounting, and shall be constructed
of Ryton® material for exceptional chemical resistance and mechanical strength.
This material shall enable the sensor to be installed in metal fittings without leakage usually
caused by heating and cooling cycles when dissimilar materials are threaded together.
3. The sensor shall have a convertible body style featuring 1-inch NPT threads on both ends
to mount into a standard 1-inch pipe tee, into a GLI adapter pipe for union mounting
with a standard 1-1/2 inch tee, or onto the end of a pipe for immersion into a vessel.
4. The built-in electronics of the sensor shall be completely encapsulated for protection
from moisture and humidity.
5. The sensor shall have a built-in preamplifier to enable the signal to be transmitted
up to 3000 ft. (914 m) with standard cabling.
C
6. The sensor signal shall have an integral temperature sensor to automatically compensate
measured values for changes in process temperature.
7. The sensor shall include a titanium ground electrode (standard) to eliminate ground
loop currents in the measuring electrodes.
8. The sensor shall be Hach Company GLI Model PD1R1 for pH measurement or GLI Model
RD1R5 for ORP measurement.
5. The sensor shall have a built-in preamplifier to enable the signal to be transmitted up
to 3000 ft. (914 m) with standard cabling.
B
B
6. The sensor signal shall have an integral temperature sensor to automatically
compensate measured values for changes in process temperature.
7. The sensor shall include a titanium ground electrode (standard) to eliminate
ground loop currents in the measuring electrodes.
8. The sensor shall be Hach Company GLI Model PDXP-series for pH measurement
or GLI Model RDXP-series for ORP measurement.
A
A
SIZE DWG. NO.
B
SCALE: 1:2 WEIGHT:
8
7
6
5
4
3
2
REV
differential_pHandORP
SHEET 10 OF 10
1
Download