Applied Ergonomics 32 (2001) 485–499 Polarized task lighting to reduce reflective glare in open-plan office cubicles D.A. Japuntich* Office Supplies Division, Office Ergonomics and Cleaning Systems, 3M Company, 3M Center, Building 230-2S-13, St. Paul, MN 55144-1000, USA Received 25 February 2000; accepted 13 March 2001 Abstract This ergonomic study deals with the common situation where a glossy document is placed between a viewer and an under-shelf task light source in a common open-plan office cubicle workstation. With a task lamp in front, when looking at a document a viewer sees two images, the document itself and specular glare, which is the reflected image of the light source. Specular glare or veiling reflection causes eye discomfort, makes it difficult to read a document and has been thought to contribute to eyestrain. This paper analyzes the application of polarized lighting for this specific situation. The use of a linear polarized light source helps to minimize specular glare by darkening the reflected image of the light source on the document. The performance and predictive optimization of the use of polarized lighting in this situation is investigated according to female and male viewer height demographics. Theoretical predictions and light measurement analysis of specular glare reduction are compared with empirical results from testing on a panel of humans on semi-gloss finish and matte finish papers. This study shows that with the right alteration of a polarized light source position, specular glare may be significantly reduced, and correlations exist between the theory, empirical measurements and the human response to specular glare reduction. r 2001 Elsevier Science Ltd. All rights reserved. Keywords: Task light; Polarized light; Glare; Eyestrain; Optometry; Architecture 1. Introduction This study deals with the common situation where a glossy document is placed between a viewer and an under-shelf light source in a constrained open-plan office cubicle workstation environment. Under-shelf and desktop task lights are commonly used in open-office plan workstations to augment or even replace inadequate over-head lighting. With a task lamp in front, a problem exists when looking at the document in that a viewer sees two images, the printed image and specular glare, the reflected image of the light source. Having a close, bright source of luminance like a reflection in the field of view negatively affects eye comfort through elevated luminance ratios. The use of a linearly polarized light source may help to minimize specular glare by darkening the reflected image of the light source on the document. The ability of a polarized light source to alter the luminance or brightness of a reflected lamp image on a *Tel.: +1-651-736-9041; fax: +1-651-736-7924. E-mail addresses: dajapuntich@mmm.com (D.A. Japuntich). document is greatly affected by the viewing angle of the observer. Studies by Allphin (1963), Crouch and Kaufman (1963), Lumsden(1976), Rea et al. (1985), and Rea et al. (1990) on task reading or viewing angles have shown a range between 01 and 401 with an average of around 251, elevating above 351 depending on the height of the viewer. The measured location of lamp images on office tasks has been studied by Siminovitch (1993), and worker glare comfort surveys were taken by Bernecker et al. (1993a, b) and by North (1991). Discomfort from reflective glare at a desk is one of the most common complaints in offices, and North (1991) found a gender difference in complaints: 32.7% for females, as compared to 21.3% for males. The degree of glossiness of a surface affects reflective glare, as does the material that the surface is made of. Glossy surfaces are dielectric materials, which include plastics, glass, waxes, transparencies, glossy paper coatings and even the toner image on photocopies. Many researchers have focused on the amount of visual performance improvement using a polarized light source to alter contrast when the specular reflection of 0003-6870/01/$ - see front matter r 2001 Elsevier Science Ltd. All rights reserved. PII: S 0 0 0 3 - 6 8 7 0 ( 0 1 ) 0 0 0 2 0 - 5 486 D.A. Japuntich / Applied Ergonomics 32 (2001) 485–499 the lamp image is superimposed over a reading task. Edwin Land (Grabau, 1940) invented the first polarized lamp to reduce veiling glare, and Marks (1959) presented an excellent explanation of contrast and polarization. Blackwell (1963, 1984), Blackwell and DiLauria, 1973 and Bernecker et al. (1993a, b) extensively studied glare and polarized light, pioneering quantitative methodologies for optimizing reading contrast resulting in higher visual performance at lower illumination, giving possible energy use reductions. Rea (1981) investigated combinations of visual task characteristics, orientation and polarization. Further polarized lighting investigators, Jones (1992) and Karpen (1995), have focused on energy savings and the use of full-spectrum fluorescent lamps with radial polarization. Boyce and Rea (1994) found little improvement in visual performance and a slight improvement in color sorting tasks using full-spectrum polarized lighting from overhead luminaires with incident light percent degree of polarization of less than 15%, measured at angles from the vertical of 401 or less. Clear and Mistrick (1996) wrote a general review of the state of the research on polarized lighting up to that time. An objective of the use of polarized lighting to reduce visual discomfort was lost in the debate of the difficulty of proving visual performance. In the discussion section of Clear and Mistrick (1996), P. Boyce asked the question, ‘‘if multi-layer polarizers are to be treated as special purpose devices, under what conditions are they useful?’’ This is an ‘‘ergonomic question’’, and it is hoped that this paper will provide some answers to that question for an application of polarized lighting for under-shelf lighting in open-plan office cubicle workstations. Certainly, at the minimum the reduction of specular glare will result in a better, more comfortable workplace. The IES RP-24 (1989) Recommended Practice for Lighting Offices Containing Computer Visual Display Terminals states, ‘‘The logical place for fixed local task lighting is under a cabinet or shelf and directly over the work station. These locations are often in the offending zone, thus producing veiling reflections at some locations on the work surface.’’ The great increase in open-plan office cubicles has created a familiar situation where uncomfortable specular glare from under-shelf lighting is always present on a document or desk surface. This is a two-dimensional situation where linear polarized light may be used to reduce glare if the proper optimized design specifications may be shown. Previous work has been done with radial polarizers (Karpen, 1995; Boyce and Rea, 1994). These not only give a wide-spread and multi-directional polarization, but also give a lower degree of polarization. The application shown in this paper uses a multi-layer linear polarizer, which in this constrained, two-dimensional use gives a very high degree of polarization of the light in the direction of the viewer. One simple, inexpensive method to see if linear polarized light is indicated to reduce specular glare is to sit at a workstation viewing the reflective glare while rotating a lens from polarized sunglasses. If the luminance from glare is altered in intensity by rotating the lens, a polarized light source may help to reduce it. If there is no change then polarized lighting is not indicated. This study shows a method of determining the location of the offending lamp reflection zone on an office desk according to ergonomic gender demographics using viewer height, lamp height and viewerlamp distance. Theoretical predictions and empirical measurements of the level of specular glare reduction using linear polarized lighting are presented, including optimization based on the location of the reflection. Further testing using human subjects was done to show their agreement with the theory and empirical measurements of specular glare reduction optimization. 2. Theoretical predictions of the location of a light source image on the surface in front of a viewer The location of the ‘‘offending zone’’ or the reflective image of the lamp on a document needs to be determined for each of a population of office workers. The problem solved in this section is the location of the lamp reflected image in the plane of vision of seated male and female workers for their wide range of height demographics. As shown in Fig. 1, lamp reflected image distance (X) is a function of the height of the viewer (h), the height of the light (Y) and the horizontal distance (L) of the viewer from the light. Viewing angle (i, equal to the incident angle) is the angle from the vertical to the line of view and is a function of the horizontal distance of the image from the lamp and the height of the lamp. If the height of the viewer increases, the lamp image moves towards the lamp, and if the height of the lamp increases, the lamp image moves away from the lamp. Solving simultaneous equations gives the Eqs. (1) and (2) for the lamp image distance from the light, X, and subsequently the viewing angle, i. In the very special case where the light and viewer heights are identical, the lamp image distance is one-half the viewer distance. The image distance from the viewer is L@X. The image distance (D) from the desk edge is the difference of W, the light distance from the desk edge, and X. X¼ YL hþY X L i ¼ arctan ¼ arctan : Y hþY ð1Þ ð2Þ D.A. Japuntich / Applied Ergonomics 32 (2001) 485–499 487 Fig. 1. Dimensions relating to forward illumination of a desk surface. Table 1 The effect of gender height demographics on image viewing angle and specular glare image distance from the front edge of a desk Seated eye height, h, above 737.6 mm desk surface (mm) Females 5th percentile Females 50th percentile Females 95th percentile Males 5th percentile Males 50th percentile Males 95th percentile 315.4 391.4 471.4 409.4 490.4 569.4 Image viewing angle, i (Deg) 36.3 33.6 31.2 33.1 30.6 28.6 Image distance from front desk edge, D (mm) 232 260 285 266 290 310 It is important to determine where the image of a fluorescent lamp falls in the range of viewing for a common reading task in an office cubicle. As an average viewing position, it is assumed that the eye of the viewer is directly above the front desk edge, as in Siminovitch (1993). A common cubicle arrangement is a lamp height (Y) of 410 mm (16 in), a lamp horizontal distance (L) of the viewer from the lamp of 533 mm (21 in), with the viewer’s eye directly above the front desk edge. The seated eye height of the viewer above a 736.6 mm (29 in) high desk surface (h) is population based for females and males as derived from the anthropometric data in Gordon et al. (1988). Eye height was calculated as the combination of popliteal height and eye height sitting to obtain the distance above the floor, and then subtracting a desk height, as shown in Table 1. It should be noted that the 50th percentile female height of 391.4 mm is below the common lamp height of 410 mm, while the male 5th percentile height of 409.4 mm is equal to this lamp height. For non-recessed luminaires, 95% of the males are not looking directly into the light source, while more than 50% of the females are exposed to direct glare from the light source. Fig. 2. The location of fluorescent lamp images on reading material for female height demographics. Another way of looking at the lamp image distance from the front edge of the desk for the population data in Table 1 is to show it the way a viewer would see it superimposed on the reading material. This is shown in Figs. 2 and 3. The location of the image of the lamp is shown for each percentile. Also added to Figs. 2 and 3 is the location of the 401 viewing angle for each of the female and male 50th percentile of the populations, respectively. A viewing angle of 401 is judged by most researchers as the maximum viewing angle for reading, although reflections above 401 can be uncomfortable, even when located on the desk surface. In all cases the specular glare images of the lamp fall on the reading material within the 401 viewing angle. The equations may also be used to calculate image location for subjects leaning back or forward and for other circumstances. They also work for other configurations, such as forward overhead lighting. From observing reading behavior, it was found that in order to keep their bodies from blocking their view a majority of readers pushed the reading material away from the edge of the desk by about 50 mm (2 in). It is interesting to note in Figs. 2 and 3 that the 401 upper limit for 50th percentile females falls exactly on the length of US letter size paper pushed up 50 mm from the desk front edge (279+50 mm=329 mm), and the same limit for men falls exactly on the length of US legal size paper pushed up 50 mm from 488 D.A. Japuntich / Applied Ergonomics 32 (2001) 485–499 Fig. 4. Reflection components from a dielectric surface for an unpolarized light source. Fig. 3. The location of fluorescent lamp images on reading material for male height demographics. the desk front edge (355.6+50 mm=406 mm). Perhaps to some extent, American paper sizes have evolved according to ergonomic attributes. A4 paper size (297+50 mm=347 mm) falls in between these two distances. 3. Light polarization and reflected images It is important to understand how light reflects and the value of the polarization of that light on reducing the intensity of a specular reflection. As seen in Fig. 4, unpolarized incident light from a lamp interacts with a reflective surface and produces un-refracted, reflected light (reflected specular glare) and refracted scattered light (image). The reflected light as specular glare is the unfocused image of the light source(s). The refracted, emitted light contains the information of color and pattern from the surface and is what we ‘‘read’’. Reflected specular glare maintains the same color as the incident light, carries no information from the surface, and has two transverse components: the vertically vibrating p-waves and the horizontally vibrating s-waves. The linear polarization characteristics of the reflected light from the surface is a function of the degree of polarization of the incident light, the viewing angle from the vertical (i), and the angle of refraction (r), which is a function of the refractive index of the surface (Z) and the viewing angle. An unpolarized ray of light consists of light waves having transverse vibrations of equal magnitude that oscillate about the line representing the direction of the light ray. For simplicity, it is common to resolve the amplitude of the light ray vibrations into components vibrating in two planes at right angles to each other along this line, explained in depth in IES (1993). For this paper with respect to light striking a horizontal surface, the two principal components are the vertically vibrating p-waves and the horizontally vibrating s-waves. When these two components are not equal, the light is partially or totally polarized. Depending on the viewing angle, the veiling glare of the reflection of an unpolarized light source on a horizontal surface in front of a viewer may consist of a majority of s-waves. If the incident light at certain viewing angles is polarized to increase the amplitude of p-waves and decrease the amplitude of swaves, the brightness or luminance of the veiling reflection may be reduced, resulting in a darker reflected light source image. Simply stated, the brightness of the specular, veiling glare is reduced with incident light made of p-waves. The effects on reflected images from light with and without polarization are well described in Shurcliff (1962) and Collett (1993). Polarized sunglasses reduce reflected glare from swaves by adsorbing them. These sunglasses have lenses analogous to vertical microscopic gratings that absorb swaves and let through p-waves. Instead of having a viewer wear sunglasses, a polarizing filter may be placed over a light source to give the same effect on the reflected image. However, although s-waves make up a proportion of reflected glare, p-waves may also be part of the glare. Contrary to common knowledge, at small angles and large angles the specular glare is composed of both s-waves and p-waves. When a vertical polarizing lens is used to eliminate the s-wave glare, the p-wave glare will still be present, superimposed on printed material on a document or on a highway while driving. This is why the glare from the image of the sun superimposed on the road in the early morning with ‘‘large’’ viewing angles approaching 80–901 is not totally eliminated by the use of polarized sunglasses while driving. Viewing angle is D.A. Japuntich / Applied Ergonomics 32 (2001) 485–499 very important, and the viewing angles most encountered in offices range between 201 and 451. 4. Theoretical predictions of glare reduction using a polarized light source In the real world of office workstations, reflective glare appears on messy, non-uniform surfaces like paper, magazines and desktops. In order to determine if polarized light theory may be applied to this problem, one needs to start with a standard surface, which may be mathematically described by the theory and later find out if the theory predicts actual measurements. In this case, that standard surface is glass. Polarized incident light may be used to reduce the glare from the reflected image of a light source on a document, depending on the viewing angle. Linear polarized light is produced by placing a polarization lens in front of a light source. No polarizing lens is perfect, and how well the light is polarized is shown by proportions of the two wave components, as described by the degree of polarization of the incident light, %Pi . The greater the degree of polarization of the incident light, the darker the reflected image of the lamp will appear to the viewer, depending on the viewing angle. When the light is not polarized, the degree of polarization is zero (%Pi ¼ 0). The theoretical intensity (IT )of the lamp image reflection is dictated by percent degree of polarization of the incident light (%Pi ), the intensity of the incident light (I0 ) and the reflection intensities for the s-waves (Is ) and the p-waves (Ip ) as given by the Augustin Fresnel (1821a, b, 1847) equations for the viewing angle (i) and the angle of refraction (r). These may be added to give the intensity of the glare of the reflected image, IT ¼ Is þ Ip at a viewing angle, in a method similar to that used in Clear and Berman (1992), as shown in Eqs. (3) and (4). 2 sin ði@rÞ IT ¼ Is þ Ip ¼ I0 ð0:50@%Pi =200Þ sin2 ði þ rÞ 2 tan ði@rÞ þ I0 ð0:50 þ %Pi =200Þ ; ð3Þ tan2 ði þ rÞ where r ¼ arcsin sin i : Z ð4Þ A percent specular glare reduction, %R, may be calculated using Eqs. (5) and (6) as the percent change of the reflected image intensity from Eq. (3) for incident light with a given percent degree of polarization, ITð%Pi Þ; to the intensity of incident light with no polarization, ITð%Pi ¼0Þ , for a given polarization film transmittance, t. All polarization films have light transmission losses. For instance, the absorbing polarizer medium for polarized 489 sunglasses typically has a transmittance of less than 0.40, while a multiplayer reflective polarizing film may have a transmittance greater than 0.70. The true glare reduction of the polarization medium alone is calculated in Eq. (5) by correcting for the transmittance in Eq. (6), the ratio of the intensities of the light transmitting through the film to the intensity without the film. tðITð%PI ¼0Þ Þ@ðITð%PI Þ Þ ; ð5Þ %R ¼ 100 tðITð%PI ¼0Þ Þ where I0ð%Pi Þ : t¼ I0ð%Pi ¼0Þ ð6Þ In Fig. 5, the theoretical performance for glass, Z ¼ 1:5, using Eq. (5) shows the percent glare reduction from incident light through polarizing lenses of different effectiveness, giving different degrees of polarization at different viewing angles, assuming the same incident light intensity. At low angles, less than 251, eliminating the s-waves and increasing the p-waves to get the same light intensity reduces the glare by less than 20%. In other words, the p-waves are part of the specular glare. As the viewing angle increases, the p-wave absorption and, therefore, the glare reduction increases as the angle increases to Brewster’s angle, where the reflected intensity of the reflected image of the lamp is darkest. Note that at Brewster’s angle of about 571 the percent glare reduction is equal to the percent degree of polarization of the incident light, showing that the condition of the polarization of the incident light is very important. 5. Empirical measurements of specular glare reduction on a standard glass surface Percent specular glare reduction of a lamp’s reflection on a standard glass surface may be calculated from light meter measurements with and without the polarizing filter covering the lamp. If the theory predicts glare reduction from measurements, the theory may be used to predict situations of optimized glare reduction in the workplace. As demonstrated in Fig. 6, empirical measurements of image luminance and lamp illuminance were taken on a desk in a typical cubicle arrangement with an undershelf T-8 fluorescent luminaire. The luminaire had a 901 metallic reflector and no diffuser lens. The polarizing filter was a polycarbonate tube, lined on the inside with an efficient multi-layer reflective polarizer film (Weber et al., 2000), mounted on and completely covering the T-8 fluorescent lamp. Familiar standard polarizing films as used in sunglasses produce polarized light by adsorbing the horizontal component of the incident light. These films 490 D.A. Japuntich / Applied Ergonomics 32 (2001) 485–499 Fig. 5. Theoretical and empirical specular glare reduction for glass at different viewing angles from polarized incident light of different degrees of polarization. have the inherent energy disadvantage of a light transmittance of less than 0.40. Multi-layer polarizing films utilize alternating layers of transparent films, which transform incident light transmitted through them into polarized light by internal reflectivity of the many layers. Past multiplayer polarizers, as in Marks (1959), were shown to be much more light transmission efficient than adsorbing polarizers, giving a light transmittance of up to 0.65. Weber et al. (2000) describes new, unique, thin, multi-layer reflective films, which utilize hundreds of alternating layers of highly birefringent polymers, giving a typical transmittance of 0.94. As shown in Fig. 6 for transmittance measurements, the Minolta Illuminance Meter was positioned at a distance of 600 mm from the lamp, perpendicular to the incident light at an angle from the vertical of 351. The 351 angle is well within the viewing field of a reader sitting at the desk. The transmittance, t, of the polarizing filter (combination of the tube and the multil-layer film) was calculated to be 0.89 from Eq. (6) from the output of the illuminance meter with and without the polarizing filter tube over the lamp. The illuminance meter was checked for polarization sensitivity for the lamp with no polarization (degree of polarization equals zero) by covering the illuminance meter lens with a polarizing lens cap cover and then rotating the lens cap cover full circle, which showed no changes in the illuminance reading. As another check, the transmittance of 0.89 for the polarizing tube was found to be identical to the transmittance of a flat laminate of polycarbonate film (same thickness as the mounting tube) and the multi-layer reflective polarizer film mounted on a flat light-box. Fig. 6 also shows the configuration of how the incident light degree of polarization, %Pi , was calculated from luminance measurements with the reflective desk surface removed so that the luminance meter could be focused directly at the surface of the lamp polarizing filter. A linear polarizing lens cap cover was placed on the luminance meter lens aperture, and the meter was focused on the lamp polarizing filter tube surface at a distance of 600 mm at an incident light angle from the vertical of 351. The meter polarizing lens cap cover was rotated to give the s-wave minimum luminance (I0s , lens cap polarization horizontal) and p-wave maximum luminance (I0p , lens cap polarization vertical). The degree of polarization was then calculated using Eq. (7). Since the polarized incident light should be Fig. 6. Experimental configuration for specular glare reduction measurements. 491 D.A. Japuntich / Applied Ergonomics 32 (2001) 485–499 dominated by the vertical p-waves, the incident light percent degree of polarization is positive according to the convention in the IES (1993) Lighting Handbook. The measured percent degree of polarization of this system was 87.5%. %Pi ¼ 100ðI0p @I0s Þ=ðI0p þ I0s Þ: ð7Þ The reflective luminance measurement methodology for glare reduction calculated by Eq. (5) follows the ASTM (1994) Standard Test Method for Specular Gloss: D52389, modified to measure specular reflection (gloss) at different angles with and without the polarization medium in place on the lamp. This method is also similar to the Blackwell (1963) methodology for reflected image luminance reduction. To measure IT as shown in Fig. 6, the Minolta LS-100 Luminance Meter was mounted on a tripod in place of the ‘‘viewer’’ and was focused on the reflected image of the lamp on the surface. The luminance meter was checked and found not to have polarization sensitivity in any orientation. Smooth glass was chosen not only for a standard, traceable reflectivity, but also because its performance may be theoretically calculated. The standard glass reflective surface (Z ¼ 1:5), traceable to ASTM D523-89, was Hunterlab D48 Glossmeter Standard: 98.8 gloss value at 751-TAPPI, manufactured by Hunter Associates Laboratory, Reston, VA. Other non-standard surfaces may give different results due to surface texture, index of refraction, light absorbency and multi-surface reflectivity. In the configuration of Fig. 6, the corrected percent glare reduction of the reflected lamp image, %R, was calculated by Eq. (5) from the luminance meter measurements (cd/m2) of the reflected lamp image at a chosen viewing angle with the polarization filter in place, giving ITð%Pi Þ , and with no polarization filter, giving ITð%Pi ¼0Þ . In Table 2 the theoretical percent glare reduction values were calculated at the measured incident light degree of polarization of 87.5%, using Eqs. (3) and (4) for the polarizing filter percent degree of polarization and transmittance, and were compared to the empirical corrected percent glare reduction. Data were gathered at two lamp heights with different viewer heights and viewer distances from the lamp. Viewer distance and height were also varied by changing the tilt angle from Table 2 Comparisons of theoretical and empirical percent specular glare reduction values at many different lamp and viewer configurations for incident light with an 87.5% degree of polarization Gender/ percentile Chair tilt (Deg) Lamp height (y) (mm) Viewer height (h) (mm) Viewer distance (L) (mm) Incident angle (Deg) Theoretical % glare reduction Measured % glare reduction Female 5th Female 50th Female 95th Male 5th Male 50th Male 95th Female 5th Female 50th Female 95th Male 5th Male 50th Male 95th Female 5th Female 50th Female 95th Male 5th Male 50th Male 95th Female 5th Female 50th Female 95th Male 5th Male 50th Male 95th Female 5th Female 50th Female 95th Male 5th Male 50th Male 95th 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 15 15 15 15 15 15 20 20 20 20 20 20 410 410 410 410 410 410 381 381 381 381 381 381 381 381 381 381 381 381 381 381 381 381 381 381 381 381 381 381 381 381 315 391 471 409 490 569 315 391 471 409 490 569 315 391 471 409 490 569 304 378 455 395 473 550 296 368 443 384 461 535 533 533 533 533 533 533 533 533 533 533 533 533 655 655 655 655 655 655 615 634 655 639 660 680 641 667 694 673 701 728 36 34 31 33 31 29 37 35 32 34 31 29 43 40 38 40 37 35 42 40 38 39 38 36 43 42 40 41 40 38 50 43 37 42 36 31 53 46 39 44 38 33 69 61 54 59 52 46 65 60 55 59 54 50 69 65 60 64 60 56 50 41 35 39 34 29 53 45 38 44 36 29 68 60 53 59 51 48 64 57 53 57 52 48 67 63 56 62 58 54 492 D.A. Japuntich / Applied Ergonomics 32 (2001) 485–499 the desk edge, as if the viewer were leaning back in a chair. The viewer’s eyes were directly over the table edge at 01 tilt angle. As seen in Fig. 5, when plotting the Table 2 measured percent specular glare reduction data points for a filter with 87.5% degree of polarization, the theory agreed very well with this data, predicting performance over a wide range of angles. Also, because of smaller height, females had larger incident angles and larger resultant glare reduction at the same configurations. 6. Optimization of glare reduction and the location of the light source image on reading material Because the theory successfully predicts the actual glare reduction, it may be used to predict optimal lamp conditions for the best glare reduction for the viewer. At the same time, the position of the reflected image may be optimized to move it further up and perhaps off the top of the page of the reading material as in Siminovitch (1993). Viewing angle is difficult to measure directly, so the easily measured and understood lamp height and lamp to desk edge distance were used in this optimization, just as they would be by an ergonomist. As people of smaller height are most affected, the height of the female 50th percentile was used. A dimensional analysis of the equations for the lamp image location on reading material and the percent glare reduction showed that the lamp height and the distance of the lamp to the desk edge may be simultaneously varied to give improved results. For the female 50th percentile of 391 mm, a grid was created within the bounds of a standard American office cubicle by varying the lamp height from 350 to 500 mm and the lamp to table edge distance from 400 to 850 mm. Location of the reflection, D, was calculated for W by Eq. (1) for X, and the percent glare reduction was calculated by Eq. (5). It was assumed that the viewer’s eyes were directly above the desk edge, and the predictive results were sorted to give output ranges as shown in Figs. 7 and 8, for the reflected image location and the percent glare reduction, respectively. Figs. 7 and 8 show that simultaneously decreasing the lamp height and increasing the lamp to table edge distance moves the image up the page and increases percent glare reduction. In some cases the image may be moved off the page completely, where glare reduction of the nuisance reflected image on the desk may be achieved. However, in actual use for a certain luminaire design, there are limits that may need to be placed on the changes to the luminaire position and installation, based upon the illumination field: the intensity of the illumination and its uniformity in the reading area (see Siminovitch, 1993, p. 239). Fig. 7. Predictive grid of fluorescent lamp specular image location optimization at different lamp heights and lamp to viewer distances for 50th percentile female height. D.A. Japuntich / Applied Ergonomics 32 (2001) 485–499 493 Fig. 8. Predictive grid of fluorescent lamp specular image glare reduction optimization at different lamp heights and lamp to viewer distances for 50th percentile female height. 7. Glare reduction testing on humans and correlation to empirical measurements This section deals with glare reduction ratings on printed-paper by a panel of people and correlation of these ratings to light meter measurements of glare reduction and, as a reference, the theoretical percent glare reduction on glass. The easily measured variables of fluorescent lamp height and lamp to desk edge distance were used to vary the viewing angle and, subsequently, the percent glare reduction. If the results for glare reduction ratings from human testing give the same optimizational trend as in Fig. 8, the use of the theory for workstation lighting design is justified. Rather than a glass surface, this study includes testing on two different papers with and without bright, diffuse ambient lighting illuminance superimposed on the illuminance of the fluorescent task light. The selection of 30 experimental subjects (17 females and 13 males) provided for near equal weighting of age and gender within the four decades of age groups (20, 30, 40, 50 years) normally encountered in the workplace. Test subjects with eye correction totaled 17 with eyeglasses and 7 with contact lenses. The Fig. 8 plot of theoretical percent glare reduction optimization for different lamp heights and distances may be regarded as a contour plot (response surface) of the change of the independent variable (% glare reduction) to the interaction of two independent variables (lamp height and lamp distance). This is a typical example of the output of an analysis of variance for a two-level factorial designed experiment. Consequently, in order to give a clear picture or contour plot of how to optimize human glare reduction response for the two paper sample types, a two-level factorial design format was chosen for the analysis of the glare reduction rating response of the test subjects to different lamp positions. For these two-level factorial designs (as shown in Table 4) there were six experimental conditions. The lamp height and the lamp distance from the viewer were varied high and low, and for the statistical analysis two center points were added at half the differences between each of these two levels. An analysis of variance or ANOVA for each response was performed to create a statistical model and equation describing the system mathematically. The experimental set up was similar to that shown in Fig. 1 with each person positioned at a 50th percentile female eye height of 391 mm (15.3 in) with their eyes directly above the desk edge. The lamp could be easily adjusted horizontally and vertically in the same space as a typical under-shelf lighting fixture. The experimental set up of the fluorescent lamp fixture was in a typical open-plan office cubicle workstation. The test lamp was a 1.2 m (4 feet) long T-12, 40 W fluorescent tube. The front of the luminaire facing 494 D.A. Japuntich / Applied Ergonomics 32 (2001) 485–499 the test subject was shaded to eliminate any direct glare from the lamp, and a 901 metallic reflector was used in the luminaire behind the fluorescent lamp. The lamp was covered except for a 0.38 m length open area facing downward in the center of the lamp, positioned directly in front of the test subject. A polycarbonate tube of 0.8 m length fitted over the lamp was shuttled back and forth across this open area. One-half of the interior of the polycarbonate tube length was lined with a multilayer reflective polarization film with an 87.5% degree of polarization (Weber et al., 2000) and the other half was lined with a polyester film of the same transmittance. In this way, the light from the lamp could be easily changed from a polarized to an unpolarized light source of the same illuminance by shuttling the tube left or right. With the polarization, Tables 3 and 4 shows the distribution of luminance parallel to the lamp at a 200 mm distance from the desk edge, measured using the luminance meter at a lamp height of 410 mm, lamp distance of 505 mm and a viewing height of 391 mm, giving a viewing angle of 271. Table 3 shows that the luminance varies less than 6% across the 20 cm width of the reading material target and varies less than 25% across the entire exposed width of the lamp from end to end. The overhead, ambient lighting was provided by two luminaires perpendicular to the test lamp, each containing two 1.2 m (4 feet) T-8 fluorescent lamps, covered with plastic diffusers. They were positioned to provide diffuse, ambient non-polarized lighting 2 m above the reading surface, each 1 m on either side of the reading surface. The illuminance provided by this lighting at the center of the reading surface, 200 mm toward the lamp from the test subject was 480 lux. This illuminance can be considered the high end of the recommendations given in IES Lighting Handbook (1981), which recommends 200–500 lux for reading tasks at office desks. The two paper samples represented typical A4 size (295 mm long) reading materials of semi-gloss and matte finish papers, the lower edge positioned 50 mm toward the lamp from the desk edge. For the convenience of the readers of this journal, the semi-gloss paper was chosen as the inside of the back cover page of the June 1998, vol. 29(3) issue of Applied Ergonomics (Notes for Authors), and the matte finish paper sample was chosen from P. 163 of the interior of the same journal volume. Both samples were two columns of text on white paper. Gloss measurements using the ANSI/ASTM (1994) Standard Test Method for Specular Gloss: D523-89 were 45.6 at 601/86.3 at 851 for an un-inked area on the semi-gloss paper, 4.5 at 601/17.2 at 851 for an un-inked area on the matte finish paper, 59.8 at 601/79.5 at 851 for an inked area on the semi-gloss paper and 10.8 at 601 /25.2 at 851 for an inked area on the matte finish paper. Papers and surfaces with this degree of specular gloss are common in any office workplace. The two-level factorial design is shown in Table 4. The experiment run order selection was random. The lamp height and lamp distance from the viewer levels are typical ranges that could be achieved in under-shelf task lights for open-plan office cubicles. The illuminance of the lamp output changed with lamp design position (see Table 4), and the illuminance with the polarizing film was the same as for the polyester film. The glare reduction rating difference at each design position was judged as the change in glare with and without polarization. In Table 4, the viewing angle of the main reflection of the lamp on the page was calculated using Eq. (2). In all positions, the glare from the lamp could be seen as a reflection on the papers. The change in glare on the paper was judged by each of the test subjects as a glare reduction rating from 0 to 10, with 10 the highest change or difference in glare. A rating of 0 was judged as no change or difference in glare at all. No explanation or definition of glare was given to the test subjects, leaving the assessment of what they were judging up to him or her. The experiments and their average (n ¼ 30) results are shown in Table 4. Design points 5 and 6 are replicates and center points, and show good reproducibility of the average glare reduction ratings from each experiment. The size of the standard deviations may be the result of the differences in sensitivity of each test subject and their picking of a different starting level for glare reduction ratings. For instance, one may start with a glare reduction of 5 and judge around that level while another started at 7. However, it is the differences between the glare reduction averages in the design points that are most important in the two-level factorial analysis. Table 5 shows the results of the analysis of variance (ANOVA) for each of the experiments as analyzed using the Design-Expert t software made by Stat-Ease t, Inc. of Minneapolis, MN, USA. Each analysis gives a predictive equation for each model. The excellent degree of linear model fit is shown by the high R2 values (R2 in Table 3 The luminance distribution with the polarizing tube from the lamp center: bond paper at a 200 mm distance from the desk edge at a lamp height of 410 mm, lamp distance of 505 mm, viewing height of 391 mm and a viewing angle of 271 Lamp edge Position (cm) in front of lamp Luminance (lux) at 200 mm from desk edge @30 49.9 Lamp center @20 67.9 @10 84.8 0 90.3 Lamp edge 10 84.3 20 69.7 30 50.9 495 D.A. Japuntich / Applied Ergonomics 32 (2001) 485–499 Table 4 Two-level factorial design of glare reduction rating for differences in lamp height and lamp to viewer distance: experimental test conditions and average results on 30 human subjects Design Run Lamp Lamp Lamp Viewing Semi-gloss point order height distance illuminance angle (Deg) paper/lights off (mm) (mm) (lux) Glare reduction rating 1 2 3 4 5 6 4 2 3 5 6 1 330 490 330 490 410 410 410 410 600 600 505 505 582 398 206 239 351 351 29.6 25.0 39.8 34.3 32.2 32.2 Matte paper/ lights off Semi-gloss paper/ lights on Matte paper/ lights on Glare reduction rating Glare reduction rating Glare reduction rating Ave n ¼ 30 SD Ave n ¼ 30 SD Ave n ¼ 30 SD Ave n ¼ 30 SD 6.3 3.8 7.9 5.6 6.5 6.4 1.8 2.0 1.6 2.0 1.7 1.3 2.8 2.0 5.2 3.5 2.9 3.3 2.5 1.6 2.3 1.9 2.1 2.4 4.4 3.1 6.3 4.9 4.9 4.4 1.8 1.5 1.7 1.6 1.8 1.7 1.5 0.9 2.4 1.0 1.2 1.5 1.1 1.1 1.7 1.1 1.3 0.9 Table 5 Two-level factorial design ANOVA models of glare reduction rating for two different papers with and without bright, non-polarized ambient lighting Test condition Model fit R2 Intercept Lamp height coefficient Lamp distance coefficient 0.998 0.976 0.951 0.859 7.532 3.217 1.502 2.684 @0.01500 @0.00844 @0.00797 @0.00625 0.00895 0.00974 0.01022 0.00263 a Predictive factorial model Semi-gloss paper/lights off Semi-gloss paper/lights on Matte paper/lights off Matte paper/lights on a Rating=intercept+Coeff (Lamp height)+Coeff (Lamp distance). this case being a measure of the amount of variation around each mean as explained by the model). These models for each of the two papers may be presented as contour plots as shown in Fig. 9 for no ambient lighting and Fig. 10 for bright ambient lighting. Figs. 9 and 10 contour plots of the results in Table 5 show the same optimization trends as the theoretical results in Fig. 8. Even though the levels of the responses are different, it is possible that both the theoretical percent glare reduction and the test subject glare reduction rating may be used to optimize an office workstation for lighting to give minimal glare. The glare reduction rating changes with the glossiness of the paper are seen by comparing the two plots in Fig. 9 for no ambient lighting. If an under-shelf task light is the sole source of lighting in front of a worker, polarized lighting is a good tool for glare reduction for both semi-gloss finish and matte finish papers. The glare reduction rating results with the addition of strong, diffuse ambient non-polarized lighting may be seen in Fig. 10. In most cases the ambient lighting was a greater source of illuminance than that of the undershelf task light. In strong ambient lighting, polarized task lighting still successfully reduces specular glare on semi-gloss finish, but to a lesser extent than without ambient lighting. However, for the matte finish paper with bright ambient lights, so little reflection from the test lamp may be perceived that glare reduction afforded by polarized light is minimal. Fig. 11 shows all of the design points of the four experiments, plotting glare reduction rating against the calculated reference of theoretical percent glare reduction on glass (task lighting degree of polarization of 87.5%) from Table 6. The data of glare reduction rating for the two papers at the two test conditions follow the same general patterns. This strongly suggests simple linear correlations for this range of viewing angles. For the two common papers studied, the glare reduction rating for most actual work conditions will fall between the plots of ambient lights off and bright, ambient lights on, with the amplitude of the rating dependent on lamp position (viewing angle). Simple linear regressions were run for each of these sets of data in Fig. 11 and are shown in Table 7. The degree of fit for each of paper and lighting condition shows that for these materials the specular glare on a chosen surface as sensed by a human may be directly related to theoretical glare calculation on a reference surface. The data in Table 7 could also be used to generate the same type of contour plots as in Figs. 9 and 10. The empirical percent specular glare reduction was measured using the luminance meter at all the same Table 4 design points that had been used in the human testing for each of the two papers at the two lighting 496 D.A. Japuntich / Applied Ergonomics 32 (2001) 485–499 Fig. 9. No ambient lighting: ANOVA model contour plots of human glare reduction rating at different lamp heights and lamp to viewer distances for 50th percentile female height for semi-gloss finish and matte finish papers. Fig. 10. Bright ambient lighting: ANOVA model contour plots of human glare reduction rating at different lamp heights and lamp to viewer distances for 50th percentile female height for semi-gloss finish and matte finish papers. conditions. When this data was compared to the human testing glare reduction ratings, Fig. 12 resulted, showing a reasonably good correlation (R2=0.91) for these very different papers and conditions. It may be concluded that the luminance (brightness) meter and the eyes of the human subjects respond in the same manner to changes in glare reduction from the use of polarized lighting. Glare reduction ratings of over 5 on papers may be achieved at measured specular glare reductions of 20% or greater. In this case, the correlation is non-specific to materials over a wide range of conditions. So, in place of using a luminance meter, one simple, inexpensive method to see if linear polarized light is indicated to reduce specular glare is to sit at a workstation viewing 497 D.A. Japuntich / Applied Ergonomics 32 (2001) 485–499 Fig. 11. Human glare reduction rating versus theoretical percent glare reduction on glass for semi-gloss and matte finish papers. Table 6 Two-level factorial design for differences in lamp height and lamp to viewer distance: theoretical percent glare reduction on glass, glare reduction rating averages and measured percent glare reduction for each design point Design point 1 2 3 4 5 6 Viewing angle (Deg) 29.6 25.0 39.8 34.3 32.2 32.2 Theoretical % glare reduction on glass 33.4 23.4 59.5 44.8 39.7 39.7 Semi-gloss paper/ lights off Matte paper/ lights off Glare reduction rating Measured % glare reduction Glare reduction rating 6.3 3.8 7.9 5.6 6.5 6.4 20.7 14.1 35.8 26.7 23.2 26.6 2.8 2.0 5.2 3.5 2.9 3.3 Table 7 Linear regressions of glare reduction rating versus theoretical percent glare reduction on glass for semi-gloss and matte finish papers at two lighting conditions Semi-gloss paper/lights off Semi-gloss paper/lights on Matte paper/lights off Matte paper/lights on Intercept Coeff R2 2.273 1.276 @0.204 0.079 0.0945 0.0844 0.0875 0.0333 0.733 0.957 0.956 0.588 the reflective glare while rotating a lens from polarized sunglasses over a 901 arc to see the magnitude of luminance difference. Semi-gloss paper/ lights on Matte paper/ lights on Measured % glare reduction Glare reduction rating Measured % glare reduction Glare reduction rating Measured % glare reduction 10.1 6.2 14.4 10.9 9.3 10.3 4.4 3.1 6.3 4.9 4.9 4.4 11.6 6.5 22.0 14.8 12.1 15.4 1.5 0.9 2.4 1.0 1.2 1.5 2.3 1.3 4.8 2.2 2.6 2.4 8. Conclusions This is a study of the specular reflection on a task placed between a viewer and an under-shelf light source in a constrained open-plan office cubicle workstation environment. Specular glare reduction and the optimization of the location of the reflected image of a lamp may be measured and accurately predicted by the methods presented in this paper. These may be achieved by proper and optimal installation of light sources producing light of a high degree of polarization in openplan office cubicles according to the height of the viewer to help in minimizing both direct glare and reflected specular glare. A diverse panel of human subjects using 498 D.A. Japuntich / Applied Ergonomics 32 (2001) 485–499 Fig. 12. Correlation of measured percent glare reduction and human glare reduction rating. a glare reduction rating at the 50th percentile female eye height has confirmed these methods. It is surmised that human subjects of progressively higher eye heights will see similar changes, but less specular glare reduction because of smaller viewing angles. This study successfully uses simple linear polarization optics for lighting optimization. However, luminaires have a definite width and length, may be covered by different light diffuser lenses and also may have different reflector designs. The effects of these dimensions on glare reduction for different viewing positions needs to be considered in future studies and in specific applications. Other lighting configurations such as side-lit lamps, different reflective systems and lenses will not follow the same optimization procedures as shown in this study. From human testing of the glare reduction rating, this study clearly shows that polarized lighting affords predictable specular glare reduction if the task light is the sole source of illuminance, even for matte finish paper. If the ambient, diffuse lighting is increased, then the benefits of the use of polarized task lighting are less, but still significant, depending on the gloss value of the viewed material and the ratio of the illuminance of ambient light to the task light. Human glare reduction ratings correlate well with empirical specular glare reduction as calculated from luminance meter measurements. For a specific surface and lighting condition, the glare reduction ratings may also be simply correlated to the theoretical percent glare reduction on a reference surface, such as glass. The ergonomics of lighting clearly needs to take worker stature into consideration for worker comfort and productivity. Because of their heights, 95% of the males are not looking directly into an undershelf cubicle light source of the common height of 410 mm, while more than 50% of the females are exposed to direct glare from the light source. Female viewers who are of smaller heights with larger viewing angles will see an increase of 6–10% reflective, specular glare over taller male viewers when using nonpolarized light, as seen in Table 2. This may help to explain the results of the research of North (1991) that shows a 50% increase of glare discomfort in women over men. These methods may be applied to other areas of study, such as in the field of education, where children are even shorter than adult females, and their heights should also be taken into consideration for proper reading and work illumination. The analysis of the geometry of lamp placement versus the viewer may be used for other illuminating studies, such as for forward over-head lighting. References Allphin, W., 1963. Sight lines to desk tasks in schools and offices. Illum. Eng. 58 (4), 244–249. ASTM Standard Test Method for Specular Gloss: D523-89, 1994. American Society for Testing and Materials, West Conshohocken, PA. Bernecker, C.A., Davis, R.G., Webster, M.P., Webster, J.P., 1993a. Task lighting in the open office: a visual comfort perspective. J. Illum. Eng. Soc. 22 (1), 18–25. D.A. Japuntich / Applied Ergonomics 32 (2001) 485–499 Bernecker, C.A., Beebe, L.A., Shi, J., 1993b. Blackwell: a look back. LD+A. 23 (8), 62–68. Blackwell, H.R., 1963. A general quantitative method for evaluating the visual significance of reflected glare, utilizing visual performance data. Illum. Eng. 58 (4), 161–243. Blackwell, H.R., 1984. Studies of visual effectiveness of polarized light, Report to the Ohio State University Institute in Vision to the Lighting Engineering Research Institute, November 30. Blackwell, H.R., DiLauria, D.L., 1973. Application procedures for evaluation of veiling reflections in terms of ESI. J. Illum. Eng. Soc. 2, 254–283. Boyce, P.R., Rea, M.S., 1994. A field evaluation of full-spectrum, polarized lighting. J. Illum. Eng. Soc. 23 (2), 85–107. Clear, R., Berman, S., 1992. Polarization and glare on VDU’s and other tilted specular surfaces. J. Illum. Eng. Soc. 21 (1), 29–40. Clear, R., Mistrick, R.G., 1996. Multilayer polarizers: a review of the claims. J. Illum. Eng. Soc. 25, 70–88. Collett, E., 1993. Mueller Matrices for Reflection and Transmission, Polarized Light: Fundamentals and Application. Marcel Dekker Inc., New York, NY (Chapter 8). Crouch, C.L., Kaufman, J.E., 1963. Practical application of polarization and light control for reduction for reduction of reflected glare. Illum. Eng. 58 (4), 277–291. Fresnel, A.J., 1821a. M!emoire sur la double r!efraction. M!em. de l’Acad. des sc. vii, 45. Fresnel, A.J., 1821b. Ann. de Chim.(2) xxviii, 263. Fresnel, A.J., 1847. Pogg. Ann. xxiii, 372, 424. Gordon, C.C., et al., 1988. Anthropometric survey of US army personnel: methods and summary statistics. Technical Report Natick/TR 89/044. Grabau, M., 1940.The Major Applications of Polarized Light. Introduction to Polarized Light. Polaroid Corporation, Cambridge, MA, 29–30 (Chapter V). 499 IES, 1993. Light and optics. In: Rea, M.S. (Ed.), Lighting Handbook, 8th Edition, Illuminating Society of North America (IESNA), New York, NY, p. 22 (Chapter 1). Jones, B.F., 1992. Polarized light: effects on vision, energy use and cost reduction. LD+A. 22, 34–36. Karpen, D., 1995. Progress in full-spectrum polarized lighting systems. Energy Eng. 92 (6), 18–48. Lumsden, W.K., 1976. Effective task lighting using polarized light. Light Light. Environ. Des. 69 (4), 160–162. Marks, A.M., 1959. Multi-layer polarizers and their application to general lighting. Illum. Eng. 54 (2), 123–135. North, V.A., 1991. Factors influencing worker’s impressions of glare in open offices. Ph.D. Thesis, University of Michigan, Ann Arbor, MI. Rea, M.S., 1981. Visual performance with realistic methods of changing contrast. J. Illum. Eng. Soc. 19 (3), 164–177. Rea, M.S., Ouellette, M.J., Kennedy, EM.E., 1985. Lighting and task parameters affecting posture, performance and subjective ratings. J. Illum. Eng. Soc. 15, 231–238. Rea, M.S., Ouellette, M.J., Tiller, D.K., 1990. The effects of luminous surroundings on visual performance, pupil size, and human preference. J. Illum. Eng. Soc. 19 (2), 45–58. RP-24. 1989. IES Recommended Practice for Lighting Offices Containing Computer Visual Display Terminals. Illuminating Engineering Society of North America, New York, NY. Shurcliff, W.A., 1962. Reflection polarizers. In: Polarized Light: Production and Use. Harvard University Press, Cambridge, MA (Chapter 6). Siminovitch, M.J., 1993. Experimental development of efficacious task source relationships for interior lighting applications. Ph.D. Thesis, University of Michigan, Ann Arbor, MI. Weber, M.F., Stover, C.A., Gilbert, L.R., Nevitt, T.J., Ouderkirk, A.J., 2000. Giant birefringent optics in multilayer polymer mirrors. Science 287, 2451–2456.