082-03-DEDE Cercon brain GB 07.04.2003 10:13 Uhr Seite 1 Cercon® brain Instructions for Use Cercon smart ceramics – The CAM-Aided All-Ceramic System 082-03-DEDE Cercon brain GB 07.04.2003 10:13 Uhr Seite 3 Table of Contents 1 Working with the Cercon brain ................................................... 5 1.1 Scope of Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 1.2 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 1.3 For Your Personal Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 1.4 General User Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 1.4.1 Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 1.4.2 Description of Device Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 1.4.2.1 Digitisation Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 1.4.2.2 Milling Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 1.4.2.3 Rotating and Turning Uniit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 1.4.2.4 START Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 1.4.2.5 STOP Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 1.4.2.6 MENU Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 2 Start-up Procedure ................................................................. 7 2.1 Cercon brain Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 2.2 Electricity Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 2.3 Getting Ready for Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 3 Instructions for Use ................................................................ 8 3.1 Operating the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 4 Safety Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Care and Maintenance 6 Error Messages 9 ............................................................. 9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 7 EC Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 3 082-03-DEDE Cercon brain GB 07.04.2003 10:13 Uhr Seite 5 GB er onbrain ® 1 Working with the Cercon brain of the device may be hazardous for users or third parties and lead to damage of the device and other material assets. 1.1 Scope of Delivery • The Cercon brain should only be operated with original 1 cercon brain device with Cercon base blanks, sizes 12, 30, 38 (5 of each) Modelling jigs, sizes 12, 30, 38 (2 of each) parts. Only in this way can the quoted performance data and required operational safety be guaranteed. • Do not open the device. The inside of the device con- Milling jigs, sizes 12, 30, 38 (1 of each) tains hazardous voltages. Only service technicians Coarse and fine cutters (2 of each) authorised by DeguDent may carry out repair and main- 1 set of positioning jigs (sizes 12, 30, 38) tenance work. Hex socket screwdrivers, small and large (1 of each) • Protect the device from dampness to prevent short- 1 set of grub screws circuiting and damage. Do not expose the device to 1 scanning powder water vapour. 2 mains cables (1 USA version) • Remove the plug from the device during long periods of disuse. • Follow the safety tips for handling milling dust. • Avoid uncontrolled contact with the milling cutters when the protective cover is open or the cutters are being changed. Uncontrolled contact with the milling cutters must also be avoided when the milling jig is being changed. The sharp milling cutters may cause an injury. • Avoid direct eye contact with the laser beam and do not manipulate the cover of the laser device. There is a risk 1.2 Technical Data Power Supply: 110 – 230 V~, 50/60 Hz of damage to the eyes. • Never operate the Cercon brain when the protective Maximum Wattage: 750 W cover is open. The working rotating and turning unit or Dimensions: 850mm (W) x 520mm (H) x 620mm (D) rotating instruments may cause an injury (risk of bruising). Weight: Approx. 120 kg 1.4 General User Information The device is for interior use only. The room temperature must be between 15 °C and 25 °C. The device’s compo- 1.4.1 Intended Use nents interact best at a constant room temperature of Cercon brain is a machine for manufacturing crowns and 20 °C ± 2 °C. Mains supply voltage should not fluctuate bridge frameworks from presintered zirconia by milling. For more than ±10 % from the nominal value. this, a standard dental wax carving is measured visually and subsequently milled from Cercon base zirconia blanks. 1.3 For Your Personal Safety The design and control of the machine are precisely tailored to Cercon base. Please read these Instructions for Use carefully before installing and starting the device. You must follow the information contained in these instructions. Please read the Safety Tips in Section 4 of these instructions. Improper use 5 082-03-DEDE Cercon brain GB 07.04.2003 10:13 Uhr Seite 6 1.4.2 Description of Device Components 2. Software Version No.: Version: e.g. 110401 1.4.2.1 Digitisation Unit MC Laser: e.g. 249 The digitisation unit comprises a Laser Class II visual measuring system with maximum power of 1 mW. The rotating 3. Units Left to Mill until Next Cutter Change: and turning unit is used to hold the model. Protect the unit Remaining for Coarse Cutter, e.g. 99 from more severe vibrations, because these may adversely Remaining for Fine Cutter, e.g. 99 influence the precise adjustment of the mechanical components. 4. Manual Cutter Change Tool Change Press Start 1.4.2.2 Milling Unit The milling unit comprises two cutting spindles for holding the coarse and fine cutters. The cutters must be replaced after 100 units have been milled. You will be given a reminder in the display. 1.4.2.3 Rotating and Turning Unit The rotating and turning unit is used to hold the modelling Pressing the START key activates the barcode reader. The cutter barcode is read and the cutters changed by loosening the cutter retention screw (threaded pin on the side of the cutting spindles). When using the cutters, please ensure that the cutter retention screw on the cutter meets the flat surface. While fastening the retention screw, push the cutter until it makes contact with the cutting spindle. and blank jigs. The unit guides the carving over the scanner and the blank over the cutters according to the calculation of the milling process. Avoid uncontrolled contact with the milling cutters while changing the cutters. The sharp milling cutters may cause an injury. 1.4.2.4 START Key With the START key, you initiate the program you have selected. 1.4.2.5 STOP Key With the STOP key, you can cancel the current operation. The fine cutter (this has a blue cover when removed from the package) must be placed in the top spindle, while the coarse cutter (this has a red cover when removed from the package) must be placed in the bottom spindle. It is crucial to place them this way, because otherwise the cutting processes will not work properly. After pressing this key, you will be asked whether you wish to cancel the procedure. If you do indeed wish to do so, return to the Start menu by pressing this key again. The procedure is thus cancelled. Pressing the STOP key followed by the START key resumes 5. Manually Switch On Suction to Clean Device: You can power up the vacuum cleaner even when the Cercon brain is not running; e.g. when you want to collect dust residue with a brush and remove this through the vacuum slits. the interrupted procedure. 1.4.2.6 MENU Key The following operations can be brought up in turn by pressing the MENU key: 1. Device No.: 6 Vacuum Cleaner START On/Off Pressing the START key switches the vacuum cleaner on and off. 6. Query Device Operating Time Serial No. Running Time e.g. 0001-6100123 e.g. 01:20:59 082-03-DEDE Cercon brain GB 07.04.2003 10:13 Uhr Seite 7 7. Select Language Select Language START German You can choose from one of the following languages after pressing the START key: German English Spanish 2.2 Electricity Supply Connect the mains cable via a wall socket with a fused circuit. This circuit should only supply the Cercon brain. Do not use an extension cable. After connecting the mains cable, hit the red key switch to turn on the device. This key switch is an emergency shutdown switch that stops all the device’s operations in the event of danger. French Italian The Cercon clean operation is guided via a control cable that should be connected at the rear of the Cercon brain. The Cercon clean’s power supply is independent of the Cercon brain’s supply and is also generated via the mains 2 Start-up Procedure cable and a wall socket with fused circuit. In this case also, you should not use an extension cable. 2.1 Cercon brain Installation 2.3 Getting Ready for Operation The device must be installed on a level and solid surface (standing base) at least 10 cm away from the wall or other objects. After opening the protective cover, insert the milling cutters in the cutting spindles (see instructions for manual cutter changing on Page 6). It is not possible to start the device Cercon brain should be positioned horizontally and with without cutters. unrestricted air circulation. Protect the device from the direct rays of the sun or other strong light sources. The cutter barcode should be read by the barcode reader. During this, the barcode of the particular cutter is shown During operation, the movements of the rotating and turning unit create energy that is converted into vibrations. on the display. The barcode reading is acknowledged by the light going out and a sound being emitted. These vibrations are largely absorbed by the Cercon brain’s cast mineral base. Please use the standing base to additionally ensure that residual vibrations do not spread further. The quality of the scanning and milling result may be adversely impaired. Even if the device’s own weight keeps it stable, there is a risk in principle that it could slide off its standing base. Please secure the device by e.g. indentations in the Avoid uncontrolled contact with the milling cutters while inserting the cutters. The sharp milling cutters may cause an injury. The Cercon clean vacuum cleaner unit, which comprises a Filter Class H vacuum cleaner, should be connected to the Cercon brain with a control cable and suction hose. See the description in Subsection 2.2. standing base, a bolted-on bracket or some other appropriate measure. For safety reasons, the device’s cover should be locked at zero current. In other words, the cover can only be opened about 1 minute after the device has been switched on. 7 082-03-DEDE Cercon brain GB 07.04.2003 10:13 Uhr Seite 8 3 Instructions for Use The blank already clamped in the blank jig is held in front of the reader so that the blank barcode can be read in. 3.1 Operating the Device Hitting the key switch, turning the key and pressing the button powers up the device. After a 30-minute stand-by period in which the device’s electronic components stab- The following appears on the display: “Insert Mat. Press Start” ilise, the Cercon brain is ready for operation. Once the device’s computer has been booted, the device name appears in the display. The blank jig with blank is clamped to the right side of the rotating and turning unit, repeating the same steps used for the modelling jig, and the cover closed. Pressing the After a short time, the following appears on the display: “Reference Point Search Press Start” START key begins the working process, and the door is locked. The following appears on the display: “Digitisation 1 After the START key is pressed, the reference point search Time Bar” is carried out. The device is then ready for operation and the cover can be opened. After digitisation of the basal side has finished, the modelThe measuring system is initialised during the reference point search. The rotating and turning unit moves into its ling and blank jigs are automatically reversed and the milling calculation performed. three maximum spatial positions. The following appears on the display: The following appears on the display: “Menu = Info Start = Work” “Milling Calculation Time Bar” Once the milling calculation has finished, the digitisation The modelling jig with wax model is clamped to the left process starts for the occlusal side. side of the rotating and turning unit. The following appears on the display: For this, the retention screw (threaded pin) is opened with “Digitisation 2 the hex socket screwdriver and the modelling jig inserted Time Bar” until it makes contact with the axis of the rotating and turning unit. The retention screw (threaded pin) is now made fast against the jig’s parallel pin. After digitisation has finished, the modelling and blank jigs are automatically reversed and the milling calculation per- Pressing the START key activates the barcode reader. formed. Now the fully automatic milling process begins. The following appears on the display: The following appears on the display: “Read Barcode “Milling Insert Mat.” Calculation Time Bar” 8 082-03-DEDE Cercon brain GB 07.04.2003 10:13 Uhr Seite 9 The basal and occlusal sides of the dental object are Health Hazard! roughened on the bottom cutting spindle. The suction unit is switched off when milling starts. • Due to the general TLV (threshold limit value) for zirconia of 1mg/m3, please ensure that local The following appears on the display: “Rough Machining suction (in this case the Cercon clean’s proper operation) is used during machining. Time Bar” • Please also make sure that you change the dust filter bag. The moment of change is shown on the Rough machining is followed by fine machining of the Cercon brain’s display. basal and occlusal sides. In addition, please note the following: The following appears on the display: • Do not let the product get in your eyes. “Fine Machining • Avoid contact with the skin and mucous Time Bar” membranes. • Wash your hands and rub cream into them after working. The cover must be opened after the milling process. The modelling and blank jigs are removed. The blank jig is opened and the finished milled framework can be removed. • Do not eat or drink while working. • Do not swallow the product. • Do not breathe in abrasive dust. The following appears on the display: • When grinding, work with a dust collection device “Menu = Info and wear mouth and face protection. Start = Work” Pressing the START key activates the barcode reader and 5 Care and Maintenance the next job can be performed. Regularly inspect the device for mechanical damage. In this way, you will prevent accidents and breakdowns. 4 Safety Tips Remove contaminants and powder residue immediately. Injury Hazard! • The milling cutters are pointed and sharp. Please do not touch the tools. This is the only way to guarantee proper operation. Strictly adhere to the requirements for changing the cutters and vacuum cleaner bag. • The rotating and turning unit may cause bruising. Please do not touch the rotating and turning unit while it is working. • Never operate the device when the protective cover is open. 9 082-03-DEDE Cercon brain GB 07.04.2003 10:13 Uhr Seite 10 6 Error Messages Message Reason, Action to Take Machine Error Switch Off General machine error. Machine must be switched off. Suction Off The suction unit is off and must be switched on. Suction Full Change Bag Filter bag is full and must be changed. Cover Open Cover was opened improperly during machining. Device must be restarted. Profile Error Machine profile cannot be read. X, Y, Z Axis Error The appropriate axis cannot be moved. Device must be restarted. Limit Switch Activated The appropriate axis cannot be moved. Device must be restarted. Barcode Reading Interrupted The STOP key was pressed while the barcode was being read or the barcode is unreadable. Barcode Read Error Barcode is unreadable. Material Already Machined Invalid barcode, or the barcode has already been used. Only use original barcodes. Laser Error Start Measurement Digitisation error. Device must be restarted. Coarse Cutter (+ Display “No.”) Only x units could be milled before the coarse cutter was changed. Fine Cutter (+ Display “No.”) Only x units could be milled before the fine cutter was changed. 10 082-03-DEDE Cercon brain GB 07.04.2003 10:13 Uhr Seite 11 7 EC Declaration of Conformity EC Declaration of Conformity Pursuant to EC Machinery Directive 98/37/EC We hereby declare that the product Cercon® brain Device for Manufacturing Zirconia Frameworks conforms to the essential requirements of the following EC directives: 1. EC Machinery Directive 98/37/EC 2. EC Directive for Electrical Equipment Designed for Use Within Certain Voltage Limits (Low Voltage Directive) 73/23/EEC 3. EC Electromagnetic Compatibility Directive 89/336/EEC The following harmonised standards were applied: EN 292-1:1991; EN 292-2:1991+A1:1995; EN 60204-1:1998; EN 61010-1:1996+A2:1996; EN 50081-1:1993; EN 50082-1:1997; EN 60825-1:1997 Dr Alexander Völcker Head of Research and Development Torsten Schwafert Head of Dental Engineering Division 11 50535156/0,5/0303/GVD 082-03-DEDE Cercon brain GB 07.04.2003 10:13 Uhr Seite 12 For more information: www.cercon-smart-ceramics.com