EMx40
SAS au Capital de 2 158 244 € - 444 871 933 R.C.S. Bourges - APE : 2651B
Headquarter : 9, rue Isaac Newton - 18000 Bourges - France
EMx40
1 st
This manual is issued specifically for the data processing racks of the TA3840 or EMx40 system.
It contains instructions detailed for functioning, operation and cleaning of this system.
To use it in an optimal way, we advise you TO READ CAREFULLY THESE
INSTRUCTIONS and to respect them throughout the life of the equipment.
Keep this manual to hand so that you can refer to it at any time. Ensure that it is complete and kept close to the equipment.
The racks TA3840C and TA3840S, as well as the TA3840R repeater module are integrated in the system for processing the tankers liquid cargo. Maintenance must be carried out by a qualified operator in communicating electronic systems.
We also draw your attention to the fact that the connection of equipments or the use of products other than those recommended by Honeywell Marine may present risks for which we will not be liable.
This manual must not be reproduced in any form whatsoever without the prior written approval of
Honeywell Marine who cannot be held responsible for any use of the information contained in this manual.
As we want you to take advantage of the most of the latest technology and new equipment, as well as to benefit from our experience, our equipments may undergo technical or design changes. As a result, some of the features and information in this manual may change without prior notice and without any obligation to up-date it.
Pictures of this document are not contractual.
Should you encounter any problems or have any questions about your TA3840 or EMx40 system, please do not hesitate to contact your nearest Honeywell Marine customer service.
The installation of EMx40 system transmitters is described in the MI5016E installation manual.
The maintenance of EMx40 system transmitters is described in the MM5016E maintenance manual.
The description and operation of the racks TA3840C/R and TA3840S for measuring data collection are described in the MT5008E technical manual.
SAFETY PRECAUTIONS:
Current regulations and legislation applicable to hazardous areas must imperatively be known and followed by personnel responsible for commissioning and operating an intrinsically safe equipment.
Take care to switch the power off before proceeding to any disconnection or removal of the transmitters.
WARNING:
Our equipments are designed and manufactured in accordance with local Our equipments are designed and manufactured in accordance with local safety regulations, and in particular European directives relative to reconciling member states' legislation:
● 89/336/EEC and 2004/108/EC "Electromagnetic compatibility",
● ATEX 94/9/EC "Equipment and protective systems intended for use in potentially explosive atmospheres",
● 96/98/EC "Marine equipment".
The EM540, EM940 radar and associated transmitters are certified for use in hazardous area according to the intrinsic safety protection type. Examination certificates are supplied in appendix.
They are intended for professional use and must be installed, used and maintained by competent staff who are qualified in this type of equipment.
In particular we wish to draw your attention to the fact that we cannot be held responsible if:
● Any technical alterations are made to our appliances without our written authorization,
● Our equipments are damaged by being operated in conditions other than the intended usage of their technical classification (power supply, temperature, environment, etc.).
The safety instructions given in this manual are merely given for guidance purposes to protect you and all those using and working on our equipments.
Honeywell Marine cannot foresee all dangerous situations that might arise.
This is why the owner and/or the operator is responsible for the operating safety of the system.
Regulations of the ship classification society may impose procedures (health and safety, fire prevention, handling of hazardous substances, etc.) which are stricter than those given in this manual. In this case, the regulations must be followed.
EM540
EM940
EC TYPE EXAMINATION CERTIFICATE, SEE APPENDIX
A
IDENTIFICATION NAMEPLATE
II 1(1) GD
Ex ia [ia Ga] IIB or IIC T4 Ga
LCIE 05 ATEX 6087X
IECEx LCIE 12.0009X
MARKING
S/N
P/N
TAG
IDENTIFICATION NAMEPLATE
II 1(1) GD
Ex ia [ia Ga] IIB or IIC T4 Ga
LCIE 05 ATEX 6087X
IECEx LCIE 12.0009X
MARKING
S/N
P/N
TAG
EC TYPE EXAMINATION CERTIFICATE, SEE APPENDIX B
IDENTIFICATION NAMEPLATE
II 1 GD
Ex ia IIC T6 Ga
Ex ia IIIC T80°C Da IP66/67
LCIE 03 ATEX 6246X
IECEx LCI 11.0039X
MARKING
S/N
P/N
TAG
IDENTIFICATION NAMEPLATE
II 1 GD
Ex ia IIC T6 Ga
Ex ia IIIC T80°C Da IP66/67
LCIE 03 ATEX 6246X
IECEx LCI 11.0039X
MARKING
S/N
P/N
TAG
EC TYPE EXAMINATION CERTIFICATE, SEE APPENDIX
C
T901-P 01TA
IDENTIFICATION NAMEPLATE
II 1 GD
Ex ia IIC T6 Ga
Ex ia IIIC T80°C Da IP66/67
LCIE 07 ATEX 6022X
IECEx LCI 11.0038X
MARKING
S/N
P/N
TAG
EC TYPE EXAMINATION CERTIFICATE, SEE APPENDIX D or
LOG3840
IDENTIFICATION NAMEPLATE
II 1 G D
Ex ia IIC T6 Ga
Ex ia IIIC T80°C Da IP66/67
LCIE 07 ATEX 6024X
IECEx LCI 11.0037X
MARKING
S/N
P/N
TAG
Model for Inert Gas pressure measurement, or other pressure for EM940 (TP901-P 01TA) ........... 13
Monitoring computer
TA3840R
24 Vdc + RS232 /
RS485 / RS422
+24 Vdc power supply
3 dry contact outputs:
- data alarm,
- system fault,
- power supply failure.
Monitoring computer when fitted with CPU board, as option
TA3840S
230 / 115 Vac
50 / 60 Hz
Power supply
RS232 /
RS485 /
RS422
TA3840C
Loading computer
RS232 / RS485
Special applications
Configuration software
Up to 96 4-20 mA inputs or On / Off inputs or On / Off outputs
2 additional analog transmitters or
Up to 63 other EMx40 radars or
EM940 EM540
Doc No: MT5016E – Revision 6 – ENG
LOG3840
(optional)
T901-P
Up to 3 temperature sensors and
1 inert gas pressure sensor
Transmitters
EMx40 Technical Manual
Pressure sensor
8
The EMx40 level radar transmitters, T901-P temperature/pressure transmitters and
LOG3840 deck digital indicator are parts of the EMx40 system, dedicated to monitor the tankers liquid cargo parameters such as level, temperature and pressure, as well as all associated measurements. They are installed in hazardous area.
The EMx40 radar level transmitters measure the liquid level inside the tank and transmit the information to the TA3840C communication unit by means of the
TA3840S safety unit located in cargo control room. Each EMx40 radar is fitted with entries for temperature/pressure transmitter and the EM940 radar with additional analog transmitters.
The T901-P temperature / pressure transmitters measure the liquid cargo temperature and the inert gas pressure. Each T901-P transmitter is directly connected to EMx40 radar.
The LOG3840 deck digital indicator displays locally the level, temperatures, pressure measured by transmitters. It is powered by the TA3840S safety unit and communicates on the same bus than the radar.
General synopsis
The general synopsis of the system is shown on the previous page.
The EMx40 radar is a stand-alone radar type level transmitter using a frequency modulated continuous wave (FMCW), electromagnetic signal to measure distance between its antenna and cargo tank content surface (ullage).
The EMx40 radar performs the following functions:
● Measures the liquid level (in mm) by processing the microwave signal thanks to its dedicated signal processor,
● Concentrates the measurements of:
Liquid temperature coming from three sensors at different levels (top, middle and bottom of the tank),
IG pressure coming from a pressure sensor,
Up to two other analog measurements for EM940 radar
● Transmits the data to the cargo control room.
The EMx40 radar is fastened on the deck, on top of a socket by means of 4 or 6 bolts. No part of the radar is in actual contact with the liquid in the tank and the antenna is the only part exposed to the tank atmosphere.
The EMx40 radar comprises:
● A 316L stainless steel circular housing protecting the electronics,
●
A FLAT-T technology antenna covered and protected against tank atmosphere by
PTFE radome,
●
A terminal box fitted with three packing-glands for electrical cables.
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 9
EM540 radar
4
1
5
EM940 radar
8 1
2
6
1 Terminal box cover
2 Three connections for cable input
3 FLAT-T technology antenna
4 316L stainless steel housing
5 Holes for M20 screws to fasten on the socket
6 Pivot hinge
2 7 3
3
5 4 6
1 Terminal box cover
2 Three connections for cable input
3 FLAT-T technology antenna
4 316L stainless steel housing
5 Holes for M20 screws to fasten on the socket
6 Pivot hinge
7 Hole for pressure sensor
8 Pressure pipe
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 10
The T901-P temperature/pressure transmitter is located on deck, on top of each tank, on a socket.
The T901-P temperature/pressure transmitter measures:
● The liquid temperature in the cargo tank at up to three different levels (high, middle and low),
● The inert gas pressure inside the cargo tank.
The T901-P temperature/pressure transmitter comprises:
● 1 to 3 platinium RTD 100
at 0 °C, located in a protecting tube immersed in the cargo tank,
● An absolute or gauge pressure sensor and its associated electronics, located inside transmitter housing,
● A 316L stainless steel housing protecting the electronics, fitted with one packing-gland for electrical cable,
● A cover for access to the terminal block.
There are several types of T901-P transmitter:
● An universal model:
With 1, 2 or 3 temperature sensor(s),
With or without pressure transmitter,
● A model for installation on pump body, with 1, 2 or 3 temperature sensor(s).
● A model for Inert Gas pressure measurement, or other pressure for EM940
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 11
Identification
1 4
2
A Absolute pressure sensor
R Gauge pressure sensor
TA Absolute pressure for Inert Gas, or other pressure for EM940
0 Without pressure transmitter
1 With pressure transmitter
0 Without temperature sensor
1 With 1 temperature sensor
2 With 2 temperature sensors
3 With 3 temperature sensors
1 Temperature sensors
2 316L stainless steel housing
3 Removable cover
4 Cable connection
5 Nuts and washers to fasten on the socket
5 5
1
Model for installation on pump body (i. e. TP901-P30F)
2 3
4
5
1 Temperature sensors
2 316L stainless steel housing
3 Removable cover
4 Cable connection
5 Screw location to fasten on the pump body
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 12
2
3
1
1 316L stainless steel housing
2 Terminal box cover
3 Cable connection
4 Flange
4
The LOG3840 deck indicator is located on deck. It operates in spy mode, reading and displaying data picked up on the communication line between the EMx40 radar
(connected to the T901-P transmitter) and the TA3840S safety unit.
The LOG3840 deck indicator calculates and displays on the LCD screen:
● The liquid level in the cargo tank expressed in innage or ullage,
● The liquid spot or average temperature in °C or
°F,
● The inert gas pressure in mbar.
3
2
2 3 4
4
1
1 Cable connection
2 LCD screen
3 Casing
4 Magnet switch
1
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 13
The LOG3840 deck indicator comprises:
● A stainless steel or polyester casing,
● A LCD screen for data display, visible through a glass window,
● Packing-glands for electrical cables,
● A magnet switch to select the displayed temperature.
Measuring range
Resolution
Accuracy
0.6 m to 40 m
0.1 mm
Instrument: +/- 3 mm
Temperature inputs
Other transmitter input, for inert gas pressure
Ambient operating temperature
Operating pressure
Protection index
Materials
Dimensions
Process connection
Typical electrical connections
Packing-glands
Weight
Operating : +/-5 mm
3 for Pt 100
at 0 °C RTD (IEC751)
Range: -165 °C to +250 °C
Signal transmission accuracy: ± 0.22 °C
3 analog input 0.5 to 2.5 Vdc
Signal transmission accuracy: ± 0.3 % FS
-35 °C to +70 °C
-200 to +1000 mbar
IP 66 / IP 67
316L stainless steel and PTFE
340 x h. 101 mm for 8" antenna
Flange ISO PN16 DN200 for 8" antenna
Two marine cables BV2 and one marine cable BV3 for connection to the TA3840S safety unit (2 pairs + common shield) and to the T901-P temperature/pressure transmitter (up to 5 pairs + common shield), or three marine cables BV4 for same connections
BV2 marine packing-glands for cable diameter min. 7 mm, max.
14.5 mm.
BV3 marine packing-gland for cable diameter min. 8.5 mm, max. 19 mm.
BV4 packing-glands for cable diameter min. 16.5 mm, max.
22.6 mm.
13 kg
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 14
SAFETY PROTECTION:
Precautions for intrinsically safe installation Ex ia:
The EM540 radar level transmitter is certified II 1 (1) G Ex ia [ia Ga] IIC or IIB T4
Ga under certificate n° LCIE 05 ATEX 6087X.
The EM540 radar must only be connected to TA3840S safety unit and T901-P temperature/pressure transmitter.
The electrical safety parameters are as follow (refer to certificate in Appendix A for
EM540)
EM540 - Class IIB
Terminals
Power supply
(J2: 1-2)
RS485 circuit
(J2: 3-4)
Temperature sensor circuit
(J6 and J7: 14-15-16,
14-17-18, 14-19-20)
Pressure sensor circuit
(J3, J4 and J5: 5-6-7, 8-
9-10, 11-12-13)
Ui
(V)
Ii (mA) Pi
(W)
Ci (µF) Li
(mH)
≤ 15 ≤ 425 ≤ 3.25 ≤ 1.1 ≈ 0
≤ 15 ≤ 100 ≤ 0.59
≤
0.001
≈ 0
Uo
(V)
/
≤ 6.2
Io
(mA)
/
≤ 139
Po
(mW)
/
≤ 215
Co
(µF)
/
≤ 790
Lo
(mH)
/
≤ 7.3
/
/
/
/
/
/
/
/
/
/
≤ 6.2
≤ 6.2
≤ 44
≤ 42
≤ 68
≤ 65
≤ 790
≤ 790
≤ 73
≤ 80
EM540 - Class IIC
Terminals
Power supply
(J2: 1-2)
RS485 circuit
(J2: 3-4)
Temperature sensor circuit
(J6 and J7: 14-15-16,
14-17-18, 14-19-20)
Pressure sensor circuit
(J3, J4 and J5: 5-6-7, 8-
9-10, 11-12-13)
Ui
(V)
Ii (mA) Pi
(W)
Ci (µF) Li
(mH)
≤ 12 ≤ 425 ≤ 3.25 ≤ 1.1
≤ 12 ≤ 100 ≤ 0.59
≤
0.001
≈ 0
≈ 0
Uo
(V)
/
≤ 6.2
Io
(mA)
/
≤ 139
Po
(mW)
/
≤ 215
/
/
/
/
/
/
/
/
/
/
≤ 6.2 ≤ 44
≤ 6.2 ≤ 42
≤ 68
≤ 65
Co
(µF)
/
≤ 34
Lo
(mH)
/
≤ 1.7
≤ 34
≤ 34
≤ 18
≤ 20
As the EM540 radar (and associated T901-P transmitter) can be installed in zone 0, a system certificate is available under number LCIE 04 ATEX 6189 X, specifying in particular the cables required characteristics:
Electrical characteristics of the cable between the power supply rack and the radar
Radar
EM540 class IIB
EM540 class IIC
Safety unit
TA840-I
TA3840S
TA840-I
TA3840S
L (µH)
≤ 100
≤ 270
≤ 22
≤ 22
L/R (µh/
)
≤ 24
≤ 44
≤ 14.2
≤ 14.2
C (µF)
≤ 7.9
≤ 3.08
≤ 3.8
≤ 3.8
The TA840-I safety unit is from a previous generation TA840 system. It is mentioned here in case of EM540 radars installed are spare on such a system.
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 15
Measuring range
Resolution
Accuracy
0.6 m to 40 m
0.1 mm
Instrument: +/- 3 mm
Temperature inputs
IG pressure transmitter
Other transmitters inputs
Ambient operating temperature
Operating pressure
Protection index
Materials
Dimensions
Process connection
Typical electrical connections
Packing-glands
Weight
Operating : +/-5 mm
3 for Pt 100
at 0 °C RTD (IEC751)
Range: -165 °C to +250 °C
Signal transmission accuracy: ± 0.22 °C
Ceramic capacitive sensor type
Standard range: 900 mbar to 1300 mbar or 800 to 2000 absolute mbar;
-70 to +350 gauge mbar
Output: 0.5 to 2.5 Vdc
Accuracy:
0.3 % FS
2 analog inputs 0.5 to 2.5 Vdc
Signal transmission accuracy: ± 0.3 % FS
-35 °C to +70 °C
-200 to +1000 mbar
IP 66 / IP 67
316L stainless steel and PTFE
395 x h. 150 mm for 8" antenna
Flange ISO PN10 DN250
Two marine cables BV2 and one marine cable BV3 for connection to the TA3840S safety unit (2 pairs + common shield) and to the T901-P temperature/pressure transmitter (up to 5 pairs + common shield), or three marine cables BV4 for same connections
BV2 packing-glands for cable diameter min. 7 mm, max. 14.5 mm.
BV3 packing-glands for cable diameter min. 8.5 mm, max. 19 mm.
BV4 packing-glands for cable diameter min. 16.5 mm, max.
22.6 mm.
16 kg
SAFETY PROTECTION:
Precautions for intrinsically safe installation Ex ia:
The EM940 radar level transmitter is certified II 1 (1) G Ex ia [ia Ga] IIB or IIC T4
Ga under certificate N° LCIE 05 ATEX 6087X.
The EM940 radar must only be connected to TA3840S safety unit and T901-P temperature/pressure transmitter.
The electrical safety parameters are as follow (refer to certificate in Appendix A).
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 16
Class IIB
Terminals
Power supply
(J2: 1-2)
RS485 circuit
(J2: 3-4)
Temperature sensor circuit
(J6 and J7: 14-15-16,
14-17-18, 14-19-20)
Pressure sensor circuit
(J3, J4 and J5: 5-6-7, 8-
9-10, 11-12-13)
Ui
(V)
Ii (mA) Pi
(W)
Ci (µF) Li
(mH)
≤ 15 ≤ 425 ≤ 3.25 ≤ 1.1
≤ 15 ≤ 100 ≤ 0.59
≤
0.001
≈ 0
≈ 0
Uo
(V)
/
≤ 6.2
Io
(mA)
/
≤ 139
Po
(mW)
/
≤ 215
Co
(µF)
/
≤ 790
Lo
(mH)
/
≤ 7.3
/
/
/
/
/
/
/
/
/
/
≤ 6.2
≤ 6.2
≤ 44
≤ 42
≤ 68
≤ 65
≤ 790
≤ 790
≤ 73
≤ 80
Class IIC
Terminals
Power supply
(J2: 1-2)
RS485 circuit
(J2: 3-4)
Temperature sensor circuit
(J6 and J7: 14-15-16,
14-17-18, 14-19-20)
Pressure sensor circuit
(J3, J4 and J5: 5-6-7, 8-
9-10, 11-12-13)
Ui
(V)
Ii (mA) Pi
(W)
Ci (µF) Li
(mH)
≤ 12 ≤ 425 ≤ 3.25 ≤ 1.1
≤ 12 ≤ 100 ≤ 0.59
≤
0.001
≈ 0
≈ 0
Uo
(V)
/
≤ 6.2
Io
(mA)
/
≤ 139
Po
(mW)
/
≤ 215
/
/
/
/
/
/
/
/
/
/
≤ 6.2 ≤ 44
≤ 6.2 ≤ 42
≤ 68
≤ 65
Co
(µF)
/
≤ 34
≤ 34
≤ 34
Lo
(mH)
/
≤ 1.7
≤ 18
≤ 20
As the EM940 radar (and associated T901-P transmitter) can be installed in zone 0, a system certificate is available under number LCIE 04 ATEX 6189 X, specifying in particular the cables required characteristics:
Electrical characteristics of the cable between the power supply rack and the radar
Radar
EM940 class IIB
EM940 class IIC
Safety unit
TA840-I
TA3840S
TA840-I
TA3840S
L (µH)
≤ 100
≤ 270
≤ 22
≤ 22
L/R (µh/
)
≤ 24
≤ 44
≤ 14.2
≤ 14.2
C (µF)
≤ 7.9
≤ 3.08
≤ 3.8
≤ 3.8
The TA840-I safety unit is from a previous generation TA840 system. It is mentioned here in case of EM940 radars installed are spare on such a system.
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 17
Temperature sensors
IG pressure transmitter
Up to 3 Pt 100
at 0 °C RTD (IEC751) class B in 316L stainless steel sheath
Ceramic capacitive sensor type
Standard range: 900 mbar to 1300 mbar or 800 to 2000 absolute mbar;
-100 to +600 gauge mbar
Output: 0.5 to 2.5 Vdc
Accuracy:
0.3 % FS
-35 °C to +70 °C Ambient operating temperature
Protection index
Material
Dimensions
Process connection
Electrical connection
Packing-glands
Weight
IP 66 / IP 67
316L stainless steel
Universal model :
220 x h. 103 mm
Pump body model :
100 x h. 70 mm
Pressure measurement model :
115 or 150 x h. 115 mm
Universal model : ISO PN20 DN50
Pump body model : provided for installation on pump body
Pressure measurement model : ISO PN40 DN25 or PN16
DN40
One marine cable BV3 (up to 5 pairs + common shield) for connection to the radar level transmitter, or BV2 on pressure measurement model
BV2 packing-glands for cable diameter min. 7 mm, max.
14.5 mm.
BV3 packing-glands for cable diameter min. 8.5 mm, max.
19 mm.
5 kg
SAFETY PROTECTION:
Precautions for intrinsically safe installation Ex ia:
The T901-P temperature/pressure transmitter is certified II 1 G D Ex ia IIC T6 Ga and under certificate n° LCIE 03 ATEX 6246X.
The T901-P temperature/pressure transmitter must only be connected to EMx40 radar.
The electrical safety parameters are as follow (refer to certificate in Appendix B).
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 18
Class IIC
Terminals
Temperature output
0.5 - 2.5 Vdc output
U (Vdc)
≤ 15
≤ 8
I (A)
≤ 0.050
≤ 0.55
Icc (A)
/
/
Ci (µF)
≤ 0.037
≤ 4
Li (mH)
≈ 0
≈ 0
As the T901-P temperature/pressure transmitter can be installed in zone 0, specifying in particular the cables required characteristics: refer to the table hereunder.
Electrical characteristics of the cable between the radar and the T901-P temperature/pressure transmitter
Radar
EM540 class IIB, IIC
EM940 class IIB, IIC
L (mH)
≤ 18
≤ 18
C (µF)
≤ 30
≤ 30
Operating temperature
Protection index
Materials
Power supply
Dimensions
Process connection
Electrical connection
Cable entries
-35 °C to +70 °C
IP 66 / IP 67
316L stainless steel and glass resistant to salty marine atmosphere
Powered by TA3840S safety unit
L x H x P (mm) 200 x 200 x 80
Flange ISO PN6 DN32
Marine cable (2 pairs + common shield) connected to the
TA3840S safety unit
3 BV2 marine packing-glands for cable diameter min. 7 mm, max. 14.5 mm; one is plugged
Operating temperature
Protection index
Materials
Power supply
Dimensions
Process connection
Electrical connection
Cable entries
-35 °C to +70 °C
IP 66 / IP 67
Black polyester glass fibber and glass resistant to salty marine atmosphere
Powered by TA3840S safety unit
L x H x P (mm) 130 x 130 x 75
Fixed on the rear face
Marine cable (2 pairs + common shield) connected to the
TA3840S safety unit
2 PG16 packing-glands for cable diameter min. 10 mm, max. 14 mm
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 19
SAFETY PROTECTION:
Precautions for intrinsically safe installation Ex ia:
The LOG3840 deck digital indicator is certified II 1 G D Ex IIC T6 Ga under certificate n° LCIE 06 ATEX 6024 X.
The LOG3840 deck digital indicator must only be connected to TA3840S safety unit and EMx40 radar.
The electrical safety parameters are as follow (refer to certificate in Appendix D).
Class IIC
Terminals
Power supply
J3 : 1/3/2/4
Power supply
J3 : 5/7/6 /8
Backlight
J4 : 1 - 2
Ui
(Vdc)
Iip (A) Iicc
(A)
≤ 14.4 ≤ 0.53 ≤ 2.2
≤ 14.4 ≤ 0.53 ≤ 2.2
Ci (µF) Li
(mH)
≈ 3.7
≤
0.276
≈
0.007
≤
0.007
≤ 14.4 ≤ 0.53 ≤ 2.2 ≈ 0 ≈ 0
Uo
(Vdc)
Io (A)
≤ 7,2 ≤ 0.21
Iocc
(A)
-
Co
(µF)
Lo
(mH)
≤ 9.8 ≤ 0.010
≤ 14.4 ≤ 0.53 ≤ 2.2 ≤ 9.8 ≤ 0.010
- - - - -
As the LOG3840 deck digital indicator can be installed in zone 0, specifying in particular the cables required characteristics.
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 20
This module is powered as soon as the TA3840S safety unit is on. The screen lights on and data are displayed, depending on the operating parameters set up.
The LCD screen indicates the following information:
● 3 main lines for the data,
● 4 spots for five messages (data and communication errors).
Line 1
5 digits + decimal dot for level display
Spot 1 communication working
(RM) or communication error (CE)
Line 2
Line 3
4 digits + decimal dot for temperature display
4 digits + decimal dot adjustable for pressure display
Note: if a decimal dot is set up, the mbar unit disappears.
Spot 2
Spot 3 level data error (LE) temperature data error
(TE)
Spot 4 pressure data error (PE)
The deck digital indicator includes a component (A) activated by an external magnet through the window, for selection of the data displayed on the second line.
The display mode for level (ullage or innage), the unit used for level and temperature, and the pressure decimal dot are set up by means of the utility software.
Only the different temperatures display, step by step, is possible.
● Touch the switch (A) with the magnet to display successively the average temperature and the temperature measured by each sensor, according to the data availability depending to the current T901-P transmitter type.
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 21
Default step: average temperature,
Step 2: upper temperature display,
Step 3: middle temperature display,
Step 4: lower temperature display.
Back to default step.
The average temperature is calculated according to the temperature recorded by each sensor immersed in the liquid cargo. When the liquid level is under the lowest sensor, its temperature is displayed.
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 22
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 23
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 24
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 25
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 26
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 27
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 28
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 29
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 30
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 31
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 32
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 33
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 34
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 35
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 36
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 37
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 38
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 39
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 40
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 41
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 42
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 43
Doc No: MT5016E – Revision 6 – ENG
Transmitters
EMx40 Technical Manual 44
Honeywell Marine SAS
9, Rue Isaac Newton
18000 Bourges
France
Tel + 33 (0) 2 48 23 79 01
Fax + 33 (0) 2 48 23 79 03
E-mail: contact.marine@honeywell.com www.honeywellprocess.com
MT5016E-rev06-ENG
February 2013
© 2013 Honeywell International Inc.